SpectraMax 190 Microplate Reader

A high-performance spectrophotometer

Introduction Accurate and precise


The SpectraMax® 190 Microplate Reader absorbance readings
from Molecular Devices is ideal for UV-Vis The unique quartz fiber optic system
life science applications, especially DNA
BENEFITS analysis. The multi-channel design truly
minimizes stray light for precise readings
across the microplate. Each channel has
• T
 une in your performance for mimics a dual-beam spectrophotometer. its own reference detector and the small
assays in UV/Vis spectrum Each sample has a discrete sample beam beam size maintains high performance
• E
 liminate the need for and reference beam so that each well is even with small sample volumes read
concentration curves with measured directly, eliminating error due to in half-area wells. The 2 nm bandwidth
PathCheck Sensor variations in light output between the optic provides the needed spectral resolution
fibers. The system consists of eight sample to ensure accuracy of DNA absorbance
• Interface with the latest state-of- beams and detectors and eight reference measurements. Up to 6 different
the-art SoftMax Pro Software beams and detectors to deliver both wavelengths can be measured during
• E
 nsure performance with superior precision and speed-of-reading one read.
validation tools across the microplate through 4.0 OD.
Optional tools for verifying the SpectraMax
190 reader performance include the PathCheck sensor eases
SpectraTest ABS1 Absorbance Plate and
the workload
IQ/OQ/PQ manual. Software tools for
For aqueous solutions, the patented*
regulatory compliance, such as FDA 21
PathCheck® Sensor senses the depth
CFR Part 11, are available through SoftMax®
of the liquid in each microplate well and
Pro GxP Software.
normalizes the absorbance value to a 1
cm pathlength. The corrected absorbance
is within 5% of the value obtained in a
conventional spectrophotometer.
The PathCheck sensor makes it is easy to:
•U
 se the extinction coefficient of a sample to calculate
concentration directly
• Correct for volume differences in the wells
• Expand the assay’s dynamic range
•C
 heck performance of pipetting devices

Spectral scanning for easy assay development


The complete spectrum of samples can be scanned to select the
optimum wavelength for reading. Spectrum scanning also allows:
•M
 onitoring of sample or reagent stability
•D
 etermination of interfering compounds or assay conditions
Figure 1. Flexible template assignment. Standards for multiple
calibration curves and unknowns can be run on separate plates.
Performs virtually any UV/VIS assay
• Bacterial identification • Entomology
• Biochemistry • Environmental testing
• Cell biology • Enzyme kinetics
• Colorimetric • Food/agricultural
• DNA quantitation • Molecular biology
• ELISAs • Pharmacology

Accurate measurements
The SpectraMax 190 reader is the tool for DNA analysis. The
ability to detect 16 ng/well of DNA and to quantitate 50 ng of
DNA provides the same sensitivity as a spectrophotometer.
The PathCheck sensor can extend the instrument linearity by a
factor of three or more. For very concentrated samples, use a
minimum volume in the microplate well (100 µL has a pathlength
of about 0.30 cm), then use the PathCheck sensor to calculate an
Figure 2. Multiple calibration curves. Multiple calibration curves can be equivalent 1-cm pathlength absorbance value.
plotted on one graph.

Ensure GxP compliance


The SpectraTest validation package provides a NIST-traceable
solution for validating the performance of the SpectraMax 190
reader automatically.

Plate stacker and robot integration


The SpectraMax 190 reader can be integrated with Molecular
Devices’ StakMax Microplate Stacker in a matter of minutes and
begin reading microplates with seven mouse clicks. For a higher
degree of automation, the Automation Vendor Partners Program
has streamlined the integration of our microplate reader systems
with all leading partner robots. The “out-of-the-box” automation
solution saves up-front integration time and resources.

Figure 3. Combined data graphing. Results from unknowns run on


different plates and different calibration curves can be plotted on
one graph.
Technical specifications

Photometric performance specifications

Wavelength range 190–850 nm


Monochromator, tunable in 1 nm
Wavelength selection
increments
Wavelength bandwidth 2 nm

Wavelength accuracy < ±2.0 nm

Wavelength repeatability ± 0.2 nm

Photometric range 0–4.0 OD


Photometric resolution 0.001 OD
Photometric accuracy < ±0.006 OD ±1.0%, 0–2.0 OD

Photometric precision < ±0.003 OD ±1.0%, 0–2.0 OD


PathCheck® sensor
< 5% compared to cuvette reading
measurement error
Stray light ≤ 0.05% @ 230 nm

Light source Xenon flash lamp


Microplate read times
Endpoint 12 seconds
Kinetics 9 second min. interval
Temperature specifications

Temperature range Ambient +4– 45°C


Temperature uniformity
±0.5°C at 37°C, well-to-well
(microplate)
30 minutes maximum upon
Temperature stabilization time
initiation
Isothermal when temperature
Reading chambers
regulation not enabled, < 1°C
General specifications

Dimensions (in.) 8.6 (H) x 22.8 (W) x 15 (D)

Dimensions (cm) 22 (H) x 58 (W) x 38 (D)

Weight 30 lbs. (13.6 kg)

Power requirements 90–250 Vac 50/60 Hz

Power consumption < 250 W


Operating temperature 10–40°C
Storage temperature 5–40°C

The PathCheck sensor is covered under U.S. Patents 5,959,738 and 6,188,476.
Ordering information VWR cat. no.
SpectraMax® 190 microplate reader, multi-mode,
MLDV190
with current version of SoftMax® Pro

Go to www.vwr.com for the latest news, special offers and details of your local VWR organisation.

Austria Belgium Denmark France Germany


VWR International GmbH VWR International bvba VWR International A/S VWR International S.A.S. VWR International GmbH
Graumanngasse 7 Researchpark Haasrode 2020 Tobaksvejen 21 Le Périgares – Bâtiment B Hilpertstraße 20a
1150 Vienna Geldenaaksebaan 464 2860 Søborg 201, rue Carnot D - 64295 Darmstadt
Tel.: +43 1 97 002 0 3001 Leuven Tel.: +45 43 86 87 88 94126 Fontenay-sous-Bois cedex Tel.: 0800 702 00 07* (national)
Email: info.at@vwr.com Tel.: +32 (0) 16 385 011 Email: info.dk@vwr.com Tel.: 0 825 02 30 30* (national) Tel.: +49 (0) 6151 3972 0 (international)
Email: vwr.be@vwr.com Tel.: +33 (0) 1 45 14 85 00 (international) Email: info.de@vwr.com
Email: info.fr@vwr.com *Freecall
* 0,18 ¤ TTC/min

Ireland / Northern Ireland The Netherlands Sweden Switzerland UK


VWR International Ltd / VWR International B.V. VWR International AB VWR International GmbH VWR International Ltd
VWR International (Northern Ireland) Ltd Postbus 8198 Fagerstagatan 18a Lerzenstrasse 16/18 Customer Service Centre
Orion Business Campus 1005 AD Amsterdam 163 94 Stockholm 8953 Dietikon Hunter Boulevard - Magna Park
Northwest Business Park Tel.: +31 (0) 20 4808 400 Tel.: +46 (0) 8 621 34 00 Tel.: +41 (0) 44 745 13 13 Lutterworth
Ballycoolin Email: info.nl@vwr.com Email: kundservice.se@vwr.com Email: info.ch@vwr.com Leicestershire
Dublin 15 LE17 4XN
Tel.: +353 (0) 1 88 22 222 Tel.: +44 (0) 800 22 33 44
Email: sales.ie@vwr.com Email: uksales@vwr.com

©2017 Molecular Devices, LLC. The trademarks used herein are the property of Molecular Devices,
LLC or their respective owners. Specifications subject to change without notice.
Patents: www.moleculardevices.com/productpatents
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
12/17 0120-1409E-EU/VWR