CORRIGENDUM-3 dtd 11.07.

2023

TENER NO.HLL/CMO/HCD/PS/ECRP-II/GH-AP/2023-24/02

SETTING UP PEDIATRIC CARE UNITS AT VARIOUS GENERAL HOSPITALS /DISTRICT HOSPITALS IN


STATE OF ARUNACHAL PRADESH ON TURNKEY BASIS

Technical Specification of Equipment


Sl Equipment
As per Tender Specification To be Read as
No: Name
Simultaneous acquisition of up to 12 Simultaneous acquisition of up to 12
1
leads leads
2 Minimum 3 inches Minimum 4 inches or more
Facility for recording in Manual or Facility for recording in Manual or
3
Auto mode or Rhythm Mode Auto mode or Rhythm Mode
Sensitivity : 2.5,5,10,20 mm/mV & Sensitivity : 5,10,20 mm/mV &
4
AGC AGC
Recording speeds of 5,10, 25 and Recording speeds of 5,10, 25 and 50
5
50 mm/sec mm/sec
Frequency Response: 0.05 Hz to Frequency Response: 0.05 Hz to
6
150 Hz 150 Hz
7 Sampling frequency : 1000 Hz Sampling frequency : 1000 Hz
User selectable filter: AC Filter, User selectable filter: AC Filter, EMG
8 EMG filter- 25 or 35 or 45 Hz or filter- 25 or 35 or 45 Hz or OFF, Anti-
OFF, Anti-Drift filter Drift filter
Print formats : Manual : 3,2,1 & Print formats : Manual : 3,2,1 &
9
AUTO: 3 Channel + Rhythm Lead AUTO: 3 Channel + Rhythm Lead
At least 80 mm paper width for the 80 mm or more paper width for the
10
thermal printer thermal printer
ECG
Printer must compatible with both Printer must compatible with Roll
11 Machine
Roll and Z fold ECG paper or Z fold ECG paper
Colour coded keys for ease of Colour coded keys for ease of
12
operation operation
Light weight – Less than 2.75 Kg
13 Should be lightweight
with Battery
Battery operation – Lithium Ion Battery operation – Lithium Ion
14 Battery -minimum 3 HRS continuous Battery -minimum 3 HRS continuous
back-up with fully charged battery back-up with fully charged battery
15 Easy to carry handle Easy to carry handle
Capability to generate any number
Capability to generate any number of
16 of ECG copies possible for
ECG copies possible for distribution
distribution
Automatic measurement and Automatic measurement and
17
interpretations of ECG data. interpretations of ECG data.
Facility to store at least 100 ECG
18 Facility to 40 or more ECG data
data
19 DF protection DF protection
PC interface facility and optional PC PC interface facility and optional PC
20 interface software (Optional ECG interface software (Optional ECG
data transfer feature) data transfer feature)
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
External storing and retrieving External storing and retrieving facility
21
facility through USB storage device through USB storage device
Direct External PC printer interface Direct External PC printer interface
22
facility facility
Display to Preview signal quality Display to Preview signal quality
23 prior to printing thereby saving time prior to printing thereby saving time
and paper and paper
24 RS232/Ethernet Port RS232/Ethernet Port
Standard accessories must be Standard accessories must be
25
provided along with the machine provided along with the machine
Power Cable – 1 no; 10 Lead
Power Cable – 1 no; 10 Lead Patient
Patient Cable – 1no; Suction
Cable – 1no; Suction Electrodes – 6
Electrodes – 6 nos; Clip-on
26 nos; Clip-on electrodes – 4 nos;
electrodes – 4 nos; ECG Gel – 1
ECG Gel – 1 bottle; Thermal
bottle; Thermal recording Paper –
recording Paper – 1no
1no
Should have BIS/CE or USFDA Should have BIS/CE or USFDA
27
certified certified
28 One-year warranty One-year warranty
Sl Equipment
As per Tender Specification To be Read as
No: Name
The machine should have facility for The machine should have facility for
ECG Monitoring, external & internal ECG Monitoring, external & internal
1
Defibrillation, transcutaneous pacing Defibrillation, transcutaneous pacing
& in-built recorder. & in-built recorder.
2 Screen Size :- 6.5 inches Screen Size :- 6.5 inches or more
The machine must have optional The machine must have optional
facility to monitor vital sign facility to monitor vital sign parameter
3
parameter such as mainstream such as mainstream EtCO2 and
EtCO2 and SpO2. SpO2.
Should work on manual and Should work on manual and
automated external defibrillation in automated external defibrillation in
4 Bi-phasic mode. The maximum Bi-phasic mode. The maximum
energy delivered by the device energy delivered by the device
Biphasic should be at least 200J or more. should be at least 200J or more.
Defibrillator Should have backlit 6 inch or more Should have backlit 6 inch or more
colour LCD display that provides colour LCD display that provides
5
clear visibility even under bright clear visibility even under bright
daylight. daylight.
Should have display of at least 4 Should have display of at least 4
6 waveforms of monitoring parameters waveforms of monitoring parameters
and their numerical data. and their numerical data.
System should have instant boot up System should have instant boot up
7 time, less than 4 seconds (to be time, less than 4 seconds (to be
demonstrated). demonstrated).
Should have easy operation of all Should have easy operation of all
8
functions through single rotary knob. functions through single rotary knob.
Should monitor ECG through Should monitor ECG through
9
external paddles and monitoring external paddles and monitoring
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
electrodes. electrodes.

Should have lightweight mainstream Should have lightweight mainstream


EtCO2 sensor (preferably <10g EtCO2 sensor (preferably <10g
weight) which can be used on both weight) which can be used on both
10
intubated and non-intubated patient. intubated and non-intubated patient.
(Optional, Price to be quoted (Optional, Price to be quoted
separately). separately).
Should have external paddles with Should have external paddles with
paddles contact indicator- for good paddles contact indicator- for good
11 paddle contact. Single adult and paddle contact. Single adult and
paediatric paddles should be paediatric paddles should be
available. available.
Should have fast charging time, Should have fast charging time,
12 charging 200J in 5 seconds or less charging 200J in 5 seconds or less
on both mains & battery. on both mains & battery.
The machine should have charge The machine should have charge
13 and discharge button on front panel and discharge button on front
& paddles. panel or paddles.
Should have self-test status
14 indicator that displays daily and Should have self test facility
monthly self-test results.
Should display alarm message and Should display alarm message and
have alarm indicator on top of the have alarm indicator on top of the
15
machine to view from distance the machine to view from distance the
alarm type and patient condition. alarm type and patient condition.
After defibrillation, the ECG After defibrillation, the ECG
waveform must recover within 5 waveform must recover within 5
16
seconds for immediately checking seconds for immediately checking
the result of defibrillation. the result of defibrillation.
The machine should be compact, The machine should be compact,
portable with built in rechargeable portable with built in rechargeable
17
battery, weight of the total machine battery, weight of the total machine
should not be more than 8 Kg. should not be more than 8 Kg.
Should have a battery capable of Should have a battery capable of
giving 80 discharges of maximum giving 80 discharges of maximum
18 energy, 3 hours of continuous energy, 3 hours of continuous
monitoring or 1 hour of external monitoring or 1 hour of external
pacing. pacing.
The battery charging time should be The battery charging time should be
19
less than 4 hours to full charge. less than 4 hours to full charge.
It should have facility to store patient It should have facility to store patient
20 data and review data on SD data and review data on SD
Card/USB. Card/USB.
Machine must be able to operate in Machine must be able to operate in
extreme environment conditions; it extreme environment conditions; it
21 should operate from 0°C to 45°C should operate from 0°C to 45°C
and should be highly resistant to and should be highly resistant to
water and dust. water and dust.
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
Should meet IP44 Level for water Should meet IP44 Level for water
22 resistance and Protection against resistance and Protection against
harmful ingress of dust. harmful ingress of dust.
Should conform to latest safety Should conform to latest safety
standards, such as IEC-60601-2-4, standards, such as IEC-60601-2-4,
23
IEC-60601-1-2, ISO 14971: 2007, IEC-60601-1-2, ISO 14971: 2007,
EN 1789: 2007 etc. EN 1789: 2007 etc.
Should have BIS/CE or USFDA Should have BIS/CE or USFDA
24
certified certified
The machine should be supplied
25 The machine should be supplied with
with
a. ECG 3 lead cable with connection a. ECG 3 lead cable with connection
26
cord – 01 No. cord – 01 No.
27 b. ECG paper roll – 2 Nos. b. ECG paper roll – 2 Nos.
28 c. Pacing cable with AED pads c. Pacing cable with AED pads
Should have advanced disposable
pads with noiseless function to
29 suppress CPR noise during chest Removed
compressions. (Optional, Price to be
quoted separately).
Should have advanced wireless
CPR audio-visual feedback system
which can be used in
synchronization with defibrillator
provides real time feedback from
30 Removed
each compression, guides
compression for depth, rate and
release. It can be used as training
assist device. (Optional, Price to be
quoted separately).
Machine must be vibration resistant, Machine must be vibration resistant,
it should meet international standard it should meet international standard
MIL-STD-810F 514.5 category 4 MIL-STD-810F 514.5 category 4 and
31
and 9 for truck ambulance and 9 for truck ambulance and Helicopter
Helicopter ambulances respectively, ambulances respectively, for transfer
for transfer of patients. of patients.
32 One-year warranty One-year warranty
Sl Equipment
As per Tender Specification To be read as
No: Name
Monitor should be of capable of Monitor should be of capable of
1 monitoring adult, paediatric and monitoring adult, paediatric and
neonatal patients neonatal patients
High visibility with large colour High visibility with large colour
2 Multipara
screen and additional alarm light. screen and additional alarm light.
Monitor
Screen size at least 12 inches TFT Screen size at least 12 inches TFT
LCD display with rotary Knob for LCD display with rotary Knob for
3
easy viewing from distance in large easy viewing from distance in large
intensive care unit. intensive care unit.
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
It should monitor ECG,
It should monitor ECG, Spo2,NIBP
Spo2(Massimo/equivalent),NIBP ,1
,1 IBP, Respiration and 2-
4 IBP, Respiration and 2-
Temprature. The monitor should be
Temprature. The monitor should
upgradable and ETCO2.
be upgradable and ETCO2.
Monitor should have minimum 6 Monitor should have 6 or
5
waveforms. morewaveforms.
Monitor should have trending facility Monitor should have trending
6
for at least 72 hours. facility for 72 hours or more.
The monitor should have 7 lead The monitor should have 7 lead ECG
ECG display (simultaneous) display (simultaneous)
Must have an inbuilt continuous Must have an inbuilt continuous
7
battery backup for minimum 3 hours. battery backup for minimum 3 hours.
The quoting firm must give The quoting firm must give
8
demonstration of the quoted model. demonstration of the quoted model.
Accessories supplied for each Accessories supplied for each
9
monitor: monitor:
10 a) 5-lead ECG Cable- 1 no. a) 5-lead ECG Cable- 1 no.
b) Spo2 sensor Adult/Paediatric -1 b) Spo2 sensor Adult/Paediatric -1
11
no. no.
12 c) NIBP Cuff Adult-1 no. c) NIBP Cuff Adult-1 no.
13 d) NIBP Cuff Paed.-1 no. d) NIBP Cuff Paed.-1 no.
e) Temp probe-Rectal & Skin-1 no. e) Temp probe-Rectal & Skin-1 no.
14
each each
Should have BIS/CE or USFDA Should have BIS/CE or USFDA
15
certified certified
16 One-year warranty One-year warranty
Sl Equipment
As per Tender Specification To be read as
No: Name
Compact portable bedside pulse Compact portable bedside pulse
1 oxymeter with minimum 4 inch and oxymeter with minimum 4 inch and
above, multicolour TFT LCD display. above, multicolour TFT LCD display.
Continuous monitoring of SpO2 Continuous monitoring of SpO2
2 (arterial blood oxygen saturation), (arterial blood oxygen saturation),
pulse rate and signal strength pulse rate and signal strength
Should have Nellcor or equivalent
technology
Pulse
Oxymeter Should have manual selection of Should have manual selection of
3
adult, paediatric and neonatal mode. adult, paediatric and neonatal mode.
5 Measuring Rage: Measuring Rage:
6 SpO2: 1 to 100% SpO2: 1 to 100%
7 Pulse Rate: 25 to 250 bpm Pulse Rate: 25 to 250 bpm
8 Pulse Amplitude: 0.03% to 20% Pulse Amplitude: 0.03% to 20%
Accuracy SpO2 for Adult, Accuracy SpO2 for Adult,
9 Paediatric& Neonate – 70% to 100% Paediatric& Neonate – 70% to 100%
(+ 2 digits) & 60% to 69% (+3 digits) (+ 2 digits) & 60% to 69% (+3 digits)
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
Display Shows: SpO2 (%), PR (HR Display Shows: SpO2 (%), PR (HR
in bpm), Plethysmographic in bpm), Plethysmographic
Waveform, Pulse amplitude with Waveform, Pulse amplitude with
visual indicators for pulse search, visual indicators for pulse search,
10
audible alarms silenced or off, audible alarms silenced or off,
interference indicator, battery interference indicator, battery
charging and Alarm management charging and Alarm management
settings settings
Multicolour display screen with ideal Multicolour display screen with ideal
11
contrast for easy readability contrast for easy readability
Should offer accurate and reliable
Should offer accurate and reliable
performance in monitoring
performance in monitoring conditions
conditions of low perfusion and
of low perfusion and signal
12 signal interference. It must show
interference. It must show SpO2
SpO2 value for low perfusion
value for low perfusion patients with
patients with accuracy – 70% to
accuracy – 70% to 100% (+ 2 digits)
100% (+ 2 digits)
Should have wide range of sensors Should have wide range of sensors
including forehead sensor for more including forehead sensor for more
13
effective monitoring option in low effective monitoring option in low
perfusion patients perfusion patients
Tenderer Should have wide range of
sensors for Adult, Paediatric,
Equipment should supply with
Neonatal and premature patient
sensors for Adult, Paediatric,
14 types separately for adhesive and
Neonatal and premature patient
software(Velcro) disposable sensors
types separately.
to accommodate diverse patient
sizes weights and needs.
Sensor Event Report, allowing Sensor Event Report, allowing
patient alarm events to be recorded patient alarm events to be
15
on sensors for memory back up of recorded on sensors for memory
24 hrs. back up of 24 hrs or more.
Power Requirements: input 200 to Power Requirements: input 200 to
240ac 50Hz supply and internal re- 240ac 50Hz supply and internal re-
16
chargeable battery with 3 hours chargeable battery with 3 hours or
battery back-up more battery back-up
It should be light weight of less than It should be light weight of less than
17 2kg and size not above 2kg and size not above
84Hx260Wx170D (mm) 84Hx260Wx170D (mm)
User preset of high/low alarm on User preset of high/low alarm on
18
SpO2 and PR monitoring SpO2 and PR monitoring
Audio visual alarm for SpO2 and PR Audio visual alarm for SpO2 and PR
in case measurements are outside in case measurements are outside
19
preset range with alarm silence preset range with alarm silence
feature feature
Should have Sat seconds alarm Should have Sat seconds alarm
management system to reduce the management system to reduce the
20
unwanted alarms which disturbs the unwanted alarms which disturbs the
sick patients. sick patients.
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
Automatic switch from mains to Automatic switch from mains to
21
batteries in case of power failure. batteries in case of power failure.
22 To be supplied with soft probe To be supplied with soft probe
Should have BIS/CE or USFDA Should have BIS/CE or USFDA
23
certified certified
24 One-year warranty One-year warranty
Sl Equipment
As per Tender Specification To be read as
No: Name
Should be Compact, Light Weight Should be Compact, Light Weight
Multipurpose ventilator for use in Multipurpose ventilator for use in
1
Ambulance, NICU, Transport & Ambulance, NICU, Transport &
Causality Causality
Ventilator weight should not be more
2 Machine should be light weight.
than 4Kg
Should be capable for use on Should be capable for use on
3
Neonatal and Paediatric Patients Neonatal and Paediatric Patients
Should have 7 inch or More Colour Should have 7 inch or More Colour
4 TFT display with Touch Screen TFT display with Touch Screen
facility. facility.
Ventilator should work on the Built-in Ventilator should work on the Built-in
5
Blower/Turbine. Blower/Turbine.
Ventilator Should be capable of Ventilator Should be capable of
6 working with High pressure Oxygen working with High pressure Oxygen
up to 6 Bar. up to 6 Bar.
Should have the following modes of Should have the following modes of
7
Ventilation: Ventilation:
Pressure Modes - IMV, PCV, Pressure Modes - IMV, PCV, PSIMV,
8 Ventilator PSIMV, CPAP/PSV, APRV, CPAP/PSV, APRV, BILEVEL, PRVC,
Invasive BILEVEL, PRVC, PRVC-SIMV PRVC-SIMV
Volume Modes – SCMV, V-A/C, V- Volume Modes – SCMV, V-A/C, V-
9
SIMV, SIMV,
Non-Invasive Modes : PSV-S/T, P- Non-Invasive Modes : PSV-S/T, P-
10
A/C, nCPAP A/C, nCPAP
11 Modes – High Flow Oxygen Therapy Modes – High Flow Oxygen Therapy
Flow Sensor Must be at Expiratory Flow Sensor Must be at Expiratory
12
End. End.
User should be able to Select Apnea User should be able to Select Apnea
13
Mode from VACV and PACV Mode from VACV and PACV
Ventilator should have 3 Real time Ventilator should have 3 Real time
14
curves and 3 Loops curves and 3 Loops
Should have Graphical trend 24 Should have Graphical trend 24
15
hours hours
Ventilator Should have the below Ventilator Should have the below
16
mentioned settings mentioned settings
17 Respiratory rate 1-120 bpm Respiratory rate 1-120 bpm
18 PEEP/CPAP-0-25CMH20 PEEP/CPAP-0-25CMH20
19 P control -5-60 cmH20 P control -5-60 cmH20
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
20 P support 0-35 cmH20 P support 0-35 cmH20
Expiratory sensitivity10–80% of Expiratory sensitivity10–80% of Peak
21
Peak Flow Flow
Tidal volume from 30 ml to 700 ml ( Tidal volume from 30 ml to 700 ml (
Only for Paediatric Patients ) ( For Only for Paediatric Patients ) ( For
22
Neonatal Patients Pressure Modes Neonatal Patients Pressure Modes
Only ( 0 to 60 cm H2o) ) Only ( 0 to 60 cm H2o) )
Should be supplied with
permanent O2 cell and shall be
23
replaced for free of cost during
the warranty and CAMC period.
Ventilator Should operate on Mains Ventilator Should operate on Mains
with a Standard Internal with a Standard Internal
24
Rechargeable Battery Back-up for a Rechargeable Battery Back-up for a
minimum of 3 Hours. minimum of 3 Hours.
Ventilator Should have the DC Ventilator Should have the DC
25 Socket for connecting external DC Socket for connecting external DC
supply if required. supply if required.
Should have Flow Triggering & Should have Flow Triggering &
Pressure Triggering facility with user Pressure Triggering facility with user
26
selectable Flow & Pressure selectable Flow & Pressure
Triggering Levels Triggering Levels
Should have Auto set Alarms
27 Should have Auto set Alarms feature
feature
Should have Alarms for Airway Should have Alarms for Airway
pressure, Tidal volume, Minute pressure, Tidal volume, Minute
28 volume, Respiratory rate volume, Respiratory rate
Apnea,O2%, Airway leak, Circuit Apnea,O2%, Airway leak, Circuit
open, Battery indicator open, Battery indicator
O2 supply pressure failure, O2 supply pressure failure, Ventilator
29
Ventilator in-operative in-operative
Should have In built Air-oxygen Should have In built Air-oxygen
30 Blending mechanism with Fi02 level Blending mechanism with Fi02 level
variable between 21%to 100% variable between 21%to 100%
Should have panel lock facility to Should have panel lock facility to
31 avoid accidental change of any avoid accidental change of any
settings settings
Should be Upgradable to ETCo2 Should be Upgradable to ETCo2 and
32 and SPO2 with PVI, SpCO, SpMet SPO2 with PVI, SpCO, SpMet and
and SpHb monitoring. SpHb monitoring.
Scope of Supply Should Include Scope of Supply Should Include
Trolley ( Same Manufacturer ), Trolley ( Same Manufacturer ), Servo
33 Servo Controlled Humidifier with Controlled Humidifier with reusable
reusable chamber, Flow Sensor -2 chamber, Flow Sensor -2 nos,
nos, Exhalation Membrane -2nos Exhalation Membrane -2nos
Ventilator Should be BIS/European Ventilator Should be BIS/European
34
CE/USFDA Approved CE/USFDA Approved
Standards: - IEC 60601-1-2, EN Standards: - IEC 60601-1-2, EN 794-
35 794-3, EN ISO15223-1 and EN 3, EN ISO15223-1 and EN 1789 for
1789 for Ambulance. Ambulance.
Technical Specification of Equipment
Sl Equipment
As per Tender Specification To be Read as
No: Name
36 Warranty for 3 year Warranty for 3 year

11.07.2023 Associate Vice President (HCD)