Specifications: STAR8000E

SHENZHEN COMEN MEDICAL INSTRUMENT CO.,LTD


Floor 10, Floor 11 and Section C of Floor 12 of Building 1A & Floor 1
to Floor 5 of Building 2, FIYTA Timepiece Building, Nanhuan Avenue,
Matian Sub-district, Guangming District, Shenzhen, Guangdong,
518106,P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email: infoatszcomen.com
Web: en.comen.com

2022 V1.5
Patient Monitor
8000E

Standard Configuration: Battery indicator


3 -lead ECG, RESP, Comen SpO2, NIBP, HR, QRS beep and alarm sound
Temp(Single Channel) Key backlight
Optional Configuration: Interface:
Dual-IBP, EtCO2, Dual-Temp, Nellcor/Masimo SpO2,
5-lead ECG, Thermal Recorder, Suntech NIBP, Parameter cable interface
Trolley, Wall mount, Ground wire AC power input
Safety Standards: USB port
IEC 60601-1 IEC 60601-1-8 IEC 60601-2-27 EN 1060-3 IEC RJ45 port
80601-2-30 IEC60601-2-34 IEC60601-2-49 ISO 80601-2-56 ISO Optional: Multi-functional interface
80601-2-61 Data storage
Physical Characteristics: Alarm Review: 200 groups
Size: 300mm×278 mm×150mm Wave Review: 6 hours (8 waves)
Weight: 2.72kg NIBP Review: 2000 groups
Screen Size: 12.1” color TFT screen Trend Graph: 160 hours
Resolution 800 × 600 Trend Table: 160hours
Waveforms: Up to 9 waveforms Power-off Yes
Ingress Protection IPX1 storage:
Operation Environment: Alarm: User-adjustable High and Low 3~level
Working Limits;
Temperature 0~40°C Prioritized audible and visual alarm
Humidity: ≤93% Network: Connected to Central Monitoring
Power Supply 100-240V~, 50/60Hz±1Hz System by hardwire/wireless
Battery Type: Rechargeable Lithium-ion battery Recorder:
Battery Capacity: 2200mAh (Option: 4400mAh) Type: Built-in; thermal array
Battery Recharging Channel: 3 channel waveforms
Time: Maximum 3 hours for charging; Speed: 25mm/s,50mm/s
Battery backup: 2 hours for continuous working Record width 50mm
Brightness: Manual from 10 to 100 External printer: Yes
Indicator: Respiration:
Alarm indicator Method: RA-LL Impedance Method
Power indicator
RR measurement NIBP:
range: Adult: 0~120rpm Method Automatic Oscillometric
Pediatric/Neonate:0~150rpm Work mode: Manual / Automatic/Continuous
Accuracy: 7~150rpm: ±2rpm or 2%, whichever is Measurement time: Adjustable(1~480min)
greater; 0~6rpm: unspecified Maximum
Resolution: 1 rpm measurement time: Adu/Ped: 120s; Neo: 85s
RESP Apnea 10s~60s(Adu); 10s~40s (Ped/Neo) Measurement
Alarm: Audible and visual alarm; alarm events types: Systolic, Diastolic, Mean
reviewable Range of systolic
Sweep Speed: 6.25, 12.5, 25mm/s pressure: Adult Mode:40~270mmHg
Gain Selection: X0.25, X0.5, X1, X2, X4 Pediatric Mode:40~200mmHg
ECG: Neonate Mode 40~135mmHg
TM
Lead Type: CardioTec 5-leads ECG Analysis, 12- Range of diastolic
Lead and 3-leads selectable pressure: Adult Mode:10~215mmHg
Lead selection 12-Lead I; II; III; aVR; aVL; aVF; V1~V6. Pediatric Mode:10~150mmHg
5-lead: I; II; III; aVR; aVL;aVF; V Neonate Mode 10~100mmHg
3-lead: I; II; III Range of mean
Gain Selection 1.25mm/mV, 2.5mm/mV, pressure: Adult Mode:20~235mmHg
5.0mm/mV, 10mm/mV, 20mm/mV, Pediatric Mode:20~165mmHg
40mm/mV, Auto. Neonate Mode 20~110mmHg
Manual replacement. Over-pressure
Plus ±750mV DC polarization voltage; protection: Both hardware and software
sensitivity change range: ±5%. over pressure protection
Sweep Speed: 6.25, 12.5, 25, 50mm/s Accuracy: ±3mmHg
Heart Rate Range: Adult: 15~300bpm Resolution: 1mmHg
Pediatric/Neonate:15~350bpm Alarm: Systolic, Diastolic, Mean
Resolution: 1 bpm PR from NIBP: 40~240bpm
Accuracy: ±1% or ±1bpm (Whichever is greater) Resolution: 1bpm
Protection: Withstand 4000VAC/50Hz voltage in Accuracy: ±3% or ±3bpm, whichever is greater
isolation; Nellcor SpO2:
Against electrosurgical interference Measurement
and defibrillation; range 0~100%
Bandwidth: MON Mode:0.5 Hz~40 Hz Resolution: 1%
DIA Mode: 0.05Hz~150 Hz Accuracy: ±2% (70~100%, Adu/Ped, non~motion)
OPE Mode:1 Hz~20 Hz ±3% (70~100%, Neo, non~motion)
ST Mode:0.05Hz~40Hz 1~69% unspecified
ST detection: -2.0mV~ +2.0 mV (Automatic) Alarm range: 0~100%
Resolution: 0.01mV PR Measurement
Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or ±10%; Range: 20~300bpm
Others: Unspecified Resolution: 1bpm
Arrhythmia Accuracy: ±3bpm (20~250bpm);
analysis: 26 Types Unspecified (251~300bpm)
pacemaker Alarm Range: 0~300bpm
detection: Detectable Masimo SpO2:
Measurement & 41 – 76 mmHg ±5% of reading
alarm range 1~100% 77 – 114 mmHg ±8% of reading
Resolution: 1% Above 80 BPM ±12% of reading
Accuracy: ±2% (70~100%, Adu/Ped, non-motion) Response Time Capnogram displayed in less than 3
±3% (70~100%, Neo, non-motion) seconds,full specifications within 3
1~69% unspecified minutes(At 25℃)
PR Measurement CO2 Resolution 0.1mmHg 0 to 38 mmHg
Range: 25~240bpm 0.25 mmHg 39 to 114mmHg
Resolution 1bpm CO2 Stability Short term drift: Drift over four hours
Accuracy: ±3bpm(non-motion) shall not exceed 1 mmHg maximum
±5bpm(motion) Long term drift: Accuracy specification
Alarm Range: 0~300bpm will be maintained over a 120-hour
PI value: 0.02~20% period
Resolution: 0.01% (0.02%~9.99%) Sampling 100 Hz
0.1% (10.0%~20.0%) Frequency
Accuracy: unspecified awRR range: 3-150 bpm
Comen SpO2: awRR accuracy: ±1 breath
Measurement Voltage 5.0 VDC ±5%, 2W
& alarm range 0~100% Requirements
Resolution: 1% Power Less than 1.0 Watts typical (Steady
Accuracy: ±2% (70~100%, Adu/Ped, non-motion) Consumption
State)
±3% (70~100%, Neo, non-motion)
Unspecified (0~69%) Up to 1.8 Watts maximum on power up
PR measure (Warm up)
Range: 20~250bpm Temperature and Operating: 0 to 40℃, 10 to 90% RH,
Resolution: 1bpm Humidity non-condensing
Accuracy: ±1bpm Storage: -40 to 70℃, <90% RH, non-
Alarm Range: 0~300bpm condensing
PI value: 0.05~20% Water Resistance IPX4 (sensor head only)
Resolution: 0.01% (0.05%~9.99%) Comen EtCO2 (Sidestream)
0.1% (10.0%~20.0%) size 101mm * 78mm * 46mm
Accuracy: unspecified Measurement & 0 – 114 mmHg, 0 – 15 %, 0 – 15.2
Temperature (Dual Channel) alarm range: kPa(at 760 mm Hg)
Measurement & Accuracy: 0 – 40mmHg ±2 mmHg
alarm range: 0~50°C 41 – 76 mmHg ±5% of reading
Sensor: Skin/rectal TEMP sensor 77 – 114 mmHg ±8% of reading
Resolution: 0.1°C Above 80 BPM ±12% of reading
Accuracy: ±0.1°C (exclusive of error of sensor) Response Time Capnogram displayed in less than 3
Channel: T1, T2, TD (Temperature Difference) seconds,full specifications within 5
Comen EtCO2 (Mainstream) minutes(At 25℃)
size 48mm * 24.3mm * 32.8mm CO2 Resolution 0.1mmHg 0 to 38 mmHg
Measurement & 0 – 114 mmHg, 0 – 15 %, 0 – 15.2 0.25 mmHg 39 to 114mmHg
alarm range: kPa(at 760 mm Hg) CO2 Stability Short term drift: Drift over four hours
Accuracy: 0 – 40mmHg ±2 mmHg shall not exceed 1 mmHg maximum
Long term drift: Accuracy specification Accuracy: Standard environment22 ±
will be maintained over a 120-hour
5 ℃ ,1013 ±
period
40hpa):
Sampling 100 Hz
Frequency 0~15%: ±(0.2%+reading×2%)
awRR range: 3 to 150 Breaths Per Minute (BPM) 15~25%: not defined
awRR accuracy: ±1 breath
All environment:
Voltage 5.0 VDC ±5%
Requirements
awRR range 0~150rpm
Power Less than 1.2 Watts typical (Steady
awRR accuracy ±1rpm
Consumption
State) Response time: <240msec (10% to 90%)
Up to 2 Watts maximum on power up Delay time: <2s
(Warm up) Masimo EtCO2 (mainstream)
Temperature and Operating: 0 to 40℃, 10 to 90% RH, Measurement &
Humidity alarm range: 0 ~190 mm Hg, 0~25% (at 760mmHg)
non-condensing
Accuracy: Standard environment 22 ±5℃,1013
Storage: -40 to 70℃, <90% RH, non-
condensing ±40hpa):
Water Resistance IPX4 (sensor head only) 0~15%: ±(0.2%+reading×2%)
Respironics EtCO2 (Sidestream)
15~25%: not defined
Measurement &
All environment:
alarm range: 0 ~150 mmHg, 0~20kPa (at 760mmHg)
Accuracy: ±2 mmHg (0~40 mmHg) ±(0.3kPa+reading×4%)
±5% of reading (41~70 mmHg)
±8% of reading (71~100 mmHg) awRR range 0~150rpm
±10% of reading (101~150 mmHg) awRR accuracy ±1rpm
awRR range: 2~150rpm Response time: <240msec (10% to 90%)
awRR accuracy: ±1rpm Delay time: <2s
Response time: <240msec (10%~90%) IBP
Delay time: <2s Channel: 2 Channels
Respironics EtCO2(mainstream) Measured Pressure: ART, PA, CVP, RAP, LAP, ICP, LV, AO,
Measurement & UAP, BAP, FAP, UVP, IAP, P1, P2,
alarm range: 0~150 mm Hg, 0~20kPa (at 760mmHg) Measurement Unit: mmHg/ kPa selectable
Accuracy: ±2mmHg (0~40 mmHg) Measurement ART: 0~300mmHg
±5% of reading (41~70 mmHg) Range: PA: -6~120 mmHg
±8% of reading (71~100 mmHg) CVP: -10~40mmHg
±10% of reading (101~150 mmHg) RAP: -10~40mmHg
awRR range: 0~150rpm LAP: -10~40mmHg
awRR accuracy: ±1rpm ICP: -10~40mmHg
Response time: <240msec (10% to 90%) LV: 0~300mmHg
Delay time: <2s AO: 0~300mmHg
Masimo EtCO2 (Sidestream) UAP: 0~300mmHg
Measurement & BAP: 0~300mmHg
alarm range: 0 ~190 mm Hg, 0 – 25% (at 760mmHg) FAP: 0~300mmHg
UVP: -10~ 40mmHg PR from IBP: 20bpm~350bpm
IAP: -10~40mmHg Resolution: 1bpm
P1, P2: -50~300mmHg Accuracy: ±1% or ±1bpm, whichever is greater
Accuracy: ±2% or ±1mmHg, whichever is greater Alarm Range: 20bpm~350bpm
Resolution: 0.1kPa or 1mmHg
(-50mmHg~300mmHg) *Notice: Specifications subject to changes without
Alarm Range: -50mmHg~300mmHg prior notice. All rights reserved by Comen