CA-104 E 07 Instrument Errors
CA-104 E 07 Instrument Errors
CA-104 E 07 Instrument Errors
No automatic printout
No manual printout
No starting measurement
Date and time setup are not OK - Check date and time in UTILITIES menu
When analyzer was switched off - Replace battery
- Change PCB Mainboard
Error Message “parameter error! Press - Press 7 key to load factory default parameters
`7`…” - calibrate thermal block temperature
- re enter conversion values
(calibration values)
Display Message “timeout” - Check for correct pipetting sample and reagent
the measuring time exceeds 420 sec, - Check for existing clot formation inside cuvette
Display Message “break: readjust” - Check for air bubbles inside cuvette
Error while adjustment. - Check for correct test volume inside cuvette
- Check for pipetting direct into cuvette
Rinsing fluid on cuvette wall is not allowed
- Check for a clean measuring chamber
- Check for foreign parts in bottom of measuring
channel (example: mixer-bar)
- Check for direct ambient light
from lamp or direct sunlight
- Close light protection cap
- Change analyzers placement
- Check for test application; learn + lag is not
appropriate as required settings.
- Check for fluid inside cuvette, turbity could
change too fast.
Display Message “break: jump” - Check for air bubbles inside cuvette
Clot detection may be caused by signal jump
maybe caused by air bubble
Note:
During the lifetime of the motor it is giving more
or less sound while spinning.
If there is no further error message in display,
the motor is not defect.
Display message “STIR // WARN” - Check if Teflon stirrer for stirring the reagent
Printout "WARN: Stirrer rpm error" has the correct size (3x13mm)
After measurement - Check if parts at magnet may block the motor
example: mixing-bar, washer, etc.
Note: - Check distance magnet vs. sensor
This warning flag appears if speed of reagent- - Check if motor is spinning in service mode
stirrer-motor was to low before or while - Check if speed sensor is detecting correctly
measurement. - Change motor
- Change PCB mainboard
Display message “STIR // WARN” - Check if parts at magnet may block the motor
Printout "WARN: Stirrer rpm error" example: mixing-bar, washer, etc.
After measurement - Check distance magnet vs. sensor
- Check if motor is spinning in service mode
Note: - Check if speed sensor is detecting correctly
This warning flag appears if speed of reagent- - Change motor
stirrer-motor was to high before or while - Change PCB mainboard
measurement.