User's Guide: Shop Online at
User's Guide: Shop Online at
TM
Title
http://192.168.1.200
TITLE
Temp. 0015.7 Dew. 0005.7 C Temp_RH_hPa RH 0040.9 Pres. 1020.8 Shop online at
omega.com
50 1200
100%
e-mail: info@omega.com
For latest product manuals:
hPa
5/Div 90/Div
www.omegamanual.info
0%
0 300
Wed Jun 01 12:00:00 PDT 2005 1 Month (1 Day/Div) Tue Jun 28 12:00:00 PDT 2005
1 Minute
1 Hour
1 Day
1 Week Main Menu
1 Month
1 Year
and
Barometric Pressure,
Temperature + Humidity
omega.com info@omega.com
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
TABLE OF CONTENTS
Part 1: Introduction
1.1 Safety and EMC Considerations........................................................................2
1.2 Before You Begin ................................................................................................2
1.3 Description ..........................................................................................................2
Part 2: Hardware
2.1 Mounting .............................................................................................................4
2.1.1 Mounting the Wall Mount iServer .......................................................4
2.1.2 Mounting the DIN Rail iServer ............................................................5
2.1.3 Removal from DIN Rail ........................................................................5
2.2 DIP Switches........................................................................................................6
2.2.1 DIP Switch Usage ................................................................................6
2.3 Parts of iServer Unit ...........................................................................................7
2.4 Network Communication Interfaces .................................................................8
2.4.1 10Base-T RJ-45 Pinout.........................................................................8
2.4.2 10Base-T Crossover Wiring .................................................................8
2.5 Industrial Probe ....................................................................................8
Part 3: Network Configuration
3.1 Network Protocols .............................................................................................9
3.2 Ethernet (MAC) Address ....................................................................................9
3.3 DHCP .............................................................................................................10
3.4 DNS .............................................................................................................10
3.5 Default IP Address.............................................................................................11
3.6 Port Number.......................................................................................................11
Part 4: Operations
4.0 Testing the Connection.....................................................................................12
4.1 iConnect Software.............................................................................................13
4.2 Setting a new IP Address over the Network ..................................................15
4.3 Setup and Operation using a Web Browser ..................................................16
4.3.1 Read Sensor .......................................................................................17
4.3.1.1 Java Runtime Environment 1.7 Setup Instructions.........................17
4.3.1.2 Browser Proxy Selection ...................................................................18
4.3.2 Chart ....................................................................................................19
4.3.3 Configuration ......................................................................................20
4.3.4 Sensor Parameter ...............................................................................22
4.3.5 Access Control ..................................................................................23
4.4 Telnet Setup ......................................................................................................25
4.5 HTTPget Program..............................................................................................25
4.5.1 HTTPget using Port 1000 ..................................................................26
4.5.2 HTTPget and ARP to setup Device IP Address ...............................27
4.6 ARP Protocol .....................................................................................................27
4.7 Remote Access (Tunneling) .............................................................................29
4.7.1 Local iServer .......................................................................................30
4.7.2 Remote iServer ..................................................................................31
4.8 iLog Software ....................................................................................................32
4.9 Mail Notifier Software .......................................................................................34
4.9.1 Installation...........................................................................................34
4.9.2 Program Options Setup and Configuration ....................................35
4.9.2.1 Sending Email Messages to a Cell Phone........................................36
4.9.3 Device Setting Setup and Configuration ..........................................36
Part 5: Specifications .................................................................................................................. 37
Part 6: Factory Preset Values ......................................................................................................40
Appendix A Glossary.............................................................................................................41
Appendix B IP Address ........................................................................................................42
i
Appendix C IP Netmask.........................................................................................................43
Appendix D ASCII Chart .......................................................................................................44
ASCII Chart Control Codes .............................................................................45
Appendix E iLog Error Messages.........................................................................................46
Appendix F Sensor System ..................................................................................................47
Part 7: Approvals Information
7.1 Electromagnetic Compatibility (EMC) ............................................................49
7.2 FCC ..............................................................................................................49
LIST OF FIGURES:
Figure 1.1 iServer with Probe on the Ethernet Network ....................................................3
Figure 2.2 Mounting - Wall Mount iServer ..........................................................................4
Figure 2.2 Mounting - DIN Rail iServer ...............................................................................5
Figure 2.3 Removal - DIN Rail iServer ...............................................................................5
Figure 2.4a DIP Switch - Wall Mount iServer ........................................................................6
Figure 2.4b DIP Switch - DIN Rail iServer ............................................................................6
Figure 2.5 Parts of iServer Unit............................................................................................7
Figure 2.6 RJ45 Pinout ........................................................................................................8
Figure 2.7 10Base-T Crossover Cable Wiring ...................................................................8
Figure 2.8 Industrial Probe Wiring Hookup ........................................................................8
Figure 3.1 Labeling ..............................................................................................................9
Figure 3.2 DIP Switch on Bottom Side of iServer ............................................................10
Figure 3.3 Telnet Login into the iServer ............................................................................11
Figure 4.1 Pinging the iServer from MS-DOS Prompt ...................................................12
Figure 4.2 Assigning an IP Address using iConnect ......................................................13
Figure 4.3 Accessing the iServer for Configuration .......................................................13
Figure 4.4 Access Control ................................................................................................15
Figure 4.5 iServer Home Page Menu ...............................................................................16
Figure 4.6 Login and Administrator Password ................................................................16
Figure 4.7 Read Sensor ....................................................................................................17
Figure 4.8 Chart ..............................................................................................................19
Figure 4.9 Configuration ...................................................................................................20
Figure 4.10 Sensor Parameter ............................................................................................22
Figure 4.11 Remote End Char ..............................................................................................23
Figure 4.12 Access Control .................................................................................................24
Figure 4.13 ARP Commands and Responses ...................................................................28
Figure 4.14 PC-to-Device Communication..........................................................................29
Figure 4.15 Device-to-Device Communication ...................................................................29
Figure 4.16 Local iServer (Serial to Ethernet) Configuration Page .................................30
Figure 4.17 Remote iServer Configuration Page ...............................................................31
Figure 4.18 iLog Software Logging Data ............................................................................32
Figure 4.19 iServer Mail Notifier Main Window ..................................................................33
Figure 4.20 iServer Mail Notifier Profile Setup ...................................................................34
Figure 4.21 iServer Mail Notifier Device Setting ................................................................35
Figure F.1 RH Accuracy Chart ...........................................................................................47
Figure F.2 Temperature Accuracy Chart ..........................................................................47
Figure F.3 Normal Range ...................................................................................................47
LIST OF TABLES:
Table 2.1 Parts of iServer Unit ....................................................................................7
ii
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by the following labels:
• NOTE
• WARNING or CAUTION
• IMPORTANT
• TIP
FEATURES
• Virtual Chart Recorder
• Web Server
• Accurate Readings
• Password Protection
• Email Alarms
• Data Logging
1
PART 1
INTRODUCTION
1.1 Safety and EMC Considerations
Refer to the CE Approval Section
EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use twisted-pair wires for signal connections.
• Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
Failure to follow all instructions and warnings may result in injury!
1.2 Before You Begin
Inspecting Your Shipment: Remove the packing slip and verify that you have received
everything listed. Inspect the container and equipment for signs of damage as soon as
you receive the shipment. Note any evidence of rough handling in transit. Immediately
report any damage to the shipping agent. The carrier will not honor damage claims
unless all shipping material is saved for inspection. After examining and removing the
contents, save the packing material and carton in the event reshipment is necessary.
Customer Service: If you need assistance, please contact the Customer Service
Department nearest you.
Manuals, Software: The latest Operation Manual as well as free iServer configuration
software (iConnect), datalogging software (iLog) and Mail Notifier are available at the
website listed on the cover page of this manual or on the CD-ROM enclosed with
your shipment.
1.3 Description
View Barometric Pressure, Temperature + Humidity with a Web Browser.
The iServer transmitter let’s you monitor and record Barometric Pressure, Temperature,
Relative Humidity and Dew Point over an Ethernet network or the Internet with no special
software except a Web Browser.
The iBTHX is for barometric pressure, temperature, relative humidity and dew
point. The iBTX is for barometric pressure and temperature. The difference
between the iBTHX and the iBTX is only the probe/sensor.
The iServer serves Active Web Pages to display real time readings, display charts of
barometric pressure, temperature and humidity, or log data in standard data formats for
use in a spreadsheet or data acquisition program such as Excel or Visual Basic.
The virtual chart viewed on the web page is a JAVA™ Applet that records a chart over
the LAN or Internet in real time. With the iServer, there is no need to invest time and
money learning a proprietary software program to log or chart the data.
Adjustable Charts: Chart scales are fully adjustable on the fly. For example, the chart
can display one minute, one hour, one day, one week, one month or one year.
Temperature and humidity can be charted across the full span (-40 to 85°C, and 0 to
100% RH) or within any narrow range (such as 20 to 30°C). Barometric Pressure can be
displayed in mbar (hectopascals hPa), millimeters of Mercury (mmHg), or inches of
Mercury (inHg).
Display and Chart Measurements: The iServer transmitters come complete with a barometric
pressure, temperature and humidity probe for measurement of a single location.
2
Award-winning Technology: The iServer is simple to install and use. It features award-
winning technology that requires no special software except a Web Browser.
The iServer connects to an Ethernet Network with a standard RJ45 connector and sends
data in standard TCP/IP packets. It is easily configured with a simple menu using a Web
Browser and can be password protected. From within an Ethernet LAN or over the
Internet, the user simply types its IP address or an easy to remember name such as
"Room 5" or "ServRoom" in any Web Browser, and the iServer serves a Web Page with
the current readings.
Typical Applications: The iServer is great for monitoring temperature + humidity in
applications such as: clean rooms, computer rooms, HVAC systems, pharmaceutical
and food processing and storage, hospitals, laboratories, semiconductor fabs, electronic
assembly, warehousing, museums, manufacturing, greenhouses, farm animal shelters,
and many more.
Email Alarms: All iServer models that are on a LAN that is connected to the Internet can
trigger an alarm that can be sent by email to a single user or to a group distribution list,
including text messages to cell phones and PDA’s.
This device can be purchased as a stand alone DIN Rail mounted unit
(iBTX-D or iBTHX-D), as a bench/wall mount unit (iBTX-W or iBTHX-W), or as a
bench/wall mount unit with an LCD display, Flash Memory Card and Back-up
Battery (iBTX-SD).
The following example illustrates how you can hookup an iServer to your network:
A standard web browser can be used to monitor and chart barometric pressure,
temperature, humidity, and dew point. The browser can also be used to configure the
device’s IP address, passwords for access and overall configuration parameters.
An iLD Big Display can display temperature and barometric pressure received from an
iServer over the Ethernet or the Internet. The following example illustrates how you can
hookup an iServer and iLD to your network:
COMPUTER COMPUTER
(iLOG application (Standard
software) Web Browser)
ETHERNET
RESET
COL!
ON!
DC POWER IN TX!
RX!
!
66.0 [2.60]
Ground Screw
33.0 [1.30]
on Rear of Case
Ethernet
27.3 [1.07]
It is recommended that you ground your unit. With one end of a wire connected
to earth ground, the other side can be wrapped around the ground screw,
located on the bottom of the case.
If unit is to be mounted on a flat surface, you may take the bottom rubber feet off
the unit.
4
2.1.2 Mounting the DIN Rail iServer
To install unit onto DIN Rail:
a) Tilt unit, position mounting slot onto DIN Rail, as shown.
b) Push unit towards DIN Rail and it will snap into place.
To enable the DHCP, besides using DIP switch #3, set the iServer’s IP address
to 0.0.0.0. An iServer with IP address of 0.0.0.0 will request an IP address,
gateway address, and subnet mask from a DHCP server over the Ethernet.
ON
4
SW1 3 SW1
OFF 2
1
1 OFF
2
3
4 ON
SW1
ON
12
3 4 OFF
SW1 4
3
2
1
OFF
ON
6
2.3 Parts of the iServer Unit
iBTHX-D / iBTX-D
Pin# Top Connector
5 Data In (B) Orange
6 CLK (B) Blue
7 SCK (H)* Red/Blk
8 Data (H)* Wht/Blk
iBTHX-W / iBTX-W DB9
Pin# Signal Pin# Signal Pin # Bottom Connector
1 Data Out (B) 6 Data In (B) 1 3.3V Red
2 N/C 7 CLK (B) 2 GND White
3 N/C 8 SCK (H)* 3 N/C
4 Data (H)* 9 3.3 Vdc 4 Data Out (B) Black
5 GND B = Barometric Pressure
* Applies only to iBTHX models *H = Humidity
METAL HOUSING
IS CONNECTED
Choose one which gives the best signal
integrity-
TO SHIELD WIRE
1) Connect Probe’s Shield to RTN if Probe Housing
is not connected to Earth Ground.
OR
2) Connect Probe’s Shield to Earth Ground if Probe
Housing is not connected to Earth Ground.
The iServer can be connected to the network using standard TCP/IP protocols.
It also supports ARP, HTTP (WEB server), DHCP, DNS and Telnet protocols.
MAC (Media Access Control) address is your computer's unique hardware number.
When you're connected to the LAN from your computer, a correspondence table relates
your IP address to your computer's physical (MAC) address. The MAC address can be
found on the label of your device and contains 6 bytes (12 characters) of hexadecimal
numbers XX:XX:XX:XX:XX:XX hex
For example: 0A:0C:3D:0B:0A:0B
Remove the small label with the default IP address and there will be room to put
your IP address. See Figure 3.1
REMOVE DEFAULT
IP ADDRESS LABEL REMOVE DEFAULT IP iSERVER'S VERSION #
AND PUT NEW ADDRESS LABEL AND PUT
CUSTOMER'S NEW CUSTOMER'S
IP ADDRESS IP ADDRESS
9
3.3 DHCP
DHCP, Dynamic Host Configuration Protocol, enables computers and network devices to
receive their IP configurations from a DHCP server.
If DHCP is enabled on your iServer, as soon as the iServer that is connected to the
network is powered on, there will be an exchange of information between the iServer and
the DHCP server. As a result, the DHCP server will assign an IP address, a Gateway
address, a Subnet Mask, and a DNS address to the iServer. Note that the DHCP server
must be correctly configured to make such assignments.
If fixed or static IP address is desired, the DHCP function must be disabled.
The iServer is shipped with DHCP disabled (factory default).
The DHCP can be enabled by setting the DIP switch #3 to ON position (refer to Figure
3.2).
It’s very important to communicate with the network administrator in order to
understand DHCP and its existing configurations on the host server before
enabling DHCP on the iServer.
The iServer is shipped with a default static IP address of 192.168.1.200 and
Subnet Mask of 255.255.255.0.
ON
DIP switch # 3 shown in “ON” position 4
3 OFF
2 4
1 3
2
1
Figure 3.2 DIP Switch on iServer
3.4 DNS
DNS, Domain Name System, enables computers and devices to be recognized over a
network based on a specific name instead of IP addresses.
For example, instead of having to use http://192.168.1.200 (IP address), you would use
http://eis03ec or any name up to sixteen alphanumeric characters defined as a Host
Name in the iServer’s web server.
The default Host Name for an iServer is "eis" followed by the last four digits of the MAC
address of that iServer unit.
On Windows servers where DHCP and DNS are separate functions it is very
important to configure the DHCP server to communicate with DNS in order for
the iServer’s Host Name to correctly respond. If you cannot access the iServer
using its Host Name, please contact your network administrator to make sure
DHCP and DNS servers are linked together
10
3.5 Default IP Address
The iServer is shipped with a default IP address of 192.168.1.200 and Subnet Mask of
255.255.255.0. If you are going to use a Web browser or Telnet program to access the
iServer using its default IP address, make sure that the PC from which you’re
establishing the connection has an IP address that is in the same range as the iServer’s
IP address (192.168.1.x, where x can be any number from 1 to 254.
You also need to make sure that your PC’s Subnet Mask is 255.255.255.0. This is a
good way to access the iServer over the network and make any configuration changes
needed. If 192.168.1.200 is already in use on your network, use an Ethernet crossover
cable between your computer and the iServer to change the IP address or any other
settings within the iServer.
3.6 Port Number
All TCP connections are defined by the IP address and a port number. A port number is
an internal address that provides a TCP/IP interface between an application software on
a computer and a device on the network or between two devices on the network.
There are three default TCP socket port numbers assigned to the iServer:
1. Port 1000 when using HTTPget program (see Section 4.5).
2. Port 2000 when trying to access the sensor (probe) connected to the port of the
iServer to receive data.
3. Port 2002 when trying to access the iServer itself for Power Recycling the iServer
remotely. This can be done using Windows standard Telnet application.
Power recycling the iServer can also be done through the iServer’s Web Server (see
Section 4.2).
Telnet stands for Telecommunications Network, it is a protocol that provides a way for
users (or clients) to connect to computers (or servers) on a network, whether in the next
building or across the world.
Example: C:\>Telnet 192.168.1.200 2002
You will then get the following screen.
C:\ Telnet 192.168.1.200
Type "reset" Firmware Version x.xx
to reboot Admin. Password:00000000
the server Admin. Login Successful
reset
The unit will reset in 5 seconds
The default _
Admin. Password
11
PART 4
OPERATIONS
This iServer can be used and configured in several ways, depending on user’s preference
and network setup. It can be configured using a Web browser, like Netscape or Internet
Explorer. It can also be configured using iConnect Configuration Software.
If DHCP and DNS servers are used, the connection is very simple, no need to find the right
IP address or watch for network conflicts, these are all done for you by your network DHCP
and DNS server. All that is left for you to do, is to enable DHCP on the iServer (see Section
2.2) and use a straight network cable to connect the iServer to a hub and power it up.
If DHCP is not the preferred method, you can configure your PC’s network connection with
an IP address of 192.168.1.x that is in the same range as the iServer’s default IP address
(192.168.1.200) and connect to the iServer using a cross-over network cable between your
PC’s network port and the iServer. After you’re done with configuring the iServer, you can
always set your PC back to its original settings.
On your computer, from the MS-DOS Prompt window type "ping 192.168.1.200” and
press Enter. If DHCP and DNS servers are used type “ping eisxxxx”, where xxxx are the
last four digits of the iServer’s MAC address, located on the back of the device. You should
get a reply as shown in Figure 4.1.
You can use the iServer’s host name (eisxxxx) instead of its IP address only if
your DHCP server is configured to communicate with your DNS. Please consult
with your IT department for details.
4.0 Testing the Connection
C:\>ping eis03ec�
Pinging eis03ec with 32 bytes of data:�
�
Reply from eis03ec: bytes=32 time=15ms TTL=60�
Reply from eis03ec: bytes=32 time=8ms TTL=60�
Reply from eis03ec: bytes=32 time=8ms TTL=60�
Reply from eis03ec: bytes=32 time=8ms TTL=60�
�
Pinging statistics for eis03ec:�
� Packets: Sent=4, Received=4, Lost=0 (0% loss)�
�
Approximate round trip times in milli-seconds:�
� Minimum=8ms, Maximum=15ms, Average=9ms
The iServer may also be assigned an IP Address by using the iConnect software.
a) Download the iConnect software from the website listed in this manual.
b) Install iConnect software on a networked PC. This software is compatible with
Windows 95, 98, NT, 2000, and XP.
c) Use iConnect to assign an IP address to the iServer and access its web pages for
configuration. You can also use any standard web browser to access the iServer’s
web pages. Consult with your IT department for obtaining an IP address.
13
4.1 iConnect Software (continued)
14
4.2 Setting a New IP Address over the Network
Besides using the iConnect software, you may use the iServer’s default IP address to
access it and assign a new IP address to it.
The iServer is shipped with a default IP address of 192.168.1.200 and Subnet Mask of
255.255.255.0. You can configure your PC’s Network connection with an IP address that
is in the same range as the iServer’s IP address (192.168.1.x) and connect to the iServer
using a crossover network cable between your PC and the iServer.
With this completed, you can go to the DOS-Prompt and ping 192.168.1.200. If you
receive responses back (Figure 4.1), you can go to the Web browser and type in
http://192.168.1.200 and it will take you to the iServer’s Home Page.
Select Access Control , button, you’ll be asked for the password. First default LOGIN
password is "12345678" and the ADMINISTRATOR password is "00000000", then you
should be on the Access Control page were you can simply type in the desired Static IP
address, and click Save.
For more details about the “Access Control” page refer to Section 4.3.5.
Access Control
http://192.168.1.200
ACCESS CONTROL
Login Password: 12345678
Admin Password: 00000000
Host Name: eis0e0f
MAC Address: 0A:0B:0C:0D:0E:0F
IP Address: 192.168.1.200
Gateway Address: 0.0.0.0
Subnet Mask: 255.255.255.0
Save Reset
Power Recycle
Main Menu
You can now connect the iServer to an Ethernet hub using a straight through cable,
power it up, and follow the ping routine mentioned in the previous section.
15
4.3 Setup and Operation Using a Web Browser
• Start your web browser.
• From the browser you type http://eisxxxx using the last four-digits from the MAC
address label located on the device if DHCP and DNS are used. If a static IP address
is used, then simply type http://x.x.x.x, where x.x.x.x is the iServer’s IP address.
• The Home Page, shown in Figure 4.5, will be displayed.
iServer Home Page
http://192.168.1.200
In order to access certain menu items of the Home Page, users may be
prompted for a password, as shown in Figure 4.6.
LOGIN ADMINISTRATOR
http://192.168.1.200 http://192.168.1.200
LOGIN ADMINISTRATOR
16
4.3.1 Read Sensor
• Click on Read Sensor . In a few seconds the following page (Figure 4.7) will appear
with all default values of 100.00. Then the actual readings of Temperature, Pressure,
Humidity and Dewpoint will display.
• This page automatically updates the Temperature, Barometric Pressure, Humidity, and
Dew Point.
While accessing the Read Sensor page, If a blank screen appears without any
“java application running” or image of a “Java logo”, please verify you have the
latest Java Runtime Environment installed and configured according to the
following instructions. If you do not have Java Runtime Environment, you may
download it from our website or contact the Customer Service Department
nearest you.
Read Sensor
http://192.168.1.200
Title
If you have
an iBTX:
Humidity and
Dewpoint are
not displayed
Main Menu
17
4.3.1.2 Browser Proxy Selection
Accessing iServer units within your internal network
• Usually when the computer and iServer are on an internal network, you will not use Proxy
server access.
• You should un-check the "Use Browser Settings" option on the "Proxy" tab.
Accessing iServer units using the internet
• Often the web browser will use Proxy server access to the internet. In such cases, the
default Java runtime settings on the "Proxy" tab should suffice. The default setting is the
"Use Browser Settings" option.
• If the default proxy setting doesn't work, then you may have a situation where the proxy
settings of the web browser are incorrect.
Diagnostics:
If the web page of the iServer appears, then the HTTP Proxy is working fine.
If the data isn't updated on the iServer upon selecting the Read Sensor web page, there may be
a problem with access through a winsock proxy server. In such cases your network
administrator will need to provide the winsock proxy server and port #s. (If the administrator
requires knowledge of the port # required on the iServer, the value is 2003).
These values should be entered into the Socks line on the "Proxy" tab (of the Java Plugin
control panel) or into the "connections" tab on the View,Internet Options dialog and make sure
that the Proxy tab shows that the "Use Browser Settings" option is not selected (i.e. when you
specify proxy connections in the Java Plugin control panel.
Accessing iServer units over Peer-to-Peer network
A simple peer-to-peer network is setup by disconnecting from the main network (as users will
often do when trying to do initial setup of the iServer) and then connecting the iServer to another
computer using a ethernet hub, an ethernet switch, or a Cross-over cable connection.
Often when using a peer-to-peer network, the Java plugin and the web browser (such as
Internet Explorer) have internet connections configured to connect through proxy servers. In
such case, you will need to simply assign the final IP address on this peer to peer network and
then view the iServer charts after connecting the iServer into the regular network. Otherwise you
can disable the Java plug-in's "Use Browser Settings" temporarily and then reconfigure the Java
plug-in settings for regular network access after testing the iServer chart access on your peer-
to-peer network.
The "Use Browser Settings" should not be selected. And the HTTP and Socks proxy entries
should be blank. Also, setup the internet browser so that proxy servers are disabled.
Java and the Java Coffee Cup Logo are trademarks or registered trademarks of Sun Microsystems, Inc. in the U.S. and other countries."
18
4.3.2 Chart
• Click on Chart , the following page (Figure 4.8) should appear. The Java™ Applet graph
displays Temperature, Pressure, and Humidity values which can be charted across the full
span (-40 to 85°C and 0 to 100% RH) or within any narrow range (such as 20 to 30ºC).
The time-base can display one minute, one hour, one day, one week, one month or one
year.
Title
Actual Temperature http://192.168.1.200 Actual Pressure
TITLE
!
Temp. 0015.7 Dew. 0005.7 C Temp_RH_hPa RH 0040.9 Pres. 1020.8
User selectable User selectable
50 1200
temperature range Pressure range
100%
Number of degrees
per division hPa Humidity Ramp
5/Div 90/Div
based on from 0 to 100%
temperature range
0%
User selectable
0 300
temperature range User selectable
Wed Jun 01 12:00:00 PDT 2005 1 Month (1 Day/Div) Tue Jun 28 12:00:00 PDT 2005
1 Minute Pressure range
1 Hour
1 Day
1 Week Main Menu
1 Month
Start Time 1 Year End Time
If you have an iBTX model, the probe will only sense Temperature and Pressure;
to indicate this, the Humidity and Dewpoint boxes on the chart will switch
between OPEN and a false number.
19
4.3.3 Configuration
CONFIGURATION
http://192.168.1.200
CONFIGURATION
No. Device Name Reading Display Display Remote Remote End Char Offset
Units Format Format (HEX)
1 Temperature SRTF F decimal T00.0F 0D 0000.0
2 Pressure SRHb mbar decimal P0000.0B 0D 0000.0
3 Humidity SRH2 % decimal H00.0% 0D 0000.0
4 Dewpoint SRDF2 F decimal D00.0F 0D 0000.0
Click on Device No. on the left to modify Sensor Parameters.
A
Secured Applet
Title Title______
Terminal Server
B
TCP/UDP TCP Server Type Command Forward CR disable Number of Connections 5 Port 02000
Take Readings
Main Menu
A) Sensor/Device No.
Clicking on the No. 1, 2, 3, or 4 allows you to modify the Sensor Parameters (See
Section 4.3.4 for more details).
Secured Applet: If checked, the LOGIN password is required to open “Read Sensor”
and “Chart” pages.
Title: editable field which will be the title on the “Read Sensor” and “Chart” page.
20
4.3.3 Configuration (continued)
B) Terminal Server
TCP/UDP*: The iServer supports TCP and UDP protocols (default is TCP). If UDP is
selected, it can be configured either for Broadcast UDP or Directed UDP. In case of
Broadcast UDP, the iServer will transmit the data to every node on the network. This
can be accomplished if the Remote IP Address is set to 255.255.255.255.
The Broadcast UDP is a practical solution when one iServer needs to communicate
with multiple nodes over the network. In the case of directed UDP, the iServer will
transmit the data to a specific node on the network. This can be accomplished if the
Remote IP Address is set to the IP address of that specific node.
Server Type: Continuous mode sends the temperature, pressure and humidity to the
Ethernet, every two seconds. It is mainly used to send readings to a remote display or
logger.
Command (default) mode needs a command to query the iServer to send the
response back to querying device.
Number of Connections: The range is from 0 to 5. If 0 is selected, the Terminal
Server feature is disabled. This means that no network connection can be made to
the sensor connected to the iServer. If 1 is selected, only one network connection
can be made to the iServer’s sensor. Any number higher than 1 would allow that
number of network hosts to read from the iServer simultaneously (default is 5).
Port: (default 2000) is the default TCP port number for the port to which the sensor is
connected. Ports 1000 (used for HTTPget, refer to Section 4.5), 2002, 2003, and 2004
are reserved for internal use.
Terminal Server usually describes a device that exchanges data between
Ethernet/TCPIP networks and RS-232/RS-485 systems. With this iServer,
the data is obtained digitally from the sensor (irrelevant to RS-232 or RS-485
interface) and can be accessed from anywhere on the network.
A computer program, such as Mail Notifier, OPC Server, iLog or HTTPget can
send TCP requests and obtain readings using the Terminal Server feature.
C) Remote Access
Remote IP Address: iServer can establish a connection to a remote device (e.g. an
iLD Remote Display with an Ethernet iServer embedded board).
Remote Port: (default 2000) the port number for the remote device to which the data
is sent (e.g. an iLD Display).
Remote Access:** Remote Access can be enabled and disabled. If enabled, the
iServer can send its data to a remote node on the same network (the “Remote IP
address” and “Remote Port” must be entered).
*TCP/UDP: when UDP mode is selected, Remote Access should be disabled
and Remote IP and Port are the UDP remote listening IP and Port. If the Remote
IP is set to 255.255.255.255, the UDP packet becomes a broadcasting packet
which will allow any device listening to the Remote port to receive the packet.
**If Remote Access is enabled, Terminal Server is automatically disabled.
21
4.3.4 Sensor Parameter
• In the first column of Configuration’s page, click on 1 to view and modify sensor
parameters. See Figure 4.9.
Sensor Parameters
http://192.168.1.200
SENSOR PARAMETERS
Device No. 1
Device Name: Temperature
Reading Command: SRTF
Display Units: F
Display Format: decimal
Remote Display Format: T00.0F
Remote End Char (HEX) 0X: 0D
Offset: 0000.0 ( C)
Update Reset
Cancel
Main Menu
You may type any ASCII characters in the first three boxes, but you must delete
the leading spaces.
Figure 4.10 Sensor Parameters
Below are some definitions of terms used in the Sensor Parameter page.
Device Name: shows on the Configuration and Read Sensor page.
Reading Command: Can be set as the following command.
SRTC Read the temperature in °C.
SRTF Read the temperature in °F.
SRHb Read the pressure in mbar / hPa.
SRHi Read the pressure in inHg (Mercury).
SRHm Read the pressure in mmHg (Mercury).
SRH2 Read the humidity.
SRDC2 Read the dewpoint in °C.
SRDF2 Read the dewpoint in °F.
Display Units: unit of Temperature in °F or °C, shows on the “Configuration” and “Read
Sensor “ pages.
Display Format:
Decimal - the leading zero’s are eliminated.
Raw - the leading zero’s remain.
22
4.3.4 Sensor Parameters (continued)
Remote Display Format: Remote display format is used for Terminal Server continuous
mode. This determines the data format sent by the iServer to a remote network node (e.g.
iLD Display)
H37.9% in humidity setting displays H, and 37.9% is the humidity value displayed.
Example: if the humidity is 37.9, then H37.9% will be seen on the remote display. If no
format is specified (blank), there is no reading sent out. If temperature is 75.7 and
T00.0F is used in temperature setting, the remote site will show T75.7F.
This format setup was originally made for the iLD Display, which has four or six LEDs.
For six LEDs, T00.00F format, and for four LEDs, 00.0F format are appropriate.
Remote End Char: The default value is 0D (Hex representation of <CR>). This means
that the iServer sends <CR> after each temperature, humidity, pressure, and dewpoint
value. This will be done either in Continuous or Command mode.
This is how the data will appear on the host with Tera Term - 206.29.25.27 VT
0D assigned: File Edit Setup Control Window Help
T75.7F T75.7F
P1014.8mbar P1014.8mbar
H37.9% H37.9%
D44.9F
D44.9F T75.7F
P1014.9mbar
H37.9%
If the end character for instance is 20 (Hex D44.9F
representation of space), the data will then appear T75.7F
as: P1014.9mbar
T75.7F P1014.8mbar H37.9% D44.9F H38.0%
D44.9F
Offset: Since the sensing probe is solid state electronics, there is no need for
calibration. If it’s determined that the readings are slightly off, the user can manually
assign numerical values to adjust the readings for temperature, humidity, pressure, and
dewpoint. For temperature offset, the unit must be in degree C. The Offset value can
either be a positive or negative number
23
4.3.5 Configure Access Control (continued)
Access Control
http://192.168.1.200
ACCESS CONTROL
Login Password: 12345678
Admin Password: 00000000
Host Name: eis0e0f
MAC Address: 0A:0B:0C:0D:0E:0F
IP Address: 192.168.1.200
Gateway Address: 0.0.0.0
Subnet Mask: 255.255.255.0
Save Reset
Power Recycle
Main Menu
24
4.3.5 Configure Access Control (continued)
Gateway Address: A gateway is a network point that acts as an entrance to another
network. A gateway is often associated with a router, which knows where to direct a
given packet of data that arrives at the gateway. If the iServer is sending packets to
another network node that is not on the same network on which the iServer is connected,
a gateway address needs to be given to the iServer. The gateway address should be the
IP address of the router connected to the same LAN to which the iServer is connected.
The iServer’s default gateway address is 0.0.0.0. Consult with your IT department for
obtaining a gateway address.
Subnet Mask: It’s a 32-bit number that is used to determine which part of the IP address
is the network portion and which part is the host portion. The iServer’s default subnet
mask is 255.255.255.0. Consult with your IT department for obtaining a subnet mask.
Changes made in the iServer’s Access Control page can be saved permanently
by pressing the Save button and power recycling the iServer (press Power
Recycle button). Pressing the Reset button will set all the fields back to their
default values.
26
4.5.2 HTTPget and ARP to setup Device IP Address
Use the iConnect software, which may be downloaded from our website, to do
these IP changes whenever possible.
Use ARP first to assign the mac address to a static IP address in computer arp table by
this command:
where:
“0000000” is admin. password. If the password is wrong, the unit will ignore the new IP.
If the new IP is taken, you will get the message " New IP is Assigned" after the HTTPget
command. The device will reset automatically. (Diagnostics LED is on for 2 second).
ARP is the Internet layer protocol responsible for matching or obtaining the MAC
(hardware) address that corresponds to a particular IP address. The ARP command
allows the user to view the current contents of the ARP cache of the local computer
(residing on the same network). Microsoft includes the ARP.EXE utility for viewing and
modifying the ARP cache with its Windows products. The following ARP commands can
be used to view cache entries:
Ping the destination computer using IP address first before using the arp -a
command.
27
4.6 ARP Protocol (continued)
28
4.7 Remote Access (Tunneling)
To "tunnel", in this context, is to transmit data between two points through a private
conduit on a shared or public network. The network could be an Ethernet LAN, a WAN,
or the Internet. There is a Serial-to-Ethernet iServer that allows for a connection between
a serial device and a PC, or between two serial devices, using an existing network rather
than dedicated wiring.
The connected serial devices to iServer’s can communicate with each other back and
forth over the networks. This characteristic is called Tunneling and it’s illustrated below.
Serial-to-Ethernet
Remote (Local iServer)
iServer
Application
Software
RS-232
COL!
ON!
TX!
RX!
!
32
RS-2
Remote Serial-to-Ethernet
iServer (Local iServer)
Barometric
Pressure
In order to use this Tunneling feature, some settings are required within the local and
remote iServer’s.
29
4.7.1 Local iServer
1. An IP address should be assigned to the iServer dynamically or statically (recommended).
2. Use a browser to access the Local iServer’s WEB page. Simply type the iServer’s IP
address at the browser’s URL location (i.e. 192.168.1.49) followed by an Enter key.
You should then see the iServer’s main WEB page.
3. Click on the Update button.
4. Click on Configuration, you will be prompted with a Password (default is 12345678).
5. On the Configuration page, under Serial Communication section, make sure the
parameters such as Baud Rate, Data Bits, Parity, Stop Bits, Flow Control, etc. match
with your attached serial device and its application software.
6. Make sure to set the End Character (Hex) to 00 and the Timeout to 0.
7. Under Terminal Server section, set Number of Connections to 0.
8. Under Remote Access section, set the Remote Access to enable, Enter the Remote
IP address (would be the IP address of the remote iServer, 192.168.1.50), and use the
default Remote Port number 2000.
9. Set Connection Control to Reconnect and set the Connection Timeout to a desired value.
The Reconnect option is used in Serial Tunneling and it applies only to the Local
iServer. If the tunneling connection between the two iServers goes down due to
network problems, power failure, etc., the Reconnect option will enable the Local
iServer to reconnect with the Remote iServer based on the specified time interval in
the Connection Timeout. For example, based on a timeout of 1000 x 10 ms (10
seconds), the Local iServer will continually attempt to reconnect and re-establish the
tunnel with the Remote iServer every 10 seconds.
10. Click on Save button for the changes to take place.
11. Initialize the serial device application software to establish the connection.
CONFIGURATION
Address http://192.168.1.49
CONFIGURATION
Serial Communication
Baud Rate 9600 Data Bit 8 Bits Parity none Stop Bits 1 bit
Terminal Server
TCP/UDP TCP Server Type slave Number of Connections 0 Local Port 02000
Save Reset
Main Menu
At this point, reset the power, first on the remote and then the local iServer and initialize
the local serial device to send or request data.
CONFIGURATION
http://192.168.1.50
CONFIGURATION
No. Device Name Reading Display Display Remote Remote End Char Offset
Units Format Format (HEX)
1 Temperature SRTF F decimal T00.0F 0D 0000.0
2 Pressure SRHb mbar decimal P0000.0B 0D 0000.0
3 Humidity SRH2 % decimal H00.0% 0D 0000.0
4 Dewpoint SRDF2 F decimal D00.0F 0D 0000.0
Click on Device No. on the left to modify Sensor Parameters.
Secured Applet
Title Title______
Terminal Server
TCP/UDP TCP Server Type Continuous Forward CR disable Number of Connections 5 Port 02000
Update
Take Readings
Main Menu
31
4.8 iLog Software
This is an Excel application software that can log temperature, humidity and dewpoint
from the iServer over the local network (Ethernet) or the internet.
a) Download the iLog software from the website listed in this manual.
b) Install iLog software on a networked PC. This software is compatible with Windows
95, 98, NT, 2000, XP, Windows Vista and Windows 7 (32 and 64-bit).
c) If you have Excel 2007 or higher, when installing iLog choose “Custom” installation
option and on the next window check the box for “Excel 2007 Apps” and continue
the installation to the end.
d) For complete information of how to use the iLog software, click on the HELP button.
e) There is a list of Error Messages in Appendix E.
32
4.8 iLog Software (continued)
Table 4.1 iLog Excel Applications
The iLog application actually consists of several Excel files, though most supported
devices can be accessed by the main iLog program.
The main program is listed as "iLog", plus a version number, under the Start Menu
program links (those links available by clicking the Start button on the Windows
taskbar).
In the following table, the iLog File column shows how many sensors are logged by the
main iLog file. If more than 3 sensors are available, then the Alternate column shows
how many sensors the device can support. Devices with more than 3 sensors will have
their own Excel file. For these files, click on the appropriate model, found in the Start
Menu\Programs\iLog.
Wireless Devices
The "Auto" column shows number of columns per remote device that are possibly
displayed.
The Full column shows number of columns allocated per device for the "Full"
spreadsheet, which will be able to display all data for all active devices.
The active wireless devices, when shown in the Excel application, will be shown with
the device number and the units returned.
33
4.9 Mail Notifier Software
The Mail Notifier Software can be used only with NEWPORT Electronics instruments.
For complete information of how to use the Mail Notifier software, click on the Help menu
of the main window.
The Mail Notifier software generates email notifications for alarm conditions. Users can
be notified automatically of alarm conditions monitored via internet connections
throughout the world. By use of the email forwarding of alarm conditions, alarm
conditions can be monitored on a network isolated from the internet and forwarded to
connections on the Internet.
The Mail Notifier utility operates under Windows 98, NT 4.0, 2000, and XP in conjunction
with existing email that supports the MAPI messaging interface. If MS Outlook has been
loaded, the MAPI support should be available.
4.9.1 Installation
The Mail Notifier must be loaded on a computer running Microsoft Windows (versions
specified earlier) using an email program that provides MAPI access. Network access
must be available between this computer and the iServer. Network access must also be
available from this computer to the appropriate email server and from the email server to
the recipient’s email server.
Alerts List
Dev1 alarm: 74.6
Dev2 alarm: 89
Dev1 alarm: 74.7
Dev1 alarm: 74.7
Dev2 alarm: 89
Dev1 alarm: 74.7
Dev1 alarm: 74.6
34
4.9.2 Program Options Setup and Configuration
Complete program setup requires: Options
• Entering a recipient for Send To Email Setup Content Startup General
the email
• Specifying connection Mail Server
Help OK Cancel
Alarm Editor
Device Info (1 of 2)
Server IP Address 192.168.1.200
OK
Cancel
Socket Number 2000
1000
Src ID Dev12
Dev Add Del
36
PART 5 SPECIFICATIONS
SENSOR SPECIFICATIONS
BAROMETRIC PRESSURE (iBTX, iBTHX)
Accuracy/Range at 25ºC: ±3.5 mbar / From 10 to 1100 mbar (1 to 110 KPa)
Resolution: 0.1 mbar
RELATIVE HUMIDITY (iBTHX)
Accuracy/Range at 25ºC ±2.75% for 10 to 90%
- Non-Condesning: ±3% for 5 to 10% and 90 to 95%
±4% for 0 to 5% and 95 to 100%
Hysteresis: ±1% RH
Non-linearity: ±3%
Response Time: 8 seconds, tau 63%
Repeatability: ±0.1%
Resolution: 0.1%, 12 bit
TEMPERATURE (iBTHX)
Accuracy/Range*:
Wand Probe: ±0.5°C to 45°C (±1°F for 41 to 113°F)
up to ±1°C for 0 to 5°C and 45 to 70°C;
(up to ±2°F for 32 to 41°F and 113 to 158°F)
Industrial Probe: ±0.5°C for 5 to 45°C (±1°F for 41 to 113°F)
up to ±1.5°C for -40 to 5°C and 45 to 85°C
(up to ±2.7°F for -40 to 41°F and 113 to 185°F)
*Note: extended temperature range is for Industrial Probe
only, the iServer’s operating temperature is 0 to 70°C.
Response Time: 5 to 30 seconds, tau 63%
Repeatability: ±0.1°C
Resolution: 0.1°C, 14bit
TEMPERATURE (iBTX)
Accuracy/Range*:
Wand Probe: ±0.8°C @ 20°C (±1.5°F @ 68°F)
±2°C for 0 to 70°C
(±3.6°F for 32 to 158°F)
Industrial Probe: ±0.8°C @ 20°C (±1.5°F @ 68°F)
±2°C for -40 to 85°C
(±3.6°F for -40 to 185°F)
*Note: extended temperature range is for Industrial Probe
only, the iServer's operating temp. is 0 to 70 °C
PROBE SPECIFICATIONS
Wand Probe: 159 mm lg x 19 mm dia (6.25” x 0.75")
Cable with DB9 connector: 152 mm long (6")
Cable operating temperature: 0 to 80°C (32 to 176°F)
Industrial Probe: 137mm lg x 16mm dia (5” x 0.63")
Housing Material: SS 316
Cable with DB9 or stripped leads: 3m long (10’)
Cable operating temperature: -55 to 105°C (-67 to 221°F)
37
iSERVER SPECIFICATIONS
Interfaces
Ethernet: 10Base-T (RJ45)
Sensor: Digital 8-wire
Supported Protocols: TCP/IP, UDP/IP, ARP, ICMP, DHCP, DNS, HTTP, and Telnet
Indicators (LED's): Network Activity, Network Link, and Diagnostics
Management: Device configuration and monitoring through embedded
WEB server
Embedded WEB Server: Serves WEB pages containing real-time data and live
updated charts within definable time intervals.
POWER
Power Input: 9 to 12 Vdc iBTHX/iBTX-W
10 to 32 Vdc iBTHX/iBTX-D
Safety Qualified (included)
ac power adapter: Nominal Output: 9 Vdc @ 0.5A;
Input: 100 to 240Vac, 50/60 Hz included for iBTHX/iBTX-W
Switching Power Supply: Sold separately for iBTHX/iBTX-D
Consumption: 2.5 W max.
ENVIRONMENTAL
Operating Temp: 0 to 70°C (32 to 158°F)
Storage Temp: -40 to 125°C (-40 to 257°F)
Material: Valox 364 PBT case with wall mount bracket: iBTHX/iBTX-W.
Polycarbonate case with DIN Rail mount: iBTHX/iBTX-D
Weight: 0.180 kg (0.4 lbs) iBTHX/iBTX-W
0.113 kg (0.25 lbs) iBTHX/iBTX-D
GENERAL
Agency Approvals: FCC-B, CE (see CE Approvals page)
Software: Compatible with Windows operating systems.
Field firmware upgradeable.
OPC Server
iConnect: Configuration software for the Ethernet interface
iLog: Macro for automatic data logging in MS Excel
Mail Notifier: Software that generates email notifications
for alarm conditions
38
PACKAGING
2.60 [66.0] 3.55 [90.2]
3.05 [77.5]
2.42 [61.6]
0.25 [6.4]
3.56
3.67 [90.3]
[93.1]
COMMUNICATIONS
ETHERNET
RESET
3.54
[90.2]
DC POWER IN
N/C
0.99
4.53 [115]
[25.1]
Material: Polycarbonate case with DIN rail mount
Figure 5.2 DIN Rail iServer Dimensions
39
PART 6
FACTORY PRESET VALUES
40
Appendix A GLOSSARY
User of this manual should be familiar with following definitions:
ARP (Address Resolution Protocol) is a protocol for mapping an Internet Protocol address (IP
address) to a physical machine address that is recognized in the local network. For example,
the IP address in use today is an address that is 32-bits long. In an Ethernet local area network,
however, addresses for attached devices are 48-bits long. (The physical machine address is
also known as a Media Access Control or MAC address.) A table, usually called the ARP cache,
is used to maintain a correlation between each MAC address and its corresponding IP address.
ARP provides the protocol rules for making this correlation and providing address conversion in
both directions.
Ethernet is a network protocol defined by the IEEE 802.3 standard. Ethernet-based
networks use MAC Address rather then IP Address to exchange data between computers.
By using ARP and adding TCP/IP support, Ethernet devices may be connected as part of the
Internet. An Ethernet LAN typically uses coaxial cable or special grades of twisted pair wires.
The most commonly installed Ethernet systems are called 10BASE-T and provide transmission
speeds up to 10 Mbps. Devices are connected to the cable and compete for access using a
Carrier Sense Multiple Access with Collision Detection (CSMA/CD) protocol.
IP (Internet Protocol) is the method or protocol by which data is sent from one computer to
another on the Internet.
IP address (Internet Protocol address) is a 32-bit number that identifies each sender or
receiver of information that is sent in packets across the Internet.
IP Netmask is a 32-bit pattern of bits used to determine which part of the IP address is the
network portion and which part is the host portion.
MAC (Media Access Control) Address is your computer's unique hardware number. When
you're connected to the Internet from your computer, a correspondence table relates your IP
address to your computer's physical (MAC) address on the LAN.
Ping is a utility that tests the network connectivity. It is used to determine if the host is capable
of exchanging information with another host.
Port number/Socket number is a way to identify a specific process to which an Internet or
other network message is to be forwarded when it arrives at a server. It is a predefined address
that serves as a route from the application to the Transport layer or from the Transport layer to
the application of the TCP/IP system.
Sockets are a method for communication between a client program and a server program in a
network and defined as "the endpoint in a connection." Information transferred across the
Internet primarily occurs between sockets.
SMTP Simple Mail Transfer Protocol is an Internet standard for electronic mail (email) transfer
across the Internet. SMTP clients usually use SMTP to send email messages by specifying the
SMTP server. The email server uses SMTP to both send and receive email messages.
SNMP Simple Network Management Protocol is a network monitoring protocol to monitor
devices connected to an Ethernet Network.
TCP/IP (Transmission Control Protocol/Internet Protocol) is the basic communication
language or protocol of the Internet. When you are set up with direct access to the Internet, your
computer is provided with a copy of the TCP/IP program just as every other computer that you
may send messages to or get information from also has a copy of TCP/IP. TCP/IP often is used
as a general term to indicate generic access to the Internet.
UDP/IP (User Datagram Protocol/Internet Protocol) is the TCP/IP standard protocol that
allows an application program on one machine to send a datagram to an application program on
another. The UDP can be either in Broadcast or Directed form. The Broadcast UDP transmits
data to every node on the same network. The Directed UDP transmits data to one node only.
41
Appendix B IP Address
All IP addresses have been divided into three smaller groups (classes) A, B and C
• Class A addresses have 8-bits of network ID and 24-bits of host ID. They can support
a large number of hosts, approximately 2 = 16,777,216 computers per network.
• Class B addresses have 16-bits of network ID and 16-bits of host ID. They can
support approximately 216 = 65,536 computers per network.
• Class C addresses have 24-bits of network ID and 8-bits of host ID. They can support
approximately 28 = 256 computers per network.
The rest of the addresses are divided into two classes, D and E.
Class D networks are not assigned to the host. They are used for multicasting.
The address range from 224.x.x.x to 239.x.x.x
42
Appendix C IP Netmask
IP Netmask or Subnet Mask is a 32-bit pattern of ones and zeros used to determine
network portion of an IP address from the host portion of the IP address. Subnet mask is
a network ID that is created by borrowing bits from host portion of IP address and using
them as part of a network ID. The table below shows a default subnet mask for address
Classes A, B, and C. Each bit that is set to "1" in the subnet mask corresponds to the bit
in the IP address that is to be used as the network ID. Each bit that is set to "0" in the
subnet mask corresponds to a bit in the IP address that is to be used as the host ID.
43
Appendix D ASCII Chart
46
Appendix F Sensor Information
F.1 Accuracy
Accuracies are tested at Manufacture’s Outgoing Quality Control at 25°C (77°F) and 3.3V.
Values exclude hysteresis and non-linearity, and is only applicable to noncondensing
environments.
F.2 Operating Conditions
Sensor works stable within
recommended normal range – see
Figure. Long term exposures to
conditions outside normal range,
especially at humidity >80%RH,
may temporarily offset the RH
signal (+3 %RH after 60h). After
return to normal range it will slowly
return towards calibration state by
itself. See Section F.4
“Reconditioning Procedure” to
accelerate eliminating the offset.
Prolonged exposure to extreme
conditions may accelerate ageing.
Figure F.3 Normal Range
47
Appendix F Sensor Information (continued)
F.6 Light
The sensor is not light sensitive. Prolonged direct exposure to sunshine or strong UV
radiation may age the housing.
48
PART 7
APPROVALS INFORMATION
7.1 CE APPROVAL
This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Basic Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min
• Input Power to Ethernet Output: 1500Vac
• Input Power to Sensor Metal Body: 1500Vac
• Ethernet to Sensor Metal Body: 1500Vac
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains
Supply (power). Unit measures Barometric Pressure, Air Temperature and Humidity.
Transients Overvoltage Surge (1.2/50uS Pulse)
• Input Power: 500V Transients Overvoltage
• Sensor : 500V Transients Overvoltage
• Ethernet: 1500V Transients Overvoltage
Note: The ac power adaptor must have Safety Qualified Agency Approvals for CE
with Double Insulation rating.
The power input rating is 10-32Vdc.
The minimum output current rating is 500mA.
EMC EN61000-6-1:2001 (Immunity) and EN61000-6-3:2001 (Emissions)
Immunity requirements for residential, commercial and light-industrial environments
• EMC Emissions Table 1, Class B
• EMC Immunity Table 1: Enclosure
Table 2: Signal Lines Ports
Table 3: Dc input/Dc output Ports
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control
and laboratory.
• EMC Emissions Table 4, Class B of EN61326
• EMC Immunity Table 1 of EN61326
Note: I/O lines / sensor cables require shielded cables and these cables
must be located on conductive cable trays or in conduits.
Refer to the EMC and Safety installation considerations (Guidelines) of this manual
for additional information.
7.2 FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules.
49
NOTES:
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
NOTES:
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
NOTES:
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal one (1) year product warranty to cover handling and shipping time. This
ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for
any damages that result from the use of its products in accordance with information provided
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the FOR NON-WARRANTY REPAIRS, consult
following information available BEFORE contacting OMEGA for current repair charges. Have
OMEGA: the following information available BEFORE
1. Purchase Order number under which the product contacting OMEGA:
was PURCHASED, 1. Purchase Order number to cover the COST
2. Model and serial number of the product under of the repair,
warranty, and 2. Model and serial number of the product, and
3. Repair instructions and/or specific problems 3. Repair instructions and/or specific problems
relative to the product. relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords
our customers the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC.
© Copyright 2019 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the
prior written consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
M
U Wire: Thermocouple, RTD & Thermistor
M
U Calibrators & Ice Point References
M
U Recorders, Controllers & Process Monitors
M
U Infrared Pyrometers
M
FLOW/LEVEL
U Rotameters, Gas Mass Flowmeters & Flow Computers
M
U Air Velocity Indicators
M
U Turbine/Paddlewheel Systems
M
U Totalizers & Batch Controllers
M
pH/CONDUCTIVITY
U pH Electrodes, Testers & Accessories
M
U Benchtop/Laboratory Meters
M
U Controllers, Calibrators, Simulators & Pumps
M
U Industrial pH & Conductivity Equipment
M
DATA ACQUISITION
U Communications-Based Acquisition Systems
M
U Data Logging Systems
M
U Wireless Sensors, Transmitters, & Receivers
M
U Signal Conditioners
M
U Data Acquisition Software
M
HEATERS
U Heating Cable
M
U Cartridge & Strip Heaters
M
U Immersion & Band Heaters
M
U Flexible Heaters
M
U Laboratory Heaters
M
ENVIRONMENTAL
MONITORING AND CONTROL
U Metering & Control Instrumentation
M
U Refractometers
M
U Pumps & Tubing
M
U Air, Soil & Water Monitors
M
U Industrial Water & Wastewater Treatment
M
U pH, Conductivity & Dissolved Oxygen Instruments
M
M4221/0319