Download as xls, pdf, or txt
Download as xls, pdf, or txt
You are on page 1of 68

Calculation of Measurment Uncertainty

PRESSURE GAUGE Date: 08-10-12


Range 30 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 29.8
2 29.8 Deviation +/- 0.7kg/sq.cm Max
3 29.8
4 30.0
5 29.8
6 30.0

Avarage = 29.8667

Standard Deviation = 0.1032796 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.046188 kg/cm²

0.046188 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.7200 0.3600 0.7200


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.2000 0.1155 0.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0021 0.1296 0.1633 0.0033 0.0133 0.0000 0.0000

UC = 0.3118

0.311773979

= 0.558367244 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.0972030140134 UA 4_
0.000 4 Inifinity 1.138E-06

Effective degrees of freedom = 85432

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.116734 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.12kg/cm² 0.18612241 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 100 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 80.1
2 80.4 Deviation +/- 0.7kg/sq.cm Max
3 80.4
4 80.4
5 80.4

Avarage = 80.3400

Standard Deviation = 0.1341641 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.060000 kg/cm²

0.060000 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.7200 0.3600 0.7200


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.4000 0.2309 0.4000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0036 0.1296 0.1633 0.0033 0.0533 0.0000 0.0000

UC = 0.3532

0.353242993

= 0.594342488 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.1247806117661 UA 4_
0.000 4 Inifinity 3.240E-06

Effective degrees of freedom = 38513

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.188685 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.12kg/cm² 0.19811416 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 300 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 299.8
2 299.6 Deviation +/- 0.7kg/sq.cm Max
3 299.6
4 299.8
5 299.8
6 299.6

Avarage = 299.7000

Standard Deviation = 0.1095445 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.048990 kg/cm²

0.048990 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.7200 0.3600 0.7200


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.2000 0.1155 0.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0024 0.1296 0.1633 0.0033 0.0133 0.0000 0.0000

UC = 0.3120

0.312040646

= 0.558605984 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.09736936458 UA 4_
0.000 4 Inifinity 1.440E-06

Effective degrees of freedom = 67618

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.117212 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.12kg/cm² 0.18620199 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 1000 bar

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Dead weight tester
Sr. No. Measured Reading (bar) Uncertainty = 0.11
1 700.4644
2 700.4587 Deviation +/- 0.4871 bar
3 700.4826
4 700.4832
5 700.4800
6

Avarage = 700.4738

Standard Deviation = 0.0114084 bar

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.005102 bar

0.005102 bar

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(bar) UOM (bar)) ESTIMATE(bar)

0.1100 0.0550 0.1100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(bar) UOM (bar) ESTIMATE(bar)
0.4871 0.2812 0.4871
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005bar :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (bar) UOM (bar) ESTIMATE(bar)


10.0000 5.7737 10.0000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(bar) UOM (bar) ESTIMATE(bar)

0.3000 0.1732 0.3000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.008572638 bar

(Rectangular Distribution) <∞>

L.H.Correction(bar) UOM (bar) ESTIMATE(bar)

0.0086 0.0049 0.0086

U6 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(bar) UOM (bar) ESTIMATE(bar)

0.0000 0.0000 0.0000

U7 : Standard Uncertainty Due To uncertanityin the value of g:100ppm

(Rectangular Distribution) <∞>

uncertainity of g UOM (bar) ESTIMATE(bar)

0.0001 0.000058 0.0001

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.000026 0.003025 0.079093 33.335 0.030002 0.000024 0.000000 0.00000000

UC = 33.4475

33.44745974

= 5.783377883 bar

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
1118.7325629977 UA 4_
0.000 4 Inifinity 1.694E-10

Effective degrees of freedom = ###

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 11.335421 bar


Reporting of Results:
For 700bar uncertainty is +/-0.651936bar 1.13354207 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 5 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 4.98
2 4.98 Deviation +/- 0.03kg/sq.cm Max
3 4.98
4 4.98
5 4.98
6 4.98

Avarage = 4.9800

Standard Deviation = 0 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.000000 kg/cm²

0.000000 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0900 0.0450 0.0900


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0300 0.0173 0.0300
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero Drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.0020 0.0003 0.00001 0.0000000 0.00003 0.00003

UC = 0.0024

0.002396772

= 0.048956836 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
5.744515112E-06 UA 4_
0.000 4 Inifinity 0.000E+00

Effective degrees of freedom = #DIV/0!

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 0.097914 kg/cm²


Reporting of Results:
For 5 kg/sq.cm uncertainty is +/-0.1kg/cm² 0.32637891 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 10 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 9.98
2 9.99 Deviation +/- 0.03kg/sq.cm Max
3 9.98
4 9.99
5 9.98
6 10.0

Avarage = 9.9833

Standard Deviation = 0.005164 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.002309 kg/cm²

0.002309 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0900 0.0450 0.0900


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0300 0.0173 0.0300
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0100 0.0058 0.0100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero Drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.0020 0.0003 0.00001 0.0000000 0.00003 0.00003

UC = 0.0024

0.002402105

= 0.049011276 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
5.770109123E-06 UA 4_
0.000 4 Inifinity 7.111E-12

Effective degrees of freedom = 811422

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 0.098023 kg/cm²


Reporting of Results:
For 5 kg/sq.cm uncertainty is +/-0.1kg/cm² 0.32674184 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 08-10-12
Range 30 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 80.10
2 80.40 Deviation +/- 0.03kg/sq.cm
3 80.40
4 80.40
5 80.40
6 29.98

Avarage = 71.9467

Standard Deviation = 20.559734 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 9.194593 kg/cm²

9.194593 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0900 0.0450 0.0900


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0300 0.0173 0.0300
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0100 0.0058 0.0100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0095 0.0055 0.0095

U6 : Standard Uncertainty Due To Zero Drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (Kg/sq.cm ESTIMATE(kg/sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

84.5405 0.0020 0.0003 0.00001 0.0000000 0.00003 0.00003

UC = 84.5429

84.54293011

= 9.194722949 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
7147.5070307633 UA 4_
### 4 Inifinity 1.787E+03

Effective degrees of freedom = 4

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 18.389446 kg/cm²


Reporting of Results:
For 5 kg/sq.cm uncertainty is +/-0.1kg/cm² 61.298153 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 07-11-12
Range -0.1 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.11
2 -0.12 Accuracy +/- 0.003kg/sq.cm
3 -0.11
4 -0.11
5 -0.11
6 -0.11

Avarage = -0.1117

Standard Deviation = 0.0040825 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.001826 kg/cm²

0.001826 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Vacuum gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0030 0.0015 0.0030


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Accuracy of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0030 0.0017 0.0030
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0100 0.0058 0.0100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Zero Drift

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/sq.cm ESTIMATE(kg/sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000033 0.000002 0.000003 0.00001 0.000033 0.000033

UC = 0.0001

8.35879E-05

= 0.009142642 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
6.986938632E-09 UA 4_
0.000 4 Inifinity 2.778E-12

Effective degrees of freedom = 2515

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 0.018285 kg/cm²


Reporting of Results:
For -0.1 kg/sq.cm uncertainty is +/-0.018kg/cm² 1.82852848 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 07-11-12
Range -0.5 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.51
2 -0.52 Deviation +/- 0.003kg/sq.cm
3 -0.51
4 -0.51
5 -0.51
6 -0.5

Avarage = -0.5117

Standard Deviation = 0.0040825 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.001826 kg/cm²

0.001826 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Vacuum gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0030 0.0015 0.0030


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Accuracy of Master Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0030 0.0017 0.0030
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0100 0.0058 0.0100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Zero Drift

(Rectangular Distribution) <∞>

Zero Drift (kg/cm²) UOM (kg/sq.cm ESTIMATE(kg/sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.0000 0.0000 0.0000 0.0000 0.0000

UC = 0.0001

8.35879E-05

= 0.009142642 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
6.986938632E-09 UA 4_
0.000 4 Inifinity 2.778E-12

Effective degrees of freedom = 2515

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 0.018285 kg/cm²


Reporting of Results:
For -0.1 kg/sq.cm uncertainty is +/-0.018kg/cm² 1.82852848 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE Date: 07-11-12
Range -0.8 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.81
2 -0.82 Deviation +/- 0.003kg/sq.cm
3 -0.82
4 -0.82
5 -0.82
6 -0.82

Avarage = -0.8183

Standard Deviation = 0.0040825 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.001826 kg/cm²

0.001826 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Vacuum gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

0.0030 0.0015 0.0030


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Accuracy of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.0030 0.0017 0.0030
U3 : Standard Uncertainty Due To Resolution of UUC 0.01/2=0.005kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R. Distribution) <∞>

Resolution(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.0050 0.0029 0.0050

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0100 0.0058 0.0100


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/sq.cm) ESTIMATE(Kg/Sq.cm)

0.0100 0.0058 0.0100

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.00000333 0.000002250 0.000003 0.00001 0.000033 0.000033

UC = 0.00008359

8.35879E-05

= 0.009142642 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
6.986938632E-09 UA 4_
0.000 4 Inifinity 2.778E-12

Effective degrees of freedom = 2515

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 0.018285 kg/cm²


Reporting of Results:
For -0.8 kg/sq.cm uncertainty is +/-0.019kg/cm² 1.82852848 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE ( AT SITE) Date: 08-10-12
Range 30 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 30.2
2 30.2 Deviation +/- 0.7kg/sq.cm Max
3 30.2
4 30.2
5 30.2

Avarage = 30.2000

Standard Deviation = 0 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.000000 kg/cm²

0.000000 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

1.2000 0.6000 1.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U5 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.3600 0.1633 0.0033 0.0000 0.0000 0.0000

UC = 0.5267

0.52670653

= 0.7257455 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.2774197688673 UA 4_
0.000 4 Inifinity 0.000E+00

Effective degrees of freedom = #DIV/0!

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.451491 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.46kg/cm² 0.24191517 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE ( AT SITE) Date: 08-10-12
Range 100 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 100.5
2 100.5 Deviation +/- 0.7kg/sq.cm Max
3 100.3
4 100.5
5 100.5
6 100.5

Avarage = 100.4667

Standard Deviation = 0.0816497 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.036515 kg/cm²

0.036515 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

1.2000 0.6000 1.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.2000 0.1155 0.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U5 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0013 0.3600 0.1633 0.0033 0.0133 0.0000 0.0000

UC = 0.5414

0.541373979

= 0.735781203 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.2930857851937 UA 4_
0.000 4 Inifinity 4.444E-07

Effective degrees of freedom = 659443

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.471562 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.46kg/cm² 0.2452604 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE ( AT SITE) Date: 08-10-12
Range 300 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 300.2
2 300.2 Deviation +/- 0.7kg/sq.cm Max
3 300.2
4 300.2
5 300.2
6 300.2

Avarage = 300.2000

Standard Deviation = 0 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.000000 kg/cm²

0.000000 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

1.2000 0.6000 1.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.2000 0.1155 0.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U5 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.3600 0.1633 0.0033 0.0133 0.0000 0.0000

UC = 0.5400

0.540040646

= 0.734874578 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.2916438990274 UA 4_
0.000 4 Inifinity 0.000E+00

Effective degrees of freedom = #DIV/0!

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.469749 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.46kg/cm² 0.24495819 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity
Calculation of Measurment Uncertainty
PRESSURE GAUGE ( AT SITE) Date: 08-10-12
Range 600 kg/sq.cm

Type A evaluation

(a) Standard uncertainty in the reading (UA)

Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 600.4
2 600.4 Deviation +/- 0.7kg/sq.cm Max
3 600.4
4 600.4
5 600.4
6 600.4

Avarage = 600.4000

Standard Deviation = 0 kg/cm²

Degree of freedom = 4

Standard deviation of mean or standard uncertainty UA = 0.000000 kg/cm²

0.000000 kg/cm²

Type B evaluation

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U1 : Standard Uncertainty Due To Uncertainty of master (Digital Pressure gauge) :-

(N.. Distribution) <∞>

Uncert. of Master(kg/cm²) UOM (kg/cm²)) ESTIMATE(kg/cm²)

1.2000 0.6000 1.2000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>
U2 : Standard Uncertainty Due To Deviation on Standard :-
(R. Distribution) <∞>
Dev. of Master(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)
0.7000 0.4042 0.7000
U3 : Standard Uncertainty Due To Resolution of UUC 0.2/2=0.1kg/cm² :-

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>


(R Distribution) <∞>

Resolution (kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)


0.1000 0.0577 0.1000

<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U4 : Standard Uncertainty Due To Hystersis :-

(Rectangular Distribution) <∞>

Hystersis(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000


<Source>, <Distribution>, Degree of Freedom>, <Calculation>, <Value>

U5 : Standard Uncertainty Due To Liquid head correction 10cm, 0.00095kg/cm²/cm

(Rectangular Distribution) <∞>

L.H.Correction(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0095 0.0055 0.0095

U5 : Standard Uncertainty Due To Zero drift

(Rectangular Distribution) <∞>

Zero Drift(kg/cm²) UOM (kg/cm²) ESTIMATE(kg/cm²)

0.0000 0.0000 0.0000

Combined uncertainty :-

UC = UA² U1² U2² U3² U4² U5² U6² U7² U8²

0.0000 0.3600 0.1633 0.0033 0.0000 0.0000 0.0000

UC = 0.5267

0.52670653

= 0.7257455 kg/cm²

Effective degrees of freedom to be calculated using Welch-Satterthwaite formula>


. UC4

Effective degrees of freedom = UA4 +


UB
UC 4 __
0.2774197688673 UA 4_
0.000 4 Inifinity 0.000E+00

Effective degrees of freedom = #DIV/0!

Expanded uncertainty :-

For 95 % level of confidence, the coverage factor k = 2 (Refer NABL 141, )

UE = +/- 1.451491 kg/cm²


Reporting of Results:
For 30 kg/sq.cm uncertainty is +/-1.46kg/cm² 0.24191517 % FS
with coverage factor k=2 for confidence level of 95 % and Effective Degree of freedom = infinity