Professional Documents
Culture Documents
1 s2.0 S0007091217457370 Main
1 s2.0 S0007091217457370 Main
APPARATUS
1. What magnitude of abberance from the vaporizer 0%rel. 0-5%rel. 5-10% rel. 10-20% rel. >20%rel.
dial setting do you find acceptable?
• • D • •
• Dial setting irrelevant, dosage according
2. Are you using Isotec or Fluotec vaporizers (Tec 3) to clinical parameters.
with the dial set between the fixed marks in the D Often • Seldom/never
dial setting range 0-1 % ?
3. Are you using Enfluratec vaporizers with (Tec 3)
the dial set at 0.2/0.4/0.6/0.8%? D Often • Seldom/never
4. If you have answered seldom/never to question D do not trust the vaporizer at these dial settings
2, is the reason that you • do not find these dial settings clinically relevant
D do not use this type of vaporizer
5. If you have answered seldom/never to question 3, • do not trust the vaporizer at these dial settings
is the reason that you • do not find these dial settings clinically relevant
• do not use this type of vaporizer
"half" and "full" marks. All were transported sent annually to the manufacturer for a complete
upright and with the dial in the "off" position. Each service check; this includes disassembling, cleaning
vaporizer was mounted on an anaesthetic machine. If and calibration. All departments stated that these
this was done using the Selectatec system, we guidelines had been followed. Apart from this, one
ensured that the manifold port valve " O " rings were department had implemented 3-monthly calibration
intact, and the lever-lock was in the "lock" position. checks using an anaesthetic agent monitor (Datex
Similarly, for every vaporizer we checked that the Capnomac, Datex, Finland). If the vaporizer output
drain port was fully closed. exceeded some specific limits (+ 20 % rel. in the dial
At the vaporizer outlet, an ISO-cone with a 50-cm setting range 0 - 1 % , +15% rel. in the dial setting
p.v.c. hose was mounted. The sampling tube of the range 2—4%), the vaporizer was sent for recali-
monitor was connected via a Y-piece at the end of the bration. Oxygen was used as carrier gas at a flow rate
hose, thus providing a homogenous gas mixture and of 5 litre min"1.
a free outlet, avoiding pressure variations. Before
each test, the vaporizer was flushed with the dial set Criteria for acceptance
at 4 % and an oxygen flow of 8 litre min"1 for a few Because of lack of explicit data (in the Operators'
minutes, in order to remove any nitrous oxide Manuals) regarding the actual calibration limits, we
dissolved in the anaesthetic liquid, as this may have decided that the output was acceptable if it was
a major influence on output [19]. In addition, we within + 10% rel. of the dial setting. Added to this
monitored the eventual output of nitrous oxide was the error in measurement, caused by the gas
during the test sequence. The vaporizers were tested monitor and calibration gas ( ± 5 % rel.). As a
at three different flow rates of oxygen: 0.5, 2 and consequence, our acceptance limits were +15% rel.
6 litre min"1. Fluotec and Isotec vaporizers (Tec 3, (or +0.1 vol%, greatest value chosen). As the
Ohmeda Ltd, U.K.) had the dials set at 0.5, 1, 2, 3.5 Fluotec 3 and Isotec 3 vaporizers do not have fixed
and 5%. As these vaporizers are used commonly marks at dial settings "0.25" and "0.75", the
with the dial set between the fixed marks, they were outputs at these dial settings are not included in the
tested also at settings of "0.25%" and "0.75%". part of the investigation dealing with acceptability.
Transparent plastic scales (log scale) with marks Consequently, the Fluotec and Isotec vaporizers
representing 0, "0.25" and "0.75" (log centre) were tested for acceptability at five dial settings and
settings for each type of vaporizer were produced. the Enfluratecs at eight dial settings; all vaporizers at
The 0 marks of the plastic scale and the dial were three flow rates.
aligned and the dial was then set at "0.25" or
"0.75", thus testing each vaporizer under uniform Statistical analysis
conditions. Enfluratec vaporizers (Tec 3, Ohmeda
Ltd, U.K.) had the dials set at 0.2, 0.4, 0.6, 0.8, 1, 2, The accuracy of the Tec vaporizers undergoing
3.5 and 5%. quarterly calibration checks and of those submitted
only to annual service and calibration was compared
The output was noted at each dial setting and flow using chi-square analysis, as was the discrepancy
rate, and readings were made continuously until between answers in the questionnaire. P < 0.05 was
stability was reached, usually after 30 s. The date of regarded as statistically significant.
the last service check was also noted. After the test
sequence at each department, the Briiel & Kjasr
RESULTS
Anaesthetic Gas Monitor Type 1304 was checked
against the calibration gas in order to evaluate any A total of 114 vaporizers was reported to be in
drift. Before our investigation of the vaporizers, we clinical use at the seven departments. One hundred
distributed a questionnaire to each anaesthetist and two vaporizers (90 %) were tested; the rest were
employed at the departments (table I). in use for anaesthetic procedures at the time of our
visit. We decided to omit the data for two Penlon
Maintenance policy vaporizers and seven Drager Vapor 19.n vaporizers
According to guidelines issued by the Danish because of small numbers. Data for the equipment
National Board of Health, all vaporizers should be tested are given in table II. The date for next
ACCURACY OF ANAESTHETIC AGENT VAPORIZERS 455
TABLE II. Type of vaporizers and time from last calibration (number
100 -
or mean (range))
Fluotec 3 (n = 58)
1
60-;
40-: [71 Fluotec 3 Isotec 3 Enfluratec 3
—
c 20^ (n=58) (n=12) (n = 24)
o
o • v, If FIG. 2. Vaporizers giving outputs beyond the acceptance limits
u
•
(±15% rel.) at one or more combinations of the five or eight dial
-20-: settings (0.2-5%) and the three flow rates of oxygen:
0.5 litre min"1 ( • ) ; 2 litre mhr 1 ( • ) ; 6 litre min"1 (0).
-40 ~ •
-60^ 1 1 1 1 1 1 1
"d.25" 0.5 "0.75" 1 2 3.5 100 - ,
i '• Dial setting (vol %)
80 -
Isotec 3 (n E
c 60 -
T3
o
>
ID
c
o
o
s Q.
substantial variation: 10% of the Fluotec 3 were in anaesthetists at the seven departments showed that
this group, compared with none of the Isotec 3 and the Fluotec 3 and Isotec 3 vaporizers were used quite
15% of the Enfluratec 3. The second condition was often (39 %) with the dial setting positioned between
output beyond the acceptance limits in the clinically the fixed marks in the low concentration range, we
relevant dial setting range (0.5 or 0.6%, 1%, 2 % ; recommend calibration and marking of the "0.25"%
three data points for each vaporizer); on this basis, and "0.75"% set points.
15 % of the Fluotec 3, 10% of the Isotec 3 and 35 % From our results we believe it is reasonable to
of the Enfluratec 3 delivered excessive output. presume that all vaporizers could achieve acceptable
One reason for the greater accuracy of the Isotec 3 accuracy by implementing regular field calibration
vaporizers could be that, because of the cost of checks, and sending for recalibration only those
isoflurane, these have been used less frequently. vaporizers with output beyond certain limits. How-
Only a relatively small number (12) of Isotec 3 ever, the regular service check of the vaporizers with
vaporizers was investigated. As the precise measure- acceptable output should not be omitted, but the
ment of anaesthetic agent concentration in the gas interval extended.
mixture delivered to the patient during anaesthesia is
still not routine, the accuracy of anaesthetic agent ACKNOWLEDGEMENTS
vaporizers is important. While cardiovascular de-
pression caused by excessive vaporizer output may We are grateful to the Briiel & Kjsr Company (DK-2850 Naerum,
be recognized and corrected, it is better avoided Denmark) and the Nellcor Company (California, U.S.A.) for
providing the gas monitors for the study.
altogether.
Carter and colleagues [17] examined 631 main-
tenance records of vaporizers in the West of REFERENCES
Scotland; among these were 236 Fluotec 3 and 214 1. Morgan M, Lumley J. Reliability of halothane vaporizers.
Enfluratec 3. All vaporizers were given 6-monthly Anaesthesia 1968; 23: 440-445.
2. Paterson GM, Hulands GH, Nunn JF. Evaluation of a new
calibration checks; apart from this there was no halothane vaporizer: The Cyprane Fluotec mark 3. British
uniform policy for maintenance. Oxygen was used as Journal of Anaesthesia 1969; 41: 109-119.
carrier gas at 5-7 litre min"1 and the monitors were 3. Noble WH. Accuracy of halothane vaporizers in clinical use.
Riken interferometers. The acceptance limits were Canadian Anaesthetists Society Journal 1970; 17: 135-144.
+ 10%rel., and to this was added the error of 4. Murray WJ, Fleming P. Fluotec mark 2 halothane output:
Nonlinearity from "off" to 0.5 per cent dial settings.
measurement. They demonstrated calibration errors Anesthesiology 1972; 36: 180-181.
in 9 % of the Fluotec 3 and in 4 % of the Enfluratec 5. Latto IP. Administration of halothane in the 0-0.5%
3. Isotec 3 vaporizers were not examined. More concentration range with the Fluotec mark 2 and mark 3
vaporizers failed the acceptance limits in our study, vaporizers. British Journal of Anaesthesia 1973; 45: 563-569.
6. Dorsch JA. Understanding Anesthetic Equipment. Baltimore:
although the acceptance limits were identical and the Williams & Wilkins Co., 1975; 92-122.
flow rates similar. The apparently smaller variation 7. Inadvertent contamination of anaesthetic circuits with halo-
in the output found by Carter's group could be a thane. British Journal of Anaesthesia 1977; 49: 745-754.
result of the procedure of 6-monthly calibration 8. Cook TL, Eger El II, Behl RS. Is your vaporizer off?
checks and recalibration of the vaporizers failing the Anesthesia and Analgesia 1977; 56: 793-800.
9. Anonymous. Health Devices 1980; 10: 31-51.
acceptance limits, compared with the general policy 10. Lin CY. Assessment of vaporizer performance in low-flow
of maintenance revealed in our study, namely annual and closed-circuit anesthesia. Anesthesia and Analgesia 1980;
service and recalibration. In addition, the vaporizers 59: 359-366.
we investigated had been in use for 10 years, as 11. Miller JM, Cascorbi HF. Yet another vaporizer hazard.
Anesthesia and Analgesia 1980; 59: 805.
opposed to 4 years in use for the vaporizers 12. Novae GD, Ursillo RC. Malfunctioning halothane vaporizer.
investigated by Carter and colleagues. Anesthesia and Analgesia 1981; 60: 121.
At the one department in our study which had 13. Steffey EP, Woliner M, Howland D. Evaluation of an
implemented 3-monthly calibration checks, the ac- isoflurane vaporizer. The Cyprane Fortec. Anesthesia and
curacy of the vaporizers was significantly better. Analgesia 1982; 61: 457^64.
14. Carter JA, McAtteer P. A serious hazard associated with the
Because of the smaller resolution of the monitor used Fluotec mark 4 vaporizer. Anaesthesia 1984; 39: 1257-1258.
for measurement at this department, compared with 15. Reike H, Raffouf EM, Zuchner K, Hauschke D, Sonntag H.
that used in the present study, the limits were Genauigkeit von Halothanverdampfern in Abhangigkeit von
somewhat less restricted in the dial setting range Temperatur, Tragergaszusammensetzung und Gasflussrate.
(Abstract in English.) Anaesthesist 1985; 34: 522-527.
0-1%, and measurements at 5% were not per- 16. Duncan JA. Equipment malfunction: possible hazard.
formed. Anaesthesia 1986; 41: 1271.
The mean output of the Fluotec 3 and the Isotec 17. Carter KB, Gray WM, Railton R, Richardson W. Long-term
3 vaporizers with the dial setting between the fixed performance of Tec vaporizers. Anaesthesia 1988; 43:
marks of 0%, 0.5% and 1%, was somewhat less 1042-1046.
18. Nielsen J, Kann T, Moller JT. Evaluation of three multigas
than that expected, especially at "0.25%". This anesthetic monitors: The Briiel & Kjaer Anesthetic Gas
could be attributable to the fact that the vaporizers Monitor 1304, the Datex Capnomac Ultima and the Nellcor
were not calibrated at these points. We could not N-2500. Journal of Clinical Monitoring 1993; 9: 91-98.
demonstrate the "cut-off" phenomenon found by 19. Scheller MS, Drummond JC. Solubility of N2O in volatile
anesthetics contributes to vaporizer aberrancy when changing
Latto [5] using Fluotec 3 vaporizers with the dial carrier gases. Anesthesia and Analgesia 1986; 65: 88—90.
setting positioned midway between "OFF" and 0.5 %, 20. Operators Manual. Tec 3. Continuous flow vaporizer.
although there might be a difference in the exact Ohmeda, England.
position of the dial setting. As the survey among the