CP-5CE-ID Guide

Do not leave the instrument connected to the SPECIFICATIONS


GENERAL
system under test when not in use.
Working Voltage: 600Vrms
The CP-5CE-ID is a split core CT meant for
operation with the Megger Power Quality Line of Always use extreme caution when connecting
the instrument around bare conductors, under Current
Instruments. This current clamp will measure AC 0.1 to 10A
current up to 5Aac. This CT is a passive CT that fault conditions, high voltage or currents may Range
does not require a power source. be present and may pose a shock hazard. Scale 200mV / Amp

SAFETY and SYMBOLS Personal protective equipment (PPE) must be Bandwidth: 40 to 3KHz
used during the installation and removal of this
Accuracy: ≤ 1% of reading +/- 5mV
WEEE instrument from hazardous live connectors.
The crossed out wheeled bin Phase Shift 1 to 5A - 1°
Do not touch circuit connections or any metal
placed on Megger products is a that is exposed due to damaged insulation. 5 to 10A – 1.5°
reminder not to dispose of the Limit
product at the end of its life with Do not use the instrument or connect it to any
Operating 20A permanently to 1KHz
general waste. Megger is external system if it shows any visible signs of
Conditions
registered in the UK as a Producer damage, malfunction or if it has been stored in
of Electrical and Electronic unfavorable conditions. Power Source NA
Equipment. The Registration No is Always inspect the instrument prior to use. Jaw Opening 0.78” ID (19.8mm)
WEE/DJ2235XR.
Replace any defective parts or return the 5.47x2.0x1.18”
Equipment complies with current Dimensions
instrument to an authorized center for repair. (139x51x30mm)
EU directives.
Do not use the instrument or connect it to any Weight 6.5oz (180g)
Application around or removal
from hazardous live conductors is external system if the enclosure is open or any Operating
-10 – 55C
permitted. parts of the enclosure are missing. Temperature
Equipment protected throughout Only use specified batteries as described by Storage
-40 – 70C
by double insulation. this document, if applicable. Temperature
The instrument shall not be used if any parts Humidity 10-35C 85% NC

G
CAUTION is defined as a condition
or practice which could result in are damaged. IEC61010-2-032
damage to or destruction of the EN50081-1 Class B
Always connect the probe to the instrument
equipment or apparatus under EN50082-2
prior to connecting to the source.
test. Electrostatic Discharge IEC 61000-4-2
This instrument is not intrinsically safe and Radiated Field IEC61000-4-3

F
WARNING is defined as a
must not be used in hazardous atmospheres. Fast Transient IEC61000-4-4
condition or practice which could
Magnetic Field at 50/60Hz IEC61000-4-8
result in personal injury or loss of If this equipment is used in the manner not 600V CAT III, Pollution Category 2
life. specified by the manufacturer, the protection *Reference conditions: 20°C to 26°C, 20 to
Safety warnings are precautions that must be provided by the equipment may be impaired. 75% RH, external magnetic field <40 A/m,
read and understood before the instrument is 48 to 65 Hz sine wave, distortion factor less
used. They must be observed during use. than 1%, no DC component, no external
current carrying conductor, test sample
centered. Load impedance > 1MΩ.
84741 Rev 2 July 2015
CP-5CE-ID Guide
OPERATING INSTRUCTIONS MAINTENANCE

G
Installation Have maintenance performed
only by qualified service

G
Personal protective equipment personnel.
(PPE) must be used during the
installation and removal of this Battery Replacement
instrument.
NA
1. Connect the CT to Megger PQ device.
Cleaning and Decontamination

To ensure optimum performance, it is important to


keep the probe jaw mating surfaces clean at all
times. Failure to do so may result in error in
NOTE: If using this CT with a PA-9 series readings.
power quality analyzer the PA-9
current full scale value MUST be To clean the probe jaws, use very fine sand paper
set to 5. (fine 600) to avoid scratching the jaw, then gently
clean with a soft oiled cloth.
2. Use the lever to open the jaw and place
the CT around the source to be measured.

Removal

1. Use the lever to open the jaw and remove


the CT from the source being measured. M
2621 Van Buren Ave
2. Disconnect the CT from PQ Analyzer Norristown, PA 19403-2329
610-676-8500

www.megger.com

84741 Rev 2 July 2015