Professional Documents
Culture Documents
Transdcer Replacement Procedure
Transdcer Replacement Procedure
Fig.2
Fig.1
“S”mark(left)
“O”mark (right)
1
4. Make sure that Shaft (5) does not lower from the highest position.
5. Detach Flange (4) with Shaft(s) from Gate valve, removing four M16×60
Socket head bolts (12), M16 spring washers (11), and M16 nuts (10).
Fig.4
6. Remove four M4×16 Socket head screws (15) and M4 shielding washers (16).
7. Remove Gland nut (19), Flat washers (8), and Gasket (7).
Flange (4)
Surface “A”
2
2. Fixing Transducer onto Shaft
Caution: For safe handling of adhesive, workspace must be well ventilated.
Nutdriver
Hose clamp
Hose clamp
Fig.6 Transducer with hose clamp Fig.7 Transducer with hose clamp
removed
2. Run the transducer cable through Shaft (5) with Flange (4).
3. Put Washer (8), Gasket (7), and Gland (19) through the transducer cable as
shown in Fig. 8.
FORE mark
Fig.8
3
5. Apply a thin coat of the adhesive, ThreeBond #1104, over the top of the
transducer. See Fig.10. The transducer top face is not visible after coating.
Safe Handling ThreeBond
1) Use only with adequate ventilation, or the workspace is well ventilated.
2) Skin Damage! Avoid contact with the skin. In case of contact, clean
immediately, wiping off using commonly available organic solvents.
6. Apply the adhesive, ThreeBond #1104 over all threads of Socket head bolt (15).
7. Fix the transducer onto Shaft (5) so that Bow mark on the transducer is
aligned with V-mark on the shaft at the angle of about 10°
, using four M4×16
Socket head screws (15) and M4 Shielding washers (16).
V-mark
Fig.12
Bow mark
Fig.11
This gap is
normal.
8. Tighten Gland (19). See Fig.8.
Fig.13
4
3. Fixing Flange Assembly to Gate Valve
2. Make sure that Flange (4), Gasket (18), and Gate valve (20) are aligned as
shown in Fig.14.
3. Using four M16×60 Socket head bolts (12), M16 Spring washers (11), and M16
Nuts (10), fix Flange assembly onto Gate valve (20).
Both FORE marks on Flange (4) and Hull flange (2) are aligned.
⑬
Not aligned.
Aligned
(4)
(18)
(20)
Fig.14
See note.
Fig.15
4. Open Gate valve (20). The distance between the handle and Shaft (5) is about
350 mm. Fig.14
350 mm Fig.16
5
5. Make sure that Gate valve indicates “0”. Fig.3
7. Make sure that V-mark on Shaft (5) is about 10 degrees CCW from FORE
mark on Hull flange (2). Fig.17
“FORE” mark
Shaft (5)
V-mark
8. Rotate the shaft several times in both directions slightly so that Transducer (1) is
in position on Hull flange (2).
9. Fix Lock ring (6) onto Shaft (5) by snug-tightening M8×30 socket head bolt (21).
Fig.15
10. Fix Lock ring (6) onto Flange (4) with the shaft, using two M8×30 socket head
bolts (13) and spring washer (14). Use non-tapped holes in Lock ring (6).
Non-tapped hole
Non-tapped hole