Installation Manual

D700 Dental Set

Technical Name: D700 Dental Set


Trade Name: D700 Dental Set
Commercial Model: D700
Table of content
1 - Introduction
2 - Preliminary installation information
3 - The Dental Unit D700
4 - Principal Dental Unit dimensions
5 - Necessary tools for installation
6 - Installing the dental Unit D700
6.1 - Opening packing box 1
6.2 - Items included in packing 1
6.3 - Undoing the setting of packing items
6.4 - Installing the auxiliary unit on the chair
6.5 - Installing the delivery unit on the chair
6.6 - Installing the reflector on the chair
6.7 - Setting the arm on the chair
6.8 - Connecting the equipment to the chair
6.9 - Connecting the unit to the chair
6.10 - Connecting the reflector to the chair
6.11 - Opening packing box 2
6.12 - Items included in packing 2
6.13 - Mounting the head rest
6.14 - Mounting the head rest filling
6.15 - Mounting the seat filling

7 - Annexes
Annex 1 - Electrical connection layout
Annex 2 - Electronic layout
Annex 3 - Hydropneumatic layout
Annex 4 - Exploded drawings
Annex 5 - Exploded drawings
Annex 6 - Exploded drawings
Annex 7 - Exploded drawings

Installation manual - Dental Unit D700 03


1 - Introduction 2.3 - The dental unit can be connected in networks
with input voltage of 110/127/220/240V +/-10%
(50 and 60Hz). Being so, it is indispensable to
This manual contains all necessary information for verify the input voltage of your network and
the Dental Set D700 installation. These activities perform the respective connection, according to
can be performed by Dabi Atlante Authorized connection layout in ANNEX-01.
Technician, by others technicians or also by the 2.4 - The equipment ground is indispensable.
customer.

It is of utmost importance the reading and


3 - The Dental Unit D700
comprehension of this manual so that the
installation and regulation procedures may be 1
successful.

The room where the equipment will be installed 2


should be in accordance with the technical 3
specifications contained in the dental unit's
installation plan.

The equipment will be subject to short circuit and


4
corrosion in case there is any penetration of water
and other liquids in its interior.

The manufacturer cannot be made responsible for


5
damage or malfunction, in case the equipment is
installed or operated in an inadequate manner.

2 - Preliminary installation Main parts 6


1 - Reflector
information 2 - Auxiliary Unit
3 - Coupled Dental unit
2.1 - In the room where the dental unit is installed, 4 - Chair
the place where the equipment will be positioned 5 - Activation pedal
should have an even floor. 6 - Control box (inbuilt in the chair)
2.2 - It is important that the room where the
dental unit is installed has an area (m²) that
permits its positioning in accordance with the
ISO/FDI ergonomic norms.
m
0c
=5
C

m
0c
=5
B

cm
50
A=
35cm

para os demais equipos,


Flex Air Eletronic,
mínimo = 80cm

mínimo = 65cm

30cm 25cm
para equipo

NORMAS RECOMENDADAS PELA ISO/FDI*


Instalação de equipamentos no consultório
*ISO - International Standards Organization
*FDI - Fédération Dentaire Internationale mínimo = 150cm

Installation manual - Dental Unit D700 04


4 - Principal Dental Unit dimensions

Dental Set D700: view in elevation.


Legend: Maximum height.

Dental Set D700: floor view.


Legend: Maximum width, maximum length.

Installation manual - Dental Unit D700 05


5 - Necessary tools for installation

Philips turnscrew Screwdriver

Universal pliers Long-nose pliers

Fixed open or ring wrench size Barrel screwdriver size


5/16”, 7/16” and 1/2” 5/16” and 1/2”

Obs: It is important to emphasize that the principal tools were listed above for the dental
unit installation and that these tools do not go with the equipment.

If the customer opted for installing the equipment without the aid of the Technician, we
advise the acquisition of the listed items or perform a “loan”.

Installation manual - Dental Unit D700 06


6 - Installing the Dental Unit D700
6.1 - Opening packing box “01”

Step 1 Step 2
Remove the packing staples (indicated by the Remove the carton box, by pulling it upwards. It
arrows), using a screwdriver or any similar tool. is advisable that this operation be executed by 2
people.

6.2 - Items included in packing “01”

Chair* E.V.A Mat Head rest

Coupled equipment Auxiliary unit Chair arm rest

* The filling in this figure is only illustrative; it is not part of this packing.

Installation manual - Dental Unit D700 07


Reflector Motor cover Support of connection items

K4 K5 K6
K1 K2 K3

Kit Support and pet deposit

6.3 - Undoing the setting of packings items

3 7
4
6

8 9
5

List of items in this packing:


1- Basin 4- Chair Arm 7- Chair
2- Equipment 5- Motor Cover 8- Equipment Support
3- Head Rest 6- Reflector 9- Unit

Installation manual - Dental Unit D700 08


Step 3 (Reflector): Remove the arm setting Step 4: Cut the arm setting clamp. Step 5: Remove the packing Reflector and
lock. place it on the floor carefully so as not to
damage it.

Step 6 (Equipment): Cut the arm setting Step 7: Remove The packing Equipment Step 8 (Cask Packing): Remove the cask
clamp. and place it in the floor carefully so as not packing and place it on the floor carefully
to damage it. so as not to damage it.

Step 9 (Chair Arm): Loosen the arm, using Step 10 (Motor Cover): Remove the motor Step 11 (Unit): Loosen the unit, using a
a 1/2” open or ring wrench, place it on the cover and place it on the floor carefully so 7/16” open or ring wrench.
floor carefully so as not to damage it. as not to damage it.

Step 12 (Unit): Remove the unit and place Step 13 (Equipment): Remove the support Step 14 (Equipment): Loosen the
it on the floor carefully so as not to damage cover, in order to facilitate the next equipment support, using a 1/2” barrel
it. operation. screwdriver, place it on the floor carefully
so as not to damage it.

Installation manual - Dental Unit D700 09


Step 15: Cut the joint setting clamp of the Step 16: Remove the joint axle-setting ring Step 17: Raise the chair’s headrest.
chair headrest. of the headrest and remove the axle.

Step 18: Position the joint hole with the Step 19: Fix the joint axle with the ring Step 20: Loosen the seat complement,
coupling hole on the chair’s headrest and (according to the photo). using a 1/2” open or ring wrench.
place the joint axle.

Step 21: Position the seat complement, Step 22: Set the seat complement, using a Step 23: Loosen the chair from the pallet,
according to the photo. 1/2” open or ring wrench. using a 1/2” open or ring wrench.
(2 screws and 2 nuts)

Step 24: Position the mat on the floor, Step 25: Remove the chair from the pallet,
observing the dental unit location inside and place it on top of the mat. It is
the room. advisable that this operation be executed
by 2 people.

Installation manual - Dental Unit D700 10


6.4 - Installing the auxiliary unit on the chair

!
AMBIDEXTROUS DENTAL UNIT: The mounting of the auxiliary unit
and equipment support should be on the right side of the chair for
left-handed dentist.

Step 26: Remove the rack cover using a Step 27: Position the unit on the seat plate Step 28: Loosen the unit body end support,
Philips screwdriver and a 5/16” fixed and set it using a 1/2” open or ring wrench, using a Philips screwdriver.
wrench. and the items K1 of the kit indicated on
page 7.

Step 29: Position the unit body point Step 30: Pass the unit hoses and cable Step 31: Mount the basin, the cup holder
support and set it using the Philips through the upper rack hole and, leaving grill; the basin grill and basin grill cover.
screwdriver. them as indicated on the photo.

6.5 - Installing the delivery unit

Step 32: Position the equipment support Step 33: Position PET support on the Step 34: Mount the support cover, by
on the seat plate and set it using a 1/2” equipment support and set it using a 5/16” positioning it as indicated in the photo.
open or ring wrench and items K2 of the kit barrel screwdriver.
indicated on page 7.

Installation manual - Dental Unit D700 11


Apply grease on the plug before executing step 35. Apply grease on the plug before executing step 36.

Step 35: Mount the plug in the “smaller” Step 36: Get the equipment mount it on Step 37: Fit correctly the arm joint parts.
tube of the equipment (please refer to the support, passing the hoses through the
position according to the photo), being plug, fit the tube into the plug, being
careful such that the lateral split of the careful such that the tube pin may fit in the
plug fits in the tube pin. plug split.

Step 38: Pass the equipment hoses above Step 39: Then pass the equipment hoses Step 40: Cut the arm protection net using a
the beams of the equipment support. through the clamp located below the seat pair of scissors.
plate and tighten the clamp, leaving
slightly free.

Step 41: Unpack the equipment head items Step 42: Pass the equipment and unit Step 43: Mount the rack cover using a
and position them on the end support, hoses and cables through the elevation Philips screwdriver and a 5/16” fixed open
according to the photo. lever. Perform an evaluation in order to wrench, operation contrary to Step 25.
Note: The high rotation instruments and micro motor assure that the hoses are not bent.
are sold separately.

Installation manual - Dental Unit D700 12


6.6 - Installing the reflector on the chair

Step 44: Mount the plug in the “larger” Step 45: Get the reflector mount it on the Step 46: Fit correctly the arm joint parts.
reflector tube (please refer to position support, passing the wires through the Cut the reflector arm safety clamp.
according to Step 34), being careful such plug, fits the tube in the plug, being careful
that the lateral split of the plug fits in the such that the pint (please refer to inside
tube pin. the tube) fits in the plug split.

Step 47: Pass the wires of the reflector Step 48: Cut the arm protection net using a
above the equipment support beams. pair of scissors.

6.7 - Setting the arm on the chair and placing rack cover finishing

Step 49: Position the arm on the seat plate Step 50: Fit the finishing rubber on the
and set it using two 1/2” Sext. screws rack cover.
removed during the execution of Step 9
and a 1/2” fixed open or ring wrench.

Installation manual - Dental Unit D700 13


6.8 - Connecting the equipment to the chair

D
G
H

H
H

B
J

Step 51: Separate the hoses that are Step 52: Connect hose H in nipple H, Step 53: Connect hose G to the upper
identified with letters, according to the connect hose B in nipple B, connect hose J nipple and the other point of hose H in the
above figure. to nipple J. Connect hose D to the pedal PET deposit support. Then tighten the
hose nipple. bottle in the support.

6.9 - Connecting the unit to the chair

2 8
F2

1
1

77

7 F3

F1 Air
Sewer

Step 54: Separate the wires and also hoses Step 55: Connect hose 1 to nipple 1,
identified with numbers and letters, connect hose 2 to nipple 2, connect hose 8
according to the above figure. to nipple 8.

77
7
Water
Air

F1
F2
Sewer

Water
F3

Step 56: Connect hose 7 to nipple 7, Step 57: Tighten the sewer Box and the
connect hose 77 to nipple 77, connect wire Black Water Nipple in the already defined
F1 to connector F1, connect wire F2 to places, then connect blue hose WATER to
connector F2 and wire F3 to connector F3. black nipple WATER setting it with clamp
K5 indicated in the kit of page 7, connect
hose SEWER to the SEWER input.

Installation manual - Dental Unit D700 14


6.10 - Connecting the reflector to the chair

Step 58: Perform the connection of the


reflector wires. There is no need for
polarization.

6.11 - Opening packing box “02”

Step 59: Remove the packing staples (indicated by the Step 60: Remove the items inside the packing in accordance
arrows), by using a screwdriver or any similar tool. with the order of mounting.

6.12 - Items included in packing “02”

Head rest filling Head rest filling Seat filling

Installation manual - Dental Unit D700 15


6.13 - Mounting the head rest

Step 61: Position the Cover above the Step 62: With the parts in position, Step 63: Mount the headrest on he
Filling, aligning the holes. Please refer to perform the setting using 4 nuts and 4 headrest structure. Please see above photo.
the above photo. Philips screws and a Philips screwdriver.

Note: The nuts and the screws are in the packing Kit.

6.14 - Mounting the head rest filling

Step 64: There are 6 points with Velcro in Step 65: Remove the protective film of the Step 66: Clean well the head rest plate.
the headrest filling and 6 points with Velcro, in two parts. Position the filling above the headrest
Velcro in the head rest plate. plate, being careful so that it is centralized
and perform pressure, as indicated, so that
there is adherence of one part to the other.

6.15 - Mounting the seat filling

Step 67: Position the seat filling above the Step 68: Perform the setting of the rear Step 69: Perform the setting of the front
seat plate, being careful so that the setting part using two 5/16” nuts and a 5/16” part using two 5/16” nuts and a 5/16”
screws fit in the holes (front and rear). fixed open wrench or barrel screwdriver. fixed open wrench or barrel screwdriver

Installation manual - Dental Unit D700 04


16
7 - Annexes

Annex 1 - Electrical connection layout


Annex 2 - Electronic layout
Annex 3 - Hydropneumatic layout
Annex 4 - Exploded drawings
Annex 5 - Exploded drawings
Annex 6 - Exploded drawings
Annex 7 - Exploded drawings

The annexes are in this CD, in the folder "Annexes".