Pressure & Vacuum
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 29.8
2 29.8 Deviation +/- 0.7kg/sq.cm Max
3 29.8
4 30.0
5 29.8
6 30.0
Avarage = 29.8667
Degree of freedom = 4
0.046188 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.3118
0.311773979
= 0.558367244 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 80.1
2 80.4 Deviation +/- 0.7kg/sq.cm Max
3 80.4
4 80.4
5 80.4
Avarage = 80.3400
Degree of freedom = 4
0.060000 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.3532
0.353242993
= 0.594342488 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.72kg.sq.cm
1 299.8
2 299.6 Deviation +/- 0.7kg/sq.cm Max
3 299.6
4 299.8
5 299.8
6 299.6
Avarage = 299.7000
Degree of freedom = 4
0.048990 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.3120
0.312040646
= 0.558605984 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Dead weight tester
Sr. No. Measured Reading (bar) Uncertainty = 0.11
1 700.4644
2 700.4587 Deviation +/- 0.4871 bar
3 700.4826
4 700.4832
5 700.4800
6
Avarage = 700.4738
Degree of freedom = 4
0.005102 bar
Type B evaluation
Combined uncertainty :-
UC = 33.4475
33.44745974
= 5.783377883 bar
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 4.98
2 4.98 Deviation +/- 0.03kg/sq.cm Max
3 4.98
4 4.98
5 4.98
6 4.98
Avarage = 4.9800
Degree of freedom = 4
0.000000 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.0024
0.002396772
= 0.048956836 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 9.98
2 9.99 Deviation +/- 0.03kg/sq.cm Max
3 9.98
4 9.99
5 9.98
6 10.0
Avarage = 9.9833
Degree of freedom = 4
0.002309 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.0024
0.002402105
= 0.049011276 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.09kg.sq.cm
1 80.10
2 80.40 Deviation +/- 0.03kg/sq.cm
3 80.40
4 80.40
5 80.40
6 29.98
Avarage = 71.9467
Degree of freedom = 4
9.194593 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 84.5429
84.54293011
= 9.194722949 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.11
2 -0.12 Accuracy +/- 0.003kg/sq.cm
3 -0.11
4 -0.11
5 -0.11
6 -0.11
Avarage = -0.1117
Degree of freedom = 4
0.001826 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.0001
8.35879E-05
= 0.009142642 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.51
2 -0.52 Deviation +/- 0.003kg/sq.cm
3 -0.51
4 -0.51
5 -0.51
6 -0.5
Avarage = -0.5117
Degree of freedom = 4
0.001826 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.0001
8.35879E-05
= 0.009142642 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-0.003kg.sq.cm
1 -0.81
2 -0.82 Deviation +/- 0.003kg/sq.cm
3 -0.82
4 -0.82
5 -0.82
6 -0.82
Avarage = -0.8183
Degree of freedom = 4
0.001826 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.00008359
8.35879E-05
= 0.009142642 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 30.2
2 30.2 Deviation +/- 0.7kg/sq.cm Max
3 30.2
4 30.2
5 30.2
Avarage = 30.2000
Degree of freedom = 4
0.000000 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.5267
0.52670653
= 0.7257455 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 100.5
2 100.5 Deviation +/- 0.7kg/sq.cm Max
3 100.3
4 100.5
5 100.5
6 100.5
Avarage = 100.4667
Degree of freedom = 4
0.036515 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.5414
0.541373979
= 0.735781203 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 300.2
2 300.2 Deviation +/- 0.7kg/sq.cm Max
3 300.2
4 300.2
5 300.2
6 300.2
Avarage = 300.2000
Degree of freedom = 4
0.000000 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.5400
0.540040646
= 0.734874578 kg/cm²
Expanded uncertainty :-
Type A evaluation
Measurment Record
Master Instrument =Digital Pressure gauge
Sr. No. Measured Reading (kg/sq.cm) Uncertainty = +/-1.2kg.sq.cm
1 600.4
2 600.4 Deviation +/- 0.7kg/sq.cm Max
3 600.4
4 600.4
5 600.4
6 600.4
Avarage = 600.4000
Degree of freedom = 4
0.000000 kg/cm²
Type B evaluation
Combined uncertainty :-
UC = 0.5267
0.52670653
= 0.7257455 kg/cm²
Expanded uncertainty :-