Interlaboratory Comparison of
Ultrasonic Backscatter Coefficient
Measurements From 2 to 9 MHz
Keith A. Wear, PhD, Timothy A. Stiles, PhD, Gary R. Frank, Ernest L. Madsen, PhD,
Francis Cheng, PhD, Ernest J. Feleppa, PhD, Christopher S. Hall, PhD, Beom Soo
Kim, BS, Paul Lee, MS, William D. O’Brien, Jr, PhD, Michael L. Oelze, PhD,
Balasundar I. Raju, PhD, K. Kirk Shung, PhD, Thaddeus A. Wilson, PhD,
Jian R. Yuan, PhD
Objective. As are the attenuation coefficient and sound speed, the backscatter
coefficient is a fundamental ultrasonic property that has been used to character-
ize many tissues. Unfortunately, there is currently far less standardization for the
ultrasonic backscatter measurement than for the other two, as evidenced by a
previous American Institute of Ultrasound in Medicine (AIUM)–sponsored inter-
laboratory comparison of ultrasonic backscatter, attenuation, and speed mea-
surements (J Ultrasound Med 1999; 18:615–631). To explore reasons for these
disparities, the AIUM Endowment for Education and Research recently supported
this second interlaboratory comparison, which extends the upper limit of the fre-
quency range from 7 to 9 MHz. Methods. Eleven laboratories were provided with
standard test objects designed and manufactured at the University of Wisconsin
Abbreviations
AIUM, American Institute of Ultrasound in Medicine; (Madison, WI). Each laboratory was asked to perform ultrasonic measurements of
MS, megasamples; RF, radio frequency; SNR, signal-to- sound speed, attenuation coefficients, and backscatter coefficients. Each labora-
noise ratio tory was blinded to the values of the ultrasonic properties of the test objects at
the time the measurements were performed. Results. Eight of the 11 laborato-
Received September 21, 2004, from the Center for ries submitted results. The range of variation of absolute magnitude of backscat-
Devices and Radiological Health, US Food and Drug
Administration, Rockville, Maryland USA (K.A.W.); ter coefficient measurements was about 2 orders of magnitude. If the results of
University of Wisconsin, Madison, Wisconsin USA 1 outlier laboratory are excluded, then the range is reduced to about 1 order of
(T.A.S., G.R.F., E.L.M.); Boston Scientific, Fremont, magnitude. Agreement regarding frequency dependence of backscatter was bet-
California USA (F.C., J.R.Y.); Philips Research,
Briarcliff Manor, New York USA (C.S.H., B.I.R.); ter than reported in the previous interlaboratory comparison. For example, when
Riverside Research Institute, New York, New York scatterers were small compared with the ultrasonic wavelength, experimental fre-
USA (E.J.F., P.L.); University of Southern California, quency-dependent backscatter coefficient data obtained by the participating lab-
Los Angeles, California (B.S.K., K.K.S.); University of
Illinois at Urbana-Champaign, Urbana, Illinois USA oratories were usually consistent with the expected Rayleigh scattering behavior
(W.D.O., M.L.O.); and University of Tennessee, (proportional to frequency to the fourth power). Conclusions. Greater standard-
Memphis, Tennessee USA (T.A.W.). Revision request- ization of backscatter measurement methods is needed. Measurements of fre-
ed November 18, 2004. Revised manuscript accept-
ed for publication April 19. 2005. quency dependence of backscatter are more consistent than measurements of
The financial support of the American Institute absolute magnitude. Key words: backscatter; interlaboratory comparison; tissue
of Ultrasound in Medicine’s Endowment for characterization.
Education and Research is gratefully acknowl-
edged. The mention of commercial products, their
sources, or their use in connection with material
A
reported herein is not to be construed as either an ttenuation, sound speed, and the backscatter
actual or implied endorsement of such products by
the Food and Drug Administration. coefficient have been used to characterize many
Address correspondence to Keith A. Wear, PhD, tissues, including blood, heart, eye, liver, kidney,
Center for Devices and Radiological Health, Food prostate, and bone.1–18 In 1998, the American
and Drug Administration, 12720 Twinbrook Pkwy,
Rockville, MD 20852 USA. Institute of Ultrasound in Medicine (AIUM) sponsored an
E-mail: kaw@cdrh.fda.gov interlaboratory comparison of methods for measure-
© 2005 by the American Institute of Ultrasound in Medicine • J Ultrasound Med 2005; 24:1235–1250 • 0278-4297/05/$3.50
Ultrasonic Backscatter Coefficient Measurements From 2 to 9 MHz
ment of attenuation, sound speed, and back- oversaw the design, fabrication, and distribution
scatter.19 Ernest L. Madsen, PhD (University of of test phantoms to participating laboratories. All
Wisconsin, Madison, WI), oversaw the design, materials were produced at the University of
fabrication, and distribution of phantoms to 10 Wisconsin by methods reported previously.30–32
laboratories. Investigators performed measure- Three batches of phantoms were produced,
ments and submitted their results to Professor labeled A, B, and C. Each of the 11 participating
Madsen. The main conclusion of this effort was laboratories received 3 phantoms, 1 from each
that there was broad agreement among the 10 batch. As in the previous interlaboratory com-
laboratories regarding attenuation and sound parison, each laboratory performed backscatter
speed measurements but considerable disparity measurements (usually at multiple frequencies,
regarding backscatter coefficient measure- depending on which transducers they had avail-
ments.19 able) on each of the phantoms in their set.
There are several factors that may contribute to Because it is necessary to know the attenuation
the increased variance for backscatter measure- coefficient and sound speed to estimate the
ments. First, attenuation and sound speed backscatter coefficient, participating laborato-
measurements are usually performed in the ries measured and reported these parameters
through-transmission mode. Because of the rela- also.
tively low attenuation coefficients of biological The 3 types of tissue-mimicking materials were
tissues and tissue-mimicking media, the acquired produced to provide variety in ultrasonic proper-
radio frequency (RF) data (from which parameter ties while still being representative of soft tissues.
estimates are derived) generally have a relatively The properties of these materials were measured
high signal-to-noise ratio (SNR). Backscatter at the University of Wisconsin. Phantom A, which
coefficient measurements, on the other hand, are mimicked human liver, had an attenuation coef-
performed in the pulse-echo mode. Because of ficient slope approximately equal to 0.4 dB ⋅ cm–1
the weak scattering properties of soft biological ⋅ MHz–1. Its frequency (f ) dependence of
tissues and speckle noise, the SNR for backscat- backscatter was about f 2.1 over the frequency
tered RF data is comparatively low. A lower SNR range of 2.5 to 5 MHz. Phantom B had an attenu-
contributes to increased variance of measure- ation coefficient slope approximately equal to 0.7
ments. Second, there is a greater variety of mea- dB ⋅ cm–1 ⋅ MHz–1. Its frequency dependence was
surement procedures and algorithms for about f 3.2 over the frequency range of 2.5 to 5
assessment of the backscatter coefficient20–29 than MHz. Phantom C included millimeter-scale scat-
for the attenuation coefficient and sound speed. terers that sometimes occur in soft tissue. Its
Third, the mathematical complexity of algo- attenuation coefficient was in the range of values
rithms for computing the backscatter coefficient determined for in vitro human breast parenchy-
generally exceeds that for the attenuation coeffi- ma reported by D’Astous and Foster.33 Its
cient and sound speed. Finally, attenuation and backscatter coefficient (in the range of 2.5–5
sound speed measurement techniques may be MHz) was near the bottom of the range for in
thoroughly tested by performing measurements vitro human breast parenchyma. Its frequency
on standard objects (eg, plastics, liquids, and rub- dependence of backscatter was about f 0.55.
bers) for which the true values are known. (Average frequency dependences are f 1.3 for
References for backscatter coefficients on such human breast parenchyma and f 0.8 for infiltrat-
easily attainable test objects are comparatively ing duct carcinoma.33)
rare (if they exist at all). Phantom properties are summarized in Table 1.
To further investigate disparities in backscatter The materials for phantoms A and B were pro-
coefficient measurements, the AIUM sponsored duced according to the method of Madsen et al.31
a second interlaboratory comparison study. The They were “macroscopically uniform,” meaning
results of that study are reported herein. that the only nonuniformity in either material
resulted from the random positioning of micro-
Materials and Methods scopic glass bead scatterers. The scatterer num-
ber densities were about 5 scatterers/mm3 (A)
Phantoms and 800 scatterers/mm3 (B). The component
As in the previous AIUM-sponsored interlabora- materials (and relative amounts by weight) for
tory comparison project,19 Professor Madsen the type A phantoms were agarose (3.5%),
ple and then exits from the sample through the Saran layer and into water. Therefore, the total effect
on the magnitude of the ultrasound signal is equal to the product of the absolute value of Equation 1
with Zi = Zwater and Zf = Zsample and the absolute value of Equation 1 with Zi = Zsample and Zf = Zwater,
namely,
(3)
where Zwater, Zsample, and Zsaran are the acoustic transducer was displayed directly on the oscillo-
impedances of water, the sample, and Saran scope, which was triggered with a pulse from the
Wrap, respectively; csaran is the speed of sound in function generator that was synchronous with the
Saran Wrap; f is the frequency (in Hertz); α(f) is start of the output voltage pulse. Received wave-
the frequency-dependent attenuation coefficient forms with and without the Saran film in the
of Saran Wrap (in nepers per meter); and l is the ultrasound path were acquired. Ten measure-
thickness of the Saran Wrap. Note that all vari- ments in 1-MHz steps were made between 1 and
ables in this expression are real numbers. 10 MHz with the use of several pairs of transduc-
The thickness of the Saran Wrap used for these ers with common nominal center frequencies.
samples was measured with a micrometer to Thirty-one measurements were conducted
measure the thickness of 32 layers of the material between 15 and 50 MHz with 1 pair of 30-MHz
and found to be 25.1 ± 0.5 µm (mean ± SD). The broadband transducers. Table 2 provides details
acoustic impedance of a material is equal to its on each pair of transducers used for these mea-
density multiplied by its speed of sound. The den- surements. The amplitude transmission coeffi-
sity of Saran Wrap was measured by computing cient was obtained by dividing the amplitude with
the ratio of the mass and volume of a large sheet the Saran present by the amplitude without the
of Saran Wrap and found to be 1.69 ± 0.02 g/cm3. Saran present.
To quantify the speed of sound and attenuation The results of these measurements and a fit of
coefficient of Saran Wrap, the amplitude trans- the amplitude transmission coefficient given by
mission coefficient of a single layer of Saran Wrap Equation 1 are shown in Figure 1. The fit parame-
in water was measured by a narrow-band tech- ters correspond to a speed of sound of 2400 m/s
nique. The experimental setup consisted of a and an attenuation coefficient of 5.0 nepers ⋅ m–1 ⋅
water tank, a Wavetek (now Fluke Corporation, MHz–1.5 ⋅ f 1.5. The results of this fit for the speed of
Everett, WA) model 81 function generator, an sound and attenuation coefficient of Saran Wrap
Amplifier Research (Souderton, PA) model were used in Equation 3 to produce a table of
75A250 RF power amplifier, a LeCroy (Chestnut transmission coefficients using Equation 3 for
Ridge, NY) model 9410 digital storage oscillo- each sample that was sent to each participating
scope, and several pairs of unfocused transducers laboratory over the range of 1 to 50 MHz.
(Panametrics, Waltham, MA; and Aerotech,
Lewistown, PA) with various center frequencies. Data Acquisition and Analysis
The function generator created a low-amplitude The laboratory numbers in the following sections
voltage signal of 30 cycles at a specific frequency, have been assigned randomly and do not corre-
which was amplified and sent to a transmitting spond to the order of institutions listed on the first
transducer. The signal detected by the receiving page of the article.
(5)
spectral difference technique.38 Potential diffrac- speed was determined by Equation 4. The refer-
tion-related errors in this insertion technique ence sound speed for water was taken to be 1488
may be ignored, provided that the speed of m/s. Transit time difference (∆t) was measured
sound in the phantoms is close to that of water.39 via a correlation integral technique40 between
Backscatter coefficients were measured by a the water path and through-tissue signals col-
reference phantom method.28 Approximately lected at the hydrophone. Sample thickness was
25 echoes were acquired from each phantom measured with a ruler. The attenuation coeffi-
by translating the transducer laterally cient was measured by the same log spectral dif-
between acquisitions. The front surface of the ference method used by laboratory 1.
phantom was aligned with the transducer Backscatter coefficients were estimated by a cal-
focal plane. Radio frequency data were gated ibrated pulse-echo technique.26 The power spec-
to isolate the interior of the phantom and to trum was determined for regions of interest (20
exclude specular echoes. The range gate was wavelengths axially by 5 beam widths laterally) by
1.8 cm long and began 2 mm beyond the spec- taking the squared magnitude of the Fourier
ular echo. The ratios of the squared moduli of transform of gated sections from the backscat-
the fast Fourier transforms of the test phantom tered time signal. A Hamming window was used
RF data to reference phantom data were com- to gate the time signals. The effects of the equip-
puted. Compensation for attenuation was per- ment on the power spectrum measurement were
formed with Equation 17 from O’Donnell and factored out by dividing by the reference spec-
Miller.21 trum from a planar acrylic surface with known
All measurements were compensated for loss- reflectivity. Also, the medium’s attenuation was
es due to transmission through the Saran Wrap compensated for.
layer.
Laboratory 3
Laboratory 2 A Panametrics model 5900PR pulser/receiver
All measurements were performed in distilled, was used. Two different single-element Panametrics
highly degassed (PO2 = 1.3 mg/L) water at 22 ± transducers were used to scan the phantoms.
0.2°C. All measurements were compensated for The 3.5-MHz transducer had an aperture of 9.5
the transmission coefficient through the sam- mm and a focal length of 42 mm; the 5.0-MHz
ples’ surfaces. transducer had an aperture of 9.5 mm and a focal
Source transducers were driven by a Panametrics length of 50 mm.
model 5800 pulser/receiver. Three spherically The phantoms were scanned in 10 planes sepa-
focused transducers were used as sources for rated by 0.5 mm. Each scan plane consisted of 40
measurements of sound speed, attenuation, and scan lines separated by 0.5 mm. Data were digi-
backscatter: (1) Panametrics model V380 (diam- tized by an Acqiris (Monroe, NY) analog-to-digi-
eter of 25.4 mm, center frequency of 3.9 MHz, tal board operating at a 50 megasamples (MS)/s
fractional bandwidth [–3 dB] of 42%, focal length sampling rate. Each scan line was digitized with
of 90 mm, –6-dB focal beam width of 1.5 mm, 3400 eight-bit samples. Data were acquired with
and –6-dB depth of focus of 25 mm), (2) the transducer focus placed at the front (near)
Panametrics model V321 (diameter of 19.1 mm, surface of the phantom, that is, at a depth of 0
center frequency of 8.4 MHz, fractional band- mm, then at depths of 5, 10, 15, 20, 25, 30, 35, and
width of 34%, focal length of 95 mm, –6-dB 40 mm, that is, at the rear (far) surface of the
focal beam width of 1.1 mm, and –6-dB depth phantom.
of focus of 38 mm), and (3) ValpeyFisher The speed of sound was calculated by measur-
Corporation (Hopkinton, MA) model IS1004HR ing the phantom thickness and then determining
(diameter of 12.7 mm, center frequency of 10 the time between zero-crossing of the RF echo
MHz, fractional bandwidth of 42%, focal length signal at the front (near) and rear (far) surfaces of
of 38 mm, –6-dB focal beam width of 0.8 mm, the phantom. Time was determined by counting
and –6-dB depth of focus of 25 mm). the number of samples between the zero-cross-
The hydrophone used for sound speed and ings. (At 50 MS/s, the duration of 1 sample is 0.02
attenuation coefficient measurements was a 1- microseconds.) The transducer was focused mid-
mm2 bilaminar membrane hydrophone (model way into the phantom. Phantom thickness was
804; Perceptron, Plymouth Meeting, PA). Sound measured with a mechanical vernier caliper
amplifier, a LeCroy model 9410 digital storage The speed of sound and attenuation were mea-
oscilloscope, and several pairs of unfocused sured by a substitution method as described by
transducers (Panametrics and KB Aerotech [now laboratory 1. The reference speed of sound in
GE Inspection Technologies], Lewistown, PA) water was 1540 m/s. Sample thickness was mea-
with various center frequencies. The function sured with calipers. A substitution method was
generator created a 30-cycle low-amplitude volt- used to measure backscatter coefficient20 (also
age signal, which was amplified and sent to a see laboratory 7 methods). A flat stainless steel
transmitting transducer. The received signal was block was used for a reference reflector.
displayed on an oscilloscope, which was trig-
gered with a pulse from the function generator Laboratory 7
that was synchronous with the onset of the out- For sound speed measurements, a Panametrics
put voltage pulse. Received waveforms with and 5-MHz, 12.7-mm-diameter transducer was used.
without the sample in the ultrasound path were The Panametrics 5052UA pulse generator was
acquired. used as a source. A LeCroy 9400A digital oscillo-
Sound speed was computed from Equation 4 scope was used to digitize RF data.
with cw =1488.4 m/s (22.0°C).41 Sample thickness Samples were placed between the transducer
was measured with calipers. A zero-crossing was and a steel reflector. The time delays for the
used as a marker for arrival time measurement. echoes from the reflector (Tm) and from both
Attenuation was measured by a log spectral dif- interfaces of the sample (t1 and t2) were mea-
ference method as discussed in the section for sured. The sound speed (cm) was then calculated
laboratory 1. from
The measurement of backscatter coefficients
was performed according to the method of Chen (8)
et al.29 Echo signals from broadband source puls-
es were compared with the echo from a plane
reflector. Data were acquired from the sample in where Tw is the propagation time measured
a tank of degassed and distilled water. A function when the sample is removed. The reference
generator (FG 5010; Tektronix, Inc, Beaverton, speed of sound in water, cw , was 1481 m/s.42
OR) created 1-cycle sinusoidal pulses with a fre- For attenuation and backscatter measure-
quency near the transducer center frequency (3.5 ments, a pair of 7.5-MHz, 6-mm-diameter trans-
MHz). The pulses were amplified with a broad- ducers (Echo Ultrasound, Reedsville, PA) was
band RF power amplifier (240L; ENI, Rochester, used. The source was a Wavetek 191 function
NY) and used to drive the transducer. This result- generator. A log spectral difference method (see
ed in a broad-bandwidth acoustic pulse with methods for laboratory 1) was used to compute
approximately 50% fractional bandwidth. Echo the attenuation coefficient.
signals were received and amplified with a For backscatter measurements, a protection
Panametrics 5052UA pulser/receiver. A 10- circuit (Expander, DEX-3; Matec, Northborough,
microsecond interval was digitized at 100 MS/s MA) consisting of a series of high-frequency
with a LeCroy 9310A digital storage oscilloscope. diodes was placed between the source and the
Echo signals from the samples were obtained for transducer. Two limiters were placed between the
98 positions by translating the sample in a raster transducer and the oscilloscope. The backscat-
fashion with the use of a stepper motor system. tered signals were digitized by a 500-MS/s analog-
to-digital converter (CS 8500; Gage Applied
Laboratory 6 Technologies, Montreal, Quebec, Canada) at a
Transducers with center frequencies of 10 and 15 250-MS/s sampling rate. The specimen was
MHz were built in-house for use in this study. placed at the Rayleigh distance of a nonfocused
The computer-controlled acoustic field scanning transducer. The backscatter coefficient in a cylin-
program was used to measure the beam pattern. drical sample volume in the specimen was deter-
A calibrated hydrophone, HP function generator mined by measuring the backscattered power
(Hewlett-Packard Company, Palo Alto, CA), ZHL- and normalizing it to the power reflected from a
32A amplifier (Mini-Circuits, Brooklyn, New perfect plane reflector (to correct for incident
York), TDS-520 oscilloscope (Tektronix, Inc), and intensity and beam characteristics).20 Phantom
DC power supply were used. thickness was measured with calipers.
to the same procedures as those used before the Nevertheless, it is perhaps worthwhile to point
samples had been sent out. Thus, any change out that laboratory 2 obtained the closest agree-
(drift) in the material properties of the samples ment with theory. Laboratories 1, 4, and 8 agreed
could be assessed. Measurements were per- with theory to within about a factor of 2.
formed at multiple frequencies throughout the Moreover, most laboratories measured frequen-
2- to 14-MHz range. cy dependence of backscatter to be fairly consis-
tent with theoretical prediction.
Results In the case of phantom B (Figure 4B), very
good agreement existed among the laboratories
In the results that follow, laboratories are anony-
mously labeled numbers 1 through 8, assigned
Figure 2. Measurements of frequency-dependent attenuation
randomly. The numbers in the figures corre- coefficients for phantom A, phantom B, and phantom C.
spond to the laboratory identification numbers
in “Materials and Methods” but do not corre-
spond to the order of the institutions listed on the
first page of this article.
Figure 2 shows measurements of attenuation
coefficients for the 3 phantoms, as functions of
frequency, for the 8 laboratories. Agreement was
within 0.15 dB ⋅ cm–1 ⋅ MHz–1 for phantom A and
within 0.2 dB ⋅ cm–1 ⋅ MHz–1 for phantom B but
only within 0.5 dB ⋅ cm–1 ⋅ MHz–1 for phantom C.
Figure 3 shows measurements of sound speed
for the 3 phantoms, as functions of frequency, for
the 8 laboratories. There was broad agreement
for most laboratories, but some measurements
(laboratories 3, 4, and 6) were at times 30 to 60
m/s greater than the majority.
Figure 4 shows measurements of backscatter
coefficient for the 3 phantoms, as functions of
frequency, for the 8 laboratories. In the case of
phantom A (Figure 4A), most measurements
showed similar rates of increase of the backscat-
ter coefficient with frequency between 2 and 5
MHz. The laboratories that performed measure-
ments between 5 and 9 MHz seemed to agree
that the backscatter coefficient was roughly con-
stant within that frequency range. The solid
curve in Figure 4A depicts the theoretical
backscatter coefficient, based on the theory of
Faran,44 assuming a scatterer diameter of 82.5
µm (the midpoint of the range from 75 to 90 µm).
The theory should not necessarily be taken to be
a perfectly accurate estimate because it is based
on numerous assumptions and approximations.
The beads were not perfect spheres, as the theo-
ry assumes, and their true material properties
were not known (generic values for glass were
used instead). The true scatterer number density
was only known to within about 10%. The theory
ignores coherent scattering and multiple scatter-
ing, which may have been present. The true dis-
tribution of scatterer sizes was not known.
(except laboratory 7) regarding the magnitude cates the frequency dependence of backscatter
and frequency dependence of the backscatter for Rayleigh scattering (frequency to the fourth
coefficient. The solid curve in Figure 4B depicts power), which is valid at frequencies less than 6
the theoretical backscatter coefficient, based on MHz. Most laboratories measured frequency
the theory of Faran,44 assuming a scatterer diam- dependence of backscatter to be fairly consistent
eter of 26 µm (the midpoint of the range from 9 with theoretical prediction.
to 43 µm). As explained above, caution must be
exercised in comparison of experimental data to Figure 4. Measurements of frequency-dependent backscatter
this theory. The dashed curve in Figure 4B indi- coefficient for phantom A, phantom B, and phantom C. Solid
curves correspond to the theoretical backscatter coefficients,
based on the theory of Faran.44 The dashed curve in B indicates
Figure 3. Measurements of sound speed (SOS) for phantom A, the frequency dependence of backscatter for Rayleigh scattering
phantom B, and phantom C. (frequency to the fourth power), which is valid at low frequencies.
seen that in the worst case (phantom C), mea- different values. Some methods use focused
surements varied by as much as about 70 m/s or transducers, whereas others use unfocused
about 5%. This is a high variability for a relatively transducers. Under some conditions, focused
easy measurement. However, it would only pro- and unfocused transducers may yield disparate
duce 5% variations in backscatter coefficient results.45 For calibration, some methods use a
measurements, a negligible fraction of the total planar reflector, whereas others use a phantom
observed (1–2 orders of magnitude). The within- that contains distributed pointlike scatterers.
batch nonuniformity (Table 3) and temporal The model for the backscattering object may be
drift (Table 4) of sound speed of the phantoms
were less than or on the order of 3 m/s, again cor-
responding to negligible variations of backscat- Figure 5. Scatterplots of attenuation coefficient measurements
versus backscatter coefficient measurements at 5 MHz for phan-
ter coefficient estimates. tom A, phantom B, and phantom C.
It is also unlikely that attenuation coefficient
variations (whether due to phantom nonunifor-
mity, temporal drift, or disparity of measure-
ments among participating laboratories)
observed in this study could explain a significant
fraction of the disparity in backscatter coefficient
measurements. Errors in attenuation coefficient
estimates lead to inaccurate compensation of
backscatter data for the effects of attenuation. If
attenuation coefficient variation were a primary
determinant of backscatter coefficient variation,
then those laboratories underestimating attenu-
ation coefficients would be expected to under-
compensate their backscatter data and report
low backscatter coefficients. Similarly, those lab-
oratories overestimating attenuation coefficients
would be expected to report high backscatter
coefficients. Thus, one might expect a positive
correlation between estimates of attenuation
and backscatter coefficients. However, as indi-
cated in Figure 5, no such positive correlation
was observed in this study. In addition, the with-
in-batch nonuniformity (Table 3) and temporal
drift (Table 4) of attenuation coefficients for the
phantoms were less than or on the order of 0.6
dB/cm at 5 MHz. Because the phantoms were
about 4 cm thick, this would lead to a range of
attenuation of less than or on the order of 1.5 dB
at 5 MHz. This is far too small to produce any-
where near the 1 or 2 order(s) of magnitude (ie,
10–20 dB) variation in the backscatter coefficient
estimate that was observed.
Within-batch nonuniformity and temporal
drift of backscatter coefficients were less than
36%. Again, this variability is small compared
with the 1 to 2 orders of magnitude variation of
measurements reported by the participating lab-
oratories.
One possible source of disparity that remains is
that different data reduction algorithms, even if
implemented correctly, may yield substantially
continuous or discrete. Under these circum- them refine their backscatter coefficient mea-
stances, production of different quantitative val- surement methods. It is hoped this will lead to
ues by different approaches should not be improved consistency in future reports of fre-
surprising. quency-dependent backscatter coefficients
Experimental error may also be an appreciable among the participating laboratories.
source of disparity. Data acquisition issues (cor-
rect positioning of transducers relative to sam- References
ples, correct positioning of reference reflectors,
avoidance of saturation of receiving electronics 1. Shung KK, Sigelmann RA, Reid JM. Scattering of
and digitizer, and optimization of transmitted ultrasound by blood. IEEE Trans Biomed Eng 1976;
power to maximize the SNR without saturation) 23:460–467.
could contribute to errors. In addition, data
2. Angelsen BAJ. A theoretical study of the scattering
reduction software errors, which may remain
of ultrasound from blood. IEEE Trans Biomed Eng
undetected in the absence of a reference stan-
1980; 27:61–67.
dard test, can also exacerbate inaccuracy. For
example, although laboratories 6 and 7 both 3. Madaras EI, Barzilai B, Perez JE, Sobel BE, Miller JG.
used the same method for backscatter coefficient Changes in myocardial backscatter throughout the
estimation, they obtained markedly dissimilar cardiac cycle. Ultrason Imaging 1983; 5:229–239.
results for phantoms B and C. At least one of
4. Campbell JA, Waag RC. Measurements of calf liver
these laboratories is likely to have had experi-
ultrasonic differential and total scattering cross sec-
mental or software error.
tions. J Acoust Soc Am 1984; 75:603–611.
Although considerable disparity existed
among the participating laboratories regarding 5. Lizzi FL, King DL, Rorke MC, et al. Comparison of
absolute levels of backscatter, the level of agree- theoretical scattering results and ultrasonic data
ment regarding frequency dependence of the from clinical liver examinations, Ultrasound Med Biol
backscatter coefficient was fair. This establishes 1988; 14:377–385.
some promise for standardized estimation of
6. Wear KA, Milunski MR, Wickline SA, Perez JE, Sobel
scatterer size.
BE, Miller JG. Differentiation between acutely
Finally, the present study offers several im-
ischemic myocardium and zones of completed
provements over the previous AIUM-sponsored
infarction in dogs on the basis of frequency-depen-
interlaboratory comparison.19 First, the upper
dent backscatter. J Acoust Soc Am 1989; 85:
range of frequencies has been extended from 7 to
2634–2641.
9 MHz (the highest frequency for which multiple
measurements are available for comparison). 7. Turnbull DH, Wilson SR, Hine AL, Foster FS.
Second, the analysis of phantom uniformity and Ultrasonic characterization of selected renal tissues.
drift has been expanded from attenuation and Ultrasound Med Biol 1989; 15:241–253.
sound speed (in the previous study) to also
8. Insana MF, Hall TJ, Fishback JL. Identifying acoustic
include backscatter (in this study). Consequently,
scattering sources in normal renal parenchyma from
the candidate mechanisms for disparity in
the anisotropy in acoustic properties. Ultrasound
backscatter coefficient measurements in this
Med Biol 1991; 17:613–626.
study could be narrowed down more convincing-
ly to algorithmic differences and software/exper- 9. Insana MF, Wood JG, Hall TJ. Identifying acoustic
imental error. Third, the results in this study scattering sources in normal renal parenchyma in
were compared with theoretically expected vivo by varying arterial and ureteral pressures.
backscatter coefficients, allowing independent Ultrasound Med Biol 1992; 18:587–599.
and objective evaluation of laboratory measure-
10. Sehgal CM. Quantitative relationship between tissue
ments. Fourth, and most important, the test
composition and scattering of ultrasound. J Acoust
phantoms in this study (unlike the previous
Soc Am 1993; 94:1944–1952.
study) have been returned to those laboratories
requesting them. Now participating laboratories 11. Wear KA, Garra BS, Hall TJ. Measurements of ultra-
have standardized test objects, with properties sonic backscatter coefficients in human liver and
documented in this article (with moderate con- kidney in vivo. J Acoust Soc Am 1995; 98:1852–
sistency if not complete unambiguity), to help 1857.
12. Feleppa EJ, Kalisz A, Sokil-Melgar JB, et al. Typing of ultrasonic tissue characterization. J Acoust Soc Am
prostate tissue by ultrasonic spectrum analysis. IEEE 1983; 73:1366–1373.
Trans Ultrason Ferroelectr Freq Control 1996;
24. Campbell JA, Waag RC. Normalization of ultrasonic
43:609–619.
scattering measurements to obtain average differ-
13. Silverman RH, Folberg R, Boldt HC, et al. Correlation ential scattering cross sections for tissues. J Acoust
of ultrasound parameter imaging with microcircula- Soc Am 1983; 74:393–399.
tory patterns in uveal melanomas. Ultrasound Med
25. Madsen EL, Insana MF, Zagzebski JA. Method of
Biol 1997; 23:573–581.
data reduction for accurate determination of acous-
14. Russell MD, Sun H, Wicks ML, et al. Ultrasonic non- tic backscatter coefficients. J Acoust Soc Am 1984;
destructive evaluation applied to prostate cancer 76:913–923.
detection. Nondestr Test Eval 1998; 14:237–256.
26. Insana MF, Hall TJ. Parametric ultrasound imaging
15. Wear KA. Frequency dependence of ultrasonic from backscatter coefficient measurements: image
backscatter from human trabecular bone: theory formation and interpretation. Ultrason Imaging
and experiment. J Acoust Soc Am 1999; 106: 1990; 12:245–267.
3659–3664.
27. Yao LZ, Zagzebski JA, Madsen EL. Backscatter coef-
16. Hall CS, Scott MJ, Lanza GM, et al. The extracellular ficient measurements using a reference phantom to
matrix is an important source of ultrasound extract depth-dependent instrumentation factors.
backscatter from myocardium. J Acoust Soc Am Ultrason Imaging 1990; 12:58–70.
2000; 107:612–619.
28. Zagzebski JA, Yao LX, Boote EJ, Lu ZF. Quantitative
17. Oelze ML, Zachary JF, O’Brien WD Jr. backscatter imaging. In: Shung KK, Thieme GA
Characterization of tissue microstructure using ultra- (eds). Ultrasonic Scattering in Biological Tissues.
sonic backscatter: theory and technique for opti- Boca Raton, FL: CRC Press Inc; 1993:451–486.
mization using a Gaussian form factor. J Acoust Soc
29. Chen JF, Zagzebski JA, Madsen EL. Tests of backscat-
Am 2002; 112:1202–1211.
ter coefficient measurement using broadband puls-
18. Oelze ML, O’Brien WD, Blue JP, Zachary JF. es. IEEE Trans Ultrason Ferroelectr Freq Control
Differentiation and characterization of rat mamma- 1993; 40:603–607.
ry fibroadenomas and 4T1 mouse carcinomas using
30. Chen X, Phillips D, Schwarz KQ, Mottley JG, Parker
quantitative ultrasound imaging. IEEE Trans Med
KJ. The measurement of backscatter coefficient
Imaging 2004; 23:764–771.
from a broadband pulse-echo system: a new for-
19. Madsen EL, Dong F, Frank GR, et al. Interlaboratory mulation. IEEE Trans Ultrason Ferroelectr Freq
comparison of ultrasonic backscatter, attenuation, Control 1997; 44:515–525.
and speed measurements. J Ultrasound Med 1999;
31. Madsen EL, Frank GR, Dong F. Liquid or solid ultra-
18:615–631.
sonically tissue-mimicking materials with very low
20. Sigelmann RA, Reid JM. Analysis and measurement scatter. Ultrasound Med Biol 1998; 24:535–542.
of ultrasound backscattering from an ensemble of
32. Chin RB, Madsen EL, Zagzebski JA, Jadvar H, Wu X-
scatterers excited by sine-wave bursts. J Acoust Soc
K, Frank GR. A reusable perfusion supporting tissue-
Am 1973; 53:1351–1355.
mimicking material for ultrasound hyperthermia
21. O’Donnell M, Miller JG. Quantitative broadband phantoms. Med Phys 1990; 17:380–390.
ultrasonic backscatter: an approach to nondestruc-
33. D’Astous FT, Foster FS. Frequency dependence of
tive evaluation in acoustically inhomogeneous mate-
ultrasonic attenuation and backscatter in breast tis-
rials. J Appl Phys 1981; 52:1056–1065.
sue. Ultrasound Med Biol 1986; 12:795–808.
22. Nicholas D, Hill CR, Nassiri DK. Evaluation of
34. Madsen EL, Zagzebski JA, Frank GR. Oil-in-gelatin
backscattering coefficients for excised human tis-
dispersions for use as ultrasonically tissue-mimicking
sues: principles and techniques. Ultrasound Med
materials. Ultrasound Med Biol 1982; 8:277–287.
Biol 1982; 8:7–15.
35. Ford RD. Introduction to Acoustics. Amsterdam, the
23. Lizzi FL, Greenebaum M, Feleppa EJ, et al.
Netherlands: Elsevier; 1970:77.
Theoretical framework for spectrum analysis in