Aroma Service Manual

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

EPILATE DIODE LASER SYSTEM

WARNNING
DO NOT OPEN THIS CONSOLE EXCEPT FOR THE AUTHORIIZED PERSON BY THE
MANUFACTURER. A/S COULD NOT BE PROVIDED BY THE MANUFACTURER.
( IF CAN NOT BE OPERATING PROPERLY AND CAUSED A PROBLEM & A FAULTY OF
THE EQUIPMENT .)
There is no DI WATER currently. Please use after add DI WATER.
When you combine body and handpiece, the both S/N number should be same.
radiator

water tank

water pump

water filter
smps RSP1500-12

smps 100-24

smps 100-12

noise filter

LCD
TOUCH PANEL

Control BOARD

Drive BOARD

Handpiece
The method of adding DI WATER and testing method

Take off board of back ( Square board connected by four volts)


.

If you see inner part, there is yellow line at WATER TANK. Add DI WATER up to yellow line.

Add DI WATER using enclosed funnel and fitting. The fitting is for removing air so you must
set fitting.

Close the square board after add DI WATER up to tallow line.

After fill DI WATER and before connect handpiece, connect enclosed test line and then
operate AROMA.

The one of important point is to check the circulation of DI WATER.


1. Parts for testing Hand piece

8pin connecting part

water line
If Aroma is delivered to your company, Add DI water in.
Don’t connect Handpiece.
Connect above two parts to the Aroma body.
After checking, if Flow switch is pressing sensor itself, connect handpiece.

Combined condition
When you combine body and handpiece, the both S/N number should be same.

Same as above photos, the S/N of sticker on back of main board should be same as S/N of
handpiece connection terminal.
Check list
1)

- Check the Hz of power cable


- Exactly combine 220v power cable to above terminal and then connect to power
2)

If the Circular red button, emergency switch (at top left ) is pressed, the machine is not
working. Move the switch to clock direction, make usable condition and then use

handpiece body
The hand piece and back of main body should be joined firmly. You have to Check the sound
and make sure that both are joined properly. ( Check the ‘Click’ sound over red arrowed
parts)
4)Right side of upper part of body, move key switch to clock direction (90 degree), the
machine will be worked.