Professional Documents
Culture Documents
JRN 2016
JRN 2016
Purpose: Since bacterial leakage along the implant-abutment interface may be responsible for peri-implant
infections, a realistic estimation of the interface gap width during function is important for risk assessment.
The purpose of this study was to compare two methods for investigating microgap formation in a loaded
dental implant, namely, microcomputed tomography (micro-CT) and three-dimensional (3D) nonlinear
finite element analysis (FEA); additionally, stresses to be expected during loading were also evaluated by
FEA. Materials and Methods: An implant-abutment complex was inspected for microgaps between the
abutment and implant in a micro-CT scanner under an oblique load of 200 N. A numerical model of the
situation was constructed; boundary conditions and external load were defined according to the experiment.
The model was refined stepwise until its load-displacement behavior corresponded sufficiently to data
from previous load experiments. FEA of the final, validated model was used to determine microgap widths.
These were compared with the widths as measured in micro-CT inspection. Finally, stress distributions were
evaluated in selected regions. Results: No microgaps wider than 13 µm could be detected by micro-CT
for the loaded implant. FEA revealed gap widths up to 10 µm between the implant and abutment at the
side of load application. Furthermore, FEA predicted plastic deformation in a limited area at the implant
collar. Conclusion: FEA proved to be an adequate method for studying microgap formation in dental
implant-abutment complexes. FEA is not limited in gap width resolution as are radiologic techniques and
can also provide insight into stress distributions within the loaded complex. Int J Oral Maxillofac Implants
2016;31:e32–e42. doi: 10.11607/jomi.4179
Keywords: dental implant, finite element analysis, microcomputed tomography, microgap formation, preload,
tightening torque
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
microgap widths were found to be in the range of only steel cylinder (inner diameter 12 mm, wall thickness
1 to 2 µm. Abdelhamed et al12 demonstrated leakage 1 mm, height 15 mm) in such a manner that, according
of toluidine blue solution through microgaps in the to the requirements of ISO 14801,24 an assumed bone
implant assemblies but could only estimate that the loss of 3 mm from the implant collar was simulated. Sub-
gap widths were less than approximately 10 µm, the sequently, the abutment was connected to the implant
resolution limit of the micro-CT device used. While in with the abutment screw and tightened with a torque
these two studies, the implants (Osseo Speed, Astra Tech) of 25 Ncm, as recommended by the manufacturer. A
were inspected unloaded, Zipprich et al16 assessed the hemispherical load cap made of cobalt chromium alloy
gap widths of these and other implants under dynamic (coron, Etkon) was then cemented to the abutment,
loading. However, up to an oblique load of 200 N, the giving a distance of 11 mm between the implant collar
gap widths were less than the resolution limit of the and the site of force application.
analytical method (approximately 4 µm). For the examination by micro-CT, the implant assembly
Because of the limited resolution of the aforemen- was inserted into a special loading jig, as shown in Fig 1.
tioned methods, no reliable conclusions are possible The loading jig basically consisted of a hollow aluminum
about the tightness of the interface between the implant cylinder (inner diameter 14 mm, wall thickness 3 mm,
and abutment against leakage of typical oral bacteria length 120 mm) surrounding the assembly, a calibrated
with diameters down to 0.1 µm.14 As an alternative coil spring (spring constant 26 N/mm) for exerting an
method, finite element analysis (FEA) can be applied, axial force to the bottom of the steel cylinder with
an approximate numerical method for solving problems implant assembly, and a loading plate, inclined by 30
in structural mechanics that is capable of calculating degrees, in contact with the hemispherical load cap. As
displacements of structures under simulated load with the loading plate rested on a linear ball bearing, only
unlimited resolution. In recent decades, FEA has been forces normal to the contact area could be transferred
used in implant dentistry to study the biomechanical to the load cap, which is, due to the inclination of the
performance of dental implants,17 but to the authors’ plate, under an angle of 30 degrees with respect to the
knowledge, only two studies18,19 have used FEA to implant axis. Microgap formation between the implant
investigate microgap formation in an implant-abutment and abutment was analyzed with an oblique load of
complex. As these latter studies neglected the mechanical 200 N resting on the specimen after adjusting the spring
effects of abutment screw tightening, their results can force via a set screw accordingly. In the micro-CT device
only be regarded as approximate. (type X8011, Viscom), the specimen was inspected at 80-kV
Thus, it was the primary goal of this study to construct acceleration voltage with 720 exposures per revolution
and validate a realistic finite element model of an implant- and 2-second duration per exposure. A 1024 × 1024
abutment assembly in an in vitro test set-up, and to charge-coupled device (CCD) detector recorded the
assess gap widths after loading by FEA and, additionally, transmitted radiation. By means of the software XVR-CT
by micro-CT for cross-checking of results. Furthermore, (Viscom) and using a cone beam algorithm, the images
stresses in selected regions were evaluated by FEA, and were reconstructed with a voxel edge length of 6.5 µm.
how far the results for microgap widths and stresses The latter was set by the program according to the
were influenced by typical simplifications made in FEA detector pixel size and the source-object and source-
regarding screw preload18–22 and contact definitions17,23 detector distances, so neither over- nor undersampling
at interfaces was investigated. of the voxel size took place except for the systematic
variation due to the cone beam geometry. The effect
of beam hardening was reduced by the software with
MATERIALS AND METHODS the corresponding beam-hardening factor set to 6.5.
Both the micro-CT experiment and three-dimensional (3D) Finite Element Analysis
FEA were set up according to the load test described in The geometrical data of implant body, abutment, and
ISO 14801.24 The implant-abutment complex investigated abutment screw, as supplied by the manufacturer in
consisted of a titanium implant (Osseo Speed 4.5, conical International Graphics Exchange Specification (IGES)
neck design, length 13 mm) and a titanium abutment format, were imported into the computer-aided design
(TiDesign 4.5/5.0; diameter 5.5, 1.5 mm, both Dentsply (CAD) part (Design Modeler) of the finite element program
Implants, formerly Astra Tech; for material properties, ANSYS (Workbench 14.0, ANSYS), which was subsequently
see Table 1).25–29 used for FEA.
Analogously to the micro-CT experiment, the implant
Micro-CT Examination was virtually embedded in polyurethane resin, and
The implant was embedded with reinforced polyurethane a load cap was placed onto the abutment. To reduce
resin (PUR, Alpha Die Top, Schütz Dental) in a stainless computational time, the abutment screw was modeled
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
1 2 3 4 5
20 mm
Fig 1 Loading jig with (1) tilted ceramic loading plate with frictionless linear ball bearing
to avoid generation of transverse force components, (2) embedded implant with load
cap, (3) loading piston, (4) loading spring, and (5) adjusting screw.
with Rm being the ultimate strength, Rp0.2 the yield 2.3 mm S Detaching
contact zones of
strength, and εm the ultimate strain. The yield strain model C (red)
εp0.2 is derived from
Implant
Rp0.2
εp0.2 = Embedding
Ε (2) resin
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
Table 2 Configuration of the Three Model degrees), dH is the head diameter (2.35 mm), dB is the
Variants According to Contact and bore diameter (2.03 mm), and µH is the coefficient of
Loading Conditions (Contact Zone friction at the screw head (0.4). In the second load step,
Designation According to Fig 2) the oblique external load of 200 N was applied as in
the analysis of model A. To ensure static equilibrium of
Contact behavior the models, all bottom nodes of the embedding resin
Contact zone I Contact zone S were held fixed. The loading conditions chosen for each
(Implant collar/ (Abutment screw/
Model embedding resin) implant ) Preload analysis are also summarized in Table 2.
Tetrahedral solid elements with quadratic trial function
A Bonded Bonded Not
applied (element type SOLID187) were used. After appropriate
B Bonded Bonded Applied
refinement at critical locations using semiautomatic mesh-
ing, the meshes consisted of a total of 347,515 (A and B)
C Detachment Detachment Applied
enabled enabled
and 399,913 (C) elements. Figure 3 shows cross-sectional
views of the meshed models used for the analyses. Very
small elements (size 0.05 to 0.1 mm) were chosen in the
that fixed contact was assumed between the load cap vicinity of contact areas between the abutment screw
and abutment, and frictional asymmetric surface-to- and the abutment, as well as between the abutment
surface contact was chosen at the screw-abutment and and implant. Due to different contact definitions, mesh
implant-abutment interfaces (marked in blue in Fig 2). sizes differ between models A/B and C.
In models A and B, the implant, embedding resin, For the validation of the finite element models, data
and abutment screw were combined to form one com- were available from five previous load experiments29 with
ponent assembly with bonded contacting faces. In identical specimens and a set-up identical to that used
contrast, detachment in the contact zones S and I in in the numerical analyses. Especially, conditions at the
Fig 2 was permitted in model C. This corresponded to implant-abutment interface (frictional contact) were the
the findings of a previous study,29 in which detachment same as defined in the FEA. The forces and correspond-
was observed in the respective areas. An overview of ing displacements (in force direction) of the point of
the contact definitions is given in Table 2. Contacting load application at the load cap had been recorded in
faces were always modeled using target and contact the experiments and could be extracted from the FEA
elements (TARGET170 and CONTA174 in ANSYS), with a calculations as well. Agreement between the experimental
friction coefficient of 0.4. For problem solving, the pure and numerical load-displacement curves was taken as
penalty method was applied with the contact stiffness evidence for the validity of the respective model.
factor set to 1.0. Furthermore, the stress results of the FEAs of model A
In the analysis of model A, the screw preload due to to C were compared at selected areas of the implant and
the tightening torque was neglected, and the implant- abutment to reveal differences due to different loading
abutment complex was directly loaded with an external and contact conditions. Additionally, abutment regions
force of 200 N, built up incrementally and acting on the were examined with high stresses after screw tightening
load cap at an angle of 30 degrees with respect to the (at the screw seat) and after external loading (at the
implant axis. To simulate the posterior implant position in abutment hexagon).
the mandible, the load was directed toward the bevelled Finally, the postprocessor of the program was used to
surface of the abutment (Fig 2). In contrast, two load evaluate the distribution of the microgap width between
steps were applied in the analyses of models B and C. In the contacting surfaces of the abutment and implant,
the first load step, the tightening torque was simulated defined as the distance between neighboring nodes of
by a preload (FP) of 271 N generated by contraction of previously specified contact and target elements in the
the screw shaft. FP was calculated from the tightening direction of the surface normal.
torque TA, as recommended by the manufacturer, using
the following formula30:
RESULTS
TA
FP =
d × μT dH + dB Evaluation of Microgap Formation with
0,159 • P + S + × μH Micro-CT
α
2 × cos (2 (4
(3) The results of the micro-CT experiment are illustrated
in Fig 4. The images show sections of the implant-
where P is the thread pitch (0.35 mm), dS is the screw abutment assembly under the oblique load of 200 N.
shaft diameter (1.83 mm), µT is the coefficient of friction Two mutually orthogonal midsections (in the xy and yz
at the thread (0.4), a is the flank angle of the thread (60 planes) are shown, together with a cross section in the
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
A/B C F
Possible start
of microgap
Y Y
a X b X
Z X
c Y d Z
10.00 mm
1 mm
Fig 3 Cross-sectional view of the meshed models Fig 4 Micro-CT sections of implant-abutment assembly ([b, c] midsec-
(left) A and B and (right) C. tions in xy and yz planes, respectively; [a] xy section enlarged 5× [red
box]; [d] cross section in xz plane approximately 65 µm below entrance
level of abutment into implant); implant under 200-N load (30 degrees).
Force components in respective planes indicated on right side. F = force.
100 Experiment 1
clearly detected. This leads to the conclusion that any 80 A
Experiment 2
putative gaps would have to be considerably smaller B Experiment 3
60 Experiment 4
than two voxel edge lengths, ie, 13 µm. C
40 Experiment 5
20
Comparison of Load-Displacement Curves of
0
FEA and Experimental Results 0.00 0.05 0.10 0.15 0.20 0.25 0.30
The force-displacement curves as determined by exper-
Displacement (mm)
iments and FEAs are compared in Fig 5. All curves run
almost linearly up to a displacement of 0.05 mm, but Fig 5 Force-displacement curves calculated in FEAs for models
A, B, and C and experimental data for five identical specimens of
then exhibit a transition to smaller slopes, possibly due the same type as used in this study according to Dittmer et al.29
to the onset of plastic deformation or beginning relative
motion of contacting parts. The experimental curves, Von Mises Stresses in Abutment and Implant
generated with identical specimens, (Fig 5, experiments The plots of von Mises stresses in the abutment (Fig
1 to 5) are less steep than the FEA graphs and vary con- 6) show similar distributions for all three models. The
siderably. Of the computed curves, those for models highest stresses arose in the hexagon region, with the
A and C are very similar and, up to a displacement of highest stress levels present in model A. Models B and
0.05 mm and a force of 80 N, are almost identical to the C exhibited almost identical distributions, with slight-
graph for experiment 4. With increasing displacement ly higher stresses at the screw base than for model A.
and force, however, the computed curves deviate more A more thorough assessment and better compara-
and more from the experimental curves. The closest bility are provided by the stress profiles given in Figs
approximation to experimental data was reached with 7 and 8, which were each taken along identical paths
model C, the model with the most realistic simulation at the abutment surface for all three models. Figure 7
of contact conditions. shows a plot of von Mises stresses after loading with
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
A B C
10.00 mm
250
200
150
A B
100
50
0
0 1 2 3 4 5 6 7
A Path (mm) B
350
300
250
200
150
Model A
100 Model B
Model C
50
0
0 0.2 0.4 0.6 0.8 1 1.2 1.4
Path (mm)
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
A B C
10.00 mm
Model C
400
300
200
100
0
0 2 4 6 8 10 12
A B
Path (mm)
200 N along a circular path at the abutment-screw profiles showed stress minima, the highest of which
seat. Different stress levels were evaluated for all mod- (approximately 380 MPa) was evaluated for model
els. The lowest stress values—between approximately A, whereas models B and C exhibited similar stress
5 and 80 MPa, with a maximum at the side facing curves, with much lower minima of approximately
toward the force (point A in Fig 7)—were determined 280 MPa.
for model A. The highest stresses, with strong local For the implant, all analyses found that the highest
variations but similar courses, occurred with models von Mises stresses developed at the implant collar
B and C. They ranged between 130 and 190 MPa, (Fig 9). Local stress maxima occurred at the side facing
were slightly higher for model C than for model B, away from the applied force, both at the collar and,
and exhibited a slight increase at the side facing away somewhat lower, on the outside of the implant at the
from the force. These values were much less than the crestal level. At the seat of the abutment screw, the
material’s yield strength (483 MPa). highest stresses were found for model C.
Figure 8 displays von Mises stresses along a path The von Mises stress profiles along a circular path
at the edge of the abutment hexagon facing toward at the implant collar demonstrate that yield strength
the external force. With all three models, the high- was reached and plastic deformation could take place
est von Mises stresses, exceeding the yield strength at the side opposite to the external oblique force of
(483 MPa) of the material, were observed at both ends 200 N (Fig 10). The area of plastic deformation was
of the path of evaluation. In the middle of the path, all larger for model A than for models B and C.
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
2.00 mm
Evaluation of Microgap Formation with FEA Another factor influencing the FEA results is the choice
Microgap widths determined by FEA for all three models of contact conditions for the interfaces between differ-
are presented in Fig 11. Gaps in all three models emerged ent parts of the implant assembly. While friction has an
in a semicircular shaped area at the side facing toward important impact on stresses in the implant components
the load. The largest gaps, with widths ranging between and on microgap formation between the implant and
2 µm at the bottom of the implant-abutment interface abutment,31–33 the exact coefficient of friction for the
and approximately 20 µm at the implant collar, were implant materials titanium and titanium alloy is unknown
calculated for model A, where screw preload had been and depends on several conditions, including surface
neglected. In contrast, gaps determined with both topography and the presence of lubricant agents. Friction
model B and model C showed the highest widths of coefficients given in the literature differ significantly,
approximately 10 µm only near the implant collar and ranging from 0.3 to 0.5.34–36 In the present analysis, a
were confined to a small semicircle with a radius of friction coefficient of 0.4 was assumed to simulate dry
approximately 1.5 mm. conditions between contacting surfaces. If lubricants
were present in the respective interfaces, the coefficient
of friction would decrease. This would result in higher
DISCUSSION screw preloads, as evaluated in a previous study,33 and
therefore lead to somewhat smaller microgap widths and
The aim of this study was to investigate microgap forma- slightly larger regions of possible plastic deformation
tion in an implant-abutment complex by FEA and micro- than determined here.
CT. Three finite element models of different complexity Different material models were chosen to represent
were generated and verified by comparing analysis the behavior of the various model constituents. Linear
results with previously presented force-displacement elastic properties were assigned to the load cap (NiCr
behavior in real load tests29 and gap measurements in alloy) and embedding resin (PUR). Because of the rela-
micro-CT experiments. To the authors’ knowledge, this tively low stresses in the load cap, this seems to be most
method for finite element model validation has not yet appropriate for NiCr. However, von Mises stresses in the
been used in other studies of this kind. embedding resin of model A, for example, turned out
To reduce the complexity of the finite element models to be in a range of approximately 30 to 130 MPa after
and to significantly save central-processing-unit time, loading with 200 N and thus sometimes exceeded the
several simplifications had to be adopted. Furthermore, compressive strengths of reinforced polyurethanes, which
specific boundary conditions had to be chosen, without lie between 80 and 170 MPa.37 This suggests that plastic
precise knowledge of relevant parameters. The possible deformation might have occurred in the resin, which is
consequences are discussed in the following. consistent with the more flattened course of experimental
Cutting off the 0.1-mm-deep microthread at the load-displacement curves beyond a force of approxi-
implant neck created a slightly more resilient support mately 80 N, as compared with the curves determined
of the abutment and may have led to slightly overes- by FEA (Fig 5). Nevertheless, this effect should have had
timated implant displacements, gap widths, and local a greater influence on the load-displacement behavior
stresses in the analyses. In contrast, modeling the abut- of the implant-abutment complex as a whole than on
ment screw thread as a cylinder, by filling the thread the stresses and gap formation within the complex.
depth (approximately 0.2 mm) with material, slightly In contrast to most other studies, in which linear
stiffened the implant assembly and counteracted the elastic material behavior was assumed for all materials
aforementioned effect. of the implant-abutment complex,18–20,38,39 the present
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
study introduced bilinear stress-strain curves for titanium and C, as compared with that for model A (Fig 10). The
and titanium alloy to simulate material behavior more region with stresses above the yield strength, however,
realistically, thus taking plastic deformation into account. stretched for about one-quarter of the path along the
This approximation, though incomplete because of the implant collar in all three models, thus indicating possible
lack of availability of true stress-strain curves for the plastic deformation in this area due to horizontal forces.
materials in question, proved to be necessary, as local It must be pointed out here that all analyses presented
stresses indeed exceeded the material’s yield strength were carried out with implants embedded only up to a
(see Figs 8 and 10). level of 3 mm below its collar, thus simulating the clinically
As expected, the most complex finite element model unfavorable case of substantial bone loss as prescribed
(C) with the most realistic contact definitions delivered by ISO 14801.24 Therefore, the numerical results are to be
the best approximation to the nonlinear experimental treated with caution and cannot readily be transferred
load-displacement behavior of the embedded implant- to the clinical situation. With an implant fully inserted
abutment complex (Fig 5). The differences between into alveolar bone and thus better supported against
the corresponding loads on the theoretical load-dis- lateral forces, maximum stresses as well as gap widths
placement curve and the nearest experimental curve and displacements are most likely to be smaller than
are negligible for small displacements, but gradually calculated in the analysis of the most realistic model
increase to a maximum of approximately 15% for higher C, and the risk of plastic deformation would therefore
displacements. This is understandable in light of the considerably diminish.
aforementioned limitations of the FEA, and the overall In comparison with the introduction of screw pre-
agreement of experimental and numerical results is load, allowing detachment at the interfaces between
regarded as reasonable evidence for the validity of the implant and embedding resin as well as between
model variant C. If, for example, possible plastic behavior the implant and abutment screw, though essential for
of PUR could have been incorporated in the model, the the quality of approximation to experimental load-
degree of approximation would have been even better, displacement curves, resulted in only a small effect on
as relatively low stiffness of the embedment would have stresses (compare results for models B and C in Figs 6
allowed higher implant displacements. In contrast, the to 10). The biggest difference can be observed at the
neglect of the possibility of detachment in the implant/ screw seat opposite to the side of force application
embedding resin and implant-abutment screw interface (Fig 7), where stresses rose by approximately 20%, as
regions in model B led to the load-displacement curve a greater share of the load must be supported at this
being approximately 15% steeper than that calculated side when detachment is allowed, and fixed contacts at
with model C. This underlines the importance of care- these interfaces were omitted that would have assured
fully defining contact conditions in complex multipart support against bending.
assemblies in this kind of FEA. Moreover, disregarding FEA revealed a detailed picture of gap width for-
the pretension of the screw (model A) allowed the abut- mation between the implant and abutment under
ment more play in its socket. This increased force-driven an oblique load. For the most realistic of the three
displacement and, finally, resulted in almost as good of models investigated, namely, model C, taking into
an approximation to experimental curves as with model account the abutment screw preload, the gap width
C, but in completely different results for stresses and gap was largest (10 µm) at the side of load application near
widths in the assembly. the implant collar, but vanished within a distance of
Of the different boundary conditions in the model approximately 1.5 mm from that point (Fig 11). Thus,
variants, the introduction of screw preload in models B continuous access for bacteria to the dead space in the
and C had the highest influence on stresses in both the implant underneath the abutment was not created.
implant and abutment. Tightening the screw understand- This would have allowed steady presence, growth,
ably resulted in higher stresses at the screw seat (Fig and colony forming of anaerobic pathogens there
7), pressed the abutment into the fixture, and thereby and would have presented by far a higher risk for
reduced the ability of the abutment to tilt under the the peri-implant health than the possible temporary
influence of oblique forces. As a consequence, the high entering of oral bacteria into the said small semicir-
stresses, which were caused by leverage effects at the cular gap. A free access to the space underneath the
lower hexagon edge of the abutment at the side of force abutment, however, was indicated in the analysis of
application in model A, were alleviated by up to 20%, model A, in which screw preload was neglected and,
and the risk of exceeding the yield strength (483 MPa) as a consequence of missing counteraction against
was considerably reduced (Fig 8). At the same time, the the external tilting force, the maximum gap width was
better fit of the abutment and socket after screw tighten- twice as high (approximately 20 µm) as for the more
ing distributed stresses more evenly and explained the realistic model C. Since Pessoa et al,18 in an FEA for a
absence of local stress maxima in curves for models B similar implant under almost identical external load
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Jörn et al
8. Brägger U, Aeschlimann S, Bürgin W, Hämmerle CH, Lang NP. Bio- 24. International Organization for Standardization (ed). Standard ISO
logical and technical complications and failures with fixed partial 14801:2007 Dentistry - Implants - Dynamic fatigue test for endosse-
dentures (FPD) on implants and teeth after four to five years of ous dental implants. 2007; International Organization for Standard-
function. Clin Oral Implants Res 2001;12:26–34. ization, Geneva, Switzerland.
9. Malathi GK, Chandra RP. Mechanical complications with implants 25. Deutsches Institut für Normung e.V. (ed). Standard DIN 17 862
and implant prostheses. IJDA 2011;3:555–558. Stangen aus Titan und Titanlegierungen, Technische Lieferbedin-
10. Piattelli A, Scarano A, Paolantonio M, et al. Fluids and micro- gungen. 1993; Beuth Verlag, Berlin, Germany.
bial penetration in the internal part of cement-retained versus 26. ASM International (ed). Metals and alloys: Materials and coatings for
screw-retained implant-abutment connections. J Periodontol medical devices: Cardiovascular. 2009; 1st ed. ASM International,
2001;72:1146–1150. Materials Park, Ohio, USA.
11. Lorenzoni FC, Coelho PG, Bonfante G, et al. Sealing capability and 27. Dittmer MP. Ermittlung der Spannungsverteilung in viergliedri-
SEM observation of the implant-abutment interface. Int J Dent gen Zirkoniumdioxidbrücken mit Hilfe der Methode der Finiten
2011;2011:864183. doi: 10.1155/2011/864183 Elemente. Dissertation (Dr. med. dent.), Medizinische Hochschule
12. Abdelhamed MI, Galley JD, Bailey MT, et al. A comparison of zirconia Hannover, Hannover, Germany, 2007.
and titanium abutments for microleakage. Clin Implant Dent Relat 28. Jörn D, Waddell J, Swain M. The influence of opaque application
Res 2015;17(suppl 2):e643–e651. methods on the bond strength and final shade of PFM restorations.
13. Quirynen M, Bollen CM, Eyssen H, van Steenberghe D. Microbial J Dent 2010;38(suppl 2):e143–e149.
penetration along the implant components of the Brånemark 29. Dittmer S, Dittmer MP, Kohorst P, et al. Effect of implant-abutment
system. An in vitro study. Clin Oral Implants Res 1994;5:239–244. connection design on load bearing capacity and failure mode of
14. Rack A, Rack T, Stiller M, et al. In vitro synchrotron-based radiogra- implants. J Prosthodont 2011;20:510–516.
phy of micro-gap formation at the implant-abutment interface of 30. Bickford JH. An Introduction to the Design and Behavior of Bolted
two-piece dental implants. J Synchrotron Radiat 2010;17:289–294. Joints. New York: Marcel Dekker, 1995.
15. Meleo D, Baggi L, Di Girolamo M, et al. Fixture-abutment connec- 31. Sakaguchi RL, Borgersen SE. Nonlinear contact analysis of
tion surface and micro-gap measurements by 3D micro-tomo- preload in dental implant screws. Int J Oral Maxillofac Implants
graphic technique analysis. Ann Ist Super Sanita 2012;48:53–58. 1995;10:295–302.
16. Zipprich H, Weigl P, Lange B, Laue HC. Erfassung, Ursachen und 32. Merz B, Hunenbart S, Belser U. Mechanics of the implant-abutment
Folgen von Mikrobewegungen am Implantat-Abutment-Interface. connection: An 8-degree taper compared to a butt joint connec-
Implantologie 2007;15:31–46. tion. Int J Oral Maxillofac Implants 2000;15:519–526.
17. Geng JP, Tan KB, Liu GR. Application of finite element analysis 33. Jörn D, Kohorst P, Besdo S, et al. Influence of lubricant on screw
in implant dentistry: A review of the literature. J Prosthet Dent preload and stresses in a finite element model for a dental implant.
2001;85:585–598. J Prosthet Dent 2014;112:340–348.
18. Pessoa RS, Muraru L, Júnior EM, et al. Influence of implant con- 34. Keith O, Kusy RP, Whitley JQ. Zirconia brackets: An evaluation of
nection type on the biomechanical environment of immediately morphology and coefficients of friction. Am J Orthod Dentofacial
placed implants - CT-based nonlinear, three-dimensional finite Orthop 1994;106:605–614.
element analysis. Clin Implant Dent Relat Res 2010;12:219–234. 35. Seitzman LE, Bolster RN, Singer IL. IBAD MoS lubrication of titanium
19. Saidin S, Abdul Kadir MR, Sulaiman E, Abu Kasim NH. Effects of alloys. Surf Coat Tech 1996;78:10–13.
different implant-abutment connections on micromotion and 36. Abkowitz S, Burke JJ, Hiltz RH. Titanium in Industry. New York: Van
stress distribution: Prediction of microgap formation. J Dent Nostand, 1995 .
2012;40:467–474. 37. Gegauff AG, Garcia JL, Koelling KW, Seghi RR. Thermoplastic
20. Çaglar A, Bal BT, Karakoca S, et al. Three-dimensional finite element composites for veneering posterior teeth—a feasibility study. Dent
analysis of titanium and yttrium-stabilized zirconium dioxide abut- Mater 2002;18:479–485.
ments and implants. Int J Oral Maxillofac Implants 2011;26:961–969. 38. Alkan I, Sertgöz A, Ekici B. Influence of occlusal forces on stress
21. Chun HJ, Shin HS, Han CH, Lee SH. Influence of implant abutment distribution in preloaded dental implant screws. J Prosthet Dent
type on stress distribution in bone under various loading condi- 2004;91:319–325.
tions using finite element analysis. Int J Oral Maxillofac Implants 39. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to
2006;21:195–202. determine implant preload. J Prosthet Dent 2003;90:539–546.
22. Wu T, Liao W, Dai N, Tang C. Design of a custom angled abutment 40. Burghardt AJ, Link TM, Majumdar S. High-resolution computed
for dental implants using computer-aided design and nonlinear tomography for clinical imaging of bone microarchitecture. Clin
finite element analysis. J Biomech 2010;43:1941–1946. Orthop Relat Res 2011;469:2179–2193.
23. Canullo L, Pace F, Coelho P, Sciubba E, Vozza I. The influence of 41. Barrett JF, Keat N. Artifacts in CT: Recognition and avoidance. Radio-
platform switching on the biomechanical aspects of the implant- graphics 2004;24:1679–1691.
abutment system. A three dimensional finite element study. Med
Oral Patol Oral Cir Bucal 2011;16:e852–e856.
© 2016 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.