Professional Documents
Culture Documents
Pulmovista Ifu 9511001 en
Pulmovista Ifu 9511001 en
PulmoVista 500
Typographical conventions
1 Consecutive numbers indicate steps of action, Any text shown on the screen and any labeling on
with the numbering restarting with "1" for each the device are printed in bold and italics, e.g., ROI 1
new sequence of actions. or Ref.
Bullet points indicate individual actions or differ- The "greater than" symbol > indicates the
ent options for action. navigation path in a dialog window, for example
System setup > System > Country. In this
– Dashes indicate the listing of data, options, or example, System setup represents the dialog
objects. window title, System represents a horizontally
(A) Letters in parentheses refer to elements in the aligned tab, and Country a vertically aligned tab.
related illustration.
A Letters in illustrations denote elements referred
to in the text. Screen reproduction
Trademarks
PulmoVista® Mikrobac®
WARNING
A WARNING statement provides important in-
formation about a potentially hazardous situa-
tion which, if not avoided, could result in
death or serious injury.
CAUTION
A CAUTION statement provides important infor-
mation about a potentially hazardous situation
which, if not avoided, may result in minor or mod-
erate injury to the user or patient or in damage to
the medical device or other property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
Contents
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Preventive maintenance. . . . . . . . . . . . . . . . . . 130
Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Disposal of the medical device. . . . . . . . . . . . . 132
Disposal of batteries. . . . . . . . . . . . . . . . . . . . . 132
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Connected devices
WARNING
Risk of electric shock and of device malfunc-
tion
Any connected devices or device combina-
tions not complying with the requirements
mentioned in these instructions for use can
compromise the correct functioning of the
medical device and lead to an electric shock.
Before operating the medical device, strictly
comply with the instructions for use of all con-
nected devices or device combinations.
Patient safety
WARNING WARNING
To properly use this medical device, the user Risk of electric shock and of device malfunc-
must obtain a full understanding of the perfor- tion
mance characteristics of this medical device
Do not use the medical device in environ-
prior to use by carefully reading these Instruc-
ments that are subject to conductive pollu-
tions for Use.
tion, e.g., metal or graphite particles or dust,
or continuous conductivity, e.g., wet condi-
WARNING tions.
Risk of incorrect use
This medical device must only be used by WARNING
physicians, respiratory therapists, and nurs- Risk of fire and of explosion
es, who have received training on the use of
PulmoVista 500 is not intended to be used in
this medical device.
areas with explosion hazards and in the pres-
ence of flammable anesthetic agents or other
WARNING flammable agents.
Risk of wrong therapeutic decisions
Do not use this medical device as the sole ba- WARNING
sis for diagnosis or therapeutic decisions. Risk of fire
Do not use the medical device in an oxygen-
WARNING enriched environment.
Risk of personal injury
Modifications to the medical device may WARNING
cause malfunctioning. Risk of fire
Do not modify this medical device without au- Do not position the medical device close to an
thorization by Dräger. oxygen outlet which opens to the ambient at-
mosphere. Keep the medical device at least
WARNING 1 m (3.3 ft) in front of the O2 outlet and 20 cm
(7.9 in) to the side of the O2 outlet.
Risk of ineffective defibrillation
Remove all parts of the patient interface from WARNING
the patient before cardiac defibrillation.
Risk of device malfunction
WARNING WARNING
Risk due to penetrating liquid Risk of device malfunction
Penetrating liquid may cause the following: If the ventilation slots or heat exchange sur-
– Damage to the device faces of the medical device are covered, the
– Electric shock air flow is blocked and the medical device may
– Device malfunctions overheat.
Ensure that no liquid penetrates the device. Do not cover the ventilation slots and heat ex-
change surfaces of the medical device.
WARNING
Risk of electric shock and of device malfunc- WARNING
tion Risk of electric shock
Penetrating liquid may cause malfunction of Do not open the housing of the medical de-
or damage to the device, which may endanger vice.
the patient.
Do not immerse or rinse the EIT system or its WARNING
components (e.g., patient interface, Cockpit, Risk of electric shock and of device malfunc-
cables, USB storage media) in liquids. During tion
surface disinfection make sure that no liquids Only connect passive storage media without
enter the device. their own power supply to the USB port.
If a liquid is accidentally spilled on the device,
disconnect the device from the power source. WARNING
Contact service personnel regarding the con- Risk of electric shock and of device malfunc-
tinued safety of the device before putting it tion
back into operation.
Do not use the medical device in areas outside
the specification for electrical supply as stat-
WARNING
ed in the section "Operating data"
Risk of device malfunction on page 135.
Do not expose the medical device to direct ra-
dioactive radiation. WARNING
Risk of device malfunction
WARNING
Do not use the medical device in the presence
Risk of device malfunction of strong magnetic fields, e.g., MRI.
Do not directly expose any part of the medical
device, except the patient interface, to X-rays. WARNING
Risk of skin burns
Disconnect the electrode belt from the device
or from the patient during electricity-based
therapies such as electrosurgery or electro-
cautery.
CAUTION
Risk of damage to the touchscreen
Do not allow sharp instruments to touch the touch-
screen of the Medical Cockpit (Infinity C500).
CAUTION
Risk of overheating
Keep the medical device away from sources of
heat such as direct sunlight or radiators. Do not
cover the device with bedclothes.
CAUTION
Risk of wrong therapeutic decisions
Check if ECG, EMG, EOG or EEG are influenced
by EIT measurements.
CAUTION
Risk due to missing anti-virus software
PulmoVista 500 does not include any anti-virus
software and thus relies on the anti-virus mea-
sures of the hospital.
Application
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Contraindications . . . . . . . . . . . . . . . . . . . . . . 17
Environments of use . . . . . . . . . . . . . . . . . . . 18
Option ADAP. . . . . . . . . . . . . . . . . . . . . . . . . . 18
Intended use
Indications
Contraindications
WARNING
Risk of pacemaker impairment
Do not use PulmoVista 500 on patients with
pacemaker.
WARNING
Risk of defibrillator impairment
Do not use PulmoVista 500 on patients with an
implantable cardioverter-defibrillator.
WARNING
Risk of excessive skin damage
Do not use PulmoVista 500 on patients whose
skin is damaged in the region where the pa-
tient interface needs to be placed.
WARNING
Do not use PulmoVista 500 on patients where
the attachment of the patient interface could
pose a risk to the patient, e.g., patients with
spinal lesions or fractures.
WARNING
Information on the safe and effective use of
the device on pregnant women is not yet avail-
able.
Do not use PulmoVista 500 on these patients.
WARNING
Information on the safe and effective use of
the device on premature babies is not yet
available.
Do not use PulmoVista 500 on these patients.
NOTE
Measurement accuracy for pediatric patients with
a tidal volume less than 20 mL was not validated.
Measurement accuracy for adult patients with a
tidal volume less than 50 mL was not validated.
Environments of use
PulmoVista 500 is intended for use in hospital PulmoVista 500 must not be used:
environments only. – During transport
– In areas of explosion hazard
When switched off and disconnected from the
– In oxygen-rich environments
patient, PulmoVista 500 may accompany a patient
– In areas where radioactive substances are used
during transport within the hospital.
The use in computed tomography examination labs WARNING
is possible, but this may result in impaired EIT data. Risk of fire and of explosion
The use in operating rooms is possible, but PulmoVista 500 is not intended to be used in
requires special safety measures regarding sterile areas with explosion hazards and in the pres-
surgery areas and electrosurgical treatment. For ence of flammable anesthetic agents or other
further information, see the following section: flammable agents.
"Product-specific safety information" on page 11.
Option ADAP
Overview
PulmoVista 500 . . . . . . . . . . . . . . . . . . . . . . . . 20
EIT module . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Right side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Underside . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Patient interface . . . . . . . . . . . . . . . . . . . . . . . 25
Electrode belt for adults . . . . . . . . . . . . . . . . . . 25
Patient cable for adults . . . . . . . . . . . . . . . . . . . 26
Electrode belt for pediatric patients . . . . . . . . . 27
Patient cable for pediatric patients . . . . . . . . . . 28
Trunk cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 29
Abbreviations in the PulmoVista 500
software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Abbreviations in these instructions for use . . . . 30
Abbreviations for labeling . . . . . . . . . . . . . . . . . 31
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
PulmoVista 500
F
A
G
B
C E
D H
010a
Front
Infinity C500 (MS18746): A Alert bar, lights up red during system start, not
otherwise used
B
EF D C
134
Back
A
B
C C
D
012
C C
D
135
A Cockpit holder
B Cover with opening for system cable
C USB ports (right and left)
D Serial communication ports (RS 232)
EIT module
D
B A
C
D
F
015
020
A Trunk cable port
F Securing bracket for system cable plug
B Label for trunk cable port
C Test connector
Back
D Label for test connector
Underside
G
E G
019
G Ventilation slots
Underside
A B C D
014b
Front
Patient interface
The patient interface is made up of the following PulmoVista 500 is intended for use on patients with
components: a chest circumference of 36 cm (14.2 in) to 150 cm
– Electrode belt for adults (sizes S-XXL) or (59 in).
pediatric patients (sizes 4XS-XS)
The electrode belt and the patient cable are
– Reference electrode
available in different versions for adults and
– Patient cable (for adults or pediatric patients)
pediatric patients.
– Trunk cable
Exterior surface
022
E Electrodes, 1 to 16
A B C B D
F Midposition marker
1 16
L L
021
NOTE
The patient cable is specifically designed for the
detection of very low voltages. Since the patient
cable is very complex, it is very sensitive and must
be handled with care. Do not bend, pull or twist the
cable.
Size Color
S Medium blue
M Dark blue
L Dark red
XL Gray
XXL Violet
A
С 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 R
L R
E D C B
028
A 16 electrode snaps
B Reference electrode snap, labeled Ref
C Right patient cable port, green colored, labeled R
D Left patient cable port, red colored, labeled L
E Closure snap, labeled C
Exterior surface
046
Sizes XS and 2XS:
Sizes 3XS and 4XS:
A B C B
1 XS 15
H F H
58 - 72 cm 22.8 - 28.3“
D E
044
047
Sizes 3XS and 4XS:
H Electrodes, 1 to 16
A BC F B
1 4XS 15
36 - 44 cm 14,2 - 17,3“
D E G
045
NOTE
The patient cable is specifically designed for the A
detection of very low voltages. Since the patient
cable is very complex, it is very sensitive and must
be handled with care. Do not bend, pull or twist the
cable.
R
on the electrode belt. The end of the patient cable
next to electrode snap 16 is equipped with a
reference electrode snap. The other end of the C
patient cable terminates with the snap for L
016
electrode 1 because the patient cable does not
A Trunk cable plug
have a closure snap.
B Right patient cable plug, green colored, labeled R
The patient cable is available in one size and is
marked with the compatible sizes of the electrode C Left patient cable plug, red colored, labeled L
belt (XS-4XS). The patient cable can be attached to
all electrode belts for pediatric patients.
A
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 R
L R B
D C
048
A 16 electrode snaps
B Reference electrode snap, labeled Ref
C Right patient cable port, green colored, labeled R
D Left patient cable port, red colored, labeled L
Color coding of the electrode snaps of the patient
cables:
Size Color
XS-4XS White
Abbreviations
Abbreviation Explanation
MB Megabyte
MEDIBUS Dräger communication protocol
for medical devices
MRI Magnetic resonance imaging
Rel. Relative
RF Radio frequency
SELV Safety extra low voltage
SIM Subscriber Identity Module
USB Universal serial bus
V Voltage
Abbreviation Explanation
C Closure snap
IPX Degree of protection against in-
gress of water
L Left
Paux Auxiliary pressure
R Right
Ref Reference electrode snap
Symbols
terial
REF Part number, sometimes
including revision index This end up
Symbol Explanation
Non-stackable
Fragile
Use by
Operating concept
Hardware elements. . . . . . . . . . . . . . . . . . . . . 36
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Rotary knob . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Fixed keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
LEDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Screen elements . . . . . . . . . . . . . . . . . . . . . . . 38
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Header bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Monitoring area. . . . . . . . . . . . . . . . . . . . . . . . . 38
Main menu bar . . . . . . . . . . . . . . . . . . . . . . . . . 39
Dialog windows. . . . . . . . . . . . . . . . . . . . . . . . . 39
Button and tab colors . . . . . . . . . . . . . . . . . . . . 40
Calibrating the touchscreen . . . . . . . . . . . . . . . 40
Hardware elements
Fixed keys
Rotary knob
Two fixed keys are located at the Cockpit. The
On/Off key (A) is located in the left corner. The
Audio Paused key (B) has no functionality.
A
027
screen.
2 Turn the rotary knob clockwise to increase a
numeric value, scroll down a list or navigate to
the right during horizontal navigation.
Turn the rotary knob counterclockwise to
decrease a numerical value, scroll up a list or
navigate to the left during horizontal navigation.
3 Press the rotary knob to confirm a setting or to
activate a function when it appears yellow.
LEDs
The Cockpit is equipped with several LEDs that A Mains power LED
report the device status.
B On/Off key and LED. The LED in the key lights
Infinity C500 (MS18746): up when the system is switched on
C Battery LED – indicates the battery status
D Ambient light sensor
A B C
003
DB A C
134
LED indications
Screen elements
Overview E
A B C D
F
503
The following diagram describes the major screen
elements of the Cockpit as the user interface of A Patient ID (requires the option ADAP)
PulmoVista 500. Each element is described in more
detail in the following sections. B System time
C Filter settings
A
D Signal quality indicator
E Technical messages
F Help text messages
B
Monitoring area
C
The monitoring area consists of the area for images
and waveforms and the parameter boxes. The
appearance and the content of the monitoring area
can be adapted to the clinical situation with the use
502
C
Header bar B
The blue header bar is always visible and always
appears along the top of the Cockpit. The header
bar is divided into several fields which are reserved
for specific types of information. The fields that are
504
A Dialog windows
B
C Dialog windows consist of one or more pages,
D
E which may in turn have subpages. Pages and
subpages are accessed by touching the
corresponding horizontal or vertical tab. Dialog
F windows contain elements for operating the system
and inform the user of current settings. Dialog
G windows can be opened by touching a button in the
H main menu bar.
I
502
A
A Views..., for accessing the predefined views B
– Main
– Full-screen image C C C C
– Ptp analysis
– End-insp. trend D
– EELI trend
– Diagnostics D
For details, see "Main view" on page 64. D
B Freeze display, see "Freeze display"
on page 82
C Export screenshot, see "Export screenshot"
on page 82
D Mark event..., see "Marking events"
on page 82
E Data review... (requires the ADAP option), see
"Data review" on page 87
505
Colors identify the availability of functions or If the touchscreen can no longer be operated, it
settings. might be maladjusted. In this case, calibrate the
touchscreen.
The colors listed in the following table indicate if a
button or a tab is available or selectable and if a NOTE
user action is required.
No screen content is displayed during touchscreen
Color Status calibration. Therefore, only calibrate the touch-
screen if no screen content needs to be displayed.
Light green The button is active, ready for se-
lection. 1 Press and hold the rotary knob and the key
Dark green The button has been selected simultaneously for at least 10 seconds.
and displays the current selec- The calibration screen is displayed.
tion.
2 Touch the markings displayed consecutively on
Yellow The button is selected. However, the touchscreen.
user input or confirmation by
pressing the rotary knob is re- Only for Infinity C500 (MS18746):
quired. 3 To confirm the calibration, touch the green tick.
Gray The button is not available for se-
lection due to a certain mode or
required connection.
Intrahospital transport . . . . . . . . . . . . . . . . . . 45
Before transportation . . . . . . . . . . . . . . . . . . . . 45
Moving the trolley . . . . . . . . . . . . . . . . . . . . . . . 45
WARNING WARNING
Risk of personal injury Risk of device malfunction, of electric shock,
and of mechanical instability
Modifications to the medical device may
cause malfunctioning. If there are any signs of damage, do not use
PulmoVista 500.
Do not modify this medical device without au-
thorization by Dräger. Contact DrägerService.
B
043
WARNING NOTE
Risk of electric shock and of device malfunc- Only use the power supply cord supplied by
tion Dräger.
Do not use the medical device outside of a
hospital. NOTE
Position PulmoVista 500 so that its mains plug can
be pulled out for disconnection without difficulty.
WARNING
Risk of electric shock Make sure that the mains power supply is in
Only connect PulmoVista 500 to a mains pow- accordance with the specifications in the
er supply with protective grounding. "Operating data" section.
Insert the mains plug into the mains power
WARNING outlet.
Risk of high leakage current The LED on the Cockpit and the LED on the
Wrong connections may jeopardize the elec- power supply module light up green.
trical safety of this device.
Only connect the mains plug to a fixed mains Power supply from the internal battery
power socket. The power supply module incorporates an internal
battery to ensure that operation of the device can
WARNING continue for at least 5 minutes without mains power
Risk of electric shock (if the battery is fully charged).
Wrong connections, e.g., to other equipment, When mains power is connected, the internal
may violate the insulation concept. battery is being charged. The internal battery will
only reach its full capacity after charging for
Do not disconnect the system cables. Contact 24 hours.
DrägerService if necessary.
NOTE
Keep PulmoVista 500 connected to the mains
power supply during storage to prevent battery de-
pletion.
Potential equalization
The power supply module contains a potential
equalization connector, which can be used in
specific cases for connections with other electrical
devices to ensure an equal ground potential. For
further details, refer to IEC 60601-1, or applicable
national or international standards.
10°
033
Intrahospital transport
WARNING
Risk of personal injury and device damage
A toppling risk exists near thresholds, on un-
even ground and on ramps.
Move the trolley at a walking pace. Securely
hold the device using the handle when pass-
ing thresholds and obstacles.
Before transportation
B A
136
B
1 Insert the connector (A) into the COM 1 port (B)
at the back of the Cockpit.
PulmoVista 500 can display the pressure data of a The pressure values measured by the
PressurePod. The PressurePod can be attached to PressurePod can be selected on the
the handle of PulmoVista 500 and connected to one Start/Standby page, see "Monitoring sessions"
of the Cockpit's USB ports. on page 59.
Getting started
Device check. . . . . . . . . . . . . . . . . . . . . . . . . . 51
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Preparing the device check . . . . . . . . . . . . . . . 51
Performing the device check . . . . . . . . . . . . . . 52
Monitoring sessions. . . . . . . . . . . . . . . . . . . . 59
Continuing the monitoring session . . . . . . . . . . 60
Starting a new monitoring session . . . . . . . . . . 60
Selecting and activating measurements. . . . . . 60
Entering patient data . . . . . . . . . . . . . . . . . . . . 60
Start/Standby
CAUTION
After extended exposure to a cold environment, B H
acclimatize the device carefully so that condensa-
tion cannot form on the electronic parts and does
not damage the device. C D
F
A
A
018
The progress bar indicates the progress of the – Device check (H), to open the device check
startup. page, see page 51
Device check
R
R
technical messages not related to known adverse TEST C
measurement conditions. L
F
Do not start the device check while the patient is L E
connected to PulmoVista 500.
031
1 Insert trunk cable plug (A) into the trunk cable
port (B).
2 Insert green patient cable plug (C) into the test
connector (D).
3 Insert red patient cable plug (E) into the test
connector (F).
Performing the device check The test results obtained from the device check are
stored until the next test, even if the device is
The device check is only possible in standby mode. switched off.
If PulmoVista 500 is not already in standby mode, If the device check was successful, measurements
switch to standby mode. may be started.
If the device check fails, do not operate the device.
Switching to standby mode Contact DrägerService.
1 Touch the Start/ Standby... button in the main
menu bar. On completion of the device check
2 Touch the Standby button on the Disconnect the patient cable plugs from the test
Start/Standby page. connector on the EIT module.
Start/Standby D
C
501
Selecting the belt size Connecting the patient cable to the electrode
belt (for adults)
The size labels and color codes of the electrode
belt for adults and the patient cable for adults are Connect the patient cable before attaching the
the same. The patient cable for pediatric patients electrode belt to the patient.
can be used in conjunction with all electrode belts
for pediatric patients. The following table shows
size and color according to chest circumference. 1 16
Adults: B
Chest circumference Color Size
70 to 85 cm Medium blue S 1 16
(28 to 33 in)
80 to 96 cm Dark blue
A
006
M
(31 to 38 in)
1 Connect snap 1 (A) of the patient cable to
92 to 110 cm Dark red L stud 1 (B) of the electrode belt. Ensure the
(36 to 43 in) orientation of the cable over the stud is
106 to 127 cm Gray XL maintained as illustrated and as marked on the
(42 to 50 in) electrode belt.
124 to 150 cm Violet XXL 2 Connect all other snaps in numerical order to
(49 to 59 in) the studs of the electrode belt. Make sure that
the patient cable changes direction above each
Pediatric patients: stud.
Chest circumference Color Size
36 to 44 cm Medium blue 4XS
(14.2 to 17.3 in)
42 to 52 cm Violet 3XS
(16.5 to 20.5 in) C
005
Male Female
F
2
С
E
140
007
1 Place the electrode belt with the patient cable
connected at the height of the 4th to 5th 4 Close the electrode belt so that all electrodes
intercostal space (medioclavicular line) around have close contact with the skin.
the chest of the patient. For female patients Depending on the position of the belt closure, the
place the electrode belt so upon the chest in EIT image may change slightly. With increasing
such a way that its position matches distance between electrodes 1 and 16 the left and
approximately the 5th intercostal space. right ventral parts of the EIT image close up. Where
Ensure that the left-right orientation of the electrode wound dressings or other material prevent skin
belt is correctly maintained. The red patient cable contact with all 16 electrodes, EIT measurements
port must be to the patient's left and the green can be performed with only 15 electrodes.
patient cable port to the patient's right. However, at least 15 electrodes must have
effective contact with the skin.
5 Connect the closure snap (E) to the closure
stud (F).
D
Connecting the reference electrode
008
Connecting the patient cable to the electrode Attaching the electrode belt to the patient (for
belt (for pediatric patients) pediatric patients)
Connect the patient cable before attaching the Infant Pediatric patient
electrode belt to the patient.
1 15
051
1 Place the electrode belt with the patient cable
connected at the height of the 4th to 5th
intercostal space (medioclavicular line) around
15
16
1
the chest of the patient.
050
052
1
Pediatric patient:
3
16
2
14
5
3
13
4
053
Ensure that the left-right orientation of the electrode
055
belt is correctly maintained. The red patient cable Infant:
port must be to the patient's left and the green
patient cable port to the patient's right.
Example for pediatric patients:
D 16
15 1 2
3
14
13 4
056
4 Close the electrode belt as follows so that all
electrodes have good skin contact:
054
2 Ensure that the midposition marker (D) in the Attach the flap of the electrode belt to
center of the electrode belt is located over the electrode stud 16 at a suitable position of
spine of the patient. the belt closure. Attach electrode snap 16 of
the patient cable to the electrode stud of the
3 Ensure that the electrode belt is positioned in electrode belt.
such a way that electrodes 1 and 16 are the
same distance from the sternum. For pediatric patients, the maximum stretch of the
Wherever possible, attach the electrodes with electrode belt must not exceed 10 %. If the
an equal distance to each other. Irregular electrode belt shows a clearly visible deformation,
electrode distances affect the quality of the EIT the recommended stretch of the electrode belt has
data. been exceeded. If a stretch of more than 10 % is
necessary, use an electrode belt with the next
higher belt size.
Depending on the position of the belt closure, the
EIT image may change slightly. With increasing
distance between electrodes 1 and 16 the left and
right ventral parts of the EIT image close up. Where
wound dressings or other material prevent skin
contact with all 16 electrodes, EIT measurements 7 To prevent infants from developing pressure
can be performed with only 15 electrodes. marks, place the patient on a soft surface such
However, at least 15 electrodes must have as a Dräger SoftBed mattress or an air
effective contact with the skin. mattress.
16
1
15
14
13
3
4
15
16
1
14
3
13
Ref
G H
Ref
057
Infant:
059
1
16
15
2
14
3
13
G H
Ref
058
1 Position PulmoVista 500 next to the bed of the 2 Connect the patient cable plugs (A) of the trunk
patient. cable to the patient cable ports (B). Ensure that
the color and the label on the patient cable
A B plugs match the corresponding color and label
on the patient cable ports.
3 Make sure there is no tension on the trunk cable
L
004
Monitoring sessions
G H
500
Touch the Start button (B). To enter the patient data, touch the Patient
data... button (H).
Trend data and reference data from the last session
are saved. Trend data is not collected while The Patient data page will open.
PulmoVista 500 is switched off or in standby mode.
Collection of trend data is resumed when the
monitoring session is continued. The measurement
starts.
Operation
Main view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Setting ROIs . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Displaying reference data. . . . . . . . . . . . . . . . . 67
Deleting reference data . . . . . . . . . . . . . . . . . . 68
Other views . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Selecting a view . . . . . . . . . . . . . . . . . . . . . . . . 69
Full-screen image . . . . . . . . . . . . . . . . . . . . . . . 69
Ptp analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
End-inspiratory trend . . . . . . . . . . . . . . . . . . . . 73
EELI trend . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Freeze display. . . . . . . . . . . . . . . . . . . . . . . . . 82
Export screenshot . . . . . . . . . . . . . . . . . . . . . 82
Marking events . . . . . . . . . . . . . . . . . . . . . . . . 82
Signal quality . . . . . . . . . . . . . . . . . . . . . . . . . 84
Signal quality indicator . . . . . . . . . . . . . . . . . . . 84
Signal check . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Standby mode. . . . . . . . . . . . . . . . . . . . . . . . . 86
Activating standby mode . . . . . . . . . . . . . . . . . 86
Ending operation . . . . . . . . . . . . . . . . . . . . . . 91
Switching off the system. . . . . . . . . . . . . . . . . . 91
Removing the patient interface. . . . . . . . . . . . . 91
Disconnecting PulmoVista 500 from
a Dräger device . . . . . . . . . . . . . . . . . . . . . . . . 92
WARNING
Risk of excessive leakage current to the pa-
tient
Do not simultaneously touch the contacts of
any connector and the patient.
WARNING
Risk of image distortion which may lead to
wrong therapeutic decisions
The following sources may distort the clinical
information content of EIT data:
– Patient movement, muscle contractions
– Changes of skin-electrode contact
– Changes of fluid accumulation in the tho-
rax or under the skin
– Movement of fluids or tissue, e.g., dia-
phragm, into or within the electrode plane
– Very low impedance of the thorax, e.g., due
to thickened skin or pulmonary edema
– Adiposity
– Temperature changes
For more information refer to the sections
"Signal quality" on page 84 and "General con-
siderations" on page 147.
CAUTION
Screenshots contain compressed information.
The image quality of screenshots might be affect-
ed by artifacts or loss of information.
NOTE
PulmoVista 500 does not provide a warning when
the battery is depleted.
Main view
The Main view is the standard monitoring page. C Global impedance waveform – this shows the
relative impedance changes of the entire
Images represent the regional distribution of
electrode plane and, if available and
ventilation, i.e., the change of lung volume within
configured, a waveform imported via
the electrode plane. Waveforms represent the
MEDIBUS, MEDIBUS.X, or from the
volume changes within the electrode plane or parts
PressurePod (as selected on the System
of it over time. Parameters represent regional
setup page)
volume changes as a percentage of the whole.
D Regional impedance waveforms – represent
Regions of interest (ROIs) are defined for all views
relative impedance changes of the defined ROI
from this page.
E Blue markers – represent beginning and end of
The Main view displays the following:
inspiration
Displaying reference data The Set ROI... page provides the following buttons
and tabs:
See "Displaying reference data" on page 67.
A Vertical slider – moves the ROI vertically within
the image (inactive when the set ROIs cover the
Setting ROIs entire height of the image)
B Horizontal slider – moves the ROI horizontally
This function is used to define ROIs of 4 equal sizes within the image (inactive when the set ROIs
within the status image. A region of interest (ROI) is cover the entire width of the image)
a user-defined area within a status image. The
image can be divided horizontally or into quadrants. C Width – changes the width of ROIs
The area covered by each ROI is represented by D Height – changes the height of ROIs
the corresponding regional impedance waveform.
E Layers – ROIs are adjacent and vertically
ROIs may overlap or be set so that the ROIs do not aligned
cover the entire thoracic contour. It is not possible
to move the ROIs beyond the border of the image. F Quadrants – ROIs are adjacent and positioned
It is possible that the ROIs do not represent all the in an upper left/right and lower left/right
ventilation in the measured plane. arrangement
Prerequisite: The Main view or End-insp. trend G Free – single ROIs can be moved
view is open. independently of each other
B 536
Set ROI
G
537
519
value by turning the rotary knob and press to
1 Touch the Free button (G).
confirm.
The buttons ROI 1 (A), ROI 2 (B), ROI 3 (C), and
4 Touch the horizontal (B) or vertical (A) slider.
ROI 4 (D) are displayed.
Move the ROIs by using the touch screen or by
turning the rotary knob. Alternatively, the ROIs 2 Touch the relevant button for the ROI.
can also be moved by touching and dragging
3 Touch the button in the Width line (E). Set the
them within the image.
value by turning the rotary knob and press to
confirm.
4 Touch the button in the Height line (F). Set the
value by turning the rotary knob and press to
confirm.
5 Touch the horizontal (G) or vertical (H) slider.
Move the ROIs by using the touch screen or by
turning the rotary knob. Alternatively, the ROIs
can also be moved by touching and dragging
them within the image.
Touch the Display change button (F). Touch the Ref button (A).
The following actions will lead to the loss of
A reference data:
– PulmoVista 500 is switched off.
– Standby mode is activated and the frame rate is
changed.
H
G – Standby mode is activated and New patient is
F I selected.
– The simulation mode is activated.
J
521
Other views
A Main E
C
B Full-screen image
E
C Ptp analysis
D End-insp. trend
E
518
E EELI trend A Small format image, according to selection in
System setup
F Diagnostics
B Global impedance waveform
C Large format image, as selected in System
Selecting a view setup
1 Touch the Views... button in the main menu bar. D Ref button
2 Touch the button to select the desired view. E Parameter fields as in the Main view
EIPtp
A Status image or dynamic image (as selected on – End-expiratory transpulmonary pressure
the System setup page) EEPtp
B Global impedance waveform and, if available, a – Difference Ptp between EIPtp and EEPtp
waveform imported via MEDIBUS, transpulmonary driving pressure
MEDIBUS.X, or from the PressurePod (as
selected on the System setup page) K Parameter field – this displays the following, if
available:
C Airway pressure waveform Paw (Pod) imported
from the PressurePod (as selected on the – End-expiratory gastric pressure EEPga
Start/Standby page) – Transdiaphragmatic pressure Pdi
D Esophageal pressure waveform Pes (Pod) L Pulmonary image – schematic, graphic
imported from the PressurePod (as selected on representation of the pressure measurements
the Start/Standby page) Paw, Pes, and Pga, and of the determination of
E Calculated transpulmonary pressure waveform the parameter Ptp
Ptp M Static analysis... button for the manual
determination of end-inspiratory and end-
F Gastric pressure waveform Pga/Paux3 (Pod)
expiratory plateau pressures
imported from the PressurePod (as selected on
the Start/Standby page) N Occlusion test... button for checking the
placement of the esophageal balloon
G Parameter field – this displays the following, if
available: O Filter... button for setting a specific low-pass
filter for the pressure waveforms of the
– Tidal Rate PressurePod
– Tidal volume VT imported via MEDIBUS or P Pes... button for checking the filling of the
MEDIBUS.X esophageal balloon
Performing the static analysis To select the plateau points, do the following:
End-inspiratory and end-expiratory plateau points 1 Place the cursor (E) at the desired position on
can be selected manually on the Static analysis the end-expiratory plateau.
page, with which a more precise determination of
2 Place the cursor (F) at the position immediately
the parameter Ptp is possible.
before or after that on the end-inspiratory
plateau.
Static analysis X
The parameter field for gastric pressure Pga A The Pes/Paw match diagram shows the
and for transdiaphragmatic pressure Pdi is not overlaid sections of the Paw (Pod) and Pes
displayed. (Pod) waveforms marked with the sliders.
D Filter... button for setting the low-pass filter for B Paw (Pod) airway pressure waveform imported
the pressure waveforms of the PressurePod from the PressurePod
E Cursor button for the end-expiratory C Pes (Pod) esophageal pressure waveform
plateau imported from the PressurePod
G Horizontal sliders for marking the occlusion Displaying the trend of the Pes parameter
phase
On the Pes page, the filling of the esophageal
H Refresh button balloon and the resulting changes to the Pes
parameter can be checked.
Defining filters
Pes
On the Filter page, cardiac artifacts can be
suppressed in the pressure waveforms that are B
imported from the PressurePod. The low-pass filter A
settings have no effect on other displayed C
561
waveforms.
A Trend waveform of the Pes parameter
Filter B Button for the time scale
C Refresh button – for updating the time scale
A
B
560
The End-insp. trend view displays the following: If signal quality is low, this is indicated at the cursor
label and then the waveform is a darker gray color.
Setting the time scale The EELI trend view displays the following:
The time scale determines the duration of the trend O B
waveforms. The time scale can be set from 1 to H I D
120 minutes. The default setting is 10 minutes. P
A C E
1 Touch the Time scale button (K).
2 Select the time by turning the rotary knob and C E
press to confirm.
C E
If the time scale is changed in the End-insp. trend
view, the time scales also change in the EELI J C E
trend view and the Diagnostics view.
K L
Refreshing EIT data N F M
G
531
Touch the Refresh button (L) (inactive when the
zoom area has been zoomed into). A Differential image EELI: C minus Ref –
displays an image based on the differences
CAUTION between the end-expiratory values at cursor
positions Ref (H) and C (I).
Risk of wrong therapeutic decisions
B Global impedance waveform – up to
The trend views are affected by various factors.
120 minutes of data are displayed as a
For information on the influencing factors see
compressed waveform. The end-expiratory
page 147.
level at each cursor position is displayed as a
horizontal dashed line.
J Table displaying the following data at the Setting the time scale
selected cursor positions Ref and C:
The time scale determines the duration of the trend
– Tidal variation values, expressed as
waveforms. The time scale can be set from 1 to
percentages
120 minutes. The default setting is 10 minutes.
– MEDIBUS, MEDIBUS.X, or PressurePod
data, if available, and as selected in the 1 Touch the Time scale button (L).
system configuration, see page 98
2 Select the time by turning the rotary knob and
– Date and time
press to confirm.
K MEDIBUS, MEDIBUS.X, or PressurePod trend
If the time scale is changed in the EELI trend view
parameters, if available, and as selected in the
and the Diagnostics view, the time scale also
system configuration, see page 98
changes in the End-insp. trend view.
L Time scale button (inactive when the zoom
area has been zoomed into)
Refreshing EIT data
M Refresh button (inactive when the zoom area
Touch the Refresh button (M) (inactive when
has been zoomed into)
the zoom area has been zoomed into).
N Event list... button - all events since the start of
measurement are displayed in a table. CAUTION
O Button for opening the Zoom page – enables Risk of wrong therapeutic decisions
the selection of a zoom area within the trend The trend views are affected by various factors.
data. For information on the influencing factors see
P Button for zooming in/zooming out – allows the page 147.
zooming in and zooming out of the zoom area
(inactive when no zoom area has been
defined).
If signal quality is low, this is indicated at the
cursor label and then the waveform is a darker
gray color.
The Conduct analysis dialog window displays the The Conduct analysis dialog window also
following elements: contains the Sections and RVD configuration
pages.
X
A B NOTE
The previous analysis is deleted every time the
C section settings are changed.
D D
E F G H
I
557
A A
B C D B C
E D
F E H
F G
555
554
On the Sections page (A), the user can make the On the RVD page (A), the user can make the
following settings: following settings:
– Automatically define sections (B) – The – Switch the calculation and display of the RVD
sections are automatically defined and can be parameter on (B) or off (C).
shifted as needed by tapping on a section on
– Set the cut-off frequency of the low-pass filter
the global waveform and dragging it across the
for RVD (D).
screen to the desired position or moving it there
by turning the rotary knob. To confirm the – Set the RVD threshold (E). All pixels with an
position, press the rotary knob or tap outside of RVD value greater than the set RVD threshold
the waveform field. are displayed in yellow.
– Manually add a section (C) – The new, light gray – Select the RVD parameter. Touch the RVD
section is added in the middle of the two sliders Ratio (F) or RVD SD (G) button. RVD Ratio, in-
and can be moved to the desired position by dicated as % of the total number of pixels of the
tapping the waveform field, dragging the section ventilated area, describes the ratio of the num-
across the screen, or by turning the rotary knob. ber of pixels with an RVD above the RVD
To confirm the position, press the rotary knob or threshold to the total number of pixels. RVD SD,
tap outside of the waveform field. To confirm the indicated as % of Ti, describes the standard de-
position, press the rotary knob or tap outside of viation of all RVD values of the ventilated area.
the waveform field.
If the calculation and display of the RVD parameter
– Manually delete a section (D) – Tap the desired is switched off, only the buttons (B) and (C) are active.
section. The section is highlighted in light gray.
The following relevant notes (H) on using and
Touch the Remove section button (D). Another
interpreting the RVD parameter must be observed:
section can be selected by turning the rotary
knob. Press the rotary knob to delete the The regional ventilation delay (RVD) is a parameter
section. which displays regional inspiration delays
compared to global inspiration and may indicate the
– Define spontaneous breaths or mandatory
cyclical collapse and re-opening of lung regions as
breaths per section (E).
well as display regionally varying time constants.
– Define the reference section for the Varying time constants have a particularly heavy
Customized function (F). impact on the display of the RVD parameter during
long inspiratory times, short expiratory times and a
Sections can only be moved within the area defined
quick rise in ventilation pressure. A “long”
by the sliders.
ventilation phase can be recognized based on the
patient flow of 0 at its end.
The influence of varying time constants can be
largely suppressed through a slow pressure rise up
to the end of the inspiration and a long expiratory
time.
M
A
C
D
K K
F
E
G
N
H I
O
J P
L Q
565
A Global impedance waveform – up to 2 hours of C Tidal images (max. 5 images are displayed
data are displayed as a compressed waveform simultaneously)
in relation to time. If available, the trend
– With an active RVD analysis: Areas in which a
parameters PEEP and EIP imported via
regional delay was detected, depending on the
MEDIBUS or MEDIBUS.X are also displayed.
configuration, are displayed in yellow. Tapping a
Sections whose tidal images are displayed in
tidal image opens the RVD analysis dialog
the analysis are highlighted in light gray.
window which displays an RVD image. The
Sections which are not displayed are displayed
RVD image displays an early inspiration
in dark gray.
compared to the global progression in cyan and
B Zoom button – only available if the horizontal a delayed inspiration in yellow. Furthermore, the
sliders are not at the start and end of the scale. progression of the impedance waveform of a
selected pixel is compared to the progression of
the global impedance waveform. By tapping the
displayed image, any pixel can be selected.
K
A
C
D
I I
F
E G
L
H M
N
J O
556
A Global impedance waveform – up to 2 hours of C Tidal images (max. 5 images are displayed
data are displayed as a compressed waveform simultaneously)
in relation to time. If available, the trend
– With an active RVD analysis: Areas in which a
parameters PEEP and EIP imported via
regional delay was detected, depending on the
MEDIBUS or MEDIBUS.X are also displayed.
configuration, are displayed in yellow. Tapping a
Sections whose tidal images are displayed in
tidal image opens the RVD analysis dialog
the analysis are highlighted in light gray.
window which displays an RVD image. The
Sections which are not displayed are displayed
RVD image displays an early inspiration
in dark gray.
compared to the global progression in cyan and
B Zoom button – only available if the horizontal a delayed inspiration in yellow. Furthermore, the
sliders are not at the start and end of the scale. progression of the impedance waveform of a
selected pixel is compared to the progression of
the global impedance waveform. By tapping the
displayed image, any pixel can be selected.
– With an inactive RVD analysis: Areas in which a G In addition to the images, the sum of the values
regional delay was detected are not marked. for C loss LP [%] or C loss [%] is displayed as
Tapping a tidal image opens the Details dialog follows:
window in which the tidal image is displayed in
– PEEP trial: Compliance loss towards lower
a larger format and the progression of the
PEEP levels (C loss LP [%]) in white
impedance waveform of a pixel is compared to
the progression of the global impedance – Customized: Compliance loss (C loss [%]) in
waveform. orange
D Display of the RVD parameter – Depending on H Diagram with the numeric values of all sections
the configuration, for every section expressed depending on the settings and analysis type.
as the standard deviation of the RVD (RVD SD)
I Scroll buttons for shifting the displayed images
or as the RVD Ratio. RVD SD, indicated as % of
– only active if more than 5 sections are
Ti, describes how inhomogeneous the regional
defined.
inspiration within the contour of the ventilated
area is. RVD Ratio, indicated as % of the total J Event list...
number of pixels, describes the ratio of the
K Help...
pixels with RVD to the total number of pixels
within the contour of the ventilated area. L Conduct analysis...
E PEEP trial: Images of the Compliance loss for M Capture analysis – The currently displayed
every displayed section. The pixel compliance analysis is saved as a screenshot.
loss is calculated in relation to the maximum
N Review analysis... – only active if at least one
compliance which is ascertained within all
analysis is saved. Saved analyses can be
sections. If there is a compliance loss towards
opened and viewed in a separate dialog
higher PEEP levels (C loss HP [%]), the loss is
window.
displayed in orange. If there is a compliance
loss towards lower PEEP levels (C loss LP O Create report – only active if a USB mass
[%]), the loss is displayed in white. The regions storage device with sufficient storage space is
without a compliance loss are displayed in dark connected to the Cockpit. An image file is
gray. created containing the entered patient data, the
Customized: Images of the compliance event list, all available analysis parameters and
changes in comparison to the selected MEDIBUS values, and a screenshot of the
reference section for every displayed section, analysis. Furthermore, an editable text file with
referred to as a compliance win (C win [%]) and the same information but without the
Compliance loss (C loss [%]). The regions with screenshot is created. Reports can also be
compliance win are displayed in cyan and the created based on previously saved analyses.
areas with compliance loss in orange. To do this, the saved analysis must be selected
with the Review analysis... button, followed by
F In addition to the images, the sum of the values
tapping the Create report button.
for C loss HP [%]or C win [%] is displayed as
follows:
– PEEP trial: Compliance loss towards higher
PEEP levels (C loss HP [%]) in orange
– Customized: Compliance win (C win [%]) in
cyan
Freeze display
Export screenshot
This function is only available if a USB storage The screen image is written to the USB storage
medium with sufficient storage space is connected medium as a bitmap file.
to the Cockpit.
The USB storage medium may be removed when
1 Connect a USB storage medium to the left or the button turns light green.
right USB port on the Cockpit. When the USB
storage medium is correctly connected, the
Export screenshotbutton turns light green.
2 Touch the Export screenshot button in the
main menu bar. When the button is touched,
it turns dark green.
Marking events
Clearing comments
Touch the Clear button (D).
Signal quality
Signal check
535
500
Restarting calibration
400
Prerequisite: PulmoVista 500 is not in standby
300 D mode.
200 C When the message Restart calibration is
100 displayed, calibration is recommended.
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 R Touch the Start button (E).
The following functions are not available during
E calibration:
– Record
534
– Mark event
The Signal check page displays the following: – Reference
A Electrode position
B Resistance of each electrode
C Resistance of reference electrode
D Defined limit of resistance
E Start button to restart calibration
The status of the electrodes (A) is indicated as
follows:
Status Explanation
Gray Skin-electrode contact is sufficient
for measurement to proceed
White Skin-electrode contact is unstable,
measurement will proceed but signal
quality may be impaired
Red Skin-electrode contact is insufficient
for measurement to proceed
Standby mode
Switch to standby mode for the following actions: Activating standby mode
– Starting a new monitoring session
– Performing a device check 1 Touch the Start/ Standby... button in the main
– Changing the following settings in System menu bar.
Setup:
– Imported MEDIBUS or MEDIBUS.X 2 Touch the Standby button (B) on the
waveforms Start/Standby page (A).
– EIT settings
Start/Standby
– Language
– Access to service settings
– Dräger default settings A
– File handling for data recording (requires
the option ADAP)
– Simulation of patient data B
– Switching off PulmoVista 500
500
B Status image – as selected in the display – Stop – stops the play function and returns
settings on the Data review page the cursor to the beginning of the current
– Tidal image: Differential image of the end of page
inspiration compared to the beginning of – Next page , opens the next page of the
inspiration, which represents regional EIT file
distribution of impedance changes of the
M Global tidal variation TV global or global minute
last detected breath.
tidal variation MTV global – depending on the
– Minute image: Tidal images averaged over the
selected status image, expressed as the sum of
last minute or less, when the cursor is located
all regional relative impedance changes of the
within the first minute of the EIT data file.
entire status image (dimensionless)
C Set ROI... – to set the ROI, see "Setting ROIs"
N Regional tidal variation or minute tidal variation
on page 90
– depending on the selected status image,
D File... – to select a file, see page 89 expressed as the sum of regional relative
impedance changes within the defined ROI
E Patient data... – displays patient data
(dimensionless)
associated with the opened file, see page 89
O Global tidal variation TV global or global minute
F Display... – to change the display settings,
tidal variation MTV global – depending on the
see page 89
selected status image, always expressed as
G MEDIBUS data... – displays the MEDIBUS or 100 %
MEDIBUS.X parameters at the cursor position if
P Regional tidal variation or minute tidal variation
the selected file contains MEDIBUS or
– depending on the selected status image of the
MEDIBUS.X data. One MEDIBUS or
defined ROI, expressed as percentage
MEDIBUS.X parameter can be selected that is
superimposed onto the global impedance Q Global relative impedance change (compared
waveform, see page 90. to the baseline level) at the cursor position,
expressed as the sum of all regional relative
H File name of the selected EIT file
impedance changes within the entire displayed
I Global impedance waveform – displays the dynamic image (dimensionless)
relative impedance changes
R Regional relative impedance change at the
J Regional impedance waveforms – represent cursor position, expressed as the sum of
relative impedance changes of the defined ROI regional relative impedance changes within the
part of the displayed dynamic image
K Cursor
(dimensionless) that corresponds to the defined
L 5 navigation buttons: ROI within the status image
– Previous page , opens the previous
S MEDIBUS or MEDIBUS.X parameter
page of the EIT file
(displayed in green color), if selected on the
– Play – this moves the cursor through the
MEDIBUS data page
current page of the EIT file. In standby mode
the cursor speed is increased with repeated T Event marker
pressing.
U Indicator for the signal quality of the selected
Depending on the selected cursor speed,
EIT file
the recorded EIT files may replay faster or
slower than real-time data. V Event comment
– Pause – pauses the cursor at its current
W Auto rescale button
location. Play can be resumed by pressing
the Pause button a second time.
File
A
542
The following patient data are displayed:
B – Patient ID (D)
– Short comment (E)
– Notes (F)
Patient data are entered and edited from System
C
setup > Data recording > Patient data. Patient
data that has been recorded as part of an EIT file
cannot be edited.
541
MEDIBUS data
A
547
Ending operation
Infinity C500 (MK31500) 3 Remove the closure snap from the closure stud.
4 Open the electrode belt.
5 Lift the patient and remove the electrode belt
A from under the patient. Make sure that the
electrode belt does not snap back towards the
018
2 Press the On/Off key (A) on the Cockpit. patient. Ensure that the patient is lifted clear of
the electrode belt so that no skin damage can
PulmoVista 500 opens the Shut down device appear when removing the electrode belt.
dialog window.
6 Remove the patient cable from the electrode
Shut down device belt.
If no further EIT measurements are intended for
B C the patient, remove the reference electrode. If
electrode gel or electrode spray has been used,
clean residue off the skin of the patient.
529
Configuration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Overview
This chapter describes the system setup functions. Changes to system setup settings become
The System setup dialog window provides the effective immediately. Changes to settings are
user with the following setup pages: retained after switching off PulmoVista 500. All
– Screen layout settings can be set to factory defaults on the
– EIT settings Service page.
– System
The Product specific page in the Service dialog
– Data recording (requires the option ADAP)
window is protected by an access code.
J V
K W X
507
Setting the time scale 1 Touch the System setup... button in the main
menu bar to open the Screen layout page (A).
The time scale setting determines the duration of
the impedance waveforms in the Main view and the 2 Select the MEDIBUS import (B) tab.
Full-screen image view.
System setup
1 Touch the button (V) in the row Time scale (J).
2 Select the value by turning the rotary knob and
press to confirm.
A
The available time scale settings vary with the C
selected frame rate.
D
Baseline update B
The baseline update setting defines at which E
interval the baseline of the impedance waveforms
is updated.
Touch the Breath button (W) to update the F
baseline after each detected breath.
G
G
Touch the Time scale (X) button to update the
baseline each time the cursor passes the zero G
position of the time scale.
For more details, see "Definition of baseline
508
frames" on page 159. PulmoVista 500 displays the MEDIBUS or
MEDIBUS.X settings (C), the selected baud rate
(D) of PulmoVista 500 and the model of the
Configuring the MEDIBUS import connected Dräger device (E).
The parity and stop bit settings for COM 1 are fixed On the MEDIBUS import page, the user can
on PulmoVista 500 and must be adjusted on the change the following settings:
connected Dräger device. The value 19200 or 9600
can be selected as the baud rate, which must be Imported MEDIBUS waveforms (F)
configured accordingly for PulmoVista 500 and the
connected Dräger device. To connect the Dräger To select which MEDIBUS or MEDIBUS.X
device, see "Preparation for using the MEDIBUS or waveforms are stored with EIT data, proceed as
MEDIBUS.X protocol" on page 46. Use the values follows:
below to configure the COM settings on the Dräger 1 Touch the button (G) to open the parameter
device according to its instructions for use. selection list.
D
G
564
On the EIT settings page, the user can change the A noise spectrum analysis is conducted during the
following settings: calibration phase. The noise spectrum is displayed
– Frame rate [Hz] (B) in the diagram (C).
– Operating frequency [kHz] (F)
– Adjustment of operating frequency (G), (H)
Operating frequency
E H I
514
2 Touch the System (A) tab. 2 Press the rotary knob to confirm.
3 Touch the Service (B) tab. The following system settings are reset to factory
settings:
System setup
Parameters Setting
A Filter Off
Status image Tidal image
Full-screen image and Ptp anal- Status image
ysis view
C Display of thoracic contour On
B Time scale 30 s
Frame rate 20 Hz
Adjustment of operating fre- Automatic
quency
D Record mode Continuous
MEDIBUS baud rate 19200
File length for recording 30 s
Saved MEDIBUS or Paw, Flow, Vol.
MEDIBUS.X waveforms
E MEDIBUS or MEDIBUS.X vol- Off
ume waveform superimposed
515
Service
H
F G H
J K L
546
The list of installed applications is displayed (L).
Installing applications
Only use SIM cards and USB-SIM card readers
545
In the simulation mode, previously recorded EIT To activate simulation mode, proceed as follows:
files can be displayed in all views. The simulation
1 Touch the relevant buttons (E), (F), (G) or (H)
mode can be used for demonstration purposes. If
to open the list of available folders, subfolders,
patient data are changed in the simulation mode,
and files. Select a folder, a subfolder, and the
e.g., for demonstration purposes, these changes
EIT files that are to be used for the simulation.
will not affect recorded data.
Selecting folders and subfolders requires the
Prerequisite: PulmoVista 500 is in standby mode.
option ADAP.
1 Touch the System setup... button in the main
Without the ADAP option, a folder and a
menu bar.
subfolder are preselected. The relevant (E) and
2 Touch the System (A) tab. (F) buttons are inactive. A preinstalled example
EIT file for demonstration purposes can be
3 Touch the Simulation (B) tab.
selected using the (G) button.
System setup 2 Touch the On button (C) and press the rotary
knob to confirm.
In the header, Simulation is displayed.
A
To stop the simulation mode, proceed as follows:
1 Touch the Start/ Standby... button in the main
C D menu bar. Touch the Standby button on the
Start/Standby page.
2 Touch the System setup... button in the main
E B menu bar.
3 Touch the System (A) tab.
4 Touch the Simulation (B) tab.
F 5 Touch the Off button (D).
H
553
H
Patient data
516
On the Patient data page, the user can change the
following data and settings:
– Patient ID (C)
– Short comment (D)
– Notes (E)
– Record mode (I)
– File length (M)
– Interval for periodic recording (N)
When data is recorded, file names are The directory and the file name are displayed in the
automatically created using the following format: (G) field. If either the patient ID or short comment is
– Folder: current date changed, a new subfolder and a new file name is
– Subfolder: Patient ID_Short comment created after the first recording.
– File: Patient ID_Short comment_Incremental
If no Patient ID or no Short comment is entered,
counter
the subfolder and the file name are both
automatically labeled ID_SC.
The incremental counter depends on the record
mode.
Clearing data Touch the Single (J), Cont. (K), or Periodic (L)
button.
Touch the Clear data button (H) to clear the text
fields.
File length
Data recording settings 1 Touch the File length button (M).
The record mode and the file length can be 2 Select the value by turning the rotary knob and
adjusted. press to confirm.
The record mode can be set as follows: Be aware that the time taken to open files increases
– Single – Single file mode to record a single file with the size of the file.
of a selected file length when the Record button
in the main menu bar is touched.
Interval for periodic recording
– Cont. – Continuous mode to record multiple
continuous files of a selected file length as long 1 Touch the button for setting the interval (N).
as the Record button in the main menu bar is
2 Select the value by turning the rotary knob and
active.
press to confirm.
– Periodic – Periodic mode to record files of a
selected file length at intervals as long as the
Record button in the main menu bar is active.
NOTE
A continuous recording is automatically ended
after 24 hours.
Troubleshooting
Reprocessing
Safety information
WARNING WARNING
Risk due to inappropriately reprocessed prod- Risk of electric shock and of device malfunc-
ucts tion
Reusable products must be reprocessed, oth- Penetrating liquid may cause malfunction of
erwise there is an increased risk of infection. or damage to the device, which may endanger
– Follow the infection prevention policies the patient.
and reprocessing regulations of the
Do not immerse or rinse the EIT system or its
health-care facility.
components (e.g., patient interface, Cockpit,
– Follow the national infection prevention cables, USB storage media) in liquids. During
policies and reprocessing regulations. surface disinfection make sure that no liquids
– Use validated procedures for reprocess- enter the device.
ing.
– Reprocess reusable products after every If a liquid is accidentally spilled on the device,
use. disconnect the device from the power source.
– Follow the manufacturer's instructions for Contact service personnel regarding the con-
cleaning agents, disinfectants, and repro- tinued safety of the device before putting it
cessing devices. back into operation.
CAUTION WARNING
Risk due to faulty products Do not use sharp tools or abrasives. Do not
immerse electrical connectors in water or oth-
Signs of wear, e.g., cracks, deformation, discolor- er liquids. Avoid the accumulation of liquid at
ation, or peeling, may occur with reprocessed the edge of the screen as this may damage the
products. device.
Check the products for signs of wear and replace
them if necessary. CAUTION
Do not clean the Cockpit using an alcoholic solu-
WARNING tion. Such solutions may damage the Cockpit.
Risk of electric shock
CAUTION
Do not sterilize PulmoVista 500 and its acces-
Risk of damage to the touchscreen
sories as this may damage the insulation of
cables. Instead, perform the cleaning proce- Do not spray the cleaning liquid directly on the
dures described in the instructions for use. touch screen. The cleaning liquid must be applied
to a cloth or tissue before cleaning the surface.
WARNING
Risk of electric shock NOTE
Do not clean the touch screen of the Cockpit while
Do not disconnect the system cables. If nec-
monitoring a patient. Before cleaning, take the de-
essary, contact service personnel.
vice out of operation.
Information on reprocessing
Follow the national infection prevention policies Follow the infection prevention policies and
and reprocessing regulations. reprocessing regulations of the health-care facility
(e.g., concerning the reprocessing cycles).
Classification Explanation
Non-critical Components that come into contact only with skin that is intact
Semi-critical Components that carry breathing gas or come into contact with mucous mem-
branes or pathologically altered skin
Critical Components that penetrate skin or mucous membranes or come into contact with
blood
Non-critical
– Device surface including screen
– Patient interface
– Trolley with modules
Before reprocessing
Device-specific components
Components Surface disin- Manual cleaning fol- Machine cleaning Description of the
fection with lowed by disinfection with thermal dis- procedure
cleaning by immersion infection
Device surface and Yes No No See page 122
trolley
Patient interface Yes No No
Components:
– Device surface and trolley
Components:
– Patient interface
After reprocessing, there are no special All further information on storage and transport
requirements for storage and transport of the included in the accompanying documents must be
product. However, the following must be observed: observed.
– Store dry and free of dust
– Avoid recontamination and damage during
transport
Disinfectants
Surface disinfectants
The manufacturers of the surface disinfectants
have verified at least the following spectra of
activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Follow the manufacturer's instructions for surface
disinfectants.
The following surface disinfectants were
compatible with the material of the device surface
and trolley at the time of testing:
After reprocessing
Prerequisites:
– All components have been reprocessed and
dried.
Procedure:
See chapter "Getting started" on page 49.
Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Definition of service terminology. . . . . . . . . . . . 129
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Overview
WARNING
Risk of fire, explosion, or excessive tempera-
ture
Replacement of the batteries must only be
performed by specialized service personnel.
WARNING
If the device is mechanically damaged or if it
is not working correctly, do not use it. Contact
service personnel.
WARNING
When servicing the device, always use re-
placement parts that are qualified to Dräger
standards. Dräger cannot warrant or endorse
the safe performance of third-party replace-
ment parts for use with the device.
Concept Definition
Service All measures (inspection, preventive maintenance, repair) intended to main-
tain and restore the functional condition of a medical device
Inspection Measures intended to determine and assess the actual state of a medical de-
vice
Maintenance Recurrent specified measures intended to maintain the functional condition
of a medical device
Repair Measures intended to restore the functional condition of a medical device af-
ter a device malfunction
Inspection
Preventive maintenance
WARNING WARNING
Risk of faulty components Risk of electric shock
Device failure is possible due to wear or mate- Before performing preventive maintenance,
rial fatigue of the components. disconnect the device from the power supply.
To maintain the proper operation of all compo-
nents, this device must undergo inspection
and preventive maintenance at specified inter-
vals.
Repair
Disposal
At the end of its service life: of at municipal collection points for waste electrical
and electronic equipment. Dräger has authorized a
Have the medical device appropriately
company to collect and dispose of this device. To
disposed of in accordance with applicable laws
initiate collection or for further information, visit
and regulations.
Dräger on the Internet at www.draeger.com. Use
the Search function with the keyword "WEEE" to
For countries subject to the find the relevant information. If access to Dräger's
EU Directive 2002/96/EC website is not possible, contact the local Dräger
Organization.
This device is subject to EU Directive 2002/96/EC
(WEEE). In order to comply with its registration
according to this directive, it must not be disposed
Disposal of batteries
Technical data
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Classification . . . . . . . . . . . . . . . . . . . . . . . . . 136
Ambient conditions
During operation
Temperature (device) 5 to 40 °C (41 to 104 °F)
Temperature (electrode belt and cables) 5 to 45 °C (41 to 113 °F)
Ambient pressure 700 to 1060 hPa (10.15 to 15.37 psi)
Relative humidity 20 to 95 %, without condensation
Height Up to 3000 m (9842 ft)
During storage and transportation
Temperature –20 to 40 °C (–4 to 104 °F)
Ambient pressure 500 to 1060 hPa (7.25 to 15.37 psi)
Relative humidity 20 to 90 %, without condensation
Settings
Performance characteristics
EIT measurement
Number of electrodes 16 electrodes plus 1 reference electrode
Feed current amplitude Max. 9 mA (rms, depending on the frequency of the
feed current)
80 to 100 % of maximum patient auxiliary current
conforming to IEC 60601-1 (3rd edition)
Frequency of the feed current and the mea- 80 to 130 kHz
surement
Pulse length of the feed current per electrode pair 2.5 to 12.5 ms (depending on the frame rate)
Display unit (Medical Cockpit Infinity C500)
Resolution 1440 x 900 pixels
Contrast ratio Min. 500 : 1
Horizontal viewing angle 130° (typical)
Vertical viewing angle 100° (typical)
Operating data
Materials used
Electrode belt Silicone rubber, conductive silicone rubber, stain-
less steel, gold-plated brass
Patient cable Plastics (thermoplastic polyurethane (TPU), poly-
amide (PA), polyurethane (PUR), polypropylene
(PP), thermoplastic elastomer (TPE), polybutylene
terephthalate (PBT))
Trunk cable Plastics (polyamide (PA), thermoplastic polyure-
thane (TPU), polyurethane (PUR))
Classification
Device combinations
This device can be operated in combination with operating organization must ensure that the device
other Dräger devices or with devices from third- combination complies with the applicable editions
party manufacturers. Follow the accompanying of the relevant standards for medical devices.
documents of the individual devices.
Device combinations approved by Dräger meet the
If a device combination is not approved by Dräger, requirements of the following standards:
the safety and the functional integrity of the – IEC 60601-1 (electrical safety, mechanical
individual devices may be compromised. The safety, software)
– IEC 60601-1-2 (EMC)
EMC declaration
General information
This device was tested for electromagnetic functions correctly in the desired arrangement
compatibility using accessories from the list of before use. The instructions for use for the other
accessories. Other accessories may only be used devices must be followed.
if they do not compromise the electromagnetic
compatibility. The use of non-compliant NOTE
accessories may result in increased In the case of transient high voltage in the mains
electromagnetic emissions or decreased power supply, it may not be possible to switch off
electromagnetic immunity of the device. the device. Contact DrägerService.
This device may be used in the direct vicinity of
other devices only if Dräger has approved this
device arrangement. If no approval has been given
by Dräger, it must be ensured that this device
Electromagnetic environment
Emissions Compliance
Radiated emissions Class A, group 2 (30 MHz to 1 GHz)
Conducted emissions Class A, group 2 (150 kHz to 30 MHz)
NOTE
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a res-
idential environment (for which CISPR 11 class B
is normally required), this equipment might not of-
fer adequate protection to radio-frequency com-
munication services. The user might need to take
mitigation measures, such as relocating or re-ori-
enting the equipment.
Connections to IT networks
Principles of operation
Information displayed by
PulmoVista 500 . . . . . . . . . . . . . . . . . . . . . . . . 152
Dynamic image. . . . . . . . . . . . . . . . . . . . . . . . . 152
Status images. . . . . . . . . . . . . . . . . . . . . . . . . . 152
Impedance waveforms . . . . . . . . . . . . . . . . . . . 153
Numeric values. . . . . . . . . . . . . . . . . . . . . . . . . 153
Image generation
034
all bioimpedance measurements at different Subsequently, the adjacent electrode pair is used
electrode pairs are referenced to the same electric for the next current injection and another 13 voltage
potential. measurements are performed. The location of the
injecting and measuring electrode pairs
Principle of measurement successively rotates around the entire thorax. One
complete rotation results in 16 voltage profiles,
PulmoVista 500 applies a known alternating current each consisting of 13 voltage measurements. The
"I1" to a pair of electrodes and measures the resulting 208 values, called a frame, are used to
resulting surface potentials "Vn" at the remaining reconstruct one cross-sectional image.
13 electrode pairs. Applying the law of Ohm, the
bioelectric impedance between the injecting and
the measuring electrode pairs can be determined
from the known current and the measured voltages.
A
C
036
PulmoVista 500 uses a FEM-based, linearized
Newton-Raphson reconstruction algorithm to
D convert the 208 voltage changes of a frame into an
035
Superposition of voltage profiles Like CT scans, the projection of the displayed EIT
images is from caudal to cranial. That means that
The reconstruction algorithm superposes the
the left side of the image displays the right side of
16 voltage profiles on each other. Reconstruction
the patient. The upper part of the image displays
artifacts are eliminated by applying selective
the ventral aspect of the patient.
boundary filtering.
Tissue Resistivity
(, m)
Blood 1.5
Heart 1.6 to 4.3
037
E
038
548
between ventilation-related impedance changes In healthy lungs there is little difference in the
and air content, for example as determined by distribution of ventilation between normal-weight
CT scans. and obese individuals. However, the ventilated lung
regions of obese individuals are displayed much
smaller in the image as the lungs are surrounded by
Some examples of EIT images from a large amount of adipose tissue.
PulmoVista 500
Pathological conditions which may lead to non-
Status image of lungs of a normal-weight ventilated lung regions
healthy adult
Various factors are associated with non- or under-
ventilated lung regions. These may include:
– overinflation (more common in the ventral
regions)
– pneumothorax (more common in the ventral
and lateral lung regions)
– atelectasis (more common in the dorsal
regions)
– pleural effusion (more common in the dorsal
regions)
– pneumonia
547
EIT and CT images from a patient with dorsal Interpretation of negative impedance changes
pleural effusion
The definition of baseline frames typically leads to
a display of solely positive impedance changes.
Nevertheless, in some circumstances negative
impedance changes may occur which need to be
interpreted correctly.
The following known effects may lead to negative
impedance changes:
– A deeper expiration than inspiration. Since the
baseline is taken at a particular end-expiratory
level, a deeper expiration in the subsequent
breath reduces the end-expiratory lung volume
and thus leads to values below the defined
549
baseline level.
A fluid accumulation such as a pleural effusion
represents a non-ventilated area and is displayed – Pendelluft: The baseline frame taken at the end
by PulmoVista 500 in black or purple color. In the of expiration defines, for every pixel, the zero
CT scan the same region is displayed in bright level. If all parts of the lung increase and
color. decrease their impedance simultaneously, then
negative values do not occur. However, when a
lung region has different filling characteristics to
other lung regions, the minimum of this
particular region may differ from the defined
zero level. This may then result in negative
values in this region.
– Artifacts: During EIT measurements, various
conditions may cause artifacts, which need to
be considered during interpretation of images.
550
Examples
Example of an artifact induced by electromagnetic
fields:
551
Tidal image
dZ
Global
ROI 2
A
ROI 1
ROI 1
B
ROI 2
C
539
Special considerations when using the The following aspects must be carefully considered
Diagnostic view when ascertaining the compliance changes
C loss LP [%] and C loss HP [%] as well as
As with any recent clinical information, experience C loss [%] and C win [%]:
with the numeric values described in this case is – Before the start of the PEEP trial function,
also limited. For this reason, users should interpret make sure that every PEEP level has been
these numeric values carefully and under correctly assigned to a section. A section
consideration of other, well-established clinical averages the selected number of spontaneous
data. breaths. For this reason, sections which contain
several breaths must be set to areas with an
The Recruitability function enables the evaluation end-expiratory lung impedance which is as
of trend data which includes recruitment constant as possible.
maneuvers. Conclusions about the potential – The change to the regional ventilation can only
recruitability of the lung can be drawn from changes be interpreted as a change to the regional
in compliance as well as in the end-expiratory lung compliance if the ventilator displays a flow of
volume [16]. Therefore, in addition to changes in 0 at the end of the inspiration and expiration.
regional compliance (C loss [%], C win [%]), This is because, if regional inspiratory times or
changes in end-expiratory lung impedance (EELI) expiratory times are shorter than the
are also shown in this analysis. corresponding regional time constants,
For PEEP maneuvers, the PEEP trial function a C loss LP [%] and C loss HP [%] as well as
enables the systematic analysis of regional C loss [%] and C win [%] may also be triggered
changes to compliance and delays in regional by varying regional time constants.
ventilation RVD in addition to the evaluation of – If the ventilator transfers no EIP value, the PIP
ventilation distribution. A compliance loss towards parameter can be used as a substitute for the
higher PEEP levels (C loss HP [%]) may indicate calculation of the compliance during volume-
overdistended regions, and a compliance loss controlled ventilation. For this, the AutoFlow
towards lower PEEP levels (C loss LP [%]) may function must be activated.
indicate collapsed regions [13]. – If no ventilation pressures are available from the
ventilator, the ventilation change can only be
In contrast, the Customized function enables the interpreted as a change to the compliance
evaluation of any other therapeutic measures. during pressure-controlled ventilation. This is
Here, the compliance win (C win [%]) and because if no ventilation pressures are detected
compliance loss (C loss [%]) compared to the during volume-controlled ventilation, the falling
selected reference section is displayed and pressure differences may cause the regional
indicated as a percentage change. tidal variation to decrease, which, in turn, would
RVD is a parameter which displays regional cause the C loss LP [%], C loss HP [%], and
inspiration delays compared to global inspiration C loss [%] to rise, without any change to the
and may indicate the cyclical collapse and re- regional compliance of the lung. The pressure
opening of lung regions [14, 15]. difference may have fallen due to rising
compliances in other lung areas at the
respective PEEP level.
Electrode plane
132
However, as the contribution of impedance
changes outside the electrode plane is relatively
small, the effect on the image is limited.
Electrode plane
122
039
deviations dgn.
As the baseline frame is continuously updated after
each detected breath, mainly impedance changes
due to tidal ventilation are displayed. Due to this
baseline definition, end-expiratory impedance
variations are suppressed to a great extent in the
dynamic image.
the last minute. The image is ideal for the To make them comparable, all regional impedance
assessment of regional distribution during waveforms are normalized using the same scaling
ventilation with varying tidal volumes. factor. This is usually smaller than the global
scaling factor.
Regional quantification can be customized in the
status image by adjusting the position and size of
the Regions Of Interest (ROIs).
Numeric values
The numeric value EELI global displays the C loss [%] and C win [%] are calculated as follows:
deviation of the global end-expiratory status at the
1 The EIT data is low-pass filtered with a cut-off
cursor positions Ref and C in relation to the global
frequency of 80 min–1. The artifact filter is acti-
tidal variation at Ref.
vated.
The numeric value ∆EELI ROI displays the 2 If a section contains several spontaneous
deviation of the regional end-expiratory status at breaths, these are averaged to ascertain the
the cursor positions Ref and C in relation to the tidal variation TV.
global tidal variation at Ref.
3 The tidal variation TVi is calculated for every
In the Diagnostics view, the C loss [%] pixel within the ventilated area.
(= Compliance Loss), ∆EELI, and RVD (= Regional
Ventilation Delay) parameters can also be 4 The compliance is calculated for every section
calculated and displayed for every defined section. and every pixel:
The algorithms used for the calculations are
partially based on publications which point out the
following:
– The method for determining the C loss HP [%] If no pressure values are available from the
and C loss LP [%] parameters enables an ventilator, the EIP – PEEP pressure difference is
estimation of lung region overdistension and set to 1.
collapse [13]
– The RVD index enables the detection of cyclical 5 Every pixel in the selected reference section is
opening and closing of lung regions [14, 15] defined as RefCompliance.
– The combined presentation of changes in 6 In every section and for every pixel, C loss [%]
compliance and lung volume can be used to is calculated as the compliance loss of the sec-
assess the recruitability of the lung [16] tion in comparison with the reference compli-
ance. C loss [%]i is set to 0 if the difference re-
sults in a positive value.
Compliance loss (C loss [%]) and Compliance
Win (C win [%]) C loss i,S o n = Compliance i,,S
i , S e c t iion o n – RefCompliance i
S e c t iion
Trend data without a PEEP maneuver can be 7 The global parameter C loss [%] is also
analyzed using the Customized function. This calculated for every section by summing up the
analysis determines changes in regional compliance loss percentage of all pixels (NV) of
compliance in comparison with a selectable the ventilated area:
reference section for every displayed section.
These changes are expressed as Compliance loss NV
i= 1
i=1 C loss i,Section
i,S e c t ion
C loss S e c ttion (% ) =
i o n (%) x 100%
100
(C loss [%]) and Compliance Win (C win [%]). NV
i= 1
i=1 RefCompliance i
– The C loss [%] parameter describes a regional 8 In every section and for every pixel, C win [%] is
compliance loss compared to the reference and calculated as the compliance loss of the section
is displayed in orange. in comparison with the reference compliance.
– The C win [%] parameter describes a regional C win [%]i is set to 0 if the difference results in a
compliance win compared to the reference and negative value.
is displayed in cyan.
C win i,Section
i,S e c t ion = Compliance i,Section
,S e c tion – RefCompliance i
Pixels with the same compliance as in the
reference section are displayed in dark gray. 9 The global parameter C win [%] is also
calculated for every section by summing up the
compliance win percentage of all pixels:
NV
i= 1
i=1 C win i,Section
i,S e c t ion
C win Section (% ) =
S e c tion (%) x 100%
100
NV
i= 1
i=1 RefCompliance i
EELI loss [%] and EELI win [%] An end-expiratory lung impedance loss
(EELI loss [%]) of e.g., 100 % thus corresponds to
Trend data with a recruitment maneuver can be
an end-expiratory lung volume loss equal to the
analyzed using the Recruitability function.
global tidal variation or the tidal volume of the
In addition to changes in regional compliance reference section.
(C loss [%], C win [%]), changes in end-expiratory
lung impedance (EELI) are also shown in this
Compliance loss (C loss LP [%] and
analysis. These changes are determined in
C loss HP [%])
comparison with a selectable reference section for
every displayed section, expressed as Trend data with PEEP maneuvers can be analyzed
EELI loss [%] and EELI win [%]. using the PEEP trial function. In this analysis,
– As in the EELI trend view, ∆EELI images are C loss [%] images only represent the PEEP-
also displayed in this analysis, each relating to induced regional compliance loss for various levels
the end-expiratory lung impedance (EELI) of of a PEEP maneuver.
the reference section.
With this method, both of the following conditions
– Pixels with an EELI loss are shown in orange.
are differentiated for every pixel:
Pixels with an EELI win are shown in cyan.
– The C loss LP [%] parameter describes the
– Pixels without EELI changes are displayed in
regional compliance loss compared to the
dark gray.The
ascertained maximum regional compliance
– The EELI loss [%] parameter is calculated by
towards lower PEEP levels and is displayed in
adding together the regional loss of end-
white.
expiratory lung impedance compared to the
– The C loss HP [%] parameter describes the
reference and is displayed in orange.
regional compliance loss compared to the
– The EELI win [%] parameter is calculated by
ascertained maximum regional compliance
adding together the regional win of end-
towards higher PEEP levels and is displayed in
expiratory lung impedance compared to the
orange.
reference and is displayed in cyan.
Pixels without compliance loss are displayed in
The EELI loss [%] and EELI win [%] parameters
dark gray.
are calculated as follows:
C loss LP [%] and C loss HP [%] are calculated as
1 The EIT data is low-pass filtered with a cut-off
follows:
frequency of 80 min–1. The artifact filter is
activated. 1 The EIT data is low-pass filtered with a cut-off
frequency of 80 min–1. The artifact filter is
2 If a section contains several spontaneous
activated.
breaths, the end-expiratory values are
averaged. 2 If a section contains several spontaneous
breaths, these are averaged to ascertain the
3 The respective amounts of the positive and
tidal variation TV.
negative pixel values of the ∆EELI images
within the ventilated areas are added up and 3 The tidal variation TVi is calculated for every
divided by the global tidal variation of the tidal pixel within the ventilated area.
image of the reference section. In contrast to
4 The compliance is calculated for every section
the EELI trend view, the values in this
and every pixel:
analysis are multiplied by a factor of 100 and
shown as a percentage increase (win) and
decrease (loss).
If no pressure values are available from the to the global inspiration (derived from the global
ventilator, the EIP – PEEP pressure difference is impedance waveform) for every pixel within the
set to 1. contour of the ventilated area.
5 The maximum compliance is defined for every The RVD parameter is calculated using the
pixel of all sections. following steps:
6 In every section and for every pixel, 1 The EIT data is low-pass filtered with a cut-off
C loss LP [%]i is calculated as the compliance frequency of 80 min–1. The artifact filter is
loss percentage of the section in comparison activated.
with the maximum compliance. C loss LP [%]i is
2 If a section contains several spontaneous
set to 0 if the observed section is at a higher
breaths, these are averaged to ascertain the
PEEP level than the section with the maximum
RVD parameter.
compliance.
3 A second, adjustable low-pass filter, which only
HighestCompliance i – Compliance i,S
(HighestCompliance on)
C loss LP i,i,SS e cti (% ) =
on (%)
c tion
i , S e c ti
tion
x 100%
100 has an effect on the impedance values used for
HighestCompliance i
the RVD parameter, is also activated.
7 The global parameter C loss LP [%] is also
4 Only pixels within the ventilated area are
calculated as the weighted average of all pixels
analyzed.
of the ventilated area:
5 The Ti start and Ti end moments are defined
NV
i=1
i= (C
C loss LP i,S o n x HighestCompliance i )
i ,See cctt iion for every section based on the global
C loss LP SSee c t iion
on =
NV
i =1
i= 1 HighestCompliance i impedance waveform by ascertaining the start
8 In every section and for every pixel, C loss HP and end of the inflation during a low-flow
[%]i is calculated as the compliance loss maneuver. If the section does not contain a low-
percentage of the section in comparison with flow maneuver, the start and end of the
the maximum compliance. C loss HP [%]i is set inspiration of the regular spontaneous breaths
to 0 if the observed section is at a lower PEEP are ascertained.
level than the section with the maximum 6 The time Ti is calculated, i.e. the time period
compliance. between Ti start and Ti end.
If the value of the RVD parameter exceeds the RVD Improvement of the algorithms used to
cut-off value set by the user, the affected pixels are calculate C loss LP [%], C loss HP [%], and RVD
marked in yellow in the corresponding tidal image.
Several limitations were identified in the algorithms
described in the referenced publications for the
Global RVD parameters definition of C loss LP [%] [13] and RVD [14, 15].
These limitations were eliminated as follows during
Depending on the configuration, PulmoVista 500
the implementation in PulmoVista 500:
calculates one of two global RVD parameters which
represent either the inhomogeneity of the regional – During PEEP maneuvers, the end-expiratory
inspiration progressions or the ratio of the area lung volume (EELI) is generally high at high
affected by the RVD to the entire ventilated area. PEEP levels and low at low PEEP levels. In
contrast, the tidal variation (TV) remains
– Standard deviation of the RVD (RVD SD) [14,
constant during volume-controlled ventilation.
15]. To calculate the RVD SD parameter,
Because of the calculation of relative
indicated as % of Ti, the standard deviation of
impedance changes in comparison with a
all RVD pixel values is defined within the
baseline, this leads to a loss in the tidal
ventilated area. Therefore, the RVD SD
variations towards higher PEEP levels, which
parameter describes the inhomogeneity of the
are not, however, linked to changes in the tidal
regional inspiration progressions. RVD SD is
volume.
calculated as follows:
Improvement: The tidal variations are adjusted
using a correction factor.
– The C loss LP [%] and C loss HP [%] values
– Ratio of the RVD (RVD Ratio) require knowledge of the pressure values (EIP,
The RVD Ratio (indicated in %), which indicates PEEP) in order to calculate the regional
the ratio of the number of pixels affected by the compliance. This requires the availability of the
RVD (NRVD) to the total number of pixels (NV) pressure values from the ventilator which are,
within the ventilated area, is calculated as the however, often not available.
second RVD parameter. The RVD Ratio
parameter is calculated as follows: Improvement: The C loss LP [%] and C loss HP
[%] values are also calculated, even if no
pressure values are available. This makes it
possible to evaluate all maneuvers which are
performed during constant pressure-controlled
ventilation.
– RVD times are always indicated as positive
values which are ascertained from the earliest
to latest pixel. This makes it difficult to
differentiate regions with quick and slow
inspiration.
Improvement: The global point in time T40glob
is subtracted from the respective T40i point in
time to allow the respective RVD to be indicated
as a positive (= late) and negative (= early)
value.
NOTE
The global RVD parameters are not affected by
these changes.
040
As all frames are referred to the smallest frame of
the last breath, usually only positive changes are A Referred to BFA
displayed in the dynamic image. B Referred to BFB
C Referred to BFC
EELI trend view The Change: C minus Ref image of the EELI
trend view displays the changes of end-expiratory
The EELI trend view provides information on lung volume within the electrode plane. In contrast,
changes in end-expiratory lung impedance which the Change: C minus Ref image of the End-insp.
are mainly due to changes of end-expiratory lung trend view displays the changes of tidal volume
volume within the electrode plane. These changes within the electrode plane.
are displayed as positive or negative offsets in the
global and regional impedance waveforms.
Data review page (optional with the
In order to display and quantify those changes, all
frames displayed in this view are referred to one
option ADAP)
single baseline frame at a fixed position. This is in
contrast to the baseline definition of the Main and As in the EELI trend view, all frames displayed in
Full-screen image and Ptp analysis view views. the data review page are referenced to one single
baseline frame at a fixed position.
The frame containing the smallest value of
summarized voltages is defined as the baseline The frame containing the smallest value of
frame, corresponding to the lowest point in the summarized voltages is defined as the baseline
Global impedance waveform. frame. The baseline frame corresponds to the
lowest point in the Global impedance waveform
dZglobal within the entire recorded EIT file.
BFFixed If the time scale is set to smaller values, the
baseline frame might not be present within the
displayed part of the EIT file.
Ref C Time
041
Main and Full-screen image and Ptp The zero position of the color scale is always
analysis view views located in a position which provides 85 % for the
display of positive changes (A). The scalar
After image reconstruction, the relative impedance endpoints of positive and negative impedance
changes that have been determined for each of the changes are of the same magnitude. However,
pixels are translated into a color scale. Regions positive impedance changes (A) are represented
with impedance changes of less than 10 % of the over 85 % of the color scale while negative
determined maximum regional impedance change impedance changes (B) are compressed to 15 % of
are represented in black. Impedance changes the color scale.
above 10 % of the maximum regional impedance
change are displayed in dark blue. As values
increase, the dark blue turns into a lighter blue. A
white color indicates the regions of maximum
regional impedance change (i.e., 100 %) within the
image.
The color scale is adjusted continuously ("auto-
scaled") to permanently display the dynamics of
regional ventilation independently of actual tidal
volumes and other bioelectric conditions. Thus all
533
end-inspiratory images typically contain some
white regions. While auto-scaling provides an optimized
continuous display of the distribution of ventilation,
After each detected breath, the minima and
it prevents these images being used for the
maxima of the impedance waveform are
quantitative analysis of ventilation.
determined. The color scale and scales of the
impedance waveforms are adjusted based on the When the reference function has been activated in
determined minimum and maximum values. the Main view, two status images can be displayed.
The scaling of these two images is based on the
Waveforms are auto-scaled so that even small
status image with the smaller maximum impedance
impedance changes are amplified. This means that
change. This scaling may result in white saturated
some conditions (e.g., apnea, very low tidal
areas – either in the reference status image or in
volumes, or damping of waveforms due to filtering)
the current status image and its corresponding
may not be recognized. In order to make the user
dynamic image. Alternatively, the image Change:
aware of this situation, waveforms that only show
Current minus Ref can be selected. The color
very small impedance changes are displayed in a
scale of this differential image is designed to
darker gray color.
display major changes between two images at
The color scales of the dynamic and status images different points in time.
are always the same, thus allowing more effective
As these changes can be positive (C) or negative
comparison. Similarly, the scales of the four
(D), the zero value (representing no change) is
regional impedance waveforms are always set to
always displayed in the midposition of this
the same value.
differential color scale. Negative values are not
As the baseline is "zeroed" after each breath, the compressed.
images usually only contain positive values.
538
Minimum and maximum values of this color scale
are the same that are taken for the dynamic image. Diagnostics view
Thus, all color scales displayed in the Main and The automatic scaling of the tidal images is based
Full-screen image and Ptp analysis view views on the maximum impedance changes of the
always contain identical end points. selected section. All tidal images and the images of
compliance changes have the same scalar
endpoints. The scaling is based on the tidal image
End-insp. trend, EELI trend and of the section with the smallest maximum
Diagnostics views impedance change. This scaling may result in white
saturated areas in the tidal images of other
The scales for images and impedance waveforms sections.
in the End-insp. trend, EELI trend and
In contrast, the compliance loss images are always
Diagnostics views are automatically scaled each
scaled to 100 %. The color scale of the images of
time these views are either opened, updated, or the
compliance changes is identical to that of the
time scale is changed. If the cursor position is
differential images in the End-insp. trend view.
changed or an analysis is started, the images are
The color scale of the images of the EELI changes
also rescaled automatically.
is identical to that of the ∆EELI images in the EELI
trend view. The color scale of the compliance loss
End-insp. trend view images shows C loss HP [%] in orange and
C loss LP [%] in white. Regions with no compliance
The auto-scaling of the two status images in the
loss are displayed in dark gray.
End-insp. trend view is based on the maximum
impedance changes in these images. The two
status images and the resulting differential image
have the same scalar endpoints. Data review page (optional with the
option ADAP)
The scaling is based on the status image with the
smaller maximum impedance change. This scaling The auto-scaling on the Data review page is based
may result in white saturated areas in the other on the minimum and maximum values contained in
status image. the selected EIT file.
When the PressurePod is connected and pressure The pressure values and parameters for lung
measurements are activated, the airway pressure, mechanics listed in the table are determined from
esophageal pressure, and gastric pressure can be the pressure waveforms.
imported in PulmoVista 500 and displayed in the
The literature contains detailed descriptions of the
Ptp analysis view.
parameters [17, 18, 19].
The Ptp waveform is determined from the Paw
The following table shows an overview of the
(Pod) waveform and the Pes (Pod) waveform.
pressure values that are displayed in the Ptp
analysis view:
Literature references
1 Heyward VH. 1998. Practical body composition assessment for children, adults, and older
adults. Int J Sport Nutr (8): 285-307
2 Faes, T. J. C., H. A. van der Meij, J. C. de Munck und R. M. Heethaar. 1999. The electric resis-
tivity of human tissues (100 Hz-10 MHz): A meta-analysis of review studies. Physiol Meas 20 (4):
R1-R10.
3 Barber, D.C. 1989. A review of image reconstruction techniques for electrical impedance tomog-
raphy. Med Phys 16 (2): 162-169.
4 Visser, K. R. 1989. Electric properties of flowing blood and impedance cardiography. Ann
Biomed Eng 17:463-473.
5 Luepschen, H., T. Meier, M. Grossherr, T. Leibecke, H. Gehring and S. Leonhardt. 2005. Clinical
applications of thoracic electrical impedance tomography. Paper presented at the 6th Confer-
ence on Biomedical Applications of Electrical Impedance Tomography, London.
6 Victorino, J. A., J. B. Borges, V. N. Okamoto, G. F. J. Matos, M. R. Tucci, M. P. R. Caramez,
H. Tanaka, et al. 2004. Imbalances in regional lung ventilation: A validation study on electrical
impedance tomography. Am J Respir Crit Care Med 169 (7): 791-800.
7 Hinz, J., P. Neumann, T. Dudykevych, L. G. Andersson, H. Wrigge, H. Burchardi, and G. Heden-
stierna. 2003. Regional ventilation by electrical impedance tomography: A comparison with ven-
tilation scintigraphy in pigs. Chest 124 (1): 314-322.
8 Frerichs, I., J. Hinz, P. Herrmann, G. Weisser, G. Hahn, T. Dudykevych, M. Quintel and G. Hellige.
2002. Detection of local lung air content by electrical impedance tomography compared with
electron beam CT. J Appl Physiol 93 (2): 660-666.
9 Riedel, T., T. Richards and A. Schibler. 2005. The value of electrical impedance tomography in
assessing the effect of body position and positive airway pressures on regional lung ventilation
in spontaneously breathing subjects. Intensive Care Med 31 (11): 1522-1528.
10 Odenstedt, H., S. Lindgren, C. Olegård, K. Erlandsson, S. Lethvall, A. Aneman, O. Stenqvist and
S. Lundin. 2005. Slow moderate pressure recruitment maneuver minimizes negative circulatory
and lung mechanic side effects: Evaluation of recruitment maneuvers using electric impedance
tomography. Intensive Care Med 31 (12): 1706-1714.
11 van Gendringen H. R., A. J. van Vught and J. R. Jansen. 2004. Regional lung volume during
high-frequency oscillatory ventilation by electrical impedance tomography. Crit Care Med 32 (3):
787-794.
12 Meier, T., H. Luepschen, J. Karsten, T. Leibecke, M. Großherr, H. Gehring and S. Leonhardt.
2008. Assessment of regional lung recruitment and derecruitment during a PEEP trial based on
electrical impedance tomography. Intensive Care Med 34 (3): 543-550.
13 Costa EL, Borges JB, Melo A, Suarez-Sipmann F, Toufen C Jr, Bohm SH, Amato MB. Bedside
estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomog-
raphy. Intensive Care Med 2009; 35:1132-1137.
14 Muders T, Luepschen H, Zinserling J, Geschus S, Fimmers R, Guenther U, et al. Tidal recruit-
ment assessed by electrical impedance tomography and computed tomography in a porcine
model of lung injury*. Crit Care Med. 2012; 40 (3): 903-11.
15 Wrigge H, Zinserling J, Muders T, Varelmann D, Gunther U, von der Groeben C, et al. Electrical
impedance tomography compared with thoracic computed tomography during a slow inflation
maneuver in experimental models of lung injury. Crit Care Med. 2008; 36 (3): 903-9.
16 Godet T, Constantin JM, Jaber S, Futier E. How to monitor a recruitment maneuver at the bed-
side. Curr Opin Crit Care 2015, 21:253-258.
17 Akoumianaki E et al., The Application of Esophageal Pressure Measurement in Patients with Re-
spiratory Failure. Am J Respir Crit Care Med Vol 189, Iss 5, pp 520-531, Mar 1, 2014.
18 Mauri T et al., Esophageal and transpulmonary pressure in the clinical setting: meaning, useful-
ness and perspectives. Intensive Care Med. 2016 Sep; 42 (9): 1360-1373.
19 Nava S et al., Patient-ventilator interaction and inspiratory effort during pressure support venti-
lation in patients with different pathologies. Eur Respir J, 1997; 10:177-183.
List of accessories
Name/Description Order-No.
Trunk cable 8420048
Patient cable for pediatric patients (for electrode 8422770
belt sizes XS-4XS)
Patient cable, size S 8420029
Patient cable, size M 8420047
Patient cable, size L 8420035
Patient cable, size XL 8420271
Patient cable, size XXL 8420273
Electrode belt, size 4XS 8422583
Electrode belt, size 3XS 8422582
Electrode belt, size 2XS 8422581
Electrode belt, size XS 8422580
Electrode belt, size S 8420059
Electrode belt, size M 8420058
Electrode belt, size L 8420057
Electrode belt, size XL 8420056
Electrode belt, size XXL 8420055
ECG electrode (pack of 50) 4527750
MEDIBUS cable (male/female) 8306488
MEDIBUS cable (female/female) 8416326
ADAP kit (option) 8420006
PressurePod 8424050
Index
B E
Baseline frames . . . . . . . . . . . . . . . . . . . . . . . . 159 EIT data
refresh . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
C EIT settings
configuring . . . . . . . . . . . . . . . . . . . . . . . . 100
C loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154 Electrode belt . . . . . . . . . . . . . . . . . . . . . . . 25, 27
C loss HP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155 attaching . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C loss LP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155 Electrode plane . . . . . . . . . . . . . . . . . . . . . . . 151
C win . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154 EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Cable connectors . . . . . . . . . . . . . . . . . . . . . . . . 29 EMC declaration . . . . . . . . . . . . . . . . . . . . . . . 137
Calibrating the touchscreen . . . . . . . . . . . . . . . . 40 Ending operation . . . . . . . . . . . . . . . . . . . . . . . 91
Calibration Environments of use . . . . . . . . . . . . . . . . . . . . . 18
restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Event
Clear data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Clinical users . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cockpit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 F
hardware elements . . . . . . . . . . . . . . . . . . . 36
prepare . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 File handling . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Color concept . . . . . . . . . . . . . . . . . . . . . . . . . . 40 delete . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 File length . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Connections to IT networks . . . . . . . . . . . . . . . 139 Filter settings . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Contour of the ventilated area . . . . . . . . . . . . . . 97 Fixed keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . 17 Frame rate . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Copying to USB . . . . . . . . . . . . . . . . . . . . . . . . 108 Freeze display . . . . . . . . . . . . . . . . . . . . . . . . . 82
Cursor position . . . . . . . . . . . . . . . . . . . . . . . . . 73 Full-screen image . . . . . . . . . . . . . . . . . . . . . . . 69
D G
Data recording General WARNINGS and CAUTIONS . . . . . . . 11
configure . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . 49
settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Data review . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Date and time settings . . . . . . . . . . . . . . . . . . . 102
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . 133
Thoracic contour . . . . . . . . . . . . . . . . . . . . . . . . 97
Time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 3
Trolley
move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
park . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
preparing . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Trunk cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
U
USB space . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
User interface . . . . . . . . . . . . . . . . . . . . . . . . . . 38
V
Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Z
Zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . 76, 78, 80
Zoom in/zoom out . . . . . . . . . . . . . . . . . . . . 73, 75
Manufacturer
Drägerwerk AG & Co. KGaA
Moislinger Allee 53 – 55
23542 Lübeck
Germany
+49 451 8 82-0
FAX +49 451 8 82-20 80
http://www.draeger.com
9511001 – GA 5667.100 en
© Drägerwerk AG & Co. KGaA
Edition: 1 – 2020-04
Dräger reserves the right to make modifications
to the medical device without prior notice.