Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

British Journal of Anaesthesia 1993; 7 1 : 453-457

APPARATUS

ACCURACY OF 94 ANAESTHETIC AGENT VAPORIZERS IN


CLINICAL USEf

J. NIELSEN, F. M. PEDERSEN, F. KNUDSEN, M. B. JENSEN AND M. IBSEN

circumstances resembling clinical use. In addition,


SUMMARY using a questionnaire, we sought the views of
Using the Bruel & Kjaer Anaesthetic Gas Monitor anaesthetists on vaporizer performance.
type 1304, we have monitored the output of 94
anaesthetic agent vaporizers (Fluotec 3: 58, En- MATERIALS AND METHODS
fluratec 3: 24, Isotec 3: 12), in seven departments
We approached the departments of anaesthesia at
of anaesthesia, at different dial settings and flow
seven hospitals in the Copenhagen area and asked
rates. The range of output, for one type of vaporizer
them to participate in this study. All agreed, thus
and dial setting (flow: 6 litre m/n~') was largest
providing three university clinics, one county hos-
with the Fluotec 3 (0.85-1.55% when dial set to
pital and three smaller urban hospitals. Each de-
1%) and smallest with the Isotec 3 (0.85-1.15%
partment stated the type, age and number of
when dial set to 1 %). In determining the number of
vaporizers in daily use and the maintenance policy.
vaporizers with unacceptable inaccuracy, we ap-
The Bruel & Kjaer Anaesthetic Gas Monitor Type
plied acceptance limits of ±15% relative on each
1304 (Bruel & Kjaer, DK-2850 Naerum, Denmark)
vaporizer and each dial setting. Using a flow of
was used to measure vaporizer output. This monitor
oxygen 6 litre min~' 17 %of Fluotec 3,8% ofIsotec 3
has been investigated in our own laboratory and
and 71% of Enfluratec 3 vaporizers had outputs
results showed excellent linearity and accuracy [18].
outside those limits. Even when some specific
The resolution of the monitor is +0.05 vol%, and
conditions (vaporizers giving output beyond the
the accuracy in determining a given concentration
limits at any two or more dial settings; output
was found to be ^ ± 5 % relative (rel.).
beyond the limits in the clinically relevant range
This monitor was warmed up for 30 min before
(0.5-2%)) were added, a substantial number of
calibration, which was performed before each test
vaporizers did not perform within the limits. We
sequence at the respective department, using a
found a significantly greater accuracy of the vapor-
calibration gas mixture (Alphagaz, Cambridge,
izers after 3-monthly calibration checks (? < 0.05)
U.K.) with certified purity (anaesthetic agent
compared with vaporizers undergoing service and
0.4 + 0.008 % rel.). The same calibration gas canister
calibration only annually. Using a questionnaire, we
was used throughout the investigation. In addition,
found that fewer than 30% of the anaesthetists
despite the fact that all vaporizers had keyed filling
using the vaporizers would accept aberrance be-
systems, the type of anaesthetic vapour delivered
yond ±10% relative of the dial setting. (Br. J.
from each vaporizer was determined using the agent-
Anaesth. 1993; 7 1 : 453-457)
specific anaesthetic gas monitor Nellcor N-2500
(Nellcor, U.S.A.).
KEY WORDS Testing was carried out in an operating room of
Equipment: vaporizers, safety. each department, the ambient air temperature being
approximately 22 °C. All vaporizers were stored at
the same temperature. If any vaporizer needed
As the concern for patient safety during anaesthesia refilling before the test, the liquid anaesthetic agent
is increasing, so is interest in the performance of used had the same temperature as the ambient air.
equipment used during anaesthesia. Performance of The level of liquid in the vaporizers was between the
anaesthetic vaporizers has been examined for many
years [1-17], but the majority of these investigations
are case reports or are based upon a small number of
vaporizers. To our knowledge, only one report [17] JACOB NIELSEN*, M.D.; FINN MOLLER PEDERSEN, M.D.; FREDDY
KNUDSEN, M.D.; MAIKEN BRITT JENSEN, M.D.; MERETE IBSEN,
described the performance of a large number of M.D. ; Department of Anaesthesia and Intensive Care Unit, Herlev
vaporizers in clinical use; this was based on main- Hospital, University of Copenhagen, Medical School, DK-2730
tenance records obtained during a 4-year period. Herlev, Denmark. Accepted for Publication: March 11, 1993.
The aim of this study was to investigate the •Present address, for correspondence: University of Copen-
hagen, Department of Anaesthesia and Intensive Care Medicine,
accuracy of a large number of vaporizers in clinical KAS Gentofte, DK-2900 Hellerup, Denmark.
use, under controlled conditions, at several flow rates fPart of this study was presented orally at the 1991 ASA
and dial settings throughout the range, under meeting, San Francisco, U.S.A.
454 BRITISH JOURNAL OF ANAESTHESIA
TABLE I. Part of questionnaire distributed to each anaesthetist employed at the departments

1. What magnitude of abberance from the vaporizer 0%rel. 0-5%rel. 5-10% rel. 10-20% rel. >20%rel.
dial setting do you find acceptable?
• • D • •
• Dial setting irrelevant, dosage according
2. Are you using Isotec or Fluotec vaporizers (Tec 3) to clinical parameters.
with the dial set between the fixed marks in the D Often • Seldom/never
dial setting range 0-1 % ?
3. Are you using Enfluratec vaporizers with (Tec 3)
the dial set at 0.2/0.4/0.6/0.8%? D Often • Seldom/never
4. If you have answered seldom/never to question D do not trust the vaporizer at these dial settings
2, is the reason that you • do not find these dial settings clinically relevant
D do not use this type of vaporizer
5. If you have answered seldom/never to question 3, • do not trust the vaporizer at these dial settings
is the reason that you • do not find these dial settings clinically relevant
• do not use this type of vaporizer

"half" and "full" marks. All were transported sent annually to the manufacturer for a complete
upright and with the dial in the "off" position. Each service check; this includes disassembling, cleaning
vaporizer was mounted on an anaesthetic machine. If and calibration. All departments stated that these
this was done using the Selectatec system, we guidelines had been followed. Apart from this, one
ensured that the manifold port valve " O " rings were department had implemented 3-monthly calibration
intact, and the lever-lock was in the "lock" position. checks using an anaesthetic agent monitor (Datex
Similarly, for every vaporizer we checked that the Capnomac, Datex, Finland). If the vaporizer output
drain port was fully closed. exceeded some specific limits (+ 20 % rel. in the dial
At the vaporizer outlet, an ISO-cone with a 50-cm setting range 0 - 1 % , +15% rel. in the dial setting
p.v.c. hose was mounted. The sampling tube of the range 2—4%), the vaporizer was sent for recali-
monitor was connected via a Y-piece at the end of the bration. Oxygen was used as carrier gas at a flow rate
hose, thus providing a homogenous gas mixture and of 5 litre min"1.
a free outlet, avoiding pressure variations. Before
each test, the vaporizer was flushed with the dial set Criteria for acceptance
at 4 % and an oxygen flow of 8 litre min"1 for a few Because of lack of explicit data (in the Operators'
minutes, in order to remove any nitrous oxide Manuals) regarding the actual calibration limits, we
dissolved in the anaesthetic liquid, as this may have decided that the output was acceptable if it was
a major influence on output [19]. In addition, we within + 10% rel. of the dial setting. Added to this
monitored the eventual output of nitrous oxide was the error in measurement, caused by the gas
during the test sequence. The vaporizers were tested monitor and calibration gas ( ± 5 % rel.). As a
at three different flow rates of oxygen: 0.5, 2 and consequence, our acceptance limits were +15% rel.
6 litre min"1. Fluotec and Isotec vaporizers (Tec 3, (or +0.1 vol%, greatest value chosen). As the
Ohmeda Ltd, U.K.) had the dials set at 0.5, 1, 2, 3.5 Fluotec 3 and Isotec 3 vaporizers do not have fixed
and 5%. As these vaporizers are used commonly marks at dial settings "0.25" and "0.75", the
with the dial set between the fixed marks, they were outputs at these dial settings are not included in the
tested also at settings of "0.25%" and "0.75%". part of the investigation dealing with acceptability.
Transparent plastic scales (log scale) with marks Consequently, the Fluotec and Isotec vaporizers
representing 0, "0.25" and "0.75" (log centre) were tested for acceptability at five dial settings and
settings for each type of vaporizer were produced. the Enfluratecs at eight dial settings; all vaporizers at
The 0 marks of the plastic scale and the dial were three flow rates.
aligned and the dial was then set at "0.25" or
"0.75", thus testing each vaporizer under uniform Statistical analysis
conditions. Enfluratec vaporizers (Tec 3, Ohmeda
Ltd, U.K.) had the dials set at 0.2, 0.4, 0.6, 0.8, 1, 2, The accuracy of the Tec vaporizers undergoing
3.5 and 5%. quarterly calibration checks and of those submitted
only to annual service and calibration was compared
The output was noted at each dial setting and flow using chi-square analysis, as was the discrepancy
rate, and readings were made continuously until between answers in the questionnaire. P < 0.05 was
stability was reached, usually after 30 s. The date of regarded as statistically significant.
the last service check was also noted. After the test
sequence at each department, the Briiel & Kjasr
RESULTS
Anaesthetic Gas Monitor Type 1304 was checked
against the calibration gas in order to evaluate any A total of 114 vaporizers was reported to be in
drift. Before our investigation of the vaporizers, we clinical use at the seven departments. One hundred
distributed a questionnaire to each anaesthetist and two vaporizers (90 %) were tested; the rest were
employed at the departments (table I). in use for anaesthetic procedures at the time of our
visit. We decided to omit the data for two Penlon
Maintenance policy vaporizers and seven Drager Vapor 19.n vaporizers
According to guidelines issued by the Danish because of small numbers. Data for the equipment
National Board of Health, all vaporizers should be tested are given in table II. The date for next
ACCURACY OF ANAESTHETIC AGENT VAPORIZERS 455

TABLE II. Type of vaporizers and time from last calibration (number
100 -
or mean (range))

Number Time from last


Vaporizer studied calibration (months)
Fluotec 3 (5%) 58 7.0(2-16)
Isotec3(5%) 12 9.0(1-12)
Enfluratec 3 (5 %) 24 8.6(1-12)

Fluotec 3 (n = 58)

1
60-;
40-: [71 Fluotec 3 Isotec 3 Enfluratec 3


c 20^ (n=58) (n=12) (n = 24)
o
o • v, If FIG. 2. Vaporizers giving outputs beyond the acceptance limits
u

(±15% rel.) at one or more combinations of the five or eight dial
-20-: settings (0.2-5%) and the three flow rates of oxygen:
0.5 litre min"1 ( • ) ; 2 litre mhr 1 ( • ) ; 6 litre min"1 (0).
-40 ~ •
-60^ 1 1 1 1 1 1 1
"d.25" 0.5 "0.75" 1 2 3.5 100 - ,
i '• Dial setting (vol %)
80 -
Isotec 3 (n E

c 60 -
T3
o
>
ID

c
o
o
s Q.

Fluotec 3 Isotec 3 Enfluratec 3


"0.25 0.5 "0.75" 1 2 (n = 58) (n=12) (n = 24)
Dial setting (vol %)
FIG. 3. Vaporizers giving outputs beyond the acceptance limits
Enfluratec 3 (n = 24) (±15% rel.) under various conditions, but all at 6 litre min"1
flow. • = Beyond limits at any one or more of the five or eight
dial settings; • = beyond limits at any two or more of the five or
eight dial settings; 0 = beyond the limits at the dial settings
0.5 %, 1 % and 2 % (Fluotec 3 and Isotec 3) or 0.6 %, 1 % and 2 %
c (Enfluratec 3).
o

Isotec 3 (dial set to 1 %: output 0.85-1.15 %). Using


the Fluotec 3 and Isotec 3 vaporizers with the dial set
at "0.25"%, the mean output was substantially less
than the dial setting, and the relative range of
0.4 0.6 0.8 1 3.5 outputs was wider, as in the fixed settings. Only one
Dial setting (vol %)
of the Fluotec 3 vaporizers gave no output at the dial
FIG. 1. Relative variation from dial setting in Tec 3 vaporizers. setting "0.25"%; thus a "cut off" phenomenon
Each column shows range of relative variation ( # = mean) at was not demonstrated generally. When the dial was
fixed vaporizer settings. Carrier gas: oxygen 6 litre min"1. set at "0.75"%, the vaporizer performance was of
the same relative proportion as in the fixed settings.
calibration was overdue in three vaporizers (by 1, 2 A small flow dependency was seen with the Tec 3
and 4 months). The mean age of the Tec vaporizers vaporizers, giving excessive output at the greatest
(time in use) was approximately 10 yr. dial setting and slowest flow rate; this diminished as
The department using 3-monthly calibration the flow increased. The output was not influenced by
checks—performed by their own medical physics changes in flow rate (0.5-6.0 litre min"1) at dial
staff—stated that two vaporizers (of a total of 33 Tec settings most relevant clinically (0.5-2%).
3) had been sent for recalibration during the past Figure 2 illustrates the number of vaporizers with
year. Figure 1 illustrates the relative range and mean an unacceptable inaccuracy at any one of the 15 or 24
deviations of the vaporizer output from the dial combinations of flow and vaporizer setting (Fluotec
settings at a flow rate of 6 litre min"1. The range of and Isotec, three flow rates and five dial settings;
outputs, for one type of vaporizer and one dial Enfluratec, three flow rates and eight dial settings).
setting, was largest with the Fluotec 3 (e.g. dial set to In order to derive the percentage of vaporizers with
1%: output 0.85-1.55%) and smallest with the an unacceptable output, especially at clinically
456 BRITISH JOURNAL OF ANAESTHESIA
TABLE III. Number of vaporizers (Tec 3, n = 94) with outputs After the test sequence at each department, the
beyond the acceptance limits { + 20% rel. at dial settings 0-1 %; monitor was again supplied with the calibration gas;
+ 15% rel. at dial settings 2-4%) at a department using 3-monthly
calibration checks, compared with vaporizers undergoing only annual
drift was =$2.5%.
calibration. Carrier gas: oxygen 6 litre min~'. P < 0.05 (x2 test) Each of the 307 anaesthetists at the seven depart-
between groups ments received a questionnaire (140 anaesthetists,
167 certified registered nurse anaesthetists (CRNA))
Unacceptable Acceptable
output output
(table I). One hundred and ninety-four (63%) were
returned (anaesthetists 55 %, CRNA 70 %). Figure 4
Vaporizers submitted to shows the degree of vaporizer inaccuracy that was
3-monthly calibration 0 30 considered acceptable by the 194 anaesthetists.
checks (n = 30) Fewer than 30% would accept variation beyond
Vaporizers undergoing
service and calibration 10 54 ± 10 % rel. of the dial setting.
once a year (n = 64) In the dial setting range 0-1 %, the anaesthetists
used those vaporizers with several, fixed settings
(Enfluratec) significantly (P < 0.01) more often
45 -, (58%) than the vaporizers (Fluotec & Isotec) with
40 -
only one fixed setting (39%). Among those anaes-
thetists who seldom/never used the vaporizers in the
35 - low range of dial settings, significantly more (P <
o 0.01) stated the reason to be lack of confidence in the
TO 30 - vaporizers without fixed set points compared with
!t
to 25 - those vaporizers possessing them.
20 -
DISCUSSION
15 -
In general, we have found that the vaporizers had an
CD
Q. 10 - unacceptable inaccuracy and the outputs varied
greatly from one vaporizer to another. The accuracy
5 -
of vaporizers in clinical use at a department using 3-
EL monthly calibration checks was significantly greater
0% 0-5% 5-10% 10-20% >20% Clinical compared with vaporizers of the same type in use at
Acceptable variation (% rel.) departments using only annual service and cali-
FIG. 4. Variation in vaporizer output from the dial setting (% rel.)
bration.
considered acceptable by 78 anaesthetists ( 0 ) and 116 Certified The variation in output within one type of
Registered Nurse Anaesthetists (D)- Clinical = using clinical vaporizer was substantial (fig. 1), with relative
criteria. variations in the range — 50 % to + 70 %, at the fixed
settings. This was most pronounced using the
relevant settings, we then applied the acceptance Fluotec 3 compared with the Enfluratec 3 and Isotec
limits (+ 15 % rel.) to two specific conditions (fig. 3): 3, but this difference was caused by one Fluotec 3
vaporizers giving output beyond the limits at any with large errors. This lack of predictability of
two or more of the five or eight dial settings, and output is not clinically acceptable, as it may mask
output beyond the limits in the clinically relevant patient responses.
range of 0.5-2 % (0.5 or 0.6 %, 1 % 2 %; three data Another consequence of variability in output is
points for each type of vaporizer). The flow rate used that it places some vaporizers beyond the limits of
was 6 litre min"1. For comparison, the percentage of acceptance of accuracy. When choosing these ac-
vaporizers failing the test at any one or more of the ceptance limits, we have used primarily criteria of
five or eight dial settings throughout the range at the ± 10 % rel., adding to this the variability of the
flow rate of 6.0 litre min"1 is shown. measurement technique setup. These limits are the
There was no difference between the percentage of same as those used by other investigators [17]. The
vaporizers with unacceptable or acceptable output resulting limits are in overall accordance with the
regarding the mean time from last service (Tec 3 specifications stated by the manufacturer of Tec 3
vaporizers (« = 94): mean time from last service 7.5 vaporizers [20]. Applying the acceptance limits of
months for both unacceptable and acceptable per- + 15% rel. on measured output, a large proportion
formers). of the vaporizers performed beyond these limits.
Applying less restrictive acceptance limits This was most pronounced at low flows, but even at
( ± 2 0% rel. in the dial setting range 0 - 1 % , 6 litre min"1, 17 % of Fluotec 3, 8 % of Isotec 3 and
+ 15 % rel. in the dial setting range 2-4%) because 71 % of Enfluratec 3 gave outputs outside these
of the lesser resolution of the monitor used by the limits. As this also includes vaporizer settings
department performing its own calibration checks unlikely to be used in ordinary clinical procedures,
every 3 months, among all the Tec 3 vaporizers we added two specific conditions to the acceptance
tested (6 litre min"1), we found a significantly greater limits and applied these to the vaporizer output at
accuracy in the vaporizers submitted to the quarterly 6 litre min"1—a flow close to that at which the
calibration checks (/> < 0.05), than in vaporizers vaporizers were calibrated. The first condition was
undergoing only annual service and recalibration excessive output at any two or more of the dial
(table III). settings, thus including the vaporizers with the more
ACCURACY OF ANAESTHETIC AGENT VAPORIZERS 457

substantial variation: 10% of the Fluotec 3 were in anaesthetists at the seven departments showed that
this group, compared with none of the Isotec 3 and the Fluotec 3 and Isotec 3 vaporizers were used quite
15% of the Enfluratec 3. The second condition was often (39 %) with the dial setting positioned between
output beyond the acceptance limits in the clinically the fixed marks in the low concentration range, we
relevant dial setting range (0.5 or 0.6%, 1%, 2 % ; recommend calibration and marking of the "0.25"%
three data points for each vaporizer); on this basis, and "0.75"% set points.
15 % of the Fluotec 3, 10% of the Isotec 3 and 35 % From our results we believe it is reasonable to
of the Enfluratec 3 delivered excessive output. presume that all vaporizers could achieve acceptable
One reason for the greater accuracy of the Isotec 3 accuracy by implementing regular field calibration
vaporizers could be that, because of the cost of checks, and sending for recalibration only those
isoflurane, these have been used less frequently. vaporizers with output beyond certain limits. How-
Only a relatively small number (12) of Isotec 3 ever, the regular service check of the vaporizers with
vaporizers was investigated. As the precise measure- acceptable output should not be omitted, but the
ment of anaesthetic agent concentration in the gas interval extended.
mixture delivered to the patient during anaesthesia is
still not routine, the accuracy of anaesthetic agent ACKNOWLEDGEMENTS
vaporizers is important. While cardiovascular de-
pression caused by excessive vaporizer output may We are grateful to the Briiel & Kjsr Company (DK-2850 Naerum,
be recognized and corrected, it is better avoided Denmark) and the Nellcor Company (California, U.S.A.) for
providing the gas monitors for the study.
altogether.
Carter and colleagues [17] examined 631 main-
tenance records of vaporizers in the West of REFERENCES
Scotland; among these were 236 Fluotec 3 and 214 1. Morgan M, Lumley J. Reliability of halothane vaporizers.
Enfluratec 3. All vaporizers were given 6-monthly Anaesthesia 1968; 23: 440-445.
2. Paterson GM, Hulands GH, Nunn JF. Evaluation of a new
calibration checks; apart from this there was no halothane vaporizer: The Cyprane Fluotec mark 3. British
uniform policy for maintenance. Oxygen was used as Journal of Anaesthesia 1969; 41: 109-119.
carrier gas at 5-7 litre min"1 and the monitors were 3. Noble WH. Accuracy of halothane vaporizers in clinical use.
Riken interferometers. The acceptance limits were Canadian Anaesthetists Society Journal 1970; 17: 135-144.
+ 10%rel., and to this was added the error of 4. Murray WJ, Fleming P. Fluotec mark 2 halothane output:
Nonlinearity from "off" to 0.5 per cent dial settings.
measurement. They demonstrated calibration errors Anesthesiology 1972; 36: 180-181.
in 9 % of the Fluotec 3 and in 4 % of the Enfluratec 5. Latto IP. Administration of halothane in the 0-0.5%
3. Isotec 3 vaporizers were not examined. More concentration range with the Fluotec mark 2 and mark 3
vaporizers failed the acceptance limits in our study, vaporizers. British Journal of Anaesthesia 1973; 45: 563-569.
6. Dorsch JA. Understanding Anesthetic Equipment. Baltimore:
although the acceptance limits were identical and the Williams & Wilkins Co., 1975; 92-122.
flow rates similar. The apparently smaller variation 7. Inadvertent contamination of anaesthetic circuits with halo-
in the output found by Carter's group could be a thane. British Journal of Anaesthesia 1977; 49: 745-754.
result of the procedure of 6-monthly calibration 8. Cook TL, Eger El II, Behl RS. Is your vaporizer off?
checks and recalibration of the vaporizers failing the Anesthesia and Analgesia 1977; 56: 793-800.
9. Anonymous. Health Devices 1980; 10: 31-51.
acceptance limits, compared with the general policy 10. Lin CY. Assessment of vaporizer performance in low-flow
of maintenance revealed in our study, namely annual and closed-circuit anesthesia. Anesthesia and Analgesia 1980;
service and recalibration. In addition, the vaporizers 59: 359-366.
we investigated had been in use for 10 years, as 11. Miller JM, Cascorbi HF. Yet another vaporizer hazard.
Anesthesia and Analgesia 1980; 59: 805.
opposed to 4 years in use for the vaporizers 12. Novae GD, Ursillo RC. Malfunctioning halothane vaporizer.
investigated by Carter and colleagues. Anesthesia and Analgesia 1981; 60: 121.
At the one department in our study which had 13. Steffey EP, Woliner M, Howland D. Evaluation of an
implemented 3-monthly calibration checks, the ac- isoflurane vaporizer. The Cyprane Fortec. Anesthesia and
curacy of the vaporizers was significantly better. Analgesia 1982; 61: 457^64.
14. Carter JA, McAtteer P. A serious hazard associated with the
Because of the smaller resolution of the monitor used Fluotec mark 4 vaporizer. Anaesthesia 1984; 39: 1257-1258.
for measurement at this department, compared with 15. Reike H, Raffouf EM, Zuchner K, Hauschke D, Sonntag H.
that used in the present study, the limits were Genauigkeit von Halothanverdampfern in Abhangigkeit von
somewhat less restricted in the dial setting range Temperatur, Tragergaszusammensetzung und Gasflussrate.
(Abstract in English.) Anaesthesist 1985; 34: 522-527.
0-1%, and measurements at 5% were not per- 16. Duncan JA. Equipment malfunction: possible hazard.
formed. Anaesthesia 1986; 41: 1271.
The mean output of the Fluotec 3 and the Isotec 17. Carter KB, Gray WM, Railton R, Richardson W. Long-term
3 vaporizers with the dial setting between the fixed performance of Tec vaporizers. Anaesthesia 1988; 43:
marks of 0%, 0.5% and 1%, was somewhat less 1042-1046.
18. Nielsen J, Kann T, Moller JT. Evaluation of three multigas
than that expected, especially at "0.25%". This anesthetic monitors: The Briiel & Kjaer Anesthetic Gas
could be attributable to the fact that the vaporizers Monitor 1304, the Datex Capnomac Ultima and the Nellcor
were not calibrated at these points. We could not N-2500. Journal of Clinical Monitoring 1993; 9: 91-98.
demonstrate the "cut-off" phenomenon found by 19. Scheller MS, Drummond JC. Solubility of N2O in volatile
anesthetics contributes to vaporizer aberrancy when changing
Latto [5] using Fluotec 3 vaporizers with the dial carrier gases. Anesthesia and Analgesia 1986; 65: 88—90.
setting positioned midway between "OFF" and 0.5 %, 20. Operators Manual. Tec 3. Continuous flow vaporizer.
although there might be a difference in the exact Ohmeda, England.
position of the dial setting. As the survey among the

You might also like