UMA CO., LTD.

MEASURE HDL
2-19-6 Yokosuka Reagent for measuring HDL Cholesterol
Matsudo, Chiba, Japan Inhibition/Direct

á 2 ~ 8 °C IVD In vitro Diagnostics Packages


R1 1 ´ 90 mL R2 1 ´ 30 mL
T DO NOT freeze 6 18 months/block from light R1 1 ´ 60 mL R2 1 ´ 20 mL

1. PURPOSE OF USE 6. MEASUREMENT PRINCIPLE


In vitro Determination of HDL cholesterol in serum or This method enables determination of HDL Cholesterol
plasma. only in a selective manner by combination of surfactant,
inorganic and organic phosphorus compounds which
2. GENERAL INSTRUCTION
inhibits reaction against LDL, VLDL and chylomicron.
1. For in vitro diagnostics use only.
st
In the 1 reaction, LDL, VDL and chylomicron except for
2. Diagnosis should be made in a comprehensive manner,
in accordance with other related test results and clinical HDL-Cholesterol are bound to surfactant, inorganic and
nd
organic phosphorous compounds. In the 2 reaction, only
symptoms by the doctor in attendance.
HDL-Cholesterol can be determined by addition of CHER
3. For guaranteed results, usage of this product must
and CHOD without doing fractionation.
comply with the instruction in this manual.
"#$%,/()*+,-.,/.,
4. For automatic analyzers: follow their instructions. 0*1,/2-/ 2/3*1,/2- -3453)/6(
HDL Cholesterol + H2O
3. MATERIALS REQUIRED BUT NOT INCLUDED
Cholesterol + fatty acid
- Saline 0.9 % and high grade purified water "#07
Cholesterol + O2 Cholestenon + H2O2
- Micropipet and other basic laboratory equipment. + 807
2 H2O2 + 4-AA + ADPS + H3O
- Calibrators and Controls (separatedly sold)
blue-violet quinone pigment + 5H2O
4. REAGENT COMPOSITION & PREPARATION (λmax = 583 nm)
- Reagent R-1: HDAOS; ascorbate oxidase, H3PO4.
7. STANDARD MEASUREMENT OPERATION
Reagent R-1 is ready for use
Specimen Calibrator Blank
- Reagent R-2: CHOD, Cholesterol Esterase (CHER),
(S) (Std) (B)
Peroxidase (POD), 4-aminoantipyrine (4-A-A).
Specimen (µL) 20 - -
Reagent R-2 is ready for use
Calibrator (µL) - 20 -
- Calibrators (separatedly sold): Ready for use
Saline (µL) - - 20
- Controls Lyo-1 & Lyo-2 (separatedly sold): Put 1 mL of
R-1 (µL) 1500 1500 1500
purified water to the vials of calibrator and controls (L, H);
Incubate at 37 °C in 5 minutes
leave at room temparature for 30 minutes before use. After
R-2 (µL) 500 500 500
reconstituted, calibrators/control can be use without dilution.
Mix well; incubate at 37 °C for 5 minutes; measure
5. SAMPLE PREPARATION & STORAGE absorbance at 600 nm
- Serum: Collect blood after 12 ~ 14 hours fasting. Wait Note: See sample preparation for details of specimen
until sample completly coagulated. Take the supernatant to
use as specimen. 8. CALCULATION & UNIT CONVERSION

- Plasma: Collect blood after 12 ~ 14 hours fasting. Treat Calculation

sample by anticoagulant (i.e heparin lithium); leave sample - Calculate ∆Abs of specimen & standards vs blank

to stand for 3 hours or centrifuge at 2000 rpm for 2 minutes; - Plot a calibration curve HDL (mg/dL) = f(∆Abs)

take the plasma layer (supernatant) and use as specimen. - Calculate HDL concentration in specimen using the

- Analyze sample soon after collection. In case, it could curve (doing same procedure for Controls)

not be analyzed soon, store sample 2 ~ 8 °C and analyze Unit conversion

within 4 days. 1 mg/dL = 0.026 mmol/L

1/2 Revision 06/2013


9. PERFORMANCE & CORRELATION TEST 13. HANDLING, USAGE & DISPOSAL
Performance Handling
- Sensitivity: Change in absorbance when using purified 1. Specimen can be potentially positive for infectious agents
water is 0.001 ~ 0.015, and change in absorbance using including hepatitis B virus and HIV. Wear glove and goggle
standard solution (cholesterol 50 mg/dL) as specimen is when needed.2. In case reagents got into skin, eye or
0.05 ~ 0.15. mouth by mistake, wash it immediately with plenty of water
- Specificity: The accuracy is within ±10.0%. and consult the doctor if needed.
- Reproducibility: CV value < 3.0%. 3. If reagents are spilled, dilute with water and wipe it out. If
- Measuring range: 5 ~ 120 mg/dL. specimen is spilled, spray 80% of alcohol over the
Correlation test specimen and wipe it out.
Same measuring principle Usage
- Serum: 1. Store reagents under specified condition. Do not use
Regression equation: Y = 1.034X-2.70 (n = 100) after expiration date.
Correlation coefficient: R = 0.999 2. Do not use the container and auxiliaries included in this
(Y: Value obtained from using this method) kit for other purposes.
Reference Materials for Calibration 3. Do not mix reagents of different lot for use.
- JCCRM 224 4. Do not add to the reagent being used even if it is the
same lot number.
10. EXPECTED VALUES
Normal reference range Disposal
1. All specimens, as well as all instruments (e.g. test tubes)
- 40 - 70 mg/dL;
Reference range should be established at each facility and that come in contact with the specimens, must be treated by
the following methods, or they must be treated according to
judgement should base on measurement results in a
the manual for infectious medical waste provided in each
comprehensive manner together with clinical symptoms
facility.
and other measurement results.
・ Sterilize with an autoclave, subjecting them to high
11. INTERFERENCES
pressure saturated steam at 121 °C for more than 20
- Ascorbic acid up to 50 mg/dL: within ±3%
minutes. Do not process waste containing sodium
- Conjugated bilirubin upto 20 mg/dL: within ±3%
hypochlorite solution with an autoclave.
- Hemoglobin 500 mg/dL: within ± 1%
・Immerse at least one hour in sodium hypochlorite solution
- Formazin turbidity upto 1600 unit: within ± 1%.
(active chloride concentration of over 1000 ppm).
- Abnormal lipoprotein contained sample may affect to the
2. This reagent contains sodium azide. Sodium azide can
determination result.
react with lead pipe and/or steel pipe and can generate
12. INFORMATION FOR AUTOANALYZERS explosive metal azide. Make sure to use plenty of water at
Calculation Method 2-point (fix) disposal. Concentration of sodium azide in R-2 is 0.05%.

Temperature 37 °C 14. OTHER INSTRUCTIONS AND CAUTION


Specimen 2.0 - Results may differ depending on the sample/reagent
Volume (μL) R1 150 ratio. Adjust parameters for different analyzer.
R2 50 - Prepare the calibration curve on the day of
Main 600 determination.
Wavelength (nm)
Sub- 800
Point 1 10
Measurement
Point 2 16
(cycle) Point 3 34

Calibration type Linear


Unit mg/dL

2/2 Revision 06/2013