Case number: CAS-0017022553 Order Number: WO-03691224 | Ser. Report Created Date: 15/02/2023
Purpose of Visit
gripper issue
Performed Activities
Checked and found that some droplets such as paper deposits are accumulated in the incubator holes and hence took out incubator
and clean incubator holes completely. Clean gripper finger with the alcohol swab and performed gripper adjustments manually and
automatic at all positions and checked gripper repetition check and observed some samples running and found ok
Service enginer remarks : the environment conditions are not ok and customer is required to maintain it on daily basis and required to
follow operator maintaining related to machine working so as reduce frequency of break down.
Time Report
Travel
Value Hours
Travel 0.5
Roche Diagnostics India Pvt. Ltd.
Field Service Report
Case number: CAS-0017022553 Order Number: WO-03691224 | Ser. Report Created Date: 15/02/2023
Signature