Professional Documents
Culture Documents
CK Vsi Tds 11.2 Manual
CK Vsi Tds 11.2 Manual
Treatment Delivery
Manual
Version 11.2.x
1075883-ENG A |i
Treatment Delivery Manual
Manufacturer
Accuray Incorporated
1209 Deming Way
Madison, Wisconsin 53717 USA
Customer Support
For more information, to request documentation, or if you have a service issue, please contact
Accuray Customer Support (North America) at +1-866-368-4807, contact your Distributor, or visit
the For more information, to request documentation, or if you have a service issue, please contact
Accuray Customer Support (North America) at +1-866-368-4807, contact your distributor, or visit
the Services tab at www.accuray.com.
NOTE: If your facility works with a third-party service provider, please contact them directly for
your service-related issues.
Copyright Information
© 2001-2021 Accuray Incorporated. All rights reserved.
This document, software (© 2001-2021) and products to which this document refers, and any other
related materials are the copyrighted and proprietary information of Accuray Incorporated, with the
exception of open source software described below, and may not be used or distributed without
written authorization of Accuray Incorporated. No part of this document may be photocopied,
reproduced, or translated into another language without written permission from Accuray
Incorporated. TomoTherapy Incorporated is a wholly owned subsidiary of Accuray Incorporated.
Any references herein to Accuray Incorporated necessarily also include reference to TomoTherapy
Incorporated by definition.
Accuray Incorporated reserves the right to revise this publication and to make changes in content
from time to time without obligation on the part of Accuray Incorporated to provide notification of
such revision or change.
Accuray Incorporated provides this guide without warranty of any kind, either implied or expressed,
including, but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Accuray Incorporated and its directors, officers, representatives, subsidiaries,
employees, agents, heirs and assigns assume no responsibility or liability, either express or
implied, for injury, death, or losses to consumers, users or service personnel resulting from
improper handling of the Accuray products by unauthorized, untrained or otherwise unqualified
personnel. Accuray Incorporated expressly denies any responsibility or liability for abuse, neglect,
misuse or tampering with Accuray System components by persons not authorized, trained or
otherwise associated with Accuray Incorporated.
1075883-ENG A | iii
Accuray Incorporated Treatment Delivery Manual
Trademark Information
The stylized Accuray logo, CyberKnife, CyberKnife VSI, CyberKnife M6, CyberKnife S7,
TomoTherapy, H Series, Tomo, TomoH, TomoHD, TomoHDA, TomoEDGE, TomoHelical,
TomoDirect, Hi·Art, PreciseART, PreciseRTX, Radixact, Accuray Precision and iDMS, Iris,
Xchange, RoboCouch, InCise, MultiPlan, Xsight, Synchrony, Synchrony Fiducial Tracking,
Synchrony Skull Tracking, Synchrony Spine Tracking Supine, Synchrony Spine Tracking Prone,
Synchrony Spine Tracking Prone with Respiratory Modeling, Synchrony Lung Tracking with
Respiratory Modeling, Synchrony Fiducial Tracking with Respiratory Modeling, Synchrony Fiducial
Tracking with InTempo Imaging, TxView, PlanTouch, QuickPlan, ClearRT, CTrue, VoLO, Planned
Adaptive, TQA, TomoLink, TomoPortal, Accuray OIS Connect, and AERO Accuray Exchange in
Radiation Oncology are trademarks or registered trademarks of Accuray Incorporated, in the
United States and other countries and may not be used or distributed without written authorization
from Accuray Incorporated. Use of Accuray Incorporated trademarks requires written authorization
from Accuray Incorporated.Warranty Information.
Warranty Information
If any Accuray products are modified in any manner all warranties associated with such products
shall become null and void. Accuray Incorporated does not assume any responsibility or liability
with respect to unauthorized modification or substitution of subsystems or components.
With proper care and maintenance, the expected service life of the Accuray System is 10 years.
The Accuray System, including each computer workstation and associated system software, has
been validated to demonstrate that the system will perform as expected. The installation of
additional software not released by Accuray Incorporated (e.g. third party, off-the-shelf, etc.) on
these computer workstations is not permitted. This includes any operating system updates. Any
effect on the safe and intended operation of the Accuray System caused by the introduction of
additional software is unknown and Accuray cannot be responsible for any impact caused by
adding such software.
Device Disposal
When an Accuray product reaches the end of its useful life and your facility desires to remove the
device, contact Accuray Customer Support to decommission, uninstall, and appropriately dispose
of the components
iv | 1075883-ENG A
Treatment Delivery Manual
WARNING: Warning statements describe possible conditions that can result in serious
fatal injury to the patient or facility personnel. Each warning gives the possible condition
and how to avoid it.
CAUTION: Caution statements describe possible conditions that can affect system performance
or cause damage to system components. Each caution gives the possible condition and how to
avoid it.
CAUTION: Federal law restricts this device to sale by or on the order of a physician.
1075883-ENG A |v
Accuray Incorporated Treatment Delivery Manual
vi | 1075883-ENG A
Treatment Delivery Manual Table of Contents
Table of Contents
Chapter 2: Safety......................................................................2-1
Introduction ...................................................................................... 2-1
Emergency Procedures ................................................................... 2-3
Emergency Stops (E-Stop) ........................................................................ 2-3
The E-Stop Button.............................................................................................. 2-4
Resetting an E-Stop ........................................................................................... 2-5
Software E-Stop Failure ..................................................................................... 2-5
E-Stop Button Failure ......................................................................................... 2-5
Emergency Power Off (EPO) ..................................................................... 2-6
Power Failure .............................................................................................. 2-7
LINAC Malfunction...................................................................................... 2-8
Fire or Fumes .............................................................................................. 2-8
Treatment Couch Failure ........................................................................... 2-8
Unexpected Robot Arm Motion ................................................................. 2-9
1075883-ENG A |vii
Accuray Incorporated Treatment Delivery Manual
viii | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |ix
Accuray Incorporated Treatment Delivery Manual
Equipment Locations......................................................................3-17
Control Room ............................................................................................ 3-18
Treatment Room........................................................................................ 3-18
Equipment Room ...................................................................................... 3-19
x| 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xi
Accuray Incorporated Treatment Delivery Manual
Function Keys.................................................................................5-21
Transferring System Data for Remote Diagnostics ............................... 5-22
Obtaining Screen Captures...................................................................... 5-22
xii | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xiii
Accuray Incorporated Treatment Delivery Manual
xiv | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xv
Accuray Incorporated Treatment Delivery Manual
xvi | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xvii
Accuray Incorporated Treatment Delivery Manual
Interrupting Treatment....................................................................8-82
Hardware E-Stop Button........................................................................... 8-82
Interlock Button......................................................................................... 8-82
Pause.......................................................................................................... 8-83
xviii | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xix
Accuray Incorporated Treatment Delivery Manual
xx | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xxi
Accuray Incorporated Treatment Delivery Manual
xxii | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xxiii
Accuray Incorporated Treatment Delivery Manual
xxiv | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xxv
Accuray Incorporated Treatment Delivery Manual
xxvi | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xxvii
Accuray Incorporated Treatment Delivery Manual
References ...................................................................................15-36
xxviii | 1075883-ENG A
Treatment Delivery Manual Table of Contents
1075883-ENG A |xxix
Accuray Incorporated Treatment Delivery Manual
xxx | 1075883-ENG A
Chapter 1: Overview
This manual contains instructions for multiple features of the CyberKnife System. Since some
features of the CyberKnife System are optional, some of the instructions in this manual may not
apply to your system.
The availability of options is dependent on regulatory approvals in a particular country and varies
from country to country.
Intended Use
The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery
and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation
treatment is indicated. The CyberKnife Treatment Delivery System may be used to treat
astrocytoma, glioma, skull base tumors, metastases (brain and bony), nasopharyngeal carcinoma,
meningioma, acoustic neuroma, schwannoma, pituitary adenoma, hemangioblastoma,
craniopharyngioma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia,
and tumors of the neck, spine, pancreas, liver, lungs, ovary, prostate, and bladder. Patients should
be examined by a team of physicians to determine if they are candidates for CyberKnife treatment.
Clinical Benefit
At high doses, radiation therapy kills cancer cells or slows their growth by damaging their DNA.
Cancer cells whose DNA is damaged beyond repair stop dividing or die. When the damaged cells
die, they are broken down and removed by the body.
1075883-ENG A | 1-1
Accuray Incorporated Treatment Delivery Manual
Residual Risks
Below are the residual risks of the device, please reference the appropriate sections of the
documentation for further details:
• Exposure to infection from cross-contamination
• High voltage warnings in service procedures, service training, and on-component labeling
• Electromagnetic compatibility and susceptibility
• Adventitious, primary, leakage, and scatter radiation levels for treatment beam
• Adventitious, primary, leakage, and scatter radiation levels for kV Imaging
• kV imaging dose
• Electromagnetic immunity
• Treatment delivery dose related to user configuration of tracking algorithm threshold
parameters
• Disabling the application of respiratory motion modeling
• Possible side-effects of radiation treatment
• Mechanical hazard from drive train
• Laser light
• Patient information privacy breach
Incident Reporting
Any serious incident that occurs while using the Accuray CyberKnife System must be reported to
the manufacturer and the competent authority of the Member State in which the user and/or patient
is established. Contact Accuray Customer Support for more information.
1-2 | 1075883-ENG A
Treatment Delivery Manual Chapter 1: Overview
Performance Characteristics
General Specifications, Commissioning, Equipment, QA Performance, and other topics related to
performance characteristics of the device can be found in the Physics Essentials Guide.
Manuals
Accuray Incorporated provides a set of user manuals (instructions for use) and a reference manual
for the CyberKnife System. Manuals in the documentation set are described below.
WARNING: Example data depicted in this manual is not intended to represent realistic
clinical data. Use of example data for treatment planning or delivery could result in patient
mistreatment. The user is solely responsible for determining appropriate data values for
any given situation.
1075883-ENG A | 1-3
Accuray Incorporated Treatment Delivery Manual
WARNING: Regardless of the language translation, all numeric data that you enter or that
is displayed in the software uses the period character (.) as a decimal separator. Be aware
of this notation convention when interpreting or entering numeric data. Incorrect
interpretation or entry of numeric data could result in patient mistreatment.
1-4 | 1075883-ENG A
Treatment Delivery Manual Chapter 1: Overview
Computer Terminology
This manual uses standard personal computer terminology. Accuray Incorporated assumes you
know how to use a standard personal computer to navigate through windows and files on your
workstation. See the documentation that came with the computer. The following conventions are
used in this manual.
Double-click Press the left mouse two times in rapid succession. If you
need to use the right mouse button, the instructions specify
double right-click.
CTRL-click Hold down the CTRL key and press the left mouse button.
SHIFT-click Hold down the SHIFT key and press the left mouse button.
Hold Press the mouse button and hold it down while you perform
another function.
Drag Position the cursor over an area of interest, click, hold the
button down, and move the mouse to select an area, create a
window, or relocate a selected item.
Select Place the cursor over a name or button and click once, or
place the cursor at the beginning of the name, click, hold, and
drag across the name until it is highlighted (changes colors).
1075883-ENG A | 1-5
Accuray Incorporated Treatment Delivery Manual
1-6 | 1075883-ENG A
Chapter 2: Safety
Introduction
This chapter provides information about the emergency procedures and safety features of the
CyberKnife Treatment Delivery System and describes safe operating practices. Review the
information in this chapter before using the equipment It is the customer’s responsibility to
schedule quarterly and annual preventative maintenance by Accuray Service personnel or other
qualified personnel designated by Accuray.
Address any questions regarding the safe and efficient use of the CyberKnife System directly to
Accuray Customer Support.
This chapter covers the following topics:
• “Emergency Procedures” on page 2-3: This section describes what to do in emergency
situations.
• “Safety Precautions” on page 2-9: This section describes how to avoid hazards.
“Facility Responsibilities” on page 2-35: This section describes user responsibilities for the training
and safety of the staff and patients who come into contact with the CyberKnife System.
WARNING: The CyberKnife System can emit lethal doses of high energy radiation. The
assemblies contain high voltage circuits that can deliver lethal electrical shocks. Always
observe safety precautions when operating or working on or around the CyberKnife
System.
WARNING: Do not alter the software or any of the components in the CyberKnife System.
Results of these actions can result in injury or death.
Emergency Procedures
You are responsible for on-site procedures that follow the guidelines outlined here. The following
emergencies and procedures are described:
• Emergency Stop (E-Stop) procedure.
• The Emergency Stop button does not work.
• Emergency Power Off (EPO) procedure.
• The LINAC X-ray beam does not shut off.
• Power failure.
• Fire or fumes are detected.
• The treatment couch does not operate properly.
• The robot arm moves in an unexpected way.
Being prepared for an emergency is essential to operate this equipment safely.
The Error Handling System window is displayed, indicating the reason for the E-Stop and
suggesting options for resolving it.
Resetting an E-Stop
After an Emergency Stop, when the status indicates that the radiation is off, you can enter the
Treatment Room.
• If the E-Stop was not caused by pressing one of the standard treatment couch E-Stop
buttons, you can manually move the standard treatment couch using the treatment couch
Hand Controller.
• If the error is recoverable, you can correct the error, reset the error, and resume
treatment. If the error is unrecoverable, you cannot reset the error. In this case, a makeup
fraction will be created. For more information on resuming treatment after an E-Stop, see
“Chapter 8: Treatment Delivery”.
WARNING: Visually observe the CyberKnife System during operation at all times.
Manually initiate an E-Stop if an unsafe condition is imminent to prevent injury or death of
the user or patient.
To manually execute an E-Stop if the Error Handling System fails to stop operation:
1. Press the nearest physical E-Stop button.
2. Contact Accuray Customer Support.
The CyberKnife System has an Emergency Power Off (EPO) system that shuts off power to all
non-computer equipment (see “Emergency Power Off (EPO)” on page 2-6).
To manually shut off power if an E-Stop button fails:
• Press the nearest EPO button, or turn off the main circuit breaker in your facility.
NOTE: All users must know the location of the main circuit
breaker in their facility.
The E-Stop button is fail-safe and does not require any operator intervention.
NOTE: The appearance of the EPO button for your facility may
differ from the above figure.
An EPO is different from an E-Stop. When an EPO is executed, power to all non-computer
equipment is shut off. An E-Stop, however, stops movement and radiation, but does not shut off
power to the equipment. For more information about an E-Stop, see “Emergency Stops (E-Stop)”
on page 2-3.
WARNING: In case of fire or electric shock, press the nearest EPO button to immediately
shut off power to all non-computer equipment and minimize injury to the user or patient as
much as possible.
Power Failure
The CyberKnife System operates from 3-phase electrical power. If the power fails, all treatment
actions stop.
In the event of a partial loss of power, resulting in abnormally low voltage on one or more phases
of the input power, the system performs an Emergency Power Off. For more information, see
“Emergency Power Off (EPO)” on page 2-6.
To recover from a power failure:
Use the following emergency procedure to protect the patient.
1. If necessary, turn on emergency lighting.
2. Enter the Treatment Room and carefully help the patient off the treatment couch.
3. When power is restored, start the CyberKnife System using normal startup procedures,
as applicable to the specific situation. For information on startup procedures, see
“Chapter 4: System Startup and Shutdown”.
• The UPS automatically switches to battery operation if utility power fails and supplies
power to connected equipment until the battery is fully discharged.
• While on battery operation, an alarm beeps every 30 seconds. Press the TEST button to
silence the alarm.
LINAC Malfunction
If the dose in the Beam Data window (see “Chapter 5: Treatment Delivery System Overview”)
exceeds by 1% or 1 cGy the treatment dose at a node and an E-Stop was not automatic, the LINAC
has not functioned properly. You must turn off the LINAC.
To turn off the LINAC:
• Press the nearest E-Stop button.
Fire or Fumes
If you smell smoke or fumes, or if you detect a fire, initiate the emergency shutdown procedure
below.
To shut down immediately:
1. Press the nearest EPO button.
2. Activate the fire alarm in the facility.
3. Attend to the patient.
4. Evacuate everyone from the facility according to facility protocols.
5. If in treatment mode, verify that the system has executed the abort sequence. All the
treatment information is recorded automatically and the remaining undelivered dose will
be saved as a makeup treatment plan for use later.
6. If the smoke or fumes are coming from the computer equipment, manually turn off the
Uninterruptible Power Supply (UPS).
If your CyberKnife System includes the optional RoboCouch® Patient Positioning System, the
RoboCouch actuators use over travel switches and software limits to prevent a runaway condition.
For instructions on operating the standard treatment couch and the optional RoboCouch System,
see “Chapter 7: Treatment Room Operations”.
Safety Precautions
This section describes the hazards and safety precautions associated with the CyberKnife System.
If the equipment is not operated safely, the following hazards could result in injury or death for
patients and personnel.
• X-ray radiation
• Radio frequency radiation
• Electromagnetic interference
• Laser radiation with power up to 1 mW at 632 nm (Class II)
• Ozone
• Sulfur hexafluoride
• Implosion
• Explosion
• Electric shock
• Heavy equipment
• Hot surfaces
This section also describes the Patient Safety Zone, which is used to determine the boundaries
within which the patient must be positioned for safe treatment (see “The Patient Safety Zone” on
page 2-22).
In addition, instructions for performing a Contact Detection Sensor check are provided. This check
should be performed as part of daily Quality Assurance (QA) procedures (see “Contact Detection
Sensor Check” on page 2-32).
It is unsafe for anyone other than trained personnel to operate the CyberKnife System. Anyone
working with the system must be aware of the hazards listed here. Only trained personnel should
perform user functions.
WARNING: Do NOT perform any maintenance not specifically listed as user functions.
Otherwise, the CyberKnife System may malfunction and could cause the permanent injury
or death of the user or patient.
X-ray Radiation
WARNING: If the equipment is not operated properly, you can receive a harmful, even
lethal, dose of radiation in a short time without knowing it is being received. Radiation
exposure can damage some types of cardiac pacemakers. Ensure that the safety
precautions described below are followed.
WARNING: For ALL CyberKnife System installations, when the CyberKnife System is
controlled manually, the LINAC radiation beam can be pointed in almost any direction,
including at locations outside the primary shielding.
The CyberKnife System emits X-rays from both the LINAC for treatment and the X-ray imaging
system to determine the patient’s real-time location.
Protection from X-rays
The following safety precautions are required:
• Radiation hazard warnings must be posted in appropriate locations.
• Only qualified personnel may operate and service the equipment.
• Personnel radiation dosimeters should be worn when operating the CyberKnife System.
• When treatment or imaging is in progress, only the patient should be allowed inside the
room.
• Before entering the room, ensure the LINAC and X-ray imaging system are not emitting
radiation. Turn the high voltage interlock key switch to the OFF position. Verify that the
LINAC BEAM ON light and the HIGH VOLTAGE light on the operator control panel are
OFF. Verify that all other installed safety systems show that the Treatment Room is safe
to enter.
• Remove the safety interlock key from the operator control panel before leaving the
Control Room area to prevent unauthorized operation. Do NOT duplicate the safety
interlock key. Use a single key only to either enable the high voltage interlock on the
operator control panel, or to disable the Treatment Room door interlock.
• Monitor the Beam Data window during treatment and commence emergency shut-off
procedures if the system does not terminate when the dose at a node exceeds the
prescribed dose by 1% or 1 cGy.
• Pay attention to objects that may lie between the LINAC and the patient. Such objects
may attenuate and scatter radiation. Such objects may include shadow trays or blocks,
side rails, spinal attachments, various panels and filters. If possible, remove these objects
or make the necessary dose corrections.
CAUTION: The CyberKnife Treatment Delivery System produces high energy X-rays. This
energy can damage sensitive electronic devices, such as pacemakers, defibrillators and
other implanted or body-worn electronic medical devices.
Implement the following precautions when patients or personnel have implanted or body-worn
electronic medical devices:
• Prominently display warnings regarding potential damage to these devices.
• If a patient or staff member has such a device, contact the device manufacturer about
possible operational damage from radiation. The frequency of the microwaves used by
the LINAC in the CyberKnife System is approximately 9 GHz.
• Even if the manufacturer indicates the device is safe to operate, monitor its operation
according to the manufacturer’s recommendations. particularly if pulse generators are
near ongoing CyberKnife System treatments.
• If possible, inhibit the operation of the device during CyberKnife System treatments.
Essential Performance
The CyberKnife Treatment Delivery System delivers a prescribed radiation dose to a patient for the
treatment of tumors. The system's robotic arm can move to deliver dose to the patient from multiple
positions. During treatment, the unit compensates for normal physiological movements of the
patient to ensure the dose is accurately delivered. The system automatically places the CyberKnife
System in an interlock protected state when any system parameters are out of specification.
The CyberKnife System meets the following essential performance specifications:
1. The deviation between the prescribed dose value and actual measured value does not
exceed ±5%.
2. The system does not permit unanticipated radiation delivery. All radiation delivered is
system controlled.
3. The kV imaging device does not produce artifacts that interrupt treatment.
4. Key components of the robot, treatment table, and collimators do not have unanticipated
motion. All motion of key components is system controlled.
5. The Target Locating System locates, monitors, and tracks the patient’s position.
6. The system cannot treat when a key component fails or a fault error occurs.
Electromagnetic Interference
WARNING: Use of accessories and cables other than those specified or provided by
Accuray Incorporated may result in increased electromagnetic emissions or decreased
electromagnetic immunity that compromises the performance of Accuray equipment.
WARNING: Failure to use Accuray equipment in the specified shielded location may
compromise the performance of Accuray equipment and interfere with other equipment or
radio services.
This equipment has been tested and found to comply with the limits of the standard for medical
devices, IEC 60601-1-2. The limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates, uses and can radiate radio
frequency energy, and, if not installed and used in accordance with the manufacturer's instructions
may cause harmful interference to other devices in the vicinity. Portable and mobile RF
communications equipment can affect medical electrical equipment. There is no guarantee that
interference will not occur in a particular installation. If this equipment causes interference with
other devices, which may be determined by turning the equipment off and on, the user is
encouraged to try and correct the interference by one or more of the following measures:
• Reorient or relocate the device receiving the interference.
• Increase the separation between the equipment.
• For more information or assistance, contact Accuray Customer Support.
The low levels of electromagnetic radiation from the LINAC may cause electromagnetic
interference (EMI) to surrounding equipment and to patients and technical personnel who have
cardiac pacemakers. In addition, EMI from other nearby equipment, such as microwave
hyperthermia and diathermy equipment, may interfere with the dose counter in the CyberKnife
System. Shielding by the walls of the Treatment Room may lessen this risk.
The following precautions further protect the CyberKnife System from electromagnetic interference
from other equipment and protect other equipment from interference from CyberKnife System
components.
• Keep all doors, panels, and covers in place while you are operating the CyberKnife
System.
• Stop any treatment if you detect fluctuations in outputs, inputs, and power, or if you notice
any obvious erroneous readings. These fluctuations indicate possible EMI between
CyberKnife System components or from other nearby equipment.
Follow the Pacemaker precautions in “Implanted or Body-Worn Electronic Medical Devices ” on
page 2-11.
Table 1
The CyberKnife System is intended for use in the electromagnetic environment specified below. The customer
or the user of the system should assure that it is used in such an environment.
Table 1 (continued)
RF emissions Group 1 The CyberKnife systems uses RF energy only for its
CISPR 11 internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
Table 2
Electromagnetic environment -
IMMUNITY test IEC 60601 test level Compliance guidance
The CyberKnife systems are intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Table 2 (continued)
Electromagnetic environment -
IMMUNITY test IEC 60601 test level Compliance guidance
Table 2 (continued)
Electromagnetic environment -
IMMUNITY test IEC 60601 test level Compliance guidance
Voltage dips, short >95% dip in UT >95% dip in UT Mains power quality should be that
interruptions and for 5 seconds for 5 second of a typical hospital environment. If
voltage variations on the user of the CyberKnife systems
power supply input require continued operation during
lines power mains interruptions, it is
IEC 61000-4-11 recommended that the system be
powered from an uninterruptible
power supply or a battery.
The CyberKnife systems are intended for use in an electromagnetic environment in which radiated RF
disturbances are not controlled. The user of the CyberKnife systems can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the CyberKnife systems as recommended below, according to the maximum output power of
the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Laser Radiation
The LINAC uses laser radiation, narrow beams of bright red light. These lasers are in the LINAC
X-ray head. These lasers aid alignment of the treatment robot for LINAC calibration. Each laser
consists of two components, the plasma tube and the power supply. The tubes used have a
maximum of 1 mW output power, thus a Class 2 rating.
WARNING: The light from a laser beam can damage the retinas of the eyes. Never look
directly into the laser beam. Advise patients not to look into the laser beam. Otherwise,
blindness can result.
WARNING: Replacement of the laser tube must be done only by Accuray Field Service.
Do not attempt to replace the laser tube. Otherwise, exposure to high voltage is possible
that may result in severe injury or death. For more information, see “Electric Shock” on
page 2-19.
Sulfur Hexafluoride
Sulfur hexafluoride (SF6) is a colorless, odorless gas used in the CyberKnife System as an
electrical insulator. It is stored in a metal container under high pressure. If the metal container
ruptures, the gas or liquid will escape rapidly. Sulfur hexafluoride is nonflammable and is
considered to be nontoxic, but it can act as a simple asphyxiate by displacing air. Contact with the
liquefied gas may cause frostbite. SF6 gas is denser than air and accumulates on the floor if
released.
If the CyberKnife System detects a loss of SF6 gas pressure when high voltage is on, it will
generate an E-Stop.
WARNING: If the gas is inhaled or if contact with the liquefied gas occurs, follow
emergency response procedures as established in your facility’s hazardous substances
training program.
Implosion
The windows that separate the vacuum in the LINAC and magnetron from the waveguide can
shatter inwardly if they are struck or jolted excessively. Damage may occur to the equipment
because of the casing. Shattering glass should not harm the operator or patient.
The thyratron tubes in the LINAC modulator could also implode.
WARNING: Only trained service personnel should service these or any part of the
CyberKnife System. Otherwise, injury or death of the user can result. If an implosion
occurs, contact Accuray Customer Support immediately.
Explosion
The system must never be operated in the presence of flammable anesthetics or other flammable
or explosive substances. Electrical arcs that can occur during the normal operation of switches,
circuit breakers, push-buttons, and other circuit components can ignite vapors and gases.
WARNING: Only trained service personnel should service any part of the CyberKnife
System. Otherwise, injury or death of the user or patient can result. If an explosion occurs,
contact Accuray Customer Support immediately.
• If flammable substances are present before the system is turned on, do not turn it on.
• If flammable substances are detected after the system has been turned on, pause the
system. Do not turn the system off.
• If flammable substances are detected during a treatment, open the Treatment Room door
to cause an E-Stop. An E-Stop must be generated to evacuate the patient, but no other
operations should be performed.
If flammable substances are detected:
1. Evacuate all persons immediately.
2. Ventilate the room to clear the air of the flammable vapor or gas.
3. Remove any volatile liquids that are producing flammable vapors to a safe storage area.
Electric Shock
Some CyberKnife System components operate with voltages up to 40,000 volts. Be cautious at all
times during operation.
WARNING: As with any equipment that uses high voltage, the residual voltage in the
components can shock you after the equipment has been turned off, which may result in
severe injury or death. Only qualified service personnel should remove equipment covers.
• Discharge all high voltage storage components prior to servicing or disconnecting from
the supply.
Heavy Equipment
Many components of the CyberKnife System are heavy, particularly the treatment robot, the
standard treatment couch, and the optional RoboCouch Patient Positioning System.
The motors that control the treatment robot, the standard treatment couch, and the RoboCouch
System are very powerful. When these components move, they have substantial inertia. Observe
utmost care to prevent moving components from hitting or crushing you, the patient, other
personnel, or other equipment.
WARNING: Collision with the treatment robot, the RoboCouch System, or any other heavy
component can result in serious injury or death to users or the patient.
WARNING: During treatment, carefully watch while the treatment robot or RoboCouch
System is moving. If you see either component moving too close (within 2 inches or 5 cm)
to any object, press the nearest E-Stop button immediately to avoid serious injury or
death to the user or patient.
WARNING: When performing operations that involve movement of the treatment robot or
RoboCouch System, never stand within the robot workspace. Otherwise, serious injury or
death can occur.
WARNING: When the RoboCouch System is in motion and under control of either the
Hand Controller or the Teach Pendant, avoid all locations that could pose a potential
crushing hazard. Such locations include (see “Chapter 7: Treatment Room Operations”):
1. The region between the upper and lower arms of the RoboCouch manipulator.
2. The region between the upper arm and nearby walls.
3. The region between the RoboCouch table top and nearby walls.
4. The area beneath the RoboCouch table top.
5. The area between the RoboCouch table top and lower arm.
Pay attention to people and other objects in the room at all times to avoid serious injury or
death.
WARNING: Observe the treatment robot safe movement envelope when you are in the
Treatment Room and when adding objects to the Treatment Room. The safe movement
envelope is the set of configurations the treatment robot can assume without colliding
with objects defined at installation. These objects include the floor, ceiling, treatment
table/patient area, and camera stands (for the X-ray imaging detectors).
WARNING: Keep adequate clearance between you and the treatment couch during
treatment couch movements to avoid possible injury.
WARNING: If any equipment collides with the patient or another component, stop the
treatment, attend to the patient, record the accumulated dosage, and call Accuray
Customer Support to service the equipment before any further use.
WARNING: If the treatment robot collides with another object, if forced motion of the
treatment robot occurs, or if slippage of the LINAC head on the treatment robot faceplate
mount occurs, you must re-verify pointing accuracy of the LINAC. Use standard Quality
Assurance (QA) methodology, including Automatic QA (AQA) and/or End-to-End testing,
to verify targeting accuracy specifications. Failure to re-verify alignment can result in
patient mistreatment. For information on Physics QA procedures, see the Physics
Essentials Guide.
CAUTION: Use caution when components are moving. Collisions among components can
damage equipment.
CAUTION: Be careful not to drop the collimator on the X-ray detectors of the imaging system.
Otherwise, damage to the X-ray detectors can occur.
For more information on interchangeable secondary collimators, see “Chapter 7: Treatment Room
Operations”.
WARNING: Personal injury may occur if you stick your fingers or other body part into the
path of a moving IRIS aperture. Do not stick your fingers or other part body part into a
moving IRIS aperture.
Before treatment, the Patient Safety Zone is used to determine the boundaries for positioning the
patient for safe treatment. For instructions on checking the Patient Safety Zone before treatment,
see Chapter 7, “Checking the Patient Safety Zone".
For the dimensions of the fixed and dynamic zones used to define the Patient Safety Zone and
detect collision risk, see “Patient Safety Zone Dimensions (Standard Treatment Couch and
RoboCouch Flat Table Top Configuration)” on page 2-24.
For instructions on responding to PDP errors, see “PDP Errors From the Treatment Delivery
Computer” on page 2-29.
WARNING: The PDP and SPDP models are based on room geometry and CyberKnife
System components. Do not place temporary objects inside the treatment room, for
example, chairs, gurneys, or IV poles. Otherwise, collision with the treatment robot or
treatment couch and serious injury or death may occur.
WARNING: The PDP model is based on the Head First Supine (HFS) patient position.
Before treatment, ensure that the patient and all devices are within the Patient Safety Zone
defined by the PDP model, even if the patient is not in the HFS position. For information
on supported patient positions, see “Chapter 8: Treatment Delivery”.
WARNING: Carefully monitor motion of the treatment robot when using the treatment
robot Teach Pendant to avoid collisions. It is possible for the treatment robot to be moved
into the patient, overhead structures such as the X-ray sources, the optional Xchange
Robotic Collimator Changer, and other objects or walls in the Treatment Room, which can
result in serious injury or death, and damage to the equipment and surrounding facilities.
B=
A=
= Machine center
E=
D=
C=
= Machine center
When aligning the patient, position the patient so that the dynamic safety zone of the treatment
couch is within the fixed safety zone relative to the machine center.
The patient size sets the dimensions of the dynamic safety zone. You can select the following
patient sizes:
• Large: Based on size data for a 99 percentile male (Figure 5 on page 2-26).
• Medium: Based on size data for a 75 percentile male (Figure 6 on page 2-27).
• Small: Based on size data for a 50 percentile female (Figure 7 on page 2-27).
WARNING: Choose a size that is as large as or larger than the patient. Otherwise,
collision with the treatment robot is possible which can result in serious injury or death of
the patient.
WARNING: Position the patient and all devices within the Patient Safety Zone defined by
the Proximity Detection Program (PDP) model. Anything extending outside the boundaries
of the PDP model increases the risk of collision with the robot and can cause serious
injury or death of the patient.
WARNING: Do NOT place pillows or other objects in an area that could cause Patient
Safety Zone violations. Do NOT position patient extremities so that safety zone violations
occur, or else the patient may be seriously injured or killed.
For example, actions that could cause Patient Safety Zone violations include:
• Placing pillows under the patient's head.
• Positioning the patient with arms raised above head; this can cause elbows to exceed the
111 cm height limit or the 108 cm superior limit.
• Using too large a wedge under the patient's knees; this can elevate the knees above the
111 cm height limit.
• Treating lesions below L-4 Head First Supine (HFS).
WARNING: Neither the PDP nor the SPDP detect a collision risk when the treatment
couch is moved from the treatment delivery computer in treatment mode (Treatment,
Phantom, Demonstration, Plan QA, Simulation, and Physics>Collimator, and
Physics>Laser Alignment modes). Use extreme caution when moving the treatment
couch from the treatment delivery computer. Visually observe all movement of the
treatment couch to prevent potential collisions. Manually driving the treatment couch or
patient into other room obstacles using the Hand Controller or Teach Pendant can result
in serious injury or death of users or the patient.
POV 10%
HOV 20%
WARNING: It is possible to move the treatment robot closer to the collision risk and
potentially collide with it when you jog at slow speed using the Teach Pendant. Use
extreme caution when moving the treatment robot towards the collision risk. For
guidelines on what to do if a collision with a patient or room object occurs, triggering an
E-Stop, see "Error Handling" in “Chapter 8: Treatment Delivery”.
The SPDP allows only slow speed movement using the Hand Controller until there is no further
collision risk. You can then resume normal jogging speeds.
WARNING: It is possible to move the RoboCouch System closer to the collision risk and
potentially collide with it when you jog at slow speed using the Hand Controller. Use
extreme caution when moving the RoboCouch System towards the collision risk. Visually
observe all movement of the treatment couch to prevent potential collisions. Manually
driving the treatment couch or patient into other room obstacles using the Hand
Controller or Teach Pendant can result in serious injury or death of users or the patient.
To perform a Contact Detection Sensor check for fixed collimators (alternative check):
1. Squeeze firmly around different areas of the sensor (Figure 9 on page 2-33).
Slot
2. Verify that a clicking sound can be heard. The area opposite the slot is most sensitive
when squeezing the sensor for sound. It may be necessary to position your ear close to
the sensor (approximately 4 in or 10 cm from the sensor) to hear the clicking sound.
Hot Surfaces
Many components of the CyberKnife System operate at high temperatures. When equipment
covers are removed, hot surfaces are accessible.
WARNING: Do not remove equipment covers. Otherwise, you can be seriously burned
from touching hot surfaces that are exposed. Only qualified service personnel should
remove any equipment covers.
During operation of the CyberKnife System and during a reasonable cool-down period after turning
off the equipment, be sure not to touch the following components when they are hot:
• Thyratron tubes
• Any light bulbs
• Transformers
• Resistors
• Magnetron
• Laser tubes
• Robot motors
WARNING: Avoid any bodily contact with the above components or any other
components that seem warm when approached. Otherwise, you can be seriously burned
from touching hot surfaces. Be aware that the thyratron, magnetron, and LINAC filaments
remain on while the equipment is in the stand-by mode.
Facility Responsibilities
This section describes your responsibilities for setting up emergency and safety procedures for
CyberKnife System operations.
General Use
Use the following guidelines for routine use of the CyberKnife System:
• Keep a logbook of regular Quality Assurance (QA) and maintenance activities and also
record all unusual events related to the system.
• Always observe the standard startup and shutdown procedures, being certain to disable
the equipment by the documented procedures only.
• Test Emergency Stop procedures monthly.
Disinfection Instructions
The only parts of the CyberKnife System that should come in contact with the patient are the
carbon fiber treatment couch table top and the cushions provided with the treatment couch. These
surfaces are not intended to be sterile. They can be cleaned with normal cleaning solutions.
Safety Operations
You must provide and train your technical, service, and support staff in the following safety
operations.
• Fire extinguishers should be present in the Treatment Room and near the operator
control panel, per local fire code.
• The Treatment Room and operator control panel area must have emergency lighting. In
addition, flashlights must be on-site as a back-up during emergencies.
• Hazard signs regarding radiation exposure, high voltage, laser light hazards, and
pacemaker hazards must be posted where appropriate, near the Treatment Room, near
the operator control panel, and on the system components.
• Radiation monitoring badges should be worn at all times while in the Treatment Room.
• Dangers to people with pacemakers must be ascertained by following the guidelines
described earlier in this chapter.
• Video and audio monitoring are mandatory whenever a patient is in the Treatment Room.
This monitoring equipment is required in addition to the radiation monitoring equipment
incorporated in the CyberKnife System.
Radiation Overdose
The CyberKnife System uses dual redundant dose monitoring systems, which stop treatment upon
reaching the prescribed dose.
If the treatment is stopped manually because of a failure in the automatic shut-off by the dose
monitoring system, the equipment must be checked and serviced by a qualified technician before
any further treatments are performed.
In the case of a radiation overexposure:
1. Seek emergency treatment for the victim.
2. Contact the local radiation safety officer according to facility protocols.
3. Notify the local, state, and federal authorities in accordance with your institutional
procedures.
4. Consult with medical experts in radiation treatment.
5. Notify the Accuray radiation safety officer at RSO@accuray.com as quickly as possible.
Introduction
The CyberKnife Treatment Delivery System combines continuous image-guidance technology with
computer-controlled robotics to deliver a prescribed dose of radiation.
The CyberKnife System utilizes a compact linear accelerator attached to a computer-controlled
robotic arm, called the treatment robot, to target a number of beams of radiation at a lesion from
many different directions. The patented image-guidance technology correlates intra-operative X-
rays with previously acquired CT images of the lesion and surrounding tissue to target each beam
of radiation. This approach results in a high dose of radiation concentrated within and conforming
to the lesion, where the beams converge.
3
3
4
6
1
1. Treatment robot 2. Patient positioning system 3. Target Locating System (TLS) 4. Linear
accelerator (LINAC) 5. X-ray detector and floor panel 6. Synchrony Respiratory Camera Array
(option).
• “Target Locating System (TLS)” on page 3-3
• “Operator Control Systems” on page 3-6
• “Accuray Precision Treatment Planning System” on page 3-7
• “LINAC and Control Systems” on page 3-8
• “Treatment Robot” on page 3-10
• “Power Distribution Unit (PDU)” on page 3-11
• “E-Stop Control Chassis (ESCC)” on page 3-11
• “Uninterruptible Power Supply (UPS)” on page 3-11
• “Miscellaneous Equipment (Supplied by Customer)” on page 3-11
X-ray
Sources
Isopost
Calibration
Tool
Digital X-ray
Detectors Localizing
Frame
Two X-ray generators (not shown in Figure 1) supply power to each X-ray source. The maximum
symmetrical radiation field is approximately 24 x 17 cm at the treatment isocenter. Depending on
the room configuration, the distance between the source and the treatment center (isocrystal) is
2200 ± 100 mm, and the distance between the detector and treatment center is 1420 ± 5 mm.
The X-ray generators and X-ray generator control chassis may be located in either the Treatment
Room or the Equipment Room.
NOTE: The localizing frame is mounted either flush with the floor
(In-Floor Detectors) or slightly above the floor (On-Floor
Detectors) in the Treatment Room.
NOTE: The isopost calibration tool is used only for calibrating the
CyberKnife System and is then removed.
Detector Windows
The 2 detector windows are made of carbon fiber material to ensure X-ray translucency and high
strength. They are designed to support up to a 500 lb (227 kg) load and protect the detectors.
Protective material covers the detector windows. The protective window coverings can be replaced
if they become scratched or damaged.
CAUTION: Do not drop any heavy or sharp objects onto the detector windows. If a heavy object
is dropped on the detector windows, call Accuray to assess the structural damage.
LINAC
The X-band LINAC provides a collimated beam of 6 MV X-rays. The LINAC consists of an electron
gun and a series of microwave cavities under vacuum. Its compact size allows robotic manipulation
with 6 degrees of freedom.
The high efficiency design of the LINAC provides a dose rate of 800 monitor units (MU) per minute.
X-ray Head
The X-ray head contains the LINAC, magnetron, microwave waveguide components, the pulse
transformer, water-circulating connections and a gas pressurizing connection for the waveguide.
The X-ray head is attached to the treatment robot.
Secondary Collimators
12 interchangeable fixed collimators with apertures (in millimeters) of 5, 7.5, 10, 12.5, 15, 20, 25,
30, 35, 40, 50, 60 allow flexibility and conformability during the treatment planning process. Pinhole
and blank (no aperture) collimators are also included. The fixed collimators attach to a secondary
collimator housing which is mounted on the X-ray head.
The optional Iris™ Variable Aperture Collimator provides the same 12 apertures as the fixed
collimators. The aperture of the Iris Collimator is adjustable under computer control. For
information, see “Iris Variable Aperture Collimator (Option)” on page 3-16.
Modulator (MOD)
The Modulator (MOD) contains high voltage pulse-generating circuits and provides a pulse to the
magnetron so the magnetron can produce radio frequency (RF) energy. The MOD is located in the
Equipment Room.
Chiller
The chiller is a closed-loop system that cools the LINAC. It consists of a water reservoir, a
recirculating pump, and a refrigeration unit. The chiller includes temperature-regulating control to
maintain a stable operating temperature, a water temperature indicator, and status lamps. The
chiller does not require external facility connections. The chiller is located in the Equipment Room.
Treatment Robot
The treatment robot is a 6-axis robot that carries the LINAC. The treatment robot provides
automated positioning of the LINAC for patient treatment with 6 degrees of freedom.
Manual operation of the treatment robot is provided by the treatment robot Teach Pendant, located
in the Treatment Room.
A treatment robot controller controls positioning of the treatment robot and is located in the
Equipment Room.
The Lung 1 View with Respiratory tracking is used in cases in which the tumor is more clearly
visible or better identified in one X-ray projection than the other. The Spine Supine tracking mode
is used in cases in which the tumor is not clearly visible in either X-ray projection. For more
information on Lung Optimized Treatment, see “Chapter 15: Lung Optimized Treatment (Option)”.
The maximum load supported by the standard treatment couch is 350 lb (159 kg). Patient weight
includes any pads, restraints, and devices that are loaded on the table top.
The standard treatment couch includes the following components:
• Standard treatment couch table top
• Standard treatment couch head base plate
• Hand Controller and readout display for manual operation of the treatment couch
The Iris Collimator nominally provides the same 12 apertures as the fixed collimators (in
millimeters): 5, 7.5, 10, 12.5, 15, 20, 25, 30, 35, 40, 50, 60. It contains 2 stacked hexagonal banks
of tungsten segments that together produce a 12-sided aperture (a regular dodecagon).
The Iris Collimator mounts on the X-ray head and requires the Xchange Robotic Collimator
Changer (see “Xchange Robotic Collimator Changer (Option Model B)” on page 3-17).
Equipment Locations
The location of components of the CyberKnife System are listed below.
Control Room
The Control Room for the CyberKnife System contains the following equipment:
• Operator control panel
• Monitors, keyboard, and mouse for the treatment delivery computer
• Quad Multiplex Monitor of CCTV system
• Intercom System
• Phone
• Physics conduit port
Treatment Room
The Treatment Room of the CyberKnife System contains the following equipment:
• LINAC
• Treatment robot
• Standard treatment couch or optional RoboCouch System
• Target Locating System (TLS). X-ray generators and X-ray generator control chassis may
be located in the Treatment Room or the Equipment Room
• Xchange table with fixed collimators and, if installed, the Iris Collimator
• Respiratory Camera Array and Respiratory interface module (if the Respiratory Modeling
System is installed)
• CCTV cameras
• Intercom System
• Phone
• Medical gas lines, if desired
Equipment Room
The Equipment Room is typically located adjacent to the shielded walls of the Treatment Room
and is intended to hold the bulk of the support equipment needed for treatment. This equipment
includes the facility power interface. The room is cooled with a customer-supplied A/C unit.
The Equipment Room contains the following equipment:
• Treatment delivery computer
• CyberKnife System data server
• Modulator (MOD)
• Modulator Control Chassis (MCC)
• Chiller in mechanical rack
• Target Locating System Control Chassis (TLSCC)
• Target Locating System (TLS) Computer
• X-ray generators and X-ray generator control chassis (if not located in the Treatment
Room)
• Interface Control Chassis (IFCC)
• Treatment robot controller
• RoboCouch controller (if the RoboCouch System is installed)
• Power Distribution Unit (PDU)
• E-Stop Control Chassis (ESCC)
• Uninterruptible Power Supply (UPS)
• Monitor, keyboard and mouse
• Firewall
• Compressor (with optional Iris Collimator)
• Automatic Temperature Control (ATC) unit (with optional Iris Collimator)
For more information on components in the Equipment Room, see the Physics Essentials Guide.
Introduction
This chapter describes procedures for taking the CyberKnife Treatment Delivery System from a
shutdown state to a powered-on condition where it is ready to deliver a treatment or be operated
in Physics mode. It also describes procedures for shutting down the CyberKnife System.
This chapter also provides equipment instructions including using the Collimator controls in
Physics mode, the Target Locating System (TLS) components, and the treatment robot Teach
Pendant.
It includes the following topics:
• “Pre-startup Checklist” on page 4-2
• “Powering On the CyberKnife System” on page 4-6
• “Error Handling” on page 4-10
• “Post-startup Checklist” on page 4-12
• “Logging into the CyberKnife System” on page 4-13
• “X-ray System Warmup” on page 4-14
• “Powering Off Equipment” on page 4-18
• “System Shutdown Checklist” on page 4-19
• “Power-on After Power Interruption” on page 4-19
• “Equipment Operating Instructions” on page 4-22
Pre-startup Checklist
It is recommended that each site develop a checklist that includes, at a minimum, the tasks below.
As each task is performed, you check it off. When all tasks are completed, initial the completed
checklist and file it with the treatment records.
1. Check that the treatment robot Teach Pendant is in the proper mode.
• Check that the Teach Pendant is in External mode.
• The display of the treatment robot Teach Pendant should be dark indicating that the
treatment robot is powered down.
2. Check the pressure (30 ± 2 psi) (2.1 Kg/cm) of SF6 gas (on the pressure gauge in the
mechanical rack in the Equipment Room) and recharge if needed.
3. Verify that the key switch on the E-Stop Control Chassis (ESCC) is not turned to
SERVICE MODE (see Figure 1).
WARNING: Do not deliver treatment using SERVICE MODE. SERVICE MODE bypasses
some safety features. Using SERVICE MODE during treatment delivery can produce
hazardous conditions that can result in injury or death to the user or patient.
The user should only use SERVICE MODE when performing an E-Stop circuit check after
a power interruption (see “Step 2: Hardware Power On” on page 4-21). Otherwise, only
Accuray Field Service should use LOCAL MODE.
4. Check the Uninterruptible Power Supply (UPS) battery charge level and verify that the
Battery disconnected/Replace battery indicator is not lit (see “Uninterruptible Power
Supply (UPS)” on page 4-35) If the battery charge indicators show less than full charge,
wait for the UPS to fully recharge before powering up the system.
Steering Magnetron
Power Control Tuning Controls Dose Rate Fault and Status
Coil Current Meter
Lock Meters and Position Indicators
Meter
7. Turn on the Treatment Room Closed Circuit TV (CCTV) cameras and intercom system.
8. Turn on the monitors for the treatment delivery computer.
You are now ready to power up the CyberKnife System using the treatment delivery system
software on the treatment delivery computer, as described in the next section.
For instructions on powering on the CyberKnife System after a power interruption, see “Power-on
After Power Interruption” on page 4-19.
To power up CyberKnife System equipment remotely:
1. Initially, the Equipment Status screen shows that the electrical power for each equipment
subsystem is OFF and the subsystem control status is Not Ready (see Figure 4).
Power Scheduler
Use the Power Scheduler to schedule automatic power-on dates for the treatment delivery system.
You can schedule automatic power-on for days of the week, select specific days to always include
or exclude, and can change or cancel the time for the next scheduled power-on. You can also set
the number of hours of inactivity after an automatic power-on before the treatment delivery system
is automatically powered off to ensure the system is not left on overnight. When an automatic
power-on is complete, a user must click OK in the message that indicates the time remaining
before the system is automatically powered off.
To open the Power Scheduler dialog box:
• On the Cyberknife System Menu screen, click Computer, and then click Power
Schedule. The Power Schedule dialog box appears.
• To cancel power-on for the next scheduled day, click to clear the Power on at check
box. Or, to activate power-on for the next scheduled day, select the Power on at check
box.
• To change the currently scheduled power-on hour, minute, or PM/AM setting, type or
select values in the appropriate boxes.
Date exceptions
The Date Exceptions area displays a calendar with the following status indicators:
• Cyan highlight box for the selected date
• Red font for today’s date
• Gray-blue highlight box for scheduled dates
• A red square for "never try to power on"
• A green square for "always try to power on"
To modify scheduling for specific dates, double click the calendar, and then do any of the following:
• To move to the month of interest, click the right or left arrows in the calendar title bar,
• To return to today’s date, click the Return to today’s date button .
• To change a date to "never try to power on," double click it. Double click it again to
change it to “always try to power on." Double click it again to clear the date’s exception
status.
Error Handling
Power-on Errors
If an error in a subsystem is encountered during the power-on sequence, a dialog box is displayed
reporting the error. When you click the OK button in the dialog box, a full power-down sequence
begins.
Post-startup Checklist
WARNING: When the CyberKnife System power-up procedure is complete, perform each
item on the post-startup checklist. Correct any discrepancy before proceeding. If
necessary, power off the equipment as described in “Powering Off Equipment” on
page 4-18. Then make the necessary corrections and restart the power-up procedure.
Failure to perform these procedures can result in damage to equipment and increased
risk of injury or death of the patient.
It is recommended that each site develop a checklist that should include, at minimum, the following
tasks:
• Check that on the ESCC no mechanical interlock LED is illuminated. If illuminated, safety
features are disabled and should be reset.
• Check that none of the bypass buttons on the ESCC are illuminated.
• Check the chiller water temperature (19 ± 1° C) on the LINAC chiller.
• Verify that the treatment robot is positioned at its perch position. Move the treatment robot
back to the perch position using the Teach Pendant as necessary. (For information on
moving the treatment robot to the perch position, see “Treatment Robot Operation” on
page 4-27.)
• Verify door interlocks and E-Stops are enabled. To do this:
Open the Treatment Room door and verify that the E-Stop status icon on the
treatment delivery computer shows that an E-Stop occurred.
Press the E-Stop button on the operator control panel and verify that the E-Stop
status icon on the treatment delivery computer shows that an E-Stop occurred.
After each daily task is performed, it should be checked off, initialed, and dated by the user.
2. Enter your username in the Username textbox. Then press the <Tab> key or move the
cursor to the next textbox.
3. Enter your password in the Password textbox.
4. Click the OK button or press the <Enter> key. Or, press the Cancel button to return to
the CyberKnife System Menu.
• When your username and password are authenticated successfully, press the
Continue button to proceed to the task or delivery mode.
• If you enter an incorrect username or password, a message asks you to try again.
Press the Continue button to return to the primary user login window and repeat
the login procedure.
As the primary user, you can add participants to your CyberKnife session. For more information on
adding participants, see “Chapter 5: Treatment Delivery System Overview”.
WARNING: During the X-ray tube warmup procedure, X-rays are produced. Make sure no
one is in the Treatment Room during this procedure to prevent unnecessary exposures.
CAUTION: Failure to perform the X-ray tube warmup procedure could damage the X-ray tubes
of the X-ray imaging system.
2. Click the X-Ray Warmup button. The X-ray Tube Warmup screen appears (see
Figure 11). The text located in the X-ray Tube Warmup section of the screen
displays the date and time that the most recent X-ray tube warmup was performed.
3. Click the Start button. The text IN PROGRESS is displayed below the yellow progress
bar, and the upper left display box flashes red, displaying the text WARMUP ON. As
images are acquired, the X-ray settings for the X-ray image detectors (Camera A and
Camera B) are displayed in the 6 text fields located in the X-Ray Tube Warmup
section (see Figure 12).
When X-ray tube warmup is complete, the upper left display box changes from red to
green again and displays the text WARMUP OFF. The progress bar displays 100%,
and the text below it displays COMPLETED. The Last Successful Warmup date
and time display is updated.
4. Click the Exit button to exit the X-ray Tube Warmup screen.
NOTE: If the X-ray imaging system has not been used for 8 or
more hours, it is recommended that you perform the X-ray tube
warmup procedure again.
If the Interlock button is pressed while X-ray tube warmup is in progress, the warmup
procedure is aborted and the Error Handling System window is displayed indicating that
the warmup procedure was interrupted (see Figure 13). Clicking the Continue button in
the Error Handling System window resets the system and allows you to restart the X-ray
tube warmup procedure by clicking the Start button in the X-Ray Tube Warmup
section.
After performing the X-ray tube warmup procedure, you are ready to perform the LINAC warmup
procedure. For more information, see the Physics Essentials Guide.
3. Click Yes to power off equipment. Or, click No to return to the Equipment Status screen
without powering off equipment.
• If you click Yes, the Power Off button on the Equipment Status screen becomes
inactive. After a few seconds equipment status changes to the powered-off state.
When power is off, the Power On button becomes active.
If you click No, the Power Off button in the Equipment Status screen remains active.
CAUTION: In the event of a power interruption, shutdown all computers from their desktop
menus. Do not turn off or reset power to the treatment delivery computer or other computers
using the power button on the computer. Doing so may cause the computer to become corrupted
and unable to reboot.
Controls in the Collimator screen are described in Table 1 on page 4-23. This screen allows you
to perform the following tasks outside of a treatment plan:
• Exchange fixed collimators.
• Change from a fixed collimator to the optional Iris Variable Aperture Collimator.
• Change the fixed aperture size or the aperture size of the Iris Collimator.
• Initialize or reset the Iris Collimator.
For more information on the Xchange System, see “Chapter 7: Treatment Room Operations”.
Chapter 6 describes how to use the Xchange System to exchange collimators automatically within
a treatment plan, in Demonstration, Plan QA, Phantom, or Treatment mode.
Control Function
Current Activity text field Displays current activity of the treatment robot and the Xchange
System, as well as the elapsed time.
Status section Displays information about the currently installed collimator and
the treatment robot position.
Collimator Type text field: Displays the currently installed
collimator type: Fixed or Iris.
Collimator Aperture text field: Displays the aperture (in
millimeters) of the currently installed collimator.
Robot Position text field: Indicates whether the treatment
robot is at or away from the perch position.
Control Function
Collimator Selection Allows you to switch between the housing for fixed collimators
section and the Iris Collimator.
Pick Up Iris button: Instructs the treatment robot to pick up the
Iris Collimator. If a fixed collimator is currently installed, the
treatment robot first drops off the fixed collimator housing (with
fixed collimator attached) in its receptacle on the Xchange table.
Pick Up Fixed button: Instructs the treatment robot to pick up
the fixed collimator housing (with previous fixed collimator
attached). If the Iris Collimator is currently installed, the treatment
robot first drops it off in its receptacle on the Xchange table.
Collimator Aperture Allows you to specify the desired collimator aperture (fixed or Iris).
Selection section
Desired Aperture drop-down list: Specifies the desired
aperture (in millimeters). In addition, allows you to select the
following options:
• Full Open: Fully opens the aperture (to approximately
68 mm) to allow attachment of the Iris pinhole collimator. For
more information, see the Physics Essentials Guide.
• Full Close: Fully closes the aperture (to just above 0 mm). If
the aperture is fully open and the Iris pinhole collimator has
just been attached, this option closes the aperture. The Iris
Collimator stalls when it contacts the pinhole collimator, thus
holding it securely. For more information, see the Physics
Essentials Guide.
Apply button: Changes the collimator aperture. If a fixed
collimator is currently installed, the treatment robot exchanges it
for the desired collimator. If the Iris Collimator is currently
installed, the Iris aperture is automatically adjusted.
Emergency Stop
Button
Emergency Stop
Indicator Lamp
Pressing the high voltage OFF button on the operator control panel turns off the high voltage to the
LINAC. When radiation is being generated, the BEAM ON light on the operator control panel is on.
When no radiation is being generated, this light will be off.
WARNING: If you suspect a problem, depress the E-Stop button to terminate radiation
delivery and cease treatment robot motion.
You may need to turn off power to the imaging detectors for maintenance or to recover from an
imaging detector device fault. If the imaging detectors have not been powered on for 4 or more
hours, they should be turned on and allowed to warm up for at least 30 minutes before resuming
system operation.
Turning on the TLS computer starts the TLS system software. After the TLS computer has been
booted and before patient treatment, the TLS computer should remain at the Windows Login
screen to ensure that only the TLS system software is operating.
CAUTION: Using TLS software tools or other programs can interfere with the performance and
proper functioning of the Target Locating System. For this reason, the TLS software tools should
only be run by Accuray personnel for maintenance or calibration.
Reboot the system if you experience problems with the TLS system software such as:
• TLS communication failures that are reported on the treatment delivery computer.
• Imaging hardware fails to power on successfully (as indicated on the Equipment Status
window on the treatment delivery computer).
• Other TLS-related errors.
Controller Power-up
The main disconnect switch is on the treatment robot controller cabinet. When this handle is in the
Up position, the switch is closed. If power is supplied to the cabinet, the controller will start. It takes
2 to 3 minutes for the treatment robot to boot up.
To power off the controller, put the handle in the Down position. It takes about 30 to 60 seconds
for the controller to shut down. Wait at least 60 seconds after the treatment robot controller has
been powered off before powering it up again.
WARNING: Maintain at least a 3 ft (1 m) distance from the treatment robot when operating
the treatment robot manually. When moving the treatment robot, focus on the treatment
robot rather than looking at the Teach Pendant to avoid collision. Collision with the
treatment robot can result in serious injury or death of the user or patient.
WARNING: Do not remove the treatment robot Teach Pendant from the Treatment Room
during treatment delivery. The Teach Pendant should only be used in the Treatment Room
in order to safely monitor operation of the treatment robot.
Mode E-Stop
Escape Selector
Start
Toggle
Program
Stop
Program
Start
Enter
Enable Switch
Enable Switch
Space Mouse
Enable Switch
Program Start
The enable switches on the Teach Pendant (also called KCP) are three-position switches (see
Figure 18). If the switches are released, the drives are disabled. If they are held too tightly, the
drives are also disabled. The switches should be held in the middle position (depressed halfway)
for the drives to be enabled. If the treatment robot Teach Pendant does not respond, check to see
if the E-Stop button is depressed halfway.
The mode selector switch is used to select the operating mode (see Table 2).
No Jogging allowed
Jogging can be performed in one of the coordinate frames shown in Table 4. Pressing the soft key
on the left side of the Teach Pendant screen changes the frame. All coordinate frames, except the
first one (Individual axes), are Cartesian coordinate systems.
Individual axes
World coordinate
frame
4. Change the speed of the treatment robot by pressing the button for the soft key located at
the lower right (second button from the bottom) of the Teach Pendant screen.
WARNING: Select 50% or less for the speed of treatment robot travel to provide enough
time to avoid collision.
5. Depress and hold one of the enable switches halfway to enable the treatment robot
drives. On the status indicator located at the bottom left of the Teach Pendant screen, the
middle indicator changes from red
to green.
When the enable switch is depressed halfway, soft keys along the right side of the Teach
Pendant screen will change to +/- corresponding to various possible motions of each of
the 6 robot joints of the treatment robot.
6. Move the treatment robot as desired using the +/- buttons for each individual joint.
WARNING: Keep your eyes on the treatment robot at all times, not the Teach Pendant,
when robot joints are in motion, to avoid collision.
WARNING: Keep your eyes on the treatment robot at all times, not the Teach Pendant,
when using the prch program to avoid collision.
CAUTION: If your system includes the optional Xchange System and the treatment robot stops
when it is in contact with or close to the Xchange collimator table, do not use the prch program
to move the treatment robot away from the Xchange collimator table. To avoid possible collision
with the Xchange table, use the MoveUp program on the Teach Pendant to move the treatment
robot safely above the table. See “Using the MoveUp Program” on page 4-33.
To move the treatment robot to the perch position using the Teach Pendant:
1. Move the mode selector switch from External mode to manual teach mode T2.
2. Press the button for the Ackn. or Ackn. All soft key located at the bottom right corner
of the Teach Pendant screen.
3. Press the blue Toggle button to highlight the top of the Teach Pendant screen.
4. Use the arrow buttons at the bottom right of the Teach Pendant to select the CKMAIN
program folder in the left panel of the Teach Pendant screen. The selected folder will be
highlighted in blue.
5. Press the Enter button (yellow button) to open the folder.
CAUTION: Do not press the Enter button (yellow button) when a program is running. Damage to
CyberKnife System program files could result.
6.
7. Use the arrow buttons at the bottom right of the Teach Pendant to highlight the prch
program in the right panel of the Teach Pendant screen.
8. Launch the prch program by pressing the button for the Select soft key displayed at the
bottom of the Teach Pendant screen.
10. Press the green Program Start button once and then release the button when the
message "programmed path reached (BCO)" is displayed in the message window in
the bottom panel of the Teach Pendant screen.
11. Press and hold the green Program Start button a second time. The prch program checks
the current position of the treatment robot to see if it can safely be moved to the perch
position.
If the treatment robot can safely be moved to the perch position, it begins moving. All
three status indicators will be green.
Keep the green Program Start button depressed until the treatment robot returns to the
perch position, as indicated by the black indicator.
12. Reset the mode selector switch to External mode by moving the switch from manual
teach mode T2 to External mode.
CAUTION: Do not use the prch program to move the treatment robot away from the Xchange
collimator table. To avoid possible collision with the Xchange table, use the MoveUp program
on the Teach Pendant to move the treatment robot safely above the table.
2. Press the button for the Ackn. or Ackn. All soft key located at the bottom right corner
of the Teach Pendant screen.
3. Press the blue Toggle button to highlight the top of the Teach Pendant screen.
4. Use the arrow buttons at the bottom right of the Teach Pendant to select the CKtool
program folder in the left panel of the Teach Pendant screen. The selected folder will be
highlighted in blue.
5. Press the Enter button (yellow button) to open the folder.
CAUTION: Do not press the Enter button (yellow button) when a program is running. Damage to
CyberKnife System program files could result.
6. Use the arrow buttons at the bottom right of the Teach Pendant to highlight the MoveUp
program in the right panel of the Teach Pendant screen.
7. Launch the MoveUp program by pressing the button for the Select soft key displayed
at the bottom of the Teach Pendant screen.
to green.
9. Press the green Program Start button once and then release the button when the
message "programmed path reached (BCO)" is displayed in the message window in
the bottom panel of the Teach Pendant screen.
10. Press and hold the green Program Start button a second time.
Keep the button depressed until the treatment robot moves vertically up and comes to a
stop, as indicated by the black indicator.
When the treatment robot has come to a stop, you can use the prch program to move to
the perch position as needed (see “Moving to the Perch Position” on page 4-32).
11. Reset the mode selector switch to External mode by moving the switch from manual
teach mode T2 to External mode.
2. Press the button for the Monitor soft key at the top of the Teach Pendant screen.
3. Use the arrow buttons at the bottom right of the Teach Pendant to select the I/O menu
item.
4. Press the Enter button (yellow button).
CAUTION: Do not press the Enter button (yellow button) when a program is running. Damage to
CyberKnife System program files could result.
5. Use the arrow buttons at the bottom right of the Teach Pendant to select the Digital
Outputs menu item. Ensure that Output number 1 is selected.
6. Press the Enter button (yellow button).
CAUTION: Do not press the Enter button (yellow button) when a program is running. Damage to
CyberKnife System program files could result.
7.
8. Depress and hold one of the enable switches halfway until the middle status indicator
changes from red
to green.
CAUTION: Do not press the Enter button (yellow button) when a program is running. Damage to
CyberKnife System program files could result.
10. Wait a few seconds to verify that the LINAC laser has turned on.
11. Reset the mode selector switch to External mode by moving the switch from manual
teach mode T2 to External mode.
Repeat the procedure above to turn the LINAC laser off.
The following table lists the icons displayed on the UPS and describes the function of each.
Overload The connected loads are drawing more than the UPS power
rating.
Introduction
The Treatment Delivery System allows you to align the patient and deliver treatment. This chapter
gives an overview of the Treatment Delivery System user interface, starting from the CyberKnife
System Menu. This chapter covers the following topics:
• “CyberKnife System Menu” on page 5-2
• “Error Handling System” on page 5-10
• “Function Keys” on page 5-21
• “Equipment Power Status” on page 5-23
• “Physics Mode” on page 5-30
• “Treatment Mode” on page 5-39
• “Demonstration Mode” on page 5-60
• “Phantom Mode” on page 5-64
• “Plan QA Mode” on page 5-65
To log into the Treatment Delivery System, follow the instructions in “Chapter 4: System Startup
and Shutdown”
To perform data management and system administration tasks, use the iDMS Data Management
System.The iDMS Data Management System allows you to perform user, patient, treatment plan,
and DICOM administration tasks for the CyberKnife System. It also includes a DRR generation
server for generating Digitally Reconstructed Radiographs (DRR images) and applications for
Image Review and Import, Patient Record Archive and Restore, and Beam Data Import. For
information on using the iDMS Data Management System, see the Data Management Manual.
The CyberKnife System Menu includes the following user interface controls:
• “CyberKnife System Menu Buttons” on page 5-3
• “Computer Menu” on page 5-7
• “Help Menu” on page 5-9
Button Function
Physics Enters Physics mode, where you perform the X-ray tube warmup
procedure, operate the LINAC manually and view current LINAC
calibration parameters, operate the Xchange Robotic Collimator
Changer outside of a treatment plan, and perform a Laser Alignment
check.
You must log in to access this mode. For more information, see
“Physics Mode” on page 5-30
Enters Demonstration mode, where you can demonstrate
Demonstration
operation of the CyberKnife System by performing a BB test. In this
mode, no treatment radiation is delivered. X-ray images can be taken
before path traversal begins, but not during the demonstration. The
treatment robot is operated manually using the Teach Pendant.
Controls are the same as those in Treatment mode, except for the
following:
Bypass X-ray imaging: You have the option of bypassing the X-ray
imaging system, allowing you to start and run the demonstration
without acquiring Live X-ray images.
Bypass collimator control: You have the option of bypassing some
controls and checks of the Secondary Collimator System. This option
allows you to operate the CyberKnife System with any secondary
collimator, regardless of the collimator type and aperture specified in
the treatment plan.
You must log in to access this mode. For more information, see
“Demonstration Mode” on page 5-60.
Plan QA Enters Plan QA mode, which provides a method to perform quality
assurance (QA) for patient-specific treatment plans. In this mode, you
can deliver a treatment plan for a specific patient to a phantom.
Button Function
Treatment Enters Treatment mode, where you load patient data, align the
patient, start treatment delivery, and monitor treatment.
You must log in to access this mode. For more information, see
“Treatment Mode” on page 5-39.
For instructions on logging into the CyberKnife System, see “Chapter 4: System Startup and
Shutdown”.
3. Enter your username in the Username textbox. Then press the <Tab> key or move the
cursor to the next textbox.
4. Enter your password in the Password textbox. Then click the OK button or press the
<Enter> key. When your username and password are authenticated successfully, the
Add Participants window is displayed (see Figure 3).
5. Press the Add Participants button. The Additional Authenticated Participants window
is displayed (see Figure 4).
6. To add a participant, click the Add button. The Participant Login window is displayed.
7. Ask the participant to log in. They should enter their username in the Participant
Name textbox and their password in the Password textbox, and then press the OK
button.
When the participant is authenticated successfully, their username is displayed in the
Username list (see Figure 4). Continue adding participants as needed.
8. To remove a participant from the list, select a name to highlight it. Then click the
Remove button. You are asked to confirm the removal. Click the OK button to remove
the name from the list. Or, click the Cancel button to return to the previous screen
without removing the name.
9. When you are finished adding or removing participants, click the Done button to proceed
to the mode. Or, click the Exit button to log out all users and return to the CyberKnife
System Menu.
6. Enter your new password in the New Password textbox. Then press the <Tab> key.
7. Enter your new password in the Repeat Password textbox to confirm it. Then click the
OK button.
Or, click the Cancel button to return to the Primary User Login window without changing
your password.
8. When your new password is accepted, the screen for the mode you selected is displayed.
You must use the new password the next time you log in.
Computer Menu
The Computer menu on the CyberKnife System Menu screen (see Figure 5) allows you to
perform the tasks described in Table 2.
Service Login This menu item is for Accuray Service use only. It allows qualified Service
personnel to log into service applications.
Restart Restarts the Treatment Delivery System software, then returns you to the
Software CyberKnife System Menu.
Reboot Reboots the treatment delivery computer. All computer processes are shut
Computer down and restarted. Then you are returned to the CyberKnife System Menu.
Shutdown Shuts down the treatment delivery computer for computer maintenance.
Computer
Shut Down Use to remotely shut down the treatment delivery system computers, the
Remote iDMS System computer, and the storage vault that are preset by Accuray
Computers service personnel according to your system configuration. This menu option
does not shut down the treatment delivery computer. To shut down the
treatment delivery computer, use the Shutdown Computer menu
option. For more information on the system computers that comprise the
CyberKnife network environment, see the Physics Essentials Guide.
3. To restart the computer, press the computer power-on switch on the computer front
panel. The treatment delivery computer is located in the Equipment Room.
Help Menu
The Help menu displays information about the CyberKnife System, including the software version
and installed options.
Treatment Treatment robot is idle and will not move unless the user
Robot initiates an action.
Treatment Treatment couch is idle and will not move unless the user
Couch initiates an action.
X-ray X-ray sources are idle and will not generate X-rays unless
Sources image acquisition is initiated by the user.
LINAC LINAC is idle and will not deliver treatment radiation unless
the user initiates an action.
E-Stop/Interlock Controls
There are three E-Stop/Interlock controls displayed at the bottom right corner of the left screen: an
Error Information button, an E-Stop status icon, and an Interlock button. Table 4 describes
these controls.
The Error Handling System window has 3 main sections: a table with information about the error
event, a section that displays the complete text for the error and its resolution, and a section that
displays icons representing the overall status of E-Stops and the E-Stop status of individual
components, including the treatment robot, the treatment couch, LINAC, treatment room door,
operator console, wall E-Stop switches and the software E-Stop state. Table 5 describes the Error
Handling System window controls:
Table 5 Error Handling System window controls
Control Description
Show All (radio When selected, shows up to 1000 acknowledged event entries for
button) every twenty-four hour period. There is no limit to the number of
unacknowledged events displayed.
Control Description
(Table) The left column displays an icon representing the type of error
event.
Indicates a notification that requires user action.
Reason/ Displays the same text that may be truncated in the error table
Resolution (section) and lists the reason the selected error occurred and the options
for resolving it.
Control Description
Status of Individual The icons used to indicate the status of individual devices have
Devices and Software essentially the same meaning as the icons used for overall status.
Interlocks (icons)
A green checkmark indicates that all E-Stops that
require user action (such as a button release, key
turn, or door close) are cleared.
Control Description
Continue (button) Signals to the Error Handling System that you have
acknowledged any listed errors. Depending on the error, user
action may be required to resolve the error in addition to
acknowledging the error.
Exit Treatment Also called Exit Application. Ends all current task
activities.
Control Description
Abort (button) Displayed in Physics mode and when loading patient data, in
which cases it ends all current task activities.
Function Keys
Function keys (keyboard shortcuts) on the keyboard allow you to perform the tasks described in
Table 6. The function keys are the top row of keys on the keyboard.
CAUTION: Do not press the function keys ALT+F5 through ALT+F8 on the keyboard while
treating a patient. The normal treatment process may be interrupted by the data collection
process.
ALT+F5 Transfer Logs Invokes a log file transfer utility to transfer a core subset of
CyberKnife System log files for the current day to the
Accuray FTP site. The subset includes the log files that are
most likely to reveal the system problem.
ALT+F6 Transfer All Logs Invokes a log file transfer utility to transfer all CyberKnife
System log files for the current day to the Accuray FTP site.
ALT+F8 Transfer Active Invokes a log file transfer utility to transfer current patient
Patient Record fraction files to the Accuray FTP site. These files contain all
information needed for an Accuray Service Engineer to
reproduce the current imaging environment, including the
patient DRR images and the last pair of Live X-ray images
captured. This function is enabled only after a patient
treatment plan has been loaded in the Alignment,
Readiness, or Delivery phases.
ALT+F11 Save Logs Creates a compressed package of current log files and
saves it in the following location: /var/tmp/
log<timestamp>.tar.gz.
F12 Save Screenshot Takes a screenshot of the current screen and saves it on the
to File treatment delivery computer in the following location:
/accuray/printscreen. Does not send the screenshot
to the printer.
Print Screen Print Screen Takes a screenshot of the current screen and saves it on the
treatment delivery computer in the following location:
/accuray/PrintScreen, in a sub-directory labeled with
the date. Simultaneously sends the screenshot to the
printer.
CAUTION: Do not press the function keys ALT+F5 through ALT+F8 on the keyboard while
treating a patient. The normal treatment process may be interrupted by the data collection
process.
An Accuray Service Engineer or Applications Support Specialist might also be given permission to
remotely log into your system. When this occurs, you will see a message on the upper right corner
of your left monitor (see Figure 8) that a remote user has connected. It will be displayed for
approximately 10 seconds and will appear on any screen that you are viewing when the login
occurs.
The Equipment Status screen shows the status of the following subsystems (see Figure 10):
• Manipulator: Treatment robot (also called the treatment manipulator)
• E-Stop: E-Stop Control Chassis (ESCC)
• LINAC: Linear accelerator
• Power Schedule: If scheduled, the date and time of the next scheduled automatic
power-on is displayed to the right of the Equipment Status panel.
Button Function
Power On Turns on electrical power for all equipment subsystems of the CyberKnife
System.
Power Off After you confirm your intent to shutdown power, turns off electrical power to
all equipment subsystems.
Power to all equipment subsystems is generally controlled remotely via the Treatment Delivery
System software. When power is controlled remotely, you turn on equipment power by clicking the
Power On button in the Equipment Status screen, as described in “Chapter 4: System Startup
and Shutdown”.
The Manipulator, ESCC, LINAC, Secondary Collimator System, and RoboCouch System may also
be powered locally via the Power Distribution Unit (PDU), or both locally and remotely via the PDU
and software. Depending on how these equipment subsystems are powered on, the Equipment
Status screen indicates one of the following power states:
• Local power only: When the subsystems listed above are powered locally via the PDU
but not powered on via the software, the Power indicator color is green and displays
ON for the Manipulator, E-Stop, LINAC, Collimator, and Couch subsystems. In addition,
the text Local Power is displayed under these subsystem names in the Equipment
Status screen.
• Remote power only: The Power indicator color is green and displays ON. The
Control Status indicator is green.
• Local and remote power: The Power indicator color is green and displays ON. In
addition, the text Local & Remote Power is displayed under the Manipulator, E-Stop,
LINAC, Collimator, and Couch subsystem names in the Equipment Status screen. The
Control Status indicator is green.
For more information on powering up and powering down the CyberKnife System, see “Chapter 4:
System Startup and Shutdown”.
During the power-up sequence, indicators in the Equipment Status screen display the power and
control states described in Table 8. Control Status indicates the status of processes needed by
the treatment application.
OFF Not Ready Indicator colors are red. The default state when an
equipment subsystem is powered off.
OFF/FAULT FAULT/Not Ready Indicator colors are red. Indicates that a fault
occurred while attempting to power up an
equipment subsystem, and the power up sequence
was not successful.
Not Available Not Available (Respiratory only) Indicator colors are gray.
Indicates that the Synchrony tracking System is not
enabled (licensed) in the current configuration of the
CyberKnife System.
Xchange Disabled Indicator colors are gray. When the Xchange System is installed,
indicates that the Xchange table has lost power or is disconnected
from the network.
For more information, see “Chapter 7: Treatment Room Operations”.
Calibrate Xchange Indicator colors are orange. When the Xchange System is installed,
indicates that the quick-check sensor on the Xchange table is unable
to confirm alignment. If this occurs, the treatment robot returns to the
perch position.
For more information, see “Chapter 7: Treatment Room Operations”.
Iris Disabled Indicator colors are gray. When the Iris Collimator is installed,
indicates that the collimator is disabled.
For more information, see “Chapter 7: Treatment Room Operations”.
Calibrate Iris Indicator colors are orange. When the Iris Collimator is installed,
indicates that the Iris Collimator needs to be calibrated.
For more information, see “Chapter 7: Treatment Room Operations”.
Physics Mode
The Physics screen in Physics mode allows Physics Quality Assurance (QA) and Calibration of
the CyberKnife System (see Figure 13). Depending on your user profile, you may not have access
to all buttons on the Physics screen.
To display this screen, click the Physics button on the CyberKnife System Menu and then log in.
The Physics button is enabled only when equipment power is on.
Button Function
LINAC Allows you to operate the LINAC manually for LINAC warmup, dosimetry
verification, and dose calibration.
For information on performing LINAC tasks, see the Physics Essentials
Guide.
Button Function
X-Ray Warmup Allows you to perform the X-ray tube warmup procedure for the imaging
X-ray sources.
For information on this procedure, see “Chapter 4: System Startup and
Shutdown”.
Collimator Allows you to exchange collimators, including the fixed collimator housing
and the Iris Variable Aperture Collimator. Controls allow you to adjust the
aperture of the Iris Collimator, outside of a treatment plan. Controls also
allow initialization of the Iris Collimator. This button is enabled only if your
configuration includes the Xchange System.
For information on collimator tasks in Physics mode, see “Chapter 4:
System Startup and Shutdown”.
Laser Alignment Allows you to perform a quick verification of robot pointing accuracy. In the
event that the laser intensity changes over time, this application allows you
to reset the baseline. It gives you options to perform the check, display the
current and calibrated values for laser intensity and position, and adjust the
baseline for intensity values when the current position offset is within
tolerance (~0.3 mm at 800 mm SAD).
For information on laser alignment tasks in Physics mode, see C“Chapter
4: System Startup and Shutdown”.
LINAC Menu
You use the LINAC menu to manually operate the LINAC for warmup, dosimetry verification, and
calibration. When you click the LINAC button, the LINAC menu is displayed as shown in
Figure 14.
Button Function
Warmup Screen
Clicking the Warmup button in the LINAC menu displays the Warmup window shown in
Figure 15. The Warmup window allows you to enable the high voltage to the LINAC and specify a
desired dose to be delivered to warm up the LINAC. This window contains the controls described
inTable 12.
Control Function
LINAC Status section Allows you to enable the high voltage to the LINAC.
HV WAIT/HV ON/HV OFF indicator: Displays the status of the
LINAC high voltage. When the high voltage is enabled but not
turned on, the red HV WAIT indicator is displayed. When the
high voltage is turned on, the yellow HV ON indicator is displayed.
When the high voltage is disabled and turned off, the green HV
OFF indicator is displayed.
BEAM ON/BEAM OFF indicator: Displays the status of the
LINAC X-ray beam. When the yellow BEAM ON indicator is
displayed, dose is being delivered.
Enable button: This button is active when the high voltage is
turned off. Clicking this button enables the high voltage.
Disable button: This button is active when the high voltage is
turned on. Clicking this button turns off the high voltage.
Beam Control section Allows you to specify a time interval for beam delivery to warm up
the LINAC.
Desired Time textbox: Allows you to enter the desired warmup
time with a value between 1 second and 10 hours.
Start button: Allows you to start beam delivery to warm up the
LINAC.
Status bar: Displays the percentage of the specified warmup time
(0 to 100%) that has elapsed.
Delivered MU: Displays the delivered dose (in monitor units)
measured for dose channels A and B.
Dose Rate (MU/min): Displays the dose rate in MU/min for
dose channels A and B. Note that during warmup calibration
factors are not applied to the dose rate.
E-STOP button Triggers an Emergency Stop (E-Stop). The high voltage is turned
off when an E-Stop is triggered.
The Calibration Check screen contains the controls described in Table 13, in addition to the
controls displayed on the Warmup screen.
Control Function
Beam Control section Automatically displays the delivered dose and dose rate for dose
channel A and B.
Delivered MU fields: Display the delivered dose (in monitor
units) measured for dose channel A and B.
Dose Rate (MU/min) fields: Display the dose rate in MU/min
for dose channel A and B.
Last Calibration display Displays the date and time of the last LINAC dosimetry calibration
area procedure.
WARNING: Do not adjust the LINAC calibration unless you are the qualified Medical
Physicist. Incorrect dose delivery can result which could cause serious injury or death of
the patient.
The Calibration Adjustment screen contains the controls described in Table 14.
Control Function
LINAC Status section Allows you to enable the high voltage to the LINAC.
HV WAIT/HV ON/HV OFF indicator: Displays the status of the
LINAC high voltage. When the high voltage is enabled but not
turned on, HV WAIT is displayed. When the high voltage is
turned on, HV ON is displayed. When the high voltage is
disabled and turned off, HV OFF is displayed.
BEAM ON/BEAM OFF indicator: Displays the status of the
LINAC X-ray beam. When the red BEAM ON indicator is
displayed, dose is being delivered.
Enable button: This button is active when the high voltage is
turned off. Clicking this button enables the high voltage.
Disable button: This button is active when the high voltage is
turned on. Clicking this button turns off the high voltage.
Control Function
Calibration section Displays nominal dose values to be delivered for each calibration
data point, and allows you to enter the corresponding measured
dose value.
Nominal MUs fields: Display 5 preset nominal dose values (in
monitor units) to be delivered.
Measured (cGy/Gy) text boxes: Allows you to enter the
measured dose (in centigray or gray depending on the iDMS
System setting) corresponding to each nominal dose.
Start/Reset button: Allows you to start dose delivery for one of
the nominal doses, or reset a data point already obtained.
Treatment Mode
This section provides additional details on the user interface screens in Treatment mode that
are not described in later chapters. It covers the following topics:
• “Load Patient Data Screen” on page 5-39
• “Overview Phase” on page 5-43
• “Left and Right Screens of the Delivery Phase” on page 5-46
For information on the workflow for patient alignment and treatment delivery, and for details on the
user interface workflow for Synchrony Skull Tracking, see “Chapter 8: Treatment Delivery”. For
details on the user interface workflow for other tracking modes, see also the following chapters:
• “Chapter 9: Synchrony Fiducial Tracking”
• “Chapter 10: Synchrony Spine Tracking (Option)”
• “Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)”
• “Chapter 12: Synchrony Respiratory Tracking System (Option)”
• “Chapter 13: Synchrony Fiducial Tracking with InTempo Imaging (Option)”
• “Chapter 15: Lung Optimized Treatment (Option)”
The Load Patient Data screen displays a list of patients with treatment plans that are approved for
delivery and allows you to select the patient to be treated. A list of approved treatment plans for
the patient is then displayed, allowing you to select a treatment plan to be delivered. Then a list of
fractions is displayed. Treatment plans are not listed until they have been approved for delivery and
DRR images have been generated on the iDMS Data Management System.
As you select the patient, treatment plan, and fraction, additional patient and treatment information
is displayed. A text message at the bottom left of the screen reports status as data is retrieved from
the database.
The screen includes the checkbox Show only OIS scheduled patients and plans (if licensed) and
the Load and Refresh buttons listed in Table 15.
Button Function
Show only OIS Select if you want to display only OIS scheduled patients and plans. Clear if
scheduled you want to display all treatable patients and plans.
patients and
plans (licensed
option)
Load After you have selected a patient, plan, and fraction, begins downloading
patient and treatment data from the iDMS Data Management System to the
treatment delivery computer.
Refresh Refreshes the Patient and Plan lists with the latest data from the iDMS
Data Management System. New patients and treatment plans are added to
the list only after the treatment plans have been approved for delivery and
DRR images have been generated on the iDMS Data Management System.
Click this button if a treatment plan has been approved for delivery but is not
yet displayed in the list.
The Patient section of the Load Patient Data screen displays the following information:
• Last Name: Last name of the patient.
• First Name: First name of the patient.
• Medical ID: Medical ID of the patient.
• Date of Birth: Date of birth of the patient
• Gender: Gender of the patient
The Plan section displays the following information:
• Name: Name of the treatment plan.
• Synchrony Method: Tracking mode specified in the treatment plan.
• Respiratory: The status of Synchrony Respiratory Tracking (yes if Synchrony
Respiratory Tracking is enabled in the treatment plan, or no if Synchrony Respiratory
Tracking is not enabled). For more information on the Synchrony tracking System, see
“Chapter 12: Synchrony Respiratory Tracking System (Option)”.
• InTempo: The status of InTempo imaging (yes if the InTempo imaging is enabled in
the treatment plan, or no if InTempo imaging is not enabled). For more information on the
Synchrony Fiducial Tracking with InTempo Imaging, see “Chapter 13: Synchrony Fiducial
Tracking with InTempo Imaging (Option)”.
• Anatomy: Anatomy to be treated.
• Orientation: Patient position as specified in the treatment plan.
• Fractions: The total number of fractions in the treatment plan.
• Delivered: The number of full fractions already delivered for the treatment plan.
• Remaining: The number of fractions that remain to be delivered for the treatment plan.
• Delivered / Prescribed (MU): Cumulative radiation amount already delivered for
the treatment plan (in monitor units) / Total prescribed radiation amount for the treatment
plan (in monitor units).
• Percentage progress bar: A bar graph displaying the percentage of the total prescribed
radiation amount for the treatment plan that has already been delivered.
The Fraction section displays the following information:
• Number: The selected fraction number to be delivered (an integer value).
If a full fraction was not completed, a makeup fraction is created that includes the
remaining radiation amount needed to complete the fraction. The makeup fraction has the
same number as the original fraction, with the text makeup displayed next to it.
• Delivered / Prescribed (MU): Cumulative radiation amount already delivered for
the fraction (in monitor units) / Total prescribed radiation amount for the fraction (in
monitor units).
If the radiation amount delivered is zero, the fraction is a full fraction. The number of full
fractions multiplied by the prescribed radiation amount for each fraction equals the total
prescribed radiation amount for the treatment plan.
If the radiation amount already delivered is non-zero, the fraction is a makeup fraction.
• Percentage progress bar: A bar graph displaying the percentage of the total prescribed
radiation amount for the fraction that has already been delivered. If the percentage is non-
zero, the fraction is a makeup fraction.
The Notifications section displays the following information:
• Message panel: Notification message about the treatment or QA plan that is displayed
if an OIS is licensed and there is an inconsistency between OIS information and the iDMS
Data Management System.
• Acknowledge checkbox: Indicates that you have read and acknowledged the
notification message. The checkbox must be selected before treatment can proceed.
Overview Phase
The left screen of the Overview phase allows you to verify that the patient is in fact the one to
be treated and the treatment plan is the correct one to use (see Figure 19). If a photo of the patient
is available, it is displayed at the left of the screen.
WARNING: Carefully verify that this patient is in fact the one to be treated and the
treatment plan is the correct one to use. Otherwise, patient mistreatment can result.
WARNING: When the safety interlock key is used to bypass the Treatment Room door
interlock and the Treatment Room door is open, potentially harmful LINAC radiation
exposure to operators can occur outside the Treatment Room. Ensure that no LINAC
radiation is delivered while the Treatment Room door interlock is bypassed and the
Treatment Room door is open.
Control Function
Data Verified Indicate that you have verified patient information, treatment plan
checkboxes information, and then treatment fraction information by selecting each
checkbox in turn. Each checkbox is enabled only after you have selected the
previous checkbox.
The patient information section of this screen displays the following information:
• First Name: First name of the patient.
• Last Name: Last name of the patient.
• Gender: Gender of the patient.
• Medical ID: Medical ID of the patient.
• Date of Birth: Date of birth of the patient.
The treatment plan information section displays the following information:
• Plan Name: Name of the treatment plan to be delivered.
• Treatment Site: Anatomy to be treated.
• Orientation: Patient position specified in the treatment plan.
• Path Set: Name of the path set.
• Synchrony Method: Tracking mode specified in the treatment plan.
• Respiratory: The status of Respiratory Tracking (Yes if Synchrony Respiratory
Tracking is enabled in the treatment plan, or No if Synchrony Respiratory Tracking is not
enabled). For more information on the Synchrony tracking System, see “Chapter 12:
Synchrony Respiratory Tracking System (Option)”.
• InTempo: The status of InTempo imaging (yes if the InTempo imaging is enabled in
the treatment plan, or no if InTempo imaging is not enabled). For more information on the
Synchrony Fiducial Tracking with InTempo Imaging, see “Chapter 13: Synchrony Fiducial
Tracking with InTempo Imaging (Option)”.
• Total Prescribed MU: Total prescribed radiation amount for the treatment plan (in
monitor units).
• Total Delivered MU: Cumulative radiation amount delivered for the treatment plan.
• Total Prescribed Dose (cGy/Gy): Total prescribed radiation dose for the
treatment plan (in centigray or gray depending on iDMS System setting).
• Total Fractions: Total number of fractions for the treatment plan.
• Plan Collimator: The type of collimator specified in the treatment plan (Fixed for
fixed collimators, or Iris for the optional Iris Collimator, followed by the aperture (in
millimeters) for fixed collimators.
The treatment fraction information section displays the following information:
• Fraction: The number of the current fraction to be delivered.
• Prescribed MU: Prescribed radiation amount per fraction (in monitor units).
Delivered MU: The radiation amount already delivered for the current treatment fraction (in
monitor units). The top of the right screen of the Overview phase summarizes current patient
and treatment data (see Figure 20).
Start Button
The Start button allows delivery of a treatment fraction or makeup fraction (see Figure 23).
Depending on the status of patient alignment, the Start button may be disabled. For more
information, see “Chapter 8: Treatment Delivery”.
When treatment is interrupted, the Start button is replaced by the Resume button. Controls are
displayed for acquiring new Live X-ray images to confirm patient alignment or to perform patient
realignment. Controls allow you to respond to treatment interruptions such as:
• Treatment is interrupted at a node that is blocked. A blocked node occurs when the
treatment robot or some other obstruction blocks the X-ray imaging system and prevents
the successful acquisition of new images.
• Treatment is interrupted when the treatment robot is at a beam position. A beam position
is a treatment robot position where the LINAC is located at a node and is pointed at the
treatment target. The beam position includes the offset of the LINAC required for
targeting corrections.
• Treatment is interrupted and treatment robot motion stops between nodes or between a
node position and a beam position. A node position is a treatment robot position where
the LINAC is located at a node and is pointed at the machine center.
Resume Button
The Resume button allows you to resume automated treatment delivery after a Soft Stop, Pause,
or recoverable E-Stop occurs. (see Figure 24).
When you click the Resume button, the Treatment Delivery System checks many factors to
determine if new Live X-ray images are required. If new images are not required, the treatment
robot moves automatically to the next node in the treatment path and treatment proceeds.
If new images are required, they are acquired automatically and image correlation is performed
prior to delivery of the next beam. If the treatment robot is not at an unblocked node position, it
moves to one for image acquisition. The treatment robot then moves to the next node in the
treatment path and treatment proceeds.
NOTE: The treatment robot may move forward (in the forward
direction of the treatment path) or backward to reach the closest
unblocked node for imaging. The closest unblocked node may
be outside the treatment path.
If it is determined by visual inspection that the new images are blocked, you can mark the images
as blocked. See “Mark Image as Blocked Button” on page 5-52.
Depending on the status of patient realignment, the Resume button may be disabled. For more
information, see “Chapter 8: Treatment Delivery”.
Pause Button
Clicking the Pause button requests a Soft Stop, which is a non-emergency treatment interruption
(see Figure 25).
A Soft Stop pauses treatment without triggering an E-Stop/interlock. Imaging errors such as an
image algorithm error or a patient out-of-bounds error also trigger a Soft Stop.
When you request a Soft Stop during treatment delivery, the following actions are performed:
• The treatment robot completes any in-progress motion, such as movement between node
positions or between node and beam positions.
• The X-ray imaging system completes any in progress image acquisition.
• If beam delivery is in progress, delivery of the current beam is completed.
• The LINAC high voltage remains ON for 180 seconds.
• When the Soft Stop is actually engaged, the elapsed time displayed in the Beam Data
window stops incrementing until treatment is resumed.
For more information on resuming treatment after a Soft Stop, see “Chapter 8: Treatment Delivery”.
Acquire Button
When you click the Acquire button in the Delivery Phase, the X-ray imaging system acquires a
pair of Live X-ray images. The Treatment Delivery System correlates the Live X-ray images with
the DRR images to determine target offsets.
The Acquire button is inactive when the treatment robot is stopped at a known blocked node. In
this case, text at the bottom of the left screen of the Delivery phase indicates that the current
position is blocked (see “Current Robot Position Indicator” on page 5-51). The Move Robot to
Imaging Node button then becomes active.
In the above case, you can click the Move Robot to Imaging Node button to automatically
move the treatment robot to the closest unblocked node for imaging, click the Acquire button to
obtain new images, and then click the Resume button to resume treatment.
NOTE: You can click the Acquire button to take images when
the treatment robot is stopped and is not at a node position.
Automatic image acquisition, however, only occurs at node
positions.
For information on acquiring images in other phases such as Couch, Technique, and Alignment,
see “Chapter 5: Treatment Delivery System Overview”.
For the remainder of the treatment plan, text at the bottom of the left screen of the
Delivery phase will indicate that the marked node is blocked (see “Current Robot
Position Indicator” on page 5-51).
NOTE: The treatment robot may move forward (in the forward
direction of the treatment path) or backward to reach the closest
unblocked node for imaging. The closest unblocked node may
be outside the treatment path.
The Move to Imaging Node button is active only when the treatment robot is not stopped at
an unblocked node position.
When the treatment robot reaches the imaging position, the Move to Imaging Node button
becomes inactive and the Acquire button becomes active, allowing you to acquire a pair of
images manually.
The Treatment Monitoring graphs available depend on the tracking mode specified in the treatment
plan. Three Treatment Monitoring graphs can be displayed at a time. All graphs are plotted as a
function of time and display 300 seconds of data at a time. Each graph includes a legend.
The graphs includes the following controls:
• To select a set of graphs for display, click a tab.
• To view a legend for the graphs, click the Toggle Legend button at the bottom left of the
graph. To hide the legend, click the Toggle Legend button again.
• To view or hide individual data graphs on a tab, click the graph legend for that data graph.
• To view coordinate values for a data point on a graph, click the data point. The coordinate
values for that data point are displayed.
• To pan the horizontal display, click and drag the mouse over the graph.
• To expand or reduce the time scale, rotate the mouse scroll wheel over the graph.
• To pause the graph, click the Pause icon.
• To resume monitoring in real-time, click the Play icon. The graph jumps to the current
time.
The available treatment graphs are described in the sections that follow.
Orientation Graph
Displays the aggregate and individual rotational displacements as a function of time (see
Figure 27):
• Aggregate (deg): Aggregate rotational displacement. The sum of the absolute values of
the rotational displacements as a function of time. This value is updated each time an X-
ray image is acquired.
• Roll (deg), Pitch (deg), Yaw (deg): Three lines, each corresponding to an individual
rotational displacement (Right/Left roll, Head Up/Head Down pitch, or CCW/CW yaw) as
a function of time.
Position Graph
Displays the distance between the actual and nominal alignment position as a function of time over
the course of the treatment path (see Figure 44). This graph indicates how much the patient is
moving over time as a function of distance from the nominal aligned position:
• Position Delta (mm): Delta translational displacement. The square root of the sum of the
squares of the three displacements along the translational axes as a function of time. This
value is updated each time an X-ray image is acquired.
• Inf/Sup (mm), Left/Right (mm), Ant/Pos (mm) translational displacements: Three
lines, each corresponding to a translational displacement in the Left/Right, Inferior/
Superior, and Anterior/Posterior directions.
Skull Graph
(Synchrony Skull Tracking mode only) Displays the value of image algorithm parameters for
Synchrony Skull Tracking mode. For more information, see “Chapter 8: Treatment Delivery”.
Fiducial Graph
(Synchrony Fiducial tracking mode only) Displays the value of algorithm parameters for Synchrony
Fiducial Tracking mode as a function of time. For more information, see “Chapter 9: Synchrony
Fiducial Tracking”.
Spine Graph
(Synchrony Spine Tracking) Displays the value of algorithm parameters for Synchrony Spine
Tracking as a function of time. For more information, see “Chapter 10: Synchrony Spine Tracking
(Option)”.
Lung Graph
(Synchrony Lung Tracking with Respiratory Modeling and Lung 1 View with Respiratory Tracking
only) Displays the value of image algorithm parameters for Synchrony Lung Tracking with
Respiratory Modeling. For more information, see “Chapter 11: Synchrony Lung Tracking with
Respiratory Modeling (Options)” and “Chapter 15: Lung Optimized Treatment (Option)”.
NOTE: The Max Image Age parameter is not the same as the
Image Interval parameter. The Max Image Age parameter
applies only to treatment plans that use the Synchrony Fiducial
Tracking with InTempo Imaging.
For more information on the Synchrony Fiducial Tracking with InTempo Imaging, see “Chapter 13:
Synchrony Fiducial Tracking with InTempo Imaging (Option)”.
Images Counter
The Images counter displays a count of the acquired and expected X-ray images for the entire
treatment plan:
• Acquired: The number of X-ray images taken so far for the treatment plan, including
images taken manually during patient alignment and automatically during treatment
delivery. Images taken by Camera A and Camera B of the X-ray imaging system are
counted separately (that is, a pair of images from Camera A and Camera B are counted
as 2 images). The number of images acquired increases by 2 for each image pair
acquired.
• Expected: The estimated total number of X-ray images for all fractions in the treatment
plan. The estimate is updated prior to each beam delivery and depends on the following:
(Synchrony Fiducial Tracking with InTempo Imaging not in use) The value of the
Image Interval parameter. The estimated total changes when this value is changed.
(Synchrony Fiducial Tracking with InTempo Imaging only) The value of the Max
Image Age parameter. The estimated total changes when this value is changed.
A fixed number of X-ray images is added for each expected patient alignment
procedure (one patient alignment procedure per treatment fraction): For a treatment
plan that does not use Respiratory Modeling, 10 X-ray images (5 image pairs) are
added. For a treatment plan that uses Respiratory Modeling, 20 X-ray images (10
image pairs) are added. The fixed number is added to the estimated total only for
fractions that have not yet been delivered. For the fraction currently being delivered,
the fixed number added to the estimated total is 0.
The estimated total increases by 2 for each image pair acquired during patient
alignment or realignment, because the system cannot know how many images will be
acquired during these procedures.
Demonstration Mode
Demonstration mode allows you to demonstrate operation of the CyberKnife System by
performing a BB test. To access this mode, click the Demonstration button on the CyberKnife
System Menu and then log in. This button is enabled only when equipment power is on.
In Demonstration mode in the delivery automation phase, no treatment radiation is delivered.
The laser turns on and off to indicate radiation beam on/off behavior.
Live X-ray images can be acquired in the patient alignment phase using a phantom. However, no
live X-ray images are acquired during the robot motion phase. You have the option of bypassing
the X-ray imaging system in the patient alignment phase, allowing you to start the demonstration
without acquiring initial Live X-rays.
The demonstration is performed using the Teach Pendant to manually move the treatment robot.
Controls are the same as those in Treatment mode (see “Treatment Mode” on page 4-48),
except as described below. In Demonstration mode, the treatment delivery computer prompts
you when it is necessary to use the Teach Pendant. When the message shown in Figure 49 is
displayed on the treatment delivery computer, you remove the safety interlock key from the
operator control panel and use it to bypass the Treatment Room door interlock. Bypassing the
Treatment Room door interlock allows you to open the Treatment Room door without producing an
E-Stop.
WARNING: When the safety interlock key is used to bypass the Treatment Room door
interlock, X-ray exposure to users and observers in the Treatment Room can occur.
Ensure that no X-ray images are taken while the Treatment Room door interlock is
bypassed. In Demonstration mode, when the Teach Pendant is used to initiate the robot
motion phase of the demonstration, no further X-ray images are taken.
WARNING: Make sure no one is in the Treatment Room when Live X-ray images are
acquired to prevent unnecessary exposures.
WARNING: Always hold the Teach Pendant while performing a treatment simulation in
Demonstration mode, so that the Emergency Stop (E-Stop) button is readily available.
WARNING: WARNING: When the safety interlock key is used to bypass the Treatment
Room door interlock and the Treatment Room door is open, potentially harmful LINAC
radiation exposure to operators can occur outside the Treatment Room. Ensure that no
LINAC radiation is delivered while the Treatment Room door interlock is bypassed and the
Treatment Room door is open.
WARNING: Make sure no one is in the Treatment Room when Live X-ray images are
acquired to prevent unnecessary exposures.
Phantom Mode
Phantom mode allows you to simulate patient treatment using a phantom to check the accuracy
and safe delivery of a patient treatment plan. To access this mode, click the Phantom button on
the CyberKnife System Menu and then log in. This button is enabled only when equipment power
is on.
Controls are the same as those in Treatment mode.
Plan QA Mode
Plan QA mode allows you to select a quality assurance (QA) plan associated with a specific
patient and patient treatment plan and deliver the QA plan to a phantom. This mode provides a
method to perform quality assurance for patient-specific treatment plans. To access this mode,
click the Plan QA button on the CyberKnife System Menu and then log in. This button is enabled
only when equipment power is on.
WARNING: Plan QA should be performed prior to patient treatment delivery to verify that
the planned dose is consistent with the measured dose. Failure to perform plan QA can
result in patient mistreatment. For more information on the plan QA procedure, see the
Physics Essentials Guide.
Controls are the same as those in Treatment mode (see “Treatment Mode” on page 5-39),
except as described below.
QA Plans
A QA plan is a special type of treatment plan that is intended to be delivered to a phantom, not a
patient. Each QA plan is associated with a specific patient and patient treatment plan. A QA plan
is derived from the patient treatment plan that it is associated with.
In Plan QA mode, you can only select QA plans for delivery. Only a single fraction is created for
a QA plan. If the fraction is partially delivered and treatment is aborted, a makeup fraction is created
that can be immediately delivered.
For information on creating a QA plan, see the Treatment Planning Manual. For information on
authorizing a QA plan and recording analysis results, see the Data Management Manual.
The Load Patient Data screen displays a list of patients with approved QA plans and allows you to
select the patient that a QA plan is associated with. After you select the patient, a list of patient
treatment plans with approved QA plans is displayed. After you select a patient treatment plan, a
list of QA plans for that treatment plan is displayed. Patient treatment plans and QA plans are not
listed until they have been approved for delivery and DRR images have been generated on the
iDMS Data Management System.
As you select the patient, patient treatment plan, and QA plan, additional patient and treatment
information is displayed. A text message at the bottom left of the screen reports status as data is
retrieved from the database.
The screen includes the buttons listed in Table 17.
Button Function
Load After you have selected a patient, patient treatment plan, and QA plan,
begins downloading patient and treatment data from the iDMS Data
Management System to the treatment delivery computer.
Refresh Refreshes the Patient, Plan, and QA Plan lists with the latest data from
the iDMS Data Management System. New patients, patient treatment plans,
and QA plans are added to the list only after the plans have been approved
for delivery and DRR images have been generated on the iDMS Data
Management System.
The Patient section of the Load Patient Data screen displays the following information:
• Last Name: Last name of the patient.
• First Name: First name of the patient.
• Medical ID: Medical ID of the patient.
• Date of Birth: Date of birth of the patient
• Remaining: The number of fractions that remain to be delivered for the patient
treatment plan.
• Delivered / Prescribed (MU): Cumulative radiation amount already delivered for
the treatment plan (in monitor units) / Total prescribed radiation amount for the treatment
plan (in monitor units).
• Percentage progress bar: A bar graph displaying the percentage of the total prescribed
radiation amount for the treatment plan that has already been delivered.
The QA Plan section displays the following information:
• Name: The name of the QA plan.
• Phantom Last Name: The last name of the phantom to which the QA plan will be
delivered.
• Phantom First Name: The first name of the phantom to which the QA plan will be
delivered.
• Phantom Medical ID: The Medical ID of the phantom to which the QA plan will be
delivered.
The Notifications section displays the following information:
• Message panel: Notification message about the [treatment or QA] plan that is
displayed if an OIS is licensed and there is an inconsistency between OIS information
and the iDMS Data Management System
• Acknowledge checkbox: Indicates that you have read and acknowledged the
notification message. The checkbox must be selected before treatment can proceed.
Overview Phase
The left screen of the Overview phase allows you to verify that the phantom is in fact the one
specified in the QA plan and that the QA plan is the correct one to use. If a photo of the phantom
is available, it is displayed at the right of the screen.
Control Function
Data Verified Indicate that you have verified patient information, treatment and QA plan
checkboxes information, and then fraction information by selecting each checkbox in
turn. Each checkbox is enabled only after you have selected the previous
checkbox.
This screen displays the same information as in Treatment mode (see “Overview Phase” on
page 5-43) for the Patient section. The Plan section displays the following information:
• Plan Name: Name of the patient treatment plan that the QA plan is associated with.
• Treatment Site: Phantom anatomy to be treated, as specified in the QA plan.
• Orientation: Phantom position specified in the QA plan.
• Path Set: Name of the path set specified in the QA plan.
• Synchrony Method: Tracking mode specified in the QA plan.
• Respiratory: The status of Synchrony Respiratory Tracking (Yes if Synchrony
Respiratory Tracking is enabled in the QA plan, or No if Synchrony Respiratory Tracking
is not enabled). For more information on the Synchrony tracking System, see “Chapter
12: Synchrony Respiratory Tracking System (Option)”.
• InTempo: The status of InTempo imaging (yes if the InTempo imaging is enabled in
the treatment plan, or no if InTempo imaging is not enabled). For more information on the
Synchrony Fiducial Tracking with InTempo Imaging, see “Chapter 13: Synchrony Fiducial
Tracking with InTempo Imaging (Option)”.
• Total Prescribed MU: Total prescribed radiation amount for the QA plan (in monitor
units).
• Total Delivered MU: Cumulative radiation amount delivered for the QA plan.
• Prescribed Dose (cGy/Gy): Prescribed radiation dose for the QA plan (in centigray
or gray depending on the iDMS System setting).
• Total Fractions: The number of fractions to be delivered in the QA plan. The total
number of fractions is always 1 for a QA plan.
• Plan Collimator: The type of collimator specified in the QA plan (Fixed for fixed
collimators or Iris for the optional Iris Collimator), followed by the aperture (in
millimeters) for fixed collimators.
• QA Phantom Name: The name of the phantom to which the QA plan will be
delivered.
• QA Phantom Medical ID: The Medical ID of the phantom to which the QA plan will
be delivered.
• QA Plan Name: Name of the QA plan to be delivered.
The Fraction section displays the following information:
• Fraction Number: The selected fraction number to be delivered.
• Prescribed MU: Prescribed radiation amount for the QA plan fraction (in monitor
units).
• Delivered MU: Cumulative radiation amount delivered for the QA plan fraction.
• Prescribed Dose (cGy/Gy): Prescribed radiation dose for the QA plan fraction (in
centigray or gray depending on the iDMS System setting).
The top of the right screen of the Overview phase summarizes current patient and QA plan data.
• First Name: First name of the patient (not the phantom name).
• Last Name: Last name of the patient (not the phantom name).
• Medical ID: Medical ID of the patient (not the phantom).
• Plan Name: Name of the QA plan (not the patient treatment plan).
• MU: Cumulative radiation amount delivered for the QA plan (updated in real time as
dose is delivered) and the total prescribed radiation amount for the QA plan, given in the
following form:
(Delivered MU) / (Prescribed MU)
where
MU is the number of monitor units.
• Treatment Site: Phantom anatomy to be treated, as specified in the QA plan.
• Orientation: Phantom position, as specified in the QA plan.
• Synchrony Method: Tracking mode specified in the QA plan.
• Mode: The mode button selected on the CyberKnife System Menu.
Introduction
This chapter describes patient preparation procedures that should occur prior to treatment using
the CyberKnife Treatment Delivery System. During pretreatment scanning, image studies are
acquired that will be used for treatment planning.
This chapter covers the following topics:
• “Patient Immobilization Devices” on page 6-2
• “Pretreatment Scanning Guidelines” on page 6-7
• “Fiducial Placement” on page 6-18
If the patient immobilization device is created on the treatment couch of the CyberKnife System, it
is important to ensure that the pitch and roll of the treatment couch are at zero degrees (0°) during
the immobilization process.
Body Immobilization
This section gives instructions for preparing a body immobilization device.
WARNING: Verify that placement of patient extremities does not violate the Patient Safety
Zone (see “Chapter 2: Safety”). Otherwise, contact can occur between the treatment robot
and the patient.
2. Make an immobilization device (an alpha cradle or vacuum bag can be used).
Upper body immobilization (bottom of sternum and superior):
• Include patient head and shoulders extending inferior toward pelvis.
• Include arm position and imprint of fingers when possible.
An example of an alpha cradle for upper body immobilization is shown in Figure 2 on
page 6-5.
To evaluate patient position and fiducial visualization for patients with fiducials:
For patients with fiducials, evaluate patient position and fiducial visualization by acquiring a Live X-
ray image using the CyberKnife System as described below.
1. Set patient in the treatment position using the newly created immobilization device.
2. Position the patient's treatment area into the imaging field of the X-ray imaging system.
3. On the treatment delivery computer, open a phantom treatment plan designed using the
Synchrony Fiducial Tracking mode.
4. Acquire Live X-ray image.
• Ignore fiducial errors. These errors occur due to the difference between patient and
phantom fiducials.
5. Verify the number and location of fiducials against details from physician operative notes
and films.
• Reposition the patient as needed to bring fiducials into the X-ray imaging field.
• Check for obscured, overlapping, or unusable fiducials.
6. Repeat steps 2 and 3 as necessary to determine the quality of fiducial placement and
patient setup.
7. Using Treatment Room lasers for reference, mark the patient position on the
immobilization device.
WARNING: Some scanners may orient patient data differently when the patient position is
other than Head First Supine (HFS). Image studies from scanners should not be used
unless the correct orientation of the patient in the Accuray Precision System has been
verified. Treatments based on incorrectly oriented patient data can lead to mistreatment of
the patient.
It is essential to have the patient in the same position for treatment as they were during
pretreatment CT scanning. A reproducible patient position is the primary requirement during
patient preparation for the CT scan.
For the standard treatment couch, if the patient will be treated using a knee bolster, the bolster
should be used during pretreatment scanning.
If a RoboCouch bolster cushion is used, the RoboCouch seat cushion should be placed on top of
it and attached using the hook and loop fasteners. The bolster cushion should be positioned on the
indexing bar using the same mounting position that will be used for treatment. For information on
the RoboCouch cushions, see “Chapter 7: Treatment Room Operations”.
WARNING: Follow the above guidelines to position the patient for pretreatment scanning.
Otherwise, the patient treatment position will not match the pretreatment scanning
position, and mistreatment can result.
For more information on the RoboCouch System, see “Chapter 7: Treatment Room Operations”.
For more information on the Patient Safety Zone, see “Chapter 2: Safety”.
Secondary Images
Secondary images are used to identify and delineate structures. Supported secondary image types
include the following:
• MRI scans
• PET scans
• 3D Angiography
• Additional CT scans
Imaging guidelines for supported image types are given in the sections that follow.
CT Imaging Guidelines
The patient position during CT scanning should be the same as the patient position that will be
used to treat the patient. Bring all immobilization devices created for the patient to the CT scanner
to replicate the patient position.
Accuray has determined that some third party CT table tops may contain embedded "guide wires"
that may cause problems when used with the CyberKnife imaging system. If these table tops are
used, Accuray recommends that the patient be positioned so that the target or Region of Interest
(ROI) for tracking purposes does not intersect the "guide wires".
You can only import CT image studies with axial slices to the CyberKnife System. Basic
parameters for CT imaging should be as follows:
• Axial or helical/spiral 1-to-1 pitch.
• IV contrast may only be used in secondary CT image studies.
• Select a CT protocol to optimize for the patient anatomy, the tracking modality (see the
following tracking sections), and reduced noise that favors the following:
High contrast image.
Reduced artifacts with Metal Artifact Reduction (MAR) if metal present.
Accurate delineation/contouring.
• The CT density model may depend on the selected protocol as well as the CT scanner.
You may need multiple density models. Refer to the Physics Essentials Guide for more
information on CT density models.
• No more than 1.5 mm slice thickness. Contiguous slices (no gaps).
Anterior gap 0 – 1 cm
Superior gap 0 – 1 cm
Patient centered left-to-right in
the FOV
4. Avoid placing external objects such as those listed below in the field of view. Avoid using
contrast agents which can interfere with Synchrony Spine Tracking.
External objects in the region of interest can interfere with Synchrony Spine Tracking
algorithm. Ensure external objects are not included in the region of interest for treatment
planning and treatment delivery.
• Indexing bar
• Cervical collar
• Bone replacement materials
• Contrast agents
5. Acquire the CT scan during breath hold at the end of normal expiration.
WARNING: Position the patient and all devices within the Patient Safety Zone defined by
the Proximity Detection Program (PDP) model. Anything extending outside the boundaries
of the PDP model increases the risk of collision with the treatment robot and may result in
serious injury or death of the patient.
Follow the guidelines for patient preparation for pretreatment scanning described in “Chapter 12:
Synchrony Respiratory Tracking System (Option)”.
3. Optimize for bone imaging and for the visualization of the lung tumor.
The Medical ID for the PET images must exactly match the medical ID used in the CyberKnife
System. Spaces and punctuation (such as dashes, slashes, commas, and so on.) may not be used
in the Medical ID or patient name.
Each image within a PET scan should meet the following conditions:
• Identical pixel dimensions
• Identical field of view
• Identical slice thickness
• Even slice spacing
• 16-bit monochrome data
• Attenuation corrected for Standardized Uptake Value (SUV) calculation
Record or enter the following information on the scanner workstation during the PET image
acquisition process, in order to compute the SUV on the Accuray Precision System:
• Patient weight (in kilograms, kg)
• Patient height (in centimeters, cm)
• Patient gender (Female, F; Male, M)
• Radionuclide half life (in seconds)
NOTE: Different vendors store the scan date and time either as
the DICOM image series date and time or the DICOM image
acquisition date and time. This difference can result in different
statistics for the SUV calculation.
NOTE: Multiple image series from the same image study will be
linked together in the CyberKnife System database. If an image
study is to be performed in multiple segments and then
combined, each segment should have the same Image Study
Unique Identification (UID). Each segment can have a different
Image Series UID.
Fiducial Placement
Fiducial Placement Principles
In order to treat the patient with ease and accuracy, a few key principles need to be taken into
consideration when fiducials are implanted for patient treatments with the CyberKnife System.
Using 3 or more fiducials significantly improves targeting accuracy. The recommended number of
fiducials is 4 - 6. An example fiducial configuration is shown in Figure 5 on page 6-18.
• The distance between the geometric centroid of the fiducial set and the geometric
centroid of the target must be 50 mm or less.
• Non-collinear placement (in the X-ray imaging plane): Right Anterior Oblique (RAO)/Left
Anterior Oblique (LAO) or Right Posterior Oblique (RPO)/Left Posterior Oblique (LPO) X-
ray images are acquired during patient treatments.
Ensure fiducial markers are not obscured in the 45 degree angle projections of the
imagining system by dense tissue or by other fiducials. (See Figure 7 “Patient coordinate
axes are based on the Head First Supine (HFS) patient position” on page 13).
WARNING: When fiducials are used for tracking, ensure that all required fiducials are
within the field of view and verify that they are visible in the DRRs.
You may find that some of the implanted fiducials are unusable. For example, implanted fiducials
may be unusable for any of the following reasons:
• Migration
• Close placement of fiducials (closer than 2.0 cm). For example, the fiducials labeled 1
and 2 in Figure 5 on page 6-18.
• Less than 15 degree angle between any grouping of 3 fiducials. For example, the angle
between the fiducials labeled 2, 1 and 3 in Figure 5 on page 6-18.
Fiducial Supplies
Information on approved fiducial supplies for soft tissue and for spine and bone is given below.
WARNING: Use only Accuray-approved fiducials for treatment with the CyberKnife
System. Fiducial tracking with the CyberKnife System has been validated using approved
fiducials only. Use of unapproved fiducials is not supported and could result in patient
mistreatment.
Dimensions
Manufacturer’s (Diameter x
Manufacturer Type Part Number Length) Description
Alpha Omega Gold Seed SMG0242-025 0.8 x 5.0 mm 25 gold seeds per vial,
packaged non-sterile. 19G thin
walled biopsy needle 10 - 11 cm
or 15 cm length.
Dimensions
Manufacturer’s (Diameter x
Manufacturer Type Part Number Length) Description
Post-placement Guidelines
For fiducials implanted for the treatment of spinal lesions, it is advised to wait 24 – 48 hours before
acquiring the CT scan therefore enabling resolution of any localized pain or swelling. For soft tissue
lesions, 1 week is recommended to allow adequate time thereby minimizing fiducial migration.
Lung
The treatment of lung lesions using the CyberKnife System requires the implantation of gold seeds
referred to as fiducial markers. This outline reviews the procedure for placement of fiducial markers
using a percutaneous approach with CT guidance. In order to track lesions in 6D (translational and
rotational movements), a minimum of 3 fiducials are required for the software to track the lesion. It
is recommended to place a minimum of 4 fiducials, 1 more than is required for tracking in 6D, and
no more than 6. This recommendation is based on a study of the tracking accuracy possible in the
presence of uncertainties in fiducial localization both in the CT scan and live radiographic images.
(Murphy MJ Fiducial-based targeting accuracy for external-beam radiotherapy Med Phys 29(3)
334-344 (2002).)
Outline of Procedure
Lung information provided by Stephen T. Kee M.D., Associate Professor of Interventional
Radiology, Stanford University Medical Center.
The fiducial placement procedure is very similar to a CT guided lung biopsy. Anyone who has
experience with CT guided lung biopsies should be well aware of the methods employed and
potential complications of the placement of needles in lung tissue. Perform the procedure utilizing
the exact same methods used for a CT guided lung biopsy; the main difference being that once the
needle is placed in the lesion, fiducials are implanted rather than taking tissue out. The
complications and challenges are almost identical.
The first step is to obtain a detailed informed consent during which the patient is made aware of
complications pertinent to placement of any needle within the lung, namely pneumothorax and
hemorrhage. The incidence of pneumothorax with a single visceral pleural puncture is
approximately 15 – 30%; this increases if multiple visceral punctures need to be performed.
Hemorrhage can occur any time a needle is placed within the lung tissue resulting in hemoptysis.
The patient is placed on the CT table depending on the best approach to be used to access the
lesion. The procedure is performed using conscious sedation and the patient should be monitored
following hospital protocol for such a procedure. Images are acquired to locate the lesion and
assess the approach and puncture site. The area is prepped using sterile technique and local
anesthetic is injected at the intended puncture site. Under CT fluoroscopic guidance, the biopsy
needle is advanced into the target lesion. The smallest gauge biopsy needle through which the
gold seeds will fit should be used.
When deciding what trajectory to use in order to place the needle within the lesion, the ideal route
allows for placement of multiple fiducials within the lesion without having to puncture the visceral
pleural multiple times. There is a significantly lower incidence of pneumothorax should the visceral
pleura be transversed once rather than multiple times. When placing the needle within the lesion
attempt to advance the needle to the back wall of the lesion and place the first fiducial in this
location. Withdraw the needle from the initial fiducial placed, verify with imaging, then implant the
second fiducial. Angle the needle medial or lateral for placement of the other fiducials repeating
the process described above on the opposing side of the lesion. Whenever possible, every attempt
should be used to place the fiducials within the lesion as long as the fiducial placement guidelines
are met (see Figure 6 on page 6-23). CT scanning can be used to verify proper fiducial placement.
Image patient at the end of the procedure to assess for complications. Obtain a chest X-ray 3 hours
after the procedure to ensure no late onset of pneumothorax.
Potential Complications
Pneumothorax — Occurs in approximately 30% of all biopsy cases; less than 10% require
treatment. A small pneumothorax can be observed and if stable in size requires no further
treatment. A pneumothorax greater than 20 – 30% requires either aspiration or placement of a
chest tube. Aspiration of air from a pneumothorax can be done through a sheathed needle while
the patient is still on the CT table. Placement of a chest tube usually requires the patient to be
supine and is often performed after the post-procedure chest X-ray. In our experience at Stanford,
placement of a chest tube has been required in a minimum of cases, less than 5%. Should a chest
tube be required, this usually requires an overnight admission to a step down unit with continuous
wall suction. A chest X-ray is performed the next day. If the pneumothorax be resolved and stable,
the patient is discharged.
Hemoptysis — Occurs in approximately 10 – 50% of cases whenever a needle is introduced into
the lung. It is less common in patients who do not have a biopsy, therefore it is rarely observed with
fiducial placement procedures. Prior to the procedure evaluate if the patient is taking anti-
coagulants. Should hemoptysis occur, the major concern is the patient’s alarm at seeing that they
are coughing up blood. Typically this complication ceases spontaneously due to the low blood
pressure within the pulmonary arteries. It is usually not life threatening. It is important to adequately
inform the patient during consent and remain calm while rendering the appropriate supportive care.
Points To Consider
The major downside with performing fiducial placement in the lung is the difficulty of obtaining a
good trajectory to place the needle within the lesion while taking into consideration the surrounding
anatomy, that is, the ribcage. Anyone who has performed multiple lung biopsies should be able to
adequately guide the placement of a needle into the lesion without great difficulty. CT and CT
fluoroscopic guidance is recommended for the placement of fiducials as it allows for better
visualization of the entire lung and lesion and enables the operator to rapidly identify complications
during the procedure.
If a pneumothorax should develop during the procedure attempt to complete the fiducial placement
procedure if at all possible. It may be necessary to place a chest tube during the procedure to re-
inflate the lung. If the patient is stable and can tolerate it, the procedure should be resumed and
completed. Should a pneumothorax develop during the placement of the first needle consider
leaving it in place and introduce a second needle into the lesion to complete the procedure.
Another point to consider is the often used statement that "the enemy of good is perfect”. During
the procedure the goal is to get multiple fiducials placed into the lesion in the best locations. It is
not absolutely necessary to get 5 fiducials in the most perfect place within the lesion to treat the
patient. Placement of 3 fiducials within or adjacent to a tumor will be sufficient to track and treat the
patient. Therefore, should the procedure become more complicated than expected, or should the
patient become unable to tolerate a long procedure rapidly placing 3 fiducials, adhering to the
fiducial placement principles, in and around the tumor should be adequate in most cases.
There is always the possibility of placing a fiducial within a significant sized pulmonary vein which
could become a serious complication as it may travel and become lodged in the heart, lungs, or
brain. This has not occurred but is possible, therefore it is mentioned in this section. It is important
to include this as a potential complication when obtaining patient consent. Documentation of the
discussion between the physician, patient, and preferably a relative is recommended.
Conclusion
The key to success is to adequately inform the patient, perform the procedure with caution using
the standard approach for needle biopsies, be prepared for potential complications, and be
familiarized with the fiducial placement principles.
Pancreas/Liver
Pancreas and liver information provided by Stephen T. Kee M.D., Associate Professor of
Interventional Radiology, Stanford University Medical Center.
Outline of Procedure
The risks of the procedure are essentially identical to those of an image guided biopsy; however,
depending upon the number of needle repositions required in order to adequately place the
fiducials these risks may be somewhat higher. Most of the risks associated with placement of
needles in soft tissues is hemorrhage; these risks are kept to a minimum if a single needle puncture
can be used in order to place multiple fiducials and if adequate imaging is obtained prior to the
procedure in order to identify potential hazards. A detailed informed consent outlining the potential
complications to the patient should be the first step.
The diagnostic imaging technique of choice is CT or CT-fluoroscopy; however, other operators
have performed this procedure using ultrasound guidance and this is quite feasible particularly for
large lesions. If the patient is considered a candidate for treatment with the CyberKnife System
while in the operating room, placement of the fiducial markers using direct visualization could be
an option.
The patient is placed on the CT table depending on the best approach to be used to access the
lesion. The procedure is performed using conscious sedation and the patient should be monitored
following hospital protocol for such a procedure. Images are acquired to locate the lesion and
assess the approach and puncture site. The area is prepped using sterile technique and local
anesthetic is injected at the intended puncture site. Under CT guidance, the biopsy needle is
advanced into the target lesion. The smallest gauge biopsy needle through which the gold seeds
will fit should be used.
Angle the needle medial or lateral for placement of the other fiducials repeating the process
described above on the opposing side (or wall) of the lesion. Whenever possible, every attempt
should be used to place 4 – 6 fiducials within the lesion as long as the fiducial placement guidelines
are met (see Figure 7 on page 6-26). This can prove to be challenging with lesions less than 3 cm
in diameter. With smaller lesions, it is acceptable to place fiducials in the organ tissue surrounding
the lesion. CT scanning should be used frequently to verify proper fiducial placement.
Potential Complications
Hemorrhage is the most likely complication when puncturing the liver, pancreas, and other soft
tissue regions in the body. Use of a small gauge needle and knowledge of critical structures
surrounding the lesion should minimize this complication. Should the concern of hemorrhage be
present for whatever reason following placement of the last fiducial, a small amount of gel-foam
slurry can be injected into the needle as it is being removed to immobilize the track. The gel-foam
slurry is created by cutting up small amounts of gel-foam, mixing with either saline or lidocaine.
Transgression of a section of bowel when advancing the needle towards a lesion particularly for
pancreatic lesions is a possibility. There are usually no necessary precautions to be taken if the
stomach or small bowel is to be transgressed, however, if the large bowel is transgressed then it
is recommended to place the patient on broad spectrum antibiotics for one week. Every attempt
should be made to avoid puncturing these structures especially the large bowel.
Points To Consider
Typically, during placement of fiducials in pancreatic lesions the imaging does not include
administration of intravenous contrast. Therefore, the contrast study from the patient’s previous
imaging should be studied to ensure that no significant vessels will be transgressed with the
placement of the needle and fiducials. If there are any doubts, a contrast CT study should be
performed during placement. There are some situations in which structures must be transgressed
to access the lesion for fiducial placement. In our practice, the preference is to cross "nothing" first,
stomach second, liver third, small bowel fourth and large bowel as an extreme last resort. As stated
above, if the large bowel is punctured the patient should be administered antibiotics.
Conclusion
The fiducial placement procedure is very similar and only slightly more elaborate than a standard
image guided biopsy. Anyone who has experience with CT guided biopsies should be well aware
of the methods employed and potential complications of the placement of needles in soft tissue
organs such as the pancreas and liver. Perform the procedure utilizing the same methods used for
an image guided biopsy, the main difference being that once the needle is placed in the lesion,
fiducials are implanted rather than taking tissue out. The complications and challenges are almost
identical. The key to success is to adequately inform the patient, perform the procedure with
caution using the standard approach for needle biopsies, be prepared for potential complications
and be familiarized with the fiducial placement principles.
Introduction
This chapter describes treatment room operations when using the CyberKnife Treatment Delivery
System. The following topics are covered:
• “Preparing the Treatment Room” on page 7-2
• “Manually Changing Fixed Collimators” on page 7-3
• “Xchange Robotic Collimator Changer” on page 7-8
• “Patient Immobilization Devices” on page 7-23
• “Standard Treatment Couch” on page 7-23
• “RoboCouch Patient Positioning System (Option)” on page 7-31
WARNING: Do not remove the treatment robot Teach Pendant from the Treatment Room
during treatment delivery. The Teach Pendant should only be used in the Treatment Room
in order to safely monitor operation of the treatment robot.
For more information on manual operation of the treatment robot using the Teach Pendant, see
“Chapter 4: System Startup and Shutdown”.
The CyberKnife System detects the size of the fixed collimator that is currently installed on the X-
ray head of the treatment robot. Depending on the treatment plan you select, you may need to
change the collimator before treatment begins. Additionally, if the treatment plan uses multiple
fixed collimators, you need to change the collimator between treatment paths.
The CyberKnife System includes 12 interchangeable fixed collimators with apertures (in
millimeters) of 5, 7.5, 10, 12.5, 15, 20, 25, 30, 35, 40, 50, and 60 (Figure 1) The aperture is stamped
on one or both faces, depending on the collimator. Each collimator is a different length and weighs
more than 12 lb (5 kg).
WARNING: Do not change the collimator when the X-ray head is over the patient. Always
move the treatment robot to the perch position before changing collimators. Otherwise,
the collimator may cause serious injury to the patient if it is dropped.
WARNING: When removing or installing a fixed collimator, use two hands and support the
collimator securely with your palm underneath the front face. Be careful not to drop the
collimator on your foot or any other part of your body, or on the X-ray detectors of the
imaging system. Otherwise serious injury to yourself or equipment damage can occur.
A fixed collimator is held in place by a spring-loaded retaining latch pin inside the fixed collimator
housing of the X-ray head. The retaining pin latches underneath the collimator lip. The collimator
is then secured on the housing by a retaining nut (see Figure 2).
WARNING: Never leave a fixed collimator hanging only by the retaining pin when the X-
ray head is over the patient or yourself. Support the collimator with your hand until the
collimator is secured on the fixed collimator housing by the retaining nut. Otherwise, the
collimator may cause serious injury to the patient if it is dropped.
Contact
Detection
Collimator System
Lip
Fixed Collimator
Housing*
Fixed
Collimator
Retaining Nut
Retaining Nut
Exposed
3. Remove the retaining nut from the fixed collimator housing. There are two styles of
retaining nut:
• Style A: Unscrew the retaining nut to remove it from the housing.
• Style B: Rotate the retaining nut counterclockwise less than one turn to release it
from the housing.
4. Place the palm of your hand firmly underneath the fixed collimator to support it. Then
release the spring-loaded retaining pin and slowly slide the collimator out of the housing.
There are two styles of retaining pin:
• Style A: Pull the collimator retaining pin to release it (Figure 4).
Collimator
Retaining Pin
• Style B: Press the collimator release lever to release the retaining pin (Figure 5).
r
Collimator
Release Lever
WARNING: Use only fixed collimators that have been commissioned with the CyberKnife
System. Otherwise, the dose delivered to the patient will be different than expected, which
could result in under- or overdosing the patient.
The Xchange collimator table Model B supports both fixed collimators and the Iris Collimator (see
Figure 7). With Model B, the Xchange, up to 3 calibration isoposts and pinhole adapters can be
stored in the drawer of the Xchange table. Alternately, isoposts and the pinhole and blank
collimators can be stored in the AQA transport case.
Guide Pins
Receptacle for Iris
Collimator*
Fixed Collimator on
Collimator Cup
Secondary Housing
in Receptacle
If your Xchange System includes Model B and the Iris Collimator, the secondary housing for fixed
collimators can be removed from the X-ray head and replaced with the Iris Collimator. These
operations are performed automatically by the Xchange System. The secondary housing and the
Iris Collimator are attached to and released from the X-ray head by a pneumatic control system.
Model B includes two receptacles where the secondary housing and the Iris Collimator are
deposited when dropped off by the treatment robot.
If your Xchange System includes Model B and fixed collimators only (no Iris Collimator), the
secondary housing is permanently attached to the X-ray head, and the receptacles are covered.
Optical
Presence
Sensors
Secondary Housing
Presence Sensors
Optical Presence
Sensors
Iris Collimator
Presence Sensors
Collimator Cups
With the Xchange table Model B, the collimator size for each collimator cup is labeled as shown in
Figure 10. Each collimator cup contains 4 presence sensors that detect when a fixed collimator is
present and whether it is upside-down. The collimator cups are spring-loaded and may move
noticeably when pressure is applied.
With Model B, the treatment robot rotates the fixed collimator after it is lowered into the cup. The
dropoff ramp pushes the collimator release lever to release the retaining pin (see Figure 10). Then
the collimator is released into the cup.
Collimator
Presence
Sensors
Dropoff Ramp
Collimator Cup
Size (mm)
WARNING: Do not attempt to lift the secondary housing for fixed collimators or the Iris
Collimator. These components are quite heavy and can cause injury or damage if
dropped. Always use the Xchange System to install these components. It is not possible
to install them manually. Store these components on the Xchange collimator table.
WARNING: Personal injury may occur if you stick your fingers or other body part into the
path of a moving IRIS aperture. Do not stick your fingers or other part body part into a
moving IRIS aperture.
Each Iris Collimator has a unique Serial Number as well as system calibration files and beam data
that are associated with a specific CyberKnife System. After your CyberKnife System is
commissioned with a particular Iris Collimator, you cannot switch to a different Iris Collimator.
CAUTION: Use only the Iris Collimator that has been commissioned with your CyberKnife
System. Your CyberKnife System will not recognize the Serial Number of a different collimator.
An Accuray Service call is required to change the Serial Number that your CyberKnife System
recognizes, and then your CyberKnife System must be recommissioned.
CAUTION: Observe the following safety precautions when operating the Xchange System.
• Do not lean, sit, or stand on the Xchange collimator table. Otherwise, the
Xchange table may require recalibration by Accuray Field Service.
• Do not operate the Xchange System with the Xchange calibration isoposts
mounted on the Xchange table. Place them in the storage drawer before
operating the Xchange System.
• With the Xchange collimator table Model B, do not place objects other than the
fixed collimators, secondary housing, and the Iris Collimator on the Xchange
table. Otherwise, the Xchange table may require recalibration by Accuray Field
Service.
• The Xchange collimator table does not include any user-serviceable
components. If problems occur, contact Customer Support.
• Do not position yourself between the Treatment Robot and the Xchange
collimator table when the Treatment Robot is in the process of exchanging
collimators.
Collimator
Collimator
Cup
Housing for Fixed
Collimators
Retaining Nut
Remover
To automatically exchange collimators before and during treatment, follow the steps below.
To use the Xchange System:
1. Power up the CyberKnife System with the Xchange System. The power-up procedure is
the same as the standard procedure. For more information, see “Chapter 4: System
Startup and Shutdown”.
2. Log into Demonstration, Plan QA, Phantom, or Treatment mode as the
primary user (see “Chapter 4: System Startup and Shutdown”).
3. Select the treatment plan on the treatment delivery computer and load patient data, as
described in “Chapter 8: Treatment Delivery”.
4. Make sure the treatment robot is located at the perch position.
Move the treatment robot to the perch position manually if necessary (see “Move
Treatment Robot to the Perch Position” on page 7-2).
5. Make sure the fixed collimators are placed correctly on the Xchange collimator table and
the storage drawer is closed.
• Collimators must be placed according to size in the correct collimator cups. The size
of each collimator is stamped on one or both faces, depending on the collimator.
• Fixed collimators should be placed face down in the collimator cups.
6. If your Xchange System includes the Iris Collimator, make sure the system is in one of the
following configurations:
• The Iris Collimator and the fixed collimator secondary housing are both present in
their receptacles on the Xchange table.
• The Iris Collimator is present in its receptacle, and the fixed collimator secondary
housing (with fixed collimator and retaining nut attached) is installed on the X-ray
head.
• The Iris Collimator is installed on the X-ray head, and the fixed collimator secondary
housing (with fixed collimator and retaining nut attached) is present in its receptacle.
7. Make sure the Xchange collimator table is free of all other objects.
8. Align the patient, as described in “Chapter 8: Treatment Delivery”.
9. Go to the Readiness phase to begin the equipment readiness check. For more
information on the equipment readiness check and confirming delivery warning message,
see “Checking Equipment Readiness” on page 8-75.
The Equipment Readiness screen is displayed (see Figure 12). For more information on
this screen, see “Chapter 5: Treatment Delivery System Overview”.
10. The Collimator section of the Equipment Readiness screen displays the type of
collimator currently installed (fixed collimator with a particular aperture, or Iris Collimator).
If the current collimator matches the collimator specified in the treatment plan, a Laser
Alignment Check is required. The Laser Alignment Check button icon is enabled
and text in the Collimator section prompts you to perform a Laser Alignment Check.
Click the Laser Alignment Check button to perform the check (Figure 13).
If the current collimator type does not match the collimator type specified in the treatment plan, text
in the Collimator section prompts you to exchange collimator housings and the appropriate
exchange button is enabled. In this case, the Laser Alignment Check is performed during the
collimator exchange, and a Laser Alignment Check is not required after the correct collimator is
installed. Text reports the status of the collimator exchange.
A B C
If a collimator type change is required, press the enabled collimator button (see
Figure 13). The Xchange System performs the exchange described in “Automatic
Collimator Housing Change” on page 7-17.
Enable Buttons
(one on each side)
WARNING: The standard treatment couch does not detect collisions with surrounding
objects. Release the Enable buttons to stop motion when you see a potential collision.
WARNING: Ensure all body parts are kept away from any moving components of the
treatment couch. Do not position yourself, the patient, or other people between moving
components of the treatment couch and any stationary object or wall in the Treatment
Room. Otherwise injury or death can result.
WARNING: Do not remove the Hand Controller from the Treatment Room. The Hand
Controller should only be used in the Treatment Room, in order to safely monitor
movement of the treatment couch.
Button Description
LEFT/RIGHT Moves the treatment couch in the patient left or right direction (along the Left-
Right axis)
ANT/POST Moves the treatment couch in the patient anterior or posterior direction (along
the Anterior-Posterior axis).
INF/SUP Moves the treatment couch in the patient inferior or superior direction (along
the Inferior-Superior axis).
ROLL L/R Rotates the treatment couch about the Inferior-Superior axis towards the left or
right of the patient (roll).
HEAD U/D Rotates the treatment couch to bring the patient head up or head down (pitch).
HOME Moves the treatment couch to the home position. This position is programmed
as a convenient position of the treatment couch for the patient get on or off the
table.
TREAT Moves the treatment couch to the nominal treatment position. The nominal
treatment position for the first treatment session for a plan is obtained from a
stored position for the type of anatomy being treated (head or body). For
subsequent sessions, the treatment position used in the previous treatment
session is used.
Enable Both Enable buttons must be pressed before any treatment couch movement
can occur.
• Yellow: Indicates that the treatment couch E-Stop is engaged. In this state, both manual
and computer-controlled motions are disabled.
The MOVE LED is green whenever any button that performs a move is depressed.
In some circumstances, if you press and hold a movement button on the Hand Controller for more
than 5 seconds without pressing the Enable buttons, the treatment couch enters a calibration mode
for that axis. When this occurs, the message, "Press ZERO to Calibrate <axis name>
axis" is displayed on the Readout Display Unit. Calibration mode should only be used for
recalibrating the zero position of each axis. If this occurs, wait for a brief period to return to the
normal mode of operation, then contact Accuray Customer Support.
WARNING: DO NOT press the ZERO button when the Readout Display Unit displays the
message "Press ZERO to Calibrate <axis name> axis" and the treatment couch is not in
the correct location for recalibrating the zero position for that axis. Doing so will
compromise the calibration of the treatment couch and will affect the system’s ability to
detect potential collisions between the treatment robot and the treatment couch or patient.
The removable foot extension attaches to the superior end of the table top using the CIVCO
Prodigy® Lok-Bar. For patient treatment in the Feet First position, the foot extension should be
attached at the “H2” index mark (see Figure 17).
WARNING: Do NOT use the foot extension for Head First treatments or attach it at index
marks other than “H2”. Otherwise, an overload condition of the treatment couch can
occur and patient injury of damage to equipment can result.
WARNING: DO NOT sit, stand, or otherwise place any loads outside of the load zone of
the standard treatment couch.
DO NOT load patients on the table top outside of the 46 inch (1180 mm) load zone.
If a load is placed on the ends of the table top away from the support column, for example,
by loading the patient or sitting at either end of the table top, there is the potential for the
table top to partially detach from the support column and descend to the floor. This could
lead to patient injury depending on the position of the patient at the time.
RoboCouch Manipulator
The RoboCouch manipulator is a 6-axis robot capable of positioning the patient in six degrees of
freedom.
The 6 axes of the RoboCouch and hazard regions are shown in Figure 20.
Hazard
Region
Between
Arms
Upper
Arm
Table Top
Lower
Arm
WARNING: When the RoboCouch System is in motion, avoid all locations that could pose
a potential crush hazard. Such locations include (see Figure 20):
1. The region between the upper and lower arms of the RoboCouch manipulator.
2. The region between the upper arm and nearby walls.
3. The region between the RoboCouch table top and nearby walls.
4. The area beneath the RoboCouch table top.
5. The area between the RoboCouch table top and lower arm.
Pay attention to people and other objects in the room at all times to avoid severe injury or
death.
The RoboCouch column cover encloses the vertical translation axis (Axis A1) of the RoboCouch
manipulator. Axis A1 provides vertical travel for the RoboCouch table top. Roller shades on the
column cover move up and down with movement of Axis A1. The following configurations are
available for the RoboCouch column cover:
• Column cover installed in a corner of the treatment room. A brake release button is
located on the column cover (see Figure 21).
Brake Release
Button
• Freestanding column covers have a door panel in that provides access to a brake release
button and the RoboCouch Teach Pendant (see Figure 22).
WARNING: Ensure that objects do not become caught in the roller shades that cover the
vertical translation axis (Axis A1). Contact with the motors and drive shaft of Axis A1 can
result in injury.
CAUTION: Observe the following safety precautions when accessing the column cover:
RoboCouch Controller
The RoboCouch controller controls the manipulator and interfaces with other components of the
CyberKnife System, including E-Stop signals. The RoboCouch controller is located in the
Equipment Room.
The flat table top provides patient support. It is fabricated from carbon fiber and designed to provide
uniform radiolucency for the Target Locating System.
The headrest for the RoboCouch Flat Table Top configuration is the same as that for the standard
treatment couch.
WARNING: Do not remove the Hand Controller from the Treatment Room. The Hand
Controller should only be used in the Treatment Room, in order to safely monitor
movement of the treatment couch.
WARNING: Do not move the RoboCouch System and the treatment robot simultaneously.
The RoboCouch System includes a Hand Controller that allows manual movement of the
RoboCouch System in the treatment room. The following components are included with the Hand
Controller:
• RoboCouch Hand Controller
• Charging Cradle
• Synchrony interface module
Hand Controller
The RoboCouch includes the following components:
• RoboCouch Hand Controller
• Charging Cradle
• Wireless Synchrony Respiratory Tracking interface module
Synchrony
Respiratory
RoboCouch Tracking
Hand Interface
Controller Module
Charging
Cradle
The Cradle is a unit for storing the RoboCouch Hand Controller, and for both storing and charging
the wireless Synchrony Respiratory Tracking interface module. The Cradle provides an interface
between the Hand Controller and the RoboCouch Controller. The Cradle is wall-mounted and
includes a hook to keep the Hand Controller cable off the floor.
An LED indicator on the front of the Cradle is lit when there is power to the Cradle for charging the
Synchrony Respiratory Tracking interface module. For more information on the Synchrony
Respiratory interface module, see “Chapter 12: Synchrony Respiratory Tracking System (Option)”.
The RoboCouch Hand Controller allows manual movement of the RoboCouch System in the
treatment room. The Hand Controller provides manual and semi-automated control of all six
degrees of freedom of the RoboCouch System. The sections that follow describe Hand Controller
controls.
LED E-Stop
Indicator
Couch E-Stop
Status Button
Indicator
Contact
Sensor Speed
Indicator Indicator
Button
Position
Control
Buttons
Enable Button
(Located on
Back of Hand
Movement Controller)
Control
Buttons
E-Stop Button
The Emergency Stop (E-Stop) button is located at the top end of the RoboCouch Hand Controller.
Pressing the E-Stop button generates a system-wide E-Stop, including the RoboCouch System.
You cannot override this E-Stop using the RoboCouch Hand Controller.
Enable Button
The Enable button located on the back of the RoboCouch Hand Controller enables manual control
of the RoboCouch System. This button must be depressed while performing any manual
movements with the Hand Controller.
WARNING: Monitor the RoboCouch System closely at all times for potential collision with
surrounding objects. If you see a potential collision, release the Enable button on the back
of the Hand Controller to stop RoboCouch movement.
• Solid green: The treatment couch has reached the indicated position.
Button Description
When you press the Position Control buttons, the nominal (default) speed is used, unless you
select another speed (see “The Translate Mode controls are shown in .” on page 7-43).
Button Description
Pitch: Rotates the RoboCouch table top to bring the patient head
up or head down (pitch).
The Yaw, Pitch, and Roll buttons include a center icon and arrows indicating the direction of
rotation. In normal operation, the center icon and the arrows are lit when the button is enabled.
If treatment couch rotation exceeds a software limit in one direction, the following occurs:
• Movement of the treatment couch stops.
• All rotation arrows turn off (unlit) except for one.
• The one remaining arrow that is lit indicates the allowed direction to reverse the rotation
so that the treatment couch is back within the software limit.
Button Description
WARNING: Do not use high speed when the patient is on the treatment couch. Sudden
stops during high speed movement may startle the patient and result in injury.
The Contact Sensor Indicator is an LED display on the Hand Controller that indicates the status of
the contact sensors located on the RoboCouch table top (see “RoboCouch System Safety
Features” on page 7-46).
If contact occurs between the table top and the lower arm of the RoboCouch manipulator, or with
an object located between the table top and the lower arm, the contact sensor triggers an E-Stop
and the Contact Sensor Indicator (a yellow exclamation point icon) blinks.
WARNING: Monitor the RoboCouch System closely at all times for potential collision with
surrounding objects. If you see a potential collision, release the Enable button on the back
of the Hand Controller to stop RoboCouch movement.
The Couch Status Indicator is an LED display on the Hand Controller that indicates the status of
the RoboCouch System. Status conditions that may occur are listed in Table 4 on page 7-45.
Depending on the status, the indicator bars are lit as follows:
• Off
• Solid green
• Green blinking
• Yellow blinking
Table Top indicator bar is solid green. Seat indicator bar and Foot
Extension indicator bar are not used.
For instructions on manually lowering the RoboCouch table top for patient unloading during power
interruption (see “Lowering the Patient During a Power Interruption” on page 7-47).
WARNING: When a RoboCouch contact sensor E-Stop is bypassed, an E-Stop will NOT be
triggered if contact occurs again. There will be NO protection against collision or crushing
of people or objects, and injury or death can result. Monitor RoboCouch movement
closely to ensure no collision occurs. If you see a potential collision, release the Enable
button on the back of the Hand Controller to stop RoboCouch movement.
When the E-Stop is bypassed, the RoboCouch System can only be moved at slow speed using the
Hand Controller.
To temporarily bypass a RoboCouch contact sensor E-Stop:
1. Push and hold the E-Stop bypass button for the RoboCouch contact sensors. As shown
in Figure 26 for, this button is located on the side of Axis A5 to the patient’s left (in the
HFS patient position).
2. While pressing the E-Stop bypass button, use the RoboCouch Hand Controller to
manually move the RoboCouch manipulator out of contact.
WARNING: If an emergency situation occurs during a power interruption, press the brake
release button to lower the RoboCouch table top to its minimum height for easier
unloading of the patient.
The brake release button is only active during a power interruption. You may choose not to use the
brake release button if it is possible for the patient to be unloaded without lowering the table top.
CAUTION: If the brake release button is pressed to lower the RoboCouch table top, a service
call to Accuray is required to reset the RoboCouch System.
• Check whether the E-Stop button on the RoboCouch Hand Controller is depressed (see
“RoboCouch Hand Controller and Cradle” on page 7-36).
• Check whether the E-Stop button on the RoboCouch table-mounted cradle is depressed
(see “RoboCouch Hand Controller and Cradle” on page 7-36)
• Check whether the E-Stop button on the RoboCouch Teach Pendant is depressed.
• Check the position of the RoboCouch Teach Pendant mode selector switch (the key
switch located on the front of the Teach Pendant). An E-Stop is generated if the switch is
not in External mode, indicated by the following icon:
• Check the contact sensors. An E-Stop is triggered if contact is made with areas covered
by the RoboCouch contact sensors (see “RoboCouch System Safety Features” on
page 7-46). To bypass a contact sensor E-Stop, see “Bypassing a RoboCouch Contact
Sensor E-Stop” on page 7-46).
• Check the safety limit switches. An E-Stop is triggered if the limits switches on Axes A1,
A4, or A5 are tripped. To bypass a limit switch interlock, press the bypass button and
move the A1, A4, or A5 axis with the RoboCouch hand control.
WARNING: Monitor the RoboCouch System closely at all times for potential collision with
surrounding objects. If you see a potential collision, release the Enable button on the back
of the Hand Controller to stop RoboCouch movement.
WARNING: When the RoboCouch System is in motion and under control of either the
Hand Controller or the Teach Pendant, avoid all locations that could pose a potential
crushing hazard. Such locations include (see Figure 20 on page 7-33):
1. The region between the upper and lower arms of the RoboCouch manipulator.
2. The region between the upper arm and nearby walls.
3. The region between the RoboCouch table top and nearby walls.
4. The area beneath the RoboCouch table top.
5. The area between the RoboCouch table top and lower arm.
Pay attention to people and other objects in the room at all times.
WARNING: The treatment couch table top can be slippery. Make sure patients are
properly secured for the type of treatment they are receiving.
Patient Alignment
When the RoboCouch System is in the nominal treatment position, you are ready for the
Alignment > Couch phase. Figure 27 shows the automatic patient positioning controls in the
Alignment > Couch phase that you use with Couch Mode to visually align the Live X-ray and
DRR images.
For information on using the Automatic Patient Positioning window, see “Chapter 8: Treatment
Delivery”. The description in Chapter 7 applies to the RoboCouch System as well as the standard
treatment couch.
WARNING: Monitor the RoboCouch System closely at all times for potential collision with
surrounding objects. If you see a potential collision, release the Enable button on the back
of the Hand Controller to stop RoboCouch movement (see “RoboCouch Hand Controller
and Cradle” on page 7-36).
WARNING: The RoboCouch System does not detect when the patient is outside the
nominal patient setup boundaries. After patient alignment or realignment with the
RoboCouch System, ensure that the patient is within the nominal alignment boundaries.
See the section, "The Patient Safety Zone," in “Chapter 2: Safety”.
WARNING: Making large movements of the treatment couch from the treatment delivery
computer outside the treatment room may increase the risk of equipment colliding with or
pinching the patient. For better visibility, make large movements of the treatment couch
inside the treatment room with the hand pendant.
Interlock
button
Treatment Delivery
NOTE: Seat angles greater than 15° must be taken into account
during treatment planning. Otherwise, the Patient Safety Zone
will be violated and you will not be able to proceed with
treatment. For more information on the Patient Safety Zone, see
“Chapter 2: Safety”.
During treatment delivery, no alignment operations are allowed using the RoboCouch System.
During treatment delivery, the treatment robot repositions the LINAC to correct the beam direction
for patient displacement. The treatment robot corrects for both translational and rotational
displacements.
For CyberKnife Systems that include the RoboCouch System, the robot corrects the beam
direction up to 10 mm along each translation axis (for treatments that do not use the Synchrony
tracking System), and up to ±1.5 degrees for rotations (yaw, pitch, and roll).
During patient alignment or realignment, the rotational correction limits can be changed to values
below these maximum values using the Imaging Parameters window. Rotational corrections during
treatment delivery can also be disabled in the Imaging Parameters window. For more information
on displacement corrections during treatment delivery, see “Chapter 8: Treatment Delivery”.
Patient Realignment
All alignment functionality provided for patient alignment is also provided during patient re-
alignment. For more information, see “Chapter 8: Treatment Delivery”.
WARNING: The RoboCouch System does not detect when the patient is outside the
nominal patient setup boundaries. After patient alignment or re-alignment with the
RoboCouch System, ensure that the patient is within the nominal alignment boundaries.
See the section, "The Patient Safety Zone," in “Chapter 2: Safety”.
Introduction
The chapter describes the general workflow for treatment delivery using the CyberKnife Treatment
Delivery System.
This chapter covers the following topics:
• “Workflow for Patient Alignment and Treatment Delivery” on page 8-3
• “Approving Plans for Treatment Delivery and DRR Generation” on page 8-5
• “Loading Patient and Treatment Data” on page 8-5
• “Initial Patient Positioning” on page 8-10
• “Visually Aligning the Patient” on page 8-15
• “Optimizing the X-ray Technique” on page 8-22
• “Performing Patient Alignment” on page 8-27
• “Checking Readiness for Treatment Delivery” on page 8-56
• “Treatment Delivery” on page 8-69
• “Interrupting Treatment” on page 8-82
• “Error Handling” on page 8-84
• “Patient Realignment After a Pause or Error” on page 8-91
• “Makeup Fractions” on page 8-92
• “Phantom Mode” on page 8-93
• “Demonstration Mode” on page 8-93
• “Plan QA Mode” on page 8-100
This chapter assumes that the CyberKnife System is powered up and that you are logged into the
treatment delivery computer. It also assumes you are familiar with "Chapter 5: Treatment Delivery
System Overview" that describes the user interface of the Treatment Delivery System.
This chapter provides general procedures that apply to all tracking modes and describes
Synchrony Skull Tracking mode. For detailed information on other tracking modes, see the
following chapters:
• "Chapter 9: Synchrony Fiducial Tracking"
• "Chapter 10: Synchrony Spine Tracking (Option)"
• "Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)"
• "Chapter 13: Synchrony Fiducial Tracking with InTempo Imaging (Option)"
• "Chapter 15: Lung Optimized Treatment (Option)"
Figure 1 Workflow for patient alignment and treatment delivery for treatments that
do not use Respiratory Modeling (Synchrony Skull Tracking mode, Synchrony
Fiducial Tracking mode, Synchrony Spine Tracking Supine).
Figure 2 shows the user interface workflow for Synchrony Fiducial Tracking with Respiratory
Modeling mode and Synchrony Spine Tracking with Respiratory Modeling mode.
Figure 2 Workflow for patient alignment and treatment delivery for treatments that
use Respiratory Modeling (Synchrony Fiducial Tracking with Respiratory Modeling
mode and Synchrony Spine Tracking with Respiratory Modeling).
Figure 3 shows the workflow for patient alignment and treatment delivery using Synchrony Lung
Tracking with Respiratory Modeling and Lung 1 View Tracking with Respiratory Modeling.
Figure 3 Workflow for patient alignment and treatment delivery for treatments that
use Respiratory Modeling (Synchrony Lung Tracking with Respiratory Modeling
mode and Lung 1 View tracking with Respiratory Modeling)
For information on using Synchrony Respiratory Tracking System, see "Chapter 12: Synchrony
Respiratory Tracking System (Option)".
The goal of each user interface workflow phase for treatments that do not use Respiratory
Modeling (see Figure 1) is as follows:
• Overview phase: Load patient, treatment plan, and treatment fraction data. Verify that
patient and treatment data are correct.
• Alignment > Couch phase: Load the patient onto the treatment couch. Visually align
the patient, using translational couch movements only. Take at least one Live X-ray
image and confirm visual alignment of the Live X-ray and DRR images.
• Alignment > Technique phase: Find the optimal X-ray technique. This X-ray
technique will be the default for future treatment fractions unless it is changed.
• Alignment > Align phase: Ensure the patient is aligned within acceptable bounds for
treatment delivery or Respiratory modeling if the SynchronyTracking System will be used.
• The Physician Setup Approval button allows a physician to electronically approve the
current patient alignment. For more information see “Physician Setup Approval of Patient
Alignment” on page 8-52.
• Readiness phase: Check that the patient is within the Patient Safety Zone. Check
equipment readiness before starting treatment delivery. Resolve, when applicable, and
confirm all delivery warning messages. Exchange collimators or perform a Laser
Alignment Check, as needed. These steps can optionally be performed earlier during the
Alignment > Couch phase.
• Delivery phase: Start treatment delivery. Monitor and verify patient alignment and
tracking results during treatment delivery.
Tasks performed during these phases are described in the sections that follow.
WARNING: Do not attempt to modify a treatment file. Improperly changed data can result
in injury or death to the patient.
On the day of treatment, on the treatment delivery computer, power on the CyberKnife System as
described in "Chapter 4: System Startup and Shutdown". The startup procedure, including X-ray
tube warmup, should be completed before the patient arrives.
Loading patient and treatment data includes the following steps:
• “Select the Patient, Plan, and Fraction” on page 8-6
• “Verify Patient and Treatment Data” on page 8-8
1075883-ENG A Approving Plans for Treatment Delivery and DRR Generation | 8-5
Accuray Incorporated Treatment Delivery Manual
2. Click to select a patient name from the Patient list. After a moment, the Plan list is
populated with the names of approved treatment plans for that patient. Treatment plans
are not listed until DRR images have been generated on the iDMS Data Management
System.
WARNING: Verify that this patient is in fact the one to be treated and the treatment plan is
the correct one to use.
The top of the right screen of the Overview phase summarizes patient and treatment data (see
Figure 6).
When this window appears, the first Data Verified checkbox is active. You must verify that
patient, treatment plan, and treatment fraction information in the three columns on the left screen
are correct.
Supine Prone
For more information on the see "Chapter 12: Synchrony Respiratory Tracking System (Option)".
WARNING: The PDP model is based on the Head First Supine (HFS) patient position.
Ensure that the patient and all devices are within the Patient Safety Zone defined by the
PDP model, even if the patient is not in the HFS position. For more information on the
Patient Safety Zone, see "Chapter 2: Safety".
WARNING: In the user interface of the Treatment Delivery System, patient coordinate
axes and screen controls are labeled based on the Head First Supine (HFS) position (see
Figure 7). Be aware of this convention if the patient position used for treatment is not HFS.
ANT
INF
RGT
During treatment planning, only treatment plans that specify supported combinations of tracking
mode and patient position are allowed. You cannot save a treatment plan that specifies an
unsupported patient position in the Accuray Precision Treatment Planning System. For more
information on the Accuray Precision System, see the Treatment Planning Manual.
For information on the Hand Controllers for the standard treatment couch and the optional
RoboCouch Patient Positioning System, see "Chapter 7: Treatment Room Operations".
WARNING: Making large movements of the treatment couch from the treatment delivery
computer outside the treatment room may increase the risk of equipment colliding with or
pinching the patient. For better visibility, make large movements of the treatment couch
inside the treatment room with the hand pendant.
WARNING: The treatment couch table top can be slippery. Make sure patients are
properly secured for the type of treatment they are receiving.
WARNING: DO NOT sit, stand, or otherwise place any loads outside of the load zone of
the standard treatment couch.
DO NOT load patients on the table top outside of the 46 inch (1180 mm) load zone.
If a load is placed on the ends of the table top away from the support column, for example,
by loading the patient or sitting at either end of the table top, there is the potential for the
table top to partially detach from the support column and descend to the floor. This could
lead to patient injury depending on the position of the patient at the time.
For more information on the load zone of the standard treatment couch, see "Chapter 7:
Treatment Room Operations".
Use the same immobilization device that was used for pretreatment scanning. Make sure
the patient is placed in the immobilization device and positioned on the table top in the
same way as during the pretreatment CT scan.
For more information on immobilization devices, see "Chapter 6: Patient Preparation".
5. Press the TREAT button on the Hand Controller to move the treatment couch to the
appropriate initial alignment position. In this position, the anatomy to be treated is in the
field of view of the X-ray imaging system.
The treatment couch moves to either the previous or default alignment position,
depending on whether this is the first treatment session or a subsequent treatment for the
treatment plan.
6. If a position measurement device or alignment lasers are available in the Treatment
Room, you can use them to help fine tune the initial alignment position. Using the Hand
Controller, move the treatment couch to align the anatomy to be treated at the machine
center.
WARNING: Before using a bolster , make sure the matching RoboCouch cushions were
used during pretreatment scanning. Otherwise, the patient treatment position will not
match the pretreatment scanning position, and mistreatment can result.
For more information on immobilization devices and pretreatment scanning, see "Chapter
6: Patient Preparation".
Follow the instructions to move the RoboCouch System to the treatment position as
described in the section, “Using the RoboCouch System to Load the Patient”, in "Chapter
7: Treatment Room Operations".
The treatment position corresponds to the initial alignment position. In this position, the
anatomy to be treated is in the field of view of the X-ray imaging system. The RoboCouch
System moves to either the previous or default alignment position, depending on whether
this is the first treatment session or a subsequent treatment for the treatment plan.
2. Follow the instructions to adjust the headrest, secure the thermoplastic mask (if needed),
and extend the footrest (if needed), as described in the section, “Using the RoboCouch
System to Load the Patient”, in "Chapter 7: Treatment Room Operations".
3. If a position measurement device or alignment lasers are available in the Treatment
Room, you can use them to help fine tune the initial alignment position. Using the Hand
Controller, move the RoboCouch System to align the anatomy to be treated at the
machine center.
In the Alignment > Couch phase, the following tasks are performed:
WARNING: Make sure no one but the patient is in the Treatment Room when Live X-ray
images are acquired.
For Synchrony Spine Tracking treatments, serial exposure is used to reduce scatter: X-ray images
are taken one at a time, first using one x-ray source and then the other, unless the Respiratory
Modeling System has been enabled.
If the Respiratory Modeling System is enabled, and for every other tracking mode, synchronous
exposure is used: X-ray images are taken using both X-ray sources simultaneously.
Figure 9 shows the left screen with acquired images, after the Live X-ray and DRR images have
been visually aligned using Couch Mode. The Live X-ray images are overlaid on the DRR images
and displayed in a teal color scale.
Using the default X-ray parameters is recommended. For Synchrony Fiducial Tracking mode, you
may need to adjust X-ray parameters for XRS A (X-ray source A) and XRS B (X-ray source B)
using the dropdown lists (see Figure 10):
WARNING: Use the minimum X-ray technique that meets all other tracking requirements.
Overexposure of X-rays can cause skin burns.
When you expand the kV dropdown list, available values are displayed in groups of 10. Click on
a group to select values in increments of 1 kV. You can also adjust mA and ms values by
selecting from available values in the dropdown lists.
The tracking algorithm does not provide feedback to the user in the Alignment > Couch phase.
Algorithm errors are not displayed.
Controls are available for handling blocked imaging nodes and situations when the treatment robot
is stopped at a blocked node or between nodes. In the Alignment > Couch phase, these
controls are useful if treatment is interrupted at a blocked imaging node and you are confirming
alignment of the patient before resuming treatment. These controls are also available in the
Delivery phase. For more information on handling blocked imaging nodes during treatment
delivery, see “Visual Inspection of Tracking Results” on page 8-76.
WARNING: During initial patient setup, visually compare the DRR and Live X-ray images
to ensure that features in are similar in both. Otherwise, patient injury or death can occur.
Table 3 on page 8-32 describes the image panel controls available in some or all of the workflow
phases to aid image analysis and visual alignment. You can use the Blend Tool (the vertical slider
to the right of each image panel) to shift between displaying the Live X-ray or DRR images. In
Figure 9 on page 8-17, the Pan/Zoom controls and the Window/Level controls have also been
used to help visually identify skeletal features in the images.
After confirming acceptable visual alignment, you can proceed to the Alignment > Technique
phase, as described in “Optimizing the X-ray Technique” on page 8-22. Otherwise, you can move
the couch as needed until the Live X-ray and DRR images align according to your specifications,
as described below.
Move Couch
Button
Couch Mode allows you to shift the Live X-ray image relative to the DRR image so that anatomical
or other features match. To do this, select the Couch icon (below each image panel; see Table 3
on page 8-32). Then click and drag in the Live X-ray image. You do this separately for Image A and
Image B.
As the relative position of the images is adjusted, the target translational offsets of the treatment
couch in the Couch Position (mm) textboxes shown in Figure 11 are automatically updated.
(The Couch Position (deg) textboxes are not updated in the Alignment > Couch phase.)
The textboxes indicate the target position of the treatment couch at which the currently shown
overlay of Live X-ray images and DRR images would be achieved. Arrows above the textboxes
point in the positive direction. Values indicate couch movement relative to its current position.
To move the Live X-ray image back to its original, unshifted position, click the Reset icon.
To move the treatment couch to the target position indicated in the Couch Position (mm)
textboxes, click the Move Couch button. Values in the textboxes change and approach zero as
the desired treatment couch position is achieved.
WARNING: Making large movements of the treatment couch from the treatment delivery
computer outside the treatment room may increase the risk of equipment colliding with or
pinching the patient. For better visibility, make large movements of the treatment couch
inside the treatment room with the hand pendant.
After moving the treatment couch, you acquire a Live X-ray image and then analyze the image to
confirm visual alignment, as described above.
If you are unable to improve patient alignment using Couch Mode, visually inspect the patient
position and try manually aligning the patient using the treatment couch Hand Controller. For more
information, see “Manual Alignment Using the Hand Controller” on page 8-51.
Interlock Button
For more information on the Interlock button, see “Interrupting Treatment” on page 8-82. For
information on recoverable E-Stops, see “Error Handling” on page 8-84.
WARNING: The automatic patient positioning system does not detect when the patient or
the treatment couch is outside the Patient Safety Zone. After patient alignment or
realignment, ensure that the patient is within the Patient Safety Zone (see “Checking the
Patient Safety Zone” on page 8-58). Otherwise collision of the treatment robot with the
patient may result. For more information on the Patient Safety Zone, see "Chapter 2:
Safety".
WARNING: Make sure no one but the patient is in the Treatment Room when Live X-ray
images are acquired.
WARNING: The Live X-ray images should have the same image quality as the DRR
images. Otherwise, the tracking algorithm may not be able to identify and track features
correctly.
Figure 13 shows the Alignment > Technique phase (left screen) for Synchrony Skull
Tracking mode with the Live X-ray images from the Alignment > Couch phase.
The top row of images corresponds to XRS B (X-ray source B) and the bottom row to XRS A (X-
ray source A):
• Raw Image: The raw (unprocessed) Live X-ray image.
• Enhanced Image: The Live X-ray image, enhanced for contrast by the tracking
algorithm.
• DRR image: The DRR image for reference.
Table 3 on page 8-32 describes the image panel controls available in some or all of the workflow
phases to aid image analysis. The goal is to determine whether to adjust the X-ray technique to
allow adequate visualization, identification, and tracking of the features to be used for alignment
and tracking.
You can also analyze image details and compare image quality with previous images in the
Image History table on the right screen (see Figure 14). The goal is to determine the optimal
X-ray technique for XRS A and XRS B independently, by evaluating image quality and Brightness
values of the last image taken in the Alignment > Couch phase and images taken in the
Alignment > Technique phase.
The Image History Camera B (top) and Image History Camera A (bottom) tables show
image details. Tables are emptied if the treatment couch is moved. Details for the last image taken
in the Alignment > Couch phase are displayed. A new row is added with each image
acquisition. Controls let you do the following:
• Click an Image ID to highlight a row. Higher numbers are more recent. The corresponding
image in the pair is automatically selected in the other table, and the Enhanced Image is
displayed to the right of the table.
A green checkmark shows which rows match the current X-ray technique selected in the
dropdown lists on the right screen.
• Select controls at the bottom of the image panels to the right of each table to aid image
analysis (see Table 3 on page 8-32).
• Select either Enhanced Image or Raw Image for image display.
• Click a column heading to sort rows. Use the scroll bar on the right or the Up and Down
arrow keys on the keyboard to scroll through a table.
Select an X-ray technique that results in X-ray images that look similar to the DRR images and
Brightness values as close to 1 as possible.
To select an X-ray technique from a previous image:
• Click the Select Technique button (located between the two Image History
tables; see Figure 14) to automatically populate the XRS A and XRS B dropdown lists
with the X-ray techniques corresponding to the currently selected image pair. These will
become the new default settings. You do not need to acquire another image.
• To select X-ray techniques corresponding to images from two different image pairs, use
the XRS A and XRS B dropdown lists to manually select X-ray parameters (see
“Acquiring Live X-ray Images” on page 8-16). Then acquire a new image to make these
values the new default settings.
If you have selected a satisfactory X-ray technique, you can proceed to the Alignment > Align
phase. Otherwise, you can modify the X-ray technique as needed until you are satisfied with the
image quality, as described below.
WARNING: Use the minimum X-ray technique that meets all other tracking requirements.
Overexposure of X-rays can cause skin burns.
You may need to modify the X-ray parameters to enable adequate visualization, identification and
tracking of the features used for alignment and tracking:
• For Synchrony Skull Tracking mode: The goal is to achieve Live X-ray images that
look similar to the DRR images.
• For Synchrony Fiducial Tracking: The goal is to achieve Live X-ray images in which
the fiducials can be clearly seen against the background.
• For Synchrony Spine Tracking: See "Chapter 10: Synchrony Spine Tracking
(Option)", for information on setting X-ray parameters.
• For Synchrony Lung Tracking with Respiratory Modeling: See "Chapter 11:
Synchrony Lung Tracking with Respiratory Modeling (Options)", for information on
setting X-ray parameters.
• For Lung 1 View with Respiratory tracking: See "Chapter 15: Lung Optimized
Treatment (Option)", for information on setting X-ray parameters.
You can use the XRS A and XRS B dropdown lists to manually select X-ray parameters or the
Select Technique button to select an X-ray technique from a previous image acquisition. The
Imaging Interval is the minimum required time interval between X-ray images (in seconds).
For more information on using the dropdown lists, see “Acquiring Live X-ray Images” on page 8-16.
After changing the X-ray technique, you acquire new Live X-ray images and then evaluate the
image quality and Brightness value, as described above. Algorithm errors are not displayed in the
Alignment > Technique phase.
After finding the optimal X-ray technique, click the Align tab to proceed to the Alignment >
Align phase, where you perform patient alignment.
The goal of the Alignment > Align phase for the Synchrony Skull Tracking mode is to make
sure the patient is aligned within acceptable bounds for treatment delivery.
The existing Live X-ray images from the Alignment > Technique phase are automatically
correlated with the DRR images. During image correlation, the Synchrony Skull Tracking algorithm
attempts to identify the features to be tracked and calculates offsets between the Live X-ray and
DRR images. The calculated offsets are displayed in the Offsets panel. The offset values indicate
ho w much the treatment couch should be moved to align features in the Live X-ray and DRR
images.
Figure 15 shows the Alignment > Align phase (left screen) for Synchrony Skull Tracking mode
after image correlation.
The left screen displays the Live X-ray images in a teal color scale overlaid on the DRR image in
grayscale for Camera B and Camera A of the imaging system. The Acquire button is enabled,
allowing you to acquire Live X-ray images. The left screen includes the following additional controls
described in the sections that follow:
• Offsets panel: After image correlation, the Offsets panel displays suggested treatment
couch movements to position the patient in correct alignment (see Figure 16 on
page 8-30). During patient alignment, you use the calculated offset values to move the
treatment couch so that features in the Live X-ray and DRR images are aligned. For more
information, see “Evaluating Calculated Offsets” on page 8-29.
• Patient position controls: Automatic patient positioning controls indicate desired motion of
the treatment couch relative to its current position (see Figure 22 on page 8-48). The
Move Couch button moves the treatment couch by this amount. The controls also
show the current absolute rotational position of the treatment couch. For more
information, see “Moving the Couch in the Alignment > Align Phase” on page 8-48.
• Algorithm parameter slider bars: Slider bars allow you to monitor the calculated value of
algorithm parameters and change the value of their thresholds (see Figure 19 on
page 8-42). The algorithm parameters depend on the tracking mode. Calculated values
are updated each time an image is acquired and image correlation is performed. When a
calculated value exceeds its threshold, a threshold violation occurs. For more
information, see “Monitoring and Adjusting Algorithm Parameters” on page 8-41.
• Rotational Bounds Checking Enabled checkbox: When this checkbox is
enabled, an interlock is triggered if patient rotational displacements computed by the
tracking algorithm exceed the maximum allowed rotational displacements. This checkbox
is enabled by default. For more information, see “Checking Readiness for Treatment
Delivery” on page 8-56.
The Alignment > Align phase has the following steps:
• “Analyzing Imaging Results” on page 8-29: You start by evaluating the calculated offset
values to determine if you need to move the treatment couch. You also review the Live X-
ray images to verify that there are sufficient skeletal features in the images and that
anatomy in the Live X-ray and DRR images matches.
• “Moving the Couch in the Alignment > Align Phase” on page 8-48: You move the
treatment couch as needed and acquire new Live X-ray images.
• (Optional) Use of the Physician Setup Approval button which allows a physician to
electronically approve the current patient alignment in the Alignment > Align phase
(treatments that do not use Respiratory Modeling) and the Alignment > Respiratory phase
(treatments that use Respiratory Modeling). For more information on Physician Setup
Approval, see “Physician Setup Approval of Patient Alignment” on page 8-52.
• (Optional) “Requesting Remote Approval of Patient Alignment (TxView Option)” on
page 8-53. With the TxView licensed option, you can remotely approve patient alignment
in the Alignment > Align phase (treatments that do not use Respiratory Modeling) and the
Alignment > Respiratory phase (treatments that use Respiratory Modeling).
Depending on the tracking mode, calculated offset values are displayed for translation only, or for
both translation and rotation:
• Synchrony Skull Tracking mode: Computed values are displayed for the 3 translations
(Left/Right, Anterior/Posterior, and Superior/Inferior directions) and 3 rotations (Left/Right
roll, Head-Up/Head-Down pitch, and Clockwise/Counterclockwise yaw).
NOTE: For trigeminal paths, the treatment robot does not correct
for rotations. All rotations are automatically disabled. However,
the Couch Corrections text fields still display rotational
offsets for the purpose of monitoring.
• Synchrony Fiducial Tracking mode: When 2 or fewer fiducials are used for tracking,
only computed values for the 3 translations are displayed. When 3 or more fiducials are
used for tracking, computed values are also displayed for the three rotations. For more
information, see "Chapter 9: Synchrony Fiducial Tracking".
• Synchrony Spine Tracking modes: Computed values are displayed for the 3
translations and 3 rotations. For more information, see "Chapter 10: Synchrony Spine
Tracking (Option)".
• Synchrony Lung Tracking with Respiratory Modeling mode: Computed values are
displayed for the 3 translations only. The values displayed for the 3 rotations are constant
values obtained from Synchrony Spine Tracking alignment. For more information, see
"Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)".
• Lung 1 View Tracking with Respiratory Modeling mode: Computed values are
displayed for the 3 translations only. The values displayed for the 3 rotations are constant
values obtained from Synchrony Spine Tracking alignment. For more information, see
"Chapter 15: Lung Optimized Treatment (Option)".
Evaluate the calculated offset values. They indicate couch movement needed to align features in
the Live X-ray and DRR images. The Live X-ray images are automatically shifted relative to the
DRR images using the calculated translational offsets. The overlaid Live X-ray and DRR images
show you how these images would align if the treatment couch is moved using the translational
offsets.
To reset the Live X-ray image displays to their original, unshifted positions, click the Reset icon
in Couch Mode (see Table 3 on page 8-32). You can then shift the Live X-ray images and DRR
images manually relative to each other to determine whether anatomical or other features are
aligned. Click the Offset icon in Couch Mode to shift the Live X-ray images again by the
translational offsets. For more information on moving the treatment couch, see “Moving the Couch
in the Alignment > Align Phase” on page 8-48.
Steps depend on the alignment scenario:
• Offsets are within tolerances required for treatment delivery, and you can verify that there
are sufficient skeletal features in the images and that anatomy in the Live X-ray and DRR
images matches.
You can proceed to the next treatment phase, the Readiness phase.
To reduce offsets, you can move the treatment couch using the Move Couch
button. See “Moving the Couch in the Alignment > Align Phase” on page 8-48.
• An orange box around an offset value indicates an offset is not within required tolerances
and an out-of-bounds error has occurred. For more information on the required
tolerances, see “Displacement Corrections During Treatment Delivery” on page 8-62. An
orange box around a tracking algorithm slider bar indicates that a threshold violation has
occurred.
You can use image analysis tools in Table 3 on page 8-32 to make sure the images
include sufficient skeletal features. Check that anatomy is aligned in the Live X-ray
and DRR images.
To reduce offsets, you can move the treatment couch using the Move Couch
button. See “Moving the Couch in the Alignment > Align Phase” on page 8-48.
You can also adjust algorithm parameters as needed. See “Monitoring and Adjusting
Algorithm Parameters” on page 8-41
Control Description
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Couch: Adjust the position of the Live X-ray image relative to the DRR image by
clicking and dragging in either the Live X-ray image or DRR image. See “Visually
Aligning the Patient” on page 8-15. Depending on the treatment phase and the
tracking mode, the following icons are displayed:
• Click the Reset icon to move the Live X-ray image back to its original,
unshifted position.
• Click the Reset icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with zero values. The Live X-ray image moves
to its unshifted position, to show existing alignment.
• Click the Offsets icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with the calculated offset values. The Live X-ray
image shifts by the translational offsets, to show the alignment that would
result if the couch is moved by these amounts.
You can toggle back and forth between these icons. The icons only shift the Live
X-ray image relative to the DRR image. They do not move the couch.
Control Description
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only. See “Using the Window/Level Controls” on
page 8-39.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
(Synchrony Fiducial Tracking mode only) You may need to modify the Window/
Level settings for the Live X-ray images to see the fiducials.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
ITTV Outline: (Lung 1 View Tracking with Respiratory Modeling only) If the
treatment plan is based on a Simulation Plan, displays a dark blue Internal Target
Tracking Volume (ITTV) contour in the DRR image that is not used for tracking.
For more information, see the Treatment Planning Manual.
The ITTV Outline is turned off by default. See "Chapter 15: Lung Optimized
Treatment (Option)".
Control Description
Tumor Outline: For Synchrony Lung Tracking with Respiratory Modeling mode
and Lung 1 View Tracking with Respiratory Modeling; lower right of image panel)
Displays a graphic overlay of the tumor outline (Target Tracking Volume, TTV) in
the DRR images and Live X-ray images that corresponds to the tumor contour
specified during treatment planning. With Lung 1 View Tracking with Respiratory
Modeling, the tumor outline is only displayed in the Live X-ray image that is used
for tracking.
• A dark blue outline indicates the reference tumor outline in the DRR images.
• A teal blue outline indicates where the tumor has been identified by the
tracking algorithm in the Live X-ray images.
• Red and green outlines indicate the reference tumor outline projected onto the
Live X-ray image in Offset Mode.
The Tumor Outline is turned on by default. See "Chapter 11: Synchrony Lung
Tracking with Respiratory Modeling (Options)", and "Chapter 15: Lung Optimized
Treatment (Option)".
(For Synchrony Skull Tracking mode only; lower right of image panel) displays
target contours projected onto the DRR in the same colors as the target volumes
of interest VOIs created during treatment planning)
The Tumor Outline is turned off by default. Once selected, use the Blend Tool to
adjust the opacity of the DRR so the VOIs are visible.
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls. See “Using the
Crosshairs” on page 8-40.
Fiducial ROI: (Synchrony Fiducial Tracking mode only) Displays a square yellow
box in the Live X-ray image for each Region of Interest (ROI) where the fiducial
extraction algorithm searches for fiducial candidates. Each box is centered on a
fiducial candidate.
The size of the box changes when the Tracking Range (mm) parameter is
adjusted. For more information, see "Chapter 9: Synchrony Fiducial Tracking".
Control Description
Align Center: (Synchrony Spine Tracking modes only) Displays a marker that
corresponds to the align center selected during treatment planning. See "Chapter
10: Synchrony Spine Tracking (Option)".
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Control Description
Full Content DRR: (Synchrony Spine Tracking modes only) Displays the
unsegmented (full content) DRR images. See "Chapter 10: Synchrony Spine
Tracking (Option)".
Full Content DRR: (Synchrony Lung Tracking with Respiratory Modeling mode
and Lung 1 View Tracking with Respiratory Modeling only) Displays the
unsegmented (full content) DRR images. See "Chapter 11: Synchrony Lung
Tracking with Respiratory Modeling (Options)", and "Chapter 15: Lung Optimized
Treatment (Option)".
Tumor Region DRR: (Synchrony Lung Tracking with Respiratory Modeling mode
and Lung 1 View Tracking with Respiratory Modeling mode only; upper right
corner of image panel) Displays the tumor region DRR images, in which all image
content that is not in the immediate region of the tumor is removed. See "Chapter
11: Synchrony Lung Tracking with Respiratory Modeling (Options)", and "Chapter
15: Lung Optimized Treatment (Option)"
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
• To set an image back to its default settings, click the Reset icon. If window and level
settings have been locked for all images in a row, clicking this icon applies to all the
images.
identified as the actual tumor. The outline remains red as you drag it. Depending on the
movement, the tumor outline in the opposing image may also move.
4. For Synchrony Fiducial Tracking mode, let go of the reference group of circle markers.
Both Live X-ray images are automatically correlated. The translational offsets determined
by moving the reference configuration of fiducials are input to the fiducial tracking
algorithm. Upon completion of the automatic correlation, green circle markers indicate the
new location of candidate fiducials in the Live X-ray image. Updated offset values are
displayed.
For Synchrony Lung Tracking with Respiratory Modeling mode or Lung 1 View Tracking
with Respiratory Modeling, let go of the tumor outline. It turns green while both Live X-ray
images are automatically correlated. The translational offsets determined by moving the
tumor outline are input to the Synchrony Lung Tracking with Respiratory Modeling
algorithm. Upon completion of the automatic correlation, a blue outline indicates the new
location of the tumor in the Live X-ray image. Updated offset values are displayed.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
These calculated values are also displayed to the left of slider bars (left screen), where you can
adjust their thresholds (see Figure 19). If a parameter value exceeds its threshold (the red error
bar extends beyond the threshold cursor), a threshold violation occurs, and an orange box is
displayed around the slider bar. When a threshold violation occurs the first time you are in the
Alignment > Align phase during a treatment fraction, you cannot move to the next phase until
the threshold violation is resolved.
Calculated parameters for Synchrony Skull Tracking mode are described below.
• “Max Brightness Error (%)” on page 8-43
• “Max Gradient Error (%)” on page 8-44
For information on algorithm parameters for other tracking modes, see the following chapters:
dxAB (mm)
The dxAB (mm) parameter gives the relative distance along the Inferior/Superior axis of the align
center in the two projections (for Camera A and Camera B of the X-ray imaging system). This
difference is called dxAB. The Synchrony Skull Tracking algorithm calculates the dxAB value for
every image acquisition. When the calculated dxAB value exceeds the dxAB threshold value, a
dxAB error occurs. The Inferior/Superior axis is used because this axis is common to both
projections.
The default value of the dxAB threshold parameter is 2.5 mm. The value of this threshold should
be small. If you change this threshold value during patient alignment, you should ideally return to
the default value when patient alignment is completed.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Skull Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before changing this parameter value.
The Max Brightness Error (%) threshold parameter sets the allowed range of the overall intensity
of Live X-ray images relative to the DRR images. This threshold parameter applies to the Live X-
ray images after they have been acquired and processed. Ideally, the calculated Brightness (%)
ratio should be less than 2.0%. The default value of the Max Brightness Error (%) threshold
parameter is 5.0%.
Adjustment of the Max Brightness Error (%) threshold parameter should only be made when it is
not possible to improve the quality of the Live X-ray images relative to the DRR images by
modifying X-ray parameters in the Alignment > Technique phase.
WARNING: If you raise the value of the Max Brightness Error (%) threshold parameter, it is
possible for the Synchrony Skull Tracking algorithm to result in larger targeting errors.
WARNING: Pediatric, elderly, or other patients with low bone density may be exposed to
increased imaging dose while compensating adjustments are made to imaging technique.
WARNING: If you raise the value of the Gradient Error (%) threshold parameter, it is
possible for the Synchrony Skull Tracking algorithm to result in larger targeting errors.
Slider bars on the left screen allows you to monitor and change the values of the following algorithm
parameter thresholds (see Figure 19 on page 8-42):
• dxAB (mm) threshold
• Max Brightness Error (%) threshold
• Max Gradient Error (%) threshold
For information on these parameters, see “Calculated Algorithm Parameters” on page 8-42.
Threshold violations are indicated when the red error bar extends beyond the adjustable threshold
cursor, and an orange box is displayed around the slider bar.
• To return to the default value, click the Reset icon. Alternatively, click and drag the
threshold cursor to the vertical line on the slider bar. The vertical line indicates the default
value.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Skull Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before changing this parameter value.
WARNING: If you raise the value of the Maximum Brightness Error (%) threshold
parameter, it is possible for the Synchrony Skull Tracking algorithm to result in larger
targeting errors.
WARNING: If you raise the value of the Maximum Gradient Error (%) threshold parameter,
it is possible for the Synchrony Skull Tracking algorithm to result in larger targeting
errors.
Click a column heading to sort rows. Use the scroll bar on the right or the Up and Down arrow keys
on the keyboard to scroll through the table.
Click on a row to highlight it and select a tab to view the following details:
• Acquisition Details: Displays acquisition time, X-ray technique, and data specific to
the tracking mode.
• Calculated Translation (mm) and Calculated Rotation (deg): The offsets
calculated by the tracking algorithm that were previously displayed in the Offsets panel.
• Absolute Couch Position (mm) and Absolute Couch Rotation (deg): The
absolute translational and rotational position of the couch for reference.
• Images tab: Displays the Live X-ray images. Select controls at the bottom of the image
to aid image analysis (see Table 3 on page 8-32).
• Calculated algorithm parameters: See “Calculated Algorithm Parameters” on page 8-42.
Depending on the tracking mode, additional tabs may be available. For information on the Image
History table for the Respiratory phase, see "Chapter 12: Synchrony Respiratory Tracking
System (Option)".
Move Couch
Button
Figure 22 Patient positioning controls in the Alignment >
Align phase (shown for standard treatment couch)
NOTE: The textbox that controls yaw rotation is disabled for the
standard treatment couch.
Offset values calculated after image acquisition and image correlation automatically fill the Couch
Position (mm) and Couch Rotation (deg) textboxes.
Arrows above the textboxes point in the positive direction. Values in the textboxes indicate couch
movement relative to its current position. Values above the rotation textboxes show the absolute
rotational position of the couch.
Depending on tracking mode, values are displayed for translations only or for both translations and
rotations.
When the Rotational Bounds Checking Enabled checkbox is enabled, an interlock is
triggered if patient rotational displacements computed by the tracking algorithm exceed the
maximum allowed rotational displacements. This checkbox is enabled by default. For more
information, see “Checking Readiness for Treatment Delivery” on page 8-56.
In the Alignment > Align phase, high initial translational offsets can lead to false rotational
values. If calculated offset values for translation and rotation are high, disable rotational movement
initially, so that only translational adjustments of the treatment couch are made. Then acquire
another image to obtain updated offsets.
If the values in the Couch Position (mm) textboxes are beyond the physical limits of treatment
couch motion, an error occurs and the Error Handling System screen is displayed. If this occurs,
you can try manually aligning the patient using the treatment couch Hand Controller. For
information, see “Manual Alignment Using the Hand Controller” on page 8-51.
Click a Lock icon to prevent a rotation axis from changing during automatic movement. Click
again to unlock it.
Click the Move Couch button to move the couch to the target position. Values in the textboxes
change and approach zero as the desired couch position is achieved. Slider bars for adjusting
algorithm parameters thresholds turn gray when the couch is moved, since previously calculated
parameter values are no longer relevant.
WARNING: The automatic patient positioning system does not detect when the patient or
the treatment couch is outside the Patient Safety Zone. After patient alignment or
realignment, ensure that the patient is within the Patient Safety Zone. Otherwise collision
of the treatment robot with the patient may result. For more information on the Patient
Safety Zone, see "Chapter 2: Safety".
You can also use Couch Mode to shift the Live X-ray image relative to the DRR image so that
anatomical or other features match. For more information, see “Using Couch Mode” on page 8-19.
In Couch Mode, to move the Live X-ray image back to its original, unshifted position, to show
existing alignment, click the Reset icon. Clicking this icon fills the Couch Position (mm)
and Couch Rotation (deg) textboxes with zero values.
In the Alignment > Align phase, you can also click the Offsets icon to fill the Couch
Position (mm) and Couch Rotation (deg) textboxes with the calculated offset values. The
Live X-ray image shifts by the translational offsets, to show the alignment that would result if the
couch is moved by these amounts.
To move the treatment couch to the target position indicated in the Couch Position (mm)
textboxes in Couch Mode, click the Move Couch button. Values in the textboxes change and
approach zero as the desired treatment couch position is achieved.
After moving the treatment couch, you acquire a Live X-ray image and then analyze the image to
confirm visual alignment, as described above.
If you are unable to improve patient alignment using Couch Mode, visually inspect the patient
position and try manually aligning the patient using the treatment couch Hand Controller. For more
information, see “Manual Alignment Using the Hand Controller” on page 8-51.
1. For initial patient alignment, perform the steps described in “Initial Alignment of the
Treatment Couch” on page 8-12. Visually align the anatomy to be tracked at the machine
center.
2. Exit the Treatment Room and close the Treatment Room door.
3. On the treatment delivery computer, click the Acquire button in the Alignment >
Align phase.
WARNING: Make sure no one but the patient is in the Treatment Room when Live X-ray
images are acquired.
WARNING: Ensure that the skeletal features, fiducials, or target tumor to be tracked is
correctly identified in the Live X-ray images by the tracking algorithm. If the features are
not correctly identified, see “Analyzing Imaging Results” on page 8-29 for troubleshooting
steps.
4. (Standard treatment couch only) If offset values are displayed in the Offsets panel, use
the Hand Controller to move the patient by the displayed amounts.
To do this, record the translational offset values displayed in the top 3 Offsets text fields
in the Alignment > Align phase (see Figure 16 on page 8-30).
Then re-enter the Treatment Room and use the Hand Controller to move the treatment
couch by the amount and in the translation directions that you recorded.
NOTE: If Offsets values are displayed, you may also use the
automatic patient positioning controls to move the treatment
couch by the displayed amounts.
5. If Offsets values are not displayed, try the following methods to align the patient:
• Enter the Treatment Room. Visually inspect the patient position and try manually
aligning the patient using the treatment couch Hand Controller. Visually verify that the
anatomy to be tracked is aligned at the machine center.
• (Standard treatment couch only) Use Couch Mode to determine how far to move the
treatment couch. Record the translational offset values displayed in the Couch
Position (mm) text boxes.
Then re-enter the Treatment Room and use the Hand Controller to move the
treatment couch by the amount and in the translation directions that you recorded.
For treatments that do not use Respiratory Modeling, once the patient has been aligned the
Physician Setup Approval button is available in the Alignment > Align phase.
For treatments that use Respiratory Modeling, once a valid Respiratory model is built, the
Physician Setup Approval button is available in the Alignment > Respiratory phase
• Click the Physician Setup Approval button and a log on dialog appears.
• Enter your credentials and click OK. The Physician Setup Approval button changes to
Approved with a green checkmark, and its tool tip displays the recorded date, time, and
approver.
The images and correlation data associated with the current patient alignment are
flagged for review.
• For treatments that use Respiratory Modeling, click the Remote Approval button in the
Alignment > Respiratory phase.
The current Respiratory model, including all images and correlation data, is flagged for
review.
Table 5 shows how the button changes with request status.
If the approving Physician does not confirm the patient alignment data (or the Respiratory model
for treatments that use Respiratory Modeling), the Remote Approval button remains in the
Pending state. In this case, proceed as per your established procedures. Direct contact with the
approving Physician by phone or other means for further consultation may be required.
If the request for approval is in the Pending (Not Current) or Confirmed (Not Current) state (see
Table 5), you can send another request at any time. The state of the Remote Approval button
applies to the most recent request.
After receiving confirmation of patient alignment from the approving Physician, you can proceed to
the next treatment phase.
NOTE: You can proceed to the next treatment phase even when
the approval request is not in the Confirmed state.
Button Description
Enabled: Remote Approval is enabled, the patient alignment is valid, and the
Respiratory model (for treatments that use Respiratory Modeling) is valid.
Remote Approval is enabled only in Treatment mode, in the Alignment >
Align phase (treatments that do not use Respiratory Modeling) and the
Alignment > Respiratory phase (treatments that use Respiratory Modeling).
Pending: A request for Remote Approval was sent and is pending approval
in TxView.
Button Description
Pending (Not Current): The pending request for Remote Approval is not
current. Since the request was sent, the patient alignment data changed in
one or more of the following ways:
A new image was acquired.
An algorithm input parameter was changed, resulting in image recorrelation.
The Respiratory model (for treatments that use Respiratory Modeling) was
updated by adding or recorrelating an image.
For treatments that use the Synchrony Fiducial Tracking with InTempo Imaging, the left screen
also includes an InTempo Settings panel. For information on the Synchrony Fiducial Tracking
with InTempo Imaging, see "Chapter 13: Synchrony Fiducial Tracking with InTempo Imaging
(Option)".
For treatments that use the SynchronyTracking System, the left screen also includes an Image
Burst Enabled checkbox, and text indicates whether the SynchronyTracking System has been
skipped. For more information on the SynchronyTracking System, see "Chapter 12: Synchrony
Respiratory Tracking System (Option)".
Figure 25 shows the right screen of the Readiness phase. The screen includes the Collimator
and Laser Alignment Check panel, where you make sure the correct collimator is installed.
It also includes the Proximity Detection Status panel, where you perform a Patient Safety
Zone Check.
WARNING: The PDP model is based on the Head First Supine (HFS) patient position.
Ensure that the patient and all devices are within the Patient Safety Zone defined by the
PDP model, even if the patient is not in the HFS position. For more information on the
Patient Safety Zone, see "Chapter 2: Safety".
For more information on the dimensions of the Patient Safety Zone and required measurements
before treatment delivery, see the section "The Patient Safety Zone," in "Chapter 2: Safety".
Always visually check the patient to confirm measurements. If you need to realign the patient
during the treatment process, it is recommended that you enter the Treatment Room and perform
a visual check of the patient position, following the patient safety guidelines in "Chapter 2: Safety".
The patient size sets the dimensions of the dynamic safety zone for treatment. You can select the
following patient sizes:
• Large: Based on size data for a 99 percentile male.
• Medium: Based on size data for a 75 percentile male.
• Small: Based on size data for a 50 percentile female.
WARNING: Choose a size that is as large as or larger than the patient. Otherwise,
collision with the treatment robot is possible which can result in serious injury or death of
the patient.
The Proximity Detection Status panel also includes a circular indicator that shows whether
the dynamic safety zone is positioned within the fixed safety zone. The circular indicator is updated
continually based on the current position of the treatment couch. The circular indicator displays the
following colors:
• Green: The dynamic safety zone is completely within the fixed safety zone. No PDP error
message is displayed.
• Orange: Some part of the dynamic safety zone is outside the fixed safety zone. There is
a higher risk of PDP errors occurring during treatment delivery. Before treatment can
commence, a confirmation message is displayed in the Delivery Warning
Messages panel on the left screen warning that the treatment couch is out of range.
To ensure that the patient is positioned within the Patient Safety Zone for treatment, follow
these steps:
For information on Patient Safety Zone dimensions, see "Chapter 2: Safety".
1. Ensure that the most anterior height of the patient is within the dynamic safety zone that
you selected based on patient size.
2. Ensure that the patient is within the fixed safety zone for head or body treatment, based
on the anatomy chosen during treatment planning.
3. In either the Alignment > Couch phase or the Readiness phase, in the Patient
Size dropdown list of the Proximity Detection Status panel, select the patient
model that is appropriate for the size of the patient (Large, Medium, or Small).
WARNING: Choose a size that is as large as or larger than the patient. Otherwise,
collision with the treatment robot is possible which can result in serious injury or death of
the patient.
WARNING: Position the patient and all devices within the Patient Safety Zone defined by
the Proximity Detection Program (PDP) model. Anything extending outside the boundaries
of the PDP model increases the risk of collision with the treatment robot and can cause
serious injury or harm to the patient.
WARNING: Do NOT place pillows or other objects in an area that could cause Patient
Safety Zone violations. Do NOT position patient extremities so that safety zone violations
occur, or else the patient may be seriously injured or killed.
For examples of actions that may cause Patient Safety Zone violations, see "Chapter 2: Safety".
Translational Corrections
The maximum corrections that the treatment robot can make along each translation axis
(X, Y, and Z) are shown in Table 6. An E-Stop is triggered if any translation displacement exceeds
the values shown below.
The maximum translational correction depends on whether the CyberKnife System uses the
standard treatment couch or the optional RoboCouch System. It also depends on whether the
treatment uses the SynchronyTracking System.
Rotational Corrections
The maximum corrections that the treatment robot can make for rotation (roll, pitch, or yaw) before
an E-Stop is triggered depend on the following:
• whether the standard treatment couch or optional RoboCouch System is used.
• the tracking mode.
• the path template.
• whether Respiratory Modeling mode is enabled.
• whether rotational corrections are disabled.
The maximum allowed rotation displacements before an E-Stop is triggered are shown in Table 7
(for the standard treatment couch) and Table 8 (for the RoboCouch System).
Table 9 Controls for exchanging collimators and performing a Laser Alignment Check
Button Description
If your configuration includes the optional Iris Variable Aperture Collimator, a wait period allows the
Iris Collimator to achieve the operating temperature set by the active temperature control system.
The typical wait time after the Iris Collimator is picked up from the Xchange collimator table is five
minutes.
The Warning Message Status area of the screen at the right of the message panel will
display an orange exclamation point if issues are found.
Each warning message must be read, understood, and when applicable, resolved before
confirming the message and proceeding to the Delivery phase.
To confirm a Warning message in the Delivery Warning Messages panel:
1. Highlight the row containing the Warning message heading.
2. Review the Warning description and, if applicable, resolve the conditions that generated
the warning.
3. Click the Confirm button to confirm you have understood and, if applicable, resolved
the condition.
4. If there is more than one Warning message, highlight the next row. Then review and
confirm the Warning message as described above.
5. To review a previously confirmed Warning message, click the row containing its message
heading.
When all Warning messages have been confirmed, the Warning Message Status area of the
screen displays a green checkmark.
When the Warning Message Status and Collimation System Status areas both
display green checkmarks, you are ready to start treatment delivery. Click the Delivery tab to go
to the Delivery phase.
For information on the optional TxView remote approval workflow, see “Requesting Remote
Approval of Patient Alignment (TxView Option)” on page 8-53.
For information on the optional Physician Setup Approval of Patient Alignment workflow, see
“Physician Setup Approval of Patient Alignment” on page 8-52.
Treatment Delivery
This section covers the following topics:
• “Starting Treatment Delivery” on page 8-71
• “Turning on the High Voltage” on page 8-72
• “Monitoring Treatment Delivery” on page 8-73
Before starting treatment delivery, perform a visual check of the patient position in the Treatment
Room. Follow the patient safety guidelines in “Checking the Patient Safety Zone” on page 8-58.
WARNING: Make sure no one but the patient is in the Treatment Room while treatment is
in progress.
WARNING: Confirm that patient anatomy is correctly aligned before treatment begins.
WARNING: Before starting treatment delivery, print the Path List report from the Accuray
Precision System. During treatment delivery, check off each beam as it is delivered using
the lines available in the report. Monitor the Beam Data window and verify that the dose
and monitor units (MU) displayed on the screen match the report. For information on
printing the Path List report, see the Treatment Planning Manual.
WARNING: WARNING: When the safety interlock key is used to bypass the Treatment
Room door interlock and the Treatment Room door is open, potentially harmful LINAC
radiation exposure to operators can occur outside the Treatment Room. Ensure that no
LINAC radiation is delivered while the Treatment Room door interlock is bypassed and the
Treatment Room door is open.
The Delivery phase for Synchrony Skull Tracking mode is shown in Figure 32 (left screen) and
Figure 33 (right screen). Controls on these screens allow you to monitor and verify patient
alignment and tracking results during treatment delivery.
Figure 32 Delivery phase (left screen) for Synchrony Skull Tracking mode
Figure 33 Delivery phase (right screen) for Synchrony Skull Tracking mode
WARNING: Visually observe the patient at all times during treatment delivery. Watch
movement of the treatment robot on the Closed Circuit TV (CCTV) for potential collision. If
the treatment robot appears to be moving too close to the patient or to any other object in
the Treatment Room, or if the treatment robot moves in an unplanned manner, push the
nearest E-Stop button.
WARNING: WARNING: When the safety interlock key is used to bypass the Treatment
Room door interlock and the Treatment Room door is open, potentially harmful LINAC
radiation exposure to operators can occur outside the Treatment Room. Ensure that no
LINAC radiation is delivered while the Treatment Room door interlock is bypassed and the
Treatment Room door is open.
When the Start button is clicked, treatment delivery automation begins and the
Treatment Delivery System begins performing a complex sequence of actions. In
general, if new Live X-ray images are required, they are acquired automatically. Before
the first beam is delivered, a message is displayed prompting you to turn on the high
voltage to the LINAC (see Figure 35).
High Voltage
High Voltage ON and OFF LINAC Laser
Status Buttons Status
Emergency Stop
Button
Emergency Stop
Status
WARNING: During treatment delivery, do not attempt to use the Teach Pendant to move
the treatment robot unless instructed to by the Treatment Delivery System user interface.
Do not remove the Teach Pendant from the Treatment Room. The Teach Pendant should
only be used in the Treatment Room, in order to safely monitor operation of the treatment
robot.
WARNING: Monitor the patient continuously during treatment delivery using the CCTV
system and the Intercom System. Watch movement of the treatment robot for potential
collision. If the treatment robot appears to be moving too close to the patient or to any
other object in the Treatment Room, or if the treatment robot moves in an unplanned
manner, push the nearest E-Stop button.
For information on the Image Burst Interval parameter for treatments that use
Respiratory Modeling, see "Chapter 12: Synchrony Respiratory Tracking System
(Option)".
Text between the Treatment Monitoring graphs on the left screen indicates the
position of the treatment robot (also called the treatment manipulator) at the time the
images were acquired and the current position of the robot.
In the Images counter, the number of acquired images increases by 2 for each pair
of images.
(For treatments that do not use Respiratory Modeling) The calculated values in the
first column of the Offsets panel values are updated continually during treatment
delivery as new images are acquired.
For information about monitoring treatments that use Respiratory Modeling, see
"Chapter 12: Synchrony Respiratory Tracking System (Option)".
• Pointing of the LINAC is adjusted for the next beam and corrected for target
displacement. The second column of the Offsets panel displays the offset corrections
that are applied to the treatment robot.
Rotational corrections are not applied to the treatment robot if the Robot Rotation
Corrections Enabled checkbox is deselected. For Synchrony Spine Tracking with
Respiratory Modeling treatments, this checkbox is deselected by default and only
translational corrections are made during treatment delivery. For trigeminal paths
templates, all rotational corrections are disabled automatically. For more information on
displacement corrections during treatment delivery, see “Displacement Corrections
During Treatment Delivery” on page 8-62.
• The Beam’s Eye View panel of the right screen displays a diagram that simulates
adjustments to the collimator aperture and shape in real time during treatment.
• Radiation beam is delivered:
The beam Sequence ID (SID) number for the current beam at the current node is
displayed as the Beam ID number in the Delivery Progress panel (see
Figure 37).
While radiation is being delivered, the BEAM ON status indicator is lit on the operator
control panel (see Figure 36 on page 8-72) and the text BEAM is ON is displayed
above the Beam Delivery panel (and on the left screen below the Acquire
button).
When no radiation is being generated, the indicator on the operator control panel is
off and the text BEAM is OFF is displayed above the Beam Delivery panel (and
on the left screen below the Acquire button).
The cumulative monitor unit (MU) amount delivered for the current beam and fraction
increases in the Beam Delivery panel.
The elapsed time that treatment has been in progress increases during treatment
delivery in the Delivery Progress panel.
The estimated total time for delivery of the current fraction in the Delivery
Progress panel and the expected total number of images in the Images counter
are updated prior to each beam delivery. These estimates become increasingly
accurate as the end of a fraction is approached.
As treatment proceeds, text between the Treatment Monitoring graphs on the left screen displays
information on the movement of the treatment robot:
• When the treatment robot arrives at an unblocked node, the Node text field displays the
current node number in white.
• When the treatment robot arrives at a blocked node, the Node text field displays the
current node number in yellow and the text "Imaging is blocked" is displayed at the
bottom of the left screen (see Figure 38).
• When the treatment robot is moving to a node, the destination node is displayed in blue.
The text "Robot moving to" is displayed next it. This text may reflect movement from
the previous node position to the next node position, or from the previous beam position
back to a node position.
The iDMS Data Management System is updated after each treatment fraction is completed. Any
screen captures obtained manually (using the F12 function key or <Print Screen> key) and a
screen shot of the final Delivery phase screens at the end of each treatment fraction are
automatically saved in the iDMS Data Management System in the Plan Administration application
for that treatment plan.
For more information on these screen displays, see "Chapter 5: Treatment Delivery System
Overview". For information on the Synchrony Fiducial Tracking with InTempo Imaging, see
"Chapter 13: Synchrony Fiducial Tracking with InTempo Imaging (Option)".
• Visually verify the patient position and verify that the target has been correctly
identified by the tracking algorithm.
• Monitor Live X-ray images for any artifacts such as surgical implants.
If at any time it appears that the features to be tracked are not correctly located in the Live
X-ray images, pause treatment by clicking the Interlock button (see Figure 40 on
page 8-82) or push the nearest E-Stop button.
If it is determined by visual inspection of the Live X-ray images that the images are blocked, you
can mark the images as blocked.
Controls are available for handling blocked imaging nodes and situations when the treatment robot
is stopped at a blocked node or between nodes (see Table 10). These controls are also available
in the Alignment > Couch phase (see “Visually Aligning the Patient” on page 8-15).
• If treatment was interrupted at a blocked imaging node and you are confirming alignment
of the patient before resuming treatment.
• If treatment was exited at a blocked imaging node and you are beginning immediate
delivery of a makeup fraction.
Button Description
Figure 38 Pause button at blocked imaging node and images marked as blocked
For more information on visually verifying tracking results, see the following:
• “Performing Patient Alignment” on page 8-27
• "Chapter 9: Synchrony Fiducial Tracking"
• "Chapter 10: Synchrony Spine Tracking (Option)"
• "Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)"
• "Chapter 15: Lung Optimized Treatment (Option)"
WARNING: For Synchrony Fiducial Tracking mode, after acquisition of Live X-ray images,
you must visually verify the patient position and the fiducial locations in the images. If it
appears that the patient anatomy or fiducials are not correctly identified, check the
following:
1. If treatment delivery is in progress, click the Pause button on the left screen.
A Soft Stop is triggered.
2. Click the Mark Image as Blocked button.
The Live X-ray images are crossed-out on the screen (see Figure 38).
3. Click the Resume button.
NOTE: The treatment robot may move forward (in the forward
direction of the treatment path) or backward to reach the closest
unblocked node for imaging. The closest unblocked node may
be outside the treatment path.
For more information on marking nodes as blocked, see "Chapter 5: Treatment Delivery System
Overview".
WARNING: For all tracking modes, acquire Live X-ray images frequently enough during
treatment for proper tracking of patient movement. The appropriate imaging frequency
depends on patient anatomy and movement. Being unaware of patient movement for a
significant period of time can result in injury or death to the patient.
You can adjust the value of the Image Interval parameter in real-time during treatment delivery to
increase or decrease the maximum imaging frequency.
To adjust the Image Interval parameter:
• Move the Image Interval (sec) slider bar under the leftmost Treatment Monitoring
graph.
• Alternatively, enter a new value for the Image Interval parameter (in seconds) in the text
box to the right of the slider bar, and then press <Enter>.
When you adjust the Image Interval value, the expected total number of X-ray images for all
fractions in the treatment plan is updated in the Images counter.
For more information on the Image Interval parameter, see "Chapter 5: Treatment Delivery System
Overview".
For more information on treatment data graphs for other tracking modes, see the following
chapters:
• "Chapter 9: Synchrony Fiducial Tracking"
• "Chapter 10: Synchrony Spine Tracking (Option)"
• "Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)"
• "Chapter 15: Lung Optimized Treatment (Option)"
Interrupting Treatment
WARNING: If any situation has the appearance of an emergency, immediately press the
nearest Emergency Stop (E-Stop) button to stop movement of the treatment couch and
treatment robot and terminate radiation delivery. For more information on Emergency
Stop procedures, see "Chapter 2: Safety".
Interlock Button
When you click the Interlock button on the bottom of the left screen on the treatment delivery
computer, a recoverable software E-Stop is triggered (see Figure 40). Treatment couch
movement, treatment robot movement, X-ray image acquisition, and radiation delivery are
immediately terminated and the high voltage to the LINAC is turned off.
Interlock Button
For information on recovering from a recoverable E-Stop after clicking the Interlock button, see
“Error Handling” on page 8-84.
Pause
If a stop is not needed immediately, the Pause button (also referred to as the Soft Stop button)
on the left screen of the Delivery phase can be used to interrupt treatment (see Figure 41).
When you click the Pause button, treatment is interrupted at the next convenient opportunity. Any
in-progress treatment robot movement, X-ray image acquisition, or beam delivery is completed
before the Soft Stop is engaged.
WARNING: Do not use the Pause button if you need to immediately terminate radiation
delivery or stop treatment robot movement. If any situation has the appearance of an
emergency, immediately press the nearest Emergency Stop (E-Stop) button. For more
information on Emergency Stop procedures, see “Chapter 2: Safety” on page 2-1.
When a Soft Stop is engaged, the button is replaced by the treatment Resume button (see
Figure 42). Clicking the Pause button begins a 180 second countdown. The LINAC high voltage
remains on during the countdown. After 180 seconds, the LINAC high voltage is automatically
turned off.
To resume treatment after a Soft Stop, see “Patient Realignment After a Pause or Error” on
page 8-91.
Error Handling
This section covers the following topics:
• “Soft Stops” on page 8-84
• “E-Stops” on page 8-85
• “Resetting a Recoverable E-Stop” on page 8-86
• “Exiting Treatment After an E-Stop” on page 8-87
During treatment delivery, the following types of errors can occur:
• a Soft Stop (a non-emergency treatment interruption)
• a recoverable Emergency Stop (E-Stop)
• an unrecoverable E-Stop.
The type of error depends on what caused the problem.
Soft Stops
Soft Stops are triggered by imaging errors such as image algorithm errors or a patient out-of-
bounds error that occur in the Delivery phase. If a Soft Stop occurs, an error message with
information about the error(s) is displayed in the Messages panel of the Delivery phase (see
Figure 43).
The information includes any user action needed to resolve the error. (Another type of message
called a Delivery Inhibitor message is also displayed in the Messages panel. Delivery Inhibitor
messages report conditions that prevent you from starting treatment (the Start button on the left
screen is disabled) until the condition is resolved.)
If more than one error occurs, click the Previous or Next button in the Messages panel to view
their descriptions. When the errors are resolved, the Resume button on the left screen of the
Delivery phase is enabled. You should confirm patient alignment before resuming treatment.
See “Patient Realignment After a Pause or Error” on page 8-91.
For a list of errors that produce a Soft Stop, and a list of Delivery Inhibitor messages, see "Chapter
8: Treatment Delivery".
E-Stops
If an E-Stop occurs, the Error Handling System screen displays information about the error(s) that
led to the E-Stop (see Figure 44). If the error is recoverable, you can correct the error, click
Continue, and resume treatment. If the error is unrecoverable, you cannot continue and must
exit the treatment. For unrecoverable errors, contact Accuray Customer Support. For information
on the Error Handling System screen, see "Chapter 5: Treatment Delivery System Overview".
• If beam was delivered and the exit sequence followed a recoverable E-Stop, the
message also reports that a makeup fraction will be generated (see Figure 60).
To begin delivery of the makeup fraction, click Cancel to cancel exiting the
treatment application.
To return to the CyberKnife System Menu without delivering the makeup fraction,
click OK.
• If the exit sequence was the result of an unrecoverable E-Stop, CyberKnife System
communications are automatically reset.
While the reset is in progress, the message window of Figure 47 is displayed with the
Yes button disabled until the reset is successful. The text equipment power not
ready is also displayed at the bottom of the message window.
If the reset is successful, the Yes button is enabled. You can then click Yes to begin
delivery of the makeup fraction, or you can click No to return to the CyberKnife
System Menu.
If the reset is not successful (within a couple of minutes), the text equipment power
not ready persists and only the No button is enabled, allowing you to return to the
CyberKnife System Menu. The CyberKnife System is then automatically powered
down.
An E-Stop (Error 623) is triggered and treatment robot movement is stopped when the Contact
Detection Sensor comes into contact with an obstacle.
WARNING: Always pay close attention to the location of the treatment robot in
relationship to the patient and obstacles in the Treatment Room, even when the Contact
Detection Sensor is extended. If the treatment robot appears to be moving too close to the
patient or to any other object, push the nearest E-Stop button. Treatment robot collision
can result in serious injury or death to the patient.
NOTE: The treatment robot may move forward (in the forward
direction of the treatment path) or backward to reach the closest
unblocked node for imaging. The closest unblocked node may
be outside the treatment path.
Makeup Fractions
A makeup fraction is automatically created when a fraction of a patient treatment plan or a QA plan
is terminated before the total dose is delivered. For example, a makeup fraction is created when
an unrecoverable error occurs or treatment delivery is exited (see “Exiting Treatment After an E-
Stop” on page 8-87).
A makeup fraction consists of all paths of the terminated fraction that were not executed. If needed,
the makeup fraction also includes a makeup path similar to the path in which the termination
occurred. The makeup path contains the same set of nodes as the terminated path, except that
nodes where dose has been delivered have a zero dose and the node where the termination
occurred contains only the dose remaining.
If treatment is exited and you immediately begin delivery of the makeup fraction from the current
position of the treatment robot (without first moving the perch position):
• If treatment was exited at a blocked node, only the Move Robot to Imaging Node
button will be enabled in the Delivery phase. Click the Move Robot to Imaging
Node button to move the treatment robot to the closest unblocked node position for
image acquisition. After successful image correlation, you can click the Resume button.
The treatment robot then proceeds to the next node in the treatment path where beam is
to be delivered.
NOTE: The treatment robot may move forward (in the forward
direction of the treatment path) or backward to reach the closest
unblocked node for imaging. The closest unblocked node may
be outside the treatment path.
Phantom Mode
Phantom mode allows you to simulate patient treatment using a phantom to check the accuracy
and safe delivery of a patient treatment plan. To access this mode, click the Phantom button on
the CyberKnife System Menu and then log in. This button is enabled only when equipment power
is on.
Controls are the same as those in Treatment mode. In Phantom mode, however, you can only
select treatment plans created for phantoms. For information on specifying whether a treatment
plan applies to a patient or phantom, see the Data Management Manual.
A phantom treatment plan has unlimited deliverable fractions. When the last fraction in a phantom
treatment plan is completed, a new fraction is automatically created. To view the new fraction in
the Overview phase, click the Refresh button. In this way, you can continue treatment delivery
to the phantom without the need to create a new treatment plan.
For more information on Phantom mode, see "Chapter 5: Treatment Delivery System Overview".
Demonstration Mode
Demonstration mode allows you to demonstrate operation of the CyberKnife System by
performing a BB test. In this mode, no treatment radiation is delivered. Live X-ray images can be
acquired in the patient alignment phase using a phantom. However, no live X-ray images are
acquired during the robot motion phase. You have the option of bypassing the X-ray imaging
system in the patient alignment phase, allowing you to start the demonstration without acquiring
initial Live X-rays. The demonstration is performed using the Teach Pendant to manually move the
treatment robot.
Controls are the same as those in Treatment mode, except that you can bypass the X-ray
imaging system and the Secondary Collimator System. For more information on controls in
Demonstration mode, see "Chapter 5: Treatment Delivery System Overview".
BB Test
The BB test is used to demonstrate operation of the CyberKnife System. It is also used to perform
a visual qualitative check of the targeting accuracy of the treatment robot.
Required Phantoms
The following phantoms are required for the BB test:
• Skull phantom with external markers (also called BBs) on the surface of the phantom for
Synchrony Skull Tracking mode. The external marker (BB) on the surface of the phantom
is shown in Figure 49.
• A skull or body phantom with 3 or more fiducials and external markers (BBs) on the
surface of the phantom for Synchrony Fiducial Tracking mode.
External
Marker
BB Test Procedure
This section gives instructions for performing the BB test.
For treatment planning and DRR generation, follow these steps:
1. Select the appropriate phantom for the BB test.
2. Obtain a CT scan of the phantom and load it as a phantom CT image into the iDMS Data
Management System.
For more information on the iDMS Data Management System, see the Data Management
Manual.
3. In the Accuray Precision System, select the phantom and CT scan for the BB test. Select
Synchrony Fiducial Tracking mode. Mark fiducials as necessary.
4. Choose a 5 mm fixed collimator and locate the target as the one of the external markers
(BBs) on the surface of the phantom.
5. Generate a simple isocentric plan. For more information, see the Treatment Planning
Manual.
6. Save the treatment plan as a deliverable plan.
To load patient data for the demonstration:
For more information on loading patient data, see “Loading Patient and Treatment Data” on
page 7-2.
1. On the treatment delivery computer, click the Demonstration button on the
CyberKnife System Menu and then log in. This button is enabled only when equipment
power is on.
The Load Patient Data screen is displayed.
2. Select the phantom, the treatment plan generated for the BB test, and the fraction. Then
click the Load button.
The screens for the Overview phase are displayed.
3. Verify the patient, treatment, and fraction information and then select the three Data
Verified checkboxes in order from left to right.
After treatment files are downloaded, click the Alignment > Couch tab to go to the
Alignment > Couch phase.
4. Click the Technique tab to go to the Alignment > Technique phase. If nothing is
changed in the phase and you are satisfied with the Technique, move on to the
Alignment > Align phase. If you are not satisfied, go back to step 3.
5. Click the Align tab to go to the Alignment > Align phase. Adjust the treatment
couch position using the values in the Offsets panel.
If no offset values are displayed, visually assess the alignment of the DRR images and
the Live X-ray images.
6. Repeat the above process until the values displayed in the Offsets panel are 1 mm or
less for translations and 1 degree or less for rotations (if available).
To perform the BB test:
1. Click the to go to the Readiness phase. An equipment readiness check is performed.
When all equipment subsystems are ready, the Warning Message Status and
Collimation System Status text fields on the left screen both display green
checkmarks. For more information on the equipment readiness check, see “Checking
Equipment Readiness” on page 8-67.
2. Click the Delivery tab to go to the Delivery phase. Then click the Start button on the
left screen of the Delivery phase.
A message is displayed in the Messages panel on the right screen that instructs you to
use the treatment robot Teach Pendant to continue the demonstration (see Figure 50).
3. Enter the Treatment Room. Bypass the Treatment Room door E-Stop and enter the
Treatment Room.
To bypass the Treatment Room door E-Stop, remove the safety interlock key Enable
High Voltage key from the operator control panel and use it to bypass the Treatment
Room door interlock. Bypassing the Treatment Room door interlock allows you to open
the Treatment Room door without producing an E-Stop.
WARNING: When the safety interlock key is used to bypass the Treatment Room door
interlock, X-ray exposure to users and observers in the Treatment Room can occur.
Ensure that no X-ray images are taken while the Treatment Room door interlock is
bypassed. In Demonstration mode, when the Teach Pendant is used to initiate the robot
motion phase of the demonstration, no further X-ray images are taken.
WARNING: Always hold the Teach Pendant while performing a treatment simulation in
Demonstration mode, so that the Emergency Stop (E-Stop) button is readily available.
WARNING: Pay close attention at all times to treatment robot movement to ensure the
safety of observers. If the treatment robot appears to be moving too close to any person
or object in the Treatment Room, push the E-Stop button on the Teach Pendant to stop
movement of the treatment robot.
5. On the Teach Pendant, select the SIM NODE button again to move to the next node in
the treatment path.
6. Verify that the treatment robot stops at the next node and that the laser is pointing at the
external marker (BB) to within 1 mm.
Adjust the phantom position a few times during the BB test and continue:
For more information on pausing and resuming treatment, see “Interrupting Treatment” on
page 7-69. For information on patient realignment, see “Patient Realignment After a Pause or
Error” on page 7-81.
WARNING: Pay close attention at all times to treatment robot movement to ensure the
safety of observers. If the treatment robot appears to be moving too close to any person
or object in the Treatment Room, push the E-Stop button on the Teach Pendant to stop
movement of the treatment robot.
1. On the left screen of the Delivery phase, click the Interlock button at the bottom of
the screen (see Figure 12 on page 8-20).
2. Verify that a CyberKnife System E-Stop occurs and the Error Handling System window is
displayed. Verify that the treatment robot stops moving.
3. Click the Continue button. Verify that the screens of the Delivery phase are
displayed.
4. Enter the Treatment Room and adjust the phantom position by about +5 mm along the X
axis, for a coordinate system defined for a patient in the HFS position on the treatment
couch:
• X axis: Inferior(+)/Superior(-)
• Y axis: Left(+)/Right(-)
• Z axis: Anterior(+)/Posterior(-)
5. Exit the Treatment Room and close the Treatment Room door. Remove the safety
interlock key from the Treatment Room door and return it to the operator control panel.
6. Click the Pause button to activate the Acquire button, and then click Acquire.
WARNING: Ensure that no users or observers are in the Treatment Room when Live X-ray
images are acquired.
7. Verify that Live X-ray images are acquired and new offset values are calculated and
displayed in the Offsets panel text fields. Verify that these values are different from the
previous values by approximately the amount that the phantom position was adjusted.
8. Click the Resume on the left screen of the Delivery phase.
A message window is redisplayed instructing you to use the treatment robot Teach
Pendant to continue the BB test (see Figure 51 on page 8-97).
9. Bypass the Treatment Room door E-Stop and re-enter the Treatment Room.
To bypass the Treatment Room door E-Stop, remove the safety interlock key from the
operator control panel and use it to bypass the Treatment Room door interlock.
Bypassing the Treatment Room door interlock allows you to open the Treatment Room
door without producing an E-Stop.
WARNING: When the safety interlock key is used to bypass the Treatment Room door
interlock, X-ray exposure to users and observers in the Treatment Room can occur.
Ensure that no X-ray images are taken while the Treatment Room door interlock is
bypassed. In Demonstration mode, when the Teach Pendant is used to initiate the robot
motion phase of the demonstration, no further X-ray images are taken.
WARNING: Always hold the Teach Pendant while performing a treatment simulation in
Demonstration mode, so that the Emergency Stop (E-Stop) button is readily available.
NOTE: In Demonstration mode, when the Teach Pendant is used to initiate the robot
motion phase of the demonstration, no further X-ray images are taken.
10. On the treatment robot Teach Pendant select the SIM NODE button to move to the next
node in the treatment path.
11. Verify that the treatment robot moves to the next node and that the laser is pointing at the
external marker (BB) to within 1 mm.
12. Repeat the BB test with following patient position adjustments:
• ±5 mm in the X direction
• ±5 mm in the Y direction
• ±5 mm in the Z direction
• +5 mm in the X, Y, and Z directions
• -5 mm in the X, Y, and Z directions
13. When a sufficient number of nodes have been observed for pointing accuracy, select the
SIM ALL button on the Teach Pendant.
WARNING: Pay close attention at all times to treatment robot movement to ensure the
safety of observers. If the treatment robot appears to be moving too close to any person
or object in the Treatment Room, push the E-Stop button on the Teach Pendant to stop
movement of the treatment robot.
The treatment robot moves through the remaining nodes in the treatment path and then
returns to the perch position, without stopping for operator intervention.
14. At the end of the demonstration, remove the safety interlock key from the Treatment
Room door, return it to the operator control panel, and close the Treatment Room door.
Plan QA Mode
Plan QA mode allows you to select a quality assurance (QA) plan associated with a specific
patient and patient treatment plan and deliver the QA plan to a phantom. This mode provides a
method to perform quality assurance for patient-specific treatment plans.
WARNING: Plan QA should be performed prior to patient treatment delivery to verify that
the planned dose is consistent with the measured dose. Failure to perform plan QA can
result in patient mistreatment. For more information on the plan QA procedure, see the
Physics Essentials Guide.
Controls in Plan QA mode are the same as those in Treatment mode, except for the additional
QA plan information described in "Chapter 5: Treatment Delivery System Overview" and as
described below.
In Plan QA mode, you can only select QA plans for delivery. Only a single fraction is created for
a QA plan. If the fraction is partially delivered and treatment is aborted, a makeup fraction is created
that can be immediately delivered.
For information on creating a QA plan, see the Treatment Planning Manual. For information on
authorizing a QA plan and recording analysis results, see the Data Management Manual.
On the day of QA plan delivery, on the treatment delivery computer, power on the CyberKnife
System as described in "Chapter 4: System Startup and Shutdown".
Loading patient and treatment data includes the following steps:
• “Select the Patient, Treatment Plan, and QA Plan” on page 8-101
• “Verify Patient and Treatment Data for the QA Plan” on page 8-102
When this window appears, the first Data Verified checkbox is active. You must verify that
patient, treatment plan and QA plan, and treatment fraction information in the three columns on the
left screen are correct.
To verify that patient and treatment information for the QA plan is correct:
1. If the patient information is correct, click the first Data Verified checkbox. The second
Data Verified checkbox in the treatment information section becomes active.
If information is not correct, click the Interlock button at the bottom of the left screen,
followed by the Exit Treatment button on the Error Handling System screen. After
clicking OK, you will return to the CyberKnife System Menu.
2. If the treatment plan and QA plan information is correct, click the second Data Verified
checkbox. The third Data Verified checkbox in the treatment fraction information
section becomes active.
If information is not correct, click the Interlock button.
3. If the treatment fraction information is correct, click the third Data Verified checkbox.
If information is not correct, click the Interlock button.
For more information on the Overview phase, see "Chapter 5: Treatment Delivery System
Overview".
Treatment files, including treatment paths, DRR images, and other treatment data for the QA plan,
are downloaded from the treatment delivery computer to the Target Locating System (TLS)
computer. The initial alignment position for the treatment couch is also downloaded to the
treatment couch Hand Controller.
After treatment files have been successfully downloaded, click the Alignment > Couch tab to
go to the Alignment > Couch phase (see “Visually Aligning the Patient” on page 8-15).
Procedures for initial positioning of the phantom and phantom alignment are the same as those
described in “Initial Patient Positioning” on page 8-10 and “Performing Patient Alignment” on
page 8-27.
Introduction
This chapter describes how to use Synchrony Fiducial Tracking mode with the CyberKnife
Treatment Delivery System.
This chapter covers the following topics:
• “Overview of the Synchrony Fiducial Tracking System Algorithm” on page 9-2
• “Workflow for Synchrony Fiducial Tracking Mode” on page 9-4
• “Verifying Patient and Treatment Data” on page 9-5
• “Visually Aligning the Patient” on page 9-5
• “Optimizing the X-ray Technique” on page 9-6
• “Performing Patient Alignment” on page 9-8
• “Checking Readiness for Treatment Delivery” on page 9-27
• “Treatment Delivery” on page 9-28
This chapter assumes you are familiar with “Chapter 8: Treatment Delivery”.
2 3 2 3
(a) CT volume (b) Live X-ray image
For a configuration of 3 fiducials, there are 3 Rigid Body Error values, one for each pair of fiducials.
In this example, the Rigid Body Error value for each fiducial pair is determined from the following
quantities:
• F1-2|CT: distance between Fiducial 1 and 2 in the CT volume
• F1-3|CT: distance between Fiducial 1 and 3 in the CT volume
• F2-3|CT: distance between Fiducial 2 and 3 in the CT volume
and
• F1-2|Live: distance between Fiducial 1 and 2 in the Live X-ray images
• F1-3|Live: distance between Fiducial 1 and 3 in the Live X-ray images
• F2-3|Live: distance between Fiducial 2 and 3 in the Live X-ray images
The Rigid Body Error, RBE, for each pair of fiducials is the difference between the corresponding
distances in the CT volume and the Live X-ray images:
RBE1-2 = | F1-2|CT - F1-2|Live |
RBE1-3 = | F1-3|CT - F1-3|Live |
RBE2-3 = | F2-3|CT - F2-3|Live |
The Rigid Body (mm) threshold sets the maximum allowed value of the Rigid Body Error for each
pair of fiducials. If the value for any one pair of fiducials exceeds this threshold, an error occurs and
treatment is interrupted.
For information on responding to a Rigid Body Error and adjusting the threshold, see “Responding
to a Rigid Body Error” on page 9-23.
Figure 3 shows the workflow for patient alignment and treatment delivery when the Synchrony
Fiducial Tracking System is used with Respiratory Modeling. For information on using the
Synchrony tracking System, see “Chapter 12: Synchrony Respiratory Tracking System (Option)”.
Aligning the patient in Synchrony Fiducial Tracking mode requires ensuring that the fiducial
extraction algorithm has correctly located the fiducials. The patient position is then adjusted until
proper alignment is reached. In the Alignment phase, you perform the following steps:
• Visually aligning the patient in the Alignment > Couch phase.
• Optimizing the X-ray technique in the Alignment > Technique phase.
• Performing patient alignment in the Alignment > Align phase.
For more information on the general workflow for patient alignment and treatment delivery, see
“Chapter 8: Treatment Delivery”.
WARNING: The Live X-ray images should have the same image quality as the DRR
images. Otherwise, the tracking algorithm may not be able to identify and track features
correctly.
X-ray parameters should be set to achieve Live X-ray images that look similar to the DRR images.
For fiducial tracking treatments, set X-ray parameters so that fiducials can be clearly seen against
the background. For more information on setting X-ray parameters and acquiring Live X-ray
images, see “Chapter 8: Treatment Delivery”.
Figure 5 shows the Alignment > Technique phase (left screen) for fiducial tracking mode
after image acquisition.
If you are satisfied with the X-ray technique, you can proceed to the Alignment > Align phase.
Otherwise, you can change the X-ray technique as needed until you are satisfied with the image
quality, as described below. For more information on this phase, see “Chapter 8: Treatment
Delivery”.
The existing Live X-ray images from the Alignment > Technique phase are automatically
correlated with the DRR images. The fiducial extraction algorithm attempts to locate the fiducials
in the images. Once located, fiducials are marked in the Live X-ray images with a green circle
marker. The locations of the reference fiducials are marked with yellow diamond markers in the
DRR images.
If fiducial extraction was successful and candidates are found for all fiducials, the Offsets panel
displays suggested treatment couch movements to position the patient in correct alignment.
The Alignment > Align phase has the following steps:
• “Analyzing Imaging Results” on page 9-9: You start by evaluating the calculated offset
values to determine if you need to move the treatment couch. You also review the Live X-
ray images to ensure that the fiducials to be used for tracking are correctly identified.
• “Moving the Couch” on page 9-26: You move the treatment couch as needed and acquire
new Live X-ray images.
• (Optional) “Requesting Remote Approval of Patient Alignment (TxView Option)” on
page 8-53. With the TxView licensed option, you can remotely approve patient alignment
in the Alignment > Align phase (treatments that do not use Respiratory Modeling) and the
Alignment > Respiratory phase (treatments that use Respiratory Modeling).
For additional information on the Alignment > Align phase, see “Chapter 8: Treatment
Delivery”.
• An orange box around an offset value indicates an offset is not within required tolerances
and an out-of-bounds error has occurred. An orange box around a tracking algorithm
slider bar indicates that a threshold violation has occurred.
You can use image analysis tools in Table 1 on page 9-11 to make sure you can
visually identify the fiducials to be used for tracking. Check that fiducials are aligned
in the Live X-ray and DRR images.
To reduce offsets, you can move the treatment couch using the Move Couch
button. See “Moving the Couch” on page 9-26.
You can also adjust algorithm parameters as needed. See “Monitoring and Adjusting
Algorithm Parameters” on page 9-16.
For more information on procedures to help visually identify fiducials, see “Chapter 8: Treatment
Delivery”.
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory Model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Control Description
Couch: Adjust the position of the Live X-ray image relative to the DRR image by
clicking and dragging in either the Live X-ray image or DRR image. See “Moving
the Couch” on page 9-26. Depending on the treatment phase, the following icons
are displayed:
• Click the Reset icon to move the Live X-ray image back to its original,
unshifted position.
• Click the Reset icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with zero values. The Live X-ray image moves
to its unshifted position, to show existing alignment.
• Click the Offsets icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with the calculated offset values. The Live X-ray
image shifts by the translational offsets, to show the alignment that would
result if the couch is moved by these amounts.
You can toggle back and forth between these icons. The icons only shift the Live
X-ray image relative to the DRR image. They do not move the couch.
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
(Synchrony Fiducial Tracking mode only) You may need to modify the Window/
Level settings for the Live X-ray images to see the fiducials.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
Control Description
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls.
Fiducial ROI: (Synchrony Fiducial Tracking mode only) Displays a square yellow
box in the Live X-ray image for each Region of Interest (ROI) where the fiducial
extraction algorithm searches for fiducial candidates. Each box is centered on a
fiducial candidate.
The size of the box changes when the Tracking Range (mm) parameter is
adjusted. For more information, see “Adjusting the Tracking Range (mm)” on
page 9-20.
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Control Description
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
Some of these calculated values are also displayed to the left of slider bars (left screen), where
you can adjust their thresholds (see Figure 8 on page 9-17). If a parameter value exceeds its
threshold, a threshold violation occurs, and an orange box is displayed around the slider bar. When
a threshold violation occurs the first time you are in the Alignment > Align phase during a
treatment fraction, you cannot move to the next phase until the threshold violation is resolved.
dxAB (mm)
The dxAB (mm) parameter gives the relative distance along the Inferior/Superior axis between a
candidate fiducial in the two projections (for Camera A and Camera B of the X-ray imaging system).
If this distance exceeds the threshold value, the fiducial extraction algorithm does not consider the
candidate fiducial in each projection to be the same fiducial and a dxAB error occurs. The X-axis
(Inferior/Superior axis) is used because this axis is common to both projections.
The default value of the dxAB threshold parameter is 2.5 mm. The default value is designed to
eliminate false identification of fiducials. However, in some cases, for example, with abnormally
long fiducials (> 5 mm) or fiducials in close proximity to another bright object in the projected
image, the displacement estimates in the Inferior/Superior direction obtained from each X-ray
image detector may differ by more than this limit. In such cases, first re-adjust the offset in order
to correctly overlap with the actual fiducials, as visually identified in the live image. Then decrease
the Tracking Range to help identify and correctly lock on to the fiducials. If the error persists, then
increase the threshold value.
WARNING: If you increase the value of the dxAB (mm) threshold, it is possible for the
fiducial extraction algorithm to incorrectly identify the fiducial configuration without
generating an error. You must therefore visually verify that the fiducial configuration has
been correctly identified.
Uncertainty (%)
The Uncertainty (%) parameter gives the detection uncertainty value for the fiducial extraction
algorithm. The fiducial extraction algorithm computes an uncertainty value that is a measure of the
incorrectness of the extracted fiducial configuration.
If the detection uncertainty value computed by the fiducial extraction algorithm is greater than the
Uncertainty threshold parameter, an Uncertainty error occurs (see “Error Handling” on page 9-30).
The default value of the Uncertainty (%) threshold parameter is 40%. If an Uncertainty error occurs,
you can consider increasing the Uncertainty (%) threshold if you are confident that the fiducials
found using a higher value are correctly identified. For more information, see “Adjusting Algorithm
Parameters” on page 9-20.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
fiducial extraction algorithm to incorrectly identify a feature as a fiducial without
generating an error. You must therefore visually verify that each fiducial has been
correctly identified and that fiducials are in the correct configuration.
Fiducial Spacing
The Spacing (mm) parameter gives the lowest calculated distance between fiducials. This distance
is also called the Fiducial Spacing.
The Spacing threshold parameter sets the minimum distance between fiducials required for
accurate estimation of rotations. The default value of the Spacing threshold parameter is 18.0 mm.
When the calculated Spacing value is less than the threshold, a Spacing error occurs, and text is
displayed at the bottom of the left screen to indicate that the threshold failed.
If fewer than 3 pairs of fiducials have Spacing values that exceed the threshold, the fiducial
extraction algorithm will calculate and display rotations, but rotational corrections will not be
applied. If fewer than 3 fiducials are selected to be used for tracking, you cannot adjust the Spacing
threshold parameter.
You can force the computation of rotations by reducing the Spacing threshold parameter. However,
you should note that the accuracy of rotational corrections increases when the fiducials are further
apart from each other.
WARNING: If you lower the value of the Fiducial Spacing Threshold parameter, it is
possible for the fiducial extraction algorithm to compute rotations inaccurately, resulting
in larger targeting errors.
Collinearity
The Collinearity (deg) parameter gives the largest calculated angle between lines joining any group
of 3 fiducials. This parameter is a measure of the closeness of a triplet of fiducials to a straight line.
The Collinearity value is determined by either the smallest calculated angle θ or 180 - θ between
lines joining any group of 3 fiducials, whichever value is lower.
The Collinearity threshold parameter sets the minimum angle required for accurate estimation of
rotations. The default value of the Collinearity threshold parameter is 15.0°. When the calculated
Collinearity value is less than the threshold, a Collinearity error occurs, and text is displayed at the
bottom of the left screen to indicate that the fiducial angle threshold failed.
If at least one fiducial triplet exceeds the Collinearity threshold, the fiducial extraction algorithm
calculates rotations, and rotational corrections are applied. Otherwise, only translational
corrections are applied. If fewer than 3 fiducials are selected to be used for tracking, you cannot
adjust the Collinearity threshold parameter.
You can force the computation of rotations by reducing the Collinearity threshold parameter.
However, you should note that the accuracy of rotational corrections is reduced when the fiducial
configuration is more collinear. However, the accuracy of rotations is not guaranteed for such
fiducial configurations.
WARNING: If you lower the value of the Collinearity threshold parameter, it is possible for
the fiducial extraction algorithm to compute rotations inaccurately, resulting in larger
targeting errors.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
fiducial extraction algorithm to incorrectly identify a feature as a fiducial without
generating an error. You must therefore visually verify that each fiducial has been
correctly identified and that fiducials are in the correct configuration.
WARNING: If you increase the value of the dxAB (mm) threshold, it is possible for the
fiducial extraction algorithm to incorrectly identify the fiducial configuration without
generating an error. You must therefore visually verify that the fiducial configuration has
been correctly identified.
WARNING: If you increase the value of the Rigid Body Error threshold, it is possible for
the fiducial extraction algorithm to result in larger targeting errors leading to
mistreatment.
WARNING: If fewer than 3 fiducials are enabled for tracking, the tracking algorithm will
not estimate rotational corrections. Only translational corrections will be applied during
beam delivery. Tracking with fewer than 3 fiducials significantly reduces the usefulness of
the Rigid Body Error value as an indicator of tracking reliability.
WARNING: Use extreme caution when considering whether to track the target using only
1 fiducial. If only 1 fiducial is enabled for tracking, the tracking algorithm will not estimate
rotational corrections and the Rigid Body Error value will no longer be available as an
indicator of tracking reliability. Tracking with only 1 fiducial increases the risk of beam
mis-targeting.
The Fiducials checkboxes on the left screen (see Figure 6 on page 9-8) allow you to turn tracking
on or off for individual fiducials:
• When the checkbox for a fiducial is selected, the fiducial extraction algorithm uses that
fiducial to track target movement. The corresponding diamond marker for that fiducial in
the DRR images is yellow.
• When a checkbox is deselected, the corresponding fiducial is not used for tracking. The
diamond marker is removed.
• Images are automatically correlated when you select or deselect a checkbox.
Up to 8 fiducial checkboxes can be displayed. The corresponding fiducials are numbered in the
images.
The following examples are cases where you could consider disabling a fiducial so that it is not
used for tracking:
• The fiducial is incorrectly identified.
• The fiducial produces Rigid Body Errors due to migration. For more information on
disabling fiducials in response to a Rigid Body Error, see “Responding to a Rigid Body
Error” on page 9-23.
• The fiducial has a high uncertainty level due to poor image quality in the Live X-ray
images, migration, obstruction by an object in the ROI, or close proximity to another
fiducial.
Once the new Live X-ray image acquisition or image correlation is complete, you should again
attempt to visually verify that the fiducials have been correctly located. If they have not, repeat the
above procedures and try again.
WARNING: If you increase the value of the Rigid Body parameter threshold, it is possible
for the fiducial extraction algorithm to result in larger targeting errors leading to
mistreatment.
When you disable the fiducial with a solid red square so that it is not used for tracking, its
contribution to the error is removed. When you disable a fiducial, a new image correlation is
performed automatically. A solid red square is then displayed for the next fiducial that is
contributing the most to the error.
After each image acquisition or correlation, the Algorithm Details tab in the Image History
table (right screen) displays the value of several fiducial tracking algorithm parameters. For the
Rigid Body Error parameter, the calculated value for each pair of fiducials is displayed. You can
refer to these values as an aid to determine which fiducials, if any, to disable for tracking in
response to a Rigid Body Error.
WARNING: If fewer than 3 fiducials are enabled for tracking, the tracking algorithm will
not estimate rotational corrections. Only translational corrections will be applied during
beam delivery. Tracking with fewer than 3 fiducials significantly reduces the usefulness of
the Rigid Body Error value as in indicator of tracking reliability.
WARNING: Use extreme caution when considering whether to track the target using only
1 fiducial. If only 1 fiducial is enabled for tracking, the tracking algorithm will not estimate
rotational corrections and the Rigid Body Error value will no longer be available as an
indicator of tracking reliability. Tracking with only 1 fiducial increases the risk of beam
mis-targeting.
If the Rigid Body Error values shown in Figure 10 are listed in descending order, the fiducial pair
that is contributing the most to the error will be at the top of the list:
RBE1-3 1.9
RBE2-3 1.7
RBE1-4 1.6
RBE1-2 1.5
RBE2-4 1.3
To determine which fiducial in the pair is contributing the most, evaluate how the error changes if
one of these fiducials is removed. If Fiducial 1 is removed, the next highest Rigid Body Error value
is 1.7. If Fiducial 3 is removed, the next highest Rigid Body Error is 1.6. Therefore, Fiducial 3 has
the most significant contribution to the Rigid Body Error.
A solid red square in the Fiducial Tracking control area indicates that Fiducial 3 is contributing the
most, as shown in Figure 11. A solid yellow square indicates that Fiducial 1 has the next largest
impact. The current value of the Rigid Body Error, 1.9 mm, is displayed under the slider bar.
If Fiducial 3 is disabled for tracking, Fiducial 1 will become the fiducial that contributes the most to
the error (Fiducial 1 contributes to the next highest value in the list). A solid red square is then
displayed for Fiducial 1, as shown in Figure 12.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
You can also use Couch Mode to shift the Live X-ray and DRR images so that anatomical or other
features match. To do this, select the Couch Mode icon (Table 1 on page 9-11). Then click and
drag in the Live X-ray image. Do this separately for each image panel.
As the relative position of the images is adjusted, the target translational values in the Couch
Position (mm) textboxes are updated. (Rotational values are not updated in Couch Mode.)
Then click the Move Couch button to move the couch to the target position.
After moving the couch, you acquire a Live X-ray image and then analyze imaging results to verify
that each fiducial has been correctly located and that the fiducials are in the correct configuration.
If the fiducials are not correctly identified, reattempt the procedures described in the previous
section “Analyzing Imaging Results” on page 9-9.
When the patient is within required translational and rotational tolerances, you can proceed with
treatment by moving to the Readiness phase. For more information on required tolerances, see
“Chapter 8: Treatment Delivery”.
Treatment Delivery
This section covers the following topics:
• “Starting Treatment Delivery” on page 9-29
• “Treatment Monitoring Graph” on page 9-29
• “Error Handling” on page 9-30
• “Patient Realignment” on page 9-30
The left screen of the Delivery phase for Synchrony Fiducial Tracking mode is shown in
Figure 13. Controls allow you to monitor and verify patient alignment and tracking results during
treatment delivery.
For more information on treatment delivery procedures, see “Chapter 8: Treatment Delivery”.
WARNING: Visually verify that the fiducials are correctly identified throughout treatment.
WARNING: After X-ray image acquisition at each node, visually verify the patient position
and verify the positions of the located fiducials. If at any time it appears that the patient or
fiducials are not correctly located in the Live X-ray images, pause treatment by clicking
the Interlock button on the left screen of the Delivery phase or push the nearest E-Stop
button.
After pausing treatment, follow the procedures described in “Analyzing Imaging Results” on
page 9-9 until fiducial location is successful and shows that the patient is within alignment bounds.
For more information on interrupting treatment, see “Chapter 8: Treatment Delivery”.
Error Handling
The fiducial extraction algorithm may fail for the following reasons:
• dxAB Error
• Rigid Body Error
• Uncertainty Error
• Spacing Error
• Collinearity Error
During patient alignment in the Alignment > Align phase, when the calculated value of a
fiducial extraction algorithm parameter exceeds the value of its threshold, a threshold violation
occurs. An orange box is then displayed around the slider bar for that parameter threshold.
If this occurs the first time you are in the Alignment > Align phase during a treatment fraction,
you cannot proceed to the next phase until the threshold violation is resolved. After the first time
(for example, if you return to the Alignment > Align phase to realign the patient), if a threshold
violation occurs, you can proceed to the next phase. However, you will not be able to start
treatment delivery until the threshold violation is resolved.
If a threshold violation occurs while treatment is in progress in the Delivery phase, a correlation
error occurs (a Soft Stop). Treatment is paused and the error is described in the Messages panel.
To determine which threshold was violated, you can return to the Alignment > Align phase to
see which slider bar is surrounded by an orange box.
For information on the parameters above and their thresholds, see “Monitoring and Adjusting
Algorithm Parameters” on page 9-16. For information on responding to a Rigid Body Error, see
“Responding to a Rigid Body Error” on page 9-23. For information on resuming treatment after a
Soft Stop, see Chapter 7, “Treatment Delivery”.
Patient Realignment
If treatment is interrupted due to a Soft Stop or E-Stop, or if you click the Interlock button or the
Pause button, you should confirm alignment of the patient before resuming treatment. For
information on patient realignment after a treatment interruption, see Chapter 7, “Treatment
Delivery”.
Introduction
This chapter describes how to use Synchrony Spine Tracking Supine and Synchrony Spine
Tracking Prone with Respiratory Modeling. When content is relevant to both tracking modes, the
generic term "Synchrony Spine Tracking" is used.
This chapter covers the following topics:
• “Overview of Synchrony Spine Tracking” on page 10-2
• “Requirements for Visual Evaluation” on page 10-3
• “Workflow for Synchrony Spine Tracking” on page 10-4
• “Verifying Patient and Treatment Data” on page 10-5
• “Visually Aligning the Patient” on page 10-7
• “Optimizing the X-ray Technique” on page 10-8
• “Performing Patient Alignment” on page 10-11
• “Checking Readiness for Treatment Delivery” on page 10-27
• “Treatment Delivery” on page 10-28
• “Synchrony Spine Tracking Prone with Respiratory Modeling” on page 10-33
Synchrony Spine Tracking enables tracking of skeletal structures in the spine region for accurate
patient positioning and radiation beam delivery using the CyberKnife System. It provides the
capability to deliver radiation treatments in all regions of the spine without the need for implanting
fiducials. Synchrony Spine Tracking is capable of accurately and automatically tracking a wide
range of skeletal structures, including all cervical, thoracic, lumbar and sacral areas.
This chapter assumes you are familiar with “Chapter 8: Treatment Delivery”.
• A region of interest (ROI) containing an 81-node grid is defined when the treatment plan
is created. During treatment delivery, Synchrony Spine Tracking computes target
displacement by monitoring the displacement of nodes of the ROI in the Live X-ray
images relative to the nodes in the DRR images. The ROI size defined during treatment
planning is used as the default ROI size during treatment delivery.
Synchrony Spine Tracking computes patient displacement in 6 degrees of freedom: 3 translations
and 3 global rotations.
During patient alignment, the computed displacements are used to adjust the patient position.
During treatment delivery, the treatment robot adjusts the position of the LINAC to correct for
translational and rotational displacement of the target. For more information on selecting the
imaging center and adjusting the ROI size during treatment planning, see the Treatment Planning
Manual.
Figure 2 shows the workflow for patient alignment and treatment delivery for Synchrony Spine
Tracking Prone with Respiratory Modeling. For information on Synchrony Spine Tracking Prone
with Respiratory Modeling, see “Synchrony Spine Tracking Prone with Respiratory Modeling” on
page 10-33. For information on using Synchrony Spine Tracking with Respiratory Modeling, see
“Chapter 12: Synchrony Respiratory Tracking System (Option)”.
WARNING: The Live X-ray images should have the same image quality as the DRR
images. Otherwise, the tracking algorithm may not be able to identify and track features
correctly.
Figure 4 shows the Alignment > Technique phase (left screen) for Synchrony Spine Tracking
after image acquisition. Table 1 on page 10-14 describes the image panel controls to aid image
analysis in this phase, including DRR display modes.
The Region of Interest (ROI) Width is represented by a yellow box in the Raw Image (far left in
Figure 4). The ROI Width (mm) value is also displayed in the Image History table on the right
screen (see Figure 5). A histogram display below each Raw Image represents the distribution of
pixel intensities in the ROI. The Image History table also displays the calculated False Nodes
value for each image.
The goal is to determine the optimal X-ray technique for XRS A and XRS B independently, by
evaluating image quality, histogram displays, and False Nodes values of images on the left screen
and in the Image History table.
Increasing the X-ray technique generally pushes the histogram peak to the left. Decreasing the X-
ray technique generally pushes the peak to the right. The ideal distribution is spread out (indicating
good contrast) and not shifted towards the extreme left or right.
If you are satisfied with the X-ray technique, you can proceed to the Alignment > Align phase.
Otherwise, you can change the X-ray technique as needed until you are satisfied with the image
quality, as described below. For more information on the ROI Width and the False Nodes
parameters, see “Monitoring and Adjusting Algorithm Parameters” on page 10-20. For more
information on the Alignment > Technique phase, see “Chapter 8: Treatment Delivery”.
The existing Live X-ray images from the Alignment > Technique phase are automatically
correlated with the DRR images.
• An orange box around an offset value indicates an offset is not within required tolerances
and an out-of-bounds error has occurred. An orange box around a tracking algorithm
slider bar indicates that a threshold violation has occurred.
You can use image analysis tools in Table 1 on page 10-14 to make sure the Region
of Interest (ROI) includes sufficient skeletal structures. Check that anatomy is aligned
in the Live X-ray and DRR images.
To reduce offsets, you can move the treatment couch using the Move Couch
button. See “Moving the Couch” on page 10-27.
You can also adjust algorithm parameters as needed. See “Monitoring and Adjusting
Algorithm Parameters” on page 10-20.
NOTE: Not all image controls are available for all treatment
modes.
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Control Description
Couch: Adjust the position of the Live X-ray image relative to the DRR image by
clicking and dragging in either the Live X-ray image or DRR image. See “Moving
the Couch” on page 10-27. Depending on the treatment phase, the following icons
are displayed:
• Click the Reset icon to move the Live X-ray image back to its original,
unshifted position.
• Click the Reset icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with zero values. The Live X-ray image moves
to its unshifted position, to show existing alignment.
• Click the Offsets icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with the calculated offset values. The Live X-ray
image shifts by the translational offsets, to show the alignment that would
result if the couch is moved by these amounts.
You can toggle back and forth between these icons. The icons only shift the Live
X-ray image relative to the DRR image. They do not move the couch.
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls.
Control Description
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Full Content DRR: (Synchrony Spine Tracking only) Displays the unsegmented
(full content) DRR images. See “Display Modes for DRR Images” on page 10-17.
Segmented DRR: (Synchrony Spine Tracking only) Displays the segmented DRR
images, in which all image content except the spine is removed. See “Display
Modes for DRR Images” on page 10-17.
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
NOTE: These display modes affect the image display only. They
do not affect the results of the Synchrony Spine Tracking
algorithm.
Synchrony Spine Tracking relies on skeletal structures in the ROI for accurate tracking. Users
should inspect the ROIs for both Camera B and Camera A of the X-ray imaging system to verify
that each ROI includes sufficient skeletal structures for tracking.
The skeletal meshes correspond to the 81-node grid that is predefined during treatment planning.
The skeletal meshes are displayed in different colors depending on the agreement between nodes
in the DRR and Live X-ray images:
• Green: The skeletal mesh in a DRR image is displayed in green.
• Dark blue: A node in a Live X-ray image is displayed in dark blue if there is good
agreement between that node and the corresponding node in the DRR image. Grid lines
projecting from that node are also displayed in dark blue.
• Orange: A node in a Live X-ray image is displayed in orange if it is a false node. Grid
lines projecting from a false node are also displayed in orange.
To identify nodes in the Live X-ray images, Synchrony Spine Tracking looks for internal anatomy
that corresponds to the region surrounding each node in the DRR images. If the estimated
translation of a node in the skeletal mesh is inconsistent with nodes in the surrounding grid, the
node is identified as a false node. The smaller the number of false nodes, the more reliable the
translational corrections from the Synchrony Spine Tracking algorithm.
In Couch Mode (see Table 1 on page 10-14), when you click the Offsets icon, the skeletal meshes
in the DRR images and the Live X-ray images for both projections (Camera A and Camera B of the
X-ray imaging system) are automatically aligned.
The Offsets icon uses only translations to perform alignment. When large rotations exist, the
rotations of the skeletal mesh in the Live X-ray images can clearly be seen. You can visually
inspect the tracking accuracy from the alignment of the mesh nodes in the area close to the image
center.
These calculated values are also displayed to the left of slider bars (left screen), where you can
adjust their thresholds (see Figure 10 on page 10-20). If a parameter value exceeds its threshold,
a threshold violation occurs, and an orange box is displayed around the slider bar. When a
threshold violation occurs the first time you are in the Alignment > Align phase during a
treatment fraction, you cannot move to the next phase until the threshold violation is resolved.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before changing this parameter value.
drAB (deg)
The drAB (deg) parameter gives the absolute difference between projected roll angles in the two
projections (for Camera A and Camera B of the X-ray imaging system). This difference is called
drAB. When the drAB value calculated by the Synchrony Spine Tracking algorithm is below the
threshold, the estimate of roll by the tracking algorithm is considered to be reliable. If the drAB
value exceeds the drAB threshold value, a drAB error occurs (see “Error Handling” on page 10-31).
The default value of the drAB threshold parameter is 2.0°. The value of the drAB threshold
parameter should be small. In general, if you change this parameter value during patient alignment,
you should return to the default value when patient alignment is completed. In some cases, for
example, poor image quality or degraded bony anatomy, it may be necessary to use a larger
threshold value than the default.
WARNING: If you increase the value of the drAB threshold parameter, you must perform
visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce unreliable
estimates of the patient roll, which in turn can lead to misdirected radiation beams. Use
caution before changing this parameter value.
WARNING: If you increase the value of the False Nodes threshold parameter, you must
perform visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before increasing this parameter value.
The ROI Width (mm) slider bar includes the following controls:
• The current value is displayed to the right of the slider bar.
• The slider bar cursor allows you to adjust the value of the parameter. Drag the cursor to
change the value. You can also enter a value directly in the textbox under the slider bar.
• When the From Planning checkbox is selected, the default value of the ROI Width
parameter specified in the treatment plan is used.
To change the ROI Width (mm) parameter from the default value:
1. Clear (deselect) the From Planning checkbox.
The textbox to the right of the slider bar is enabled with the default value 40.0 mm.
2. Click and drag the slider bar cursor to adjust the value. Alternatively, enter a new value in
the textbox under the slider bar and press <Enter>.
3. When the ROI Width value is changed, a blue box representing the ROI in the DRR
image dynamically changes size.
Images are automatically correlated when the ROI Width (mm) parameter is changed.
To return to the default ROI Width value specified in the treatment plan:
1. Select the From Planning checkbox. The ROI Width value from the treatment plan
is displayed and the textbox is disabled.
The ROI Width should be adjusted during treatment planning based on the region of the spine
being tracked. For example, you may consider increasing the value of this parameter for the lumbar
and lower thoracic region because vertebrae are larger, and also increasing it for small instrument
implants and decayed skeletal structures. For information on how to size the ROI appropriately,
see the Treatment Planning Manual.
In most cases the ROI Width specified during treatment planning should be used. The ROI Width
can be adjusted during treatment delivery to include more detailed spine structure that may not
have been seen in the DRR images generated during treatment planning.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before changing this parameter value.
WARNING: If you increase the value of the False Nodes threshold parameter, you must
perform visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before increasing this parameter value.
WARNING: If you increase the value of the drAB threshold parameter, you must perform
visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce unreliable
estimates of the patient roll, which in turn can lead to misdirected radiation beams. Use
caution before changing this parameter value.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
When the patient is within required translational and rotational tolerances, you can proceed with
treatment by moving to the Readiness phase. For more information on required tolerances, see
“Chapter 8: Treatment Delivery”.
Treatment Delivery
This section covers the following topics:
• “Visual Inspection During Treatment DeliverySynchrony Spine Tracking relies on the
nodes in the center columns of the ROI to estimate the patient roll rotations (rotation
about the Inferior/Superior axis).” on page 10-29
• “Treatment Monitoring Graph” on page 10-30
• “Error Handling” on page 10-31
The Delivery phase (left screen) for Synchrony Spine Tracking is shown in Figure 13. Controls
on this screen and the right screen allow you to monitor and verify patient alignment and tracking
results during treatment delivery using Synchrony Spine Tracking .
For additional information on using Synchrony Spine Tracking Prone with Respiratory Modeling,
see “Synchrony Spine Tracking Prone with Respiratory Modeling” on page 10-33.
For more information on treatment delivery procedures, see “Chapter 8: Treatment Delivery”. For
information on treatment planning for treatment using Synchrony Spine Tracking , see the
Treatment Planning Manual.
Visual Inspection During Treatment DeliverySynchrony
WARNING: During treatment delivery, after acquisition of Live X-ray images at each node,
visually verify the patient position and verify that there are sufficient skeletal features in
the images. Make sure that anatomy in the Live X-ray and DRR images matches. For more
information on visually identifying features to be tracked, see “Chapter 8: Treatment
Delivery”.
WARNING: Lack of skeletal features in the Region of Interest (ROI) can result in incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Inspect the ROIs in the Live X-ray images for both Camera A and Camera B to verify that
the ROI includes sufficient skeletal structures.
Spine Tracking relies on the nodes in the center columns of the ROI to estimate the patient roll
rotations (rotation about the Inferior/Superior axis).
WARNING: Lack of features in the center columns of the ROI can cause the Synchrony
Spine Tracking algorithm to produce incorrect estimates of the patient roll rotation, which
in turn can lead to misdirected radiation beams. Inspect the ROIs in the Live X-ray images
from both Camera A and Camera B to verify that the nodes in the center columns of each
ROI include sufficient features.
Error Handling
The Synchrony Spine Tracking algorithm may fail for the following reasons:
• “False Nodes Error” on page 10-31
• “dxAB Error” on page 10-32
• “drAB Error” on page 10-32
During patient alignment in the Alignment > Align phase, when the calculated value of a
Synchrony Spine Tracking parameter exceeds the value of its threshold, one of the above
threshold violations occurs. An orange box is then displayed around the slider bar for that
parameter threshold.
If this occurs the first time you are in the Alignment > Align phase during a treatment fraction,
you cannot proceed to the next phase until the threshold violation is resolved. After the first time
(for example, if you return to the Alignment > Align phase to realign the patient), if a threshold
violation occurs, you can proceed to the next phase. However, you will not be able to start
treatment delivery until the threshold violation is resolved.
If a threshold violation occurs while treatment is in progress in the Delivery phase, a correlation
error occurs (a Soft Stop). Treatment is paused and the error is described in the Messages panel.
To determine which threshold was violated, you can return to the Alignment > Align phase to
see which slider bar is surrounded by an orange box.
For information on resuming treatment after a Soft Stop, see Chapter 7, “Treatment Delivery”.
WARNING: If you increase the value of the False Nodes threshold parameter, you must
perform visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before increasing this parameter value.
dxAB Error
A dxAB Error due to overly large dxAB values may be caused by large yaw rotations during the
Delivery phase. In this case, an error message is displayed in the Messages window.
To resolve this error, first check patient alignment. The treatment couch yaw position may need to
be adjusted. Then consider increasing the value of the dxAB threshold parameter. After the patient
is aligned, return to the default value of this parameter.
WARNING: If you change the value of the dxAB threshold parameter, you must perform
visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Spine Tracking algorithm to produce incorrect
estimates of the patient position, which in turn can lead to misdirected radiation beams.
Use caution before changing this parameter value.
drAB Error
A drAB Error due to overly large drAB values may be caused by image noise, low contrast, or a
deformed spine region during the Delivery phase. In this case, an error message is displayed in
the Messages window.
To resolve this error, first check patient alignment. Next consider adjusting X-ray parameters to
improve X-ray image quality. Then for very low contrast images, consider increasing the value of
the drAB threshold parameter to resolve the Soft Stop generated by the error. After the patient is
successfully aligned, return to the default value of this parameter.
If the previous solutions do not resolve the error, then after visually confirming tracking accuracy,
try increasing the drAB threshold value and disable (clear) the Robot Rotation Corrections
Enabled checkbox in the Readiness phase. Perform visual inspection to monitor and confirm
tracking results in both projections (Camera A and Camera B).
WARNING: If you increase the value of the drAB threshold parameter, you must perform
visual inspection after tracking results are displayed. Increasing the value of this
parameter may cause the Synchrony Spine Tracking algorithm to produce unreliable
estimates of the patient roll, which in turn can lead to misdirected radiation beams. Use
caution before changing this parameter value.
CT Acquisition
For Synchrony Spine Tracking Prone with Respiratory Modeling treatments, it is essential to have
the patient in the same position for treatment as they were during pretreatment CT scanning,
including all immobilization devices. A reproducible patient position is the primary requirement
during patient preparation for the CT scan. Bring all immobilization devices created for the patient
to the CT scanner to replicate the patient position.
It is very important that the CT scan be taken during natural (not forced) inhalation or exhalation,
and that the same immobilization applied to the patient is used for both CT scanning and treatment
delivery. If any compression device is placed on the patient, it is essential that it be present for both
CT scanning and treatment delivery.
The size of any immobilization device placed between the hard surface of the treatment couch
table top and the patient should be minimized to reduce the distance from the couch surface to
highest point of the patient. Using an immobilization device that is thicker than the couch cushions
provided by Accuray increases the risk that the patient will be positioned outside the Patient Safety
Zone. For more information on the Patient Safety Zone, see “Chapter 2: Safety”.
WARNING: Position the patient and all devices within the Patient Safety Zone defined by
the Proximity Detection Program (PDP) model. Anything extending outside the boundaries
of the PDP model increases the risk of collision with the treatment robot and can cause
serious injury or harm to the patient.
Follow the guidelines for patient preparation for pretreatment scanning described in “Chapter 6:
Patient Preparation”, and “Chapter 12: Synchrony Respiratory Tracking System (Option)”.
Because spine rotations are not corrected for by default for Synchrony Spine Tracking Prone with
Respiratory Modeling treatments, the default settings for Rotational Bounds Checking that
trigger a Soft Stop are reduced to 2.5 degrees for all three rotations (roll, pitch, and yaw). You can
override these limits down to 0.1 degrees or up to 5 degrees for each rotation.
When target motion is not clinically relevant as determined by a clinician, it may be advantageous
to disable respiratory modeling. For Synchrony Spine Tracking Prone, when respiratory modeling
is disabled, rotational corrections are enabled by default and can be corrected for during treatment
delivery. To disable respiratory modeling, select the Skip Respiratory checkbox. The Robot
Rotation Corrections Enabled checkbox is then automatically selected in the Readiness
phase. The settings for Rotational Bounds Checking will be determined by the treatment
couch, among other factors. For more information, see “Chapter 8: Treatment Delivery”.
Introduction
This chapter describes how to use the Synchrony Lung Tracking with Respiratory Modeling with
the CyberKnife System. This chapter covers the following topics:
• “Overview of the Synchrony Lung Tracking with Respiratory Modeling” on page 11-2
• “Requirements for Visual Evaluation” on page 11-5
• “Workflow for Synchrony Lung Tracking with Respiratory Modeling” on page 11-6
• “Verifying Patient and Treatment Data” on page 11-8
• “Visually Aligning the Patient” on page 11-9
• “Optimizing the X-ray Technique” on page 11-10
• “Performing Patient Alignment” on page 11-11
• “Treatment Delivery” on page 11-34
The Synchrony Lung Tracking with Respiratory Modeling enables radiation beam delivery in the
lung with for accurate tracking without the need for implanting fiducials. It is used in combination
with Synchrony Spine Tracking to perform patient alignment. During treatment delivery, the
Synchrony Lung Tracking with Respiratory Modeling may be used with the Synchrony tracking
System to track movement of the tumor due to patient breathing patterns.
This chapter assumes you are familiar with the following chapters:
• “Chapter 8: Treatment Delivery” on page 8-1
• “Chapter 10: Synchrony Spine Tracking (Option)” on page 10-1
• “Chapter 12: Synchrony Respiratory Tracking System (Option)” on page
12-1
11-2 | Overview of the Synchrony Lung Tracking with Respiratory Modeling 1075883-ENG A
Treatment Delivery Manual Chapter 11: Synchrony Lung Tracking with Respiratory Modeling
Quality Metrics
The Synchrony Lung Tracking with Respiratory Modeling algorithm operates by locating an
intensity pattern in the Live X-ray images that is similar to that in the DRR images. It is possible
that more than one such pattern exists in the search area. The algorithm takes into account several
factors to assess the registration quality:
• The maximum of the similarity measure between the Live X-ray and DRR images.
• The number of other local maxima found within the Tracking Range (for information on
the Tracking Range parameters, see “Adjusting Tracking Range X (mm) and Tracking
Range Y (mm)” on page 11-30).
• The tumor size and the tumor location relative to the prediction made by the Synchrony
tracking System. For more information on the Synchrony tracking System, see “Chapter
12: Synchrony Respiratory Tracking System (Option)”.
These factors are combined to calculate the detection uncertainty. The threshold for detection
uncertainty is user-defined (see “Uncertainty (%)” on page 11-29). If the detection uncertainty is
higher than the threshold, treatment delivery is interrupted by a Soft Stop.
The Synchrony Lung Tracking with Respiratory Modeling algorithm also calculates a dxAB value
as a quality metric. For information on this parameter, see “dxAB (mm)” on page 11-29.
1075883-ENG A Overview of the Synchrony Lung Tracking with Respiratory Modeling | 11-3
Accuray Incorporated Treatment Delivery Manual
Preferred Projection
In some cases, the tumor is more clearly visible or better identified on one Live X-ray image than
the other. Depending on tumor location, the tumor in one projection may partially or completely
overlap with the spine, chest wall, heart, diaphragm, or other anatomic structures. In these cases,
the Preferred Projection feature of the Synchrony Lung Tracking with Respiratory Modeling
algorithm can improve the accuracy and robustness of target localization (see “Enabling Preferred
Projection” on page 11-31).
When the preferred projection method is enabled, the Synchrony Lung Tracking with Respiratory
Modeling uses the shared Inferior/Superior direction in the two Live X-ray images to combine the
similarity measures from both images when attempting to find the target. In this way, poor target
visibility in one image can be overcome by combining the similarity measure from that image with
the well-defined similarity maximum in the image where the target is clearly visible.
WARNING: For image correlations when the preferred projection method is invoked, the
dxAB value can no longer be used as an indication of tracking reliability. It is possible for
a tracking failure to occur without producing a dxAB error.
11-4 | Overview of the Synchrony Lung Tracking with Respiratory Modeling 1075883-ENG A
Treatment Delivery Manual Chapter 11: Synchrony Lung Tracking with Respiratory Modeling
In the Accuray Precision System, sample DRR images are generated and should be inspected to
determine whether the case is suitable for Synchrony Lung Tracking with Respiratory Modeling. If
the case is confirmed to be suitable, the treatment plan is saved. Prior to treatment delivery,
multiple DRR images are generated using the iDMS Data Management System. The DRR images
are used during patient alignment with Synchrony Spine Tracking and during treatment using the
Synchrony Lung Tracking with Respiratory Modeling.
For more information on the DRR images that are generated, see “Generating DRR Images” on
page 11-8. For more information on treatment planning using the Synchrony Lung Tracking with
Respiratory Modeling, see the Treatment Planning Manual.
Two treatment couch positions are used during treatment delivery with the Synchrony Lung
Tracking with Respiratory Modeling:
• Spine position: The alignment center defined during treatment planning in the Align step
of the Accuray Precision System. This position is used for alignment by Synchrony Spine
Tracking.
• Lung position: The treatment center automatically determined during treatment planning
during tumor segmentation. This is the spine position plus the alignment-to-treatment
offset from planning. This position is used for tracking by the Synchrony Lung Tracking
with Respiratory Modeling.
The alignment center is typically different from the treatment center. The alignment center is used
during an initial spine alignment procedure using Synchrony Spine Tracking. The spine alignment
is performed in the Alignment > Align phase. This procedure determines the rotational
alignment corrections that will be used during treatment delivery with the Synchrony Lung Tracking
with Respiratory Modeling.
After spine alignment is completed, you transition from Synchrony Spine Tracking to the Synchrony
Lung Tracking with Respiratory Modeling. This transition occurs when you move to the
Alignment > Lung phase. You then move the treatment couch to the treatment center.
WARNING: It is strongly recommended that Synchrony Spine Tracking be used for initial
patient alignment (by performing spine alignment in the Alignment > Align phase)
before treatment is delivered using the Synchrony Lung Tracking with Respiratory
Modeling. Otherwise, rotational misalignment during patient positioning will not be
compensated for.
11-6 | Workflow for Synchrony Lung Tracking with Respiratory Modeling 1075883-ENG A
Treatment Delivery Manual Chapter 11: Synchrony Lung Tracking with Respiratory Modeling
Treatment delivery using the Synchrony Lung Tracking with Respiratory Modeling includes the
following general steps:
1. Load patient and treatment plan data in the Overview phase, as described in “Chapter
8: Treatment Delivery”.
2. Find the optimal X-ray technique for spine alignment using Synchrony Spine Tracking in
the first Alignment > Technique phase, as described in “Chapter 10: Synchrony
Spine Tracking (Option)”.
3. Position the patient on the treatment couch and move the patient to the alignment center
for initial spine alignment using Synchrony Spine Tracking in the Alignment > Align
phase, as described in “Chapter 10: Synchrony Spine Tracking (Option)”.
4. Use Synchrony Spine Tracking to align the patient translationally and rotationally at the
spine position, as described in “Chapter 10: Synchrony Spine Tracking (Option)”. After
this step, the treatment couch is at Synchrony Spine Tracking alignment center.
5. Switch from Synchrony Spine Tracking alignment center (the spine position) to the
Synchrony Lung Tracking with Respiratory Modeling treatment center (the lung position)
in the Alignment > Lung phase.
This step involves treatment couch movement. The difference between the alignment
center and the treatment center includes the alignment-to-treatment offset defined from
treatment planning plus any additional offset the user applied during the current or
previous treatment fractions. For more information, see “Transition to the Synchrony
Lung Tracking with Respiratory Modeling” on page 11-13.
6. Find the optimal X-ray technique for treatment delivery using the Synchrony Lung
Tracking with Respiratory Modeling in the second Alignment > Technique phase.
7. Use the Synchrony Lung Tracking with Respiratory Modeling to translationally align the
patient sufficiently for tracking, in the Alignment > Respiratory phase
If the Synchrony tracking System is used, establish a Respiratory correlation model for
patient respiratory motion.
1075883-ENG A Workflow for Synchrony Lung Tracking with Respiratory Modeling | 11-7
Accuray Incorporated Treatment Delivery Manual
Figure 2 Alignment > Couch phase (left screen) for Synchrony Lung Tracking with Respi-
ratory Modeling after visual alignment using Couch Mode
For Synchrony Lung Tracking with Respiratory Modeling treatments, selecting the Skip Spine
checkbox allows you to skip spine alignment and proceed directly to the Alignment > Lung
phase. This checkbox is deselected by default. You should not skip spine alignment if it is possible
to track using Synchrony Spine Tracking algorithm. If you do decide to skip spine alignment, you
should align to the best of your visual ability before proceeding.
WARNING: It is strongly recommended that Synchrony Spine Tracking be used for initial
patient alignment (by performing spine alignment in the Alignment > Align phase)
before treatment is delivered using the Synchrony Lung Tracking with Respiratory
Modeling. Otherwise, rotational misalignment during patient positioning will not be
compensated for.
After aligning the patient properly within acceptable bounds, you are ready to move to the
Alignment > Lung phase.
WARNING: Do not reposition the patient or adjust treatment couch rotation after
switching to the Synchrony Lung Tracking with Respiratory Modeling System. If it is
necessary to perform either of these actions, return to the Alignment > Align phase
and repeat Synchrony Spine Tracking alignment procedure, as described in “Chapter 10:
Synchrony Spine Tracking (Option)”.
The goal of this step is to move the treatment couch from the spine position to the lung position.
You then take an image and make sure that rigid structures are aligned in the Live X-ray and DRR
images.
Figure 4 shows the Alignment > Lung phase (left screen) before moving to the lung position.
The full content DRR images corresponding to the treatment center for the Synchrony Lung
Tracking with Respiratory Modeling are displayed.
Table 1 shows the treatment couch position display in the Alignment > Lung phase (below
Image A). The display includes a position indicator which shows the treatment couch at one of 4
positions:
• At the spine position: The spine align center from treatment planning plus the final offsets
from the Alignment > Align phase.
• At the lung position: This is the spine position plus the shift to the lung align center from
treatment planning.
• Between the spine and lung position: The treatment couch is paused between or moving
between these two positions.
• Near but not at the lung position.
The display also includes 2 buttons for moving the treatment couch, as described in Table 2.
Move Couch Moves the treatment couch to the lung position (treatment
to Lung position).
To move to the spine or lung position after treatment couch motion is interrupted:
• In the Alignment > Lung phase, click the Move Couch to Spine button to move
to the spine position (alignment center). Or, click the Move Couch to Lung position to
move to the lung position (treatment center).
If it is necessary, you can add an offset to the alignment-to-treatment offset from planning, as
described below.
To add an offset to the lung position from planning:
1. Enter target translational values in the Couch Position (mm) textboxes.
Arrows above the textboxes point in the positive direction. Values indicate target
treatment couch movement relative to its current position.
2. Click the Move Couch button to move the treatment couch. (The couch rotations from
spine alignment are locked in the Alignment > Lung phase.)
The additional offset is automatically saved and will be applied during treatment delivery
for the current fraction and future fractions. (The Apply additional relative offset
checkbox is selected by default.)
3. To remove the additional offset that you generated in Step 2, or to remove the additional
offset that was saved and applied for previous fractions, deselect the Apply
additional relative offset checkbox. Then click the Move Couch to Lung
button.
This step returns the couch to the lung position (the treatment position from planning).
The lung position will then be used for the current fraction and future fractions.
After moving the treatment couch to the lung position, taking an image, and verifying that rigid
structures in the Live X-ray and DRR images are aligned, you are ready to move to the second
Alignment > Technique phase.
WARNING: The Live X-ray images should have the same image quality as the DRR
images. Otherwise, the tracking algorithm may not be able to identify and track features
correctly.
Figure 5 shows the Alignment > Technique phase (left screen) for the Synchrony Lung
Tracking with Respiratory Modeling after image acquisition. A histogram display below each Raw
Image represents the distribution of pixel intensities in the entire image. Table 3 on page 11-23
describes the image panel controls to aid image analysis in this phase, including DRR display
modes.
Figure 6 shows the Alignment > Technique phase (right screen) for the Synchrony Lung
Tracking with Respiratory Modeling after image acquisition. The image history displays the
technique used to acquire the selected image as well as a histogram representing the distribution
of pixel intensities in the selected image,
The goal is to determine the optimal X-ray technique for XRS A and XRS B independently, by
evaluating image quality and histogram displays of images on the left screen and in the Image
History table.
Increasing the X-ray technique generally pushes the histogram peak to the left. Decreasing the X-
ray technique generally pushes the peak to the right. The ideal distribution is spread out (indicating
good contrast) and not shifted towards the extreme left or right.
If you have selected a satisfactory X-ray technique, you can proceed to the
Alignment > Respiratory phase. Otherwise, you can change the X-ray technique as needed
until you are satisfied with the image quality, as described below. For more information on the
Alignment > Technique phase, see “Chapter 8: Treatment Delivery”.
The Alignment > Respiratory phase has the following steps, described in the sections
below:
• “Analyzing Imaging Results” on page 11-21: You start by evaluating image correlation
results to determine if you need to move the treatment couch. You also review the Live X-
ray images to verify that the tumor has been correctly located and that anatomy in the
Live X-ray and DRR images matches.
• “Alignment for Treatment Using the Synchrony Lung Tracking with Respiratory Modeling”
on page 11-20: You build a Respiratory correlation model for respiratory motion.
If the Synchrony Lung Tracking with Respiratory Modeling algorithm cannot locate the tumor, the
displayed tumor outline in the Live X-ray images will not be aligned with the actual tumor. An out-
of-bounds error may also occur, indicated by an orange box around the row in the Respiratory
Model (mm) table that is not within required tolerances. Also tracking algorithm threshold
violations may occur, indicated by an orange box around the row in the table and around the slider
bar for a threshold parameter.
In this situation, you can attempt one or more of the following:
• Make the following adjustments as appropriate:
Use Offset Mode to drag the tumor outlines in the Live X-ray images and determine
translational offsets. For more information, see “Using Offset Mode” on page 11-27.
You can increase the Tracking Range X and Tracking Range Y parameters. For
more information, see “Adjusting Tracking Range X (mm) and Tracking Range Y
(mm)” on page 11-30.
• You can also consider increasing the value of the Uncertainty threshold parameter if you
are confident that the tumor found using a higher value is correctly identified. For more
information, see “Uncertainty (%)” on page 11-29.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
Synchrony Lung Tracking with Respiratory Modeling algorithm to incorrectly identify the
tumor without generating an error. You must therefore visually verify that the tumor has
been correctly identified throughout the treatment.
• If the tumor is visible in one projection but blocked (for example, by radiopaque structures
such as the spine or heart) in the other projection, or if a large dxAB value and large
detection Uncertainty are calculated for one of the projections, you can consider enabling
the preferred projection method. For more information, see “Enabling Preferred
Projection” on page 11-31.
For more information on image analysis tools to verify that the tumor is correctly located by the
tracking algorithm, see Table 3 on page 11-23. For information on calculated algorithm
parameters, parameter thresholds, and algorithm input parameters, see “Correlation Error (mm)”
on page 11-30.
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory Model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Control Description
Couch: Adjust the position of the Live X-ray image relative to the DRR image by
clicking and dragging in either the Live X-ray image or DRR image. For more
information, see “Chapter 8: Treatment Delivery”. Depending on the treatment
phase, the following icons are displayed:
• Click the Reset icon to move the Live X-ray image back to its original,
unshifted position.
• Click the Reset icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with zero values. The Live X-ray image moves
to its unshifted position, to show existing alignment.
• Click the Offsets icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with the calculated offset values. The Live X-ray
image shifts by the translational offsets, to show the alignment that would
result if the couch is moved by these amounts.
You can toggle back and forth between these icons. The icons only shift the Live
X-ray image relative to the DRR image. They do not move the couch.
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
Control Description
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls.
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Full Content DRR: (Synchrony Lung Tracking with Respiratory Modeling mode)
Displays the unsegmented (full content) DRR images. For more information, see
“Display Modes for DRR Images” on page 11-26.
Control Description
Tumor Region DRR: (Synchrony Lung Tracking with Respiratory Modeling mode;
upper right corner of image panel) Displays the tumor region DRR images, in
which all image content that is not in the immediate region of the tumor is
removed. For more information, see “Display Modes for DRR Images” on
page 11-26.
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
NOTE: These display modes affect the image display only. They
do not affect the results of the Synchrony Lung Tracking with
Respiratory Modeling algorithm.
You can select different display modes for each image (corresponding to Camera B and Camera A
of the X-ray imaging system).
and the Move Couch button as needed. For more information on moving the couch,
see “Chapter 8: Treatment Delivery”.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
These calculated values are also displayed to the left of slider bars (left screen), where you can
adjust their thresholds (see Figure 9). If a parameter value exceeds its threshold, a threshold
violation occurs, and an orange box is displayed around the slider bar. When a threshold violation
occurs the first time you are in the Alignment > Respiratory phase during a treatment
fraction, you cannot move to the next phase until the threshold violation is resolved.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Lung Tracking with Respiratory Modeling algorithm
to produce incorrect estimates of the patient position, which in turn can lead to
misdirected radiation beams. Use caution before changing this parameter value.
If the detection uncertainty value calculated by the Synchrony Lung Tracking with Respiratory
Modeling algorithm is greater than the Uncertainty threshold parameter, an Uncertainty error
occurs (see “Error Handling” on page 11-37).
The default value of the Uncertainty (%) threshold parameter is 40%. If an Uncertainty error occurs,
you can consider increasing the Uncertainty (%) threshold if you are confident that the tracking
algorithm correctly located the target. For more information, see “Correlation Error (mm)” on
page 11-30.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
Synchrony Lung Tracking with Respiratory Modeling algorithm to incorrectly identify the
tumor without generating an error. You must therefore visually verify that the tumor has
been correctly identified throughout the treatment.
The Correlation Error indicates how well a particular model point agrees with the current
Respiratory Model. The user can lower the correlation error threshold value so that a Soft Stop is
triggered at less than the default value of 5 mm. For more information on Respiratory Correlation
Error, see “Chapter 12: Synchrony Respiratory Tracking System (Option)” on page 12-1.
The Tracking Range parameters define the search area within which the Synchrony Lung Tracking
with Respiratory Modeling algorithm searches for the target. During image registration, a matching
window is moved throughout this search area to search for the desired intensity pattern. For more
information on the search area and the matching window, see “Synchrony Lung Tracking with
Respiratory Modeling Algorithm” on page 11-2.
The default Tracking Range X and Tracking Range Y values are 12.0 mm. The Tracking Range is
the distance from the center of the Live X-ray images that the tracking algorithm will search for the
target. Depending on target location and/or motion, users may find it helpful to increase and/or
decrease this value during treatment.
In most cases, the default value will be sufficient.
To adjust the Tracking Range (mm) parameter:
• Click and drag the adjustable cursor on the slider bar to adjust the value (see Figure 9 on
page 11-29). Alternatively, enter a new value in the textbox to the right of the slider bar
and press <Enter>.
• To return to the default value, click the Reset icon. Alternatively, click and drag the
adjustable cursor on the slider bar to the vertical line. The vertical line indicates the
default value.
Images are automatically correlated when a Tracking Range parameter is changed.
Enabling Preferred Projection
The Preferred Projection checkbox (see Figure 9 on page 11-29) allows you to enable or
disable the preferred projection method for locating the target. This checkbox is disabled by
default. When you select this checkbox, an image correlation is performed automatically. For more
information on preferred projection, see “Preferred Projection” on page 11-4.
WARNING: For image correlations when the preferred projection method is invoked, the
dxAB value can no longer be used as an indication of tracking reliability. It is possible for
a tracking failure to occur without producing a dxAB error.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Lung Tracking with Respiratory Modeling algorithm
to produce incorrect estimates of the patient position, which in turn can lead to
misdirected radiation beams. Use caution before changing this parameter value.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
Synchrony Lung Tracking with Respiratory Modeling algorithm to incorrectly identify the
tumor without generating an error. You must therefore visually verify that the tumor has
been correctly identified.
If the Synchrony tracking System will be used to track respiratory motion, you must also build a
Respiratory correlation model before starting treatment delivery. The Respiratory Correlation Error
for all model points is updated dynamically when an image is added, removed, or recorrelated in
the model.
The Skip Respiratory checkbox (Figure 7 on page 11-20) is deselected (cleared) by default so
that the Synchrony tracking System is enabled.
Selecting the Skip Respiratory checkbox disables use of the Synchrony tracking System for
treatment delivery and changes the appearance of the left screen of the Respiratory phase. The
Respiratory Model (mm) table is replaced by the Offsets panel displayed in the
Alignment > Align phase. Offset values for only the most recent image are displayed, and
dashes are displayed for the calculated rotations.
For information on building a Respiratory correlation model and using the Synchrony tracking
System during treatment delivery, see “Chapter 12: Synchrony Respiratory Tracking System
(Option)”.
For information on the optional TxView remote approval workflow, see “Requesting Remote
Approval of Patient Alignment (TxView Option)” on page 8-55.
For information on the optional Physician Setup Approval of Patient Alignment workflow, see
“Physician Setup Approval of Patient Alignment” on page 8-54.
Treatment Delivery
This section covers the following topics:
• “Visual Inspection During Treatment Delivery” on page 11-35
• “Treatment Monitoring Graph” on page 11-36
• “Error Handling” on page 11-37
• “Patient Realignment” on page 11-39
The Delivery phase (left screen) for the Synchrony Lung Tracking with Respiratory Modeling is
shown in Figure 11. Controls on this screen and the right screen allow you to monitor and verify
patient alignment and tracking results, as well as the Respiratory Model, during treatment delivery
using the Synchrony Lung Tracking with Respiratory Modeling.
Because the Synchrony Lung Tracking with Respiratory Modeling algorithm does not estimate
rotational offsets, only translational offsets are displayed and updated on the Left screen in the
Calculated column during treatment delivery automation (treatments that do not use Respiratory
Modeling only). Dashes are displayed for the calculated rotational offsets. In the Applied column,
the rotational offsets that are displayed and used to perform rotation adjustments of the LINAC
during treatment are the rotational corrections obtained from initial Synchrony Spine Tracking
alignment.
For more information on treatment delivery procedures, see “Chapter 8: Treatment Delivery”. For
information on the Synchrony tracking System, see “Chapter 12: Synchrony Respiratory Tracking
System (Option)”. For information on treatment planning for treatment using the Synchrony Lung
Tracking with Respiratory Modeling, see the Treatment Planning Manual.
WARNING: Visually compare the DRR and Live X-ray images to ensure that features are
similar in both. For example, if you attempt Synchrony Lung Tracking with Respiratory
Modeling treatment delivery at the alignment center position instead of the treatment
center position, the images will not be similar.
WARNING: For image correlations when the preferred projection method is used, the
dxAB value can no longer be used as an indication of tracking reliability. It is possible for
a tracking failure to occur without producing a dxAB error.
WARNING: During treatment delivery, after acquisition of Live X-ray images, visually
verify the patient position and verify that tumor is visible in the images. Make sure that
anatomy in the Live X-ray and DRR images matches. For more information on visually
identifying features to be tracked, see “Chapter 8: Treatment Delivery”.
These graphs allow you to observe the values of the above parameters during treatment delivery
as an indication of tracking reliability.
Error Handling
The Synchrony Lung Tracking with Respiratory Modeling algorithm may fail for the following
reasons:
• “Uncertainty Error” on page 11-37
• “dxAB Error” on page 11-38
During patient alignment in the Alignment > Align phase, when the calculated value of an
Synchrony Lung Tracking with Respiratory Modeling algorithm parameter exceeds the value of its
threshold, one of the above threshold violations occurs. An orange box is then displayed around
the slider bar for that parameter threshold.
If this occurs the first time you are in the Alignment > Align phase during a treatment fraction,
you cannot proceed to the next phase until the threshold violation is resolved. After the first time
(for example, if you return to the Alignment > Align phase to realign the patient), if a threshold
violation occurs, you can proceed to the next phase. However, you will not be able to start
treatment delivery until the threshold violation is resolved.
If a threshold violation occurs while treatment is in progress in the Delivery phase, a correlation
error occurs (a Soft Stop). Treatment is paused and the error is described in the Messages panel.
To determine which threshold was violated, you can return to the Alignment > Align phase to
see which slider bar is surrounded by an orange box.
For information on resuming treatment after a Soft Stop, see Chapter 7, “Treatment Delivery”.
Uncertainty Error
An Uncertainty Error occurs when the computed value of the Uncertainty parameter exceeds the
value of the Uncertainty threshold parameter during the Delivery phase. For more information
about the detection uncertainty, see “Uncertainty (%)” on page 11-29.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
Synchrony Lung Tracking with Respiratory Modeling algorithm to incorrectly identify the
tumor without generating an error. You must therefore visually verify that the tumor has
been correctly identified.
dxAB Error
A dxAB error occurs when the calculated value of dxAB exceeds the value of the dxAB threshold
parameter during the Delivery phase. When this error occurs, a message is displayed in the
Messages window. To address a dxAB error, consider reducing the tracking range and/or using
Offset Mode, and enabling the preferred projection method. For more information on dxAB, see
“dxAB (mm)” on page 11-29.
NOTE: Values for the dxAB parameter are not calculated for
image correlations when the preferred projection method is
invoked.
WARNING: If you change the value of the dxAB threshold parameter, it is important to
perform visual inspection after tracking results are displayed. Changing the value of this
parameter can cause the Synchrony Lung Tracking with Respiratory Modeling algorithm
to produce incorrect estimates of the patient position, which in turn can lead to
misdirected radiation beams. Use caution before changing this parameter value.
WARNING: For image correlations when the preferred projection method is invoked, the
dxAB value can no longer be used as an indication of tracking reliability. It is possible for
a tracking failure to occur without producing a dxAB error.
Patient Realignment
The same as for patient alignment, patient realignment has two phases:
• Realignment using Synchrony Spine Tracking.
• Realignment using the Synchrony Lung Tracking with Respiratory Modeling.
An example case when this is necessary is when the patient moves during treatment and global
patient rotational alignment needs to be recalculated.
If Synchrony Lung Tracking with Respiratory modeling is being used and treatment delivery is
interrupted by a Soft Stop, an E-Stop/Interlock, or a Pause, if the Respiratory correlation model is
still valid, you can click the Start/Resume button to resume treatment.
Introduction
The Synchrony Respiratory Tracking System is an option to the CyberKnife Treatment Delivery
System and is intended to enable dynamic image-guided stereotactic radiosurgery and precision
radiotherapy of lesions, tumors and conditions that move with respiration.
This chapter describes how to use the Synchrony Respiratory Tracking System with the
CyberKnife System. It covers the following topics:
• “Synchrony Respiratory Tracking System Components” on page 12-2
• “Synchrony Respiratory Tracking Interface Module” on page 12-2
• “Patient Preparation for Pretreatment Scanning” on page 12-51
• “Patient Setup for Treatment” on page 12-54
• “Patient Alignment and Tracking” on page 12-57
• “Building a Respiratory Correlation Model” on page 12-60
• “Treatment Delivery” on page 12-81
The Synchrony Respiratory Tracking System tracks and compensates for the motion of targets in
areas affected by respiration. These may include targets in the lung, liver, pancreas, and kidneys,
among others. The Synchrony Respiratory Tracking System compensates for the respiratory
motion of the target by correlating target motion with the respiratory motion of the patient and
synchronizing beam delivery with target motion. The treatment robot adjusts and compensates for
these movements to ensure more accurate treatment.
This chapter assumes you are familiar with “Chapter 8: Treatment Delivery”.
Wireless
Synchrony
RoboCouch Respiratory
Hand Tracking
Controller Interface
Charging
Cradle
For more information on the RoboCouch Patient Positioning System, see “Chapter 7: Treatment
Room Operations”.
Charging Cradle
The charging Cradle is a wall-mounted unit for storing and charging the wireless Synchrony
Respiratory Tracking interface module, and also for storing the RoboCouch Hand Controller. An
LED indicator on the front of the Cradle is lit when there is power to the Cradle.
To charge the wireless Synchrony Respiratory Tracking interface module:
• Place it in the Cradle in the recess provided.
Battery Indicator
When the wireless Synchrony Respiratory Tracking interface module is fully charged, it has
approximately 8 hours of battery life. A battery indicator on the front of the interface module
indicates when battery power is low. The interface module also includes an adapter plug for
auxiliary power. The auxiliary power supply replaces battery power but does not charge the battery.
The battery indicator has different states depending on whether the Synchrony Respiratory
Tracking interface module is sitting in the charging Cradle or whether an infrared communication
link has been established with the treatment delivery computer. If the interface module is sitting in
the Cradle, the battery indicator has the following states:
• Blinking yellow: Battery power is low and the battery is recharging.
• Blinking green: The battery is recharging but has not reached its full charge.
• Solid green: The battery is fully charged.
During treatment, when the wireless Synchrony Respiratory Tracking interface module is mounted
in the bracket underneath the RoboCouch headrest, the battery indicator is lit only if the battery
charge is nearing a low charge condition. When the battery indicator is in a blinking yellow state,
the battery has a residual charge of 1 - 2 hours and requires recharging soon.
Communication Indicator
The wireless Synchrony Respiratory Tracking interface module includes a green LED
communication indicator that is lit when there is communication between the IR transmitter and the
interface module. If there is no communication link, the communication indicator will either be off,
indicating no communication, or will blink erratically, indicating a poor communication link. The
position of the Synchrony Respiratory Tracking interface module underneath the RoboCouch
headrest is designed to provide a good communication link. If you observe no communication or a
poor communication link, check the system or call Accuray Customer Support.
Cameras
The Synchrony Respiratory Camera Array includes 3 LED detectors that detect motion of the
Synchrony Respiratory External LED Markers on the Synchrony Respiratory Vest.
As the Synchrony External LED Markers move with patient breathing, signals from the Synchrony
Respiratory Tracking Camera Array are used to determine breathing waveforms. The breathing
waveforms are displayed on the treatment delivery computer.
The patient wears the Synchrony Respiratory Vest during pretreatment scanning and during
Synchrony Respiratory Tracking treatment. The Synchrony Respiratory Vest is shown in Figure 5.
The 3 Synchrony Respiratory External LED Markers attach to the vest during Synchrony
Respiratory Tracking treatment and move with patient breathing.
Cable Management
Velcro Tabs
Velcro Strips
(Supine and Prone)
Tracking Markers
Drawstrings
The Synchrony Respiratory Vest is available in 3 sizes (see Table 1 on page 12-9).
NOTE: The general size guidelines below may not apply to all
patients. Patient body shape may also affect vest size selection.
In such cases, it may be necessary to select a vest that is one
size larger or one size smaller.
Overview Phase
When the Synchrony Respiratory Tracking System is used, the right screen of the Overview
phase includes Synchrony Respiratory Tracking controls and displays (See Figure 6). These
displays allow you to monitor the signals from the Synchrony Respiratory External LED Markers
and also to monitor the Respiratory Model if you have returned to the Overview phase from a
later phase in which you have built a Respiratory Model.
The following Synchrony Respiratory Modeling controls and displays are available:
• Markers control area: 3 Marker checkboxes allow you to enable or disable Tracking
Markers. A solid green square to the left of a checkbox indicates that a Tracking Marker is
currently visible to the Synchrony Respiratory Camera Array. A solid yellow square
indicates that it is currently not visible. A status bar above each checkbox shows the
visibility history of each Tracking Marker for the last 120 seconds. For more information,
see “Markers Control Area” on page 12-36.
• Summary Info section: A Respiratory Phase indicator shows which of the 4
Respiratory Model points required to proceed with treatment delivery (the Peak and
Valley model points, and the Center inhale and exhale model points) have been acquired.
For more information, see “Respiratory Phase Indicator” on page 12-16".
• Model Information section: Values for Coverage (%), Average Standard Error (mm),
Maximum Standard Error (mm); the number of Model Points; and the Model State are
displayed for the current Respiratory Model. For more information on these Respiratory
Model details, see “Respiratory Model Table” on page 12-13 and “Respiratory Details
Tab” on page 12-31.
• Tracking Markers graph: Waveforms representing breathing patterns for each Tracking
Marker enabled for tracking are displayed. Model points are shown as dots on the
waveforms. If at least 2 model points have been acquired, Amplitude indicators are also
displayed showing the respiratory range of motion covered by the current Respiratory
Model.
Controls allow you to pause and resume monitoring of Tracking Marker motion. You can
also pan the graph and change the time scale. For more information, see “Tracking
Markers Graph” on page 12-37".
• Comet graphs: If at least 2 model points have been acquired, 2 Comet graphs provide a
virtual see-through view of the patient from the coronal and sagittal perspectives,
indicated by a human body icon. These views give a qualitative picture of how well the
model points fit the current Respiratory Model. For more information, see “Comet Graphs”
on page 12-30.
Respiratory Phase
This section covers the following topics:
• “Respiratory Model Table” on page 12-13
• “Image Display Area” on page 12-17
• “Algorithm Parameters and Thresholds” on page 12-22
• “Manual Setup Controls” on page 12-22
• “Automatic Setup Controls (Option)” on page 12-24
• “Rotational Bounds Checking Enabled Checkbox” on page 12-25
• “Skip Respiratory Checkbox” on page 12-25
• “Reset Respiratory Model Button” on page 12-26
• “Image History Table” on page 12-28
• “Breathing Waveform Snippet” on page 12-29
• “Comet Graphs” on page 12-30
• “Respiratory Details Tab” on page 12-31
• “Acquisition Details Tab” on page 12-35
• “Images Tab” on page 12-35
• “Markers Control Area” on page 12-36
• “Tracking Markers Graph” on page 12-37
This section describes the controls and displays in the Alignment > Respiratory phase for
building and monitoring a Respiratory Model.
The table has 15 rows, corresponding to all 15 model points. Model points are listed in the phase
order they would appear in a single respiratory cycle. (They are not listed in the order they were
acquired.) This is useful when you use Movie Mode to improve the visibility of tumor motion by
cycling through all the images in respiratory phase order. Movie Mode helps you verify that the
tracking algorithm has correctly identified and is correctly tracking the target. For more information
on Movie Mode, see Table 2 on page 12-17.
The currently selected row is highlighted in yellow and the corresponding image is displayed. A red
X indicates that an image is marked as blocked.
Rows in the Respiratory Model (mm) table display the following information:
• Respiratory Correlation Error (column 1): This value (in millimeters) indicates how well a
particular model point agrees with the current Respiratory Model.
The Respiratory Correlation Error for all model points is updated when an image is
added, removed, or recorrelated in the model.
• Offsets (columns 2 to 4): Calculated Offset values (in millimeters) between the Live X-ray
and DRR images, for each image acquisition in the Respiratory Model.
Arrows for translations point in the positive direction. The human body icon matches the
patient orientation in the treatment plan.
Calculated Offset values in the Respiratory Model (mm) table indicate couch movement that
would be needed to align features in the Live X-ray and DRR images. In the Respiratory phase,
images are not automatically shifted by the Offset amounts.
An orange box around a row indicates:
• an offset is not within required tolerances (an out-of-bounds error), or
• a tracking algorithm threshold violation has occurred.
Rows with an orange box around them are considered candidates for the Respiratory Model but
are not included in the model. If the error or threshold violation is resolved, the model point is then
automatically added to the Respiratory Model.
Selecting a row in the Respiratory Model (mm) table highlights the corresponding row in the
Image History table. For more information, see “Image History Table” on page 12-28.
Rotation Corrections
The Respiratory Model (mm) table displays rotational corrections which depend on the
tracking mode:
• Average Rotation Correction (deg): (Synchrony Fiducial Tracking with
Respiratory Modeling and Synchrony Spine Tracking Prone with Respiratory Modeling)
The average of the rotational values for images in the current Respiratory Model (see
Figure 7 on page 12-13).
• Alignment Rotation Correction (deg): (Synchrony Lung Tracking with
Respiratory Modeling mode) The rotational values for the last image acquired in the
Alignment > Spine phase (see Figure 8). If spine alignment was skipped, dashes are
displayed.
At a minimum, the 4 model points corresponding to the Peak and Valley and Center Inhale and
Center Exhale are required to proceed to the Readiness phase, as well as a Valid (Optimal or
Sub-Optimal) Model State.
Model Information
The Model Information section displays the following information:
• Avg Std Err (mm): The average standard error over all enabled Tracking Markers of
the Respiratory Model.
• Max Std Err (mm): The maximum standard error for all enabled Tracking Markers of
the Respiratory Model.
The standard error of the Respiratory correlation model is a good indicator of the overall
difference between the Respiratory Model curve and the individual model points that
represent each correlation image. Standard Error, estd, is defined in Equation (1):
e
2
i (1)
i
e std = ------------
n
where:
ei is the difference between the model curve and the model point, and n is the number of
model points.
• Model Information Menu Button: clicking this will display more detailed
information about the model state.
• Model State: Optimal, Sub-Optimal, or Invalid.
For information on the different model states, see “Building a Respiratory Correlation
Model” on page 12-60.
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory Model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked
correctly.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Control Description
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only. For more information, see “Chapter 8:
Treatment Delivery”.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
(Synchrony Fiducial Tracking mode with Respiratory Modeling only) You may need
to modify the Window/Level settings for the Live X-ray images to see the fiducials.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
ITTV Outline: (Lung 1 View Tracking with Respiratory Modeling only) If the
treatment plan is based on a Simulation Plan, displays a dark blue Internal Target
Tracking Volume (ITTV) contour in the DRR image that is not used for tracking.
For more information, see the Treatment Planning Manual.
The ITTV Outline is turned off by default. See “Chapter 15: Lung Optimized
Treatment (Option)”.
Control Description
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls. For more information,
see “Chapter 8: Treatment Delivery”.
Fiducial ROI: (Synchrony Fiducial Tracking with Respiratory Modeling mode only)
Displays a square yellow box in the Live X-ray image for each Region of Interest
(ROI) where the fiducial extraction algorithm searches for fiducial candidates.
Each box is centered on a fiducial candidate.
The size of the box changes when the Tracking Range (mm) parameter is
adjusted. For more information, see “Chapter 9: Synchrony Fiducial Tracking”.
Control Description
Align Center: (Synchrony Spine Tracking Prone with Respiratory Modeling only)
Displays a marker that corresponds to the align center selected during treatment
planning. See “Chapter 10: Synchrony Spine Tracking (Option)”.
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Full Content DRR: (Synchrony Lung Tracking with Respiratory Modeling mode
and Lung 1 View Tracking with Respiratory Modeling mode only) Displays the
unsegmented (full content) DRR images. See “Chapter 11: Synchrony Lung
Tracking with Respiratory Modeling (Options)”, and “Chapter 15: Lung Optimized
Treatment (Option)”.
Tumor Region DRR: (Synchrony Lung Tracking with Respiratory Modeling mode
and Lung 1 View Tracking with Respiratory Modeling mode; upper right corner of
image panel) Displays the tumor region DRR images, in which all image content
that is not in the immediate region of the tumor is removed. See “Chapter 11:
Synchrony Lung Tracking with Respiratory Modeling (Options)”, and “Chapter 15:
Lung Optimized Treatment (Option)”.
Control Description
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
Move
Couch
Button
The Phase Selection icon allows you to turn phase selection on. With phase selection, the
Synchrony Respiratory Tracking System attempts to acquire model points at targeted phases of
the respiratory cycle by automatically controlling the timing of image acquisition. The No Phase
Selection icon allows you to turn phase selection off.
The Waveform display represents a single respiratory cycle. Circles in the waveform represent
phases of the respiratory cycle that you click to select when using the User-defined Respiratory
Phase Selection method. For step-by-step instructions on building a Respiratory Model using the
Manual Setup tab, see “Using Manual Setup” on page 12-63
Click the Move Couch button to move the couch to the target position. Values in the textboxes
change and approach zero as the desired couch position is achieved.
After you move the couch, you must build a new Respiratory Model.
For more information on moving the couch using the automatic patient positioning controls, see
“Chapter 8: Treatment Delivery”.
Peak/Valley
Icon
Move Couch
Button
Initially only the Acquire Peak and Valley Image button is enabled, allowing you to perform
Respiratory alignment. After Peak and Valley Images are acquired, the Dataset Acquisition
button is enabled. For more information on the Automatic Setup option, see “Using Automatic
Setup (Option)” on page 12-73.
WARNING: Use caution before disabling Synchrony Respiratory Tracking mode. If you
disable Synchrony Respiratory Tracking mode, respiratory motion will not be modeled or
corrected for during treatment.
To disable Synchrony Respiratory Tracking mode to deliver treatment that does not use
Respiratory Modeling:
• Click to select the Skip Respiratory checkbox on the Respiratory screen.
Selecting the Skip Respiratory checkbox changes the appearance of the left screen
of the Respiratory phase. The Respiratory Model (mm) table is replaced by the
Offsets panel displayed in the Alignment > Align phase. Offset values for only the
most recent image are displayed.
To re-enable Synchrony Respiratory Tracking mode to deliver treatments that use
Respiratory Modeling:
• Click to deselect the Skip Respiratory checkbox on the Respiratory screen. The
existing Respiratory Model is re-used.
WARNING: You must reset the Respiratory Model if you make any of the above changes.
Otherwise, targeting accuracy will be compromised.
WARNING: If you adjust the position of the Synchrony Respiratory Camera Array, you
must reset the Respiratory Model even if you have not yet established a Respiratory
Model. Otherwise, it is possible for the treatment robot to exceed displacement limits
without generating an error, and beam mistargeting can result.
To invalidate the current Respiratory Model and delete all model points:
1. Click the Reset Respiratory Model button in the Respiratory phase (see
Figure 12).
2. Click the Yes button to confirm deletion of the current Respiratory Model.
The current Respiratory Model will be invalidated and all model points deleted from it. The
Respiratory Model (mm) table will be cleared. You can then build a new model.
Deleted model points will still be listed in the Image History table but can no longer be
re-used in the Respiratory Model.
WARNING: When treating using the Synchrony Respiratory Tracking System, you must
reset the Respiratory Model after any adjustment of the patient or treatment couch
position using the treatment couch Hand Controller. Then you must rebuild and validate a
new Respiratory Model. Failure to reset and reestablish the Respiratory Model can result
in inaccurate radiation delivery and harm to the patient.
Comet Graphs
The Comet graphs to the right of the Image History table show a virtual see-through view of
the patient from the coronal and sagittal perspectives, indicated by the human body icon (see
Figure 15 and Figure 16).
The vertical axis represents the Inferior/Superior direction in the patient coordinate system. The
two views give a qualitative picture of how well the model points fit the current Respiratory Model:
• The moving “comet” represents average movement of the Tracking Markers in real-time
and shows where the Respiratory Model indicates the tumor is right now and where the
treatment robot would be delivering beam with the current model.
• The static points represent where images in the Respiratory Model were acquired.
• The “tail” of the comet shows the shape of the Respiratory Model. Figure 15 shows an
example for a Linear model. Figure 16 shows an example for a Dual Poly model. The
quality of the Respiratory Model is indicated by how far the static points fall from the tail.
For more information on model types, see “XYZ vs. R Correlation Graphs” on
page 12-32.
Each correlation graph shows movement of the target (fiducial configuration or tumor) vs.
movement of each enabled Tracking Marker, for the current Respiratory Model. Each Tracking
Marker is represented by a different color:
• red: Tracking Marker 1
• green: Tracking Marker 2
• lavender: Tracking Marker 3
On each graph, a point is displayed for each model point in the Respiratory Model. The Synchrony
Respiratory Tracking System chooses one of three model types to minimize the overall correlation
error by determining the best-fit straight line, curvilinear line, or dual polynomial through these
points:
• Linear: A linear best-fit line, when target motion is linear. (See example in Figure 17.)
• Curved Poly: A curvilinear line, when the target moves back and forth along an arc.
• Dual Poly: A dual polynomial, when the target moves along an arc and uses different
paths during inspiration and expiration. (See example in Figure 17.)
The model type for each axis of the patient coordinate system is displayed below each graph. For
a Linear model, a good visual indication of a high-quality Respiratory Model is a correlation graph
with all of its points lying on the line.
The vertical axis of the correlation graphs represents the target offsets calculated by the imaging
system for each of the three principal axes of the patient coordinate system.
The horizontal axis of each correlation graph represents movement, R, of each Tracking Marker,
as described in the section “Tracking Marker Movement "R"” on page 12-34.
The following considerations are important when evaluating the correlation graphs:
• Whether all model points for a Tracking Marker lie on a line or curve. You can examine
any outlier model point by clicking the corresponding row in the Image History table
to view the Live X-ray image. Verify that the target has been correctly identified. Evaluate
any changes in the breathing pattern to determine whether to reset the Respiratory Model
and rebuild it. The model point in question can also be removed from the Respiratory
Model if needed (see “Image History Table” on page 12-28).
• Whether the slope of the line or curve is steep (more vertical in the graph). A steep slope
indicates that a small Tracking Marker motion is being used to predict large target motion.
In this case, you can evaluate placement of the Tracking Markers (see “Evaluating
Placement of Tracking Markers” on page 12-79).
Marker 2
Camera
Coordinate
Marker 1 System
Patient
Coordinate
System
Marker 3
Target
Couch
The horizontal axis of each correlation graph represents the movement of each Tracking Marker
projected onto a single fixed direction in space. The projected movement is labeled R in Figure 19.
The fixed direction chosen for one Tracking Marker may be independent of the direction chosen
for another Tracking Marker.
The direction in space for each Tracking Marker is chosen as follows: Over several breathing
cycles, Over several breathing cycles, the marker positions are collected and used to determine
the direction of the marker’s major motion. This direction is used as the reference direction for that
Tracking Marker. If the reference direction is close to parallel with the Yc axis of the Synchrony
Respiratory Camera Array (see Figure 19), the camera Yc axis is used as the reference direction
instead. The Yc axis typically exhibits the most Tracking Marker motion. In general, depending on
positioning of the Synchrony Respiratory Camera Array, the Yc axis is not far from parallel with the
Z axis of the patient coordinate system (Anterior/Posterior).
Images Tab
The Images tab below the Image History table displays the Live X-ray images and the X-ray
technique corresponding to the currently selected row in the table. Select controls at the bottom of
the image to aid image analysis (see Table 2 on page 12-17).
Status bars show the visibility history of each marker for the last 120 seconds. History data fills in
from right-to-left:
•
Green: Tracking Marker is visible to the Synchrony Respiratory Camera Array.
•
Yellow: Tracking Marker is not visible to the Synchrony Respiratory Camera Array.
•
Mixed green/yellow: Tracking Marker is intermittently visible/not visible.
A solid green square to the left of a checkbox indicates that a Tracking Marker is currently visible
to the Synchrony Respiratory Camera Array. A solid yellow square indicates that it is currently not
visible.
An orange box around the Markers checkboxes indicates that a marker is not visible. The
visibility history for a Tracking Marker should be solid green. If not, you should disable that Tracking
Marker so it is not used when building a Respiratory Model. You can also evaluate the positioning
of the Tracking Markers and the Synchrony Respiratory Camera Array.
In some cases, disabling a Tracking Marker can cause one or more existing model points to be
removed from the Respiratory Model. This occurs if the remaining enabled Tracking Markers were
not enabled when those model points were acquired. Table 4 gives an example of a Respiratory
Model in which 4 model points were acquired when one or more Tracking Markers were disabled.
If Marker 1 is now disabled, Model Point 1 will be removed from the Respiratory Model. If both
Markers 1 and 2 are now disabled, Model Points 1 through 4 will be removed from the Respiratory
Model. The Coverage (%) value can also change (see “Respiratory Details Tab” on
page 12-31).
Amplitude Indicators
Three Amplitude indicators to the left of the Tracking Markers graph (see Figure 22) show how
much of the total respiratory range of motion (maximum to minimum amplitude) is covered by the
current Respiratory Model. At least 2 model points are required to see the Amplitude indicators.
Colors indicate the covered respiratory range as follows (see Figure 23):
• Orange: The orange bar represents the sampled maximum to minimum range of the
respiratory markers during patient respiration.
• Green: The superimposed green bar represents the amplitude range of the camera array
images in the current Respiratory Model relative to the orange sampled range.
There is one Amplitude indicator for each Tracking Marker (from left-to-right, for Tracking
Markers 1, 2 and 3). Each Amplitude indicator bar moves up and down as the Tracking Marker
moves.
Readiness Phase
When the Synchrony Respiratory Tracking System is used, the left screen of the Readiness
phase includes the Image Burst Enabled checkbox (see Figure 24).
Selecting this checkbox allows the Image Burst option to be used during treatment delivery
automation. The Image Burst option is disabled (the checkbox is inactive) if the Synchrony
Respiratory Tracking System is disabled. For more information on the Image Burst option during
treatment delivery, see “Left and Right Screens of the Delivery Phase” on page 12-41".
If the Skip Respiratory checkbox is selected in the Alignment > Respiratory phase, the
text Respiratory Skipped is displayed in the Readiness phase (see Figure 25). For more
information on the Skip Respiratory checkbox, see “Skip Respiratory Checkbox” on page 12-25.
In Figure 27, the Respiratory Model (mm) table displays offsets for all Respiratory Model
points. The Treatment Monitoring graphs below the image panels include graphs for Lung tracking
parameters, couch Position (mm), couch Orientation (deg), and Respiratory Model
Standard Error (mm). In this example, Image Burst is enabled.
Figure 28 shows the left screen of the Delivery phase when the Synchrony Respiratory Tracking
System is disabled.
The left screen is similar to the screen in the Alignment > Align phase. The Respiratory
Model (mm) table is replaced by the Offsets panel. Offsets for only the most current image
acquisition are displayed. The Image Burst option is disabled and the Image Interval (sec)
slider bar is displayed instead.
For the Synchrony Lung Tracking with Respiratory Modeling, rotational offsets are not calculated
during treatment. The rotational corrections obtained from initial Spine alignment are used to
perform rotation adjustments of the LINAC during treatment delivery. To indicate this, dashes are
displayed for the calculated rotations in the Offsets panel.
Figure 29 shows the right screen of the Delivery phase with the Synchrony Respiratory Tracking
System. Information about the current Respiratory Model is displayed even when the Synchrony
Respiratory Tracking System is disabled.
The sections below describe the Respiratory controls and displays available on these screens.
If a Respiratory Model becomes invalid, beam delivery is interrupted. When treatment resumes,
image acquisition and image correlation are automatically performed before the next beam
delivery and the Respiratory Model is refreshed. The Respiratory Model may become invalid
because of patient motion, camera shift, marker shift, excessive marker motion, if a targeted phase
is missed by newly acquired images, the model type changes from Linear or Curved Poly to Dual
Poly, or when disabled markers are re-enabled.
The Image Burst option can only be enabled if the Synchrony Respiratory Tracking System is
enabled. If you disable the Synchrony Respiratory Tracking System in the Alignment >
Respiratory phase, the Image Burst Interval (sec) slider bar changes to the Image
Interval (sec) slider bar, with a default Image Interval of one image every 15 seconds (see
Figure 33). Also, if you disable imaging in the Readiness phase by selecting the Disable
Imaging checkbox, the Image Burst option cannot be enabled.
• Respiratory Model Standard Error (mm) graph: This graph shows the Average
Standard Error (mm) and the Maximum Standard Error (mm) over enabled markers for
the Respiratory Model as a function of time. The Average Standard Error (mm) and
Maximum Standard Error (mm) for the current Respiratory Model are also displayed In
the Model Information.
A solid green square to the left of a checkbox indicates that a Synchrony Respiratory External LED
Marker is currently visible to the Synchrony Respiratory Camera Array. A sold yellow square
indicates that it is currently not visible. A status bar above each checkbox shows the visibility
history of each Tracking Marker for the last 120 seconds. For more information, see “Markers
Control Area” on page 12-36.
For more information on these Respiratory Model details, see “Respiratory Model Table” on
page 12-13 and “Respiratory Details Tab” on page 12-31.
Model points are shown as dots on the waveforms. If at least 2 model points have been acquired,
Amplitude indicators are also displayed showing the respiratory range of motion covered by the
current Respiratory Model.
Controls allow you to pause and resume monitoring of Tracking Marker motion. You can also pan
the graph and change the time scale. For more information, see “Tracking Markers Graph” on
page 12-37".
Comet Graphs
If at least 2 model points have been acquired, the Comet graphs provide a virtual see-through view
of the patient from the coronal and sagittal perspectives, indicated by a human body icon.
These views give a qualitative picture of how well the model points fit the current Respiratory
Model. For more information, see “Comet Graphs” on page 12-30.
NOTE: Only unzip the vest to remove it. DO NOT readjust the
drawstrings.
Pretreatment Scanning
The patient must wear the Synchrony Respiratory Vest during all pretreatment scanning, including
during CT, MRI, and PET scans.
Follow standard protocol guidelines for the appropriate pretreatment scan for extracranial
applications. For more information on pretreatment scanning guidelines, see “Chapter 6: Patient
Preparation”.
To acquire pretreatment scans for treatment planning when Synchrony Respiratory
Modeling will be used:
1. Instruct the patient to breathe normally until scan acquisition begins, at which time the
patient must hold their breath at end of normal expiration.
Treatment Planning
The Synchrony Respiratory Tracking treatment option must be specified in the patient treatment
plan. For information on creating a treatment plan that specifies Synchrony Respiratory Modeling
treatment, see the Treatment Planning Manual.
• In the supine position, this location is typically the diaphragm area, but possibly may
be the lower thoracic area (for "chest breathers") or the abdominal area (for "stomach
breathers").
• In the prone position, this location is typically in the thoracic area on the back of the
patient, but can also be in the lumbar area.
2. Attach the 3 Tracking Markers to the white Velcro strips on the Synchrony Respiratory
Vest at the location of maximum body motion due to respiration (see Figure 41).
3. Secure the cables of the Tracking Marker harness using the cable management Velcro
tabs on the vest. (See Figure 41.)
4. Connect the Tracking Marker harness connector to the Synchrony Respiratory interface
module.
CAUTION: Moving the Synchrony Respiratory Camera Array forward (closer to the patient) can
increase the potential for collision between it and the treatment robot.
Later, after building a Respiratory correlation model, you should evaluate the placement of the
Tracking Markers by reviewing the breathing waveforms and the correlation error values (see
“Evaluating Placement of Tracking Markers” on page 12-79).
Final evaluation of the placement of the Tracking Markers should take place after patient alignment
and identification of the target have been completed.
WARNING: Verify that this patient is in fact the one to be treated and the treatment plan is
the correct one to use.
• If the plan information is correct, click the second Data Verified button. The
fraction information in the Data Verified checkbox becomes active.
• If the fraction information is correct, click the third Data Verified button.
• If any information is not correct, click the Error Information button (see
Figure 43) and exit the treatment application.
Error Information
Button
Peaks and valleys in the waveforms represent the two ends of the respiratory cycle (Start
Inspiration/Expiration and End Inspiration/Expiration). Depending on the motion of each Tracking
Marker, peaks may correspond to full inspiration in one waveform and full expiration in another
waveform. The respiratory cycle is divided into 8 non-overlapping time regions called phases.
Each Live X-ray image acquired by the Treatment Delivery System adds a point to the Respiratory
correlation model. Each model point is classified as belonging to one of the 8 phases. For an
accurate and robust model, it is critical that the model points are distributed evenly and cover the
whole range of respiratory motion. Accomplishing this distribution requires timing X-ray image
acquisition with breathing motion.
A minimum of 6 model points (including the Peak and Valley model points and the Center Inhale
and Center Exhale model points) is required to build a complete Respiratory Model.
Acquisition of at least 12 images is recommended. A maximum of 15 images (15 model points) can
be stored in the model dataset at a time covering some or all of the 8 phases in the respiratory
cycle. As additional images are acquired, a first-in, first-out (FIFO) approach is used to update the
dataset. After a Respiratory Model is built, the Respiratory Model dataset is updated during
treatment delivery with each new X-ray image acquired.
In Figure 47, the top waveform shows model points (representing X-ray images) acquired at full
inspiration and expiration (the Peak and Valley model points). The middle waveform shows model
points acquired at the midpoint of inspiration and expiration (the Center Inhale and Center Exhale
model points) and additional model points acquired in the respiratory cycle. The bottom waveform
shows the automatic image burst of three model points when treatment delivery begins.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information will be added to
the Respiratory correlation model.
With the above criteria, a Respiratory Model is categorized as one of the following Model States in
the Respiratory Model (mm) table of the Respiratory phase (see Figure 47):
• Optimal: Contains 8 or more model points (for Linear and Curved Poly model types), 12
or more model points (for Dual Poly model type), including 4 specific respiratory phases
(the Peak/Valley and Center Inhale/Exhale model points and the midpoints between
them). When an Optimal model is established, the Readiness tab is active and you can
proceed with treatment.
• Sub-optimal: This model state is only relevant for Linear and Curved Poly model types.
Contains 8 or more model points, including 4 specific respiratory phases (the Peak/Valley
and the Center Inhale/Exhale model points) and covers greater than 40% of the initial
respiratory marker range. When a Sub-Optimal model is established, the Readiness
tab is activated, however a Delivery Warning Message must be resolved or
acknowledged and confirmed before you can proceed with treatment. For information on
confirming delivery warning messages see, “Checking Equipment Readiness” on
page 8-67.
• Invalid: Contains fewer than 6 model points (for Linear and Curved Poly model types), or
fewer than 12 model points (for Dual Poly model type), or covers less than 40% of the
initial respiratory marker range, or is missing one of the 4 following respiratory phases
(the Peak/Valley and Center Inhale/Exhale model points and the midpoints between
them). The Readiness tab is disabled and you cannot proceed with treatment.
Detailed information about the state of the model and the actions required to achieve an Optimal
model state are displayed when clicking the Model Information Menu Button. The
Synchrony Respiratory Tracking System provides two ways to build a Respiratory Model,
described in the sections that follow:
• Manual Setup: You initiate each individual image acquisition and build a Respiratory
Model one model point at a time. See “Using Manual Setup” on page 12-63
• Automatic Setup (option): With this option, you can acquire a complete Respiratory
dataset automatically in two steps. See “Using Automatic Setup (Option)” on page 12-73
• No Phase Selection Phase selection is disabled so that timing of the next image
acquisition is not phase-driven.
With Image-by-Image Phase Selection, model points are acquired in the following order (see
Figure 48):
1. The first 2 model points are acquired at the maximum and minimum amplitudes of the
respiratory cycle (the Peak and Valley model points) to obtain good coverage at the two
extremes.
2. Next, 2 model points are acquired at the midpoints of the respiratory cycle (the Center
Inhale and Center Exhale model points).
3. Finally, the last 4 model points are acquired at amplitudes halfway between the above 2
phases and the midpoints of the respiratory cycle, to obtain good coverage on the inhale
and exhale sides.
If the Respiratory Model is missing one or more of the 8 phases, the Synchrony Respiratory
Tracking System targets the missing phases in the order given above. If the model contains the
maximum of 15 model points, the next model point replaces the oldest model point in the dataset.
If any phases are missing in the dataset, the missing phases are targeted in the order given above.
If there are no missing phases, the next model point is acquired at a random phase. The oldest
model point is discarded before the new model point is acquired.
The Synchrony Respiratory Tracking System continuously re-evaluates the existing model points,
regardless of whether a new model point has been acquired. The existing model is compared with
the actual breathing pattern determined by the current motion of the Tracking Markers. Changes
in the breathing pattern can result in changes in the phase distribution of existing model points. For
example, a model point that has been classified as a Peak may no longer be at the maximum
amplitude of respiratory motion. This re-evaluation of the existing model is used to reclassify the
phases of the model points to determine the next phase that will be triggered by Phase Selection.
It is possible for the Synchrony Respiratory Tracking System to miss a phase it has targeted. In
this case, the new model point is still used in the model, even though one or more holes remain in
the distribution of phases. The missing phase will be retargeted in the next image acquisition.
You can enable and disable Phase Selection at any time in the Respiratory phase when using
Manual Setup.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
2. Use User-defined Respiratory Phase Selection to acquire both the Peak and Valley
model points (full inspiration/expiration) and either the Center Inhale or Center Exhale
model point (the midpoint of the respiratory cycle). For instructions, see “User Defined
Respiratory Phase Selection” on page 12-69.
Visually compare the Live X-ray and DRR images to ensure that features are similar in
both. Ensure that the target to be tracked is correctly identified in the Live X-ray images
by the tracking algorithm.
3. Review the translational offsets of the Center Inhale or Center Exhale image in the
Respiratory Model (mm) table.
If the values are close to zero, it is not necessary to move the couch.
If the values are high in one or more axes, you can move the couch to center the range of
respiratory motion. Enter the translational offsets of the Center Inhale/Exhale image into
the Couch Position (mm) textboxes.
4. Click the Move Couch button to move the patient to the respiratory center. For more
information, see “Move Couch Controls” on page 12-23.
5. Re-acquire the Center Inhale or Center Exhale image to verify that the patient is aligned
near the respiratory center.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
Visually compare the Live X-ray and DRR images to ensure that features are similar in
both. Ensure that the target to be tracked is correctly identified in the Live X-ray images
by the tracking algorithm.
4. Repeat the above step to acquire images throughout the respiratory cycle.
5. Monitor Respiratory modeling as you acquire images:
• Check the Respiratory Phase indicator. As the Peak/Valley and Center Inhale/Exhale
model points are acquired, the Respiratory Phase indicator is updated (see
Figure 51).
Verify using Movie Mode (see Table 2 on page 12-17) that the tracking algorithm has
correctly identified the target to be tracked.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information will be added to
the Respiratory correlation model.
To build a Respiratory Model with the User Defined Respiratory Phase method:
1. Click the Manual Setup tab (see Figure 49).
2. Click a circle in the waveform display to select a phase in the respiratory cycle. (see
Figure 52).
Visually compare the Live X-ray and DRR images to ensure that features are similar in
both. Ensure that the target to be tracked is correctly identified in the Live X-ray images
by the tracking algorithm.
4. Repeat Step 2 and Step 3 above to continue to acquire images throughout the respiratory
cycle at phases you select.
5. Monitor Respiratory modeling as you acquire images:
• Check the Respiratory Phase indicator. As the Peak/Valley and Center Inhale/Exhale
model points are acquired, the Respiratory Phase indicator is updated (see
Figure 51).
At a minimum, the Respiratory Model must be valid (containing at least 6 model
points for Linear and Curved Poly models, or 12 model points for Dual Poly models,
and must include the Peak/Valley and the Center Inhale/Exhale model points) before
you can proceed with treatment delivery.
• Review the Respiratory Model (mm) table. See “Respiratory Model Table” on
page 12-44.
• Evaluate the breathing waveforms and the Amplitude indicators in the Tracking
Markers graph. See “Tracking Markers Graph” on page 12-37.
6. After acquiring a Respiratory dataset, you analyze the Respiratory Model. See “Analyzing
the Respiratory Model” on page 12-77.
Verify using Movie Mode (see Table 2 on page 12-17) that the tracking algorithm has
correctly identified the target to be tracked.
No Phase Selection
With No Phase Selection, timing is not automatically phase-driven. The Synchrony Respiratory
Tracking System acquires a Live X-ray image as soon as it can after you click the Acquire
Image button, without waiting for a particular phase (see Figure 54). This method is helpful for
patients with irregular breathing patterns, when it can be difficult for the Synchrony Respiratory
Tracking System to acquire a model point at a desired phase using Phase Selection.
To build an accurate and robust Respiratory Model with Phase Selection disabled, the model
points should be distributed evenly and cover the whole range of respiration for maximum
coverage. Follow these guidelines:
• Acquire 2 - 4 images spaced evenly throughout the inspiration phase.
End Expiration
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
• Acquire the remaining 4 model points at amplitudes halfway between the above
locations.
The Readiness phase is activated when you have built a valid Respiratory Model
(containing at least 6 model points for Linear and Curved Poly models, or 12 model points
for Dual Poly models, including the Peak/Valley and the Center Inhale/Exhale model
points). The Model State in the Model Information section of the Respiratory
Model (mm) table will be listed as Optimal or Sub-optimal.
6. Monitor Respiratory modeling as you acquire images:
• Check the Respiratory Phase indicator. As the Peak/Valley and Center Inhale/Exhale
model points are acquired, the Respiratory Phase indicator is updated (see
Figure 51).
At a minimum, the Respiratory model must be valid (containing at least 6 model
points for Linear and Curved Poly models, or 12 model points for Dual Poly models,
and must include the Peak/Valley and the Center Inhale/Exhale model points) before
you can proceed with treatment delivery.
• Review the Respiratory Model (mm) table. See “Respiratory Model Table” on
page 12-44.
• Evaluate the breathing waveforms and the Amplitude indicators in the Tracking
Markers graph. See “Tracking Markers Graph” on page 12-37.
7. Continue to acquire images to refine the Respiratory Model.
After the Respiratory Model is validated, you can acquire additional images at different
phases of the respiratory cycle to add to the Respiratory dataset, up to a maximum of 15
model points.
8. After acquiring a Respiratory dataset, you analyze the Respiratory Model. See “Analyzing
the Respiratory Model” on page 12-77.
Verify using Movie Mode (see Table 2 on page 12-17) that the tracking algorithm has
correctly identified the target to be tracked.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
2. Click the Acquire Peak and Valley Image button in the Respiratory
Alignment section.
Two Live X-ray images are taken corresponding to full inspiration/expiration. Images are
displayed on the left screen as they are acquired.
Visually compare the Live X-ray and DRR images to ensure that features are similar in
both. Ensure that the target to be tracked is correctly identified in the Live X-ray images
by the tracking algorithm.
The Peak/Valley icon shows green dots when the Peak and Valley model points of the
respiratory cycle have been acquired
5. If you moved the couch, repeat Step 2 to reacquire the Peak/Valley model points.
After Respiratory alignment, the Acquire Dataset button is enabled in the Dataset
Acquisition section (see Figure 56), allowing automatic acquisition of remaining Respiratory
model points.
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
NOTE:
Acquire
Dataset
Button
You only need to click the button once. The 2 Peak and Valley model points must exist to
enable this button.
Live X-ray images are acquired in a rapid burst at 1 second intervals. 3 images are taken
at the end for any missing respiratory phases.
As images are acquired to complete the Respiratory Model, a Wait time in seconds may
be displayed when acquisition is used for extended periods, which is the countdown time
remaining before the Acquire Dataset button is re-enabled for another burst of
imaging (see Figure 57).
WARNING: When Live X-ray images are being acquired, you must verify that the target is
being identified and tracked correctly. Otherwise, inaccurate information may be added to
the Respiratory correlation model.
The user can lower the correlation error threshold value so that a Soft Stop is triggered at less than
the default value of 5 mm.
For Synchrony Fiducial Tracking with Respiratory Modeling and for Synchrony Lung Tracking with
Respiratory Modeling, the threshold value can be set between 2.0 mm and 5.0 mm. For Synchrony
Spine Tracking Prone with Respiratory Modeling, the threshold can be decreased from 2.5 mm to
1.5 mm.
If the correlation error is close to the threshold value, consider turning off a Tracking Marker, or
repositioning the Tracking Markers or the Synchrony Respiratory Camera Array. Then rebuild the
Respiratory correlation model and reevaluate the correlation error.
When the Synchrony Respiratory External LED Markers are placed properly and an acceptable
Respiratory correlation model has been built, patient setup is complete and you are ready to start
treatment delivery, as described in “Treatment Delivery” on page 12-81.
For information on the optional TxView remote approval workflow, see “Requesting Remote
Approval of Patient Alignment (TxView Option)” on page 8-55.
For information on the optional Physician Setup Approval of Patient Alignment workflow, see
“Physician Setup Approval of Patient Alignment” on page 8-54.
Treatment Delivery
This section covers the following topics:
• “Starting Synchrony Respiratory Tracking Treatment” on page 12-81
• “Visually Verifying the Target” on page 12-84
• “Displacement Corrections During Treatment” on page 12-84
• “Monitoring Synchrony Respiratory Tracking” on page 12-85
• “Interrupting Treatment” on page 12-85
• “Error Handling” on page 12-85
• “Patient Realignment” on page 12-86
For more information on treatment delivery, see “Chapter 8: Treatment Delivery”.
3. If there is more than one Delivery Warning Message, review, resolve when
applicable, and confirm all messages as necessary.
When all messages have been confirmed and the Collimation System Status check has
been performed, you can proceed to the Delivery phase.
4. Click the Start button in the Delivery phase to begin treatment delivery, as described
in “Chapter 8: Treatment Delivery”.
WARNING: While treatment is in progress, watch the movement of the treatment robot on
the Closed Circuit TV (CCTV) system for monitoring the patient.
WARNING: During treatment using the Synchrony Respiratory Modeling System, Live X-
ray images are used to verify that the Respiratory Model is still valid for the patient
position. Image frequently enough during treatment to ensure that the Respiratory Model
is still valid.
During treatment delivery, a new model point is added to the Respiratory Model dataset
with each image acquisition, as described in “Overview of Respiratory Modeling” on
page 12-60. When the Respiratory Model contains the maximum of 15 model points and
the oldest model point is discarded before a new model point is acquired, it is possible for
the Model State to change from Optimal to Sub-optimal. With Phase Selection, it is
also possible for the model to change from Optimal to Sub-optimal if the Synchrony
Respiratory Tracking System misses the targeted phase. During treatment delivery
automation, if the Image Burst option is available on your CyberKnife System and
enabled, the Synchrony Respiratory Tracking System will take a burst of 3 images every
60 seconds, which is the default value of the Image Burst Interval parameter. Image Burst
can be disabled in the Readiness phase. The Image Burst Interval can be changed
using the Image Burst Interval (sec) slider bar on the left screen of the Delivery
phase (see “Image Burst Interval Slider Bar (Option)” on page 12-45).
If the Respiratory Model becomes Invalid during treatment delivery, an E-Stop occurs.
You must then rebuild the Respiratory Model to obtain at least a Sub-optimal model state
before you can resume treatment. It is possible for the model to become Invalid if the
model type changes from a Linear or a Curved Poly model to a Dual Poly model, and an
Interlock will occur. In order to proceed with treatment, you must acquire more images for
additional model data points.
If you need to add images to the Respiratory Model during treatment delivery, click the
Pause button to produce a Soft Stop and interrupt treatment at the next convenient
opportunity, or click the Interlock button to trigger an interlock and stop treatment
delivery immediately. For more information on interrupting treatment delivery, see
“Chapter 8: Treatment Delivery”.
WARNING: Visually verify that the target (fiducials or tumor) is correctly identified
throughout treatment. Make sure that anatomy in the Live X-ray and DRR images matches.
For more information on visually identifying features to be tracked, see “Chapter 8:
Treatment Delivery”.
It is possible that the tracking algorithm will incorrectly identify the target without returning an error.
In this case, Offset values and the average Respiratory Correlation Error value will still be
displayed on the Offsets window on the treatment delivery computer, However, the markers in
the Live X-ray images will not be positioned over the actual target:
• In Synchrony Fiducial Tracking with Respiratory Modeling mode with the Display
Fiducial Markers checkbox selected, the green circle markers will not be positioned
over the actual fiducials.
• In Synchrony Lung Tracking with Respiratory Modeling mode with the Display Marker
checkbox selected, the tumor outline will not be positioned over the actual tumor.
For more information on adjusting parameters used by the tracking algorithms to correctly identify
the target, see the following:
• “Chapter 8: Treatment Delivery”.
• “Chapter 9: Synchrony Fiducial Tracking”.
• “Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)”.
When an acceptable Respiratory Model has been established, the Synchrony Respiratory
Tracking System assists the Treatment Delivery System to correlate the Live X-ray images with
the DRR images. The Respiratory Model is used to predict the target position based on respiratory
motion and adjust the location of the centroid of the Region of Interest (ROI) around the target
accordingly.
For more information on treatment robot displacement corrections, see “Chapter 8: Treatment
Delivery”.
Interrupting Treatment
If treatment is interrupted due to a Soft Stop or E-Stop, or if you click the Interlock button, and
the Respiratory correlation model is still valid, you can click the Start/Resume button to resume
treatment.
For more information on interrupting treatment using an E-Stop button, the Pause button, or the
Interlock button, see “Chapter 8: Treatment Delivery”.
Error Handling
If an error occurs during treatment delivery, an E-Stop occurs and the CyberKnife System is
paused. If the error is recoverable, you can correct the error, reset the E-Stop, and resume
treatment. If the error is unrecoverable, you cannot reset the E-Stop and must exit the treatment.
Patient Realignment
If treatment is interrupted due to a Soft Stop or E-Stop, or if you click the Pause button, you must
confirm alignment of the patient before treatment can resume. For information on patient
realignment after treatment has been paused or an error occurs, see “Chapter 8: Treatment
Delivery”.
Introduction
This chapter describes how to use the Synchrony Tracking with InTempo Imaging with the
CyberKnife Treatment Delivery System. The Synchrony Tracking with InTempo Imaging is a time-
based motion tracking technology used to compensate for intrafraction motion of the target. It can
be used with Synchrony Skull Tracking, Synchrony Spine Tracking Supine, or Synchrony Fiducial
Tracking. The most common tracking mode used with InTempo is Synchrony Fiducial Tracking
mode. This is the mode referenced throughout this chapter unless otherwise indicated.
This chapter covers the following topics:
• “Overview of the Synchrony Fiducial Tracking with InTempo Imaging” on page 13-2
• “Workflow for Treatments with the InTempo System” on page 13-4
• “InTempo in the Overview Phase” on page 13-4
• “InTempo in the Patient Alignment Phase” on page 13-4
• “InTempo Settings in the Readiness Phase” on page 13-5
• “InTempo in the Treatment Delivery Phase” on page 13-6
This chapter assumes you are familiar with “Chapter 9: Synchrony Fiducial Tracking” and “Chapter
10: Synchrony Spine Tracking (Option)”. For information on Synchrony Skull Tracking mode, see
“Chapter 8: Treatment Delivery”.
Image Age
Image Age is the elapsed time between acquisition of an X-ray image and completion of delivery
of the beam (or beams) that use that image for targeting. The Synchrony Fiducial Tracking with
InTempo Imaging allows the user to specify the maximum allowed Image Age, by adjusting the
Max Image Age parameter. The Max Image Age parameter may be set to any value between 15
and 150 seconds (between 30 and 150 seconds for the Synchrony Spine Tracking Supine mode).
It may be adjusted at any time during treatment without interrupting treatment delivery.
By setting the value of the Max Image Age, the user controls the frequency of X-ray imaging. In
this way, the Synchrony Fiducial Tracking with InTempo Imaging guarantees that no beam is
delivered based on an image that is older than the user-defined value of this parameter. The
Synchrony Fiducial Tracking with InTempo Imaging decides whether to acquire a new image
based on the duration of the upcoming beam, by determining whether delivery of the beam will fit
within the specified value of the Max Image Age parameter.
This decision does not imply that only one beam will be delivered between two image acquisitions.
The Synchrony Fiducial Tracking with InTempo Imaging will always aim to maximize the number
of beams delivered within the duration set by the Max Image Age parameter. This is accomplished
by "splitting" the beams into manageable segments, as described below, and maximizing the
number of segments delivered within the image age specified.
Beam Segments
In order to comply with the desired Max Image Age set by the user, beams generated during
treatment planning are "split" into segments if their duration exceeds the minimum allowed value
of this parameter. A beam with a long beam-on time will therefore be split into and delivered as
several segments. The length (in time) of each segment is optimized so as to always fit within the
minimum allowed value of the Max Image Age parameter (15 seconds; 30 seconds for the
Synchrony Spine Tracking Supine mode).
13-2 | Overview of the Synchrony Fiducial Tracking with InTempo Imaging 1075883-ENG A
Treatment Delivery Manual Chapter 13: Synchrony Tracking with InTempo Imaging (Option)
The beam Sequence ID number is displayed during treatment in the Delivery Progress panel
of the Delivery phase (right screen) and provides a check to ensure that the beam number
generated at the time of treatment planning is maintained. If a beam is split into segments, the
segment number that is currently being delivered is also displayed (see “Starting Treatment” on
page 13-7).
Adaptive Imaging
With the Synchrony Fiducial Tracking with InTempo Imaging, in the event that observed motion
between two consecutive images is larger than a preset, user-defined threshold, adaptive imaging
is triggered. Adaptive imaging, if enabled, will cause the imaging frequency to be maximized by
automatically setting the Max Image Age parameter to its smallest value. This adjustment occurs
without triggering a treatment interruption, thus allowing treatment delivery to continue.
1075883-ENG A Overview of the Synchrony Fiducial Tracking with InTempo Imaging | 13-3
Accuray Incorporated Treatment Delivery Manual
2 2 2 (1)
RSS = X +Y +Z
• Max rotational shift (deg) textbox: Sets the maximum allowed rotational shift of
the target. If any single rotation (roll, pitch, or yaw) exceeds this threshold, a response is
triggered, as described below.
NOTE: The InTempo prostate path set does not include image-
blocked nodes.
Starting Treatment
If the Start button of the Delivery phase (left screen) is enabled, the CyberKnife System is ready
to start treatment. Follow the treatment instructions for the type of treatment being delivered.
During treatment using the Synchrony Fiducial Tracking with InTempo Imaging, the treatment
robot moves from node to node in the treatment path. At each node where one or more beams will
be delivered, the CyberKnife System does the following:
• Live X-ray images may be acquired and correlated with the previous X-ray image. In the
Images counter at the bottom of the Treatment Delivery screen, the number of images
acquired increases by 2 for each image pair acquired.
• Adjusts the pointing of the LINAC for the next beam, and corrects for target displacement
as described in “Chapter 8: Treatment Delivery”.
• When radiation is being generated, the BEAM ON status indicator is lit on the operator
control panel and the text BEAM is ON is displayed on both the left and right screens.
When no radiation is being generated, the indicator on the operator control panel is off
and the text BEAM is OFF is displayed on both the left and right screens.
• If the next treatment beam has been divided into segments, the first segment of the next
beam is delivered. As many segments as possible that fit within the current Max Image
Age setting are delivered. If all segments fit, all segments will be delivered.
• If the predicted image age of the current image at the end of the next beam delivery
exceeds the value of the Max Image Age parameter, a new image is taken before
delivery of the next beam begins.
• The Beam Delivery and Delivery Progress panels display information on the
radiation dose delivered for the current beam and fraction (see Figure 2).
The elapsed time that treatment has been in progress and the estimated total time for
delivery of the current fraction are updated prior to each beam delivery. For more
information on the Beam Delivery and Delivery Progress panels, see “Chapter 5:
Treatment Delivery System Overview”.
• In the Images counter, the estimated total number of images (labeled Expected) may
change during treatment delivery, because the estimate depends on variable details of
beam delivery timing. At the beginning of a long treatment path, the estimated total time is
not as accurate as at the end of the path, so the estimated total number of images may
not be as accurate.
During treatment delivery, if adaptive imaging is enabled and target motion exceeds one or both of
the thresholds set in the InTempo Settings panel of the Readiness phase, the Max Image
Age (sec) panel below the leftmost Treatment Monitoring graph flashes:
• Flashing aqua: Indicates that the Max Image Age parameter has been set to its minimum
value.
• Flashing orange: Indicates that the Max Image Age parameter was already set at its
minimum value and an excursion has occurred.
Pausing Treatment
For information on pausing treatment, see “Chapter 8: Treatment Delivery”.
Interrupting Treatment
If a Soft Stop or E-Stop occurs, or if you click the Interlock button or the Pause button,
treatment is interrupted. For more information on interrupting treatment and error handling, see
“Chapter 8: Treatment Delivery”.
Introduction
The Simulation Application provides a workflow for determining which lung tracking mode(s) are
viable for a given patient prior to creating a deliverable treatment plan. This chapter gives an
overview of the Simulation Application user interface and describes the screen controls and image
displays in each simulation phase and can be used as a reference as you perform a simulation.
The chapter also describes the workflow of the Simulation Application. This chapter covers the
following topics:
• “Starting the Simulation Application” on page 14-2
• “Overview of Simulation Application Screen” on page 14-4
• “Automatic Mode User Interface” on page 14-10
• “Manual Mode User Interface” on page 14-53
• “Simulation Application Workflow” on page 14-69
This chapter assumes that the CyberKnife System is powered up and that you are logged into the
treatment delivery computer.
This chapter assumes you are familiar with “Chapter 8: Treatment Delivery”, which describes
procedures for patient alignment, X-ray image acquisition, and error handling.
This chapter also assumes the Synchrony Fiducial Tracking with Respiratory Modeling System is
enabled on your CyberKnife System. You should also be familiar with “Chapter 12: Synchrony
Respiratory Tracking System (Option)”.
This chapter assumes you are familiar with the following chapters and treatment delivery tracking
modes:
• “Chapter 10: Synchrony Spine Tracking (Option)” on page 10-1
• “Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)” on
page 11-1
• “Chapter 15: Lung Optimized Treatment (Option)” on page 15-1
For instructions on logging into the CyberKnife System, see “Chapter 4: System Startup and
Shutdown”. You can add participants to your Simulation Application session, as described in
“Chapter 5: Treatment Delivery System Overview”.
When you click the Simulation button on the CyberKnife System Menu and log in, the
Overview phase (left screen) is displayed).
The Overview phase displays a list of patients with Simulation Plans and allows you to select a
patient for the simulation session. A list of Simulation Plans for the patient is then displayed.
Simulation Plans are not listed until DRR images have been generated on the iDMS Data
Management System.
NOTE: DRR generation will not begin until the Simulation Plan is
closed on the Accuray Precision System.
As you select the patient and Simulation Plan, additional information is displayed. A notification
message may be displayed which you must acknowledge before the simulation can proceed. Text
at the bottom left of the left screen reports status as data is retrieved from the database.
The Overview phase includes the controls listed in Table 1.
Button Function
Load After you have selected a patient and Simulation Plan and acknowledged
any notification messages, begins downloading patient and simulation data
from the iDMS Data Management System to the treatment delivery
computer.
Refresh Refreshes the Patient and Plan lists with the latest data from the iDMS
Data Management System. New patients and Simulation Plans are added to
the list only after DRR images have been generated on the iDMS Data
Management System. Click this button if a Simulation Plan has been
generated but is not yet displayed in the list.
The Patient section of the Load Patient Data window displays the following information:
• Last Name: Last name of the patient.
• First Name: First name of the patient.
• Medical ID: Medical ID of the patient.
• Date of Birth: Date of birth of the patient.
• Gender: Gender of the patient.
The Plan section displays the following information:
• Name: Name of the Simulation Plan.
• Anatomy: Anatomy to be treated.
• Orientation: Patient position as specified in the Simulation Plan.
• Fractions: Selected fraction number to be delivered.
The Notifications section displays the following information:
• Message panel: Notification message that is displayed if an OIS is licensed and there
is an inconsistency between OIS information and the iDMS Data Management System.
• Acknowledge checkbox: Indicates that you have read and acknowledged the
notification message. The checkbox must be selected before treatment can proceed.
2 4
The right screen of the Simulation Application displays controls for using the Synchrony tracking
System to build and monitor a Respiratory correlation model. For more information about the
Synchrony tracking System, see “Chapter 12: Synchrony Respiratory Tracking System (Option)”.
The treatment data display shows general patient and Simulation Plan information, including the
following:
• First Name: First name and middle name of the patient.
• Last Name: Last name of the patient.
• Medical ID: Medical ID of the patient.
• Plan: Name of the Simulation Plan.
• Session Images: Number of images acquired during the current session.
• Plan Images: Total number of images for the plan.
• Estimated Imaging Dose: Displays the estimated imaging dose for the plan and for
the current session in cGy or Gy depending on the iDMS System setting.
• Time: Current date and time as set on the treatment delivery computer.
Tabs are enabled depending on which phases of the workflow are currently available. The currently
selected phase is highlighted in yellow. The selected tab displays an in-progress indicator when
automatic operations are in progress, for example (see Figure 5):
• Live X-ray image acquisition
• image correlation
• treatment couch movement using the Move Couch button
• treatment couch movement using the Auto Align button
In general, when the in-progress indicator is displayed, controls on the screen are disabled except
for the Interlock button.
You can switch between Automatic and Manual mode. If you have an incomplete simulation
dataset and switch modes, a dialog box is displayed asking you to confirm whether you want to
discard the incomplete dataset. If you click the OK button, the dataset will be discarded. If you have
complete simulation dataset and you switch modes, the dataset will not be discarded. However,
you will no longer be able to review the dataset in the acquired dataset panel in the Offset phase
or Confirmation phase of Automatic mode.
For more information on controls in Automatic mode and Manual mode, see “Automatic Mode User
Interface” on page 14-10 and “Manual Mode User Interface” on page 14-53. For information on the
workflow in these modes, see “Simulation Application Workflow” on page 14-69.
These icons are displayed on every screen in the Treatment Delivery System, including the
Simulation Application. For information, see “Chapter 5: Treatment Delivery System Overview”.
These controls are displayed on every screen in the Treatment Delivery System, including the
Simulation Application. For information, see “Chapter 5: Treatment Delivery System Overview”.
Spine Phase
This section describes the controls and displays in the Spine phase. It covers the following topics:
• “Image Panels in the Spine Phase” on page 14-11
• “Skip Spine Checkbox” on page 14-16
• “dxAB, drAB, and False Nodes Slider Bars” on page 14-16
• “Image Correlation Parameters Panel” on page 14-17
• “X-ray Parameters Panel” on page 14-19
• “Algorithm Errors Panel” on page 14-20
• “Patient Position Display” on page 14-21
• “Patient Positioning Controls” on page 14-22
For a description of the workflow in the Spine phase using these controls, see “Simulation
Application Workflow” on page 14-69.
The top row corresponds to Camera A of the X-ray imaging system. The bottom row corresponds
to Camera B. The first image in each row is the DRR image. The center image is the most recently
acquired Live X-ray image. The last image in each row is an overlay of the DRR image and the
Live X-ray image. The green human body icon in each image shows the patient orientation relative
to the X-ray source.
Each image panel includes controls which allow you to change the image display, as well as
perform other functions, described below. Some controls are hidden in a row at the bottom of each
image panel (see Figure 10). These controls are described in Table 3.
Table 3 describes the image panel controls available in the Spine phase.
View Full Screen Displays a full screen view of the image. The image is
automatically zoomed to fill the width of the screen. All
cursor modes, including couch cursor mode, are available
in full screen view.
Live X-ray Image Display Skeletal Mesh: Toggles on or off the display of
Display Modes the skeletal mesh and ROI. For more information on this
display mode, see “Chapter 10: Synchrony Spine Tracking
(Option)”.
Cursor Modes Window and Level: Allows you to adjust the contrast and
brightness of the image. For information, see “Window and
Level Controls” on page 14-14.
Zoom and Pan: Allows you to zoom in and out and pan
the image. For information, see “Zoom and Pan Controls”
on page 14-15.
View Crosshairs
Displays a pair of crosshairs in the image to aid in the alignment of DRR and Live X-ray images.
The vertical crosshair is green, and the horizontal crosshair is red.
To move the crosshairs:
• Click and drag the intersection of the crosshairs to move both crosshairs together.
• Click and drag either the vertical or horizontal crosshair to movie it independently.
Clicking anywhere else in the image produces the behavior of the currently selected cursor mode
(see Table 3).
For information on the effect of adjusting these parameters during spine alignment, see “Chapter
10: Synchrony Spine Tracking (Option)”.
The slider bars include the following displays and controls:
• The current value of each parameter (the error value determined by the imaging
algorithm) is displayed under the left end of each slider bar. The border between the red
and green portions of the slider bar also indicates the current value. The more red, the
higher the error value.
• The threshold cursor allows you to adjust the value of the threshold. Values above this
threshold trigger an imaging algorithm error. Drag the cursor to change the threshold.
You can also enter a threshold directly in the textbox under the right end of each slider
bar.
• Dragging the threshold cursor across the red/green border of the slider bar and back
automatically displays and hides the Algorithm Errors panel.
• The default value of the threshold is indicated by a dashed line on the slider bar.
To change the value of an error threshold:
You can change the value of a threshold during spine alignment in response to errors that occur
after image correlation is performed.
1. Click and drag the threshold cursor to adjust the threshold value. Alternatively, enter a
new value in the textbox under the right end of the slider bar and press <Enter>.
When you change one of the above parameters, all controls in the Spine phase except the
Interlock button and the other image correlation parameters are disabled. Image correlation
results will be displayed after a time delay. Continued changes to parameters may increase the
delay. Once the correlation results are displayed, all Spine phase controls will be re-enabled.
For more information on the effect of changing these parameters, see “Chapter 10: Synchrony
Spine Tracking (Option)”.
To return to the default ROI Width value specified in the Simulation Plan:
1. Select the From Planning checkbox. The ROI Width value from the Simulation Plan
is displayed and the textbox is disabled.
For more information on the ROI Width parameter, see “Chapter 10: Synchrony Spine Tracking
(Option)”.
Contrast Textboxes
The Contrast A and Contrast B parameters correspond to Camera A and Camera B of the
imaging system. The range of parameter values is from 1 to 10.
When the Automatic checkbox is selected, Contrast values are determined automatically by
the Target Locating System (TLS). Pixel intensities in the Live X-ray images are adjusted
automatically to match the intensities in the DRR images. This checkbox is selected by default.
You can manually specify values. The manually specified contrast values will be applied the next
time image correlation is performed.
The final set of X-ray parameters you select in the Spine phase will become part of the simulation
dataset sent to the iDMS Data Management System at the end of the simulation session. If the
results of this dataset are selected during the Review phase, these X-ray parameters will be used
as the default parameters for spine alignment during treatment delivery.
Toggle buttons allow you to switch between the Overlay images and the Algorithm Errors
panel, as described in Table 4.
The 3 values on the left are the translational offsets (in millimeters) in the Inferior(+)/Superior(-),
Left(+)/Right(-), and Anterior(+)/Posterior(-) directions, as indicated by the arrows. The 3 values on
the right are the rotational offsets (in degrees) for Left(+)/Right(-) roll, Head-up(+)/Head-down(-)
pitch, and Counterclockwise(+)/Clockwise(-) yaw, as indicated by the arrows. The arrows point in
the positive direction and do not change. An orange box around values indicates the values are
out of bounds and the patient must be repositioned.
The values in the patient position display are updated after each Live X-ray image is acquired.
Images are acquired using the Acquire button, as described in Table 5.
The controls include a diagram showing the current absolute rotational position of the treatment
couch above a row of textboxes. The 3 non-editable values give rotations (in degrees) for Left(+)/
Right(-) roll, Head-up(+)/Head-down(-) pitch, and Counterclockwise(+)/Clockwise(-) yaw, as
indicated by the arrows. The arrows point in the positive direction and do not change.
The textboxes underneath the diagram indicate desired motion of the treatment couch relative to
the current couch position. During the next automatic movement of the treatment couch using the
Move Couch button (see Table 6), the couch moves by this amount. As the treatment couch
moves, the values in the textboxes change and approach zero as the desired couch position is
achieved.
The desired motion can also be specified using couch cursor mode which automatically updates
the textboxes, as described in Table 3 on page 14-12.
Each time a successful image correlation is performed, the textboxes are automatically updated
with the position of the treatment couch at the time of image acquisition plus the calculated patient
offsets.
Each rotation axis can be independently "locked" during automatic movement to prevent it from
changing. To lock a rotation axis, click the icon next to a rotation axis. To unlock a rotation, click
the icon. The locks also apply to the Auto Align button (see Table 6).
Couch Move Moves the treatment couch to the target position indicated
by the values entered in the textboxes in the patient
positioning controls.
Lung Phase
This section describes the controls and displays in the Lung phase. It covers the following topics:
• “Image Panels in the Lung Phase” on page 14-24
• “X-ray Parameters Panel” on page 14-26
• “Treatment Couch Position Display” on page 14-27
• “Patient Positioning Controls” on page 14-30
For a description of the workflow in the Lung phase using these controls, see “Simulation
Application Workflow” on page 14-69.
Similar to the Spine phase, the top row corresponds to Camera A of the X-ray imaging system.
The bottom row corresponds to Camera B. The first image in each row is the DRR image. The
center image is the most recently acquired Live X-ray image. The last image in each row is an
overlay of the DRR image and the Live X-ray image. The green human body icon in each image
shows the patient orientation relative to the X-ray source.
Each image panel includes controls which allow you to change the image display, as well as
perform other functions, described below. Some controls are hidden in a row at the bottom of each
image panel (see Figure 21).
Controls are similar to those in the Spine phase (see Table 3 on page 14-12) except for the DRR
display modes in the top left corner, described in Table 7.
• At the lung position: The treatment position specified in the Simulation Plan. The position
is determined by combining the spine position with the alignment-to-treatment offset
provided by the Simulation Plan.
• Between the spine position and the lung position. This position is indicated when the
treatment couch is moving between or paused between these two positions.
• Near but not at the lung position.
The display also includes 3 buttons for moving the treatment couch, as described in Table 9.
Table 9 describes the buttons associated with the treatment couch position display.
Move to Lung Moves the treatment couch to the lung position (treatment
position).
Move Couch Moves the treatment couch to the target position indicated
by the values entered in the textboxes in the patient
positioning controls.
This button is enabled when the treatment couch is at or
near the lung position and when the target couch position
in the textboxes does not match the current position. For
information, see “Patient Positioning Controls” on
page 14-30.
The row of textboxes underneath the diagram allows you to enter a target position for the next
automatic movement of the treatment couch using the Move Couch button described in Table 9.
The target location can also be specified using couch cursor mode which automatically updates
the textboxes, as described in Table 3 on page 14-12.
X-ray Phase
This section describes the controls and displays in the X-ray phase. It covers the following topics:
• “Image Panels in the X-ray Phase” on page 14-31
• “X-ray Parameters Panel” on page 14-32
• “Image Acquisition Controls” on page 14-33
For a description of the workflow in the X-ray phase using these controls, see “Simulation
Application Workflow” on page 14-69.
The top row corresponds to Camera A of the X-ray imaging system. The bottom row corresponds
to Camera B. The first image in each row is the raw (unprocessed) Live X-ray image. The raw
images have not been enhanced for contrast by the tracking algorithm. The center image is also
the Live X-ray image but it has been enhanced for contrast by the tracking algorithm. The last
image in each row is the DRR image.
A histogram display under each raw image represents the distribution of pixel intensities and
provides an aid during refinement of the X-ray technique (see Figure 24). Increasing the X-ray
technique will generally push the histogram peak to the left. Decreasing the X-ray technique will
generally push the peak to the right. The ideal histogram distribution is spread out (indicating good
contrast) and shifted towards the right.
Each image panel includes controls which allow you to change the image display. Some controls
are hidden in a row at the bottom of each image panel (see Figure 25). A subset of the controls in
the Spine phase is available, described in Table 3 on page 14-12.
Offset Phase
This section describes the controls and displays in the Offset phase. It covers the following topics:
• “Image Panels in the Offset Phase” on page 14-34
• “Dataset Acquisition Controls” on page 14-38
• “Offset Specification Controls” on page 14-40
• “Camera Visibility Controls” on page 14-42
For a description of the workflow in the Offset phase using these controls, see “Simulation
Application Workflow” on page 14-69.
4B
4A
NOTE: The order in which the image pairs is displayed does not
correspond to their location in the respiratory cycle.
One image at a time can be displayed in the large offset specification panel. The currently
displayed image is outlined in yellow (see Figure 28). You can also use the Up and Down arrow
keys on your keyboard and the scroll wheel on the mouse to scroll through the images and display
them one at a time in the offset specification panel. When you hold down the arrow keys or use the
scroll wheel, the rapid view of the image series may help to identify the tumor.
An image outlined in blue can be used for an offset specification. To use the image, click on it to
display it in the offset specification panel. An image that is not outlined in blue cannot be used, for
one of the following reasons:
• 2 images in that column have already been used for the offset specification.
• The image is too near the respiratory phase of another image that has already been used
for the offset specification.
For more information on specifying offsets in the Offset phase, see “Simulation Application
Workflow” on page 14-69.
The panel also includes controls which allow you to change the image display, as well as perform
other functions, described below. Some controls are hidden in a row at the bottom of each image
panel (see Figure 30).
A subset of the controls in the Spine phase is available (described in Table 3 on page 14-12), as
well as an additional Live X-ray image display mode, described in Table 11
The offset specification panel also includes controls that allow you to specify the visually identified
tumor location. For more information, see “Offset Specification Controls” on page 14-40.
Figure 31 shows the respiratory waveform display. This display shows a waveform representing a
respiratory cycle. The waveform contains 8 positions representing the 8 phases of the respiratory
cycle. The position with the highest amplitude represents either full inspiration or full expiration, and
the position with the lowest amplitude represents the opposite. The waveform displays checkmarks
for phases in the respiratory cycle where images have been acquired.
Acquiring Images
When the Acquire Dataset button is clicked, the Synchrony tracking System is used to time
image acquisition so that the images are spread throughout the respiratory cycle. At least 8 image
pairs are acquired before a dataset is considered complete. Images are acquired repetitively until
7 of the 8 respiratory phases are covered, or until a maximum of 12 image pairs are acquired.
Respiratory phases are targeted as follows (see Figure 32): The first two image acquisitions target
the peak and valley of the respiratory waveform (full inspiration and expiration). Next, the midpoints
of inspiration and expiration are targeted. Lastly, the center positions of the respiratory waveform
are targeted.
Taken First
Taken Second
Taken Last
As image pairs are acquired, they are displayed in the acquired dataset panel (see “Acquired
Dataset Panel” on page 14-35.
If dataset acquisition is interrupted by an E-Stop, then after the E-Stop is reset, you can click the
Acquire button to continue adding images to the dataset, as long as less than 7 minutes has
elapsed since the first image was acquired.
Discarding Datasets
The following actions will discard acquired datasets in the Offset phase, after a confirmation
dialog box is displayed asking you to confirm that you want to discard the dataset:
• Switching back to a previous phase and changing the X-ray technique.
• Switching back to a previous phase and changing the treatment couch position.
• Allowing too much time to pass while the system is paused (for example, due to errors),
before a dataset is complete.
Datasets that are discarded are not available for later review. Datasets must be completed in the
Confirmation phase in order to be available for review. For information, see “Confirmation
Phase” on page 14-43.
Table 14 describes the buttons used to confirm or remove a user-provided offset in the offset
specification panel.
Offsets can only be specified for images that have at least one respiratory phase position vertically
between them in the respiratory waveform, as shown in Figure 34.
Confirmation Phase
This section describes the controls and displays in the Confirmation phase. It covers the
following topics:
• “Image Panels in the Confirmation Phase” on page 14-43
• “Controls for Confirming Correlation Results” on page 14-47
• “Uncertainty and dxAB Bar Graphs” on page 14-48
For a description of the workflow in the Confirmation phase using these controls, see
“Simulation Application Workflow” on page 14-69.
2A 2B
Icon Description
Image pairs are displayed as their correlations are completed. The lung tracking algorithm
automatically selects the 2 image pairs to be correlated first and they may not be the first 2 image
pairs. As a result, the first 2 image pairs to be displayed may appear in lower positions in the
acquired dataset panel.
If the tumor is specified as not visible in both the Camera A and Camera B views in the Offset
phase (that is, the Tumor A Visible and Tumor B Visible buttons are deselected), all
images in the acquired dataset panel have the icon superimposed on them.
If the tumor is specified as visible in one of the camera views in the Offset phase, images in that
view in the acquired dataset panel have the icon superimposed on them in the Confirmation
phase. As the image correlations are performed, it may take time for the images to populate the
panel. The Confirmation tab displays an in-progress indicator as the correlations are
performed.
One image pair at a time can be displayed in the image A and image B panels (see “Image A and
Image B Panels” on page 14-45). The currently displayed images are outlined in yellow. You can
also use the Up and Down arrow keys on your keyboard and the scroll wheel on the mouse to scroll
through the image pairs and display them one at a time in the image panels.
Each panel also includes controls which allow you to change the image display, as well as perform
other functions, described below. Some controls are hidden in a row at the bottom of each image
panel (see Figure 37).
A subset of the controls in the Spine phase is available (described in Table 3 on page 14-12), as
well as an additional Live X-ray image display mode, described in Table 17
Table 18 describes the buttons that switch between Image A and Image B viewing options.
Image A|B Displays Image B in the left panel and Image A in the right
Panel panel.
This is the default view when both image views are
confirmed to be trackable in the Offset phase (that is,
when the tumor is confirmed to be visible in both Camera A
and Camera B of the X-ray imaging system).
When a DRR image is displayed in the left image panel, controls for confirming correlation results
are not displayed beneath it (see “Controls for Confirming Correlation Results” on page 14-47).
You can switch between the same DRR display modes as in Offset mode (see Table 12 on
page 14-37).
One purpose of the lung tracking algorithm in the Confirmation phase is to determine
appropriate settings (enabled or disabled) for the Preferred Projection algorithm parameter. The
user has no input in this determination. If this parameter is enabled, it is enabled for the entire
dataset.
For more information on these parameters, see “Review Phase” on page 14-48 and “Chapter 11:
Synchrony Lung Tracking with Respiratory Modeling (Options)”.
Correlation A Indicates that you cannot see the tumor to confirm whether
Uncertain and the correlation was successful or unsuccessful.
Correlation B
Uncertain
Review Phase
Figure 39 shows the main panel of the Review phase in Automatic mode. For a description of the
workflow in the Review phase, see “Simulation Application Workflow” on page 14-69
The top portion of the panel displays a table listing results for each simulation dataset in the current
simulation session as well as datasets from previous login sessions. Each dataset corresponds to
one row of the table. If at least one dataset from a previous session exists, you can go straight to
the Review phase after logging into the Simulation Application. Datasets acquired in Automatic
mode are added to the table when the user has provided feedback for every image correlation in
the Confirmation phase (see “Confirmation Phase” on page 14-43). Columns in the table are
described in Table 20.
NOTE: The term "2-View" used in the main panel of the Review
phase is equivalent to Synchrony Lung Tracking with Respiratory
Modeling (Lung with Respiratory).
A For Image A, the percentage (%) and number (out of the total
number of images acquired) of false positives identified by the
user.
For example, for a dataset of type Automatic, a value in this
column of 12% (1/8) indicates that 12% of the images (1 out of 8
images) acquired by Camera A in the Offset phase were
identified as false positives by the user in the Confirmation
phase.
B For Image B, the percentage (%) and number (out of the total
number of images acquired) of false positives identified by the
user.
1-View A For Image A, the percentage (%) and number (out of the total
number of images acquired) of successful correlations
confirmed by the user.
1-View B For Image B, the percentage (%) and number (out of the total
number of images acquired) of successful correlations
confirmed by the user.
2-View For both Image A and Image B pairs, the percentage (%) and
Lung with number (out of the total number of image pairs acquired) of
Respiratory successful correlations for both images in the pair confirmed by
the user.
1-View A For Image A, the percentage (%) and number (out of the total
number of images acquired) of user-confirmed successful
correlations plus user-indicated uncertain correlations.
1-View B For Image B, the percentage (%) and number (out of the total
number of images acquired) of user-confirmed successful
correlations plus user-indicated uncertain correlations.
2-View Similar to the table column under the icon above in the
Lung with table, except that user-indicated uncertain correlations are
Respiratory counted as if they are successful correlations.
That is, for both Image A and Image B pairs, this column shows
the percentage (%) and number (out of the total number of
image pairs acquired) of user-confirmed successful correlations
plus the number of user-indicated uncertain correlations for
both images in the pair.
For information on the criteria used by the Simulation Application for determining a certain tracking
mode result, see “Simulation Application Workflow” on page 14-69.
In the Result column, the selected dataset icon is displayed next to the dataset that is
currently selected by the user to be used as a candidate for treatment planning.
Clicking a row in the table highlights that row in yellow and displays information about the selected
dataset in the lower portion of the panel.
Spine Phase
The Spine phase in Manual mode is the same as in Automatic mode. For information, see “Spine
Phase” on page 14-10.
Lung Phase
The Lung phase in Manual mode is the same as in Automatic mode. For information, see “Lung
Phase” on page 14-24.
X-ray Phase
The X-ray phase in Manual mode is the same as in Automatic mode. For information, see “X-ray
Phase” on page 14-31.
Respiratory Phase
This section describes the controls and displays in the Respiratory phase. It covers the
following topics:
• “Image Panels in the Respiratory Phase” on page 14-54
• “Offset Specification Controls” on page 14-56
• “dxAB and Uncertainty Slider Bars” on page 14-57
• “Image Correlation Parameters Panel” on page 14-58
• “X-ray Parameters Panel” on page 14-60
• “Algorithm Errors Panel” on page 14-61
• “Patient Position Display” on page 14-61
• “Patient Positioning Controls” on page 14-62
• “Respiratory Modeling Controls” on page 14-63
For a description of the workflow in the Respiratory phase using these controls, see “Simulation
Application Workflow” on page 14-69.
The top row corresponds to Camera A of the X-ray imaging system. The bottom row corresponds
to Camera B. The first image in each row is the DRR image. The second image is the most recently
acquired Live X-ray image. The last image in each row is an overlay of the DRR image and the
Live X-ray image.
Each image panel includes controls which allow you to change the image display, as well as
perform other functions, described below. Some controls are hidden in a row at the bottom of each
image panel (see Figure 41). These controls are described in Table 3.
Controls are similar to those in the Spine phase (see Table 3 on page 14-12) except for those
described in Table 22.
The upper right corner of each Live X-ray image panel displays the respiratory phase during which
it was acquired (see Figure 42).
For information on the effect of adjusting these parameters, see “Chapter 11: Synchrony Lung
Tracking with Respiratory Modeling (Options)”.
The slider bars include the same displays and controls described in “dxAB, drAB, and False Nodes
Slider Bars” on page 14-16.
To change the value of an error threshold:
You can change the value of a threshold in response to errors that occur after image correlation is
performed.
Click and drag the threshold cursor to adjust the threshold value. Alternatively, enter a new value
in the textbox under the right end of the slider bar and press <Enter>.
When you change one of the above parameters, all controls in the Respiratory phase except
the Interlock button and the other image correlation parameters are disabled. Image correlation
results will be displayed after a time delay. Continued changes to parameters may increase the
delay. Once the correlation results are displayed, all Respiratory phase controls will be re-
enabled.
For more information on the effect of changing these parameters, see “Chapter 11: Synchrony
Lung Tracking with Respiratory Modeling (Options)”.
Contrast Textboxes
The Contrast A and Contrast B parameters correspond to Camera A and Camera B of the
imaging system. The range of parameter values is from 1 to 10. Values apply to image correlations
in the Respiratory phase.
When the Automatic checkbox is selected, Contrast values are determined automatically by
the Target Locating System (TLS).
These parameter adjustments are similar to those for the Spine phase. For more information, see
“Contrast Textboxes” on page 14-19.
Preferred Projection ON
The Preferred Projection ON checkbox allows you to enable or disable the Preferred
Projection parameter. When enabled, the preferred projection method is enabled for locating the
tumor in the Respiratory phase. For more information on the preferred projection method, see
“Chapter 11: Synchrony Lung Tracking with Respiratory Modeling (Options)”.
By default, the final X-ray technique used in the Lung phase is used in the Respiratory phase.
Changes made to the X-ray technique in the Respiratory phase are applied to the all other
phases except the Spine phase.
The 3 values on the left are the tracking target position relative to the machine center (in
millimeters) in the Inferior(+)/Superior(-), Left(+)/Right(-), and Anterior(+)/Posterior(-) directions, as
indicated by the arrows. The arrows point in the positive direction and do not change. An orange
box around values indicates the values are out of bounds and must be adjusted.
The value on the far right is the Respiratory correlation error (in millimeters).
The values in the patient position display are updated after each Live X-ray image is acquired.
Images are acquired using the Acquire button, as described in “Respiratory Modeling Controls”
on page 14-63.
The row of textboxes underneath the diagram allows you to enter a target position for the next
automatic movement of the treatment couch using the Move Couch button described in
Table 23. The target location can also be specified using couch cursor mode which automatically
updates the textboxes, as described in Table 3 on page 14-12.
Each time a successful image correlation is performed, the textboxes are automatically updated
with the position of the treatment couch at the time of image acquisition plus the calculated location
of the tracking target.
Move Couch Moves the treatment couch to the target position indicated
by the values entered in the textboxes in the patient
positioning controls.
Review Phase
Figure 48 shows the main panel of the Review phase in Manual mode. For a description of the
workflow in the Review phase, see “Simulation Application Workflow” on page 14-69.
The panel is similar to the Review phase in Automatic mode (see “Review Phase” on page 14-48)
with the following changes:
• In Manual mode, the columns in the table for false positives identified by the user and for
user-confirmed successful correlations plus user-indicated uncertain correlations do not
apply. No values are listed in these columns for datasets acquired in Manual mode.
• In Manual mode, the values listed in the table under the icon for the percentage/
number of successful correlations are not relevant.
• In Manual mode, the Simulation Application does not use an algorithm to determine a
simulation result. Instead, the user selects a tracking mode from the Result dropdown
list at the bottom of the panel to apply to the highlighted dataset (see Figure 49).
Figure 49 Result dropdown menu in the Review phase for Manual mode
Acquire
Couch X-ray Confirm
Spine Dataset and Review
Positioning Technique Algorithm
Alignment Specify Dataset
for Lung Refinement Results
Offsets
The workflow in Automatic mode includes the following general steps, described in the sections
below:
1. Set up the patient so that the Field of View is centered on the respiratory cycle and the X-
ray technique is optimized.
2. Acquire a simulation dataset which contains Live X-ray images and image algorithm
correlation results that sample the whole respiratory cycle. Image correlation is performed
to determine which image views are appropriate for tracking.
3. Review the simulation dataset. Repeat Step 2 until satisfied.
5. Click the Load button to begin downloading patient and Simulation Plan data from the
iDMS Data Management System to the treatment delivery computer. The Spine tab of
the Simulation Application will be displayed.
Or, click the Exit button to return to the CyberKnife System Menu.
6. If there is a pre-existing simulation dataset for the patient from a previous simulation
session, you can go directly to the Review phase to view results for this dataset.
By clicking the Skip Spine checkbox, you can switch directly to the Lung phase. Otherwise, the
Lung tab is disabled until spine alignment is performed.
Once spine alignment is complete, click the Lung tab to switch to the Lung phase.
• If spine alignment was not performed, the spine position is the position of the treatment
couch at which the last image was taken in the Spine phase.
• The lung position is the spine position plus the difference between the alignment center
and the ITTV center from treatment planning.
In the Lung phase you first click the Move to Lung button and then wait until the treatment
couch has completed movement to the lung position. You then click the Acquire button to take a
pair of Live X-ray images using the default X-ray parameters. In general, you should defer
changing X-ray parameters until the X-ray phase.
After acquiring a pair of Live X-ray images and determining that there are no unexpected features
or obstructions in the images, click the X-ray tab to switch to the X-ray phase.
After reviewing simulation results, you select one dataset as your candidate using the controls
described in “Review the Simulation Dataset” on page 14-73 and then exit the Simulation
Application.
In Automatic mode, the Simulation Application result is chosen for a particular dataset based on
the following criteria:
• If the percentage of successful 2-View correlations is 75% or greater, then the result for
that dataset will be Lung with Respiratory.
• If the percentage of successful 2-View correlations is less than 75% and you have not
skipped spine alignment, then the result will be Spine Supine, Lung 1-View with
Respiratory, or non-trackable.
If the percentage of successful 1-View A correlations is 75% or greater and spine
alignment was not skipped, then that is the result, Lung 1-View A with Respiratory.
If the percentage of successful 1-View B correlations is 75% or greater and spine
alignment was not skipped, then that is the result, Lung 1-View B with Respiratory.
If the percentage of successful 1-View and 2-View correlations are less than 75% and
you have not skipped spine alignment, then the result will be Spine Supine.
• Otherwise, the result will be Not Trackable.
If in the Confirmation phase, the user specified that the tumor was not visible in
both Camera A and Camera B, the result will be Spine Supine tracking.
This motion is
tracked.
ITTV
While the ITV described above is based on the Clinical Target Volume (CTV) and is delineated by
the Physician, the Internal Target Tracking Volume (ITTV) is based on the Target Tracking Volume
(TTV) and is generally delineated by the Physicist or Dosimetrist during creation of a Simulation
plan.
The ITTV represents the expected respiratory excursion of the TTV. The ITTV is defined from the
delineation of the TTV on the two respiratory phases represented by the CT scans used to create
a Simulation plan. The shape and position of the ITTV is independent of the tracking mode
subsequently selected when creating a treatment plan.
During treatment of a case where Lung 1-View Tracking with Respiratory Modeling is selected and
the treatment plan was based on a Simulation plan, the projected overlay of the ITTV is displayed
in the untracked Live X-ray image.
PTV
It is important to recognize that there are additional geometrical uncertainties involved in treatment
delivery which should be combined into a setup margin that is applied to the ITV to form a Planning
Target Volume (PTV) [Ref. 1]. These should include an estimate of the Respiratory modeling
uncertainty in the Motion Tracking Plane with Lung 1-View with Respiratory, together with an
estimate of the uncertainty in the position of the ITV relative to the spine region covered by the
Synchrony Spine Tracking Region of Interest (ROI) in the untracked out-of-plane direction for Lung
1-View Tracking or in all directions for Lung with Respiratory Tracking.
The Lung 1-View with Respiratory algorithm searches for and locates the target in either the A or
B image (corresponding to Camera A or Camera B of the X-ray imaging system) during treatment.
The image that is used for tracking (A or B) is specified during treatment planning. Target
localization is performed by image registration between the DRR image and the Live X-ray image
for the specified image. Specifically, the image intensity pattern of the tumor region in the DRR
image is matched to the most similar region in the Live X-ray image. A matching window for the
tumor is defined based on the 2D tumor outline. The matching window is the smallest rectangle
that encompasses the entire tumor. Since typical tumor shapes are not rectangular, the matching
window will also include some of the surrounding region.
Registration is performed by moving the matching window throughout a user-defined search area
(see “Tracking Range X (mm) and Tracking Range Y (mm)” on page 15-30). The region of the
search area with maximum similarity to the corresponding DRR image is considered to be the
target. To assess the quality of the registration, the Synchrony Lung Tracking with Respiratory
Modeling algorithm uses quality metrics (see “Quality Metrics” on page 15-7).
The 1-View with Respiratory algorithm computes three translational components of the target
patient displacement. The rotational components are provided by the roll displacements estimated
by the Synchrony Spine Tracking Supine during patient alignment and are assumed to be constant
throughout the treatment. Successful initial patient alignment using the Synchrony Spine Tracking
Supine is therefore necessary. The treatment robot continually adjusts the position of the LINAC
to correct for translational offsets of the target and applies the constant rotational correction.
1-View with Respiratory is used in conjunction with the Synchrony tracking System to track and
compensate for target respiratory motion in the tracking plane. Treatment delivery uses 1-View
with Respiratory for target localization and the Synchrony tracking System for compensating for
respiratory motion.
Quality Metrics
The 1-View with Respiratory algorithm operates by locating an intensity pattern in the Live X-ray
image that is similar to that in the DRR image. It is possible that more than one such pattern exists
in the search area. The algorithm takes into account several factors to assess the registration
quality:
• The maximum of the similarity measure between the Live X-ray and DRR images.
• The number of other local maxima found within the Tracking Range (for information on
the Tracking Range parameters, see “Tracking Range X (mm) and Tracking Range Y
(mm)” on page 15-30).
• The tumor size and the tumor location relative to the prediction made by the Synchrony
tracking System. For more information on the Synchrony tracking System, see “Chapter
12: Synchrony Respiratory Tracking System (Option)”.
The threshold for detection uncertainty is user-defined (see “Uncertainty (%)” on page 15-29). If
the detection uncertainty is higher than the threshold, treatment delivery is interrupted by a Soft
Stop.
Two treatment couch positions are used during treatment delivery with 1-View with Respiratory:
• Spine position: The alignment center defined during treatment planning in the Align step
of the Accuray Precision System. This position is used for alignment by the Synchrony
Spine Tracking Supine.
• Lung position: The treatment center automatically determined during treatment planning
during tumor segmentation. This is the spine position plus the alignment-to-treatment
offset from planning. This position is used for tracking by the Synchrony Lung Tracking
with Respiratory Modeling.
The alignment center is typically different from the treatment center. The alignment center is used
during an initial spine alignment procedure using the Synchrony Spine Tracking Supine. The spine
alignment is performed in the Alignment > Align phase. This procedure determines the
rotational alignment corrections that will be used during treatment delivery with Lung 1-View
Tracking with Respiratory Modeling.
After spine alignment is completed, you transition from the Synchrony Spine Tracking Supine to 1-
View with Respiratory. This transition occurs when you move to the Alignment > Lung phase.
You then move the treatment couch to the treatment center.
Treatment delivery using 1-View with Respiratory includes the following general steps:
1. Load patient and treatment plan data in the Overview phase, as described in “Chapter
8: Treatment Delivery”.
2. Find the optimal X-ray technique for spine alignment using the Synchrony Spine Tracking
Supine in the first Alignment > Technique phase, as described in “Chapter 10:
Synchrony Spine Tracking (Option)”.
3. Position the patient on the treatment couch and move the patient to the alignment center
for initial spine alignment using the Synchrony Spine Tracking Supine in the Alignment
> Align phase, as described in “Chapter 10: Synchrony Spine Tracking (Option)”.
1075883-ENG A Workflow for Lung 1-View Tracking with Respiratory Modeling | 15-9
Accuray Incorporated Treatment Delivery Manual
4. Use Synchrony Spine Tracking Supine to align the patient translationally and rotationally
at the spine position, as described in “Chapter 10: Synchrony Spine Tracking (Option)”.
After this step, the treatment couch is at the Synchrony Spine Tracking alignment center.
5. Switch from the Synchrony Spine Tracking alignment center (the spine position) to the 1-
View with Respiratory treatment center (the lung position) in the Alignment > Lung
phase.
This step involves treatment couch movement. The difference between the alignment
center and the treatment center includes the alignment-to-treatment offset defined from
treatment planning plus any additional offset the user applied during the current or
previous treatment fractions. For more information, see “Transition to Lung 1-View
Tracking with Respiratory Modeling” on page 15-15.
6. Find the optimal X-ray technique for treatment delivery using Lung 1-View Tracking with
Respiratory Modeling in the second Alignment > Technique phase.
7. Use 1-View with Respiratory to translationally align the patient sufficiently for tracking, in
the Alignment > Respiratory phase.
If the Synchrony tracking System is used, establish a Respiratory correlation model for
patient respiratory motion.
15-10 | Workflow for Lung 1-View Tracking with Respiratory Modeling 1075883-ENG A
Treatment Delivery Manual Chapter 15: Lung Optimized Treatment (Option)
After aligning the patient properly within acceptable bounds using the Synchrony Spine Tracking
Supine, you are ready to move to the Alignment > Lung phase.
WARNING: Do not reposition the patient or adjust treatment couch rotation after
switching to Lung 1-View with Respiratory. If it is necessary to do this, return to the
Alignment > Align phase and repeat the Synchrony Spine Tracking alignment
procedure, as described in “Chapter 10: Synchrony Spine Tracking (Option)”. Otherwise,
incorrect patient displacement corrections will be used during treatment.
Table 1 shows the treatment couch position display in the Alignment > Lung phase (left
screen, below Image A). The display includes a position indicator which shows the treatment couch
at one of 4 positions:
• At the spine position: The spine position (alignment center) from treatment planning plus
the final offsets from the Alignment > Align phase.
• At the lung position: The treatment position from planning. This is the spine position plus
the alignment-to-treatment offset from planning.
• Between the spine and lung position: The treatment couch is paused between or moving
between these two positions.
The display also includes 2 buttons for moving the treatment couch, as described in Table 2.
Table 2 describes the buttons associated with the treatment couch position display.
Move Couch Moves the treatment couch to the lung position (treatment
to Lung position).
2. Click the Acquire button to acquire a Live X-ray image. For more information on
acquiring Live X-ray images, see “Chapter 8: Treatment Delivery”.
3. Analyze imaging results to verify that rigid structures are aligned in the Live X-ray and
DRR images. Verify that there are no unexpected features or obstructions in the images.
For more information, see “Analyzing Imaging Results” on page 15-21.
If treatment couch motion is interrupted during the transition between Synchrony Spine Tracking
and Lung 1-View with Respiratory, for example, if an E-Stop button is pressed, you can use the
treatment couch position display to move to the spine or lung position, as described below.
To move to the spine or lung position after treatment couch motion is interrupted:
• In the Alignment > Lung phase, click the Move to Spine button to move to the
spine position (alignment center). Or, click the Move to Lung position to move to the
lung position (treatment center).
If it is necessary, you can add an offset to the alignment-to-treatment offset from planning, as
described below.
To add an offset to the lung position from planning:
1. Enter target translation values in the Couch Position (mm) textboxes.
Arrows above the textboxes point in the positive direction. Values indicate target
treatment couch movement relative to its current position.
2. Click the Move Couch button to move the treatment couch. (The couch rotations from
spine alignment are locked in the Alignment > Lung phase.)
The additional offset is automatically saved and will be applied during treatment delivery
for the current fraction and future fractions. (The Apply additional relative offset
checkbox is selected by default.)
3. To remove the additional offset that you generated in Step 2, or to remove the additional
offset that was saved and applied for previous fractions, deselect the Apply
additional relative offset checkbox. Then click the Move to Lung button.
This step returns the couch to the lung position (the treatment position from planning).
The lung position will then be used for the current fraction and future fractions.
After moving the treatment couch to the lung position, taking an image, and verifying that rigid
structures in the DRR image and Live X-ray image that is used for tracking are aligned, you are
ready to move to the second Alignment > Technique phase.
WARNING: The Live X-ray images should have the same image quality as the DRR
images. Otherwise, the tracking algorithm may not be able to identify and track features
correctly.
In the Alignment > Technique phase (left screen), a histogram display below each Raw
Image represents the distribution of pixel intensities in the entire image. Table 3 on page 15-23
describes the image panel controls to aid image analysis in this phase, including DRR display
modes.
The goal is to determine the optimal X-ray technique for XRS A and XRS B independently, by
evaluating image quality and histogram displays of images on the left screen and in the Image
History table.
Increasing the X-ray technique generally pushes the histogram peak to the left. Decreasing the X-
ray technique generally pushes the peak to the right. The ideal distribution is spread out (indicating
good contrast) and not shifted towards the extreme left or right.
If you have selected a satisfactory X-ray technique, you can proceed to the
Alignment > Respiratory phase. Otherwise, you can change the X-ray technique as needed
until you are satisfied with the image quality, as described below. For more information on the
Alignment > Technique phase, see “Chapter 8: Treatment Delivery”.
The Alignment > Respiratory phase has the following steps, described in the sections
below:
• “Analyzing Imaging Results” on page 15-21: You start by evaluating image correlation
results to determine if you need to move the treatment couch. You also review the Live X-
ray images to verify that the tumor has been correctly located and that anatomy in the
Live X-ray and DRR images matches.
• “Respiratory Alignment and Building a Respiratory Model” on page 15-32: You build a
Respiratory correlation model for respiratory motion.
If the 1-View with Respiratory algorithm cannot locate the tumor, the displayed tumor outline in the
Live X-ray image will not be aligned with the actual tumor. An out-of-bounds error may also occur,
indicated by an orange box around the row in the Respiratory Model (mm) table that is not
within required tolerances. Also tracking algorithm threshold violations may occur, indicated by an
orange box around the row in the table and around the slider bar for a threshold parameter.
In this situation, you can attempt one or more of the following:
• Make the following adjustments as appropriate:
Use Offset Mode to drag the tumor outline in the Live X-ray image used for tracking to
determine translational offsets in the tracking plane. For more information, see “Using
Offset Mode” on page 15-27.
If you are unable to correlate successfully using Offset Mode, you can increase the
Tracking Range X and Tracking Range Y. For more information, see “Tracking
Range X (mm) and Tracking Range Y (mm)” on page 15-30.
• Couch Mode can be used to move the Live X-ray image used for tracking relative to the
corresponding DRR image to determine translational offsets in the tracking plane. For
more information, see Table 3 on page 15-23 and “Chapter 8: Treatment Delivery”.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
• You can also consider increasing the value of the Uncertainty threshold parameter if you
are confident that the tumor found using a higher value is correctly identified. For more
information, see “Uncertainty (%)” on page 15-29.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
1-View with Respiratory algorithm to incorrectly identify the tumor without generating an
error. You must therefore visually verify that the tumor has been correctly identified in the
image used for tracking throughout the treatment.
For more information on image analysis tools to verify that the tumor is correctly located by the
tracking algorithm, see Table 3 on page 15-23. For information on calculated algorithm
parameters, parameter thresholds, and algorithm input parameters, see “Adjusting Algorithm
Parameters” on page 15-30.
Control Description
Movie Mode: Use Movie Mode to cycle through all Live X-ray images in the
Respiratory Model one at a time in respiratory sequence. This makes tumor
motion more visible and helps you verify that tumor motion is being tracked.
• Click the Play icon to start Movie Mode.
• Click the Pause icon to pause Movie Mode.
• Use the zoom slider bar to increase or decrease playback speed.
• Or, click the Plus icon or the Minus icon to change playback speed.
• Click the Reset icon to reset the playback speed to its default.
Control Description
Couch: Adjust the position of the Live X-ray image relative to the DRR image by
clicking and dragging in either the Live X-ray image or DRR image. For more
information, see “Chapter 8: Treatment Delivery”. Depending on the treatment
phase, the following icons are displayed:
• Click the Reset icon to move the Live X-ray image back to its original,
unshifted position.
• Click the Reset icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with zero values. The Live X-ray image moves
to its unshifted position, to show existing alignment.
• Click the Offsets icon to fill the Couch Position (mm) and Couch
Rotation (deg) textboxes with the calculated offset values. The Live X-ray
image shifts by the translational offsets, to show the alignment that would
result if the couch is moved by these amounts.
You can toggle back and forth between these icons. The icons only shift the Live
X-ray image relative to the DRR image. They do not move the couch.
Window/Level: Adjust the window (contrast) and level (brightness) of the image.
A histogram representing the distribution of pixel intensities is displayed. Controls
help you visually identify anatomical or other features in the Live X-ray images.
They affect the image display only.
• Click in the image and drag up or down to adjust the window. Or, click and
drag the left and right edges of the histogram to move them towards the
center to adjust both the window and level.
• Click in the image and drag left or right to adjust the level. Or, click and drag
the vertical line at the center of the histogram to the left or right.
• Click the Invert Contrast icon to invert image contrast. Inverting image
contrast may make it easier to visually identify anatomical and other features.
• Click the Lock icon to apply the same window and level settings to all
images. Click again to unlock settings.
• Click the Reset icon to reset window and level settings to their defaults.
Checkerboard: Displays the Live X-ray and DRR image in a pattern of alternating
squares to help visually evaluate image alignment. Click the icon once to display
the checkered pattern. Click again to display the pattern rotated 90°. Click once
more to turn off the checkered pattern.
Drag the red dot at the center of the checkered pattern to move the intersection of
squares in the image. Cannot be used with the Blend Tool.
Control Description
ITTV Outline: (Lung 1-View Tracking with Respiratory Modeling only) If the
treatment plan is based on a Simulation Plan, displays a dark blue Internal Target
Tracking Volume (ITTV) contour in the DRR image that is not used for tracking.
For more information, see the Treatment Planning Manual.
The ITTV Outline is turned off by default.
Tumor Outline: (Lung 1-View Tracking with Respiratory Modeling; lower right of
image panel) Displays a graphic overlay of the tumor outline (Target Tracking
Volume, TTV) in the DRR images and Live X-ray images that corresponds to the
tumor contour specified during treatment planning. With Lung 1-View Tracking
with Respiratory Modeling, the tumor outline is only displayed in the Live X-ray
image that is used for tracking.
• A dark blue outline indicates the reference tumor outline in the DRR images.
• A teal blue outline indicates where the tumor has been identified by the
tracking algorithm in the Live X-ray images.
• Red and green outlines indicate the reference tumor outline projected onto the
Live X-ray image in Offset Mode.
The Tumor Outline is turned on by default.
Crosshairs: Displays a pair of crosshairs to aid in alignment of the Live X-ray and
DRR images. Can be used with other image panel controls.
Blend Tool: Use the Blend Tool to change the blending of the Live X-ray and
DRR images.
• Drag the cursor over the green marks at the right of each image to display
the hidden slider. Drag the slider up and down to change the blending.
• Click the Automatic Blend icon to alternate continuously between the Live
X-ray and DRR images as you watch. The Live X-ray image will fade in and
out automatically.
• Click the up or down arrow below the slider to increase or decrease the speed
of the automatic blend (represented by a number).
• Click the Pause icon to pause the automatic blend.
• Click the Reset icon to reset the blend to the default (50% Live X-ray
image and 50% DRR image).
Cannot be used with Checkerboard.
Control Description
Full Content DRR: (Lung 1-View Tracking with Respiratory Modeling) Displays
the unsegmented (full content) DRR images. See “Display Modes for DRR
Images” on page 15-27.
Tumor Region DRR: (Lung 1-View Tracking with Respiratory Modeling; upper
right corner of image panel) Displays the tumor region DRR images, in which all
image content that is not in the immediate region of the tumor is removed. See
“Display Modes for DRR Images” on page 15-27.
X-ray Technique Changed: (left edge of image panel) Indicates that the X-ray
technique has changed since the currently displayed image was acquired.
Couch Moved: (left edge of image panel) Indicates that the couch position has
changed. Therefore the current image correlation results are invalid and a new
image is required.
scroll wheel Use the Blend Tool. Use the mouse scroll wheel to change the
blending of the Live X-ray and DRR images.
CTRL + Zoom in and out. (Click in this Workflow Guide help window
scroll wheel to scroll through text.)
NOTE: These display modes affect the image display only. They
do not affect the results of the 1-View with Respiratory algorithm.
You can select different display modes for each image (corresponding to Camera B and Camera
A of the X-ray imaging system).
Upon completion of the automatic correlation, a blue outline indicates the new location of
the tumor in the Live X-ray image. Updated offset values are displayed.
WARNING: Couch movement is not monitored by the collision detection system of the
CyberKnife System. Visually observe all movement of the couch to prevent potential
collisions.
6. Click the AutoCouch button in the Alignment and Tracking window. The Automatic
Patient Positioning window is displayed. You can use the automatic patient positioning
system to move the treatment couch by the amounts indicated in the Couch
Corrections text fields.
For more information on the automatic patient positioning system, see “Chapter 8:
Treatment Delivery”.
This calculated value is also displayed to the left of slider bar (left screen), where you can adjust
its threshold (see Figure 8). If the parameter value exceeds its threshold, a threshold violation
occurs, and an orange box is displayed around the slider bar. When a threshold violation occurs
the first time you are in the Alignment > Respiratory phase during a treatment fraction, you
cannot move to the next phase until the threshold violation is resolved.
Uncertainty (%)
The Uncertainty (%) parameter gives the calculated detection uncertainty value for the Lung 1-
View Tracking with Respiratory Modeling algorithm. For more information on the detection
uncertainty, see “Quality Metrics” on page 15-7.
The 1-View with Respiratory algorithm calculates a detection uncertainty value for the image
(Image A or B) that is used for tracking. If the calculated value is higher than the Uncertainty
threshold parameter, an Uncertainty error occurs (see “Error Handling” on page 15-35).
The default value of the Uncertainty (%) threshold parameter is 40%. If an Uncertainty error occurs,
you can consider increasing the Uncertainty (%) threshold if you are confident that the tracking
algorithm correctly located the target. For more information, see “Adjusting Algorithm Parameters”
on page 15-30.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
1-View with Respiratory algorithm to incorrectly identify the tumor without generating an
error. You must therefore visually verify that the tumor has been correctly identified
throughout the treatment.
• Click and drag the adjustable cursor on the slider bar to adjust the value (see Figure 8 on
page 15-29). Alternatively, enter a new value in the textbox to the right of the slider bar
and press <Enter>.
• To return to the default value, click the Reset icon. Alternatively, click and drag the
adjustable cursor on the slider bar to the vertical line. The vertical line indicates the
default value.
Images are automatically correlated when a Tracking Range parameter is changed.
Adjusting Algorithm Parameter Thresholds
A slider bar on the left screen allows you to monitor and change the value of the following algorithm
parameter threshold (see Figure 9):
• Uncertainty (%) threshold
For information on this parameter, see “Calculated Algorithm Parameters” on page 15-28.
Threshold violations are indicated when the red error bar extends beyond the adjustable threshold
cursor, and an orange box is displayed around the slider bar.
• To return to the default value, click the Reset icon. Alternatively, click and drag the
threshold cursor to the vertical line on the slider bar. The vertical line indicates the default
value.
WARNING: If you increase the value of the Uncertainty (%) threshold, it is possible for the
Lung 1-View with Respiratory algorithm to incorrectly identify the tumor without
generating an error. You must therefore visually verify that the tumor has been correctly
identified.
Treatment Delivery
This section covers the following topics:
• “Visual Inspection During Treatment Delivery” on page 15-34
• “Treatment Monitoring Graph” on page 15-34
• “Error Handling” on page 15-35
• “Patient Realignment” on page 15-36
The Delivery phase (left screen) for 1-View with Respiratory is shown in Figure 10. Controls on
this screen and the right screen allow you to monitor and verify patient alignment and tracking
results, as well as the Respiratory Model, during treatment delivery using 1-View with Respiratory.
Because the 1-View with Respiratory algorithm does not estimate rotational offsets, only
translational offsets are displayed and updated in the Calculated column during treatment.
Dashes are displayed for the calculated rotational offsets. In the Applied column, the rotational
offsets that are displayed and used to perform rotation adjustments of the LINAC during treatment
are the rotational corrections obtained from initial Synchrony Spine Tracking alignment.
For more information on treatment delivery procedures, see “Chapter 8: Treatment Delivery”. For
information on the Synchrony tracking System, see “Chapter 12: Synchrony Respiratory Tracking
System (Option)”. For information on treatment planning for treatment using 1-View with
Respiratory, see the Treatment Planning Manual.
WARNING: Visually compare the DRR and Live X-ray images to ensure that features are
similar in both.
WARNING: During treatment delivery, after acquisition of Live X-ray images, visually
verify the patient position and verify that the tumor is visible in the image used for
tracking. Make sure that anatomy in the Live X-ray and DRR images matches. For more
information on visually identifying features to be tracked, see “Chapter 8: Treatment
Delivery”.
Error Handling
The 1-View with Respiratory algorithm may fail for the following reason:
• “Uncertainty Error” on page 15-35
During patient alignment in the Alignment > Align phase, when the calculated value of this
parameter exceeds the value of its threshold, a threshold violation occurs. An orange box is then
displayed around the slider bar for that parameter threshold.
If this occurs the first time you are in the Alignment > Align phase during a treatment fraction,
you cannot proceed to the next phase until the threshold violation is resolved. After the first time
(for example, if you return to the Alignment > Align phase to realign the patient), if a threshold
violation occurs, you can proceed to the next phase. However, you will not be able to start
treatment delivery until the threshold violation is resolved.
If a threshold violation occurs while treatment is in progress in the Delivery phase, a correlation
error occurs (a Soft Stop). Treatment is paused and the error is described in the Messages panel.
To determine which threshold was violated, you can return to the Alignment > Align phase to
see which slider bar is surrounded by an orange box.
For information on resuming treatment after a Soft Stop, see “Chapter 8: Treatment Delivery”.
Uncertainty Error
An Uncertainty Error occurs when the computed value of the Uncertainty parameter exceeds the
value of the Uncertainty threshold parameter during the Delivery phase. For more information
about the detection uncertainty, see “Uncertainty (%)” on page 15-29.
WARNING: If you increase the value of the Uncertainty (%) threshold, the 1-View with
Respiratory algorithm may incorrectly identify the tumor without generating an error. You
must therefore visually verify that the tumor has been correctly identified.
Patient Realignment
The same as for patient alignment, patient realignment has two phases:
• Realignment using the Synchrony Spine Tracking Supine.
• Realignment using 1-View with Respiratory.
An example case when this is necessary is when the patient moves during treatment and global
patient rotational alignment needs to be recalculated.
If the Synchrony tracking System is being used and treatment delivery is interrupted by a Soft Stop,
an E-Stop, or a Pause, if the Respiratory correlation model is still valid, you can click the Start/
Resume button to resume treatment.
References
1. ICRU Report 62 (1999) Prescribing, Recording and Reporting Photon Beam Therapy.
International Commission on Radiation Units and Measurements, Bethesda, MD.
(Supplement to ICRU Report 50.)
The CyberKnife System displays three types of errors: recoverable Emergency Stops (E-Stops),
unrecoverable E-Stops, and Soft Stops. The Error Handling System screen displays information
about E-stop s.
Figure 1 shows an example Error Handling System screen for a recoverable E-Stop. The example
shows the E-Stop that is active and lists the actions you may take to resolve it. If there is more than
one option for resolving an E-Stop, generally, the first item in a list of resolutions is the main cause
of the problem. However, it is important to review the entire list because another item may have
triggered the error. You can correct recoverable errors. If several errors have the same timestamp,
correct the earliest error in the list first, because it may be the cause of the others. The most recent
error is listed at the top of the list.
Some errors can be caused by hardware or software. All software-related errors and hardware
errors can be either recoverable or unrecoverable. An Accuray Service Engineer must correct
unrecoverable errors. An example of an unrecoverable software error is an incorrect data file.
The following types of error messages are described below:
• “Error Handling System Screen Errors” on page A-2
• “Readiness Phase Error Messages” on page A-64
• “Delivery Phase Error Messages” on page A-68
Figure 1
• Error numbers 4xxxx are unrecoverable errors that require a software recycle and power
cycle.
10010 Error reading RAID status. Contact Accuray. Contact Accuray Customer Support.
10011 A disk of the RAID array is faulty. Contact Accuray. Contact Accuray Customer Support.
10012 Invalid RAID status. Contact Accuray. Contact Accuray Customer Support.
10013 iDMS Error: Data server RAID status is Critical. See the System Administration
application of the iDMS Data
Management System for more
information.
10014 iDMS Warning: Data server RAID status could not See the System Administration
be determined. application of the iDMS Data
Management System for more
information.
10015 iDMS Error: Unable to determine data server RAID See the System Administration
status. application of the iDMS Data
Management System for more
information.
10016 iDMS Error: Storage Vault RAID status is Critical. See the System Administration
application of the iDMS Data
Management System for more
information.
10017 iDMS Warning: Unable to determine Storage Vault See the System Administration
RAID status. application of the iDMS Data
Management System for more
information.
10020 Error reading power supply status. Contact Contact Accuray Customer Support.
Accuray.
10021 At least one computer power supply is unplugged Contact Accuray Customer Support.
or faulty. Contact Accuray.
10022 Invalid power supply status. Contact Accuray. Contact Accuray Customer Support.
10023 iDMS Warning: Storage Vault disk status is See the System Administration
Warning. application of the iDMS Data
Management System for more
information.
10024 iDMS Error: Storage Vault disk status is Failure. See the System Administration
application of the iDMS Data
Management System for more
information.
10025 iDMS Warning: Unable to determine Storage Vault See the System Administration
disk status. application of the iDMS Data
Management System for more
information.
10026 iDMS Warning: Storage Vault system status is See the System Administration
Warning. application of the iDMS Data
Management System for more
information.
10027 iDMS Error: Storage Vault system status is Failure. See the System Administration
application of the iDMS Data
Management System for more
information.
10028 iDMS Warning: Unable to determine Storage Vault See the System Administration
system status. application of the iDMS Data
Management System for more
information.
10029 Error writing file to disk. Contact Accuray Customer Contact Accuray Customer Support.
Support.
10030 Error reading disk space status. Contact Accuray. Contact Accuray Customer Support.
10031 Warning: The disk is 70% full. Please take action to Contact Accuray Customer Support.
reduce disk usage. Normal operation is allowed
until the disk is 95% full, at which point further
treatment activities will not be allowed.
10032 Warning: The disk is 75% full. Please take action to Contact Accuray Customer Support.
reduce disk usage. Normal operation is allowed
until the disk is 95% full, at which point further
treatment activities will not be allowed.
10033 Warning: The disk is 80% full. Please take action to Contact Accuray Customer Support.
reduce disk usage. Normal operation is allowed
until the disk is 95% full, at which point further
treatment activities will not be allowed.
10034 Warning: The disk is 85% full. Please take action to Contact Accuray Customer Support.
reduce disk usage. Normal operation is allowed
until the disk is 95% full, at which point further
treatment activities will not be allowed.
10035 Warning: The disk is 90% full. Please take action to Contact Accuray Customer Support.
reduce disk usage. Normal operation is allowed
until the disk is 95% full, at which point further
treatment activities will not be allowed.
10036 WARNING: The disk is 95% full! Any ongoing Contact Accuray Customer Support.
patient treatment is allowed, but new patient
treatments will not be allowed until usage is below
95% full (maximum). If the disk becomes
completely full, no system operations will be
allowed.
10037 Invalid disk space status. Contact Accuray. Contact Accuray Customer Support.
10038 Warning: The free space on the disk is low. Please Contact Accuray Customer Support.
take action to reduce disk usage. Syncing data for
service access will resume once the disk space
frees up.
10040 Error determining UPS status. Contact Accuray. Contact Accuray Customer Support.
10041 UPS reports running out of battery power. System Contact Accuray Customer Support.
is shutting down.
10042 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 15 or more
minutes remaining. If power is not restored, the
system will shutdown automatically.
10043 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 12 minutes
remaining. If power is not restored the system will
shutdown automatically.
10044 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 10 minutes
remaining. If power is not restored the system will
shutdown automatically.
10045 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 9 minutes
remaining. If power is not restored the system will
shutdown automatically.
10046 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 8 minutes
remaining. If power is not restored the system will
shutdown automatically.
10047 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 7 minutes
remaining. If power is not restored the system will
shutdown automatically.
10048 Uninterruptible Power Supply (UPS) reports Contact Accuray Customer Support.
system is running on battery power with 6 minutes
remaining. If power is not restored the system will
shutdown automatically.
10049 Failed to communicate with Uninterruptible Power Contact Accuray Customer Support.
Supply (UPS). Contact Accuray.
10050 Uninterruptible Power Supply (UPS) reports battery Contact Accuray Customer Support.
fault. Contact Accuray.
10051 Invalid UPS status. Contact Accuray. Contact Accuray Customer Support.
10052 iDMS Warning: The iDMS Uninterruptible Power See the System Administration
Supply (UPS) is now on battery power. application of the iDMS Data
Management System for more
information.
10053 iDMS Error: Unable to determine iDMS See the System Administration
Uninterruptible Power Supply (UPS) status. application of the iDMS Data
Management System for more
information.
10054 iDMS Error: Last backup failed. See the System Administration
application of the iDMS Data
Management System for more
information.
10055 iDMS Error: Unknown backup status. See the System Administration
application of the iDMS Data
Management System for more
information.
10056 iDMS Error: Minimum disk capacity not available. See the System Administration
application of the iDMS Data
Management System for more
information.
10057 iDMS Warning: Minimum disk capacity could not See the System Administration
be calculated. application of the iDMS Data
Management System for more
information.
10058 iDMS Warning: Previous data server shutdown See the System Administration
was unexpected. application of the iDMS Data
Management System for more
information.
10060 Failed to read Security Configuration file. Contact Contact Accuray Customer Support.
Accuray.
10061 Note: System datafile security check is currently Contact Accuray Customer Support.
disabled.
10062 Invalid datafile security status. Contact Accuray. Contact Accuray Customer Support.
10063 System Data Logging is not synchronized with Contact Accuray Customer Support.
iDMS. Contact Accuray.
10064 Failed to read System Data Logging status. Contact Accuray Customer Support.
Contact Accuray.
10065 System Data Logging synchronization status is Contact Accuray Customer Support.
unknown.
10066 Invalid System Data Logging status. Contact Contact Accuray Customer Support.
Accuray.
10070 Error determining iDMS connection status. Contact Contact Accuray Customer Support.
Accuray.
10071 The network connection to iDMS is unavailable. Contact Accuray Customer Support.
The current fraction (if any) and newly generated
make-up fractions will be treatable, but other
treatments will not be possible until the connection
is re-established.
10072 Invalid iDMS connection status. Contact Accuray. Contact Accuray Customer Support.
10073 OIS error: Failed to open appointment from OIS. Check Worklist service.
10074 Unable to send RT plan to OIS. Export to OIS from See the System Administration
iDMS is required. See iDMS System Administration application of the iDMS Data
for more information. Management System for more
information.
10075 Unable to send RT record to OIS. Manual input of See the System Administration
associated records into OIS is required. See iDMS application of the iDMS Data
System Administration for more information. Management System for more
information.
10076 RT plan from OIS does not match CyberKnife See the System Administration
treatment plan. See iDMS System Administration application of the iDMS Data
for more information. Management System for more
information.
10077 OIS error: Failed to send treatment record. Use Plan Administration to resend.
10078 OIS error: OIS general error. Contact Accuray Customer Support.
10080 Failed to read Supplemental Proximity Detection Contact Accuray Customer Support.
Program status. Contact Accuray Customer
Support.
10081 Invalid Supplemental Proximity Detection Program Contact Accuray Customer Support.
status. Contact Accuray Customer Support.
10082 Supplemental Proximity Detection Program is not Contact Accuray Customer Support.
currently functioning.
10084 E-Stop Interlock Control Chassis (ELCC) and If problem persists, contact Accuray
treatment robot show an interlock status mismatch. Customer Support.
10085 E-Stop Interlock Control Chassis (ELCC) and If problem persists, contact Accuray
treatment couch show an interlock status Customer Support.
mismatch.
10086 E-Stop Interlock Control Chassis (ELCC) and If problem persists, contact Accuray
LINAC application in Physics mode show an Customer Support.
interlock status mismatch.
10087 E-Stop Interlock Control Chassis (ELCC) and If problem persists, contact Accuray
Secondary Collimator System show an interlock Customer Support.
status mismatch.
10088 E-Stop Interlock Control Chassis (ELCC) and If problem persists, contact Accuray
Collision Avoidance System show an interlock Customer Support.
status mismatch.
10089 iDMS Error: The cis log file is too large. See the System Administration
application of the iDMS Data
Management System for more
information.
10090 iDMS Warning: The cis log file does not exist. See the System Administration
application of the iDMS Data
Management System for more
information.
10091 iDMS Error: Status of the cis log file is unknown. See the System Administration
application of the iDMS Data
Management System for more
information.
10094 Collimator housing ready. You can now move the If problem persists, contact Accuray
treatment robot. Customer Support.
20100 E-Stop button on the LINAC Modulator Control Release E-Stop button on Modulator
Chassis (MCC) is depressed. Control Chassis (MCC) and then
press Fault Reset button on the
MCC.
20102 One or more E-Stop buttons on the walls are (or Make sure all E-Stop buttons on the
were) depressed. walls are released.
20104 Treatment Room door E-Stop detected. Close Treatment Room door.
20106 The E-Stop button on the operator control panel is Make sure the E-Stop button on the
(or was) depressed. operator control panel is released.
20108 E-Stop button on the treatment robot Teach Release E-Stop buttons.
Pendant or manipulator controller cabinet is
depressed.
20111 E-Stop has been cleared. Try to reset the system No corrective action required
again.
20112 LINAC driver process communication error If problem persists, contact Accuray
(socket) Customer Support.
20113 LINAC driver process communication error (ping) If problem persists, contact Accuray
Customer Support.
20114 LINAC driver process communication error (CVT) If problem persists, contact Accuray
Customer Support.
20115 LINAC process communication error: LINAC If problem persists, contact Accuray
Control Computer (LCC) to Programmatic Logical Customer Support.
Circuit (PLC).
20116 LINAC process communication error: Treatment If problem persists, contact Accuray
delivery computer to LINAC Control Computer Customer Support.
(LCC)
20118 LINAC dose measurement channels differ by too If problem persists, contact Accuray
much. Customer Support.
20119 LINAC beam indicator is ON but no dose is If problem persists, contact Accuray
detected. Customer Support.
20120 LINAC outbound dose rate error If problem persists, contact Accuray
Customer Support.
20121 LINAC time limit for beam delivery has been If problem persists, contact Accuray
exceeded. Customer Support.
20122 Error: A local application is running on the LINAC Terminate the local application on
Control Computer (LCC). the LINAC Control Computer (LCC).
20124 LINAC SF6 gas pressure is low. Add SF6 gas. If problem persists,
contact Accuray Customer Support.
20128 LINAC reports an unknown hardware fault. If problem persists, contact Accuray
Customer Support.
20129 High Voltage key on the operator control panel is Turn the High Voltage key on the
not in the ON position. operator control panel to the ON
position.
20151 Linac SF6 gas is low Recharge the system with SF6 gas
20189 Dose rate A higher than limit Contact Accuray Customer Support.
20190 Dose rate B higher than limit Contact Accuray Customer Support.
20218 Beam status is on but no dose is being delivered. Contact Accuray Customer Support.
20220 No Software E-Stop detected during power up Check that the software E-Stop
sequence. bypass switch is not in maintenance
mode
20223 Dose rate A lower than limit Contact Accuray Customer Support.
20224 Dose rate B lower than limit Contact Accuray Customer Support.
20225 High voltage went off unexpectedly Contact Accuray Customer Support.
20303 System E-Stop is unexpectedly clear. Make sure Software E-Stop is not
bypassed on E-Stop Control
Chassis (ESCC).
20304 Error Handling System timed out attempting to Review prior events in this table. If
clear System E-Stop. necessary, click the Show All button
to view prior events. Resolve all
unresolved errors. Then click
Continue and try again. If problem
persists, contact Accuray Customer
Support.
20305 Error Handling System timed out while attempting Make sure Software E-Stop is not
to set System E-Stop. bypassed on E-Stop Control
Chassis (ESCC).
20306 Error Handling System failed attempting to clear Try again. If problem persists,
System E-Stop. contact Accuray Customer Support.
20307 Error Handling System internal queue error. Click Continue button. If problem
persists, contact Accuray Customer
Support.
20308 Software E-Stop unexpectedly clear. Make sure Software E-Stop is not
bypassed on E-Stop Control
Chassis (ESCC).
20312 A Secondary Collimator System device fault may If the problem persists, contact
have occurred. Accuray Customer Support. If
another E-Stop occurred first, the
problem is not due to a Secondary
Collimator System device fault.
20315 Treatment couch reports a violation of joint angle If problem persists, contact Accuray
limit(s). Customer Support.
20316 Treatment couch console issued a local interlock/ If problem persists, contact Accuray
E-Stop via the treatment couch. Customer Support.
20317 Secondary Collimator System issued a local If problem persists, contact Accuray
interlock/E-Stop via the treatment couch. Customer Support.
20318 Remote client (TMT adapter) issued a local If problem persists, contact Accuray
interlock/E-Stop via the treatment couch. Customer Support.
20319 A treatment couch client issued a local interlock/E- If problem persists, contact Accuray
Stop. Customer Support.
20320 The key on the treatment couch Teach Pendant is Put the treatment couch Teach
not in External mode. Pendant key in the External mode
position.
20322 E-Stop button on RoboCouch Hand Controller is Release E-Stop button, if button is
depressed, or there is a failure to communicate depressed. Otherwise, continue.
with Hand Controller.
20323 Unexpected RoboCouch Teach Pendant program Turn the key on the RoboCouch
execution mode. Teach Pendant back and forth. If
problem persists, cycle RoboCouch
controller power or contact Accuray
Customer Support.
20324 RoboCouch Axis A4 and/or Axis A5 limit switch is Examine the treatment couch. When
engaged. it is safe to do so, move the
treatment couch away from the limit.
20325 RoboCouch Axis A1 limit switch is engaged. Examine the treatment couch. When
it is safe to do so, move the
treatment couch away from the limit.
20326 RoboCouch forearm or offset bracket collision Examine the treatment couch. When
detection engaged. it is safe to do so, move the
treatment couch away from the limit.
20327 Contact sensor on RoboCouch upper arm has Examine the treatment couch. When
detected a pinch condition. it is safe to do so, move the
treatment couch to relieve the pinch
condition.
20328 RoboCouch contact sensor has detected a pinch Examine the treatment couch. When
condition at the foot of the table. it is safe to do so, move the
treatment couch to relieve the pinch
condition.
20329 RoboCouch contact sensor has detected a pinch Examine the treatment couch. When
condition at the foot of the table or at the offset axis it is safe to do so, move the
(Axis A7). treatment couch to relieve the pinch
condition.
20330 RoboCouch Seat Angle and/or Seat Height Make sure Seat Angle and/or Seat
Adjustment knobs are not latched. Height Adjustment knobs are fully
latched.
20331 Treatment couch received a stop message from If problem persists, contact Accuray
the Supplemental Proximity Detection Program. Customer Support.
20332 Treatment couch interlock is inconsistent with E- If problem persists, recycle power to
Stop Interlock Control Chassis (ELCC). the treatment couch.
20333 Treatment couch E-Stop is inconsistent with E- If problem persists, recycle power to
Stop Interlock Control Chassis (ELCC). the treatment couch.
20335 RoboCouch failed to set joint angle limits. Move the RoboCouch to the treat
position. If problem persists, contact
Accuray Customer Support.
20336 RoboCouch joint torque limit exceeded. If problem persists, contact Accuray
Customer Support.
20340 Xchange process canceled due to invalid sensor If problem persists, contact Accuray
triggered. Customer Support.
20341 Xchange process canceled due to invalid sensor If problem persists, contact Accuray
readings. Customer Support.
20342 Treatment robot was unable to complete a move If problem persists, contact Accuray
during the Xchange process. Customer Support.
20343 A system error has interrupted the automated If problem persists, contact Accuray
Xchange process. Customer Support.
20344 Background light on Xchange table too high. If problem persists, contact Accuray
Customer Support.
20345 An invalid process is requesting SCS task client If problem persists, contact Accuray
services. Customer Support.
20346 Two or more processes are requesting SCS task If problem persists, contact Accuray
client services. Customer Support.
20347 The request is not valid for the current status of the If problem persists, contact Accuray
collimator housing. Customer Support.
20348 Secondary Collimator System general error. If problem persists, contact Accuray
Customer Support.
20349 Iris collimator information does not match the data Pick up a different housing and then
file. pick up the Iris housing again. If
problem persists, contact Accuray
Customer Support.
20350 Air pressure for housing is too low Contact Accuray Customer Support.
20353 Xchange table drawer is open. Close the Xchange table drawer.
20357 Treatment robot is not at the designated starting Collimator Change only available
position. from perch.
20358 Problem communicating with the Xchange table. If problem persists, contact Accuray
Customer Support.
20361 Problem with pneumatics for Iris collimator Contact Accuray Customer Support.
housing.
20362 Invalid collimator cup type (Xchange table Model A Contact Accuray Customer Support.
or Model B).
20366 Invalid Xchange configuration file detected. Contact Accuray Customer Support.
20371 DVRT2 reading is not consistent with DVRT1 If problem persists, contact Accuray
reading. Customer Support.
20373 An invalid process is requesting SCS task client If problem persists, contact Accuray
services. Customer Support.
20374 Two or more processes are requesting SCS task If problem persists, contact Accuray
client services. Customer Support.
20375 The request is not valid for the current status of the If problem persists, contact Accuray
collimator housing. Customer Support.
20376 Secondary Collimator System general error. If problem persists, contact Accuray
Customer Support.
20377 ICCU DVRT Signals Latched up. Reset required. Contact Accuray Customer Support.
20378 Either collimator is missing or PTCU potentiometer If problem persists, contact Accuray
signals are latched up. Insert collimator or reset Customer Support.
power.
20380 ICCU DMOD Reference voltage operational error. If problem persists, contact Accuray
Customer Support.
20381 Iris Collimator motor brake failure. If problem persists, contact Accuray
Customer Support.
20382 Iris Collimator temperature control system failed, or If problem persists, contact Accuray
temperature out of range. Customer Support.
20383 Iris Collimator middle limit switch failure. If problem persists, contact Accuray
Customer Support.
20384 Collimator housing detection problem occurred. If problem persists, contact Accuray
Customer Support.
20387 Secondary Collimator System device fault. If problem persists, contact Accuray
Customer Support.
20388 Two or more processes are requesting SCS task If problem persists, contact Accuray
client services. Customer Support.
20398 Secondary Collimator System device on E-Stop If problem persists, contact Accuray
Interlock Control Chassis (ELCC) is in bypass Customer Support.
state.
20399 Path file and Secondary Collimator System If problem persists, contact Accuray
requests do not match. Customer Support.
20400 Internal logic error: Error Handling System (EHS) If problem persists, call Accuray
to E-Stop Interlock Control Chassis (ELCC) Customer Support.
processing
20401 Treatment delivery computer cannot communicate If problem persists, call Accuray
with E-Stop Interlock Control Chassis (ELCC). Customer Support.
20402 Treatment delivery computer command to E-Stop If problem persists, call Accuray
Interlock Control Chassis (ELCC) failed. Customer Support.
20403 Treatment delivery computer cannot connect to E- If problem persists, call Accuray
Stop Interlock Control Chassis (ELCC). Customer Support.
20404 E-Stop Interlock Control Chassis (ELCC) is not in Turn key on E-Stop Interlock Control
Remote Mode. Chassis (ELCC) to Remote Mode.
20405 E-Stop Interlock Control Chassis (ELCC) Call Accuray Customer Support.
configuration is invalid.
20406 E-Stop Interlock Control Chassis (ELCC) has a If problem persists, call Accuray
fault. Customer Support.
20408 The LINAC E-Stop button is (or was) depressed. Make sure the LINAC E-Stop button
is released.
20416 LINAC has set its interlock. If problem persists, call Accuray
Customer Support.
20417 Secondary Collimator System has an interlock. If problem persists, call Accuray
Customer Support.
20418 Treatment robot has set its interlock. If problem persists, call Accuray
Customer Support.
20419 Treatment couch has set its interlock. If problem persists, call Accuray
Customer Support.
20421 Collision Avoidance System has an interlock. Inspect the Treatment Room and
correct the problem.
20423 Operator control panel has an E-Stop Interlock If problem persists, call Accuray
Control Chassis (ELCC) fault. Customer Support.
20424 LINAC has an E-Stop Interlock Control Chassis If problem persists, call Accuray
(ELCC) fault. Customer Support.
20425 Secondary Collimator System has an E-Stop If problem persists, call Accuray
Interlock Control Chassis (ELCC) fault. Customer Support.
20426 Imaging system has an E-Stop Interlock Control If problem persists, call Accuray
Chassis (ELCC) fault. Customer Support.
20427 Treatment robot has an E-Stop Interlock Control If problem persists, call Accuray
Chassis (ELCC) fault. Customer Support.
20428 Treatment couch has an E-Stop Interlock Control If problem persists, call Accuray
Chassis (ELCC) fault. Customer Support.
20429 Wall switch has an E-Stop Interlock Control If problem persists, call Accuray
Chassis (ELCC) fault. Customer Support.
20430 Treatment Room door has an E-Stop Interlock If problem persists, call Accuray
Control Chassis (ELCC) fault. Customer Support.
20431 Collision Avoidance System has an E-Stop If problem persists, call Accuray
Interlock Control Chassis (ELCC) fault. Customer Support.
20432 Unknown E-Stop Interlock Control Chassis (ELCC) If problem persists, call Accuray
hardware fault. Customer Support.
20433 Actual and commanded interlock configurations If problem persists, call Accuray
differ. Customer Support.
20434 Failure to enable or disable the interlock of the If problem persists, call Accuray
Target Locating System (TLS). Customer Support.
20435 E-Stop Interlock Control Chassis (ELCC) reports Recycle power to the E-Stop
an incorrect mode or status. Interlock Control Chassis (ELCC),
then recycle system power. If
problem persists, contact Accuray
Customer Support.
20436 Logic error: The Error Handling System received a If problem persists, contact Accuray
command that is not allowed in the current state. Customer Support.
20441 Treatment robot reports an unknown hardware If problem persists, contact Accuray
fault. Customer Support.
20442 Treatment couch reports unknown hardware fault. If problem persists, contact Accuray
Customer Support.
20443 LINAC reports an unknown hardware fault. If problem persists, contact Accuray
Customer Support.
20444 Secondary Collimator System reports an unknown If problem persists, contact Accuray
hardware fault. Customer Support.
20445 Collision Avoidance System reports an unknown If problem persists, contact Accuray
hardware fault. Customer Support.
20449 Secondary Collimator System communication error If problem persists, contact Accuray
Customer Support.
20450 Collision Avoidance System communication error. If problem persists, contact Accuray
Customer Support.
20451 Collision Avoidance System reported a hardware If problem persists, contact Accuray
fault. Customer Support.
20469 Collimator power failure or power filter problem If problem persists, contact Accuray
detected. Customer Support.
20471 Secondary Feedback System status reporting has If problem persists, contact Accuray
timed out. Customer Support.
20472 Error in treatment workflow for Secondary If problem persists, contact Accuray
Feedback System check. Customer Support.
20481 Multi-Gain Calibration Object Obstruction Error. If problem persists, contact Accuray
Verify that no object (for example, the treatment Customer Support.
robot or treatment couch) is blocking the image
detectors.
20482 Multi-Gain Calibration Image Intensity Error. Verify If problem persists, contact Accuray
that no objects are blocking the X-ray imaging Customer Support.
system.
20483 Multi-Gain Calibration Image Acquisition Error. If problem persists, contact Accuray
Customer Support.
20484 Multi-Gain Calibration Timeout. Reset and try If problem persists, contact Accuray
again. Customer Support.
20486 Multi-Gain Calibration initialization timed out. Reset If problem persists, contact Accuray
and try again. Customer Support.
20504 Simulation Application failed to move the treatment If problem persists, reset (cycle)
couch. equipment power. If problem still
persists, contact Accuray Customer
Support.
20505 Simulation Application failed to acquire an image. If problem persists, reset (cycle)
equipment power. If problem still
persists, contact Accuray Customer
Support.
20506 Simulation Application failed to correlate images. If problem persists, reset (cycle)
equipment power. If problem still
persists, contact Accuray Customer
Support.
20507 Simulation Application failed to save information to If problem persists, contact Accuray
the file system. Customer Support.
20508 Simulation Application failed to change the X-ray If problem persists, reset (cycle)
technique. equipment power. If problem still
persists, contact Accuray Customer
Support.
20509 Simulation Application failed to achieve the If problem persists, contact Accuray
requested X-ray technique. Customer Support.
20510 Simulation Application failed to record progress. If problem persists, contact Accuray
Customer Support.
20511 Tracking offsets are not available for all images. Make sure the Respiratory External
LED Markers are visible throughout
the breathing cycle, return to the
Offset phase, and try acquiring
another image set.
20512 Simulation Application cannot move the treatment Try moving the treatment couch a
couch to the desired position because it exceeds smaller distance. If problem persists,
couch limits. contact Accuray Customer Support.
20513 No Tracking Marker LED was visible when image Check LED placement and check for
was acquired. obstacles in the room.
20514 The simulation plan was created using Contact Accuray Customer Support.
configuration or calibration data that differs from
current system configuration or calibration data.
20516 Timed out waiting for acknowledgment from the If problem persists, contact Accuray
Respiratory Tracking System. Customer Support.
20570 X-ray source A has a recoverable fault Try again and if problem persists,
call Accuray.
20571 X-ray source B has a recoverable fault Try again and if problem persists,
call Accuray.
20582 TLS not responding to Imaging Xray Parameter Send parameters again. If a timeout
message. occurs again, contact Accuray
Customer Support.
20585 Unable to generate TLS diagnostic X-rays. Check status of E-Stops and
interlocks, and retry.
20617 Treatment robot failed to set joint angle limit(s). Move the treatment robot to the
perch position. If problem persists,
contact Accuray Customer Support.
20618 Treatment robot motion stopped unexpectedly (RSI If problem persists, contact Accuray
motion). Customer Support.
20620 Treatment robot has a local Interlock but the If problem persists, contact Accuray
reason could not be determined. Customer Support.
20621 The key on the treatment robot Teach Pendant is Make sure the key on the treatment
not in External mode. robot Teach Pendant is in External
mode.
20623 A contact detection sensor has detected contact. Contact Accuray Customer Support.
20624 The Respiratory Tracking System generated Check the patient position and, if
signals that indicate the patient is out of bounds. necessary, reset the Respiratory
Model.
20625 Treatment robot EtherRSI module is not able to Contact Accuray Customer Support.
connect with Motion Tracking System (MTS).
20626 Treatment robot has not received the 'Beam On' Contact Accuray Customer Support.
signal.
20627 Treatment robot motion digital output on E-Stop Contact Accuray Customer Support.
Control Chassis (ESCC) is in unexpected state.
20628 Treatment robot received a stop message from the If problem persists, contact Accuray
Supplemental Proximity Detection Program. Customer Support.
20629 Treatment robot interlock is inconsistent with E- If problem persists, recycle power to
Stop Interlock Control Chassis (ELCC). the treatment robot.
20630 Treatment robot E-Stop is inconsistent with E-Stop If problem persists, recycle power to
Interlock Control Chassis (ELCC). the treatment robot.
20631 Fiducial Location Optimization Error Retry correlation. If this fails again,
contact Accuray Customer Support.
20632 Fiducial Location Constraint Error: Excessive Retry Acquisition and/or Attempt
fiducial migration detected. Realignment. Consider disregarding
errant fiducial.
20633 Fiducial Extraction Error: Fiducial could not be Retry Acquisition and/or Attempt
found in live image. Realignment. Check X-Ray
parameters and extraction
thresholds.
20634 Fiducial Pairing Error: Corresponding fiducials from Retry Acquisition and/or Attempt
A and B were not found. Realignment. Check X-Ray
parameters and extraction
thresholds.
20635 Fiducial Identification Error: Could not match Retry Acquisition and/or Attempt
fiducials to reference. Realignment. Check X-Ray
parameters and extraction
thresholds.
20636 Fiducial extraction confidence level is less than the Retry target acquisition. If problem
required threshold. persists try the following error
resolutions: (1) Ensure that fiducials
are within tracking range (2) Check
Image quality and/or X-ray exposure
parameters (3) Consider reducing
the confidence level threshold
20639 The Respiratory Tracking System has generated Contact Accuray Customer Support.
an unspecific error.
20641 The Respiratory Tracking System reports an The Respiratory External LED
excessive target offset estimate. Markers may have moved too far
outside the range covered by
Respiratory model images. Consider
acquiring more images or resetting
the Respiratory model. Verify that
the marker movement is as large as
reasonably possible. For Fiducial or
Lung with Respiratory, place the
markers where they move most, not
necessarily close to the target. For
Spine Prone with Respiratory, place
the markers close to the target.
20642 The Respiratory Tracking System generated Check the patient position and reset
signals that indicate the patient is out of bounds. the Respiratory Model. If problem
persists, contact Accuray Customer
Support.
20643 Communication timing between the Respiratory If problem persists, recycle power
Tracking System and treatment robot failed. and/or contact Accuray Customer
Support.
20644 The Respiratory Tracking System reports Check Respiratory External LED
Respiratory External Markers not visible. Marker placement and check for
obstacles in the room.
20645 The Respiratory Tracking System reports a Reset and re-establish the
modeling error. Respiratory Model. If problem
persists, contact Accuray Customer
Support.
20646 Imaging subsystems are out of synch. Simultaneous XRS exposure failed.
Reset can be attempted. If problem
persists, contact Accuray Customer
Support.
20647 The Respiratory Tracking System was given an If problem persists, contact Accuray
invalid command. Customer Support.
20648 The Respiratory Tracking System is operating in The Respiratory Tracking System is
Demonstration mode. configured for demonstration mode.
A Respiratory Model cannot be built.
If this is unexpected, contact
Accuray Customer Support.
20649 The Respiratory Tracking System reports a Check Respiratory Camera Array
communication error between the treatment hardware. If problem persists,
delivery computer and optical hardware. contact Accuray Customer Support.
20650 The Respiratory Tracking System reports Check the patient position and, if
excessive change in LED motion pattern. necessary, reset the Respiratory
Model.
20651 The Respiratory Tracking System failed to remove Reset Respiratory Model.
most recent model point.
20652 Target Locating System provided corrections that Re-acquire image. If problem
are outside of the tracking plane. persists, contact Accuray Customer
Support.
20657 IP dxAB magnitude error: Measured value of dxAB Check patient rotation & attempt
over threshold. realignment.
20658 The Respiratory Tracking System reports The Respiratory Model was reset.
excessive change in LED motion pattern. Check the patient position.
20659 The Respiratory Model expired. The most recent Add images (model points) to the
image (model point) is too old. Respiratory Model. If necessary,
reset the Respiratory Model first.
20661 Error detected during IP rotational calculations Contact Accuray Customer Support.
20663 Camera gains cannot be adjusted automatically. Contact Accuray Customer Support.
Try to manually adjust the gains and acquire
another image. If the adjustment fails again,
contact Accuray Customer Support.
20685 TLSC image acquisition error Camera failure occured. Reset can
be attempted.
20686 TLSC general communication failure. Restart TLS Computer. Reset can
be attempted.
20691 TLSC camera A has a recoverable fault Target Locating System (TLS)
power will be recycled. If problem
persists, contact Accuray Customer
Support.
20692 TLSC camera B has a recoverable fault Target Locating System (TLS)
power will be recycled. If problem
persists, contact Accuray Customer
Support.
20693 TLSC system has a recoverable fault Target Locating System (TLS)
power will be recycled. If problem
persists, contact Accuray Customer
Support.
20696 Image Brightness Gradient Error Take another image. If this fails
again, contact Accuray Customer
Support.
20700 The Respiratory Camera was bumped or moved. Adjust the Respiratory Camera if
The Respiratory model was reset. necessary.
20705 LINAC calibration (LinCal) Beam command failed. If problem persists, contact Accuray
Customer Support.
20706 LINAC calibration (LinCal) High Voltage command If problem persists, contact Accuray
failed. Customer Support.
20711 A treatment robot client issued a local E-Stop. If problem persists, contact Accuray
Customer Support.
20712 Treatment robot console issued a local E-Stop via If problem persists, contact Accuray
the treatment robot. Customer Support.
20713 Remote client (TMT adapter) issued a local E-Stop If problem persists, contact Accuray
via the treatment robot. Customer Support.
20714 Secondary Collimator System issued a local E- If problem persists, contact Accuray
Stop via the treatment robot. Customer Support.
20715 Treatment robot reports violation of wrist angle If problem persists, contact Accuray
limit(s). Customer Support.
20716 Treatment robot reports violation of joint angle If problem persists, contact Accuray
limit(s). Customer Support.
20717 Treatment robot joint torque limit exceeded. If problem persists, contact Accuray
Customer Support.
20732 Treatment robot and/or Secondary Collimator If problem persists, contact Accuray
System not ready to perform Laser Alignment Customer Support.
Check activities.
20735 Laser Alignment Check scan of sensor failed. Make sure laser is visible when
powered on and check for items
blocking the sensor. If problem
persists, contact Accuray Customer
Support.
20736 Treatment robot motion was interrupted. Clear Clear interlocks and try again.
interlocks and try again.
20755 Couch violates bounding box. Verify that the couch position is
within bounds for the selected path
set.
20767 The Respiratory Tracking System reports the Check the patient position and, if
prediction error exceeds acceptable threshold. necessary, reset the Respiratory
Model.
20775 Laser did not trigger the sensor over the collimator Move the treatment robot to the
retaining nut remover. perch position manually using the
Teach Pendant and try again. If
problem continues, check the
sensor for obstructions or dust.
Change collimators manually and
contact Accuray Customer Support.
20776 Laser did not meet minimum intensity threshold for Try again. If problem continues,
the quick-check sensor during exchange. check the sensor for obstructions or
dust. If problem persists, contact
Accuray Customer Support.
20777 Laser did not meet minimum intensity threshold for Try again. If problem continues,
the quick-check sensor during exchange. check the sensor for obstructions or
dust. Change collimators manually
and if problem persists, contact
Accuray Customer Support.
20778 Laser did not trigger the sensor over the collimator Enter the Treatment Room,
holder. manually change the fixed
collimator, reattach the collimator
retaining nut, and return the
treatment robot to the perch
position. If problem persists, contact
Accuray Customer Support.
20779 Laser did not trigger the sensor in the housing Move the treatment robot to the
receptacle. perch position manually using the
Teach Pendant and try again. If
problem continues, check the
sensor for obstructions or dust. If
problem persists, contact Accuray
Customer Support.
20780 Xchange System detected the incorrect collimator Enter the Treatment Room, install
after completing the exchange. the correct collimator, and verify
placement of collimators on the
table. If the correct collimator was
picked up, contact Accuray
Customer Support.
20781 Xchange table reports that there is an object in the Enter the Treatment Room and
drop-off location. verify that collimators are in their
proper locations. Verify that no other
objects are on the Xchange table.
Move the treatment robot to the
perch position and try again.
20782 Xchange table reports that desired fixed collimator Enter the Treatment Room and
is not located in the proper location on the table. verify that the fixed collimators are in
the proper locations on the Xchange
table prior to performing another
exchange.
20783 The Xchange System was unable to automatically Enter the Treatment Room,
release the fixed collimator from the collimator manually change the fixed
housing. collimator, reattach the collimator
retaining nut, and return the
treatment robot to the perch
position. If problem persists, contact
Accuray Customer Support.
20784 The Xchange System was unable to automatically Enter the Treatment Room,
pick up the desired fixed collimator. manually change the fixed
collimator, reattach the collimator
retaining nut, and return the
treatment robot to the perch
position. If problem persists, contact
Accuray Customer Support.
20785 The Xchange System was unable to automatically Enter the Treatment Room,
remove the collimator retaining nut. manually change the fixed
collimator, reattach the collimator
retaining nut, and return the
treatment robot to the perch
position. If problem persists, contact
Accuray Customer Support.
20786 The collimator retaining nut may not have been Ensure that the collimator retaining
properly installed after the collimator exchange. nut is in the proper orientation and
attached to the collimator housing.
Return the treatment robot to the
perch position to continue. If
problem persists, contact Accuray
Customer Support.
20787 The Xchange table reports invalid sensor readings Check the Xchange table to ensure
for the desired collimator housing. the collimator housing is located in
the receptacle in the proper
orientation and no foreign objects
are on or in the table. Move the
treatment robot to the perch position
and try again. If problem persists,
contact Accuray Customer Support.
20788 The Xchange table reports invalid sensor readings Check the Xchange table to ensure
for the collimator housing drop-off location. there are no foreign objects on or in
the table. Move the treatment robot
to the perch position and try again. If
problem persists, contact Accuray
Customer Support.
20789 Xchange table overtravel sensor was triggered Contact Accuray Customer Support.
before picking up or dropping off the collimator
housing.
20790 Xchange Collimator Tray not in the expected Xchange table recalibration may be
position required. Targeting accuracy check
recommended. If problem persists,
contact Accuray Customer Support.
20791 Collimator Retaining Collar could not be removed Xchange Collimator Tray not in the
expected position
20792 Collimator Retaining Collar not present or loose. Make sure that the collimator
retaining collar is present and that its
sensor is engaged.
20793 Collimator arrangement error (Xchange Tray needs Collimators are not in the expected/
resorting) appropriate spot. Please resort the
collimators.
20797 Failed to send motion command to the Treatment Reset the error and try again. If
Robot. problem persists, contact Accuray
Customer Support.
20798 Failed to drop off/pick up the collimator Manually change the collimator and
try again. If problem persists,
contact Accuray Customer Support.
20800 Manipulator is not responding to a resume. Reset the error and try again. If
problem persists, contact Accuray
Customer Support.
20801 Equipment readiness check was interrupted. Return the treatment robot to the
perch position before proceeding.
20802 Cannot launch application: Software ESTOP is Make sure Software E-Stop is not
bypassed or its state cannot be determined. bypassed on E-Stop Control
Chassis (ESCC).
20803 Error updating the position for Treat button on the If problem persists, contact Accuray
treatment couch Hand Controller. Default position Customer Support.
will be used.
20804 The center of tumor motion provided by treatment You may proceed with treatment.
planning does not lie in the tracking plane.
Treatment can proceed, but couch movement may
be necessary after entering the Lung phase.
20805 TREATMENT ERROR: Path error list is non- Contact Accuray Customer Support.
existent or all zero. Path calibration is required.
20806 Missing or incomplete error list for path file. Contact Accuray Customer Support.
20807 High Voltage key is not in the ON position. High Voltage key must be in the ON
position to enable treatment
delivery.
20808 Robot cannot move to an imaging node from the Click Resume to resume treatment
current position. delivery automation. An image will
be acquired for the next beam if
needed.
20809 Image was acquired with invalid Respiratory phase When the Respiratory Tracking
selection setting. System is being used, automatic
phase selection must be in effect.
Return to the Respiratory phase to
select automatic phase selection.
20810 Image correlation algorithm inputs are invalid. Try again and if problem persists,
call Accuray.
20811 The Respiratory Tracking System rejected a model Confirm the error and continue. If
point. problem persists, call Accuray
Customer Support.
20812 Image cannot be added to the Respiratory Model. Verify that the Respiratory External
LED Markers are visible and
enabled.
20813 Treatment timed out while downloading path file. If problem persists, reset (cycle)
equipment power. If problem still
persists, contact Accuray Customer
Support.
20814 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from Error Handling System adaptor. Customer Support.
20815 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from Image Processing adaptor. Customer Support.
20816 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from Respiratory adaptor. Customer Support.
20817 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from Data Management Bridge. Customer Support.
20818 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from treatment robot adaptor. Customer Support.
20819 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from LINAC adaptor. Customer Support.
20820 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from treatment couch adaptor. Customer Support.
20821 Timeout waiting for initialization acknowledgment If problem persists, contact Accuray
from collimation adaptor. Customer Support.
20823 Cannot enter the Delivery phase without valid Return to the Alignment phase and
offsets. acquire a new image.
20824 Respiratory Model is not valid for delivery. Return to the Respiratory phase and
establish a valid Respiratory Model.
20825 Cannot enter the Delivery phase because the Return to the Alignment phase and
treatment couch has moved since the most recent acquire a new image.
image acquisition.
20826 Cannot enter the Delivery phase because the Return to the Alignment phase and
current correlation will be discarded. acquire a new image.
20827 Cannot enter the Delivery phase because the Return to the Alignment phase and
current image is blocked. acquire a new image.
20828 Imaging during treatment delivery is disabled and Return to the Readiness phase and
there is no current image. re-enable imaging during treatment
delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20829 Imaging during treatment delivery is disabled and Return to the Readiness phase and
the current image is blocked. re-enable imaging during treatment
delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20830 Imaging during treatment delivery is disabled and Return to the Readiness phase and
the current correlation will be discarded. re-enable imaging during treatment
delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20831 Imaging during treatment delivery is disabled and Return to the Readiness phase and
there are no correlation results. re-enable imaging during treatment
delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20832 Imaging during treatment delivery is disabled and Return to the Readiness phase and
the current correlation results are not valid for re-enable imaging during treatment
treatment delivery. delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20833 Imaging during treatment delivery is disabled and Return to the Readiness phase and
the treatment couch has moved since the last re-enable imaging during treatment
image was acquired. delivery, and/or return to the
Alignment phase and obtain valid
correlation results.
20834 Timeout waiting to configure Treatment Room door If problem persists, contact Accuray
interlock. Customer Support.
20835 An error occurred during treatment delivery Reset the error and try again. If
automation. problem persists, contact Accuray
Customer Support.
20836 Imaging subsystem reports that it is not ready for If problem persists, contact Accuray
delivery automation. Customer Support.
20837 LINAC reports that it is not ready for delivery If problem persists, contact Accuray
automation. Customer Support.
20838 Timeout waiting for treatment robot subsystem to If problem persists, contact Accuray
be ready. Customer Support.
20839 Treatment robot reports that it is not ready for Return the treatment robot to the
delivery automation. perch position and try again. If
problem persists, contact Accuray
Customer Support.
20841 Secondary Collimator System timed out If problem persists, contact Accuray
Customer Support.
20842 Timeout waiting to launch delivery automation. If problem persists, contact Accuray
Customer Support.
20843 Request to move treatment robot to imaging node If problem persists, contact Accuray
was rejected. Customer Support.
20844 Treatment timed out waiting for acknowledgment If problem persists, contact Accuray
from the Respiratory Tracking System. Customer Support.
20845 Target is not in tracking plane. Go to Lung phase, move the couch
to the lung position, and perform
realignment as needed.
20846 The couch reports an invalid position. Return the couch to the treatment
position using the Teach Pendant.
20847 Treatment couch position has changed during If problem persists, contact Accuray
delivery automation. Customer Support.
20848 Simulation Application failed to move the treatment If problem persists, reset (cycle)
couch. equipment power. If problem still
persists, contact Accuray Customer
Support.
20849 Delivery automation timed out waiting for robot If problem persists, contact Accuray
movement to be completed. Customer Support.
20850 Delivery automation timed out waiting for If problem persists, contact Accuray
collimator aperture change. Customer Support.
20851 Delivery automation timed out waiting for response If problem persists, contact Accuray
from the Respiratory Tracking System or received Customer Support.
unexpected response from the Respiratory
Tracking System.
20852 Delivery automation timed out waiting for response If problem persists, contact Accuray
from LINAC susbsystem. Customer Support.
20853 Treatment cannot move the couch to the desired If problem persists, contact Accuray
position because the couch exceeds its limits. You Customer Support.
may retry after adjusting the patient position on the
couch.
20854 Treatment couch position has changed. If problem persists, contact Accuray
Customer Support.
20855 Secondary Collimator System: Unexpected status If problem persists, contact Accuray
reported. Customer Support.
20856 An error occurred while recording Secondary If problem persists, contact Accuray
Feedback System data. Customer Support.
20857 Collimator retaining nut is missing or loose. Attach collimator retaining nut and
ensure retaining nut sensor is
engaged.
20858 Treatment application failed to change the X-ray If problem persists, reset (cycle)
technique. equipment power. If problem still
persists, contact Accuray Customer
Support.
20859 Treatment application failed to achieve the If problem persists, contact Accuray
requested X-ray technique. Customer Support.
20860 Treatment application failed to acquire an image. If problem persists, reset (cycle)
equipment power. If problem still
persists, contact Accuray Customer
Support.
20864 Treatment application failed to correlate images. If problem persists, reset (cycle)
equipment power. If problem still
persists, contact Accuray Customer
Support.
20865 Delivery automation error. Attempted to acquire an If problem persists, contact Accuray
image at a blocked node. Customer Support.
20868 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error enabling high voltage. Customer Support.
20869 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error turning high voltage on or off. Customer Support.
20870 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error trying to turn on beam. Customer Support.
20871 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error trying to turn on laser. Customer Support.
20872 LINAC beam directed at insufficient vault shielding. In Treatment mode, you may retry
after adjusting the patient position.
In Physics mode, retry after
adjusting the treatment robot
position.
20873 Communication error in vault shielding check If problem persists, contact Accuray
program. Customer Support.
20874 A Peak/Valley image acquisition is not needed. Peak/Valley image acquisition is not
allowed because the Respiratory
Model already contains Peak and
Valley model points.
20875 LINAC Subsystem Control (LSC) software If problem persists, contact Accuray
encountered an error setting or clearing the beam Customer Support.
table.
20876 Treatment application timed out waiting for an If problem persists, contact Accuray
image acquisition. Customer Support.
20877 Treatment application timed out waiting for a If problem persists, contact Accuray
correlation result. Customer Support.
20878 Time out waiting for couch position update. If problem persists, contact Accuray
Customer Support.
20879 Fixed initialization did not succeed. If problem persists, contact Accuray
Customer Support.
20880 Iris initialization did not succeed. If problem persists, contact Accuray
Customer Support.
20882 Treatment delivery automation could not start If problem persists, contact Accuray
because the Treatment Room door interlock is Customer Support.
unexpectedly bypassed.
20883 Treatment delivery automation could not start If problem persists, contact Accuray
because the Error Handling System indicated that Customer Support.
the System E-Stop state is not clear.
20884 Treatment delivery automation could not start If problem persists, contact Accuray
because an acknowledgment that treatment is Customer Support.
active was not received from the Respiratory
Tracking System.
20885 Treatment delivery automation could not start. If problem persists, contact Accuray
Customer Support.
20890 Respiratory model is not valid for delivery. Review the Treatment messages in
the Delivery phase. If necessary,
return to the Respiratory phase and
build a valid Respiratory model.
20901 One or both of the E-Stop buttons on the treatment Make sure both E-Stop buttons on
couch are (or were) depressed. either end of the treatment couch
(under the table top) are released.
20902 RoboCouch E-Stop. Please check the following: Please release the treatment couch
- Hand Controller E-Stop button E-Stop button(s).
- Teach Pendant E-Stop button
- Teach Pendant External Mode switch
- RoboCouch contact sensors
- RoboCouch safety limit switches
20903 The treatment couch reports the device interlock is Make sure both E-Stop buttons on
engaged. either end of the treatment couch
(under the table top) are released.
20950 Patient Data Loader encountered an unknown If problem persists, contact Accuray
error. Customer Support.
20951 Patient Data Loader reports unknown startup If problem persists, contact Accuray
information. Customer Support.
20952 Patient Data Loader startup information is stale. If problem persists, contact Accuray
Customer Support.
20954 Patient Data Loader reports unrecognized startup If problem persists, contact Accuray
information. Customer Support.
20955 Patient Data Loader state machine received If problem persists, contact Accuray
unexpected input. Customer Support.
20958 Patient Data Loader state machine in unknown If problem persists, contact Accuray
state. Customer Support.
20960 Data Management Bridge request is stale. If problem persists, contact Accuray
Customer Support.
20961 Patient Data Loader successfully unloaded a If problem persists, contact Accuray
fraction while recovering from a previous error Customer Support.
condition.
20968 Treatment Data Manager request is stale. If problem persists, contact Accuray
Customer Support.
20969 Treatment Data Manager reported an error during If problem persists, contact Accuray
finalization. Customer Support.
20970 Fraction IDs do not all agree. If problem persists, contact Accuray
Customer Support.
20971 Data Management Bridge reports that TDCS owns If problem persists, contact Accuray
no fraction. Customer Support.
20972 Treatment Data Manager reported inconsistent If problem persists, contact Accuray
fraction. Customer Support.
20973 Dose shift not sent in Delivery phase. Send dose shift in Delivery phase.
20974 Dose shift adjustments not allowed after starting Exit treatment and send dose shift in
treatment delivery. Delivery phase of makeup fraction.
20975 Dose shift sent during treatment delivery Send dose shift in Delivery phase
automation. before treatment automation.
20976 Dose shift adjustments not allowed for Respiratory Contact Accuray Customer Support.
treatment plans.
20977 Dose shift enabled in configuration for VSI system. Dose shift is not supported for this
system, remove or disable the
setting.
20978 Tumor contour generation failed for Skull Tracking. Contact Accuray Customer Support.
20979 Tumor contour generation timed out for Skull Contact Accuray Customer Support.
Tracking.
31008 sproc creation of a process failed error Contact Accuray Customer Support.
31016 Failure to connect to TLS PC for configuration Contact Accuray Customer Support.
verification.
31017 Invalid or no license found for current system Contact Accuray Customer Support.
configuration.
31021 Failure to connect to LCC PC for configuration Contact Accuray Customer Support.
verification.
31022 Inconsistency detected in Gantry LINAC Vault type Contact Accuray Customer Support.
key.
31025 Respiratory Tracking System process died Reset (cycle) equipment power. If
unexpectedly problem persists, contact Accuray
Customer Support.
31026 Respiratory Tracking System communication error Reset the Respiratory Tracking
when reading Tracking Marker data. System (Flashpoint firmware) and
recycle power. If problem persists,
contact Accuray Customer Support.
31031 LINAC has an inconsistent beam state. Contact Accuray Customer Support.
31032 LINAC delivered dose that is larger than the Contact Accuray Customer Support.
required amount.
31034 LINAC Stop counter does not match termination Contact Accuray Customer Support.
requirement.
31035 LINAC beam symmetry error is out of tolerance. Check beam symmetry.
31036 The LINAC high voltage is locally controlled by the Exit from all LINAC Control
LINAC Control Computer (LCC). Computer (LCC) applications.
31037 The LINAC adapter is not supervised by an If problem persists, contact Accuray
appropriate application process. Customer Support.
31038 Linac power failure has been detected. If problem persists, contact Accuray
Customer Support.
31161 LCF warmup not completed in time Contact Accuray Customer Support.
31164 LCF wrote to registers while beam was on Contact Accuray Customer Support.
31165 Write to dose counters outside diagnostic mode Contact Accuray Customer Support.
31172 Could not turn off the beam Contact Accuray Customer Support.
31176 Under dose for node detected by LCF Contact Accuray Customer Support.
31177 Attempt to set stop register A failed Contact Accuray Customer Support.
31178 Attempt to set stop register B failed Contact Accuray Customer Support.
31179 Reset kill beam failed; beam on denied Contact Accuray Customer Support.
31184 LCF could not spawn a task Contact Accuray Customer Support.
31186 Attempt to reset dose while beam is on Contact Accuray Customer Support.
31187 Attempt to resume LCF while E-Stop is on Contact Accuray Customer Support.
31188 Mus from counter A and mus from counter B are Contact Accuray Customer Support.
too far apart
31192 Leakage from counter A greater than limit Contact Accuray Customer Support.
31193 Leakage from counter B greater than limit Contact Accuray Customer Support.
31194 Attempt to turn beam on in wrong state Contact Accuray Customer Support.
31195 New stop register setting less than old Contact Accuray Customer Support.
31196 Attempt to set registers while beam on Contact Accuray Customer Support.
31197 Attempt to set registers in wrong mode Contact Accuray Customer Support.
31200 DoseA greater than stopA and beam is on Contact Accuray Customer Support.
31201 Beam delivery ended because the dose counter B Contact Accuray Customer Support.
reached the stop register B before the dose
counter A reached the stop register A.
31202 Attempt to set diagnostic mode failed Contact Accuray Customer Support.
31203 Attempt to set CPU mode failed Contact Accuray Customer Support.
31206 Start treatment to LCF in wrong state Contact Accuray Customer Support.
31207 Linac controller has detected inconsistent system Contact Accuray Customer Support.
command and system status
31208 A/D dose rate too high Contact Accuray Customer Support.
31209 A/D dose rate too low Contact Accuray Customer Support.
31210 Cumulative under dose detected by LCF Contact Accuray Customer Support.
31211 AFC positive current too high Contact Accuray Customer Support.
31212 AFC positive current too low Contact Accuray Customer Support.
31213 AFC negative current too high Contact Accuray Customer Support.
31214 AFC negative current too low Contact Accuray Customer Support.
31215 New A count is less than old Contact Accuray Customer Support.
31216 New B count is less than old Contact Accuray Customer Support.
31222 New node information was received before Exit treatment and move the
previous node was completed treatment robot to the Perch position
using the Teach Pendant. Contact
Accuray Customer Support.
31320 RoboCouch offset axis (Axis A7) clutch has been Contact Accuray Customer Support.
released. RoboCouch mastering is lost. Do not
move the treatment couch or shut down or reboot
the RoboCouch controller. Contact Accuray
Customer Support.
31321 Treatment couch is not mastered. Do not move the Contact Accuray Customer Support.
treatment couch or shut down or reboot the
treatment couch controller.
31322 Treatment couch offset axis (Axis A7) is not Contact Accuray Customer Support.
coupled. Do not move the treatment couch or shut
down or reboot the treatment couch controller.
31325 RoboCouch serial number data is inconsistent. Contact Accuray Customer Support.
31326 RoboCouch controller unrecoverable error. Reset power for the RoboCouch
controller. If problem persists,
contact Accuray Customer Support.
31355 Hardware Fault: Iris collimator initialization Contact Accuray Customer Support.
required.
31378 Either collimator is missing or PTCU potentiometer If problem persists, contact Accuray
signals are latched up. Insert collimator or reset Customer Support.
power.
31379 Fixed collimator status check failed. If problem persists, contact Accuray
Customer Support.
31402 Excess dose at node detected by software Contact Accuray Customer Support.
31403 Excess cumulative dose detected by software Contact Accuray Customer Support.
31404 Under dose at node detected by software Contact Accuray Customer Support.
31405 Cumulative under dose detected by software Contact Accuray Customer Support.
31410 An unexpected Linac status has been detected. Contact Accuray Customer Support.
31411 A requested treatment robot move is not allowed. Contact Accuray Customer Support.
31413 Total dose does not match sum of doses in Contact Accuray Customer Support.
treatment plan
31415 Prescribed cGy for node returned from LCF does Contact Accuray Customer Support.
not match saved value
31416 Prescribed mus for node returned from LCF does Contact Accuray Customer Support.
not match saved value
31417 Prescribed cumulative cGy returned from LCF Contact Accuray Customer Support.
does not match saved value
31418 Prescribed cumulative mus returned from LCF Contact Accuray Customer Support.
does not match saved value
31419 Prescribed A counts at node returned from LCF Contact Accuray Customer Support.
does not match saved value
31420 Prescribed B counts at node returned from LCF Contact Accuray Customer Support.
does not match saved value
31421 Prescribed A stop count returned from LCF does Contact Accuray Customer Support.
not match saved value
31422 Prescribed B stop count returned from LCF does Contact Accuray Customer Support.
not match saved value
31423 Node not ready for treatment Contact Accuray Customer Support.
31424 Connection to the linac controller can not be Power down the CyberKnife System
established and wait fifteen minutes before
attempting to reconnect.
31512 No images are available for correlation. If problem persists, contact Accuray
Customer Support.
31606 Manipulator cannot continue Clear all estops and try to resume
again
31608 Linac power went off unexpectedly Contact Accuray Customer Support.
31609 Manipulator moved while paused. Moving the treatment robot manually
using the Teach Pendant is not
permitted during treatment. If the
treatment robot was not moved
manually, contact Accuray
Customer Support immediately.
31628 Treatment robot is not mastered. Do NOT move Contact Accuray Customer Support.
the treatment robot or shut down or reboot the
treatment robot controller.
31641 Treatment robot mastering data is inconsistent. Contact Accuray Customer Support.
31642 Treatment robot serial number data is inconsistent. Contact Accuray Customer Support.
31643 Treatment robot controller unrecoverable error. Reset the treatment robot controller.
If problem persists, contact Accuray
Customer Support.
31668 DRR download taking too long. Retry. If downloading takes too long
again, exit treatment and contact
Accuray Customer Support.
31669 The ISCAL configuration file is invalid. Contact Accuray Customer Support.
31674 Connection to TLS computer closed unexpectedly Contact Accuray Customer Support.
31678 TLSC file transfer protocol err Contact Accuray Customer Support.
31680 TLSC illegal Drr index reference Contact Accuray Customer Support.
31681 TLSC illegal patient tag error. Contact Accuray Customer Support.
31690 TLSC has generated an unspecified error Contact Accuray Customer Support.
31691 TLSC camera A has an unrecoverable fault Contact Accuray Customer Support.
31692 TLSC camera B has an unrecoverable fault Contact Accuray Customer Support.
31693 TLSC system has an unrecoverable fault Contact Accuray Customer Support.
31694 Target Locating System Computer (TLSC) has an If problem persists, contact Accuray
unrecoverable fault. Customer Support.
31730 Laser Alignment Check application general error. Contact Accuray Customer Support.
31731 Failed to communicate with required subsystems. Contact Accuray Customer Support.
Reset equipment power and try again.
31733 Laser Alignment Check: Failed to communicate Contact Accuray Customer Support.
with treatment robot.
31734 Laser Alignment Check communication error. Contact Accuray Customer Support.
31735 Invalid Laser Alignment Check configuration. Contact Accuray Customer Support.
31736 Laser Alignment Check: Invalid status reported by Contact Accuray Customer Support.
Secondary Collimator System.
31749 TLS communication adaptor task failed to start. If problem persists, contact Accuray
Customer Support.
31750 Equipment Power Status encountered an error If problem persists, contact Accuray
attempting to communicate with the LINAC. Customer Support.
31759 Error reading Couch obstacle files. Contact Accuray Customer Support.
31761 Failure to spawn LINAC communication adaptor If problem persists, contact Accuray
task. Customer Support.
31762 Proximity Detection Program (PDP) active Contact Accuray Customer Support.
monitoring of treatment robot position failed
unexpectedly.
31763 Failure to spawn SCS communication adaptor task. If problem persists, contact Accuray
Customer Support.
31765 Failure to communicate with the Proximity Contact Accuray Customer Support.
Detection Program (PDP).
31766 Xchange System communication adaptor task Contact Accuray Customer Support.
failed to start.
31767 The Respiratory Tracking System reports an invalid Contact Accuray Customer Support.
transformation matrix for imaging to robot
coordinates.
31768 Invalid Respiratory Tracking System Configuration Check the configuration files on the
Respiratory Tracking System
computer
31769 Incorrect Respiratory Tracking System firmware Contact Accuray Customer Support.
version or communication problem.
31770 System time mismatch between the imaging Contact Accuray Customer Support.
computer and treatment delivery computer. The
Time Synchronization Service (SNTP) may not be
running or may be misconfigured.
31771 The Respiratory System reports a data sampling First try recycling system power, or
problem. reboot the treatment delivery
computer. If problem persists,
contact Accuray Customer Support.
31790 Xchange Collimator Tray not in the expected System positioning check failed.
position Contact Accuray Customer Support.
31800 Cannot launch application: Image Processing Contact Accuray Customer Support.
adaptor process is stubbed or configured for
calibration.
31802 Cannot launch application: Data Management Contact Accuray Customer Support.
Bridge process is stubbed or configured for
calibration.
31803 Cannot launch application: Treatment robot Contact Accuray Customer Support.
adaptor process is stubbed or configured for
calibration.
31804 Cannot launch application: LINAC adaptor process Contact Accuray Customer Support.
is stubbed or configured for calibration.
31805 Cannot launch application: Treatment couch Contact Accuray Customer Support.
adaptor process is stubbed or configured for
calibration.
31806 Cannot launch application: Collimation adaptor Contact Accuray Customer Support.
process is stubbed or configured for calibration.
31807 Cannot launch application: Respiratory Tracking Contact Accuray Customer Support.
System process is stubbed or configured for
calibration.
31808 Error setting the interlock configuration (timeout). If problem persists, contact Accuray
Customer Support.
31809 Error setting the interlock configuration (mismatch). If problem persists, contact Accuray
Customer Support.
31812 Reference DRR file load failed. Contact Accuray Customer Support.
31815 Treatment timed out while downloading path file. If problem persists, contact Accuray
Customer Support.
31816 Cannot launch application: Error Handling System Contact Accuray Customer Support.
process is stubbed or configured for calibration.
31818 The treatment plan requires Respiratory Modeling, If problem persists, contact Accuray
but the Respiratory Tracking System is not Customer Support.
powered on or has a fault.
31819 Treatment application timed out while waiting for If problem persists, contact Accuray
the LINAC status. Customer Support.
31820 No response received from PDP for couch position Error is unrecoverable. If problem
check persists, contact Accuray Customer
Support.
31821 Error loading plan: One or more error list entries is Error is unrecoverable. If problem
missing for path file. persists, contact Accuray Customer
Support.
31822 Error loading plan: Error list file does not have at Error is unrecoverable. If problem
least one non-zero entry for path file. persists, contact Accuray Customer
Support.
31833 Treatment timed out waiting for acknowledgment Error is unrecoverable. If problem
from the automation thread. persists, contact Accuray Customer
Support.
31834 Robot cannot move to an imaging node from the Error is unrecoverable. If problem
current position. persists, contact Accuray Customer
Support.
31840 Invalid node alignment state or tracking mode for Contact Accuray Customer Support.
treatment.
31841 An error occurred while recording the dose during If problem persists, contact Accuray
treatment. Customer Support.
31845 Treatment timed out waiting for acknowledgment If problem persists, contact Accuray
from Treatment Data Manager. Customer Support.
31846 Robot moved during beam on. Contact Accuray Customer Support.
31851 Secondary Collimator System: General Error Contact Accuray Customer Support.
31853 Secondary Collimator System: Action was not Contact Accuray Customer Support.
completed within timeout period.
31855 Secondary Collimator System: Unexpected status Contact Accuray Customer Support.
reported.
31856 Xchange table drawer is open. Close the Xchange table drawer.
31857 Collimator retaining nut is missing or loose. Attach collimator retaining nut and
ensure retaining nut sensor is
engaged.
31858 Iris collimator has an accessory attached to Remove the Iris collimator
accessory flange. accessory attached to accessory
flange.
31865 Iris collimator information does not match the data Pick up a different housing and then
file. pick up the Iris housing again. If
problem persists, contact Accuray
Customer Support.
31866 The treatment plan was created using configuration Contact Accuray Customer Support.
or calibration data that differs from current system
configuration or calibration data.
31867 Treatment encountered an error while attempting If problem persists, contact Accuray
to turn off high voltage. Customer Support.
31868 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error enabling high voltage. Customer Support.
31869 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error turning high voltage on or off. Customer Support.
31870 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error trying to turn on beam. Customer Support.
31871 LINAC Subsystem Control software encountered If problem persists, contact Accuray
an error trying to turn on laser. Customer Support.
31873 Communication error in vault shielding check If problem persists, contact Accuray
program. Customer Support.
31874 Configuration error in vault shielding check Contact Accuray Customer Support.
program.
31875 LINAC Subsystem Control (LSC) software If problem persists, contact Accuray
encountered an error setting or clearing the beam Customer Support.
table.
31876 Treatment Data Manager process encountered an If problem persists, contact Accuray
error. Customer Support.
31877 Treatment Data Manager encountered error while If problem persists, contact Accuray
attempting to load current fraction. Customer Support.
31878 Treatment Data Manager had error while storing Contact Accuray Customer Support.
data.
31880 LINAC calibration (LinCal) Beam command failed. If problem persists, contact Accuray
Customer Support.
31882 LINAC Subsystem Control (LSC) software denied Reset (cycle) equipment power and
control request. try again. If problem persists,
contact Accuray Customer Support.
31883 LINAC calibration (LinCal) failed to communicate If problem persists, contact Accuray
with LINAC. Customer Support.
31884 LINAC calibration (LinCal) Beam command timed If problem persists, contact Accuray
out. Customer Support.
31885 LINAC calibration (LinCal) High Voltage command If problem persists, contact Accuray
failed. Customer Support.
31886 LINAC calibration (LinCal) High Voltage command If problem persists, contact Accuray
timed out. Customer Support.
31895 Response from LINAC Control Computer timed Contact Accuray Customer Support.
out.
31898 Error setting the Interlock configuration (mismatch). If problem persists, contact Accuray
Customer Support.
31906 Simulation Application could not save the If problem persists, contact Accuray
appropriate files. Customer Support.
31907 Simulation Application failed to initialize tracking If problem persists, contact Accuray
with the Target Locating System. Customer Support.
31908 Simulation Application failed to initialize the Motion If problem persists, contact Accuray
Tracking System. Customer Support.
31909 Simulation Application has detected an Synchrony Contact Accuray Customer Support.
Lung Tracking algorithm version mismatch.
32001 Dose shift not configured error. Contact Accuray Customer Support.
41023 Uninterruptible Power Supply (UPS) is on battery Wait for system shutdown. Contact
power and battery power is running out. System Accuray Customer Support for help
shutdown is imminent. with recovering from power outage.
41434 Command to E-Stop Interlock Control Chassis If problem persists, call Accuray
(ELCC) timed out. Customer Support.
41560 A process not responding to the status request Contact Accuray Customer Support.
41894 LINAC calibration (LinCal) failed to initialize. Contact Accuray Customer Support.
41910 Master version configuration missing from data Contact Accuray Customer Support.
files.
41911 Local version configuration missing from system. Contact Accuray Customer Support.
41913 Local version did not match master. Contact Accuray Customer Support.
41953 Patient Data Loader startup information contains If problem persists, contact Accuray
unrecognized work scenario. Customer Support.
41973 Data Management Bridge reports that download is If problem persists, contact Accuray
in progress. Customer Support.
41974 Data Management Bridge reports that TDCS If problem persists, contact Accuray
currently owns a fraction. Customer Support.
41975 Attempt to create the active fraction file failed. If problem persists, contact Accuray
Customer Support.
41976 An active fraction file already exists. If problem persists, contact Accuray
Customer Support.
42010 Collimator housing action incomplete. Treatment If problem persists, contact Accuray
robot motion inhibited. Customer Support.
Description Detail
Imaging has been disabled. Images will NOT automatically be acquired during
treatment delivery. New translation and rotation offsets
will NOT be acquired.
The Respiratory Tracking System has Respiratory motion will NOT be compensated for
been disabled. during treatment delivery.
The Respiratory Model quality is Sub- An Optimal Respiratory Model that contains 15 images
Optimal. is best for treatment.
Robot Rotation Corrections are disabled. Only translation offsets will be applied during beam
delivery. Make sure that the current rotation offsets are
small enough to be safely ignored during beam
delivery without affecting beam pointing accuracy.
Robot Rotational Bounds Checking is Rotation offsets will be measured but not applied. Only
disabled. translation offsets will be applied during beam delivery.
Large rotation offset results will NOT interrupt delivery.
Trigeminal plan: Robot Rotation Robot Rotation Corrections are disabled during
Corrections are disabled. trigeminal treatments. Only translation offsets will be
applied during beam delivery. Make sure that the
current rotation offsets are small enough to be safely
ignored during beam delivery without affecting beam
pointing accuracy.
Spine alignment was skipped. Rotation offsets based on the spine position have not
been measured and will not be applied. Only
translation offsets will be applied during beam delivery.
Description Detail
Fewer than three fiducials are selected. Rotation offsets will not be measured. Only translation
offsets will be applied during beam delivery.
Single fiducial tracking Only one fiducial is selected. Rotation offsets will not
be measured. Only translation offsets will be applied
during beam delivery. Use extreme caution! Tracking
with only one fiducial increases the risk of beam
mistargeting.
The Patient Safety Zone has been Verify that patient positioning is correct. Consider
violated. repositioning the treatment couch or selecting a
different Patient Size.
Treatment couch away from planned lung The treatment couch position has been moved away
tracking position. from the planned lung tracking position by %.1f mm in
the %s direction and %.1f mm in the %s direction.
While these values will be corrected for during delivery
there may be a similar magnitude of error in the
untracked axis that will not be corrected for. Continuing
with treatment without evaluating the impact of this
error may result in mistreatment and patient injury. 1)
Re-examine the patient setup and immobilization
devices to ensure that the CT setup has been
reproduced. 2) Carefully evaluate whether the
anatomy in the Live images is properly aligned with the
anatomy in the DRR images. Pay special attention to
the tumor position in the untracked image. 3) If the
treatment plan provides margins that allow for this
magnitude of error in the untracked axis then this
condition may be acceptable. Consult the treatment
planner. Note that repeating patient setup may help
reduce these values.
Laser Alignment Check failed. Laser Alignment Check is not calibrated. Beam
pointing accuracy cannot be verified. Call Accuray
Customer Support to calibrate the Laser Alignment
Check. Exit treatment delivery and perform an AQA
test to verify beam pointing accuracy.
Description Detail
Laser Alignment Check failed. Preparation for the Laser Alignment Check failed.
Beam pointing accuracy cannot be verified. Call
Accuray Customer Support. Exit treatment delivery
and perform an AQA test to verify beam pointing
accuracy.
Laser Alignment Check failed. Beam pointing accuracy cannot be verified. Exit
treatment delivery and perform an AQA test to verify
beam pointing accuracy. If the AQA test passes and
the Laser Alignment Check fails again call Accuray
Customer Support.
Treatment robot not at the perch position. Treatment delivery will resume with the treatment robot
away from the perch position.
Trip test of 'Beam On Too Long' error The LINAC Beam duration limits have been artificially
shortened. The duration limits have been set to
progressively shorter values to ensure that a 'Beam On
Too Long' error will be tripped.
High error threshold for dxAB. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for Brightness. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for Gradient. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for dxAB. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for drAB. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for False Nodes. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
Description Detail
High error threshold for dxAB. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for Uncertainty. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for Rigid Body. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
Low error threshold for Spacing. The error threshold for %s is set to %.1f%s. This
threshold may be too low to prevent incorrect
correlation results.
Low error threshold for Collinearity. The error threshold for %s is set to %.1f%s. This
threshold may be too low to prevent incorrect
correlation results.
High error threshold for dxAB. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
High error threshold for Uncertainty. The error threshold for %s is set to %.1f%s. This
threshold may be too high to prevent incorrect
correlation results.
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
SS User pressed soft stop button. User pressed soft stop button.
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
DI The current image has been marked Treatment delivery automation cannot
blocked. proceed with a blocked image.
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
DI The treatment robot is not at perch. The treatment robot is not at the perch
position and is not at a position from
which delivery automation can start.
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
DI New spine alignment image needed. The current spine alignment results are
not based on the most recent image. A
new spine alignment image is required
before treatment delivery automation
can start.
DI PDP couch check failed. The most recent PDP couch check did
not use the present couch position.
DI Couch position is too far Superior. Couch is too far superior to begin
treatment. This is caused by either:
1)Incorrect placement of the patient on
the couch. 2)Incorrect path selection (i.e.
head path for body treatment)
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
I The system is ready to turn on High Push the High Voltage ON button.
Voltage.
I High Voltage key is not in the ON High Voltage cannot be enabled until
position. key is in the ON position.
Type
(SS=SoftStop
DI=Delivery
Inhibitor
I=Informational) Description Details
I LINAC Control Computer (LCC) is High Voltage cannot be enabled until the
not configured for remote control LINAC Control Computer (LCC) is
from the treatment delivery configured for remote control by the
computer. treatment delivery computer.
I You may enter the Treatment Room. You may enter the Treatment Room.
Wait for the 'Fraction Completed'
message before returning to the
CyberKnife System Menu.
Classification
The CyberKnife System is classified by the following:
• Protection against electric shock: Class I, permanently connected.
• Applied part: Patient treatment table only, Type B.
• Methods of sterilization or disinfection: Not required.
• Degree of safety in the presence of flammable mixtures: Not suitable for use in the
presence of flammable mixtures.
• Mode of operation:
Continuous operation (as per 3rd edition of IEC 60601-1 standard).
Continuous operation with intermittent loading (as per 2nd edition of IEC 60601-1
standard).
Standards Compliance
System certification labels are issued by independent testing organizations. The system is
compliant with the following standards:
• EN/IEC 60601-1
• EN/IEC 60601-1-2
• EN/IEC 60601-1-3
• EN/IEC 60601-1-6
• EN/IEC 60601-2-1
• EN/IEC 60601-2-68
• EN/IEC 60601-2-54
• EN/IEC 60950-1
• EN/IEC 60825
• EN/IEC 62366
• EN/IEC 62304
• CAN/CSA C22.2 No. 60601-1
• AAMI ES 60601-1
Radiiation Leakage
Per IEC 60601-2-1, Clause 29.4.1, LINAC manufacturers must inform users if leakage radiation in
an area that can be brought into close proximity to the patient during irradiation exceeds 0.5% of
the maximum absorbed dose at 5 cm from the LINAC enclosure surface. All such points are shown
in the following figure. Of these points, the highest is 0.77% of maximum absorbed dose (at 5 cm
from the enclosure surface). In the CyberKnife System, the worst case proximity of this point to the
patient is 16.5 cm, at which distance the leakage is reduced to 0.39%.
% Leakage
System Labels
The CyberKnife System labeling includes several symbols designed to convey safety and usage
information. Table 1 lists the symbols and their description.
2D – 2-Dimensional
E2E – End-to-End (as in End-to-End test) RTSSET – Radiation Therapy Structure Set
Definitions
A. B. C. D. E. F. G. H. I. J. K. L. M. N. O. P. Q. R. S. T. U. V. W. X. Y. Z
6 Degrees of freedom
Refers to motion in 3D space defined using a Cartesian coordinate system with 6 parameters
representing translation (X, Y, and Z) and rotation (r for roll, p for pitch, and w for yaw). X, Y, and
Z represent the three Cartesian axes. See Degrees of freedom (DOF).
Accelerator
The portion of the LINAC which accelerates electrons from a gun into a high-energy electron beam.
See LINAC and Waveguide.
Alignment center
The alignment center is a point defined in the IECP (International Electrotechnical Commission
patient) coordinate system by the Accuray Precision System, based on the different requirements
of each tracking mode. During the patient alignment phase of treatment delivery, the alignment
center is aligned with the machine center in the robot frame. See Machine center.
Axial
An axial plane divides the body into upper and lower segments. This is similar to “slicing” the body
like a loaf of bread starting at the top of the head and working down.
Ball-cube
The Ball-cube is a QA tool designed by Accuray Incorporated for measuring stereotactic targeting
accuracy. Physically it is an ABS plastic film cassette used for holding two pieces of radiochromic
film. The notched films are placed orthogonally in the Ball-cube. The Ball-cube is 2.5 inches
(6.35 cm) on each side and contains a 1.25 inch (3.175 cm) acrylic ball, which is visible in a CT
image and used as a treatment target.
Beam
The CyberKnife System uses three X-ray beams: the high therapeutic energy (6 MV) radiosurgical
beam emanating from the LINAC and two low diagnostic energy beams used for image generation.
See LINAC, X-ray source (XRS), and Collimator.
Beam position
A beam position is a position of the treatment robot where the LINAC is located at a treatment node
and is pointed at the treatment target. The beam position includes the offset of the LINAC required
for targeting corrections. Where no targeting corrections are required, the beam position is the
same as the node position. See also Node position.
Beam selection
One of the primary algorithms of the Treatment Planning System (TPS). Beams are selected
according to the given lesion size and shape. The selection uses a reachability model of the
treatment robot workspace. Multiple beam directions with a high degree of collimation are used to
accurately target the lesion while sparing normal tissue. See Treatment Planning System (TPS).
Blocked node
A blocked node is a node at which the treatment robot or some other obstruction blocks the X-ray
imaging system and prevents the successful acquisition of new images. See also Node.
Centigray (cGy)
A unit of radiation dose equal to the rad, 1/100 of a Gray. Radiation dose is a measure of radiation
energy deposited per unit mass.
Centroid
The point about which the beams are assumed to move relative the respiration cycle. This point is
only relative for cases using Respiratory Modeling, because it is assumed for cases that do not use
Respiratory Modeling the moving image deforms but does not translate relative to the targeting of
the beams.
Cold spot
A region being irradiated which receives an under-dose of radiation.
Collimator
X-rays radiate in all directions when an electron beam strikes the target. Because X-rays cannot
be focused, heavy metal collimators are used to block the unwanted portion of the radiation to
create a narrow beam.
The LINAC has two collimators, a fixed primary collimator and a changeable secondary collimator.
The primary collimator surrounds the X-ray target and limits the X-ray beam to a narrow cone in
the forward direction. Secondary collimators may be attached to the X-ray head to produce
different beam sizes.
12 interchangeable fixed collimators with apertures (in millimeters) of 5, 7.5, 10, 12.5, 15, 20, 25,
30, 35, 40, 50, 60 (defined at a distance of 80 cm from the X-ray source) allow flexibility and
conformability during the treatment planning process. The fixed collimators attach to a secondary
collimator housing which is mounted on the X-ray head. The fixed collimators produce a beam with
a circular cross section. Each fixed collimator has a different length in 1/16 in (1.5875 mm)
increments based on its aperture. The LINAC contains dual independent linear potentiometers to
detect the length of the currently installed fixed collimator. Treatment cannot proceed with the
wrong collimator installed.
The optional Iris Variable Aperture Collimator provides the same 12 apertures as the fixed
collimators. The aperture of the Iris Collimator is adjustable under computer control. It contains 2
stacked hexagonal banks of tungsten segments that together produce a 12-sided aperture (a
regular dodecagon). The Iris Collimator allows consolidation of multiple path sets with multiple
fixed collimators into a single path set. Using the tungsten segments to rapidly adjust the aperture,
the Iris Collimator can deliver multiple-sized beams from each LINAC position.
The optional Multileaf Collimator is a secondary beam collimator composed of two opposing banks
of beam attenuators. Each bank is composed of 41 individual leaves which can be moved
independently. Used to shape the radiation beam to irregular apertures defined during treatment
planning.
A collimator is also used on each of the imaging X-ray sources to minimize patient exposure.
Conformal
The extent to which the radiosurgical treatment dose is restricted to the target volume, sparing
surrounding healthy tissue, which may be critical or sensitive to radiation.
See also Dose.
Coronal
A frontal image of the body. A coronal plane divides the body into front and back segments.
Couch
See Patient treatment table.
Critical region
An important anatomical structure that is especially sensitive to radiation such as the eyes, optic
nerve, or brain stem.
Deformable registration
A registration model that allows for relative internal motion between the two images.
Deformed image
A moving image that is a CT image series that has been transformed using a deformable
registration algorithm so that it aligns with a fixed CT series.
Dmax
The depth at which the radiation dose reaches a maximum in water or water-equivalent material,
such as soft tissue.
Dose
A measure of accumulated radiation exposure.
Dose Calculation
Calculation of Monitor Units (MUs) to be delivered by the LINAC for each beam to irradiate the
tumor with the prescribed dose.
Dosimetry system
Dosimeters measure the energy carried by a beam over time. The dosimetry system in the LINAC
uses dual independent dosimeters (named A and B or primary and secondary) at the X-ray source
to measure its output. They are configured in a redundant dose monitoring combination to
safeguard against incorrect dose delivery. Irradiation at a node is terminated by the dosimeter
hardware when the delivered dose at the node is equal to the planned dose. The first dosimeter to
reach the desired dose causes the LINAC to shut off. The software uses the dose from the primary
dosimeter (dosimeter A) as the measure of delivered dose.
Electron gun
See Gun.
Error recoverable
An error condition that causes an E-Stop or interlock. Treatment can resume once the error
condition is cleared.
Error unrecoverable
An error condition that causes an E-Stop or interlock. Equipment power must be cycled to reset
the E-Stop after the error condition has cleared.
E-Stop
Emergency Stop. Pressing an E-Stop button manually executes an Emergency Stop. An E-Stop
button performs the same function as a software interlock. However, it interacts directly with
CyberKnife System hardware to stop operation. Treatment couch movement, treatment robot
movement, X-ray image acquisition, and radiation delivery are immediately terminated and the
high voltage to the LINAC is turned off.
Fiducials
Artificial references (such as gold seeds or stainless steel screws) that are implanted in or near
tissues undergoing radiosurgery treatment. These markers are used as artificial landmarks by the
imaging system.
Fixed image
Primary CT image used for planning and DRR generation. Also referred to as Image A.
Forward planning
Treatment planning for a given beam selection and dosage. The computer finds the dose
distribution. See Treatment Planning System (TPS).
Fractionated radiosurgery
Treatment that has been partitioned into several doses repeated over time (such as several days).
Frame
The coordinate reference frames used to locate the treatment robot, LINAC, and patient. The
coordinate frames used with the CyberKnife System are in general 3D Cartesian coordinate frames
with 6 degrees of freedom: 3 translations (X, Y, and Z) and 3 rotations (yaw, pitch, and roll; or w,
p, and r). See IECP (IEC patient) coordinate system, Robot frame, and Patient coordinate system.
Frameless
Radiosurgery systems such as the Gamma Knife use a frame rigidly attached to the skull to ensure
accurate radiation targeting and delivery. The CyberKnife System is considered frameless since
only a non-intrusive restraint is used. The CyberKnife System targets off of bony landmarks and
fiducials. The imaging system on the CyberKnife System allows the treatment robot to correct for
small movements of the target during treatment delivery.
Function
An operation the user can perform.
Gafchromic Film
A class of self-developing radiochromic film (http://www.ashland.com) commonly used for
measuring ionizing radiation dose. See Ball-cube and EBT type film.
Gray
A unit of radiation dose equal to 100 cGy. Radiation dose is a measure of radiation energy
deposited per unit mass (1 Gray = 1 Joule/Kg). See Centigray (cGy).
Gun
The electron gun in the LINAC that injects a low energy electron beam into the accelerator. The
gun is a vacuum tube that consists of the following:
• Cathode with a filament heater to produce the electrons
• Control grid to turn the electron beam on and off
• Anode to provide initial acceleration for the beam.
The high voltage power supply provides a negative voltage to the cathode.
Hot spot
A region being irradiated that receives an abnormally high dose of radiation.
Image registration
Spatial transformation to align images.
Imaging system
The imaging system locates the treatment center by comparing real-time radiographs of the patient
with DRRs.
The imaging system contains two nearly identical X-ray imaging units. Each unit has the following
components:
• Standard medical radiographic X-ray tube
• High-voltage generator
• Tungsten copper X-ray collimator
• X-ray detector
Inverse planning
In the Accuray Precision System, the user specifies the following constraints when performing
inverse planning:
• Minimum lesion dose (in cGy)
• Maximum lesion dose (in cGy)
• Maximum beam weight (in MU)
The computer uses a linear programming algorithm to find a dose for each beam to satisfy the
constraints. See Treatment Planning System (TPS).
Ionizing radiation
A radiation beam that, when passing into a biological absorbing medium such as body tissue,
transfers some of the energy carried in the beam to the medium sufficiently to ionize the tissue at
the molecular level and therefore to disrupt cellular reproduction processes. The energy
transferred may produce a biological effect in the medium. Use of the word “radiation” in this
document will mean ionizing radiation.
Isocenter
The point in a target where the central axes of treatment beams intersect.
Isocentric planning
Creating treatment plans using isocentric geometry.
Isocrystal
Target located at the tip of the isopost that is used with the laser in the LINAC to perform treatment
robot pointing calibration. Indicates the precise position of the machine center. See Isopost.
Isopost
Calibration tool used to establish the machine center. See Isocrystal.
Key switch
The radiation enable switch in the operator control panel. When turned on, it provides the “high
voltage enable” to the LINAC for treatment delivery, simulation, or calibration.
LINAC
Acronym for linear accelerator. A linear accelerator is a device that uses high frequency
electromagnetic waves to accelerate charged particles such as electrons to high energies through
a linear tube. In the CyberKnife System, the LINAC uses a magnetron to produce X-band
microwaves to accelerate an electron beam to a nominal energy of 6 MeV. The beam strikes a
tungsten target where X-rays are produced. Here the X-ray beam energy is measured by two
dosimeters. The X-rays are reduced to a narrow beam by a fixed primary collimator and then by a
secondary changeable collimator. The entire apparatus is mounted on the end of the treatment
robot arm that aims the X-ray beam at the treatment site in the patient. Radiation is only fired from
the LINAC when the treatment robot is stopped and located at a node.
Table 1 LINAC safety interlocks: Conditions when status indicators are illuminated
DOOR INTLK (Door Interlock) All treatment room doors are closed.
WATER FLOW (Water Flow Interlock) Cooling water or air flow is adequate.
DOSE P.S. (Dose Power Supply) Dosimetry high voltage power supplies are operating
properly.
HVOC (High Voltage Over-Current) Excessive current is not being drawn from the high
voltage power supply (to magnetron).
MAN OVERRIDE (Manual Override) When the CyberKnife System operates under
computer control this light is illuminated.
GRID PWR SPLY (Grid Power Supply) Accelerator electron gun power supply is working
properly.
HV ENABLE (HV Enable) Treatment delivery computer has enabled the high
voltage.
EXT E-STOP (External E-Stop) No E-Stops and fault conditions monitored by the E-
Stop Control Chassis (ESCC) are currently
triggered.
HEAD FANS (X-ray Head Fans) The fans in the LINAC X-ray head are operational. If
the fans are not working this light will extinguish and
a Water Flow Fault will occur.
Machine center
The machine center is a physical reference point chosen during system installation in the center of
the patient treatment area. The origin of the robot frame is located at the machine center.
The machine center provides a reference point for treatment robot calibration. During system
installation, and at other times when system calibration is required, the machine center is used to
provide a precise reference point to correct the computed node file. That is, all beams from the
LINAC are corrected to point at precisely the same point, the machine center. The Laser Alignment
Unit (LAU) is used to perform this calibration. See Robot frame, Alignment center, and Isopost.
Magnetron
A magnetron is a vacuum tube device that produces microwaves. It functions as a high-power
oscillator, generating microwave pulses of several microseconds duration and with a repetition rate
of less than 240 pulses per second. The LINAC uses a mechanically tunable magnetron to provide
the energy necessary to accelerate the electron beam. See LINAC.
Makeup fraction
A new treatment fraction created when an executing plan is aborted during treatment delivery. The
makeup fraction contains the remaining dose to be delivered to the patient. It consists of all paths
and beam that were not executed. Monitor Unit fractions are generated at any time after first image
acquisition if the selected treatment plan fraction is not fully treated.
Mass density
The mass per unit volume in units g/cm3.
MeV
One million electron volts, a unit of energy used to specify the energy of particles such as the
electron beam in the LINAC. One electron volt is the energy equivalent of an electron being
accelerated in free space through a field of one volt. The LINAC produces a nominal 6 MeV
electron beam.
Modulator
For the VSI system, contains the pulse power modulator generator for the magnetron and the high
voltage supply for the LINAC gun.
For the S7 and M6 system, contains the pulse power modulator generator for the magnetron.
Moving image
A medical image used to assist delineation and evaluation of specific Volumes of Interest. Also
referred to as Image B.
MV
A medical radiation term that is a measure of the energy of a spectrum of X-ray photons. See MeV.
Node
One of a set of predefined nominal points where the source of the radiation beam (the LINAC) can
be placed. Node locations are fixed and cannot be altered for individual treatment plans. See
also Node position.
Node position
A position of the treatment robot where the LINAC is located at a node and is pointed at the
machine center. See also Node, Beam position.
Non-isocentric planning
Common in CyberKnife treatment plans, non-isocentric beam geometry utilizes beams aimed at
various points on a target(s). This is useful for dose painting.
Penumbra
The edge region surrounding a beam where the energy level drops from 80% to 20% of its
maximum.
Perch position
Rest position of the treatment robot.
Phantom
A physical or mathematical substitute for the patient. A 3D photographic film cube is placed in the
phantom in the particular volume. The exposed film shows the dose distribution applied during a
simulation run.
The phantom with film cube is used to perform various QA tests. In an End-to-End test, a CT image
is generated of a dosimetric phantom and a plan is created to treat a specified location in the film
pack. After irradiating the phantom, the film pack is read and a comparison is made of the location
of the planned treatment with the center of the delivered dose. In this way the total system accuracy
is tested. See Ball-cube.
Phantom mode
One of the operational modes of the CyberKnife System, a physical execution of the treatment plan
by the system with a dosimetric phantom instead of a patient. The primary purpose is to check for
the accurate and safe delivery of a patient treatment plan.
Prescription
Dose to be delivered to the tumor.
Rad
A unit of radiation dose, replaced by the centigray (cGy).
Radiosurgery
The treatment of lesions using stereotactically directed convergent beams of external irradiation,
usually in 1 to 5 fractions.
Reference image
The Digitally Reconstructed Radiographs (DRRs). See Digitally Reconstructed Radiograph (DRR).
Reference plan
A plan that is displayed for comparison.
Robot frame
(The treatment robot is also referred to as the treatment manipulator.) The origin of the robot frame
is the machine center. Plans created by the Accuray Precision System are saved in path files that
use the robot frame. Its X and Y axes depend on where the base of the robot is with respect to a
patient in the HFS position.
The base of the robot may be at the Left-Superior or the Right-Superior of the patient. The
orientation and rotation of the robot are calibrated during system installation.
If the robot is at the patient's Right-Superior, X is positive to the patient's Left-Inferior (see
Figure 1). This axis is roughly 45° from the patient's Inferior. Y is positive to the patient's Left-
Superior and is roughly 45° from the patient's Left. The actual orientation of the robot relative to the
treatment couch is measured during system installation.
If the robot is at the patient's Left-Superior, X is positive to the patient's Right-Inferior and is roughly
45° from the patient's Right. Y is positive to the patient's Left-Inferior and is roughly 45° from the
patient's Inferior.
The Z axis is positive to the patient's Anterior regardless of where the robot is. See Machine center
and IECP (IEC patient) coordinate system.
A = Anterior
P = Posterior
S = Superior
I = Inferior
R = Right
L = Left
A
L I
S
R
Machine Center
P
Robot Base
at Patient
Right-Superior
Sagittal
A side image of the body. A sagittal plane divides the body into left and right segments.
Seed point
A point placed on the patient images to obtain a starting position for the image registration.
Sequential Optimization
Sequential optimization is an algorithm used to optimize a Accuray Precision System treatment
plan during the planning stage.
Soft Stop
A non-emergency treatment interruption that switches from treatment delivery to patient
realignment without generating an E-Stop.
Step
A group of functions in the Accuray Precision System organized into a single user interface layout
(screen).
Stereotactic
Any device that accurately guides a surgical tool in space. Adjective form of stereotaxis.
See Stereotaxis.
Stereotaxis
Stereotaxis is a branch of neurosurgery that utilizes spatial information provided by neuroradiologic
studies to treat certain disorders of the central nervous system with great accuracy.
See Stereotactic.
Target
There are two different targets referred to in the CyberKnife System:
• Inside the LINAC, a 6 MeV electron beam is aimed at a tungsten target to generate 6 MV
X-rays.
• The X-ray beam is aimed at a second target, the tumor in the patient.
Task
A group of step in the Accuray Precision System. Each task is displayed on a tab.
Teach Pendant
A hand held device used to manually control the treatment robot or to program sequences of
movements for it. In the CyberKnife System, the operator uses the Teach Pendant to position the
treatment robot for calibration or maintenance, and to manually power up components such as
alignment lasers or imaging X-ray sources. Maintenance mode can only be accessed with a
special key. The CyberKnife System is used for patient treatment only when the treatment robot is
under computer control.
Tool frame
One of the coordinate frames used by the CyberKnife System. The tool frame is a Cartesian
coordinate system with 6 degrees of freedom: 3 translations (X, Y, and Z) and 3 rotations (yaw,
pitch, and roll). The tool frame location (X, Y, and Z) is defined relative to the faceplate of the
treatment robot. The faceplate is located at the end of the “wrist” of the treatment robot (Axes A4,
A5, and A6) where the LINAC is mounted.
The Z axis is perpendicular to the faceplate (positive pointing straight out). The origin of the tool
frame is called the tool center point (TCP). The TCP for the CyberKnife System is at the nominal
point of the LINAC X-ray source. Its position moves with the treatment robot and is continually
being recalculated during treatment delivery. Rotations are defined around the axes of the tool
frame, which depend on the particular orientation of the treatment robot at that time. See Robot
frame and World frame.
Treatment path
The path the LINAC would follow in moving through a set or a subset of nodes. All defined nodes
are visited during radiation treatment or simulation. No nodes are skipped even if radiation is not
activated at all nodes. Normally radiation is only activated at a subset of the defined nodes. For the
CyberKnife System, there are single- or 3-path options available.
Voxel
A 3-dimensional pixel, an atomic element of a 3D image. An abbreviation for “volume pixel.”
Water phantom
A precise radiation measurement device used for characterization and calibration of the LINAC
beams. Every LINAC is different. In the CyberKnife System, the beam from each collimator is
characterized using a water phantom and the data collected is stored in a system file. This beam
data is used in treatment planning to determine how much dose to deliver.
Waveguide
The structure in the accelerator that develops electric fields necessary to accelerate electrons.
Utilizes microwaves of the appropriate frequency and power to provide the necessary energy.
Workspace
The workspace is the set of configurations that the treatment robot can assume without entering a
forbidden region. A forbidden region is an obstacle plus any buffer around the obstacle. Defined
obstacles in the treatment area include the patient, couch, camera stands (for X-ray image
detectors), floor, and ceiling.
World frame
One of the coordinate frames used by the CyberKnife System. The World frame is a Cartesian
coordinate system (X, Y, and Z expressed in millimeters) whose origin is predefined by the
treatment robot manufacturer for the particular model of treatment robot. For the CyberKnife
System, the origin of the World frame is located at the intersection of the 2 joints (Axis 1 and Axis
2) at the base of the treatment robot. See Tool frame and Robot frame.
X-band
A frequency spectrum in the microwave band. The LINAC magnetron operates in the X-band. X-
band microwaves are used in the CyberKnife System LINAC because it allows for smaller
treatment robot-mounted LINAC components than would a lower frequency band. Conventional
LINAC accelerators typically use the lower-frequency S-band.
X-ray detectors
Amorphous silicon flat panel X-ray image detectors used by the Target Locating System computer
to obtain images of the patient for use in patient alignment and motion tracking.
X-ray head
The portion of the LINAC that is mounted to the end of the treatment robot arm and provides the
radio-surgical X-ray beam.
X-rays
Photons in the energy band above ultraviolet light that have been produced by high energy
particles colliding into a target.