M110 - 0007 - Jqa - M4X0.7-6H GP NP - Unc 23 - Thread Plug Gauge.

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Calibration Report of JQA M4X0.

7-6H GP NP Thread Plug Gauge

MPC Control #: M110_0007 Serial Number: M110_0007

Asset ID: N/A Calibration Date: August 18, 2023

Section 1 - Simple Pitch Diameter


Measured Values
Status of Nominal Lower Upper Uncertainty
Measuring (mm)
Thread Plug Size Limit Limit Result (±)
Position
Gauge (mm) (mm) x-axis y-axis (mm) (mm)

1/5 T 3.5519 3.5519


GP 3.545 1/2 T 3.5470 3.5520 3.5520 3.5550 PASS 0.0013
4/5 T 3.5500 3.5500
1/5 T 3.6665 3.6665
NP 3.663 1/2 T 3.6630 3.6660 3.6660 3.6710 PASS 0.0013
4/5 T 3.6672 3.6672

Section 2 - Major Diameter


Measured Values
Status of Nominal Lower Upper Uncertainty
Measuring (mm)
Thread Plug Size Limit Limit Result (±)
Position
Gauge (mm) (mm) x-axis y-axis (mm) (mm)

1/5 T 4.0052 4.0052


GP 4.000 1/2 T 3.9980 4.0052 4.0052 4.0140 PASS 0.00032
4/5 T 4.0066 4.0066
1/5 T 3.8066 3.8066
NP 3.807 1/2 T 3.7990 3.8062 3.8062 3.8150 PASS 0.00032
4/5 T 3.8073 3.8073

4/5T 4/5T
1/2T 1/2T
1/5T 1/5T
y-axis
(90 degree)
(0 degree)
Diameter

Major
Pitch

Diameter
Diameter

x-axis

Nominal Size, x-axis


Pitch

GO NOGO

Measuring Position of Thread Plug Gauge

Confidential: Controlled Document. Page 1 of 2


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev05
Calibration Report of JQA M4X0.7-6H GP NP Thread Plug Gauge

MPC Control #: M110_0007 Serial Number: M110_0007

Asset ID: N/A Calibration Date: August 18, 2023

Statements of Pass or Fail Conformance


The uncertainty of measurement has been taken into account when determining compliance with specification.
All measurements and test results guard banded to ensure the probability of false-accept does not exceed 2% in compliance with ANSI/NCSL Z540.3-2006.
The status of compliance with the acceptance criteria is reported as:
PASS - Compliant with specification.
Z
PASS - The measured value is within acceptance limits.
However, a portion of the expanded uncertainty of measurement at 95% exceeds the specified tolerance.
Z
FAIL - The measured value is not within the acceptance limits.
However, a portion of the expanded uncertainty of measurement at 95% is within the specified tolerance.
FAIL - Not compliant with specification.

The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds
to a coverage probability of approximately 95% , unless otherwise stated.
This calibration report complies with ISO/IEC 17025:2017 and ANSI/NCSL Z540.3 Method 6-Guard Bands based on Test Uncertainty Ratio.

- End of Calibration Report -

Confidential: Controlled Document. Page 2 of 2


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev05