Professional Documents
Culture Documents
DICOM Connection Guide
DICOM Connection Guide
1/3
Communication Settings for IP Address MWM Communication Test by
1 Set the IP address, subnet mask, default gateway of the
Receiving the Patient Information
electrocardiograph. When all the settings are completed, check the MWM
communication by receiving the patient information.
For details, refer to the Operater’s Manual of ECG-2250
or ECG-2350.
1 Get the test order from the person in charge of connection
2 Set the “Setup-Communication settings-Common settings- destination.
Communication mode” to “DICOM”.
2 Press the “ID” key to display “Enter the ID” screen. Enter
3 Set the “DICOM settings” in “Setup-Communication the patient ID and press the “Search” function key.
settings-Server settings”.
NOTE When need to connect execute MWM, set the
“Setup-Patient information-Input settings-Query
settings-Patient server” to “On”.
2/3
STORAGE Communication Test by Save and Manage the Local
Sending the DICOM Image Environment
When all the settings are completed, check the STORAGE When all the operation is completed, save the system settings
communication by sending DICOM image. and fill in the check sheet.
Please manage the system settings and check sheet.
1 Confirm the available patient ID, name with the person
in charge of connection destination before you send the
DICOM image.
NOTE • Use the test order after you set to receive the
patient information.
• Enter the weight, birth date (age) manually if
these information cannot be acquired.
Europe
Manufacturer
European Representative
SHANGHAI KOHDEN MEDICAL NIHON KOHDEN EUROPE GmbH NIHON KOHDEN UK LTD NIHON KOHDEN SINGAPORE PTE LTD NIHON KOHDEN KOREA, INC.
ELECTRONIC INSTRUMENT CORP. Raiffeisenstrasse 10 Trident Court 118, 1 Oakcroft Road 1 Maritime Square, #10-34 HarbourFront 5F Miso Bldg.890-47 Daechi-dong,
567 Huancheng Bei Road, Shanghai D-61191 Rosbach, Germany Chessington, Surrey KT9 1BD, UK Centre Singapore 099253 Gangnam-gu, Seoul, 135-280 Korea
Comprehensive Phone +49 6003-827-0 Phone +44 20-8391-6800 Phone +65 6376-2210 Phone +82 2-3273-2310
Industrial Development Zone, Shanghai China Fax +49 6003-827-599 Fax +44 20-8391-6809 Fax +65 6376-2264 Fax +82 2-3273-2352
Post Code: 201401
Phone: +86 (21) 57436998 NIHON KOHDEN DEUTSCHLAND GmbH NIHON KOHDEN ITALIA S.r.l. NIHON KOHDEN FRANCE SARL NIHON KOHDEN IBERICA S.L.
Fax : +86 (21) 57436939
Raiffeisenstrasse 10 Via FratelliBronzetti 28 8, rue Francois Delage, 94 230 Cachan, C/Ulises 75A
D-61191 Rosbach, Germany 24124 Bergamo, Italy France E-28043 Madrid, Spain
Phone +49 6003-827-0 Phone +39 035-219543 Phone +33 1-49-08-05-50 Phone +34 91-7-161080
Fax +49 6003-827-599 Fax +39 035-232546 Fax +33 1-49-08-93-32 Fax +34 91-3-004676
3/3