SPR2-330 815 02 02 02

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 18

w

rth
ü
W
s
n
m
ie
S

ARCADIS Avantic
SP

Startup
System

Laser Light Localizer Generator

© Siemens, 2005

10047655

© Siemens, 2005
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: SPR2-330.815.02.02.02 English


Replaces: SPR2-330.815.02.01.02 Doc. Gen. Date: 10.09
CS SD 24
N/A

2005
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2005“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

ARCADIS Avantic SPR2-330.815.02.02.02 Page 2 of 18 © Siemens, 2005


10.09 CS SD 24
Table of Contents 3
0Table of Contents

1 _______ Safety and General Information ____________________________________ 4

Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General safety information (in existing documents) . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Laser light localizer option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Validity of these instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required measuring equipment and aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 _______ Startup ________________________________________________________ 6

Configuration and functional test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6


ARCADIS Avantic configuration: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Adjusting and checking the laser light fan beams. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Adjusting the laser light fan beams to the I.I. grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Checking the laser light beam at the image intensifier . . . . . . . . . . . . . . . . . . . . . . . 10
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Check the identification and warning labels for USA . . . . . . . . . . . . . . . . . . . . . . . . 12
Check the identification and warning labels (other than the USA) . . . . . . . . . . . . . . 14
Installing the covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Storing documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

3 _______ Maintenance ___________________________________________________ 16

Maintenance activities on the laser light localizer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16


Mechanical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operating values, inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

4 _______ Changes to Previous Version _____________________________________ 17

© Siemens, 2005 SPR2-330.815.02.02.02 Page 3 of 18 ARCADIS Avantic


10.09 CS SD 24
4 Safety and General Information
1-
Safety information
1Safety and General Information

General safety information (in existing documents) 0

WARNING Danger of injury, death, or material damage.


Non-compliance can lead to death, injury, or material dam-
age.

Observe:
¹ The product-specific safety information in these
instructions and the installation instructions for the
ARCADIS Avantic Laser Light Localizer Generator,
print no. SPR2-330.812.02...
¹ The general safety information in TD00-000.860.01...,
and
¹ The safety information in accordance with ARTD Part 2.

Laser light localizer option 0

CAUTION Laser emissions!


This product contains class 2 lasers. (USA: Laser class 2)
Disregarding safety precautions can lead to bodily injury,
especially to the retina of the eye, resulting in irreversible
damage to vision.
¹ Observe the safety information in ARTD-002.731.03...
When working with the laser light localizer, do not look
directly into the laser beam.

NOTE Laser emissions!


There is no direct hazard to the eye (blinking reflex). Never-
theless, do not look directly into the laser beam.

Protective conductor resistance test 0

After all work has been completed and all covers have been installed, perform the protec-
tive conductor test according to ARTD-002.731.17... The protective conductor resistance
may not exceed 0.2 Ohms.

ARCADIS Avantic SPR2-330.815.02.02.02 Page 4 of 18 © Siemens, 2005


10.09 CS SD 24
Safety and General Information 5

General information 1.1

Validity of these instructions 0

• These adjustment instructions pertain to the startup of a laser light localizer at an


ARCADIS Avantic system.
• They also serve as adjustment instructions for laser light localizer service and mainte-
nance.
¹ Therefore, keep them in a safe place.

Required measuring equipment and aids 0

• Tool kit (including a set of metric Allen keys)


• Protective conductor meter, e.g. Bender UNIMET 1000, material number 05138727.

© Siemens, 2005 SPR2-330.815.02.02.02 Page 5 of 18 ARCADIS Avantic


10.09 CS SD 24
6 Startup
2-
Configuration and functional test
2Startup

ARCADIS Avantic configuration: 0

• Switch on the system and wait for it to boot up.


• Open local service and log in.
• Click menus <Main System>-<Configuration>-<Options>.
• Set "Laser Existing" to <yes> and click <Save>.
• After the change, save a new backup of the parameters on the system backup CD.

Functional Test 0

The ARCADIS Avantic system is operational.


• Switch on the laser light localizer via the laser light localizer button on the base unit con-
trol console.
¹ The two laser diodes light up.
• Switch off the laser light localizer again with the laser light localizer button.
¹ The two laser diodes go off.
• Switch on the laser light localizer again with the laser light localizer button and leave it
on for 5 minutes.
¹ The two laser diodes remain on for approx. 5 minutes, and then switch off auto-
matically.

ARCADIS Avantic SPR2-330.815.02.02.02 Page 6 of 18 © Siemens, 2005


10.09 CS SD 24
Startup 7

Adjusting and checking the laser light fan beams 2.1

Adjusting the laser light fan beams to the I.I. grid 0

C-arm position
• Set the C-arm to the 0 degree angulation position.
• Set the C-arm to the 0 degree orbital position.
¹ The C-arm is perpendicular, POWERPHOS below, I.I. above.

Laser light localizer, design up until October 2009:

Fig. 1: Laser_unit

© Siemens, 2005 SPR2-330.815.02.02.02 Page 7 of 18 ARCADIS Avantic


10.09 CS SD 24
8 Startup

Fig. 2: Laser light fan beams


• Move the excenter (1 und 2 /Fig. 1 / p. 7) to the "zero position".
• Loosen the screws (3/Fig. 1 / p. 7) until the laser diodes can be turned in their holders.

NOTICE Important!
¹ Do not turn the lenses of the laser diodes.

• Turn the diodes in the holder until the laser light fan beams are reflected on the I.I. as
shown in (Fig. 2 / p. 8). The laser light fan beams begin at (X /Fig. 2 / p. 8). This setting
ensures that the laser light fan beams cross near the laser light localizer.
• Retighten the screws (3/Fig. 1 / p. 7).

ARCADIS Avantic SPR2-330.815.02.02.02 Page 8 of 18 © Siemens, 2005


10.09 CS SD 24
Startup 9

Laser light localizer, design beginning in November 2009:

Fig. 3: Laser light localizer adjustment


Pos. 1 Setting for the transverse laser position
Pos. 2 Setting for the longitudinal laser position
Pos. 3 Laser diode holders

Fig. 4: Laser light fan beams


• Move the excenter (1 und 2 /Fig. 3 / p. 9) to the center position.
• Loosen the screws (3/Fig. 3 / p. 9) until the laser diodes can be turned in the holders.

NOTE Do not turn the lenses of the laser diodes.

© Siemens, 2005 SPR2-330.815.02.02.02 Page 9 of 18 ARCADIS Avantic


10.09 CS SD 24
10 Startup

• Turn the diodes in the holders until the laser light fan beams are reflected on the I.I. as
shown in (Fig. 4 / p. 9). The laser light fan beams begin at (X /Fig. 4 / p. 9). The setting
ensures that the laser light fan beams cross near the laser light localizer.
• Retighten the screws (3/Fig. 3 / p. 9)

Checking the laser light beam at the image intensifier 0

This check is to be performed according to (Fig. 5 / p. 10).


¹ The laser light fan beams have to intersect at the center of the mounting screw.
¹ The laser light fan beams should begin at the (1 und 2 /Fig. 5 / p. 10) positions.

NOTE After placing the POWERPHOS cover back on, the fan beams
of the laser light localizer are shadowed slightly at the (1 und
2 /Fig. 5 / p. 10) positions (I.I. edge). This is a design issue
and cannot be corrected. However, it does not negatively
impact use of the laser light localizer in normal operation.

Fig. 5: Laser light_I.I.

ARCADIS Avantic SPR2-330.815.02.02.02 Page 10 of 18 © Siemens, 2005


10.09 CS SD 24
Startup 11

NOTE After placing the POWERPHOS cover back on, the fan beams
of the laser light localizer are shadowed slightly at the (1 und
2 /Fig. 5 / p. 10) positions (I.I. edge). This is a design issue
and cannot be corrected. However, it does not negatively
impact use of the laser light localizer in normal operation.

© Siemens, 2005 SPR2-330.815.02.02.02 Page 11 of 18 ARCADIS Avantic


10.09 CS SD 24
12 Startup

Final steps 2.2

Check the identification and warning labels for USA 0

Fig. 6: Stickers_other_countries
• Visually inspect the identification label, warning logo type "Laser class 2", DHHS certifi-
cation label, and the laser aperture label as in (Fig. 6 / p. 12) and (Fig. 7 / p. 13).

ARCADIS Avantic SPR2-330.815.02.02.02 Page 12 of 18 © Siemens, 2005


10.09 CS SD 24
Startup 13

Fig. 7: Sticker_laser aperture

© Siemens, 2005 SPR2-330.815.02.02.02 Page 13 of 18 ARCADIS Avantic


10.09 CS SD 24
14 Startup

Check the identification and warning labels (other than the USA) 0

Fig. 8: Label_not_usa
• Check warning label according to (Fig. 8 / p. 14).
Other countries: Check the warning label according to IEC 825 (text) in the local lan-
guage, the warning label according to IEC 825 (symbol), and the
identification label.
Great Britain: Check the warning label according to IEC 825 (text) in the local lan-
guage, the warning label according to IEC 825 (symbol), and the
identification label. Also check laser aperture labels (Fig. 7 / p. 13).

ARCADIS Avantic SPR2-330.815.02.02.02 Page 14 of 18 © Siemens, 2005


10.09 CS SD 24
Startup 15

Installing the covers 0

• Install the POWERPHOS cover and secure it with the cover screws. Reseal the POW-
ERPHOS cover with silicone sealing compound.

NOTE The POWERPHOS cover slightly shadows the laser light fan
beams at the I.I. edge. This is a design issue and cannot be
corrected. However, it does not negatively impact use of the
laser light localizer in normal operation.

Storing documents 0

• Store these adjustment instructions in register 4, "Start-up" of the ARCADIS Avantic


system folder. These adjustment instructions can be used for adjustment procedures
during subsequent service and maintenance work.

Functional Test 0

The ARCADIS Avantic system is operational.


• Switch on the laser light localizer via the laser light localizer button on the base unit con-
trol console.
¹ The two laser diodes light up.
• Switch off the laser light localizer again with the laser light localizer button.
¹ The two laser diodes go off.
• Switch on the laser light localizer again with the laser light localizer button and leave it
on for 5 minutes.
¹ The two laser diodes remain on for approx. 5 minutes, and then switch off auto-
matically.

Protective conductor resistance test 0

• Perform the protective conductor resistance test for the entire ARCADIS Avantic sys-
tem.
¹ The protective conductor resistance must not exceed 0.2 ohms.

© Siemens, 2005 SPR2-330.815.02.02.02 Page 15 of 18 ARCADIS Avantic


10.09 CS SD 24
16 Maintenance
3-
Maintenance activities on the laser light localizer
3Maintenance

Mechanical safety 0

• Ensure that the laser light localizer is labeled as described in this document. Replace
any labels that are damaged.
• Check the I.I. ring as well as the removable grid for mechanical damage.
• Check the laser beam exit window at the tube cover for damage.
• Ensure that the laser light fan beams are set as described in this document.

Operating values, inspection 0

• Check the function of the "Laser light localizer on/off" button on the base unit control
console.
• Switch the laser light localizer back on.
¹ The laser diodes must shut off automatically after 5 minutes.

ARCADIS Avantic SPR2-330.815.02.02.02 Page 16 of 18 © Siemens, 2005


10.09 CS SD 24
Changes to Previous Version 17
4Changes to Previous Version 4-
Chapters 1 through 3
Re-edited.
Changes in the laser light localizer beginning in November 2009 incorporated

© Siemens, 2005 SPR2-330.815.02.02.02 Page 17 of 18 ARCADIS Avantic


10.09 CS SD 24
18 Changes to Previous Version

ARCADIS Avantic SPR2-330.815.02.02.02 Page 18 of 18 © Siemens, 2005


10.09 CS SD 24

You might also like