Professional Documents
Culture Documents
Hematl1 MG Activity 45
Hematl1 MG Activity 45
References:
Glowm. (n.d.). White Blood Cell Differential Count (Differential, Diff). https://glowm.com/
resources/glowm/cd/pages/resources/Lab/white.htm
Sight Dc. (n.d.). Neutrophils: definition, absolute high & low, ranges.
https://www.sightdx.com / knowledge-center/neutrophils
Manual counting of RBCS is unreliable because there are too many RBCs shown in the
counting chambers of a hemocytometer, and their relatively minute size also adds to the
manual method’s disadvantage. This could lead to miscounting errors such as recounting
the same RBC or failing to count (skipping) an RBC or two. These errors would produce
erroneous results like a falsely increased and falsely decreased counts of RBC, respectively.
Further, manual RBC counting relies on human visualization, judgment, and skill, so it is
even more unreliable and inaccurate results may be produced.
References:
Greer, J. et. al (2014) Wintrobe’s Clinical Hematology 13th edition. pp. 32. Wolters Kluwer I
Lippincott Williams and Wilkins.
Science Direct. (2020). Errors in Red Blood Cell count. https://www.sciencedirect.com/topics
/biochemistry-genetics-and-molecular-biology/erythrocyte-count
3. Why is it necessary to compute for corrected WBC count in the presence of significant
number of nucleated red blood cells? Give the formula (5 points)
If five or more nucleated RBCs per 100 WBCs are observed on the differential count on a
stained peripheral blood film, the WBC must be corrected for these cells. Because
nucleated RBCs are indistinguishable from WBCs when seen on the hemocytometer, they
could be erroneously counted as such, and the result would show a false increased in
WBC count. This could lead to misdiagnosis and improper treatment of the patient, so it’s
important to compute for correct WBC count. It is accomplished by using the formula:
Reference:
Keohane, E., Otto, C., & Walenga, J. (2016). Rodak’s Hematology Clinical Principles and Applications (6th
ed). Elsevier Inc. ISBN: 978-0-323-53045-3
4. Describe proper way of cleaning and drying (Hematology) diluting pipettes (4 points)
1) Attach the end of the pipette with the suction (pump) device.
2) Fill the pipette with water and thoroughly wash out the diluted blood until traces are
removed.
3) Draw and expel a bleach solution for a few times, followed by a rinse of distilled
water to clean the pipette from stains.
4) Rinse the pipette with absolute (95%) alcohol to remove the water within, then, use
either acetone or ether to remove alcohol.
5) Pass air through to dry the pipette, so that the bead rolls freely (free movement) in
the bulb.
** It is a very laborious process to clean the pipettes by blowing through them, further, one
cannot dry them properly by blowing, as expired air is laden with moisture.
References:
AUN Edu. (n.d.). III. Errors in counting total RBCs and WBCs by haemocytometer.
www.aun.edu.eg/developmentvet/clinical%20laboratory%20diagnosis/D_1_c.htm.
Klimud. (n.d.). Fundamentals of Hematology Cell Counts and Measurements. https://www.klim
ud.org/public/atlas/idrar/web/www.irvingcrowley.com/cls/fund.htm?fbclid=IwAR325UZ
nkfgUgghg2cXnYVkIMtNOtSEkSvnJ5gxHmsbST0xpin5UmQ3mpeo
5. Give 5 sources of errors in manual hemocytometer (cell counting). Briefly explain each (10
points)
4) Human Subjectivity
Each person performing the manual cell count adheres to a certain set of criteria that
defines a cell along with the stain intensity threshold to count it as viable or dead. Such
variations in human perception when counting manually can be extremely detrimental
to experimental setup and analysis when counting cells manually. If multiple users
count the same sample, it is not uncommon to see a variance significantly higher than
the mean of a Poisson distribution.
5) Mathematical Errors
With any type of data collection, a good deal of math is involved. The use of
mathematics is essential in calculating the average number of cells in a suspension.
Math performed manually is time consuming and is also error-prone, as it relies on the
skill of the person involved. Improper formulas, miscomputations, and typographical
errors like mistyped numbers and mathematical functions, can lead to huge
inaccuracies in the values of the cells.
References:
Chemometec. (n.d.). Manual vs. Automated Cell Counting. https://chemometec.com/resourc
es/mini-reviews/manual-vs-automated-cell-counting/
Corning. (n.d.). 10 Most Common Errors Made in Cell Counting. https://www.corning.com/cata
log/cls/documents/application-notes/CLS-AN-495.pdf