Professional Documents
Culture Documents
Complete Guide - ISO 13485 - 2016 QMS For Medical Device-1
Complete Guide - ISO 13485 - 2016 QMS For Medical Device-1
A long time ago working for a company where I was about to transition from leading R&D
projects to moving into head of Quality, I was told by a wise colleague to be just aware that
a lot of companies consider having a Quality department and a quality management system
as a necessary evil, and a must have just to meet regulations.
Over time I have seen many companies to different degrees, believe and act as if this were
indeed a fact and not the misconception that it really is. Fortunately there are ways to
change this thinking of management, and I can share some best practices on how this can
be achieved in other articles that I have available.
The point of me bringing this up at the beginning of this article is to encourage you, as you
read through each section covering all elements of the ISO 13485:2016 Standard, is not to
just to think of the items as regulatory requirements, but how they can be beneficial to
achieving high quality of your medical device and provide the user and patient with a safe
product that meets their needs and intended use.
My recommendation is for you to think about how each element of ISO 13485:2016 and your
quality management system can help guide you on not only meeting the regulatory
requirements, but also the safety, product quality, and business needs, including cost and
time savings, customer and employee satisfaction and competitiveness.
Ok, so now let’s get into each of the 5 elements of ISO 13485:2016, with the goal to help you
understand and comply with the requirements, and so you can focus on true quality.
4 5 6 7 8
Quality Management Resource Product Measurement
management responsibility management realization analysis &
system improvement
https://fasttrackiso13485.com 1 of 37 P020.1
Your Complete Guide to ISO 13485:2016 Quality
Management System for Medical Devices
CONTENTS
https://fasttrackiso13485.com 2 of 37 P020.1
Introduction
One frequently asked question from start-up medical device companies, and even
some established business, is what countries require the ISO 13485:2016
No simple answer to that question, but generally. compliance with ISO 13485:2016
is required by countries in Europe and Asia, UK, Canada, Australia, and many
other countries. ISO 13485 is accepted as the basis for CE marking in the EU.
Medical devices marketed in the United States, however, must meet the
requirements of the FDA’s Quality System Regulation (QSR). This also will be
changing in the not too distant future as the FDA is working towards harmonizing
its medical device quality system regulation, 21 CFR Part 820, to ISO 13485:2016
ISO 13485:2016 is a globally reconized regulatory and quality standard for medical
device companies, and is based on ISO 9001 with additions for companies
involved in any part of the life cycle of medical devices. It adopts the process
model concepts of Plan, Do, Check, Act, and is designed for medical device
companies of all sizes and all product classifications.
The ISO Certificate helps to assure customers that the company has a good robust
quality management system and helps to provide them with the confidence in your
medical device quality.
https://fasttrackiso13485.com 3 of 37 P020.1
Effective Quality Management System
Key processes in your quality management system need to include; quality goals
and regular monitoring of those goals, a feedback process, document controls,
design and development process, internal audits and a corrective action process,
along with many other elements.
ISO 13485:2016 has all of this covered within the 5 elements, and with total
involvement from senior management on down, it can be very effective in helping
to provide quality products and producing a continues improvement culture.
Continuous monitoring, measuring and analysis within the QMS, in order to drive
improvements in product and processes, and to ensure conformity and
effectiveness of the quality management system is a key element of the ISO
13485:2016 Standard.
https://fasttrackiso13485.com 4 of 37 P020.1
General Requirements
Element 4.1 General requirements, sets the tone and intent, for the
ISO 13485:2016 medical devices quality management system standard, and
states that the company shall document a quality management system, and
maintain its effectiveness in accordance with the requirements of the Standard and
applicable regulatory requirements.
These are all part of the 5 elements of the standard. Another requirement is for the
organization to document the role(s) undertaken by the organization.
Documentation Requirements
Most people are not big fans of documentation, and it’s critical that all involved in
following the processes covered in the QMS procedures, and ISO 13485:2016
Standard, and those responsible for completing the records, understand fully the
importance and value of these requirements.
https://fasttrackiso13485.com 5 of 37 P020.1
In today’s high tech world it may well be worth considering using one of the many
eQMS systems that are available. This is becoming a very popular option even for
start-up medical device companies.
Just need to keep in mind that any software eQMS applications need to be
validated prior to use.
https://fasttrackiso13485.com 6 of 37 P020.1
Quality Manual
Reading the Quality Manual is a good way to start off the training for new
employees that will be involved in the quality management system. Also for any
external audit, the auditor will be asking right up front for a copy of the manual.
The Quality Manual should cover a summary of each section of ISO 13485:2016
and how the company meets that requirement, and be written in a general policy
format.
The ISO 13485:2016 Standard includes the following requirements for the
Quality Manual
• Description of the scope of your QMS, including details of and justification for
any exclusion or non-application.
For more on a Quality Manual for ISO 13485, download our eBook by
clicking on the link below:
Writing a Quality Manual that meets ISO 13485:2016
https://fasttrackiso13485.com 7 of 37 P020.1
Medical Device File
For every medical device type or device family there needs to be Medical Device
File established, and maintained that includes, or referencing documents, that
contain the following:
• Description of the product, its use, purpose, labeling and instructions for use.
• Specifications of the product.
• Specs and procedures for manufacturing, storage, handling and distribution.
• Procedures for measuring and monitoring
• As appropriate procedures for installation and servicing
Control of Documents
Having a good documentation procedure which defines your controls for all quality
management system documents and records is essential and needs to define the
following:
• Ensure documents remain legible, identifiable and do not deteriorate over time
• Control of all required external documents that support your QMS
Best practice for QMS documentation is to establish and follow a thorough and
functional document management system for your company, that meets your
requirements and the ISO 13485:2016 Standard, as well as applicable regulatory
requirements, but without being burdensome.
For a start-up medical device company this can be a challenge if they don’t have
the resources and/or experience to develop such a quality management system.
https://fasttrackiso13485.com 8 of 37 P020.1
This is where an experienced consultant like Fast-Track QMS Consultants can be
a major benefit.
Also worth considering for both a start-up and established medical device
company, is using an eQMS which can manage your documentation and allow you
to better focus on developing and manufacturing high quality products.
Fast-Track QMS Consultants can work with you on this and help in the selection
and implementation of your eQMS. See our Partner page on our website.
Control of Records
Records require very similar controls, as those for documents, and provide the
evidence that specific procedures were followed and in some cases the results of
carrying out the procedure, i.e., inspection and testing procedures and records.
One additional requirement for QMS records is defining the required record
retention period. This needs to be document and controlled and using an eQMS
system can be a very efficient way of controlling your QMS records.
Management Responsibility
A good and effective quality culture has to be supported and promoted from the
top down, and needs to have clear, communicated and ongoing commitment from
executive management.
Fortunately I have experienced more companies where the full support and
commitment for the quality management system and product quality was present
and visible to all employees. This was a significant enabler, in making these
companies successful, and resulting in a reputation for high quality products.
https://fasttrackiso13485.com 9 of 37 P020.1
Management Commitment
Two very important parts to management commitment, and I would suggest the
first, and critical component is the true vision and understanding of just how
important the top-down commitment of quality is to a medical device company, and
second, is being able to show evidence of this commitment as required by the ISO
Standard.
Customer focus
Top management, often referred to as executive management, needs to set the
example for all employees on just how important it is for the company and for
customers to have an effective quality management system. A big part of this is for
management to ensure that customer requirements and applicable regulatory
requirements are determined and met.
The final customer for a medical device is the patient, and this needs to be the
focus all through the QMS from design and development of the medical device,
through manufacturing, distribution, feedback on the device, post market
surveillance, and all aspects of the quality management system.
https://fasttrackiso13485.com 10 of 37 P020.1
Quality Policy
Having a Quality Policy is both a requirement of ISO 13485:2016 and a good way
to communicate management’s commitment to quality, and to maintain the
effectiveness of the quality management system. It needs to be applicable to the
purpose of the organization. The policy should be designed not just by the quality
leader but include in its drafting the members of the senior management team.
It may not be practical for all employees to be able to recite the policy word for
word, but it needs to be communicated throughout the organization and everyone
should know where its posted in their area and be able to point to it if asked.
I have found one best practice is to include the Quality Policy in the Quality Manual
and to include as one of the inputs for your Management Review meeting, at least
once a year to review the Quality Policy for continued suitability.
QUALITY POLICY
Company xyzzy will consistently provide products that meet or exceed the
requirements and expectations of our customers. We will meet all
appropriate international quality standards, and regulatory requirements.
We will actively pursue continuous quality improvement and programs,
that enable each employee to perform their job right the first time, every
time.
Top management has the responsibility at the highest level, to ensure the quality
management system is fully implemented and is effective. This includes appointing
a member of management to be the Management Representative, and who has
the responsibility and authority to ensure the QMS is fully documented, and for
reporting to top management on the effectiveness of the QMS, and areas that
need improvement.
https://fasttrackiso13485.com 11 of 37 P020.1
Another important responsibility for top management, and the Management
Representative, is communication throughout the organization of the required
regulatory requirements and the effectiveness of the QMS.
Here are some of the most effective ways and best practices I have seen for
communicating the above, just to provide you with some ideas:
Management Review
All too often, and perhaps more than just about any other quality system
requirement, these meetings are considered by management as a must do, check
the box activity.
I would encourage any organization that fits that description, to find ways of
making these management reviews constructive, informative, and where required,
action oriented.
https://fasttrackiso13485.com 12 of 37 P020.1
My experience has been that there should not be any big surprises from
management reviews, if, the company is doing a good job of communicating and
discussing ongoing quality updates on a regular basis.
However conducting an overall review of all of the required inputs can still be
beneficial at least annually.
Some would suggest it should be every quarter, and for myself having experience
with companies that conducted reviews quarterly, twice a year and annually, I
would recommend, that for an established mature quality management system,
that once or twice a year is likely ok. and for start-up medical device companies
once a quarter might be fine.
Executive management who are on board with the companies quality culture will
embrace reviews of the overall quality management system, including review of
the quality policy, quality objectives and results and overall status and
effectiveness of the QMS.
https://fasttrackiso13485.com 13 of 37 P020.1
Management Review Outputs should include
Resource Management
All of the above resources are very important and play a major part in effectively
managing your medical device business, and producing a high quality, conforming
product that you and your customers expect and need.
This includes the need for the company to determine and provide the resources
required to implement and maintain the effectiveness of the quality management
system, including meeting applicable regulatory and customer requirements.
https://fasttrackiso13485.com 14 of 37 P020.1
For human resources that will be performing work that can affect product quality,
the focus rightly so, is on hiring competent personnel based on appropriate
education, training, skills and experience.
Another key point is the need for methods to check the effectiveness of training
based on the level of risk associated with the work being performed.
Best practice I can recommend, is to start with a baseline on training needs and
document in a training matrix, covering all employee job functions, including
management positions, and listing the specific training required for those positions.
Details on education, experience etc. can be part of the Job Description for each
position, and training plans for key positions should also be documented, as well
as having an annual training plan for the company.
I know of a medical device manufacturing plant that was closed down for weeks
due to contamination. Fungi was detected on wooden pallets on a raw material
incoming shipment, not immediately being removed from the facility, which then
led to potential contamination of the whole facility.
https://fasttrackiso13485.com 15 of 37 P020.1
Product Realization- Planning
This is a major element of ISO 13485:2016 and describes how your company
designs and develops, manufactures and delivers the product.
Customer-Related Processes
So I think everyone would agree, that understanding what the customer needs
from your product is one key to success. Defining and documenting these
requirements is fundamental to your quality management system. Failing to do this
well increases the risks to your business and to the quality of the product.
You could have a great medical device design but if does not meet the customers
needs then you’re not going to be successful. Let’s look at the customer related
processes as defined in element 7.2
https://fasttrackiso13485.com 16 of 37 P020.1
Design and Development
https://fasttrackiso13485.com 17 of 37 P020.1
Risk management and design reviews are also key parts of the design and
development methodologies and I cover those in more detail below.
Also worth noting is that the design and development process as defined in the
ISO 13485:2016 Standard is very similar to that with the FDA 21 CFR Part 820.30
requirements for design controls
So your medical device design and development project proposal has executive
management approval and now it’s time to start the formal design project, and one
following each of the ISO 13485:2016 7.3 requirements.
The best practice that I have seen frequently, and have been part of, is the use of
a cross functional team for Design and Development projects. It should not be a
solo effort on the part of R&D.
The full-time team members will most often be only from R&D with other
department representatives getting involved as required. The size of the team and
represented areas will depend on the complexity of the device design and may
also require that outside suppliers be involved.
All team members need to be trained in the procedures and requirements for
Design and Development.
https://fasttrackiso13485.com 18 of 37 P020.1
Design and Development Inputs
Design Inputs come from the User Requirements, and define performance criteria,
functional requirements, safety and usability, as well as features of device design.
It is best practice to consider how you will verify or validated each of these design
inputs at the time you are establishing those inputs. Its referred to sometimes as
“simultaneous engineering” with can provide benefits including multidisciplinary
collaboration and a fast-track design process.
As your Design and Development project progresses you will start to develop your
Design Outputs. These are the specifications for the device and provide the
information for purchasing, manufacturing and any installation and service
provision. It can be in the form of engineering drawings, specification sheets etc.
Each of the design Outputs need to correspond and meet the documented and
approved Design Inputs. Design Outputs should also include:
Design Outputs also need to be in a form suitable for verification against the
Design Inputs and need to be documented and approved before release.
https://fasttrackiso13485.com 19 of 37 P020.1
Design and Development Review
They are also a way to ensure and document the design and development process
is on track and meeting planned expectations.
Design Reviews are conducted throughout the project as shown in the diagram
below.
Design
User
Needs Review
Design
Input
Design
Process
Design
Verification Output
Medical
Device
Validation
FDA Design Controls Waterfall Diagram
https://fasttrackiso13485.com 20 of 37 P020.1
Design and Development Verification
First let’s clarify any potential confusion between the meaning of verification versus
validation. According to the FDA, they define them as follows:
So as mentioned earlier, when you’re developing the Design Inputs you should
have a good idea and plan how to conduct Design Verification. The basic goal is to
prove by inspection and testing that the Design Outputs meet the Design Inputs.
Plan your Design Verification and document that plan before you begin the
inspection and testing process.
Document the results and indicate pass or failure of requirements along with any
required follow up actions.
Design Validation is to confirm that the medical device you have designed,
consistently meets the design specifications, user needs and intended uses. This
means you should be involving the potential customers and end users to provide
you documented feedback on the design, and if it meets their needs.
Validation testing can be performed under actual or simulated conditions and must
be performed using product that was manufactured using final production
equivalent documentation, materials and processes. This also means that the
manufacturing, testing and inspection process need to have been completed, and
as appropriate, all of the required Process Validations.
https://fasttrackiso13485.com 21 of 37 P020.1
It’s always best practice to have established early on in the project, a Master
Validation Plan, that would include all the required Process Validations and Design
Validations. This MVP needs to include for the Design Validation activities, how
many end users are required for the validation, what type of testing is to be
performed, acceptance criteria, etc.
Design and development transfer is not a single event that occurs once a design is
complete and validated, but should start early in the design and development
process.
As shown in the diagram below it is more of a simultaneous activity and can start
as soon as parts of the design and manufacturing process become finalized.
For any company this approach makes sense, and for a start-up medical device
company this is critical for the fast-track approach. Any time a partial design
transfer takes place, a design review should be held first, and the final transfer also
preceded by a design review and all documented.
https://fasttrackiso13485.com 22 of 37 P020.1
Design and Development Changes
As a normal part of your design and development process there will be changes
you make to your medical device design during development. It is necessary to
control these changes during the design and development pre-market phase.
These changes need to be controlled, validated, reviewed, and documented in
your DHF.
Once the design is finalized and released through the Design Transfer process,
any new changes being made will need to follow a documented procedure
for control of design changes.
That means you need to have processes and procedures in place to ensure:
A risk assessment should be carried out for all but the simplest of changes and
completed before the design change process begins. The formal risk assessment
can help guide the design as well as its verification and validation plans.
A review of design changes along with the risk assessment should include
evaluation of the effect of the change on the product, including product in process,
and product already distributed.
For more on Design and Development, you can download our eBook by
going to our website: Click here
https://fasttrackiso13485.com 23 of 37 P020.1
Design and Development Files
All activities related to the design and development project are to be maintained in
these files and kept current with any design changes.
The Design and Development File, or Design History File (DHF), which it is
commonly referred to, will have been updated continually as the team works its
way through the product design and development.
Purchasing
https://fasttrackiso13485.com 24 of 37 P020.1
Unfortunately there are some medical device companies, including start-ups as
well as mature companies, that don’t give supplier approval and management the
priority it needs.
This can cause lengthy delays in product production and distribution, as the
supplier works on corrections to rejected material, and even more serious, product
recalls due to failure of supplier material on distributed product.
This is another area of the quality management system where risk needs to be
considered, and the selection of suppliers and criteria used should be based on
the risks associated with the medical device and the material being purchased.
The diagram below outlines the basic steps that should be followed for supplier
selection and management.
https://fasttrackiso13485.com 25 of 37 P020.1
Production and Service Provision
This element, for control of production, and as appropriate for service provision, is
to ensure that the manufacturing of your medical device is planned, carried out,
monitored, and controlled, all to ensure that the product conforms to the required
specifications.
The level of production controls should be based on risk and communicated and
controlled through various documentation including, procedures, work instructions,
process monitoring, including SPC, specifications and inspection and test
procedures.
Each product and production process should be well defined and documented,
including acceptance criteria. Not until the release criteria has been met should the
work in-process be released to the next process, or for final product, released for
distribution.
Cleanliness of Product
For any type of product you would expect it to be clean, but of course for a medical
device this can be especially critical and depending on the type of medical device
will require certain levels of cleanliness. This needs to be well defined in
specifications and the manufacturing cleaning processes defined and documented.
https://fasttrackiso13485.com 26 of 37 P020.1
Installation and Service Provision
If your medical device is of the type that requires installation and/or servicing, you
need to have documented procedures covering these activities. These procedures
will be part of your medical device file.
If your medical device requires servicing, then you need to define servicing
requirements, specifications, and procedures for doing so; these items are part of
the medical device file.
Servicing also requires review of records for activities carried out by the company
or by an approved service supplier to see if any of the information should be
considered as a complaint and/or to be used as an input to the improvement
process.
Sterilization:
For medical devices that are sterilized and/or use sterile barrier packaging, the
sterilization process must be validated prior to implementation in production.
Records of each production lot or batch after sterilization and including sterilization
process parameters shall be documented and traceable to each production lot or
batch and be part of the products manufacturing records.
https://fasttrackiso13485.com 27 of 37 P020.1
Validation of Processes for Production and Service Provision
ISO 13485:2016 7.5.6 states the following requirement for process validation:
The organization shall validate any processes for production and service provision
where the resulting output cannot be or is not verified by subsequent monitoring or
measurement and as a consequence, deficiencies become apparent only after the
product is in use or the service has been delivered.
Very often it can be beneficial due to risk, overall efficiency and costs to carry out
process validation even though process verification is carried out. The decision
tree below is a guide to the decision process.
https://fasttrackiso13485.com 28 of 37 P020.1
In most cases, other than the simplest validations, a cross-functional team should
be formed to plan and oversee the validation activities. Just a very general
guideline, in my experience the magic number for team efficiency is trying to not
go over 5 to 7 team members. Its ok to bring in other area experts occasionally as
required, e.g. regulatory, microbiologist, marketing, etc.
Members of the validation team will depend on the process being validated but can
include personnel with expertise in, and representing:
• Quality Assurance
• Engineering
• Manufacturing
• Research & Development
Team members need to be qualified in the areas they represent, and who know
the product and process very well. They also need to be aware of defects and
errors that may be encountered as part of their area of responsibility and specific
to the product being validated.
The team leader needs to be trained in the process validation procedure, and other
team members be familiar with the overall validation process. Each of the
validation team members will be listed in the protocols with the areas represented.
Process validation depending on the process being validated will follow the steps
shown in the below diagram, and are Installation Qualification (IQ), Operational
Qualification (OQ) and Performance Qualification (PQ)
https://fasttrackiso13485.com 29 of 37 P020.1
Identification
The reason for clear identification is to help avoid mix-ups and to provide indication
of the product status and its inspection and/or testing status.
Today many global markets require a unique device identification (UDI) and is
required to be part of the device labeling. For more on this click on the link below
for guidance on requirements for UDI.
Traceability
Traceability requirements depend on the type of medical device and not all require
it. Implantable medical devices is one example where procedures are required to
meet regulatory requirements for traceability.
Records need to be maintained for product after distribution and include details of
the name and address of when and where the product was shipped. If there is a
need to act due to some adverse event this traceability information is critical.
https://fasttrackiso13485.com 30 of 37 P020.1
Customer Property
If you have any customer property in your manufacturing site you need to have
established processes in place to control, identify, verify, protect, and safeguard
this property. Records need to be maintained for these activities if applicable.
Preservation of Product
This requirement is for documented procedures for preserving and protecting the
conformity of the product during processing, storage, handling and distribution.
You may not have a separate procedure for this, but rather include these
requirements in your other documented procedures specific to your manufacturing
process procedures, environmental control procedures, packaging design, etc.
All of the measuring equipment you use, throughout your manufacturing processes
to determine acceptance of work in process and final product release, must be
controlled, calibrated, or verified to standards.
For some applications, i.e. use of software, may need to be validated before use.
Any external calibration service needs to go through a service supplier approval
process.
https://fasttrackiso13485.com 31 of 37 P020.1
Measurement, Analysis and Improvement
Feedback also can be gathered from production and post-production activities and
all sources of feedback can be potential input into risk management for elements
of the quality management system.
The diagram below shows the basic required steps in the complaint handling
process.
https://fasttrackiso13485.com 32 of 37 P020.1
Internal Audit
I have always regarded internal audits as a good opportunity to not only monitor
the compliance to your procedures, as well as to regulatory requirements, but also
to check the effectiveness of the quality management system, and to find
opportunities for making corrections and improvements.
The diagram below outlines the basic steps in planning, conducting the audit and
follow up actions:
Internal audits should also be planned based on risk. If parts of your quality system
are lower risk, and past audits and ongoing performance has met requirements,
then these areas may need auditing less frequently. On the other hand, for higher
risk areas, or areas where there have been procedure changes or issues, these
should be audited more often.
https://fasttrackiso13485.com 33 of 37 P020.1
Control of Nonconforming Product
The procedure needs to clearly define the responsibilities, authorities and controls,
for the identification, documentation, segregation, evaluation and disposition of
nonconforming product.
A good question to always ask is, if the nonconformity has occurred with the
product being only slightly outside of the specification and is still acceptable under
deviation, then is the spec tighter than it needs to be?
Rework procedures are also a requirement, and depending on the type of product,
could also justify having a separate procedure for sorting the conforming from the
nonconforming, in a batch or production lot of product.
https://fasttrackiso13485.com 34 of 37 P020.1
Analysis of Data
Procedures are required for the analysis of appropriate data to demonstrate the
suitability, adequacy and effectiveness of the quality management system. I have
always believed this should be a routine process for a start-up medical device
company as well as a mature business.
The use of data analysis and statistical tools and trend analysis were appropriate,
can be very beneficial not only for meeting the requirements of the ISO 13485 and
regulatory requirements but also for improving the product quality, by generating
corrective and preventive actions, improving customer satisfaction and reducing
costs.
https://fasttrackiso13485.com 35 of 37 P020.1
Corrective and Preventive Action
I encourage any medical device company, to give the CAPA process the attention
and priority it deserves, to resolve customer complaints and feedback, internal
audits, nonconforming product ,and any other significant issues with your QMS.
The diagram below outlines the CAPA process that should be used as far as the
inputs, procedure steps and resulting outputs:
Corrective action and preventive action are two very distinctive steps with
corrective action being reactive and preventive action being proactive. The ISO
13485 Standard does differentiate between the two as follows:
https://fasttrackiso13485.com 36 of 37 P020.1
Final Thoughts
So hopefully reading this article has answered some of your questions on the
requirements of the ISO 13485:2016 Standard, plus some best practices that will
help provide a quality management system that delivers more than just the
minimum of meeting the regulatory requirements.
A robust well managed QMS supported by top management can provide a real
business advantage, by resulting in high quality medical devices that are safe and
meet the requirements of the customer.
A QMS is not just to get certification and check the box, but is also to facilitate a
more efficient medical device company, and one that can include a focus on the
quality of your product, and to understand your customers, and what they need for
the patients.
If you’re a start-up medical device company, I know from experience there are
many pressing issues that take up your time and resources, and implementing a
quality management system is just one of those. If that is the case, then you might
want to consider using outside experienced consulting help.
That’s where we can work with you and assist with your ISO 13485:2016 gap
analysis and implementation planning. We can also provide you with the
documentation templates we have available, along with the consulting services we
offer, we can save you considerable time and overall cost by getting your quality
system implementation fast-tracked and certified, so you can get your products to
market faster, and start earning a return on your investment.
https://fasttrackiso13485.com 37 of 37 P020.1