Cobas b221 Kgmu 28151 Sample Distributor
Cobas b221 Kgmu 28151 Sample Distributor
Case number: CAS-0017534005 Order Number: WO-11325966 | Ser. Report Created Date: 16/05/2023
: Contact Phone:
Purpose of Visit
hardware test error and heating issue
28151 - cobas b 221 6 Roche OMNI S6 system
Performed Activities
Checked and found that sd unit has problem hence ordered new sd unit , but optical sensors are also dislocated from their slots.
Hence recorrected them, there are multiple problems in the machine and now it has started working. We had ordered bg measuring
chamber and it is unused now we are returning same. Except mss parameters , machine got calibrated and as there is no s3 pack
solution , those parameters were left not calibrated
Time Report
Travel
Value Hours
Travel 5.5
Roche Diagnostics India Pvt. Ltd.
Field Service Report
Case number: CAS-0017534005 Order Number: WO-11325966 | Ser. Report Created Date: 16/05/2023
Spare Parts
Case number: CAS-0017534005 Order Number: WO-11325966 | Ser. Report Created Date: 16/05/2023
Signature