CA-104 E 07 Instrument Errors

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 15

Edition: 3

Chapter 7: Instrument Errors


List of Instrument Errors .......................................................................................................................... 1
Instrument Errors Details ......................................................................................................................... 3

CA-104 S/M November 2013


List of Instrument Errors

The analyzer does not start


The display shows no characters
The display shows block characters

The analyzer does not heat up


The analyzer gets very hot
Display shows “COOL DOWN”

No correct Keyboard function 2-Channel


No correct Keyboard function 4-Channel

Date and time setup are not OK

No automatic printout
No manual printout

No automatic printout on external printer


No manual printout on external printer
Printer prints “crazy” characters on external printer

Display shows “meas // plot”

Display shows flashing “P”


Display shows flashing “E”
Display shows flashing “e”
Display shows flashing “F”
Display shows flashing “W”
Display shows flashing “D”

Error message “break: motor”

Error message “ERROR: xxx”


Error message “ERROR recursive…”
Error message “ERROR REGLER: re_adc(): adc_err=3 $0,$0”
Error message “WATCHDOG: eewrite(3): EEPROM faulty!$0,$0”
Error message “parameter error! Press `7`…”

Display message “timeout”


Display message “bottom”
Display message “dark”
Display message “top”
Display message “drift”
Display message “break: readjust”
Display message “break: jump”

CA-104 S/M 7-1 November 2013


List of Instrument Errors contd.

Self starting measurement

No starting measurement

Liquid has pipetted direct into measuring channel

Motor is making sound while spinning

Display message “TEMP // WARN”


Printout “Warn: Temp instable”
Display message “STIR // WARN”
Printout “Warn: Stirrer rpm error”

CA-104 S/M 7-2 November 2013


Instrument Errors Details

Problem Possible solution


The analyzer does not start - Check power connections
- Echeck external power supply
- Check secondary fuse (internal)
- Check power supply 5VDC / 24VDC
Change power supply (external)
Change PCB Ext. DC-Power (internal)
- Check PCB mainboard Voltage LEDs
5VDC / 12VDC / 24VDC
- Check PCB mainboard
Change PCB mainboard

The display shows no characters - Check power connections


There are no beep tones by pressing keyboard - Echeck external power supply
There are no beep tone while starting up - Check secondary fuse (internal)
- Check power supply 5VDC / 24VDC
Change power supply (external)
Change PCB Ext. DC-Power (internal)
- Check PCB mainboard Voltage LEDs
5VDC / 12VDC / 24VDC
- Check PCB mainboard
Change PCB mainboard

The display shows no characters - Check display contrast


There are single beep tones while starting up
The display shows no characters - Check internal cable connection between
There is a continuous beep tone every second display and PCB Mainboard
- Check LCD-Module (display)
Change display
- Check PCB mainboard
Change PCB mainboard

The display shows “block” characters - Check LCD-Module (display)


There are no beep tones by pressing keyboard Change display
There are no beep tone while starting up - Check PCB mainboard
Change PCB mainboard

CA-104 S/M 7-3 November 2013


Instrument Errors Details contd.

Problem Possible solution


The analyzer does not heat up - Recalibrate thermal block temperature
(Thermal Block) - Check thermal block Connector to PCB
Mainboard
- Check thermal block heating
- Check thermal block sensor (100K at 25°C)
- Change thermal block
- Change PCB Mainboard

The analyzer gets very hot - Recalibrate thermal block temperature


(Thermal Block) - Check thermal block Connector to PCB
Mainboard
- Check thermal block sensor (100K at 25°C)
- Change thermal block
- Change PCB Mainboard

Display shows “COOL DOWN” - Check if analyzer is placed in direct sunlight


- Check if analyzer is placed next to a heating
- check present thermal block temperature
with external thermometer
- Recalibrate thermal block temperature
- Check thermal block connector to PCB
Mainboard
- Check thermal block sensor (100K at 25°C)
- Change thermal block
- Change PCB Mainboard

CA-104 S/M 7-4 November 2013


Instrument Errors Details contd.

Problem Possible solution


No correct Keyboard function 4-Channel - Check if 2-Channel Software is installed
- Check if keyboard is damaged
- Check if keyboard is connected properly
(inside analyzer)
- Check if keyboard cable is damaged
(inside analyzer)
- Change keyboard
- Change PCB Mainboard

Problem Possible solution


Date and time setup are not OK - Check date and time in UTILITIES menu
- Change PCB Mainboard

Date and time setup are not OK - Check date and time in UTILITIES menu
When analyzer was switched off - Replace battery
- Change PCB Mainboard

CA-104 S/M 7-5 November 2013


Instrument Errors Details contd.

Problem Possible solution


No automatic printout - Check if printer is switched “Auto”
Results should be printed when clotting is found - Check printer connection (cable connection)
- Check printer paper
- Check paper sensor (printer)

No manual printout - Check if printer is switched “ON”


Results should be printed when displayed and - Check printer connection (cable connection)
Ch x-key is pressed - Check printer paper
- Check paper sensor (printer)

Problem Possible solution


No automatic printout on external printer - Check if printer is switched ON
Results should be printed when clotting is found - Check if printer is ONLINE
- Check printer connection (cable connection)
- Check analyzer setup
Port A = ext. printer
printer = Auto
- Check correct setup of external printer

No manual printout on external printer - Check if printer is switched ON


Results should be printed when displayed and - Check if printer is ONLINE
Ch x-key is pressed - Check printer connection (cable connection)
- Check analyzer setup
Port A = ext. printer
printer = ON
- Check correct setup of external printer

Printer prints “crazy” characters - Check analyzer setup


on external printer Port A = ext. printer
printer = Auto
- Check correct setup of external printer

CA-104 S/M 7-6 November 2013


Instrument Errors Details contd.

Problem Possible solution


Display shows “meas // plot” - Press CH x – key to show result
No result is shown / printed
- Check analyzer setup
Port A = OFF (!)

CA-104 S/M 7-7 November 2013


Instrument Errors Details contd.

Problem Possible solution


Display shows flashing“P” - Remove SD-Card
P = write protected - disable write protection on SD-Card (switch)

Display shows flashing “E” - Check Size of SD-Card


E = error reading / writing 128MB, 256MB, 512 MB, 1GB
- Check Type of SD-Card
no SD-HC-Card alowed
no SD-XD-Card allowed
- Check file structure FAT on SD Card
- Check for enough memory space on SD-Card
- Format SD-Card with FAT file system on
digital photo camera
- Change SD-Card

Display shows flashing “e” - Check Size of SD-Card


e = error reading / writing 128MB, 256MB, 512 MB, 1GB
- Check Type of SD-Card
no SD-HC-Card alowed
no SD-XD-Card allowed
- Check file structure FAT on SD Card
- Check for enough memory space on SD-Card
- Format SD-Card with FAT file system on
digital photo camera
- Change SD-Card

Display shows flashing “f” - Check Size of SD-Card


f = failure reading / writing 128MB, 256MB, 512 MB, 1GB
- Check Type of SD-Card
no SD-HC-Card alowed
no SD-XD-Card allowed
- Check file structure FAT on SD Card
- Check for enough memory space on SD-Card
- Format SD-Card with FAT file system on
digital photo camera
- Change SD-Card

Display shows flashing “W” - writing on SD-Card


W = writing remove SD-Card if required

Display shows flashing “D” - writing on SD-Card


D = deleting remove SD-Card if required

CA-104 S/M 7-8 November 2013


Instrument Errors Details contd.

Problem Possible solution


Error Message “break: motor” - Check of dirt in measuring channel
Less than three beep tones after switching on - Check of dirt on magnet (inside analyzer)
- check if parts at magnet may block the motor
example: mixing-bar
- Check distance magnet vs. sensor
- Check if motor is spinning in service mode
- Check if speed sensor is detecting correctly
- Change motor
- Change PCB mainboard

Problem Possible solution


Error Message “ERROR: xxx” - Contact LABiTec’s
Xxx = specific note service & support department

Error Message “ERROR recursive” - Contact LABiTec’s


service & support department

Error Message - Contact LABiTec’s


>>> ERROR <<< service & support department
REGLER: re_adc(): adc_err=3 $0,$0 - Change PCB mainboard

Error Message - Press Enter to quit message


WATCHDOG: eewrite()3: EEPROM faulty!$0,$0 - Press 7 key to load factory default parameters
ERROR - calibrate thermal block temperature
- re enter conversion values
(calibration values)

Error Message “parameter error! Press - Press 7 key to load factory default parameters
`7`…” - calibrate thermal block temperature
- re enter conversion values
(calibration values)

CA-104 S/M 7-9 November 2013


Instrument Errors Details contd.

Problem Possible solution


Display Message “timeout” - Check for continuous workflow while pipetting
the “wait for reagent time” (xxx ul // GO-S)
exceeds 180 sec,

Display Message “timeout” - Check for correct pipetting sample and reagent
the measuring time exceeds 420 sec, - Check for existing clot formation inside cuvette

Display Message “bottom” - Check for turbid fluid inside cuvette

Display Message “dark” - Check for turbid fluid inside cuvette

Display Message “top” - Check for clear fluid inside cuvette


- Check for direct ambient light
from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement

Display Message “drift” - Check for air bubbles inside cuvette


- Check for correct test volume inside cuvette
- Check for pipetting direct into cuvette
Rinsing fluid on cuvette wall is not allowed
- Check for a clean measuring chamber
- Check for foreign parts in bottom of measuring
channel (example: mixer-bar)
- Check for direct ambient light
from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement

Display Message “break: readjust” - Check for air bubbles inside cuvette
Error while adjustment. - Check for correct test volume inside cuvette
- Check for pipetting direct into cuvette
Rinsing fluid on cuvette wall is not allowed
- Check for a clean measuring chamber
- Check for foreign parts in bottom of measuring
channel (example: mixer-bar)
- Check for direct ambient light
from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement
- Check for test application; learn + lag is not
appropriate as required settings.
- Check for fluid inside cuvette, turbity could
change too fast.

Display Message “break: jump” - Check for air bubbles inside cuvette
Clot detection may be caused by signal jump
maybe caused by air bubble

CA-104 S/M 7-10 November 2013


CA-104 S/M 7-11 November 2013
Instrument Errors Details contd.

Problem Possible solution


Self starting measurement - Remove cuvette from measuring channel
Incubation timer starts without cuvette - Close light protection cap
- Switch OFF analyzer
- Switch ON analyzer
- Wait until message “remove cuvette…”
- Make sure no cuvette is in measuring channel
- close light protection cap
- Press Start-key for “autoblanking” process
- keep light protection cap is closed while
processing

Self starting measurement - Check for direct ambient light


Incubation timer starts without cuvette from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement
- repeat “autoblanking” process as described
before

Self starting measurement - Remove used cuvette immediately after result


Incubation timer starts without cuvette appeared
- keep light protection cap closed when waiting
for measurement and in standby
- repeat “autoblanking” process as described
before

Self starting measurement - Check for correct test-setup


Incubation timer starts without cuvette (at each test application)
Cuv remove detection = OFF
- repeat “autoblanking” process as described
before

Self starting measurement - Check for air bubbles inside cuvette


Measuring timer starts without adding start - Check for correct test volume inside cuvette
reagent - Check for pipetting direct into cuvette
Rinsing fluid on cuvette wall is not allowed
- Check for a clean measuring chamber
- Check for foreign parts in bottom of measuring
channel (example: mixer-bar)
- Check for direct ambient light
from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement

Self starting measurement - Check if cuvette detection is “confused”


Measuring timer starts without adding start - repeat “autoblanking” process as described
reagent before

CA-104 S/M 7-12 November 2013


Instrument Errors Details contd.

Problem Possible solution


No starting measurement - Check for enabled start-pipette
Measurement does not start after adding start set Port A = OFF
reagent

No starting measurement - Dispense straight into the cuvette from centre of


Measurement does not start after adding start the light protection cap
reagent - Care for correct handling while adding
start reagent

Problem Possible solution


Liquid has pipetted direct into measuring - Aspirate fluid with a pipette out of the channel
channel - Disassemble analyzer
There was no cuvette inside before pipetting - Clean thermal block with alcohol and fleece
- Clean optics on PCB with alcohol and fleece
carefully
- Clean magnet with alcohol and fleece
- Check / adjust distance between magnet and
speed sensor
- Reassemble analyzer

Problem Possible solution


Motor is making sound while spinning - replace the motor

Note:
During the lifetime of the motor it is giving more
or less sound while spinning.
If there is no further error message in display,
the motor is not defect.

CA-104 S/M 7-13 November 2013


Instrument Errors Details contd.

Problem Possible solution


Display message “TEMP // WARN” Warning flag
Printout "WARN: Temp. instable" - Check if result is valid
After measurement - Check if analyzer is placed in direct sunlight
- Check if analyzer is placed next to a heating
- check present thermal block temperature
Note: with external thermometer
- Recalibrate thermal block temperature
This warn flag appears if - Check thermal block connector to PCB
thermalblocktemperature was too hot while Mainboard
measurement. - Check thermal block sensor (100K at 25°C)
- Change thermal block
- Change PCB Mainboard

Display message “TEMP // WARN” Warning flag


Printout "WARN: Temp. instable" - Check if result is valid
After measurement - check present thermal block temperature
with external thermometer
- Recalibrate thermal block temperature
Note: - Check thermal block connector to PCB
This warn flag appears if Mainboard
thermalblocktemperature was too cold while - Check thermal block sensor (100K at 25°C)
measurement. - Change thermal block
- Change PCB Mainboard

Display message “STIR // WARN” - Check if Teflon stirrer for stirring the reagent
Printout "WARN: Stirrer rpm error" has the correct size (3x13mm)
After measurement - Check if parts at magnet may block the motor
example: mixing-bar, washer, etc.
Note: - Check distance magnet vs. sensor
This warning flag appears if speed of reagent- - Check if motor is spinning in service mode
stirrer-motor was to low before or while - Check if speed sensor is detecting correctly
measurement. - Change motor
- Change PCB mainboard

Display message “STIR // WARN” - Check if parts at magnet may block the motor
Printout "WARN: Stirrer rpm error" example: mixing-bar, washer, etc.
After measurement - Check distance magnet vs. sensor
- Check if motor is spinning in service mode
Note: - Check if speed sensor is detecting correctly
This warning flag appears if speed of reagent- - Change motor
stirrer-motor was to high before or while - Change PCB mainboard
measurement.

CA-104 S/M 7-14 November 2013