Case number: CAS-0017766721 Order Number: WO-13812075 | Ser. Report Created Date: 21/06/2023
: Contact Phone:
Purpose of Visit
Waste separator issue
28071 - cobas b 221 6 Roche OMNI S6 system
Performed Activities
Checked and found that waste docking mechanism was little bit broken and hence fixed waste docking mechanism because waste
system was leaking. Checked and found that vaccum release valve v13 in vaccum pump was not working and hence repaired valve
working. After Doing maintenance run and aspirated all fluidic routines and system calibration done. Run samples and observed
machine. Found k
Service enginer remarks - please operate machine properly so that consumables donot get broken