Professional Documents
Culture Documents
BP 2
BP 2
4.2.2. Materials
- BIOPAC Pressure Cuff (SS19L with gauge dial for MP3X or SS19LA with
onscreen gauge display for MP45)
- BIOPAC Stethoscope (SS30L)
- BIOPAC Electrode Lead Set (SS2L)
- Electrodes – disposable, 3 per Subject (EL503)
- Rubbing alcohol and swab (to clean stethoscope earpieces and stethoscope
diaphragm)
- Tape measure (for pulse speed calculation)Optional: washable felt pen (to mark
stethoscope placement on arm)
- Biopac Student Lab System: software BSL 3.7.7 or above data acquisition unit
MP36, MP35, MP30 (Windows only), or MP45
- Computer System
4.2.3. Method
4.2.3.1. Calibration
Step 1: Double check the electrodes and stethoscope, and make sure Subject is relaxed
and sitting down.
Step 2: Click Calibrate.
Step 3: Confirm cuff is deflated and click OK.
Step 4: Roll the cuff onto itself such that when completed, the Velcro will prevent it from
unrolling with inflated.
Step 5: When prompted, Director inflates cuff to 100 mmHg and Recorder clicks “OK”.
Step 6: At the next prompt, Director deflates cuff to 40 mmHg and Recorder clicks
“OK”.
Step 7: Director taps the stethoscope diaphragm twice.
Step 8: Check calibration data:
- If similar, proceed to Data Recording.
- If different, Redo Calibration.
4.2.3.2. Recording lesson data
Step 1: Prepare for the recording.
Step 2: Review the Blood Pressure measurement procedure provided in the Introduction
Step 3: Make sure all the air in the sphygmomanometer cuff is expelled and close the
pressure release valve.
Step 4: Place the cuff on Subject’s LEFT arm. (Fig. 4.2.3.1a)
- “Artery” label should be over the brachial artery (with the arrow on the label
facing down).
- Lower edge of cuff should be 1.5 to 2 inches above the antecubital fossa (inner
aspect of elbow).
- Wrap the cuff evenly and snugly on Subject’s arm.
* Cycle measurements: If ECG was recorded, use , if ECG was not recorded,
use .
Cycle * Calculate the Mean
Condition Trial
1 2 3 of Cycles 1-3 of Trial 1-2 means
1 78,43 81,39 78,02 79,28
Left arm, seated 79,04
2 77,3 79,98 79,12 78,8
1 72,48 71,06 72,37 71,97
Right arm, seated 71,825
2 72,58 70,92 71,54 71,68
1 50,32 51,38 51,48 51,06
Right arm, lying down 51,275
2 49,47 52,58 52,42 51,49
Right arm, after exercise 1 96,23 97,88 96,2 96,77
Table 4.2.4.1d
E. Timing of Korotkoff Sounds
Timing of Sounds
Condition Trial
Mean (calc)
1 0,73
Left arm, seated 0,76
2 0,79
1 0,537
Right arm, seated 0,544
2 0,551
1 0,398
Right arm, lying down 0,4305
2 0,463
Right arm, after exercise 1 0,813 0,813
Table 4.2.4.1e
F. Calculation of Pulse Speed
4.2.5. Questions
1. Note the difference in systolic pressure value between when the sound actually began,
was detected by the stethoscope transducer, and was recorded, and the time when the
observer first heard the sound and pressed the marker button. (Example: 141 mmHg –
135 mmHg = 6 mmHg). What factors could account for this difference? Would the
observed difference be the same if measured by another observer? Explain your answer.
There would be a difference, the recorded data would be less accurate than those detected
by Stethoscope transducer, because of the communication lag, there would be difference
between reaction time of hearing and recording the data, which is type of human error.
2. a) Does your systolic and/or diastolic arterial pressure change as your heart rate
increases?
Systolic and diastolic arterial pressure increases with increase in heart rate. Because
diastolic pressure is the pressure that take place on your artery walls between the heart
beats while systolic pressure refers to the pressure arises during heart beat. So, increase in
systolic pressure is greater than diastolic pressure.
b) How does this change affect your Pulse Pressure?
This change will result in two ways:
First way, with good cardiovascular health: Pulse pressure likewise increases as heart rate
increases after exercise. After exercise, supply routes are more versatile.
Second way, with hardened artries, they don't extend also with increased pressure; Pulse
pressure stays unaltered.
3. Give three sources of error in the indirect method of determining systemic arterial
blood pressure:
Wrong size cuff used: Large cuff will give false low blood pressure reading while Small
cuff will give false high blood pressure reading. Cuff size should be recommended that
the bladder length and width should be 80% and 40% respectively.
Not proper functioning of the electronic equipment: The correct working of the electronic
euipment can be checked by verifying the displayed pulse with an actual patient pulse.
Differences greater then 10% will affect the calculations seriously and results in incorrect
diastolic and systolic values on the display screen. Non Invasive Blood Pressure (NIBP)
machines is used as electronic machine that recognizes the changes in the air pressure in
the cuff caused due to the blood flowing through the BP cuff extremity. Sensors calculate
the patient’s pulse rate and Mean Arterial Pressure (MAP). Software use these two
variables for calculating the systolic and diastolic BP.
Incorect position of pateint body: Most accurate blood pressure is obtained by upright
seated position, in this the pressure remains at the patient side for longer. The problem in
measuring accurate pressure take place as pateint lying on their side or in other position.
For measuring correct blood pressure Bp cuff is put at mid heart level at the time of
pressure. Transducer should also place at mid heart level for accurate measuring of
pressure.
4. Use an equation that relates flow, pressure, and resistance to define mean arterial
pressure:
The equation for mean arterial pressure is:
MAP=(1/3) PP (pulse pressure) + Diastolic
Example: if someone's blood pressure is 125/65, then Pulse Pressure is 60. Now that you
have this info you can calculate MAP (mean arterial pressure)
MAP= (1/3) 60 + 65
MAP= 85
CO is cardiac output (CO is defined as stroke volume, the amount of blood pumped out
in a single beat, times heart rate-- CO = SV x HR)
*MAP is mean arterial pressure
* SVR is systemic vascular resistance
* CVP is central venous pressure
Or, different equation, same answer: MAP = DP + (1/3)PP
Where DP is diastolic pressure and PP is pulse pressure. (Systolic minus Diastolic)
5. Blood flow (liters per min.) through the pulmonary circuit equals blood flow through
the systemic circuit, but pulmonary resistance to flow is 5 times less than the systemic
resistance to flow. Using the equation in Question 4, show that mean pulmonary pressure
is 5 times less than mean systemic pressure.
As we have known,
Pressure
Pressure = Flow x Resistance => Flow = (1)
Resistance
Given,
Systemic flow = Pulmonary flow (2)
Systemic Resistance = 5 x Pulmonary Resistance
So, we replace the equation (1) into (2):
Systemic Pressure Pulmonary Pressure
=
Systemic Resistance Pulmonary Resistance
Systemic Pressure Pulmonary Pressure
( ¿) =
5 x Pulmonary Resistance Pulmonary Resistance
Systemic Pressure
( ¿) =Pulmonary Pressure
5
6. Define the first and second sounds of Korotkoff. Which sound is used to approximate
systolic pressure and which sound is used to approximate diastolic pressure?
The first sound occurs when the cuff pressure is same to the systolic arterial pressure.
When this sound occurs, the vibrations generated as blood is forced past the area of
occlusion. These vibrations are dependent on the systolic Korotokoff sound. The diastolic
sound is different in pitch of vibration. When the occlusion pressure cuff reduces below
the diastolic pressure and there is no more occlusion of the artery. The second sound
represents the diastolic pressure. The first sound is used approximate systolic pressure.
The first sound of Korotkoff occurs when the cuff pressure equals the systolic pressure.
The second sound is used approximate diastolic pressure. The second Korotkoff sound
occurs when the cuff pressure equals the diastolic pressure.
7. Why is mean arterial pressure not equal to (systolic pressure – diastolic pressure)/2?
The heart spends more time in diastole than in systole, therefore the MAP has to take into
account the extra amount of time spent in diastole. The lengths of systole and diastole are
different. The amount of time spent in diastole is longer than the amount of time spent in
systole. The length of time the heart is in diastole is approximately twice as long as it is
in systole. As one's heart-rate increases and the length of diastole shortens, the mean
arterial pressure is much closer to just the average of systolic blood pressure and diastolic
pressure.
8. Define pulse pressure. Explain, in terms of changes in systolic and diastolic pressures,
why pulse pressure increases during exercise.
Pulse pressure is a measurement obtained by calculating the difference the systolic and
diastolic pressure. During exercise flow needs to increase in order to supply the body
with more oxygen. In this process the systolic pressure increases while the diastolic
pressure stays the same or increases by a small amount at best. Using the equation stated
earlier to find pulse pressure, knowing that systolic pressure increases and diastolic does
not change during exercise it can be found that pulse pressure increases during exercise.
9. Give one reason why blood pressure in the left arm may be different than blood
pressure in the right arm of a Subject at rest
A blood pressure difference of less than 5 mmHg between the left and right arm is
considered normal. This minor difference is usually due to the asymmetry in how the
blood vessels come off the aorta, the main vessel leaving the heart.the left arm is closest
to the aorta.
10. Name an artery other than the brachial that could be used for an indirect
measurement of blood pressure and explain your choice.
Brachial artery is used most often for checking the blood pressure.Another artery used for
indirect measurement of blood pressure is femoral artery .
Explanation: Femoral artery pressure gives standard for calibrating auscultatory technics.
As this pressure is widely used cardiac catheterization.
One advantage of of femoral artery is that the vessel is larger and gives stronger
impulsation than radial artery.
Note that, while measuring the blood pressure from femoral artery a cuff which 25%
wider should be used.