REPAIR MANUAL
3D OCT-1
(Type:Maestro2)
TOPCON CORPORATION
Y.Kojima
1.00 2019/01/12 Creation of new document
K.Tsubokura
Y.Kojima
1.01 2019/01/22 The changing of the authorizer
H.Okada
Y.Kojima
1.02 2019/0/ Revision of the indicated matter with the design review
H.Okada
1. No part of this manual may be copied or reprinted, in whole or in part, without prior written permission.
2. The contents of this manual are subject to change without prior notice and without legal obligation.
3. The contents of this manual are correct to the best of our knowledge. Please inform us of any ambiguous
or erroneous descriptions, missing information, etc.
4. Original Instructions
This manual was originally written in English.
Page
Confidential, duplication and reproduction prohibited
㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
R-OCT1-1-1901-04
5.UPDATE OF SOFTWARE ................................................................................................. 5-1
5.1 Update of software ................................................................................................ 5-1
5.2 How to back up with fundus calibration data and anterior calibration data ............. 5-4
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04 㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
6.1.39 Removing and assembling the TRC/OCT unit................................................. 6-21
6.1.40 Removing and assembling the LIGHT SOURCE UNIT ................................... 6-21
6.1.41 Removing and assembling the Y-axis detection sensor................................... 6-22
6.1.42 Removing and assembling the Y-axis drive unit .............................................. 6-22
6.1.43 Removing and assembling the XY-axis drive unit............................................ 6-23
6.1.44 Removing and assembling the X-axis sensor ................................................. 6-23
6.1.45 Removing and assembling EXTERNAL I / F PCBs ......................................... 6-24
6.1.46 Removing and assembling 3P SW CABLE UNIT ............................................ 6-24
6.2 Inspection and Adjustment Procedures ............................................................... 6-25
6.2.1 Adjustments and Cautions when replacing each unit ...................................... 6-25
6.2.2 Adjustment of the SLD light intensity at objective lens .................................... 6-27
6.2.3 Adjustment of the light intensity sensor at objective lens................................. 6-30
6.2.4 Confirmation of PD current value .................................................................... 6-35
6.2.5 Replacement / Check of Xenon PCB .............................................................. 6-40
6.2.6 Adjustment of the observation light intensity ................................................... 6-46
6.2.7 Adjustment of blink detection .......................................................................... 6-49
6.2.8 Adjustment of the photography observation camera ....................................... 6-58
6.2.9 Adjustment of white balance ........................................................................... 6-73
6.2.10 Stereoscopic camera orientation..................................................................... 6-79
6.2.11 Adjustment of TRC Focus............................................................................... 6-92
6.2.12 Installing the split unit/adjustment ................................................................. 6-100
6.2.13 Adjustment of the split bar focus position.......................................................6-111
6.2.14 Split analysis frame adjustment/Luminance adjustment ................................ 6-120
6.2.15 Adjustment of the center ghost ..................................................................... 6-127
6.2.16 TRC_M2-PCB Adjustment ............................................................................ 6-138
6.2.17 Adjustment of OCT-PCB ............................................................................... 6-143
6.2.18 Adjustment when replacing GPU_BOARD.................................................... 6-151
6.2.19 Pupil position adjustment .............................................................................. 6-160
6.2.20 Galvano adjustment...................................................................................... 6-164
6.2.21 Adjustment of the 0D focus for focusing lens ................................................ 6-172
6.2.22 Adjustment of OCT focusing ± 15D focus ..................................................... 6-176
6.2.23 Check of optical axis deviation for focusing lens ........................................... 6-179
6.2.24 Check of image unstableness for focusing lens ............................................ 6-183
6.2.25 Scan length adjustment ................................................................................ 6-186
6.2.26 Adjustment of the internal fixation LCD position............................................ 6-197
6.2.27 Adjustment of the aperture selector unit position .......................................... 6-203
6.2.28 Adjustment of Spectrometer and Line CCD................................................... 6-211
6.2.29 CCD shift ...................................................................................................... 6-223
6.2.30 Rescaling...................................................................................................... 6-233
6.2.31 Adjusting the optical path length ................................................................... 6-256
6.2.32 Confirmation and adjustment of interference sensitivity ................................ 6-264
6.2.33 Replacement of fiber unit .............................................................................. 6-272
6.2.34 Anterior OCT Calibration............................................................................... 6-277
6.2.35 Replacing the OCT IMG Board ..................................................................... 6-324
Confidential, duplication and reproduction prohibited
㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
R-OCT1-1-1901-04
7.TROUBLESHOOTING....................................................................................................... 7-1
7.1 You may not get high-sensitive OCT image.Proceed㸽[xxxxxx]............................. 7-1
7.2 SLD light source error(xx) Please turn the power switch OFF................................ 7-3
7.3 Trouble in the parts related with B-scan image ...................................................... 7-4
7.4 Trouble in the parts related with Control Panel(LCD Touch panel)......................... 7-5
7.5 Trouble in the parts related with Camera ............................................................... 7-7
7.6 Troubles in the parts related with Galvano mirror .................................................. 7-8
7.7 Troubles in the parts related with Fundus image ................................................... 7-9
7.8 Troubles in the parts related with boot up ............................................................ 7-10
7.9 Troubles in the parts related with polarization controller ...................................... 7-10
7.10 Xe light source error㸦xx㸧. Please turn the power switch OFF........................... 7-11
7.11 Others ................................................................................................................. 7-11
11.Appendix....................................................................................................................... 11-1
11.1 Cleaning of the objective lens.............................................................................. 11-1
11.2 Galvano mirror cleaning ...................................................................................... 11-3
11.3 OCT optical cleaning ........................................................................................... 11-5
11.4 Pasting dustproof tape · sponge.......................................................................... 11-9
11.5 Cable wiring method.......................................................................................... 11-13
11.6 Items and Functions of Graph tool Software...................................................... 11-22
11.7 Connecting CCD-Camera.................................................................................. 11-25
11.7.1 SONY XC-EI50 ............................................................................................. 11-25
11.7.2 Basler(USB3.0 type) ..................................................................................... 11-27
11.8 GENERAL INFORMATION ON USAGE AND MAINTENANCE ......................... 11-33
11.8.1 INTENDED PATIENT POPULATION ............................................................ 11-33
11.8.2 INTENDED USER PROFILE ........................................................................ 11-33
11.8.3 ENVIRONMENTAL CONDITIONS FOR USE ............................................... 11-33
11.8.4 STORAGE, USAGE PERIOD ....................................................................... 11-33
11.8.5 ELECTRIC RATING...................................................................................... 11-34
11.8.6 SYSTEM CLASSIFICATION ......................................................................... 11-34
11.8.7 DIMENSIONS AND WEIGHT........................................................................ 11-34
11.9 Electromagnetic compatibility ............................................................................ 11-35
11.9.1 Requirements for the EXTERNAL DEVICE................................................... 11-39
11.9.2 PATIENT’S ENVIRONMENT......................................................................... 11-40
11.9.3 SPECIFICATIONS OF THE PERSONAL COMPUTER
(COMMERCIALPRODUCT) TO BE CONNECTED....................................... 11-41
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04 㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
11.9.4 SPECIFICATIONS OF THE INSULATION TRANSFORMER
(COMMERCIALPRODUCT) TO BE CONNECTED....................................... 11-41
11.10 3D OCT-1 SOFTWARE LICENSE TERMS........................................................ 11-42
11.10.1 PLEASE READ THE FOLLOWING “LICENSE TERMS” CAREFULLY BEFORE
USING OR INSTALLING THIS SOFTWARE!................................................ 11-42
11.10.2 LICENSE TERMS........................................................................................ 11-42
11.11 Installation of Clean Booth during Repair of Medical Devices ............................ 11-45
11.12 Service News History ........................................................................................ 11-48
Confidential, duplication and reproduction prohibited
㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
R-OCT1-1-1901-04
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04 㻾㻙㻰㻾㻵㻻㻯㼀㻝㻙㻜㻙㻝㻜㻝㻜㻙㻜㻝㻌
1-1
3D OPTICAL COHERENCE TOMOGRAPHY 䇵㻌3D OCT-1(Type:Maestro2) – Repair Manual㻌
1. OVERVIEW
PRODUCT CHARACTERISTICS
The product is a special imaging device that observes, photographs, and records fundus images,
anterior segment images, and fundus tomographic images, and provides them as electronic images for
diagnostic purposes.
By attaching the attachment for the anterior segment, which is an optional accessory, to the product,
tomographic images of the anterior segment can be observed, photographed, and recorded. (The anterior
segment attachment is optional.)
Images of the fundus and anterior segment are taken with a built-in digital camera.
The photographed fundus, anterior segment, fundus, and anterior segment images can be recorded on a
personal computer connected to the product (hereinafter referred to as the "PC"). The control panel
displays the observation image and the shot image. Auto alignment function and manual alignment
function are available.
In some product, fundus and anterior segment images cannot be taken (hereinafter referred to as
"fundus images cannot be taken").
Some product images cannot be taken except fundus, anterior segment, and fundus tomography (3D
wide scan only) (hereinafter referred to as "screening images").
SPECIFICATIONS BY ITEMS
Item Specifications
Observation & photographing of the fundus
• Type of photography Color (Note 3), Red-free (Note 1) & IR (Note 4)
• Picture angle for 45° ±5% or less
observation/photogr 30° or equivalent (digital zoom) (Note 3)
aphy
• Operating distance 34.8 ±0.1mm (when taking a picture of fundus)
• Normal pupil diameter : I4.0mm or more
Observable/photograp Small pupil diameter : I3.3mm or more
hable diameter of pupil
• Fundus image Center : 60 lines/mm or more
resolution(on fundus) Middle (r/2) : 40 lines/mm or more
Middle (r) : 25 lines/mm or more
IR photography : Center: 5 lines/mm or more (Note 4)
Observation & photographing of the fundus tomogram
• Scan range Horizontal direction 3 – 12mm ±5% or less
(on fundus) Vertical direction 3 – 9mm ±5% or less
• Scan pattern (Note 5) 3D scan (horizontal/vertical)
Linear scan (Line-scan/Cross-scan/Radial-scan)
• Scan speed 50,000 A-Scans per second
• Lateral resolution ȝP
• In-depth resolution ȝP
• Photographable I2.5mm or more
diameter of pupil
Observation & photographing of the fundus image/fundus tomogram
• Fixation target Internal fixation target:
Dot matrix type organic EL
The display position can be changed and adjusted. The displaying
method can be changed.
Peripheral fixation target:
This is displayed according to the internal fixation target displayed
position.
External fixation target
(Note 1) Digital red-free photography that processes a color image and displays it in pseudo-red-free condition
(Note 2) When the attachment for anterior segment is included in the system configuration
(Note 3) Concerning this item, the instrument without the fundus photography function is excluded.
(Note 4) This is used only for recording the position where a tomogram is captured.
(Note 5) In the instrument for screening, it is possible to use only "3D Wide Scan (scan size: Vertical 9mm ×
horizontal 12mm)".
(Note 6) Excludes the instrument for screening.
SAFETY PRECAUTIONS
DISPLAYS AND SYMBOLS FOR SAFE USE
The main body of the equipment and the instruction manual contain important information in order to
prevent harm to users and others and damage to property from occurring and to ensure safe and
correct use.
Be sure to understand the following labels and symbols before reading "General information on
safety" and observing the instructions.
Users should not rely solely on images made using the device in making decisions regarding diagnosis
or other therapeutic procedures, but should rely on their own expertise and judgment.
DISCLAIMERS
• TOPCON shall not take any responsibility for damage due to fire, earthquakes, actions by third
persons and other accidents, or damage due to negligence and misuse by the user and any
use under unusual conditions.
• TOPCON shall not take any responsibility for damage derived from inability to properly use this
instrument, such as loss of business profit and suspension of business.
• TOPCON shall not take any responsibility for damage caused from using this instrument in a
manner other than that described in this Instruction Manual.
• Diagnoses made shall be the responsibility of pertaining doctors and TOPCON shall not take
any responsibility for the results of such diagnoses.
PREPARATIONS
EQUIPMENT INSTALLATION
x The instrument should be moved by two people holding the bottom of the device. Be sure
to hold the bottom with two people.
To avoid injury, be aware of projections at the bottom.
Carrying by one person may cause harm to his/her back or injury by
falling parts. Also, holding areas other than the bottom and holding the
External I/O terminal cover may cause injury, as well as damage to the
instrument.
x To prevent damage and injuries, do not install the instrument on an
uneven, unsteady or sloped surface.
x When setting an instrument on an instrument table, pay attention not
to catch the patient's fingers between the instrument and the table.
(1) Hold the specified position of the instrument firmly with both hands and place it on the optical table.
(1) Make sure the POWER switch of the instrument is turned OFF.
(2) Connect the power cord to the Power inlet.
(3) Insert the power cord plug into the 3-pin AC grounding receptacle.
Data output
The product can be connected to a personal computer or the like via local area networks.
(1) Connect the cable (LAN cable) to the output terminal of the product.
The time to start the power save status can be changed in the System setting
menu "Power Save Timer (Min)".
You can change the duration of the transition to the power save mode by "Setting up function" or
"Setting up system" in "POWER SAVE TIMER (minutes)".
PRIOR CONFIRMATION
Ɣ%HVXUHWRUHDGWKHLQVWUXFWLRQVLQ6HFWLRQSULRUWRXVLQJWKHSURGXFW
Ɣ,I\RXZDQWWRUHSDLUILUVWUHDGIURP&KDSWHU5HSDLU3UDFWLFHVRU&KDSWHU7URXEOHVKRRWLQJ
Working procedure Applicable item Page described
Trouble shooting
ng Chapter 7
Initial inspection Failures
Normal
Inspection
Abnormality
Repair execution
Abnormality
Inspection
In
Normal
Ending
Chapter 1 Overview
Learn product characteristics and safety precautions.
2. PRODUCT OUTLINE
2.1 Nomenclature
䐧㻌
䐦㻌
䐯㻌
䐥㻌
䐮㻌
䐤
䞤㻌
䐰
䐭㻌
䐣㻌
䞢㻌
䐫㻌 䐬㻌
VIEW A
䐢㻌
䞣㻌
䐡㻌
䐪㻌
䐨㻌
䐲㻌
䞡㻌
䐱㻌
䐠㻌
䐩㻌
䐟㻌
2.3
CAN
2.3.1
AC input TRC focusing motor
㻌 Base X motor TRC focusing position sensor
㻌 Base X position sensor Diopter compensation motor
R-OCT1-1-1901-04
Power supply unit 㻌 Base Y motor Diopter compensation position sensor
࣭Inlet +24V BASE_Sub_CPU-PCB
࣭Power switch Base Y position sensor Aperture selector motor
࣭SWpower supply Base Z motor A
Aperture selector
l sensor
Base Z position sensor Xenon lamp Split IN/OUT motor
Trigger/Release Xenon power supply PCB Chin-rest motor
Chi S li IN/OUT sensor
Split
Chin-rest controll PCB
Chi C External fixation Split motor
Fundus observation/ lampp LED Split position sensor
TRC_M2
AC INPUT
to 3P inlet +24V
to BASE_SUB_CPU_PCB
CN201
[Function]
x Supplies power to POWER RELAY PCB/TRC PCB/OCT POWER PCB.
to SW POWERSUPPLY
+24V
CN201
to XENON POWER
PCB to CHIN REST
CN226 CN225 PCB
P
CN206
[Function]
x Drive of X, Y, Z axis motor and zero position detection
x Drives the chin-rest.
x Power supply to GPU_BOARD, OCT-PCB, XENON PCB
[Check pin]㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 [Switch]
SW1: Reset switch
Pin Voltage value Pin Voltage value
TP1 +5V(4.86-5.22V) TP 3 (or Z 10) +12V(11.60-12.36V)
Z122 +5V(4.86-5.22V) TP4(or Z9) +12V(11.63-12.55V)
TP2 +3.3V(3.20-3.43V) TP5(or Z5) +8.5V(8.33-8.88V)
Z123 +3.3V(3.20-3.43V) Z6, Z8, Z11, DG
Z150, Z157, etc.
[Function]
x Power supply to OCT-PCB and image calculation
x Image calculation of line CCD camera
x LAN communication
J11(Back
( side))
to GPU_BOARD CN5700
[Check pin]㻌
Pin Voltage value Pin Voltage value Pin Voltage value
TP1 +3.3V±0.16V Z15 +1.8V±0.05V Z20 +1.8V±0.05V
TP2 DG Z16 +1.8V±0.05V Z21 +1.8V±0.05V
TP3 +1.8V±0.05V Z17 +1.8V±0.05V Z22 +1.8V±0.05V
TP4 +1.8V±0.05V Z18 +1.8V±0.05V
Z12 +3.3V±0.16V Z19 +1.8V±0.05V
J121
J120
to OCT IMG BOARD CN1
[Function]
x USB communication
x LAN communication
[Function]
Lighting/control of each light source
Anterior segment Anterior segment SPLIT LED PCB IR / TRC correction
fixation LED observation LED switching sensor
TRC focusing sensor Diopter compensation Aperture selector Split IN/OUT sensor
sensor sensor
Split drive sensor Observation LED Fixation LED Cover detection sensor
LCD reverse sensor Blink detection sensor
㻌
Supplies power to each PCB.
COLOR CAMERA
ANTERIOR CAMERA
CPU FAN UNIT
LCD UNIT
[Check pin] 㻌
Pin Voltage value Pin Voltage value Pin Voltage value
TP1 DGND TP4 +24V±1.2V TP8 +12V±0.6V
TP2 +12VA±0.6V TP5 DGND TP9 +3.3V±0.3V
TP3 AGND TP7 -12VA±0.6V TP11 Blink Sensor
Observation light level 3 (closed eyelid state)
0.6V䡚0.8V
J240 J210
J212 J245 J281 J287
to GALVANO_UNIT
_ J3 to Y-AXIS_MOTOR
_
to OCT_IMG-BOARD
_ CN7 to Y-AXIS_SEN_UNIT
to BASE_SUB_CPU_PCB CN203 to OCT_IMG-BOARD CN6
[Function]
Supplies power to each PCB.㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 Each unit control
Optical path length correction motor Galvano UNIT
OCT focusing motor Y axis sensor unit
Polarization control motor SLD MODULE PCB
Y axis motor SLD PWR CHK PCB
㻌 OCT_IMG_BOARD䠄CIBT4䠅
㻌 Line CCD CAMERA
[Check pin]
Pin Voltage value
TP1 DGND
TP3 AGND
TP11 SLD_ANODE
[Check pad]
Pad Contents Pad Contents
CP+24 +12±1.2V CP+5.2VA +5.2±0.25V
CP-12VA -12±1.2V CP+15V +15.0±1.5V(Power supply for galvano driver)
CP+12V +12±1.2V CP-15V -15.0±1.5V(Power supply for galvano driver䠅
CP+12VA +12±1.2V CP+15Vccd +15.0±1.5V(Line CCD voltage)
CP+3.3V +3.3±0.15V CPVREFA +3.0±0.05V
CP+3.3Vcibt +3.4±0.20V CPVref_DAC 5.0±0.10V
CP+1.2V +1.2±0.10V CPVref_OS +2.0V(Galvano offset voltage)
㻌
[Switch]㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 [LED]
SW Contents 㻌 LED Contents
SW1 Reset switch 㻌 LED1 +24V
SW2 Unused (All off) 㻌 LED5 SLD_ANODE(TP11)
㻌
to XENON unit
[Function]
x Supplies the XENON lighting power.
[Volume]
VR1: Adjustment of the XENON light intensity
J292
to the polarization control motor
J293
to OCT CTRL PCB J280
[Function]
x Relays the polarization control motor drive.
J292
to the diopter compensation motor
J293
to TRC PCB J260
[Function]
x Relays the diopter compensation motor drive.
[Function]
x Detects the split IN/OUT.
x Relays the drive motor drive.
to DC_MOTOR_UNIT1
J228
to TRC_M2-PCB J262
J227
[Function]
x Detects the aperture “0”.
x Relays the selector motor drive.
to DC_MOTOR_UNIT1
JJ290
J291
to TRC_M2-PCB J262
to DC_MOTOR_UNIT1
JJ290
J291
to TRC_M2-PCB J260
[Function]
x Detects the fundus focusing lens position.
x Split drive position detection.
to DC_MOTOR_UNIT1
JJ290
J291
to TRC_M2-PCB J262
[Function]
x Detects the OCT unit focusing lens position.
to DC_MOTOR_UNIT1
J
J290
J291
to OCT-PCB J280
[Function]
x IR / TRC correction switching detection.
3. OPTICAL PRINCIPLE
The patient's eye is illuminated by near infrared light, which is emitted by the fundus illumination
optical system (IR LED and xenon lamp (*1)). The fundus observation/photography optical system
forms an image on the image pick-up element (fundus observation camera), and the image can be
observed on the control panel.
The system works as follows: when operating the buttons which are displayed on the control panel
by the software, the photography light, which is emitted by the fundus illumination optical system,
illuminates the patient's eye. The image, which is received by the image pick-up element (fundus
photography camera) built in the fundus observation/photography optical system, is recorded as an
electronic fundus image (*2). You can also set "digital zoom" that extracts and saves the center of
the photographed image (equivalent to angle of view of 30°) through software. (*1)
The anterior segment/fundus tomogram is obtained by using the optical interference
phenomenon.
Near infrared light, which is emitted by the super luminescence diode (SLD), is separated by the
fiber coupler. One portion of the separated light is guided to the prism and is reflected by it, before
returning to the fiber coupler. The other portion of light is guided into the eye and is reflected by the
cornea/fundus tissues, before returning to the fiber coupler. When the two reflected portions of lights
join and overlap, a low interference wave with differing amplitude is generated. This wave is
separated by the diffraction grating and is then converted into the electric signal by the line CCD.
This signal is then processed to observe, photograph and record the anterior segment/fundus
tomogram.
Using the sensor built in the auto alignment optical system (anterior segment observation LED and
camera), the auto alignment function detects the positions of the pupil and instrument.
Then, the instrument is moved to a position applicable to photography.
There are two auto focus functions, "Split auto focus function" and "OCT auto focus function". The
user should select and use one of them. The split auto focus function detects the reflected image of
the split lines projected to fundus. Then, the lens in the observation/photography optical system is
moved to a proper position by the auto focus mechanism to adjust focus correctly. In the case of the
OCT auto focus function, the lens in the observation/photography optical system is moved to a
proper position by the auto focus mechanism in order to get the optimum image quality level
(coefficient calculated by S/N ratio) of the observed fundus tomogram. So focus is adjusted
correctly.
By continuously photographing, recording and processing the fundus/anterior segment
tomogram, analysis such as three dimensional display, fundus image display or quantitative
measurement is conducted.
4. SETTING AT SHIPMENT
4.1 Menu setting at shipment from factory
Photography Setting
Utility button Item button Item selector button At shipment
Fixation TYPE CONSTANT
• Changes lighting/blinking of the internal (Lighting)
fixation target. FLICKERING {
(Blinking)
SPEED NORMAL (Hz) 1-8 1
• This is valid when the internal fixation target
is set to "FLICKERING" (Blinking).
• Set the blinking speed for all other states
except when photographing.
SPEED MEASURE (Hz) 1-8 2
• This is valid when the internal fixation target
is set to "FLICKERING" (Blinking).
• Set the blinking speed for photographing.
Brightness 1-10 10
• Sets the brightness of the internal fixation
target.
Shape (Default) 1 dot {
• Sets the shape of the internal fixation target.
X
Shape (Additional) Not in use {
4 dot rectangle
Cross
Screen Display Constant {
Given time
Fundus photo Blink Detection Level OFF
*3
LOW {
NOMAL
OCT+Color Mode High ISO {
㺃Changes the color fundus image quality Normal ISO
(High/Normal/Low).
Low ISO
Flash Level (OCT+Color Mode) 1.0Ws to 16.0Ws *1
Gain (OCT+Color Mode) 0 to 9,BS,AC*2 *1
Color Mode High ISO {
Normal ISO
Low ISO
Flash Level (Color Mode) 1.0Ws to 16.0Ws *1
Gain (Color Mode) 0 to 9,BS,AC*2 *1
Adjust Flash at Peripheral Photo 0 to +4 +1
*1: For the flash level and Gain of each image quality at shipment, refer to the table below.
Flash level Gain
Image
Setting at Setting at
quality Settable range Settable range
shipment shipment
High 1.0㹼16.0Ws 4.0Ws 0㹼63 0
Normal 1.0㹼16.0Ws 2.0Ws 0㹼63 8
Low 1.0㹼16.0Ws 1.0Ws 0㹼63 21
Auto Operation
Utility button Item button Item selector button At shipment
Auto Alignment ON/OFF OFF
ON {
ACTION Discontinuous
Continuous {
Timeout (Sec) OFF,10,20,30 20
Manual Mode two cameras {
one camera
Focus ON/OFF OFF
ON {
one time {
sequence
Split+OCT
OCT: Retry OFF
ON {
Small Pupil *1 ON/OFF OFF {
• This function is usable only when the
color fun- dus capture icon is selected. ON
• When the pupil diameter of the patient is
Magnification Mode 45°Mask {
small (pupil diameter is approx. ij4.0mm
or less), the small pupil diaphragm is 30°Mask
automatically set and the instrument takes
a color picture of the fun- dus. After
taking a picture, the small pupil dia-
phragm is automatically removed. When
you set the small pupil diaphragm selector
button to "ON" on Area 1 of the
photography screen (color photography),
the above automatic setting is not done.
Auto Optimize ON/OFF OFF
ON {
Auto Z OFF
•Detects the retina posi- ON {
tion automatically, and
always displays the fun-
dus tomogram clearly
during observation.
Z-lock After Auto Z OFF
•After "Auto Z", the ON {
photographed fundus
image is fixed.
Auto Pola OFF
•Adjusts the output ON {
sensitivity of the fundus
tomo- gram to the
optimum automatically.
Auto Operation Auto Shoot Timer (sec) 1.3.5.8.10.15 3
OU Capture Right eye ĺLeft eye {
Left eye ĺRight eye
OFF
Show Preview OFF
ON {
Preview method *2 Both eyes
Each eye (with timer)
Each eye {
Preview Timer (sec) *3 3, 5, 10 5
Check Alignment OFF {
Result ON
Check Alignment 3, 5, 10 5
Timeout *4
AutoShoot OFF
ON {
Auto Cataract ON/OFF OFF
ON {
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04
4-4
3D OPTICAL COHERENCE TOMOGRAPHY 䇵㻌3D OCT-1(Type:Maestro2) – Repair Manual㻌
*1: In the instrument without the fundus photography function, this is not displayed.
*2: When "ON" is set for "Preview", it is valid to change setting for this item.
*3: When "ON" is set for "Preview" and "Each eye (with timer)" is set for "Preview method", it is valid to change
setting for this item.
*4: When "ON" is set for "Alignment check", setting for this item is valid.
System setting
Utility button Item button Item selector button At shipment
System Setting Measurement Sound OFF {
Sets ON/OFF for the measurement sound. ON
Power Save Timer (Min) 0 to 60 10
Sets the time until the power save function is ON.
Display Brightness 1 to 10,send 5
Sets the brightness of control panel.
Packing Mode Start
Tap the [Start] button, and the head unit and
chinrest move to be in the packaged status.
Language Japanese(JP)
Set the display language English(US) {
4.2 How to start the Tool Mode and “CAN SEND TOOL”
4.2.1 Tool Mode
(1) Start the main unit.
(2) When the main unit is started, long-press the [SETUP] at the top right of the screen.
(3) The password input screen is displayed, so enter the password and press the [OK].
5. UPDATE OF SOFTWARE
5.1 Update of software
Purpose
Update for customer’s request or after the release of latest version.
Necessary Tools
Tool Name Tool No. Image
Version up file - -
USB memory
*Please do not use USB memory with security function
Procedure
(1) Preparation
(a) Store the software for UPDATE into the USB memory.
*File configuration
x MAESTRO2_****.zip (**** means version.)
x MAESTRO2_Update
x OCT_UpdateAppPro
x OCT_UpdatePro
(b) Turn on the instrument.
Connect the tool USB memory to the USB port of the instrument.
(c) The software update check screen appears. Tap the [OK] button.
Update of software starts.
(d) Tap the [OK] button when the dialog box is displayed.
Update of software starts.
(f) Tap the [Please turn off the power] button to turn OFF the power.
5.2 How to back up with fundus calibration data and anterior calibration
data
Purpose
Back up for Fundus calibration data and Anterior Calibration data
Necessary Tool
PC -
Procedure
(1) Enabling export of anterior segment CLB file
x If the procedure described in (1) is not performed and the procedure described in (2) is
performed, the item "anterior segment CLB file" is not displayed on the screen described
in (2)(c).
(a) Insert the USB memory into the instrument and turn the power ON.
(b) Enter [Settings], set [Options] ĺ>$QWHULRU2&7 seg.] in [Page 7 /9] to “ON”, and tap [OK].
(2) Exporting rescaling files (fundus CLB files) and anterior segment CLB files
(a) Enter the setting menu in tool mode.
(b) Tap [Info] ĺ[Export] and tap [Export] in [Rescale File] and [Anterior Eye CLB File], respectively.
(c) Remove the USB memory, Insert the USB memory into the PC.
(d) Make sure that the "oct" folder is created immediately below the USB memory and the CLB files of the
anterior segment and fundus are exported.
ŊņRFW
Ōņ;;;;;;B'2&7-1_ANTERIOR.CLB
Ŋņ;;;;;;B'2&7-1_FU NDUS.CLB
x "XXXXXX" at the top of the CLB file is the body serial number.
6. Repair practice
Use the torque driver to tighten the screws in the disassembly and assembly according to the screw
tightening torque table below.
Since the mirror base and Y-axis drive unit of the main unit are made of resin, it is dangerous that
the screw hole will be damaged if the tightening exceeds the torque shown below.
ͤ Reference
The following table shows the torque when the conventional iron screw is attached to the other metal
part.
Screw tightening torque table (screw hole material is metal)
Nominal Tapping screw [cN.m] Subject
M2 26.5 -
M2.5 51.9 -
M3 90.2 147
M4 225.4 343
M6 754.6 1176
2 5
6
1 Assembly point
x Reverse the disassembly procedure.
x When replacing the CHIN REST PCB, measure the
resistance between No.1 and No.2 of the mounted
CHIN REST PCBVR1 with a tester, and adjust VR1
2 so that it is equal to the measured resistance before
mounting.
3
4
2 Assembly point
x Reverse the disassembly procedure.
10 9
2
(a) Remove the two cover caps ո.
8
(c) Remove the middle cover right շ.
7 Assembly point
9 x Reverse the disassembly procedure.
4 2
Assembly point
x Reverse the disassembly procedure.
1
9
(a) Remove the four screws պ.
(b) Remove the front չ of the monitor cover.
(2) Removal of the monitor cover lower part ջ
10 (a) Remove the two screws ռ.
(b) Turn the monitor up.
(c) Remove the monitor cover lower part ջ.
11 Assembly point
x Reverse the disassembly procedure.
12
2
Assembly point
x Reverse the disassembly procedure.
Conductive tape 1 x Gasket 1
Gasket1 13 mm from the center and top of the monitor
x Gasket 2
Place just above the bar code seal. The ears extend
evenly.㻌
Bar code seal x Conductive tape 1
Cut to 25mm × 35mm: 2 pieces
Gasket2 ͤ Close contact between sheet metal and LCD
The amount of tape Conductive tape 2
protrusion is 5 mm. x Conductive tape 2
Cut to 25mm × 30mm: 2 pieces
Paste from the end of the center sheet metal R.
ͤ Adhesion between sheet metal and LCD.
The amount of tape protrusion is 5 mm.
ͤ To be caught by the rotating plate.
Assembly point
3
x Reverse the disassembly procedure.
2
Assembly point
㻝 x Reverse the disassembly procedure.
2
6.1.13 Removing and assembling the CPU FAN unit, BATTERY PCB and TEMP
PCB
Schematic representation Removing method
(1) Removal of the CPU FAN unit ձ
2
(a) Remove the four screws ղ.
(b) Remove CPU FAN unit ձ.
1 (2) Removal of the BATTERY PCB ճ
(a) Be connected to the BATTREY PCB
(b) Remove the connector.
(c) Remove the two screws մ.
(d) Remove the BATTERY PCB ճ.
Assembly point
x Reverse the disassembly procedure.
6.1.16 Removing and assembling the anterior segment viewing LED PCB
Schematic representation Removing method
(1) Removal of the anterior segment viewing LED
PCB (right)ձ
(a) Remove the connected connector.
(b) Remove the two screws ղ.
1 (c) Remove the right 1 of the anterior segment
viewing LED PCB.
(2) Removal of the anterior segment observing LED
PCB left ճ
2
(a) Remove the connected connector.
(b) Remove the two screws մ.
3
6.1.18 Removing and assembling the anterior segment fixation lamp LED PCB
Schematic representation Removing method
(1) Removal of the anterior segment fixation lamp
LED PCB ձ
(a) Remove the connectors connected to the TRC
PCB J238.
(b) Remove the two screws ղ.
(c) Remove the anterior segment fixation lamp LED
PCB (right)ձ.
1
Assembly point
x Reverse the disassembly procedure.
Assembly point
x Reverse the disassembly procedure.
2
㻠
㻡㻌
6.1.23 Removing and assembling the optical path length correction unit, collimator
flange, and galvano holder
Schematic representation Removing method
(1) Removal of the Optical Path Length Correction
Unit ձ
(a) Remove the connected cable.
(b) Remove the three screws ղ.
1 (c) Remove the optical path length correction unit ձ.
(2) Removal of the collimator flange ճ
(a) Remove the two screws մ.
(b) Remove the collimator flange ճ.
2
5
(3) Removal of the Galvano Holder յ
(a) Remove the two screws մ.
(b) Remove the Galvano holder յ.
3 Assembly point
6
x Reverse the disassembly procedure.
4
6.1.24 Removing and assembling the camera for shooting and observation
Schematic representation Removing method
(1) Removal of the camera for shooting observation ձ
(a) Remove the two screws ղ.
(b) Remove the shooting observation camera ձ.
Assembly point
x Reverse the disassembly procedure.
1 Assembly point
x Reverse the disassembly procedure.
3 Assembly point
x Reverse the disassembly procedure.
6.1.28 Removing and assembling the BASE SUB CPU PCB䞉OCT PCB
Schematic representation Removing method
(1) Removal of the OCT PCB ձ
(a) Remove the connected connector.
3 (b) Remove the four screws ղ.
(c) Remove OCT PCB ձ.
(2) Removal of the BASE SUB CPU PCB ճ
(a) Remove the connected connector.
4
(b) Remove the four screws մ.
(c) Remove the BASE SUB CPU PCB ճ.
Assembly point
x Reverse the disassembly procedure.
1
Assembly point
x Reverse the disassembly procedure.
x Mount the OCT focusing unit in the upper right
direction.
x Adjust and press the end surface of the sponge ղ
Align the sponge end with the
against the upper part.
upper part. x BASE SUB CPU PCB mounting screws do not use
* Paste it so as to press it. FWSs because the screw hole land diameter is small.
Make sure that the convex
portion of the focus is covered. Fill the gap with
the motor with a
Cover the "cover sheet metal"
thin tool etc.
and the gap.
6.1.31 Removing and assembling the monitor drive unit, TRC_M2 PCB
Schematic representation Removing method
1 (1) Removal of the monitor drive unit ձ
2 (a) Remove the four screws մ.
Dowel
(b) Remove the monitor drive unit ձ.
(2) Removal of the TRC_M2 PCB ճ
(a) Remove the connected connector.
(b) Remove the four screws մ.
(c) Remove TRC_M2 PCB3ճ.
2
Assembly point
4 x Reverse the disassembly procedure.
6.1.32 Removing and assembling the lower light shielding plate metal
Schematic representation Removing method
(1) Removal of the Lower Light Shield Metal ձ
(a) Remove the two screws ղ.
(b) Remove the light-shielding plate ձ.
(c) Remove the dust-proof part ճ
3
1
Assembly point
x Reverse the disassembly procedure.
2
2 Assembly point
x Reverse the disassembly procedure.
6.1.35 Removing and assembling the split unit and internal fixation lamp
Schematic representation Removing method
(1) Removal of the Split Unit ձ
(a) Remove the three screws ղ.
(b) Remove the split unit ձ.
8 (2) Removal of the POS SEN PCB2 ն
6.1.36 Removing and assembling the diopter compensation unit and ELBOW unit
Schematic representation Removing method
(1) Removal of the diopter correction unit ձ
(a) Remove the two screws ղ.
2 1
(b) Remove the diopter correction unit ձ.
(2) Removal of the ELBOW unit ճ
(a) Remove the four screws մ.
(b) Remove the ELBOW unit ճ.
Assembly point
3
x Reverse the disassembly procedure.
4
x Insert the diopter correction unit into two convex
positions of the mirror body and tighten the screws.
Diopter compensation unit ELBOW unit x Insert the ELBOW unit into the three convex positions
of the mirror, and tighten the screws.
6.1.37 Removing and assembling the branch mirror unit and blink detection sensor
unit
Schematic representation Removing method
(1) Removal of the Branch Mirror Unit ձ
(a) Remove the two screws ղ.
(b) Remove the branch mirror unit ձ.
(2) Removal of the blink detection sensor unit ճ
1
(a) Remove the screw մ.
3
Assembly point
4
x Reverse the disassembly procedure.
1
2
Assembly point
x Reverse the disassembly procedure.
Assembly point
x Reverse the disassembly procedure.
Assembly point
*Remove the mirror body from x Reverse the disassembly procedure.
the stand and remove the
hexagonal screw on the back of
the "airplane sheet metal"
3
Assembly point
x Reverse the disassembly procedure.
3 Assembly point
x Reverse the disassembly procedure.
4
2
(1) Removal of the XZ drive unit ձ
(a) Remove the four screws ղ.
Assembly point
x Reverse the disassembly procedure.
Assembly point
x Reverse the disassembly procedure.
Assembly point
x Reverse the disassembly procedure.
Assembly point
x Reverse the disassembly procedure.
Replacement Parts
or Phenomenon
Adjustment Item
Galvano Error
(1024779-01)
Estimated repair time (minutes) 240 210 330 240 240 330 240 210 240 300 700 900 300 300 155 335
Adjustment of 6.2.17 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐥㻌 㻌 㻌 㻌 䐡㻌
OCT-PCB
Adjustment when 6.2.18
replacing 㻌 䐠㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌
GPU_BOARD
Pupil position 6.2.19 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐣㻌 㻌 䐧㻌 䐧㻌 㻌 㻌 㻌 㻌
adjustment
Galvano adjustment 6.2.20 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐥㻌 㻌 䐤㻌 䐨㻌 㻌 㻌 㻌 㻌
Adjustment of the 6.2.21
0D focus for 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐦㻌 㻌 㻌 㻌 䐣㻌
focusing lens
Adjustment of OCT 6.2.22
focusing ± 15D 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐩㻌 㻌 㻌 㻌 䐤㻌
focus
Check of optical 6.2.23
axis deviation for 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐪㻌 㻌 㻌 㻌 䐥㻌
focusing lens
Check of image 6.2.24
unstableness for 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐫㻌 㻌 㻌 㻌 䐦㻌
focusing lens
Scan length 6.2.25 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐦㻌 㻌 䐥㻌 䐲㻌 㻌 㻌 㻌 㻌
adjustment
Adjustment of the 6.2.26
internal fixation 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐦㻌 㻌 㻌 㻌 㻌 㻌
LCD position
Adjustment of the 6.2.27
aperture selector unit 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐡㻌 㻌 㻌
position
Adjustment of 6.2.28
Spectrometer and 㻌 㻌 㻌 㻌 㻌 㻌 䐡㻌 㻌 㻌 㻌 䐨㻌 䐭㻌 㻌 㻌 㻌 㻌
Line CCD
CCD shift 6.2.29 㻌 㻌 㻌 㻌 㻌 㻌 䐣㻌 㻌 㻌 㻌 䐬㻌 䞡㻌 㻌 㻌 㻌 㻌
Rescaling 6.2.30 㻌 㻌 㻌 㻌 㻌 㻌 䐢㻌 㻌 㻌 㻌 䐪㻌 䐯㻌 㻌 㻌 㻌 㻌
Adjusting the optical 6.2.31 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐤㻌 㻌 㻌 䐰㻌 㻌 㻌 㻌 䐢㻌
path length
Confirmation and 6.2.32
adjustment of
㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐩㻌 䐮㻌 㻌 㻌 㻌 䐧㻌
interference
sensitivity
Replacement of fiber 6.2.33 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐤㻌 㻌 㻌 㻌 㻌
unit
Anterior OCT 6.2.34 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐧㻌 䐡㻌 䐭㻌 䞢㻌 㻌 㻌 㻌 㻌
Calibration
Galvano mirror 11.2 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐡㻌 㻌 䐣㻌 䐣㻌 㻌 㻌 㻌 㻌
cleaning
OCT optical 11.3 㻌 㻌 㻌 㻌 㻌 㻌 㻌 㻌 䐢㻌 㻌 㻌 䐢㻌 㻌 㻌 㻌 㻌
Cleaning
x At the adjustment place where numerals are shown, perform the adjustments in the
specified order.
Necessary tools
Tool name Tool No. Image
Power meter OCT-1000-04
㻌
Objective Lens Cap for RGB 㻙㻌
Procedure
(1) Preparation
(a) Attach the Objective Lens Cap for RGB and probe of Power meter to the instrument.
(b) 6HWWKHSRZHUPHWHULQ=HURZDYHOHQJWKȜ QP
x If changing settings of SLD Pulse and Level, set [OFF] in on/off of SLD setting.
x If [POWER CONTRL] is set to “HI” without setting “LOW” , the SLD light source might
be broken. When [POWER CONTRL] is set to “OFF”, set it in the order of ”HI”
ĺ´LOW´ĺ³OFF”.
(b) Adjustment
i Adjust Pulse Width, so that the objective lens power is within 580 to 600uW
ii Tap [Save] of Pulse Width after adjustment.
Necessary tools
Tool name Tool No. Image
Maestro䊡_LightAdjustment -
LAN cable -
Prepare other necessary tools which are described in the reference section.
Procedure
(1) Preparation
(a) Connects the main body to PC using LAN cable.
(b) Turn ON the main body.
(c) Make [Transmission of the setting data] of LightAdjustment effective.
i Create the shortcut of Maestroϩ_LightAdjustment.
ii Display the property screen of the shortcut.
iii Add “ / Mode-Send” in the end of the link(Top enter a space), Click Apply㸦㐺⏝(A)㸧.
Ex.: C:\Users\user1\Desktop\3DOCT-1_LightAdjustment.exe /Mode-Send
(c) When the center position detection is completed, the coordinates of the center position and the graph
are displayed.
(d) Click [Gain Adjustment] and perform the auto adjustment to display the light intensity.
x When out of standard, it can be mounted again after confirming that there is no dust in
front of the sensor.
Standard
Item Standard Notes
Sensor width for Galvano Y 700 or more
x Click the [Show / Update Stop] to stop each of the display of "SLDLight".
(b) When the following message is displayed, restart the body by clicking the [OK].
Necessary tools
Tool name Tool No. Image
Superlum power tool OCT-2000-23
Anti-static band -
Procedure
(1) Preparation
x Use anti-static band on your hand. If the operation described in this section was done
without using anti-static band, SLD module might be damaged.
(b) Confirm the SLD rated current value and SLD serial number described on the SLD module.
Serial number㻌
(c) Turn ON Superlum power tool. Press [Select] to show “SLD Current”.
(e) Press [Select] to display “Current Limit”. Set the display value to “SLD rated current value” +1mA.
x Basically, set the current limit value to a current value of +1.0mA. If an error occurs and
the PD current cannot be measured, check the value between +2.0 mA and 5.0 mA. If the
error cannot be measured even if the error is changed, the SLD module may be damaged
or the power supply of the Superlum may be changed.
x Connect Superlum power tool after turning it OFF. If Superlum power tool is connected
without turning OFF, SLD module may be broken.
(h) Turn ON Superlum power tool. Press [Select] to display “Pd Monitor”.
(i) Press [On/Off] twice. SLD lamp will be lit in green, then the current value will be displayed in “Pd
Monitor”.
If SLD lamp is lit in red (or changed from green to red), reconnect SLD cable 2K.
(j) Compare the PD current value with the one at the shipping from maker.
Item Standard Notes
PD current value Within ±5% against the value If out of standard, replace SLD
at the shipping from maker module. (*)
*Carry out the SLD module replacement after taking the repair training ()section 6.2.33).
If you have not taken the repair training, ask TOPCON for the repair.
x Do not adjust VR2. The charge voltage (VR1) and overcharge voltage (VR2) have been
adjusted in the factory.
x Before performing this adjustment, make sure that the cable is connected to the Xe power
supply PCB. (Refer to the following figure.)
Before connecting the cable, make sure that the instrument is not turned on.
x When replace the Xenon lamp, wait more than 5minutes for the discharge of Xenon
Capacitor(Xenon POWER PCB).
Purpose
Repair for the Xe unit failure concerning the error related with Xe unit.
Necessary tools
Tool name Tool No. Image
Object holder OCT-1-01
㻌
Digital voltmeter or Digital multi meter -
Procedure
(1) Replacement of Xenon lamp(XE_FLASH_UNIT_M2)
(a) Remove the lamp house cover, the head cover, the middle cover, and the communication cover, the
base cover. ( ) section 6.1.4~6.1.6)
(b) Remove the Xenon lamp connector.
(c) Loosen two screws using the flat head screw driver and Remove the Xenon lamp after slide to left.
(d) Attach the Xenon lamp for replacement.
(c)Loosen the screw and Remove
the Xenon lamp after slide to left
heat sink
(b) Remove two connectors, four screws and two spacers on Xenon POWER PCB.( ) section 6.1.38)
(c) Remove the Xenon POWER PCB.
(d) Attach Xenon POWER PCB for replacement (Do not forget to assemble the protection cover as well).
(e) Assemble the heat sink of Line CCD .
(f) Attach the base cover, the communication cover, the middle cover, the head cover, and lamp house
cover.
1024631-01
Xenon_CBL_UNIT
_M2 1022217-01
Xenon_POW_CABLE
VR1
Push
Push
Figure D
(d) Measurement
i Select "CAN SEND TOOL" - "Lamp Ctl" - "Xe" - "Level" - "0: 4.0" and tap [Set].
ii Tap "Flash" - [Set] to make Xe emit light and confirm that the light intensity is within the standards
shown in the table below.
iii Similarly change "Level" to "-4", "-2", "+ 2", "+ 4" to cause Xe to emit light and confirm that each is
within the standard.
Standard
Item Standard Notes
-4: 2.0 A × 0.5±15% Let A be the measured value of
-2: 2.9 A × 0.7±15% “0: 4.0”
0: 4.0 A: 0.37mJ±30%
+2: 5.7 A × 1.4±15%
+4: 8.1 A × 2.0±15%
Necessary tools
Tool name Tool No. Image
Power meter OCT-1000-04
㻌
Objective lens cap for RGB -
Procedure
(1) Preparation
(a) Put the objective lens cap onto the objective lens in the main body.
(b) Turn ON the main body.
(c) Activate CAN SEND TOOL and make the following settings.
iii Attach the probe of the power meter to the objective lens cap.
(b) Confirm the measured value(Level 1,2,4) as referring to the Observation light intensity setting same as
(a).
Observation light intensity 1 2 3 4
Measured value (μW) 100±10 120±10 155±10 210±10
(c) After confirming each value are adjusted in the previous procedure, Tap the [Save].
Necessary tools
Tool name Tool No. Image
Object holder OCT-1-01
PC -
㻌
LAN cable -
Procedure
(1) Preparation
(a) Turn on the main body.
(b) Launch CAN SEND TOOL.
(c) Set as follows.
Tab Zone Item Value
Motor Ctl 1/2 Split IN / OUT - OUT
Fixation Ctl Internal/External Fixation LCD Setting Patern 0:External ĺ
Set
Lamp Ctl SLD ON/OFF OFF
Lamp Ctl IR LED - OFF
Lamp Ctl IR LED Level3 Set
Blink Ctl Blink Adjustment IR Adjustment Set *1
Blink Ctl Blink Adjustment Setting Read ĺ
“Gain 255”㻌
ĺ㻌 Save *2
(d) Looking through the objective lens, check the "retreat of the split bar", "SLD Off", "IR LED Off" and
"internal fixation Off".
(2) Adjustment
(a) Eyelid opening adjustment
i Attach the Blink detection adjustment tool (A) and the Object holder onto the objective lens.
ii Adjust "Offset" and set the value of "Blink Monitor" to be within 240 DN ± 35 DN.
Standard
Item Standard Note
Eyelid Opening 240DN±35DN Adjust by Offset Value.
Adjust the DN value to
the lowest value within
the range of the standard.
ii Adjust "Gain" and set the value of "Blink Monitor" to be 820DN -0DN/+1910DN.
x The value of "Blink Monitor" is displayed by tapping [ON].
x Adjust so that it approaches 820 DN as much as possible at 1910DN or less.
Standard
Item Standard Note
Eyelid Closure 820DN -0DN/+1910DN Adjust by Gain value
Make it as close as possible to 820
DN between 820 and 1910 DN.
v Connect the main unit and the PC with the LAN cable.
vi Startup “IMAGEnet6”.
x Tap the bottom part on “Camera 1” and “Camera 2” screen to display pupil mark.
xi Tap the [Capture Start] button.
Pupil mark㻌
xii Tap [Capture start], then close the shutter of the blink detection confirmation tool within 1 second, and
confirm that the following message is displayed.
Necessary tools
Tool Name Tool No. Image
Object holder OCT-1-01
pylon Viewer -
CameraPositionAdjust -
(pylon Viewer must be installed.)
Torque driver(18cN䞉m) -
Procedure
(1) Preparation
Install Basler Camera and Personal Computer System settings.
(a) Pylon viewer installation procedure
ii Check "I agree to the pylon Terms & Conditions" and click "Next".
vi Click [Install].
x Use torque driver (18 cN䡡m) when attaching camera link cable to OCT_IMG - BOARD.
ii Loosen the screws on both upper sides and flip the Holder Plate (OCT_IMG - BOARD) about 90
degrees.
IMG_B_POW_CABLE
Loosen the screws.㻌 There is a latch on the
There is a latch on the back back side
side of the LAN cable.㻌
iii Adjust the angle and depth of the sheet metal so that IMG_B_POW_CABLE can be connected to
OCT_IMG_BOARD.
iv Tighten the upper and lower screws and fix the sheet metal.
x If it is forcibly connected, the board / cable may be damaged or broken.
x IMG_B_POW_CABLE may touch to the edges of the metal plate depending on the
position of the sheet metal jumping up. As a result, there is a possibility of damaging the
coating of the cable, so attach the insulating tape to the contacting part.
x Find the position where the sheet metal is the largest open.
Flip up.
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04
6-63
3D OPTICAL COHERENCE TOMOGRAPHY 䇵㻌3D OCT-1(Type:Maestro2) – Repair Manual㻌
Attention to short
Disconnect the USB cable of the
Basler cameras connected to the GPU
board. Connect the USB cable to the
PC using the USB3. 0 extension cable.㻌
Objective lens
front holder
* 1. If the live image is dark, change the value and tap [Set].
(Do not tap [Save])
x Restart CameraPositionAdjust if the image becomes dark while not in use during use.
(d) Adjust mask position to center using the three screws CR3×8S+FW_3BZn.
Loosen 6 SU3x4S3BZn㻌
Focusing tool㻌
(b) Loosen the two screws 6SU3×4S3BZn for coarse camera inclination adjustment. Adjust camera
inclination and position in the green frame.
(c) After adjustment, apply the TA lock. Adjustment frame㻌
3 pcs of CR3×8S+FW_3BZn 㻭㻌
Center position
x Camera inclination: Galvano scale should
be located between reference lines for Reference lines
inclination. for inclination
x Camera position: Adjust camera position 㻰㻌
referring to the adjustment frame (green
circle).
x The distance between the mask position 㻯㻌
and the edge of the screen must be equal
the vertical and horizontal positions.
AҸB䚸CҸD
㻮㻌
Coarse camera
inclination adjustment㻌
CR3X8S+FW_3BZn
6SV3X4S 3BZn㻌 Apply screw locking
TA lock
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04
6-68
3D OPTICAL COHERENCE TOMOGRAPHY 䇵㻌3D OCT-1(Type:Maestro2) – Repair Manual
ii Click [ZOOM].
iii Confirm the error between the galvano scale and green line.
Read the X, Y error between the
galvano scale and green line
(0.05mm / division).
Error =¥;2+Y2)
ii
Confirm the error between the galvano scale and green line.
Item Standard Check Points
Difference of Within ijmm Check the difference of Galvano scale and
Observation Optical axis (within 3 divisions) Green Line in tool.
If it is out of the standard, carry out step (4) and
(5) (a) again.
(c) Remove the Galvano-OCT Optical Alignment Check Tool and terminate the CameraPositionAdjust.
(b) When the communication is established between Basler and PC, double click “Basler ac......”.
If the communication is not established, confirm the connection between Basler and PC.
(c) Confirm Basler information.
(d) Select "Guru" if the User Level is Beginner or Expert.
(e) Open “Acquisition Controls” clicking “(+)”.
(f) Set “Trigger mode” to ON. Screen will be changed to Live image.
(g) Click “Continuous Shot”.
(h) Set “FeatureSets” to “Full”.
A Standard:
|A-B|=1mm
|C-D|=2mm
C D
x "Section 6.2.8 Adjustment of the photography observation camera " has been completed
before this adjustment.
Necessary Tools
Tool Name Tool No. Image
Object holder OCT-1-01
LAN cable -
Maestroϩ_AdjustBaslerCamera -
Procedure
(1) Preparation
(a) Attach the Object holder to the WB (White Balance) compensation tool.
(b) Connect the main unit with the PC using the LAN cable.
(c) Turn on the main unit.
(g) Tap [Advance]-[Focus]-[-] in capture monitor. Then move ³ź´Iocus indicator bar mark to around -10
and adjust the focus.
x Check that the diopter correction lens switch button is set to "0".
Focus
Advance
ii Set the RGB value range used for WB adjustment. Set a point near the center of white, the leftmost one,
to within the range of the 120×120 analysis frame. It must not stick out from the white range.
1䠅㻌 Change the value of [RGB values in the specified range] to move the position of the analysis frame 㻌 㻌 㻌 㻌
(yellow).
㻌 㻌 䈜Move the analysis frame (yellow) to the vicinity of the center of the white chart.
2䠅㻌 Set the difference between the right and left and the top and bottom of the [RGB values in the
specified range] to 120.
㻌 㻌 䈜Setting the difference between the right, left, and top and bottom to 120 makes the analysis frame
(yellow) 120×120 squares in size.
*Set the position of the analysis frame so that it does not protrude from the white chart.
iii Click [Set range]
120
120
x Flash level is set 4.0Ws (default) which is high quality of the photograpy light intensity
basically. In case the RGB adjustment value is not standard value, change the flash level to
make the adjustment value is in the standard value.
Standard
Item Standard Notes
RGB adjustment value R: 192±31 -
G: 96±9
B: 207±19
Change "adjustment
value" and shoot.
x Do not change the RGB adjustment value when checking the Cyan chart.
"Cyan" chart
Analysis frame
(b) When the dialog “Is adjustment parameter set up?” is displayed, click “Yes”. Then the dialog “Write of
adjustment parameter was completed.” appears. Click “OK” and close Adjust Basler Camera.exe. Then
turn off the power of the main unit.
Necessary Tools
Tool Name Tool No. Image
Object holder OCT-1-01
㻌
Stereo matching adjustment B tool OCT-1-31
Including calcXYZ.csv in USB-memory
(Use the same serial number as OCT-1-34)
Anterior CCD filter OCT-1-33
㻌
SONY XC-EI50/2 40423 68000
LAN cable -
3DOCT-1_SendAlignmentParameter.exe -
(for the data transfer application software)
IC Capture -
Procedures
(1) Analysis of orientation㻌
(a) Preparation
x When you transfer calcXYZ.csv stored in the tool attached USB memory and the setting
file "Rism5.cmr" "TargetSheet2.gcp" etc. to the PC, prepare in advance.
i Attach “Stereo matching adjustment B tool” for the orientation work to the objective lens part of the
main unit.㻌At this time, it is necessary to detach the external board for peripheral fixation (on the circle)
and the rubber cap.㻌
Insert the pin into the upper part of the objective lens properly and fix it securely with the fixing screw
so that there will be no deviation. Be careful because the precision will deteriorate, if orientation work is
done with deviation in the tool.
ii Attach the anterior segment CCD filter to each of the left and right stereo cameras.
Attach Anterior
CCD ,Stereo camera filter㻌
x The direction of the stereo camera is defined by the right camera when viewed from the
examiner's side, and the left camera when viewed from the left side..
x If the tool for orientation work is not attached in the proper position, the measurement
accuracy will be affected.
iii Connect the main unit to the PC (with the tool software installed) using the LAN cable.
iv Execute QMcalib.exe in StereoCameraMeasuringTool (ver1.2.0 or later).
v Click the [New Project]. While the dialog of project creation is displayed in the software for orientation
analysis, make sure that the live video of the left and right cameras is displayed on the main unit screen.㻌
ii Secure with the set screw at the bottom of the lens, make sure the lens does not move and apply TA lock.
Switch the display of the image, and perform it on both the right and left lenses.
*Please be aware that if the lens and the camera move after fixing the lens, the measurement accuracy
will go wrong.
iii Click "load" of "Left image cmr" and "Right image cmr" respectively, specify "Rism5.cmr" file, and
click "Save".
iv Click "load" of "gcp", specify "TargetSheet2.gcp" file, and click "Save".Sapo9899
v Click the [OK].
vii When the analysis is completed, the result is displayed. Confirm that "Auto Detect Points" is "50/57" or
higher and "Auto Calib Result" is "OK".
x If "Auto Calib Result" is "NG" or "Auto Detect Points" is less than 50, the process starts
again from acquiring images ()step (b)).
x Even if "OK" is displayed in "Auto Calib Result", readjustment is required if "Auto Detect
Points" is not "50/57" or more ()step (b)).
ix Check that the [Camera Calibration Determination] dialog box is displayed and OK is displayed in the
[Determination] column, and click [OK].In the case of NG, start again from step (b).
* The data does not always match the following figure data.
Camera position
coordinate
X, Y, Z
Camera rotation
angle
Omega, Phi, Kappa
Left camera position parameter - Output Left camera position parameter - Output
contents contents
iv The internal fixation light at the aperture position will be observed like a circular form. Read the error of
X direction and Y direction using the scale of the check tool. Referring to the drawing below, read the
scale at the edge of the circular form of the internal fixation light. Decide the adjustment direction and
quantitiy.
v One division corresponds to 100μm. Convert the result to the unit of “μm”. Record the value for the
further step.
vi Transfer the value to the application software on the main body in step (3).
ĺWKHPDLQERG\DXWRDOLJQPHQW;<YDOXHLQ³7DUJHW3RVLWLRQ2IIVHW´㸧
-Y scale reading: 3
Conversion data (Y):
-100μm
X(+) direction
+Y scale reading: 5
e.g.) In the case of the above figure, since it is necessary to move -ȝPLQWKH;GLUHFWLRQDQG-
ȝPLQWKH<GLUHFWLRQLQRUGHUWRPDNHWKHcenter of the fixation light and the center of the
scale coincide with each other, the target position offset X is "-100 "And Target Position Offset Y
is" -100 ".
(c) Measurement with Difference of Cornea aperture and Iris aperture (TargetPositionOffset: Z)
Note the values of "Auto Alignment2"-"Target Position Offset Z" on the page 9 of tool mode.
iii As in the previous step, specify Camera position <Right > (.EXT file) and Target position (.csv) (.csv
file) to ensure that the file is displayed.
x Target position (.csv) specifies the stereo-matching-adjustment-B tool and the set of
calcXYZ.csv files that you used to correction.
iv Enter the values of X, Y measured in (2)-(b) and Z checked in (2)-(c) in "Target position Offset".
(a) Reboot up the instrument after transfer the orientation data in step (3).
Then update the sending data
(b) Launch IMAGEnet6.
(c) Enter the tool mode and make the following settings.
i Set[Macula]of fixation position in [Fundus].
ii Tap [ON] at [Auto Alignment] tab in Page 5.
iii Tap [ON] at [Auto Focus] tab in Page 5.
iv Tap [OK].
(d) Tap [Fundus Photo].
(e) Attach tool filter(OCT-1-33) in front of Anterior observation camera on left and right or attach the main
cover, front cover and filter, etc.
(f) Capture Left and Right.
No problem to movement that clear fundus image within mask as below capture images.
* If the accuracy check tool is not attached in the proper position, the measurement accuracy will be
affected.
(c) Enter the tool mode and make the following settings.
i Enter “scan for Fundus shooting” in the Fundus shooting mode, set internal fixation to “Macula”.
ii Turn on the Auto Alignment Adjust in page 9/9 Adjustment mode to enable the function.
iii Tap [OK],then close Serviceman mode.
(d) Confirm that the reading value of Z direction in micrometer is set to the standard value shown on the
tool.
(e) Tap [Fundus Photo].
(f) By tapping [Capture start], you will tap the pupil part manually to specify the screen, so tap the pupil
part of the model eye that appears in each image to the left and right, and tap [Capture start].
(g) When the pupil portion of the model is detected, the XYZ figure (value of the difference from the target
position) will be displayed at the bottom left of the IR observation screen.
(h) Confirm that the value of XYZ displayed in step (g) is within the value of X and Y: 0.3 mm and the
value of Z is within 0.1 mm.
(i) After confirmation, set Auto Alignment Adjust to OFF in page 9/9 Adjustment mode set in SetUp.
(j) If out of standard in step (11) -(i), proceed as follows.
i Remove the tool once and attach it again. Then repeat the same procedure. If still out of standard, carry
out next step ii.
ii Carry out “orientation work” again. Confirm the detected points in orientation target image. Adjust focus
position rotating the focusing lens to focus the target ,which was (were) not detected.
Necessary Tools
Tool Name Tool No. Image
Object holder OCT-1-01
㻌
pylon Viewer -
Procedure
(1) Preparation
(a) Turn on the main unit.
(b) Connect the Basler camera and the PC with a USB 3.0 extension cable.
Click “Basler ac ͐”
(e) Move “TRC Focus” to around “0” position. (About 1500 in “CAN SEND TOOL”㸧
(f) Tap the [Capture] button in “Fundus Capture” frame in “Motor Ctl” to capture the image.
(g) Confirm the image clear resolution up to chart “I”.
x If the resolution is not good, repeat steps (e) ~ (f) to find the suitable focus position.
x Focus precisely moving around “20” in “CAN SEND TOOL”.
x When confirming the focus position, please adjust the magnification on the pylon
viewer to about 300% so that the resolving power chart of the split focusing test eye can
be seen clearly.
(3) Confirmation and adjustment of the potentiometer in -10D position of split focus
(a) Attach the split focusing test eye of -10D.
Attach “-10D” in
the tip of the tool.
(b) Tap [-10D] of TRC-OCT Focus column in Motor Ctrl 1/2 tab.
(c) Move “TRC Focus” to around “-10D” position. (About 860 in “CAN SEND TOOL”)
(d) Tap the [Capture] button in “Fundus Capture” frame in “Motor Ctl 1/2” to capture the image.
(e) Confirm the image clear resolution up to chart “G”.
x If the resolution is not good, repeat steps (c) ~ (d) to find the suitable focus position.
x Focus precisely moving around “20” in “CAN SEND TOOL”.
x When confirming the focus position, please adjust the magnification on the pylon
viewer to about 300% so that the resolving power chart of the split focusing test eye can
be seen clearly.
(4) Confirmation and adjustment of the potentiometer in +10D position of split focus
(a) Attach the split focusing test eye of +10D.
Attach “+10D” in
the tip of the tool.
(b) Tap [+10D] of TRC-OCT Focus column in Motor Ctrl 1/2 tab.
(c) Move “TRC Focus” to around “+10D” position. (About 2780 in “CAN SEND TOOL”)
(d) Tap the [Capture] button in “Fundus Capture” frame in “Motor Ctl 1/2” to capture the image.
(e) Confirm the image clear resolution up to chart “I”.
x If the resolution is not good, repeat steps (b) ~ (c) to find the suitable focus position.
x Focus precisely moving around “20” in “CAN SEND TOOL”.
x When confirming the focus position, please adjust the magnification on the pylon viewer
to about 300% so that the resolving power chart of the split focusing test eye can be seen
clearly.
Necessary tools
Tool name Tool No. Image
Object holder OCT-1-01
㻌
Torque screwdriver (20 ~ 120cNm) OCT-1-35
IC Capture.exe -
Projector -
(Penlight, etc., Alternative)
Procedure
(1) Replacement of split unit
(a) Remove the split unit.
i Remove the CN7 cable of OCT_IMG - BOARD (with latch on the back side), LAN cable of CN 1 (with
latch on the back side), camera link cable, and "2 screws fixing the sheet metal".
Remove while holding the
latch on the back side
CN1
CN7
ii Loosen the two screws as shown in the figure below, rotate the sheet metal 90 degrees, and fix with the
one screw removed in (b).
iii Remove the two screws and remove the light shielding metal plate.
㻶㻞㻢㻡㻌 㻶㻞㻣㻝㻌
Confidential, duplication and reproduction prohibited
R-OCT1-1-1901-04
6-102
3D OPTICAL COHERENCE TOMOGRAPHY 䇵㻌3D OCT-1(Type:Maestro2) – Repair Manual
v Remove the six screws and remove the light shielding metal plate.
* Aligning direction
Mirror body base
(d) Attach the split 2 hole aperture tool to the Object holder.
(e) Attach CCD camera and the capture parts (camera lens is not necessary).
Object holder
x Adjust the Split LED brightness level so that the tool's scale is observed clearly.
x It is also possible to make it easier to see by illuminating the split unit with the projector.
x If the image with split 2 hole aperture image is not within the specification, carry out the
following adjustment in mirror body.
(b) Check that the left and right 2 hole aperture images are located at the same position (Target reference:
19 scales). Turn off the power to the main unit and remove the split unit if the right and left positions
are not equal, in the rotation adjustment, readjust the right and left positions so that they are at the same
position.
Connect to PC
x Adjust the brightness so that the light-shielding bar and tool scale can be observed clearly.
(d) Rotate the gear manually to move to the position where both ends of the light shielding bar are well
focused.
Carry out the procedure (e) ~ (h) by referring the standard in the diagram.
Standard
Items Standard Remarks
Misalignment of light Left and right: within 0.3mm
shielding bar
Vignetting of split image No vignetting
by light shielding bar
Inclination of light Within ± 1° (patient side)
shielding bar
Split image location Horizontal center: ±0.35 mm or Check shift of lower from upper side equal to 1/6
less of emission line width to the left observing from
Vertical center (Split line position): objective lens side.
WRPP
(e) Confirm that the misalignment of light shielding bar is within the standard.
(f) Confirm that the vignetting of split image by light shielding bar is within the standard.
(g) Confirm that the inclination of light shielding bar is within the standard.
(h) Confirm that the split image location is within the standard. (If out of standard, proceed to next step.)
Misalignment of light
shielding bar
Check the dust on the split bar.
Inclination of light
shielding bar
Verify 1/6 shift of the emission line.
i. Adjust Z direction to align ii. Re-adjust the split image iii. Adjust Z direction and
horizontal positions of upper position by eccentric shaft shift the emission line to
and lower emission lines. and eccentric bush. the left by 1/6.
Save 0D value.
Necessary Tools
Tool Name Tool No. Image
Object holder OCT-1-01
pylon Viewer -㻌
Procedure
(1) Preparation
(a) Turn on the main unit.
x Turn on the main unit and connect Baslercamera and PC.
x Caution the USBextension cable and Basler camera not to contact to PCB.
(protect the metal part of USB cable by wrapping the aircap)
(b) Connect the PC to the Basler camera through a USB 3.0 extension cable.
Monochrome image㻌
(o) Remove “0D” tool from the attached split focusing test eye.
(b) Tap "CAN SEND TOOL" - "Motor Ctl 1/2 "-" TRC - OCT Focus "-" Move "- [- 10 D].
(c) After entering split movement amount (approximate: 3400) to "Motor Ctl 1/2" - "Split Unit" - "Move",
tap [Move], Look for a position where the light shielding bar is in focus.
(Focus position: a position where the upper split is taken as a reference and the lower split is shifted by 1/6
right)
(b) Tap "CAN SEND TOOL" - "Motor Ctl 1/2 "-" TRC - OCT Focus "-" Move "- [+ 10 D].
(c) After entering split movement amount (approximate: 1000) to "Motor Ctl 1/2" - "Split Unit" - "Move",
tap [Move], Look for a position where the light shielding bar is in focus.
(Focus position: a position where the upper split is taken as a reference and the lower split is shifted by 1/6
right)
(d) Confirm that the split bright line is within 1/4 from 1/6 offset.
Item Standard
split bright line within 1/4 from 1/6 offset
㻗㻝㻞㻰㻌
Item Standard
split bright line within 1/4 from 1/6 offset
㻙㻝㻟㻰㻌
(h) Remove the diopter compensation check tool (-13/+12).
(i) Turn OFF main unit.
Necessary Tools
Tool Name Tool No. Image
Object holder OCT-1-01
LAN cable -
Procedure
(1) Preparation
(a) Connect PC and the main unit with LAN cable.
(b) Attach the Object holder and Split alignment schematic eye(0D) to the main unit.
䈜0D㻌
(h) Tap the bottom part on the alignment screen of 㺀Camera 1㺁 and 㺀Camera 2㺁 to display pupil marks.
(i) Tap [Capture START] again.
Pupil marks㻌
(j) Tap [Focus]. Align the upper and lower split operating the purple indicator bar.
Focus
Standard
Item Standard Notes
Deviation of upper and lower splits Within 1/4 of the split width See Figure B
(d) Tap [Auto Focus Measure]. Then confirm matching of the split and measurement time within 0.931
second.
(e) Confirm the split LED value(Red letters value) saved in section 6.2.17(3).
Standard
Item Standard Notes
Autofocus measurement time Within 0.931s Focus (+) side
(f) Tap [Focus]. Set ³ź´ mark to the left end of the bar operating the purple indicator bar.
(g)Tap [Auto Focus Measure]. Then confirm matching of the split and measurement time within 0.931
second.
Standard
Item Standard Notes
Autofocus measurement time Within 0.931s Focus (-) side
Attach the black spot adjustment check tool to the front side of the objective lens and carry out the
positioning of the black spot, center ghost measurement, and bright spot measurement.
This instruction describes the methods to perform the adjustments mentioned above.
The following settings and adjustment procedures are described in this section.
x Preparation
x Adjustment of black spot
x System environment setting
x Confirmation of ghost
Necessary tools
Tool name Tool No. Image
Object holder OCT-1-01
M118FM08 OCT-1-52
IC capture.exe -
NWTool.exe -
(Tool Software for RGB measurement)㻌
Procedure
(1) Preparation
(a) Remove the cover of the main unit ()section 6.1.4).
(b) Remove the CN7 cable of OCT_IMG - BOARD (with latch on the back side), LAN cable of CN 1
(with latch on the back side), camera link cable, and "2 screws fixing the sheet metal".
(c) Loosen the two screws as shown in the figure below, rotate the sheet metal 90 degrees, and fix with the
one screw removed in (b).
(d) Remove the two screws and remove the light shielding metal plate.
(f) Remove the six screws and remove the light shielding metal plate.
J291
ii Remove the xenon unit and illuminate the scale with the outside lamp.
iii Insert black spot adjustment driver (two) and hexagonal 1.5mm wrench in the black spot adjustment
screw.
Then loosen the two screws which fix the unit for the light axis in about one turn.
Use the hexagonal wrench with short tip since the work space is narrow in this operation.
x Befor removing the black spot adjustment screws, dissolve TA lock with solvent in
advance.
(b) Adjust the black spots to the center of the scale while watching the monitor with black spot adjustment
driver (two) and hexagonal 1.5mm wrench.
(c) After adjustment is completed, tighten the two screws and apply TA lock.
(d) Return the plate, attach the cable to CN 7 of OCT_IMG - BOARD, and put it in a state that can turn on
the power.
ii Put in the light shielding sheet of Black spot lens/center ghost check tool.
v Tap the bottom part on the alignment screen of 㺀Camera 1㺁 and 㺀Camera 2㺁 to display pupil marks.
vi Tap [Capture START] again.
Pupil marks㻌
viii Maximize the shooting light amount and tap [Capture START].
Xe emits light and capture is done.
㻯㼑㼚㼠㼑㼞㻌㼛㼒㻌㼓㼔㼛㼟㼠㻌
(h) Confirm the bright spot (confirmation of the defect of the objective lens).
If the figure of the “Counts over RGB200” is 0, it means that there is no bright spot whose maximum
brightness value exceeds 200.
In such a case, exchange the s an object lens.
(i) After completing the work, do "split unit installation, adjustment" ()section 6.2.12).
Necessary tools
Tool name Part No. Image
IR card OCT-1000-03
Digital voltmeter -
Manufacturer: FLUKE
Type name: 111 TRUE RMS MULTIMETER
DC voltage range: 0 to 500 V or more
Measurement accuracy:0.5 [%] or more
Procedure
(1) Pre-inspection confirmation
Check the setting of the DIP SW of the substitute TRC_M2- PCB()section 2.3.3 step (7)).
Bit Setting Function
Bit1 OFF Unused
Bit2 OFF Unused
Bit3 OFF Unused
Bit4 OFF Unused
(2) Inspection procedure
Test Items and Content List
Step
Inspection item Details of inspection
numbers
(3)- (b) Check each output voltage See section 2.3.3 (7)
࣭ TRC focus motor
࣭ Diopter correction motor
࣭ Split motor
(5)- (a) Motor operation and sensor value check
࣭ Split I/O motor (motor operation only)
࣭ Aperture changeover motor (motor operation only)
࣭ IR correction motor
࣭ Check that the anterior segment fixation LED is lit.
࣭ Check that the internal fixation lamp OLED is lit.
(5)- (b) LED lighting check ࣭ Check that the anterior segment observation LED is lit.
࣭ Check that the split LED is lit.
࣭ Check that the fundus observation LED is lit.
Check that the LCD reversal sensor value can be obtained
(5)- (c) Check LCD reversal sensor value
correctly.
(5)- (d) Checking the version Check that the version is written.
Check that error handling works when the light blocking state of
(5)- (e) Check operation of cover detection sensor
the cover detection sensor is changed.
Adjusting blink detection and checking Check that the blink detection voltage changes when the light
(5)- (f)
operation blocking state of the blink detection sensor is changed.
(5)-(g) Checking TRC focus operation Check that the focus positions of 0D and ±10D are aligned.
iii Turn on the power supply of the main unit and confirm that the voltage of each part falls within the
standard ()section 2.3.3 step (7)).
iv Turn OFF the power.
v Connect each unit of the main unit to the TRC_M2 PCB ()section 2.3.3 step (7)).
(c) Update the software()section 5.1).
x When an error message is displayed in the above operations, repeat the operations described
in (2) and thereafter.
x Split I/O motor (motor operation only) - Tap [IN] Ѝ [OUT] at "Split IN/OUT" to check operation
sound.
x Aperture selector motor(motor operation only) - Tap [Normal] Ѝ [SPupil] at "Small Pupil" to check
operation sound.
x IR correction motor (motor operation only) - Tap [IN] Ѝ [OUT] at "TRC IN/OUT" to check operation
sound.
Necessary tools
Tool name Part No. Image
Oscilloscope (hereinafter referred to as oscilloscope) -
Procedure
(1) Replacement of OCT-PCB
(a) Replace the OCT-PCB ()section 6.1).
x Check that the DIPSW of SW2 is turned off from Bit1 to 4.
(h) Confirm that the TP14 does not exceed the SLD maximum current. Repeat [/HYHO@ĺ>6HW@WRJUDGXDOO\
increase the maximum current to a level equivalent to the rated current of the SLD. Record Level.
x The SLD rated current is described in the fiber unit.
x For a SLD-rated current of 160 mA: TP14 = 1.6 V
For 200mA: TP14 = 2.0V㻌
Rated current of
(i) Increase "Level" so that the TP14 voltage becomes the rated current of "SLD+0.05 [V]".
Standard
Item Standard Remark
TP14 SLD rated current+0.05V
(j) Lower the [SLD Limit], adjust the TP14 voltage to be within the range of "(rated current of
SLD+0.02[V])±0.01[V]", and record "SLD Limit".
Standard
Item Standard Remark
TP14 (SLD rated current + 0.02V) ± 0.01V Record the SLD Limit
(k) Increase "Level" to ensure that the TP14 does not rise above "(rated current for SLDs) + 0.05V.
(l) After checking, lower "Level" to the values recorded in (i). At this time, make sure that the TP14
voltage is the original voltage (rated current of SLD).
(m) Tap [Save] in the [Level] field to Save.
x Go to step (4) with the oscillos connected.
(4) Check the light intensity control: Check/set the PD reference value.
(a) Records the TP12 values. (PD value when CW is lit)
(b) Select "Lamp Ctl"-"SLD"-"ON/OFF" from "Low" to "OFF".
(c) Set "Lamp Ctl"-"SLD"-"SLDPulse" to "1:Brinking", tap [Set], and then set "ON/OFF" to
/RZĺ+,*+
(d) Set the vertical axis of the channel connected to the TP12 of the oscilloscope to 500 [mV/div] and the
KRUL]RQWDOD[LVWR>ȝVGLY@DQGPHDVXUHWKHYROWDJHLQDFFRUGDQFHZLWKWKH73+,-level voltage in
the following measuring range.
x TP12 between 2.0 [V] and 1.0 [V]: 500 [mV/div]
x TP12 between 1.0 [V] and 0.5 [V]: 200 [mV/div]
x TP12 between 0.5 [V] and 0.0 [V]: 100 [mV/div]
i Adjust the GND level of the measurement channel of the oscillos to the second scale from the bottom.
ii As shown in the figure below, the time cursor 1 is aligned with the rise of the pulse, and the time cursor
2 iVPRYHGWRWKHSRLQW>ȝVHF@DIWHUWKHULVHRIWKHSXOVH
iii Adjust the amplitude cursor 1 to the GND level of the channel connected to the TP12. Adjust the
amplitude cursor 2 to the intersection point of the time cursor 2 and the pulse waveform. Record the
voltage value to the third decimal place. (PD value when the PW is lit)
Record the value rounded to the nearest fourth decimal place when the voltage value [V] of the oscillos
is indicated to the nearest fourth decimal place.
Amplitude cursor 22
MEASURED
VALUE [V] Amplitude
>ȝV]