Step 8 – Recommended Doses and Exposure Factors
TM 726-722-G1 Step 8 - 2
Step 8 – Recommended Doses and Exposure Factors
1. KVP
Excellent-quality digital images are best achieved with settings of 100KVP or less due to the
inherent characteristics of the image intensifier and digital camera combination. 120 KVP should
not be used for most digital imaging systems.
In addition, lower dose exposures are a significant advantage of digital systems, but when
combined with high KVP cause inadequate exposure technique with many generators, due to
insufficient exposure times.
2. Recommended Ranges
NOTE: Failure to install the x-ray generator signal line for RAD EXPOSE ON will result
in exposure termination at 180 milliseconds.
TM 726-722-G1 Step 8 - 3
Step 8 – Recommended Doses and Exposure Factors
Recommended Doses
Continuous (μR/S)
Pulsed-15pps (μR/Fr) 2.1 2.9 3.8 5.1
Pulsed- 7.5pps (μR/Fr) 2.7 3.7 5.0 6.7
Pulsed 3.75pps (μR/Fr) 3.2 4.4 5.9 7.9
Pulsed-1pps (μR/Fr) 5.0 6.9 9.2 12.3
TM 726-722-G1 Step 8 - 4