Sinohero-S80 (Excello D) Data Sheet

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

MULTI-PARAMETER

PATIENT
MONITOR
touching for life

S80

Features:
10.4" color TFT display with maximum 10 waveforms
Power software function include Large Font, Short Trend, Arrhythmia and ST analysis
Pitch tone to provide audibly recognized SpO2
Maximum 120-hour trends to review all parameters information
Wire net working with Central Monitoring Systems
Technical Specification S80
Size and Weight ST segment EtCO2 (Sidestream)
- Product size: 31.8x26.4x15.2cm - Measurement range: -2.0mV~2.0mV - Measure method: Infrared spectrum
- Packing size: 43x32x35cm - Resolution:0.01mV - Measure Range: 0.0~13.1 %( 0~99.6mmHg)
- Net weight: 4.5kg - Resolution: 1mmHg
- Gross weight: 6.5kg RESP - Unit: %,mmHg,kpa
- Method: Impedance variation between RA-LL(R-F) - Accuracy: 0% to 4.9%, ±0.3 %( ±2.0mmHg)
Power supply - Gain: ×1, ×2, ×4 5.0%to 13.1 %,< ±10%of the reading
- Power Voltage: AC 100-240V 50/60Hz - Bandwidth:0.25Hz to 2.0Hz (-3dB) - Measurement range of awRR: 3~150rpm
- Power Input: ≦85VA - Sweep Speed: 6.25mm/s,12.5mm/s,25mm/s - Calibration: Offset calibration:auto,manual,Gain calibration
- Safety class: Category I - Measurement Range:0~150 rpm
EtCO2(Mainstream)
Display NIBP - Measure method: Infrared spectrum
- 10.4”Color TFT display - Method: Automatic oscillometry - Warm up time: Capnogram displayed in less than 15
- Resolution:800x600 pixels - Range of measurement: seconds,At an ambient temperature of 25℃,full specifications
Adult: 10~270mmHg within 2 minutes.
Battery Child: 10~235mmHg - Measure Range: 0.0~19.7 %(0~150mmHg)
- Type: Rechargeable Acid Battery(Option Lithium Neonate: 10~135mmHg - Resolution: 1mmHg
battery) - Rise time(10 l/min):≦60ms
- Operating time under the normal use and full charge:≧ - Accuracy: Static:±2% or ±3mmHg,whichecer is greater - Unit: %,mmHg,kpa
120minutes - Unit: mmHg, kPa - CO2 Accuracy: 0-40mmHg, ±2mmHg
- 2 batteries ≧240 minutes (Option Lithium battery will - Intervals for AUTO measurement 41-70mmHg, ±5% of reading
double operate time) time:1,2,3,4,5,10,15,20,30,45,60,90 minutes; 2,4,8 hours 71-100mmHg, ±8% of reading
- Pulse rate range:40~240bpm 101-150mmHg, ±10% of reading
(at 760 mmHg,ambient temperature of 25℃)
Thermal Recorder Standard SpO2(Digital) - awRR measurement range:0~150rpm
- Method: Thermal dot array - Measurement Range:0~100% - awRR measurement Accuracy: ±1rpm
- Paper width:50mm(1.97 in) - Accuracy: At70~100%,±2%
- Paper Speed:12.5/25/50(mm/sec) At0~69%,unspecified EtCO2 (Microstream)
- Traces: Maximum 3 tracks PR - Measure method: Infrared spectrum
- Measurement Range:25~250bpm - Warm up time: Capnogram displayed in less than 20 seconds,
Alarm - Accuracy: ±1% or ±1bpm,whichever is greater At an ambient temperature of 25℃,full specifications within 2
- Three Levels: Low, medium and high minutes.
- Indication: Auditory and visual Nellcor-SpO2 - Measure Range: 0.0-19.7%(0-150mmHg)
- Setup: Default and custom - Measurement Range:0~100% - Resolution: 1mmHg
- Silence: All alarm can be silenced - Resolution:1% - Unit: %,mmHg, kpa
- Volume: 45-85 dB measured at 1 meter - Accuracy:At70~100%:±2%(Adult) - CO2 Accuracy: 0-40mmHg, ±2mmHg
At70~100%:±3%(Neonate) 41-70mmHg, ±5% of reading
Trend At70~100%:±2%(Low Perfusion) 71-100mmHg, ±8% of reading
- Trend: 120 hours At0~69%,unspecified 101-150mmHg, ±10% of reading
- Parameters option: HR, SpO2, NIBP, PR, RESP, PR (At 760 mmHg, ambient temperature of 25℃)
EtCO2, Temp1, Temp2, AA, N2O,O2, IBP1, IBP2, ST. - Measurement Range:20~300bpm (When RR>80 rpm, all the rang is ±12% of reading)
- Cycle intervals of trend storage1 min, 2 min, 3 min, 4 - Resolution:1bpm CO2 response time: <3S
min, 5 min, 10 min, 15 min, 20 min, 25 min, 30 min. - Accuracy: 20bpm to 250bpm: ±3bpm - awRR measurement range:2~150rpm
251bpm to 300bpm: unspecified - awRR measurement Accuracy: ±1rpm
Storage & Reviewing - Sample Flow Rate:50ml/min ±10ml/min
- ECG: 10 minutes one important lead’s ECG waveform Masimo-SpO2
- Alarm: 1000 groups alarm events reviewing - Measurement Range: 0% to 100% Environment
- NIBP: 750 groups NIBP measurement - Resolution:1% - Operating temperature:0~+40℃
- Arrhythmia: 128 groups data(8 seconds ECG - Accuracy: - Transportation and Storage temperature:-20℃ to +50℃
waveform) 70% to 100%:±2%(adult/pediatric, on-motion conditions) - Relative humisity:Working≦85%
- Power-off storage: 72 hours trend data & 1 ECG wave 70% to 100%:± 3 %( neonate, on-motion conditions) Transportation and storage ≦93%
form (Option) 70% to 100%:± 3 %( motion conditions) - Atmospheric pressure:860hPa to 1060hPa
0% to 69%, unspecified Transportation and storage 500-1060 hPa
ECG - Average time:2-4s,4-6s,8s,10s,12s,14s,16s
- Lead Mode:3-leads ECG input Standard configuration of S80:
5-leads ECG input PR ECG, HR, RESP, SpO2, NIBP, 1-TEMP, Lithium Battery, PR
- Lead selection:Ⅰ,Ⅱ,Ⅲ - Measurement Range:25bpm to 240bpm
Ⅰ,Ⅱ,Ⅲ,aVR,aVL,aVF,V- - Accuracy: ±3bpm(non-motion conditions) Optional parameter& configuration of S80:
- Gain: 2.5mm/mV(×0.25), 5mm/mV(×0.5), 10mm/m ± 5bpm (motion conditions) Nellcor SpO2, Masimo SpO2, 2-TEMP, 2-IBP, EtCO2(Side-
V(×1), 20mm/mV(×2), 40mm/mV(×4),Auto - Resolution: 1bpm stream, Microstream, Mainstream)
- CMRR: Monitor mode≧105dB Multigas/O2, VGA output, Thermal recorder.
Surgery mode≦105dB TEMP
Diagnostic mode≧90dB - Measurement Range: 25.0℃~50.0℃
- Frequency response(-3d B): - Accuracy: ±0.1℃ (exclusive of probe)
Monitor mode 0.5~40Hz - Unit: Celsius(℃),Fahrenheit(℉)
Surgery mode 1~25Hz - Connecting cable: Compatible with YSI-400 serial
Diagnostic mode 0.05~150Hz
- input impedance: ≧5.0Mohm IBP
- ECG signal range:±10.0mV - Channel: 2
- Electrode offset potential: ±500mV - Measurement way: Directly invasive pressure measure
- Patient Leakage Current:<10uA ment
- Standardizing signal:1mV±5% - Sensitivity of transducer: 5uV/V/mmHg, ±2%
- Baseline recovery:<5s after Defibrillation.(Mon or Surg - Impedance of transducer: 300 to 3000Ω
mode) - Measurement Range: -50~+300mmHg
- Indication of electrode separation: Every electrode(ex - Resolution: 1mmHg
clusive of RL) - Unit: mmHg, kpa, cmH2O
- Protection: Breakdown Voltage 4000AVC 50/60Hz;de - Accuracy:
fibrillator proof Static: ±1mmHg or ±2%, whichever is grater(exclusive of
- Sweep speed:12.5mm/s,25mm/s,50mm/s transducer)
±4mmHg or ±4%, whichever is grater(inclusion of
HR transducer)
- Range: Adult 10~300bpm Dynamic:±4mmHg or 4%, whichever is grater
Pediatric & Neonate 10~350bpm Transducer sites: Arterial Pressure (ART)
- Refreshing time:≦50 bpm Per2 pulses Pulmonary Artery Pressure (PA)
50~120bpm per 4 pulses Left Atrium Pressure (LAP)
≧120bpm per 6 pulses Right Atrium Pressure (RAP)
- Resolution: 1bpm Central Venous Pressure (CVP)
- Accuracy: ±1% or ±1bpm,whichever is greater Intracranial Pressure (ICP)

Specification subject to be changed without prior notice.

Sino-Hero(Shenzhen)Bio-Medical Electronics Co., LTD.


Tel : + 86-755-26408541 | 26404878 Email: sales@sinoheromed.com Fax: + 86-755-26420566 Web:www.sinoheromed.com
Add: A area, Second Floor, First Building, Tongkangfu Industrial Park, Yingrenshi Community, Shiyan Sub-district, Bao'an District, 518108 Shenzhen,
touching for life
PEOPLE'S REPUBLIC OF CHINA Version: V1.0