Professional Documents
Culture Documents
YFD360 Flaw Detector Instruction v3.2
YFD360 Flaw Detector Instruction v3.2
Operating Manual
2022-03-01
1
1 Introduction··················································· ·································4
1.1 Features of the Instrument...........................................................................4
1.2 Specifications................................................................................................. 7
1.3 Base Instrument Package............................................................................ 7
2 Understanding the Keypad, Menu System and Display····························8
2.1 Structure Feature...........................................................................................8
2.2 Screen Display...............................................................................................9
2.3 Keys and Rotary Knob Features...............................................................12
2.4 Menus and Functions................................................................................. 13
2.5 Alarm Lights..................................................................................................14
3 Initial Start-up··················································· ······························15
3.1 Power Supply...............................................................................................15
3.2 Connecting a Probe.................................................................................... 15
3.3 Starting the Instrument............................................................................... 16
4 Operation··················································· ···································1 ··7·
4.0 Display mode selection.............................................................................. 17
4.1 Parameter Item Selection.......................................................................... 17
4.2 Parameter setting........................................................................................ 17
4.3Adjusting the Display Range...................................................................... 17
4.4 Setting the Material Velocity...................................................................... 18
4.5 Setting the Display Delay...........................................................................18
4.6 Selecting the Probe Test Mode................................................................. 19
4.7 Selecting a Rectification Mode..................................................................19
4.8 Setting the A-Scan Reject Level............................................................... 20
4.9 Setting the Gain...........................................................................................21
4.10 Changing the Gain-Adjustment Increment (DB STEP).......................21
4.11 Auto Gain Feature..................................................................................... 22
4.12 Configuring the Gates.............................................................................. 22
4.13 Setting the Damping Level...................................................................... 26
4.14 Setting the Pulse Energy Level...............................................................26
4.15 Setting the Pulse Width............................................................................26
4.16 Adjusting the Pulse Repetition Frequency (PRF)................................ 26
4.17 Specifying the Probe Frequency............................................................ 27
4.18 Specifying the Crystal Size......................................................................27
4.19 Setting the Probe X-Value....................................................................... 27
4.20 Probe Delay Calibration........................................................................... 27
4.21 Setting the Angle of Incidence................................................................ 27
4.22 Magnifying the Contents of a Gate.........................................................27
4.23 Freezing the A-Scan Display...................................................................28
4.24 Setting the Display Grid........................................................................... 28
4.25 Selecting Units...........................................................................................28
4.26 Scale Setting..............................................................................................29
4.27 Fill function................................................................................................. 29
4.28 Setting the Display Brightness................................................................ 29
4.29 Skin Selection............................................................................................ 30
4.30 Key Sound..................................................................................................30
4.31 Date and Time Setting..............................................................................30
4.32 Display the System Information.............................................................. 30
4.33 Resetting the Instrument..........................................................................30
4.34 Connecting to a Computer.......................................................................31
2
5 Calibration and Measurement··················································· ·········3··2····
5.1 Calibration with Straight- and Angle-Beam Probes................................32
5.2 Calibration with Dual-Element (TR) Probes............................................ 34
5.3 Distance Amplitude Curve(DAC/TCG).....................................................35
5.4 Measuring with DGS/AVG..........................................................................37
5.5 Curved Surface Correction........................................................................ 40
5.6 AWS D1.1 Weld Rating Feature............................................................... 41
5.7 Flaw Sizing Feature.................................................................................... 42
5.8 Crack Height Measuring Feature..............................................................42
5.9 Envelope Function...................................................................................... 42
5.10 Peak Hold Feature.................................................................................... 43
5.11 B-Scan Feature..........................................................................................44
5.12 Weld Figure Feature.................................................................................44
5.13 Coded Echo color......................................................................................45
5.14 Data Storage..............................................................................................45
6 Maintenance and Care··················································· ··················4··8·······
6.1 Care of the Instrument................................................................................48
6.2 Care of the Batteries...................................................................................48
6.3 Maintenance.................................................................................................48
6.4 Warranty....................................................................................................... 48
6.5 Tips on Safety.............................................................................................. 48
Appendix··················································· ······································4··9·
Appendix A: Charging the Batteries................................................................ 49
Appendix B: Menu Structure............................................................................ 50
3
1 Introduction
The YFD360 is an advanced digital ultrasonic flaw detector featuring a multi-
color LCD and a host of new features to meet challenging inspection
requirements. It combines powerful flaw detection and measurement capabilities,
extensive data storage, and the ability to transfer detailed inspection data to the
PC via its high-speed USB port and bluetooth.
The instrument incorporates many advanced signal processing features
including a 20MHz RF bandwidth to permit testing of thin materials, narrowband
filters to improve signal to noise in high gain applications.
The instrument can be widely used in locating and sizing hidden cracks, voids,
disbands, and similar discontinuities in welds, forgings, billets, axles, shafts,
tanks and pressure vessels, turbines, and structural components.
1.1 Features of the Instrument
The instrument extends the performance and range of applications that are
capable of being satisfied by a portable instrument. The quality, portability,
durability, and dependability that you have come to expect from the popular leeb
Series of instruments remain.
Display
Hi-resolution multi-color LCD with continuously brightness control provides high
contrast viewing of the waveform from bright, direct sunlight to complete
darkness.
The hi-resolution multi-color TFT LCD display with fast 70 Hz update gives an
“analog look” to the waveform providing detailed information that is critical in
many applications including nuclear power plant inspections.
Range
Up to 10000 mm in steel; range selectable in fixed steps or continuously
variable. Suitable for use on large work pieces and in high-resolution
measurements.
Pulser
Pulse Energy selectable among 70V、150V、300V、400、500V.
Pulse Width tunable from 30ns~510ns to match the probes with different
frequency.
Pulse Repetition Frequency adjustable from 30 Hz to 2 KHz in 1 Hz increments.
Damping selectable among 50Ω、150Ω、250Ω and 400Ω for optimum probe
performance
Test Modes: Pulse echo, dual element and thru-transmission
Probe Types: Single Straight, Single Angle, Daul Straight, Daul Anlge, Thrugh,
Creeping, Surface
Receiver
Sampling:12 digit AD Converter at the sampling speed of 300 MHz
Rectification:Positive Halfwave, Negative Halfwave, Fullwave and RF
Analog Bandwidth: 0.2MHz to 20MHz capability with selectable frequency
ranges (automatically set by the instrument) to match probe for optimum
performance.
4
Gain: 0 dB to 110 dB adjustable in selectable steps 0.1 dB, 1dB, 2 dB, 6 dB,
and locked.
Gates
Two fully independent gates offer a range of measurement options for signal
height or distance using peak triggering.
The echo-to-echo mode allows accurate gate positioning for signals which are
extremely close together.
Gate Start: Variable over entire displayed range
Gate Width: Variable from Gate Start to end of displayed range
Gate Height: Variable from 0 to 99% Full Screen Height
Alarms: Threshold positive/negative
Memory
Memory of 500 set files to store calibration set-ups.
Memory of 1000 test report files to store. All the files can be stored, recalled
and cleared. Connectting U-Disk,the unit can get unlimited memory.
Functions
Semiautomatic two point calibration: Automated calibration of transducer
zero offset ,angle, front end and/or material velocity
Flaw Locating:Live display Sound-path, Projection (surface distance), Depth,
Amplitude,
Flaw sizing: Automatic flaw sizing using AVG/AVG or DAC, speeds reporting
of defect acceptance or rejection.
Digital Readout and Trig. Function: Thickness/Depth can be displayed in
digital readout when using a normal probe and Peam path, Surface Distance
and Depth are directly displayed when angle probe is in use.
Both the DAC and the AVG method of amplitude evaluation are available.
Blue tooth connection to mobile device to generate and print detection reports on-site.
(Option)
AWS D1.1.
Curved Surface Correction feature
Crack Height Measure function
Magnify gate:spreading of the gate range over the entire screen width
Auto-gain function
Envelope: Simultaneous display of live A-scan at 70 Hz update rate and
envelope of A-scan display
Peak Hold: Compare frozen peak waveforms to live A-Scans to easily
interpret test results.
A Scan Freeze:Display freeze holds waveform and sound path data
Capture mark.
Echo color within the Gate.
5
Real Time Clock
The instrument clock keeps running tracking the time.
Communication
High speed USB3.0 port. Connectting U-Disk , the unit can get unlimited
memory.
Blue tooth connection to mobile device to generate and print detection reports on-site.
6
Battery
Internal rechargeable Li-ion battery pack rated 7.2V at 8000 mAh
15 hours nominal operating time depending on display brightness
4 hours typical recharge time
Knob
Operating adjustments are easily and quickly made using the rotary knob.
1.2 Specifications
Range: 0 to 10000 mm, at steel velocity
Material Velocity: 100 to 20000m/s
Display Delay: -20 to 3400 µs
Probe Delay/Zero Offset : 0 to 99.99µs
Sensitivity: 110 dB max in selectable resolution 0.1, 1.0, 2.0, 6.0 dB and
locked.
Test Modes: Pulse echo, dual element and thru-transmission
Pulse Repetition Frequency ranges from 10 Hz to 2000 Hz
Pulse Energy: 70V、150V、300V、400、500V. selectable
Bandwidth (amplifier bandpass ): 0.5 to 15 MHz
Gate Monitors: Two independent gates controllable over entire sweep range
Rectification: Positive half wave, negative half wave, full wave, RF
System Linearity: Horizontal: +/-0.2% FSW, Vertical: 0.25% FSH, Amplifier
Accuracy +/-1 dB.
Reject (suppression): 0 to 80% full screen height
Units: Inch or millimeter
Transducer Connections: BNC or LEMO
Power Requirements: AC Mains 100-240 VAC, 50-60 Hz
Operating Temperature: -10℃ to 50℃
Storage Temperature: -30℃ to 50℃
1.3 Base Instrument Package
Portable Ultrasonic Flaw Detector with multi-color LCD Display
Straight Beam Transducer (2.5 MHz, Φ20)
Angle Beam Transducer (4 MHz, 8 mm×9 mm, 60°)
Interconnect Cable for the transducer (Q9-C5,or optional C9- C5)
Rechargeable Li-Ion Battery Package, 8.0 amp hour
Power supply/charger unit
Supporting pillar
U-Disk :
1.Operating Manual in
2.English Datapro Software
7
2 Understanding the Keypad, Menu System and Display
8
2.2 Screen Display
The instrument displays are designed to be easy to interpret.
9
Submenus
Main menus
10
The main menus are individual page.
The Submenus are displayed at the right of the screen. The selected menu or
submenu is highlighted. We can use cancel key to show it and hide it.
The status icons at the top-right corner and the A-Scan area shows the system
status. The following status icons will be shown:
Probe types icon:
Straight beam probe : ,
Angle beam probe : ,
Dual element probe : ,
Through-transmit mode:
Creeping probe: ,
Surface probe:
11
2.3 Keys and Rotary Knob Features
The instrument is designed to give the user quick access to all of the
instrument’s functions. Its easy-to-use menu system allows any function to be
accessed with no more than three key presses.
To access any function:
Press menu key select a functional group. The menus across the right of the
display will immediately be replaced with the submenus contained in the
selected menu.
Use Rotary Knob or Press and to move the cursor to select the desired
function.
Change the value listed in the dialog box with the rotary knob. Some values
can also be adjusted with pressing right and left key.
Special keys such as Gain, Gate, Range, Probe, DAC/AVG, Freeze, Auto-
Gain, Envelop, Peak Hold ,Mem and SET are grouped together for easy
thumb control. This design allows direct access to important instrument set-
up parameters and provides easy and fast operating in difficult inspection
environments
Rotary knob makes it convenient for handling device parameters and functions,
It support three operations: turn clockwise, turn counter-clockwise and single
press.
Turn clockwise and Turn counter-clockwise:
Move the gate in full screen mode.
Adjust the selected parameter in dialog box,
change the selection in submenus mode and main menus mode.
Single press: changes mode between coarse and fine parameter adjustment or
changes item selection.
2.4 Menus and Functions
The instrument menu system consists of several menus, submenus, and
functions. It allows the operator to select and adjust various features and
instrument settings.
Available menus are accessed via the main Menu. Note that the menus
visible on your particular instrument depend on which options are installed.
Each menu contains several submenus.
Menus and submenus are selected by pressing the key enter.
When a submenu is selected, the functions contained in that submenu are
listed in the Function Bar down the right-hand side of the display screen.
12
Functions are then selected by knob or pressing and .Press knob to
bring up dialog box of the function.
Turning the Function Knob will change the value shown in the selected
function’s box.
Note that some functions, like RANGE, have coarse and fine adjustment modes.
There are three steps in the fine adjustment mode. Coarse and fine modes are
selected by pressing right and left key. When the function is in coarse
adjustment mode, turning the function knob will produce large changes in the
selected function’s value. When the function is in fine adjustment mode, rotating
the function knob will change the value by smaller amounts.
13
3 Initial Start-up
Mode A: Cancel key and function keys is pressed and display mode will change to B.
Menu key is pressed and display mode will change to C.
Mode B: Cancel key is pressed and will change to A. Menu key is pressed,
Display mode will change to C.
Mode C: Cancel key is pressed and display mode will change to B.
Enter key is pressed and display mode will change to B.
16
Calibration requires the use of calibrated standard made of the same material as
the test piece. Prior to calibrating the instrument/probe combination, the A-Scan
display-screen range (the material thickness value represented by the full
horizontal width of the screen) will normally be set to a value equal to or slightly
larger than the calibrated standard.
- Activate the range menu by pressing the range key.
- Press knob to bring up dialog box of range.
- To change the range, turn the knob.
- The display’s horizontal range will remain as set.
17
D-DELAY range: -20~3400 µs
- Activate the PROBE submenu (located in the CAL menu) by pressing the
menu key below it.
- Select the function item titled PROBE.
- To change the probe mode, turn the knob. Each available probe mode is
represented by an icon on screen display whenever that probe mode is
indicated.
- The probe mode will be set to the last one displayed.
4.7 Selecting a Rectification Mode
You can select the rectification mode of the echo pulses according to your
application in the function RECTIFY (P/R→RECTIFY).
Rectification affects the orientation of the A-Scan on the display screen. The A-
Scan represents the pulse (echo) that’s returned from the material being tested
to the instrument. The series of echoes looks like the Radio Frequency (RF)
signal.
Full-Wave Rectification
combines the positive and
negative rectified signals
together, and displays both of
them in a positive orientation.
Full-wave rectification is
recommended for most
inspections.
The offset gain is for the purpose of flaw evaluation. The DAC/AVG curves
remain unchanged when adjusting the offset gain.
Sensitivity correction gain: Compensate for the transfer losses in the material
under test. This correction gain is necessary if test object and reference block
have different surface qualities. You have to find out the adjustment value for the
compensation of transfer losses by experiments. The gain is varied accordingly
in this correction, while the curve line remains unchanged. To set the sensitivity
correction gain, select the function T-CORR and then adjust its value according
to the experiment.
When DAC is active, adjusting the basic gain setting will result in an equal
adjustment of the DAC curve position. To increase the instrument gain without
changing the DAC curve position, use the T-CORR function to compensate for
varying couping/surface conditions.
20
peak echo inside the currently selected gate to a target height.
- Set the target echo height using the AUTO function located in the GAIN
submenu.
AUTO
- Press GAIN to start the Auto Gain function. “AUTO-XX%” will be shown
on the screen.
- The Auto Gain function will end when the echo height reaches the target
value. Or you can stop the function by pressing again.
AUTO
GAIN
Positioning Gates
Use the following procedures to set the vertical and horizontal position of the A
and B-Gates. Remember that gate position has the following effects on
instrument performance:
A-Scan echoes on the right side of the display screen represent features that
occur at a greater depth from the test-material surface than those on the left
of the display screen. Therefore, moving a gate to the right means that the
gate is evaluating a deeper portion of the test material.
A wider gate will simply span the equivalent of more test-material depth
21
Increasing the vertical height (called threshold) of a gate means that only
reflected signals of sufficiently large amplitude will cross the gate.
Setting a Gate’s Starting Point
You can fix the starting point of the gates A or B within the display range.
- Activate the GATE submenu (located in the main menu).
- Select the gate START function and adjust the starting point by turning
the knob. Increasing and decreasing the value of the starting point moves
the gate to the right and left, respectively.
- The gate starting point will remain as set, even when width adjustment is
made.
You can also set the selected gate’s starting point using the knob in full
screen mode.
You can also adjust gate’s width by simply pressing the and . Note that
the instrument will only search the echo whose peak crosses the selected gate
in this function.
When a gate alarm is activated, one or more of the following will occur:
An alarm indication light on the front of the instrument will illuminate
An audible alarm (HORN) will sound
Note that the detection method chosen will be reflected by a small icon. This icon
is displayed on the status bar containing the measured reading.
23
4.13 Setting the Damping Level
This function serves for matching the probe. You can use it to adjust the damping
of the probe’s oscillating circuit and to consequently change the height, width
and resolution of the echo display.
- Activate the Probe submenu.
- Select the function item titled DAMPING. And bring up dialog box.
- To change the specified damping level and optimize the A-Scan signal
appearance, continue pressing key or turn the knob. You’ll note that
the following damping levels are available: 50 Ω,100Ω,150Ω,400Ω.
- The damping level will be set to the one last displayed.
4.14 Setting the Pulse Energy Level
Use the function ENERGY to set the pulser voltage. The relative energy with
which the pulser fires can be set to70V,120V, 250V, 300V, 350V, 400V or500V.
The setting 350V and 400V is recommended for most inspections. HIGH is used
for inspections in which maximum sensitivity is import, e.g. for the detection of
small flaws. Choose the setting LOW for broadband probes or if narrow echoes
are required (better lateral resolution).
To set the pulser energy
- Activate the P/R submenu by pressing the menu key.
- Select the function item titled ENERGY. Press knob to bring up the dialog box.
- Select the energy level by turning the knob.
4.15 Setting the Pulse Width
Use the function P-WIDTH to set the pulser signal width to match the installed
probe.
Adjustment range: 30~510 ns
4.16 Adjusting the Pulse Repetition Frequency (PRF)
The pulse repetition frequency indicates the number of times an initial pulse is
triggered per second. You can determine whether you need the highest possible
PRF value, or whether you are satisfied with a low value.
The PRF value ranges from 20 Hz to 3000 KHz. The larger your work piece, the
smaller PRF values are needed in order to avoid phantom echoes. In the case of
smaller PRF values, however, the A-Scan update rate becomes lower, for this
reason, high values are required if a work piece should be scanned fast.
The best way to determine the suitable PRF value is by experimenting: start from
the highest step and reduce the value until there are no more phantom echoes.
To set the PRF level.
24
4.17 Specifying the Probe Frequency
In this function, you can specify the probe frequency according to the
frequency of your probe. The instrument will automatically utilize a built-in filter
to match the probe frequency.
- Activate the PROBE submenu by pressing the probe key .
- Select the function item titled FREQ.
- To change the specified frequency, turn the knob.
- The probe frequency will be set to the last one displayed.
4.18 Specifying the Piezo Crystal Size
Use P_SIZE function to set the Piezo Crystal Size of your probe. This value is
required when programming the AVG.
4.19 Setting the Probe X-Value
25
Before Magnifying After Magnifying
26
4.26 Scale Setting
27
THICKNESS
THICK: Use the THICK function to set the material’s wall thickness. This value is
required for the automatic calculation of the real depth.
Adjustment range: 0~5000 mm
Diameter (Outside diameter of the test object): You will need the CSC function
for tests on circular curved surfaces, for example when testing longitudinally
welded tubes. In order to make the instrument carry out the corresponding
correction of (reduced) projection distance and depth, you should enter the
outside diameter of your test object in this function.
Adjustment range: 0~5000 mm
Select the function SCALE. Then use the knob to set the required display mode..
4.27 Fill function
The function FILL toggles between the filled and the normal echo display mode.
The filled echo display mode improves the echo perceptibility due to the strong
contrast, especially in cases where workpieces are scanned more quickly.
28
4.29 Skin Selection
The SKIN SEL value determines the color of the background, the gate and the
DAC/AVG curves. All color schemes are recommended for indoor operation
while S1 and S4 are best suited for outdoor operation.
4.30 Key Sound
The key sound can be turned on or off using the KEY SOUND function located in
the CFG submenu (CFG→→KEY SOUND).
4.31 Date and Time Setting
For a correct documentation you should always make sure that you are using the
correct date and time setting.
-
Time is displayed in Hour-Minute format. And the time setting procedure is
similar to that of the date.
Once set, the internal clock of the instrument will maintain the current date and
time.
4.32 Resetting the Instrument
In case the instrument can no longer be operated, or you need to make a basic
initialization (factory setting), you can reset the instrument to original.
The instrument can be reset by the SET RECALL function. Load a setting with
default set. Any empty set is a setting with default set.
29
30
during system reset. And the instrument settings will be reset to default.
But the saved settings and the reports will be reserved.
NOTE:
The effects of resetting the instrument may not be reversed.
No key action should be performed during resetting process.
31
5 Calibration and Measurement
Before working with the instrument, you have to calibrate the instrument: you
have to adjust the material velocity and display range and allow for the probe
delay depending on the material and dimensions of the test object.
To ensure a safe and proper operation of the instrument, it is necessary that the
operator be adequately trained in the field of ultrasonic testing technology.
Below you will find some examples of common calibration methods for certain
test tasks.
5.1 Calibration with Straight- Beam and Angle-Beam Probes
Case A: With Known Material Velocity
Calibration process
- Set the known material velocity in VEL function.
- Couple the probe to the calibration block.
- Set the required display range in RANGE. The calibration echo must be
displayed on the screen.
- Position the gate on one of the calibration echoes until the sound path of
the echo is indicated in the measurement line.
- After this, change the adjustment of the function P-DELAY until the
correct sound path for the selected calibration echo is indicated in the
measurement line.
Example:
You are carrying out the calibration with a straight probe for the calibration range
of 300 mm via the function group BASE using the calibration block DB-P
(thickness 225 mm) which is laid flatwise.
VEL.
- Set the gate so that it is positioned on the first
calibration echo (from 225 mm).
- Read the sound path on the status line. If this
value is not equal to 225 mm, change the
adjustment for the function P-DELAY until it is at
225 mm.
- Save the calibration result to current set file.
This completes the calibration of the instrument to the material velocity of 5920
m/s with a calibration range of 300 mm for the straight-beam probe used.
Case B: With Unknown Material Velocity(Straight- Beam Probes)
Use the semiautomatic calibration function of the instrument via the function
CAL PROBE for this calibration case.
32
The Thickness of two calibration must be entered.. The instrument will then carry out a
plausibility check, calculate the material velocity and the probe delay, and automatically
set the parameters using the calculation result.
Calibration Process
- Set PROBE type to STRAIGHT.
- The Thickness of two calibration must be entered.
33
- Show calibration result.
The calibration is confirmed by the result message. The instrument will now
automatically determine the sound velocity and the probe delay and set the
corresponding functions accordingly. The value of the function P-DELAY will be
set to the correct value.
If the instrument is not able to carry out any valid calibration on the basis of the
input values and the echoes recorded, a corresponding error message is
displayed. In that case, please check the values of your calibration lines and
repeat the process of recording the calibration echoes.
Case C: With Unknown Material Velocity(Angle- Beam Probes)
For this calibration case, two methods is supported (TWO HOLE and V2).
Calibration Process(TWO HOLE)
- Set PROBE type to Angle
34
- The depth of two hole must be entered.
35
- Position the other gate on the second calibration echo.
36
5.2 Calibration with Dual-Element (TR) Probes
Dual-element (TR) probes are especially used for wall thickness measurement.
37
The following peculiarities must be taken into account when using these probes:
Echo Flank. Most dual-element (TR) probes have a roof angle (transducer
elements with inclined orientation toward the test surface). This causes mode
conversions both at beam index (sound entry into the material) and at the
reflection from the backwall, which can result in very jagged echoes.
V-Path Error. Dual-element (TR) probes produce a v-shaped sound path from
the pulser via the reflection from the backwall to the receiver element. This so-
called “V-path error” affects the measuring accuracy. You should therefore
choose two wall thicknesses that cover the expected thickness measurement
range for the calibration. In this way, the V-path error can be corrected to the
greatest possible extend.
Higher Material Velocity. Due to the V-path error, a higher material velocity
than that of the test material is given during calibration, especially with small
thicknesses. This is typical of dual-element (TR) probes and serves for
compensation of V-path error.
With small wall thicknesses, the above-described effect leads to an echo
amplitude drop which has to be especially taken into account with thicknesses
less than 2 mm.
A stepped reference block having different wall thicknesses is required for
calibration. The wall thicknesses must be selected so that they cover the
expected readings.
Calibration Process
We recommend to use the semiautomatic calibration function for the calibration
with T/R probes.
• Set PROBE type to DAUL S.
• The Thickness of two calibration must be entered.
• Position one gate on the first calibration echo.
• Position one gate on thesecond calibration echo.
• Show calibration result.
If necessary, check the calibration on one or several known calibration lines, e.g.
using the stepped reference block.
38
When displayed, the DAC curve visually represents a line of constant reflector
peaks over a range of material depths. A new feature of the instrument is a
multiple-curve option that displays three dB offset DAC curves simultaneously.
Each curve represents constant reflector size at varying material depth.
Remember that in DAC mode, the only deviation from traditional display and
operation is the appearance of the DAC curve. All A-Scan echoes are displayed
at their non-compensated height.
A DAC curve is programmed using a series of same-reflector echoes at various
depths covering the range of depths to be inspected in the test material.
Because near field and beam spread vary according to transducer size and
frequency, and materials vary in attenuation and velocity, DAC must be
programmed differently for different applications. A DAC curve can be based on
up to 16 data points (material depths).These points are recorded from the DAC
menu as described below.
Recording the DAC Curve
DAC Curve points are typically taken from a standard with equally sized
reflectors (holes) located at various material depths. The primary echo from each
of these points (for up to a total of 16 echoes) are recorded. When DAC is active,
the instrument displays a curve that represents echo peaks for constant
reflectors at varying material depth.
39
- Select a standard for DAC.
When operating with DAC turned on, three DAC curves including RL, SL and EL
are typically displayed. The three curves can be offset from the original DAC
curve (ML) by a user-inputted amount. The unit support one to six DAC
curves,when we
Select a standard, The offset of
DAC curves is fixed. In custom
Mode, we can adjust offset freely.
The offset range from -50dB to
50dB.
40
- Start building DAC curve.Input max depth of reflactor.
41
- Input the correct gain.
-
- Finished.
42
Editing DAC Curve
After reference points are recorded, their values may be manually adjusted, or
some points may be manually deleted. To edit or delete points:
- With the DAC menu accessed, select EDIT submenu.
- Select EDIT function. Continue to press enter key to start the EDIT
process. When DAC EDIT process is triggered, “DAC” and a digit
number appear indicating the point index on the top right corner of the
display.
43
- When complete select FINISH and press key and follow the
on-screen prompts to finish the editing.
Specifying the DAC Reference Curve
Once the DAC curve is recorded and displayed, echoes are automatically
compared to the size reference curve, which can be specified in the SIZE REF
selection. The options include RL, SL, EL, and GL.
The dB equivalent height of the signal above or below the corresponding DAC
Reference Curve amplitude is displayed as a comparison result.
TCG curve
When we have a DAC curve, we can use open TCG function to building a tcg
curve automatically.
44
5.4 Measuring with DGS/AVG
Using the AVG (the same as DGS) function, you can compare the reflecting
power of a natural flaw in the test object with that of a theoretical flaw (circular
disk-shaped equivalent reflector) at the same depth.
Note: You are comparing the reflecting power of a natural flaw with that of a
theoretical flaw. No definite conclusions may be drawn on the natural flaw
(roughness, inclined position, etc.).
The so-called AVG diagram forms the basis for this comparison of the reflecting
power. This diagram consists of a set of curves showing the correlation of three
influencing variables:
Distance between the probe and circular disk-shaped equivalent reflector
Difference in gain between various large circular disk-shaped equivalent
reflectors and an infinitely large backwall
Size of the circular disk-shaped equivalent reflector. The influencing variable
always remains constant for one curve of the set of curves.
The advantage of the AVG method lies in the fact that you can carry out
reproducible evaluations of small discontinuities. The reproducibility is most of all
important, for example, whenever you aim to carry out an acceptance test.
Apart from the influencing variables already mentioned, there are other factors
determining the curve shape: sound attenuation, transfer losses, amplitude
correction value, probe and etc.
The following probe parameters affect the curve shape:
Element or crystal diameter
Frequency
Delay length
Delay velocity
You can adjust these parameters on the instrument in such a way that you can
use the AVG method with many different probes and on different materials.
Note: Before setting the AVG function, the instrument must first be calibrated
because all functions affecting the AVG evaluation mode (VEL, P-DELAY,
DAMPING, ENERGY, P-WIDTH, FREQ, DIAMETER, RECTIFY) can no longer
be changed after the reference echo has been recorded.
Settings for the AVG measurement
Before using the AVG feature to evaluate reflectors in test pieces, the
characteristics of the attached probe must be specified, certain characteristics of
the reference standard must be input, and a reference echo must be stored.
45
Record the Reference Echo that Defines the AVG Curve
Prior to generating the AVG curve, a test standard with a known reflector must
be used to define a reference point. Acceptable test standards include these
reference types:
BW – Backwall echo with reference defect size defined as infinity
SDH – Side Drilled Hole with a reference defect size defined as the hole’s
diameter.
FBH – Flat Bottom Hole with a reference defect size equal to the hole’s facial
diameter.
46
47
Once a reference echo has been recorded, the AVG curves (the AVG-UL, the
AVG-ML and the AVG-DL) are automatically displayed using default settings.
Creating AVG Offset Curves
When operating with AVG turned on, three AVG curves including the AVG-UL,
the AVG-ML and the AVG-DL are typically displayed. The three curves can be
offset by user-inputted equivalent flaw size values in the AVG-UL, AVG-ML and
AVG-DL selection. The offset values will usually depend on the largest
acceptable flaw size.
Evaluating Test Samples in AVG Mode
Once the AVG curve is recorded and displayed, echoes are automatically
compared to the AVG curve and evaluated based on the recorded reference.
The result is shown as ERS (Equivalent Reflector Size), which evaluates the
reflected echo and calculates the equivalent reflector size.
Before activating the AWS D1.1 weld rating feature, be sure that all instrument
settings are properly adjusted for the specific measurement application. Then
access the AWS D1.1 submenu and follow this procedure:
- Step 1: Apply couplant and couple the probe to a suitable reference test
48
standard.
- Step 2: Ensure that the A-Gate is positioned over the desired echo. Then
adjust the gain until the peak of the desired reference echo reaches 50%
of full screen height (FSH). Note that if the echo’s peak amplitude does
not fall between 47.5% and 52.5%, the inputted point will not be
accepted.
- Step 3: Select the REFRNCE function and then press to define the
reference dB level.
- Step 4: To evaluate a reflector in a test piece, couple the probe to the test
piece and adjust the gate position so that it is over the desired echo.
- Step 5: Adjust the gain until the peak of the test-piece’s echo reaches
50% of full screen height (FSH). Note that if the echo’s peak amplitude
does not fall between 47.5% and 52.5%, the inputted point will not be
accepted.
- Step 6: Return to the AWS D1.1 submenu and select the INDICAT
function (then press again to confirm). Then press . This will record
the current dB setting and will automatically calculate and display values
for the AWS variables ATTEN and RATING. RATING can then be applied
to the appropriate acceptance criteria fond in AWS D1.1.
49
5.7 G-COLOR
Activated the function, the wave in the gate will have the gate color .It is in toolkit function
groups.
50
5.8 Crack Height Measuring Feature
This feature allows measuring the height of a crack found inside the workpiece.
Before starting to use the Crack Height Measure function, the instrument and the
probe must be correctly calibrated.
Note that PEAK HOLD and ENVELOPE can not function simultaneously.
-
Also you can repeatedly press to turn on/off the PEAK HOLD function.
PEAK
HOLD
Note that PEAK HOLD and ENVELOPE can not function simultaneously.
52
5.11 B-Scan Feature(Option)
In B-SCAN mode, the
instrument displays a graphical
cross-section of the workpiece.
Use the following procedure to
switch to B-SCAN mode.
- Select the BSCAN
function
.
- Switch BSCAN to ON by
means of turning the
knob.
53
Report Files
Wave files are used to store test report files, along with the instrument settings.
Data in the wave files can be stored, recalled and erased.
The unit can save 1000 reports, the report can be export to U-disk any time.
SET Files
SET files (CH001~CH500) are used to store a specific instrument setup
configuration. This means that whenever you recall a stored channel file, your
instrument is again set up exactly the same as it was at the moment when the
file was stored. This makes each one of your tests reproducible. You will find the
following functions:
54
55
6 Maintenance and Care
6.3 Maintenance
The instrument requires basically no maintenance.
Attention:
Repair work may only be carried out by authorized Service staff of us.
6.4 Warranty
When used in accordance with the manufacturer’s written instructions and under
normal operating conditions, the instrument is conditionally guaranteed to be
free from defects in material and workmanship for a period of two years from
date of shipment.
This warranty shall not apply to equipment subjected to misuse or abuse,
improper installation, alteration, neglect, or accident. Excluded from this
warranty are expendable items such as transducers, interconnecting cables, and
batteries.
This warranty is limited to the original purchaser and is not transferable. No other
warranty, expressed or implied, is made.
56
Appendix
Note:
The battery pack should be charged over a period of time, even not used.
The instrument can only work with the specially designed battery pack and
power adapter (charger) supplied by us. Working with others may result in
damage of the instrument, battery leakage, fire or even explosion.
57
Appendix B: Menu Structure
50
51