ARCADIS Avantic
SP
Startup
System
© Siemens, 2005
10047655
© Siemens, 2005
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2005
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2005“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General safety information (in existing documents) . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Laser light localizer option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Validity of these instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required measuring equipment and aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Observe:
¹ The product-specific safety information in these
instructions and the installation instructions for the
ARCADIS Avantic Laser Light Localizer Generator,
print no. SPR2-330.812.02...
¹ The general safety information in TD00-000.860.01...,
and
¹ The safety information in accordance with ARTD Part 2.
After all work has been completed and all covers have been installed, perform the protec-
tive conductor test according to ARTD-002.731.17... The protective conductor resistance
may not exceed 0.2 Ohms.
Functional Test 0
C-arm position
• Set the C-arm to the 0 degree angulation position.
• Set the C-arm to the 0 degree orbital position.
¹ The C-arm is perpendicular, POWERPHOS below, I.I. above.
Fig. 1: Laser_unit
NOTICE Important!
¹ Do not turn the lenses of the laser diodes.
• Turn the diodes in the holder until the laser light fan beams are reflected on the I.I. as
shown in (Fig. 2 / p. 8). The laser light fan beams begin at (X /Fig. 2 / p. 8). This setting
ensures that the laser light fan beams cross near the laser light localizer.
• Retighten the screws (3/Fig. 1 / p. 7).
• Turn the diodes in the holders until the laser light fan beams are reflected on the I.I. as
shown in (Fig. 4 / p. 9). The laser light fan beams begin at (X /Fig. 4 / p. 9). The setting
ensures that the laser light fan beams cross near the laser light localizer.
• Retighten the screws (3/Fig. 3 / p. 9)
NOTE After placing the POWERPHOS cover back on, the fan beams
of the laser light localizer are shadowed slightly at the (1 und
2 /Fig. 5 / p. 10) positions (I.I. edge). This is a design issue
and cannot be corrected. However, it does not negatively
impact use of the laser light localizer in normal operation.
NOTE After placing the POWERPHOS cover back on, the fan beams
of the laser light localizer are shadowed slightly at the (1 und
2 /Fig. 5 / p. 10) positions (I.I. edge). This is a design issue
and cannot be corrected. However, it does not negatively
impact use of the laser light localizer in normal operation.
Fig. 6: Stickers_other_countries
• Visually inspect the identification label, warning logo type "Laser class 2", DHHS certifi-
cation label, and the laser aperture label as in (Fig. 6 / p. 12) and (Fig. 7 / p. 13).
Check the identification and warning labels (other than the USA) 0
Fig. 8: Label_not_usa
• Check warning label according to (Fig. 8 / p. 14).
Other countries: Check the warning label according to IEC 825 (text) in the local lan-
guage, the warning label according to IEC 825 (symbol), and the
identification label.
Great Britain: Check the warning label according to IEC 825 (text) in the local lan-
guage, the warning label according to IEC 825 (symbol), and the
identification label. Also check laser aperture labels (Fig. 7 / p. 13).
• Install the POWERPHOS cover and secure it with the cover screws. Reseal the POW-
ERPHOS cover with silicone sealing compound.
NOTE The POWERPHOS cover slightly shadows the laser light fan
beams at the I.I. edge. This is a design issue and cannot be
corrected. However, it does not negatively impact use of the
laser light localizer in normal operation.
Storing documents 0
Functional Test 0
• Perform the protective conductor resistance test for the entire ARCADIS Avantic sys-
tem.
¹ The protective conductor resistance must not exceed 0.2 ohms.
Mechanical safety 0
• Ensure that the laser light localizer is labeled as described in this document. Replace
any labels that are damaged.
• Check the I.I. ring as well as the removable grid for mechanical damage.
• Check the laser beam exit window at the tube cover for damage.
• Ensure that the laser light fan beams are set as described in this document.
• Check the function of the "Laser light localizer on/off" button on the base unit control
console.
• Switch the laser light localizer back on.
¹ The laser diodes must shut off automatically after 5 minutes.