Professional Documents
Culture Documents
Operator Manual For ELAN 30s - v2016.00.02
Operator Manual For ELAN 30s - v2016.00.02
Operator Manual For ELAN 30s - v2016.00.02
For
Transasia Bio-medicals Limited, Transasia House, 8 Chandivali Studio Road, Andheri (East),
Mumbai-400072, India
Operator Manual ELAN 30s
Foreword
This manual is organized in a progressive sequence for easy study and reference. It is an
instructional aide to provide a reference for easy operation and general maintenance of
this analyzer. It contains detailed description of the analyzer features and specifications.
The analyzer is used with operational PC and Printer, and can interact with the host
computer. The operational PC consists of the application software for the user to operate
the analyzer.
All the samples and reagents for measurements including samples obtained from patients
are controlled by barcodes enabling the analyzer to perform the entire process of the
analysis automatically.
Use of the analyzer with proper knowledge will ensure quality test results and trouble free
analyzer operation and performance.
This operation manual is prepared based on the assumption that the user has knowledge
of immunology.
ii
Operator Manual ELAN 30s
Legal Information
All rights are reserved by manufacturer of this product.
The contents of this document are subject to change without prior notice and without legal
obligation.
This document and the information herein are provided for the sole use of the intended
recipient(s) and for information purposes only.
This document contains contents which are the confidential and proprietary information of
the manufacturer.
No part of this document can be copied or reprinted, in whole or in part, without prior
written permission from the manufacturer.
iii
Operator Manual ELAN 30s
Contents
1. Safety Information......................................................................................................... 1-1
1.1. Safety Instructions ............................................................................................... 1-1
1.2. Warning Labels.................................................................................................... 1-2
Precautionary Warning .............................................................................................. 1-3
1.3............................................................................................................................... 1-3
2. Introduction to the Analyzer......................................................................................... 2-1
2.1. Introduction .......................................................................................................... 2-1
2.1.1. Operating Principle..................................................................................................... 2-2
2.1.2. Sequence of Operation .............................................................................................. 2-5
2.2. Analyzer Overview ............................................................................................... 2-6
2.2.1. Front View with Dome Closed .................................................................................... 2-6
2.2.2. Front View with Dome Open ...................................................................................... 2-6
2.2.3. Left Side View ............................................................................................................ 2-7
2.2.4. Right Side View .......................................................................................................... 2-7
2.3. Technical Specification ........................................................................................ 2-8
2.3.1. General Specifications ............................................................................................... 2-8
2.3.2. Reader Specification*................................................................................................. 2-8
2.3.3. Washer Specification.................................................................................................. 2-9
2.3.4. Incubator Specification ............................................................................................... 2-9
2.3.5. Product Specification.................................................................................................. 2-9
2.3.6. Sample Pipetting ........................................................................................................ 2-9
2.3.7. Reagent Pipetting..................................................................................................... 2-10
3. Pre-installation Requirements...................................................................................... 3-1
3.1. Installation Conditions.......................................................................................... 3-1
3.2. Site Requirements ............................................................................................... 3-4
3.2.1. Proper Room Temperature and Humidity .................................................................. 3-4
3.2.2. External Interferences ................................................................................................ 3-4
3.2.3. Proper Room Lighting ................................................................................................ 3-5
3.2.4. Proper Laboratory Cleanliness (Practice) .................................................................. 3-5
3.2.5. Electrical Requirements ............................................................................................. 3-5
3.3. Floor Requirements ............................................................................................. 3-6
4. Installation ..................................................................................................................... 4-1
4.1. Accessories Checklist .......................................................................................... 4-1
4.2. Basic Operational Information.............................................................................. 4-3
4.2.1. Receiving Instructions ................................................................................................ 4-3
4.2.2. Warranty Information.................................................................................................. 4-3
4.3. Unpacking the Analyzer ....................................................................................... 4-4
4.4. Installing the Components of the Analyzer......................................................... 4-13
4.4.1. Installing the Sample and Reagent Tray (SRGT)..................................................... 4-13
4.4.2. Installing Handheld Barcode Scanner ...................................................................... 4-15
4.4.3. Installing DI Water, Wash Buffer and Waste bottles ................................................ 4-15
4.5. Connecting the Analyzer with Computer............................................................ 4-18
4.6. Printer Installation .............................................................................................. 4-18
4.7. Turning on the System....................................................................................... 4-19
4.8. Software Installation Procedure ......................................................................... 4-20
4.8.1. Minimum System Configuration ............................................................................... 4-20
iv
Operator Manual ELAN 30s
v
Operator Manual ELAN 30s
vi
Operator Manual ELAN 30s
vii
Operator Manual ELAN 30s
Document Conventions
Before reading the manual, please get familiarized with the following icons used in this
manual.
Bio-hazard
Electric Shock
High Temperature
Injury
viii
Operator Manual ELAN 30s
1. Safety Information
Before operating the analyzer, please read the safety information given in the
manual
1-1
Operator Manual ELAN 30s
1-2
Operator Manual ELAN 30s
1-3
Operator Manual ELAN 30s
1-4
Operator Manual ELAN 30s
2-1
Operator Manual ELAN 30s
2-2
Operator Manual ELAN 30s
Incident or white light contains the entire spectrum, objects that appear colored,
absorb light at a particular wavelength, and reflect others, thus giving different
colors. That is color is a function of its wavelength.
Light having a wavelength of less than 400 nm is termed Ultraviolet, whereas light
having a wavelength greater than 800 nm is described as Infrared, both ultraviolet
and infrared lights are invisible to the human eyes. Light corresponding to
wavelengths between 400 nm and 800 nm is visible to the human eye and is
termed as visible light.
2-3
Operator Manual ELAN 30s
Since the total incident light = light absorbed + light transmitted, it follows that:
Therefore as the absorbed light (Absorbance) increases, the transmitted light will
decrease
As we increase the concentration of the colored solution, the light absorbed
increases, and we find that the transmittance varies inversely and logarithmically
with concentration.
Using the above formula, for different percentage of transmittance we obtain the
following absorbance.
Percentage Transmittance Absorbance
100 0.000
50 0.301
25 0.602
12.5 0.903
2-4
Operator Manual ELAN 30s
The study of the correlation between the concentration of a colored liquid, the
intensity of its color and the amount of light absorbed by the colored solution is
termed as Colorimetry.
Useful Information:
Percent Absorbance Absolute Error Per Cent Relative Error
10 1.000
11 0.959 0.041 4.1
45 0.347
46 0.337 0.001 2.9
90 0.046
91 0.041 0.005 10.9
95 0.022
96 0.018 0.004 8.2
The relative error is minimal at an absorbance of 0.434 (36.8 %T). Consequently,
methods should be designed such that readings fall near the center of the scale,
preferably within an absorbance of approximately 0.1 and 0.7 (20 and 80%T).
2-5
Operator Manual ELAN 30s
2-6
Operator Manual ELAN 30s
2-7
Operator Manual ELAN 30s
2-8
Operator Manual ELAN 30s
2-9
Operator Manual ELAN 30s
2-10
Operator Manual ELAN 30s
3. Pre-installation Requirements
Before installing the analyzer, you must follow the pre-installation requirements to
ensure the accuracy and precision of the analyzer.
The proper location is an important consideration. Poor location can lead to
malfunction of the analyzer.
3-1
Operator Manual ELAN 30s
3-2
Operator Manual ELAN 30s
8. The following cautions should be taken when using and handling the
reagents:
a. After unpacking the reagents, be sure not to allow contamination of
reagents (e.g. dust, dirt etc.).
b. Do not use reagents that are expired.
c. Handle reagent gently to avoid formation of bubbles.
d. Take care not to spill the reagent. If it spills, wipe it off immediately using
a wet cloth.
e. Follow other instructions described in the package insert on each
reagent.
f. If a reagent happens to enter your eye, wash it off immediately using
plenty of water, and take medical treatment at once.
g. If you swallow it inadvertently, call for a doctor immediately and drink
plenty of water.
Some reagents are strong acids or alkalis. Exercise great h
care so that your hands and clothing do not come into
contact with reagents. If your hands or clothing come into
contact with either reagent, immediately wash them off
with soap and water. If a reagent comes into contact with
your eye(s), immediately rinse with water for at least 15
minutes.
Prohibit any alteration and/or modification to the analyzer without
permission from the manufacturer.
9. The following precautions should be taken for preventing infection due to
sample handling:
Do not touch the samples, mixtures, & waste liquids with
bare hands. Be sure to wear gloves to protect you from
coming in contact with infectious sample. In case any
samples come in contact with your skin, thoroughly rinse
the area that came in contact with the sample & consult a
physician. Immediately wipe off any contaminants from
the system.
10. The following precautions should be taken for disposing the bio-hazardous
waste:
Treat the drain water as infectious waste. Collect the drain
water in reserve can & allow it to be disposed off by expert
distributors.
11. Space Requirements:
Analyzer dimensions:
3-3
Operator Manual ELAN 30s
Dome close
754mm ( L) x 532mm (B) x 535mm (H)
Dome Open
754mm ( L) x 532mm (B) x 888mm (H)
3-4
Operator Manual ELAN 30s
3.2.5.2. Grounding
It is absolutely necessary that a perfect earthing must be provided at the
power source with all applicable local requirement (Only a certified
grounded, 3 pin power plug should be used).
3-5
Operator Manual ELAN 30s
The 3 pin power cord plugs shipped with the analyzer, computer,
printer and monitor may not be compatible with the local electrical
sockets of some countries. Procure it from the local market prior
to the installation.
3-6
Operator Manual ELAN 30s
4. Installation
4.1. Accessories Checklist
The manufacturer reserves right to change the accessories
supplied with ELAN 30s analyzer without notice.
The main unit and accessories are packed in separate cartons. Authorized
representative is responsible for unpacking, installing, and initial setting up of the
analyzer.
Serial Item
Description Quantity
Number Code
FLS BOTTLE BOX
FLS BOTTLE (4 LIT.) ASSEMBLY WITH LEVEL SENSOR 1 No.
1
108317 FOR DI WATER
FLS BOTTLE (4 LIT.) ASSEMBLY WITH LEVEL SENSOR 1 No.
2
106540 FOR WASTE
FLS BOTTLE (1 LIT.) ASSEMBLY WITH LEVEL SENSOR 1 No.
3
106532 FOR BUFFER - 1
FLS BOTTLE (1 LIT.) ASSY. WITH LEVEL SENSOR FOR 1 No.
4
108323 BUFFER - 2
FLS BOTTLE (1 LIT.) ASSY. WITH LEVEL SENSOR FOR 1 No.
*5
108326 BUFFER - 3
6 111787 TUBING SET FOR FLS BOTTLE ASSY 1 Set
7 108026 BOTTLE TRAY 1 No.
8 106490 ASSEMBLY FOR SRGT 1 No.
9 107878 SRGT COVER 1 No.
10 107877 STRIP TRAY COVER 1 No.
11 120958 ERBA Auto Wash (10x100ml) 1 No.
SHIPPER BOX
12 120161 PLASTIC TEST TUBES(12MM X 75MM) 60 Nos.
13 114452 TIP BIN CONTAINER FOR ELAN 30 Nos.
14 130053 VAILS CAP: 2ML 50 Nos.
15 111967 20ml WIDE MOUTH REAGENT BOTTLE 50 Nos.
SAMPLE TIPS - BLACK NIGHT , ARTICLE NO. 49000-
16 113591 192 Nos.
0000; 300ul, One pack/rack of 96 tips : RITTER MAKE
17 111784 WHITE COLOUR UNCOATED STRIP 1 No.
18 115177 TEST TUBE (12mm X 75mm LONG) BLACK 3 Nos.
114769 CABLE USB JUMPER USB2.0 28x1P+ 24x2C - USB A to 1 No.
19 B 2.0 version WITH FERRITE CORE (BOTH SIDE A,B )
1.8 meter
POWER CORD INDIAN PLUG ST 250VAC/6A - IA6A3 3 x 1 No.
20
106046 0.75 sq. mm 1.8 mtr. Length
4-1
Operator Manual ELAN 30s
Serial Item
Description Quantity
Number Code
*21 109788 HAND HELD BARCODE SCANNER 1 No.
22 170016 UNCOATED STRIP PLATE 1 No.
ADAPTOR FOR MICROCENTRIFUGE TUBE 10 Nos.
#23
113651 (EPPENDROF)
TOOL KIT
24 101494 PLASTIC TOOL BOX 1 No.
25 101688 SCREW DRIVER X100L CR-V CARBON TIP (1 No) 1 No.
26 101687 SCREW DRIVER X100L CR-V CARBON TIP (2 No) 1 No.
27 101689 SCREW DRIVER IMPORTED ( - ) NO.3 1 No.
28 101675 NUT DRIVER 5.5 MM FOR M3 1 No.
29 101676 NUT DRIVER 7.0 MM FOR M4 1 No.
30 101495 BOX SPANNER (10 -11MM) 1 No.
31 101692 ALLEN DRIVER (M3) 2.5 MM T TYPE-9" LONG 1 No.
32 101693 ALLEN DRIVER (M4) 3.0 MM T TYPE-9" LONG 1 No.
33 101694 ALLEN DRIVER (M5) 4.0 MM T TYPE-9" LONG 1 No.
HEX BALL ALLEN KEY SET CONSISTING OF 1.5mm,
34 111604 2mm, 2.5mm,3mm,4mm ,5mm ,6mm ALLEN KEYS WITH 1 Set
PLASTIC HOLDER & BLISTER PACKING
35 101514 NOSE PLIER 1 No.
36 101515 FLAT SPANNER 10 / 11 1 No.
37 101524 FLAT SPANNER 14-15 MM 1 No.
38 100323 METALLIC FORCEP 2A-SA (NOSE WITH RADIUS) 1 No.
39 100295 TRIMMER 933 NO 1 No.
40 101516 SPANNER SET SIZE: 1 to 8 Nos. 1 No.
41 101678 6" ADJUSTABLE SPANNER 1 No.
42 100294 TUBE CUTTER --( LEGRIS-MAKE) 1 No.
43 109350 KNOB FOR STRIP TRAY 2 Nos.
PM KIT
44 101677 PROBE CLEANER 1 No.
45 111755 PHOTOMETER LAMP ASSEMBLY 2 Nos.
46 107815 SET OF FUESES 1 Set
47 107180 25micron FILTER FOR DI 4 Nos.
48 107179 25micron FILTER FOR BUFFER 10 Nos.
1/8-27 NPT Thread with 7/16" Hex to Classic Series Barb, 1 No.
49 107170
3/16" (4.75 mm) ID Tubing;. 1850-1
NUT PANEL MOUNT 1/4-28 UNF WITH 7/16" HEX
50 100884
PMSN-X0 5 Nos.
PANEL MOUNT CONNECTOR, 1/4-28 UNF TO 200
51 107881
SERIES BARBS, 1/8"(3MM) ID TUBING,; PMS230-1 5 Nos.
4-2
Operator Manual ELAN 30s
Serial Item
Description Quantity
Number Code
53 -- SOFTWARE CD 1 No.
54 109375 HYDRAULIC DIAGRAM (ELAN 30s) With BOM Rev.07 1 No.
55 -- FQC report 1 No.
* - Optional (As per the customer’s requirement)
# Recommended Micro centrifuge Tube of 2.0ml “Tarsons/Eppendorf” make to be used with the ADAPTOR
FOR MICROCENTRIFUGE TUBE (EPPENDROF)
4-3
Operator Manual ELAN 30s
4-4
Operator Manual ELAN 30s
3. Cut all the packing tape and remove the top, side, and outer buffer from
covering the analyzer.
4-5
Operator Manual ELAN 30s
4-6
Operator Manual ELAN 30s
5. Remove the clean film and masking tape from the analyzer outer body
7. Screw the lifting handle (4 Nos) in the thread provided on the both side of
the analyzer (to shift the analyzer in working place).
4-7
Operator Manual ELAN 30s
9. Remove the packaging buffer sheet placed below the pump mounting &
stickers from the rear panel & dome front side.
4-8
Operator Manual ELAN 30s
10. Open the dome for unpacking the Robotic ARM, Photometer and Washer
11. Unscrew (2 Nos pan head screw) strip tray-washer packing bracket bearing
caution sticker.
12. Remove the masking tape and front packing buffer of the washer by slightly
lifting washer in upward direction.
4-9
Operator Manual ELAN 30s
13. Remove washer needle packing cap from washer manifold by slightly lifting
the pull out the cap slowly so that needle should not get damage
16. To remove Buffer1 from robotic ARM rotate the clockwise direction inside
the ID of the container .Remove Buffer2 by pulling up side after removing
the Buffer 1.
4-10
Operator Manual ELAN 30s
17. Unmask the tape from robotic ARM to remove Probe packing cap from its
location
18. Remove the masking tape from right side panel to unfix drain collection tray
4-11
Operator Manual ELAN 30s
20. Connect the DI, Waste, Buffer bottles to the analyzer. Follow the color
coding for tube & connector placement
21. Connect the power supply & PC, Analyzer will initialize. The disposable tip
used for packing probe will be ejected at the tip ejected at the tip ejecting
location
22. Place the analyzer at the desired place keeping in view the floor
requirement given in section 3.3 Floor Requirements.
4-12
Operator Manual ELAN 30s
4-13
Operator Manual ELAN 30s
4. Gently place the sample and reagent tray into the container on the
aluminum shaft.
5. After placing the sample and reagent tray into the container, hold the nut on
SRGT tray and slowly rotate it in clockwise/anticlockwise direction. The
index pin will now slide into the index pin hole.
6. Tighten the nut on SRGT tray by rotating it clockwise. Note that the nut will
not get engaged in the shaft if the index pin has not slide properly into the
index pin hole.
7. Unpack the Sample and Reagent tray lid, and place it over the Sample and
Reagent Tray.
8. Make sure that the notch on the SRGT lid fits into the cut provided on the
container.
4-14
Operator Manual ELAN 30s
4-15
Operator Manual ELAN 30s
3. Unpack Float Sensors cables and connect one end to switch provided on
the right side of the panel and connect other end to the respective
BOTTLES (float sensor with blue sleeves for DI WATER, yellow sleeves for
WASH BUFFER 1, green sleeves for WASH BUFFER 2, and black sleeves
for WASH BUFFER 3, red sleeves for WASTE).
4. Fill the DI Water can with 4 liters of fresh DI water and unpack the tubing
from the accessory box.
5. Take two separate tubes and connect one end of one tube to the blue
ringed outlet marked IN and connect one end of another tube to the white
ringed outlet marked OUT. Connect other ends to the 4 liter DI WATER
BOTTLE on the steel nozzle and blue ringed nozzle. (Tube coming from the
blue ringed outlet marked IN should be connected to steel nozzle and the
tube from the white ringed outlet marked OUT should be connected to blue
ringed nozzle on the BOTTLE).
It is must to have proper connection of DI Water tubing to
proper water flow in washer manifold & probe..
6. Now, fill 1 liters Bottles with the cleaning solution. (Prepared as directed in
KIT insert)
7. Take another separate tube and connect one end to the yellow ringed outlet
marked IN and other end to the 1 liter WASH BUFFER 1 BOTTLE on the
steel nozzle.
8. Now take a tube and connect one end to the green ringed outlet marked IN
and other end to the 1 liter WASH BUFFER 2 BOTTLE on the steel nozzle.
9. Similarly, use the same procedure to connect the WASH BUFFER 3
BOTTLE (optional) on the steel nozzle, if required.
10. Now connect the silicon tube to the analyzer outlet marked WASTE marked
OUT and connect the other end of the tubing into the nozzle of 4 liter Waste
bottle.
4-16
Operator Manual ELAN 30s
4-17
Operator Manual ELAN 30s
2. Connect one end of the power cord to the analyzer’s power inlet and other
end to the main power supply as shown in figure above.
Make sure that the On/Off power button should be in OFF
condition before turning on the main power supply.
After connecting the power cord and USB cable with the analyzer, turn on the
main power switch, and then turn on the On/Off power button on the analyzer.
4-18
Operator Manual ELAN 30s
6. Start the ELAN 30s software, if installed in the computer else install the
software. For communicating with the analyzer, you must install the
software. (Refer section 4.8 Software Installation Procedure for installation
procedure in more detail).
When the ELAN 30s software is started, the screen resolution will
automatically initialize to 1024*768 to give best operating experience.
Software makes this setting on start-up automatically. Only if
the PC driver is not supporting the change then the user will
have to do it manually.
4-19
Operator Manual ELAN 30s
4.8.2. Pre-requisite
4.8.2.1. PC Settings Required for Windows 7
PC Settings will be applied automatically while installing the
customized image of Windows 7 Embedded as Operating System.
* A customized image of Windows 7 Embedded is provided as
Operating System for Analyzer PC.
Remove all memory resident software including anti-virus software from the
Analyzer PC if available. To do this use the following procedure:
Click Start button, and click Control Panel. Now, under the Category view,
go to Programs > Programs and Features. Select the desired anti-virus
software from the list, and click Uninstall.
Remove firewall, automatic update, other security software and from the
Analyzer PC.
Click Start button, and click Control Panel. Now, under the Category view,
go to System and Security > Windows Firewall, and click on the link Turn
off Windows Firewall on or off. Now select the option Turn off Windows
Firewall (not recommended), and then click OK button.
To turn off automatic updates, under Control Panel, go to System and
Security > Windows Update, click on the link Change settings, and select
the Never check for updates (not recommended) from the drop down list,
and click OK.
Do not run any other application on the Analyzer PC during batch run on
Analyzer PC.
4-20
Operator Manual ELAN 30s
c. Now click on the link Change power settings (Applicable only if option
is available). The following screen will be displayed.
4-21
Operator Manual ELAN 30s
4-22
Operator Manual ELAN 30s
Similarly apply same settings to power saver option. Repeat the step c
to e.
User account control settings
f. Open User Account Control Settings by clicking the Start button, and
then clicking Control Panel. Under Small or Large icons view, click on
User Accounts and then click Change User Account Control
settings.
4-23
Operator Manual ELAN 30s
4-24
Operator Manual ELAN 30s
4-25
Operator Manual ELAN 30s
4-26
Operator Manual ELAN 30s
4-27
Operator Manual ELAN 30s
4-28
Operator Manual ELAN 30s
4-29
Operator Manual ELAN 30s
Wait until the progress bar is complete and the following screen appears on
the screen.
4-30
Operator Manual ELAN 30s
4. Click Next.
The following screen will be displayed. The installer will install the software
in the default location C:\Elan30s\. It is recommended to install the software
in the default location. To install to a different location enter the appropriate
location or click Browse.
The installation in a location, different from the default one is
not recommendable.
4-31
Operator Manual ELAN 30s
4-32
Operator Manual ELAN 30s
Once the installation completes, the software icon will be created on the
desktop and the following screen will be displayed.
4-33
Operator Manual ELAN 30s
8. Click on Close or (on the top-right corner of this screen) to close the
screen and restart the computer.
4-34
Operator Manual ELAN 30s
6. The above screen indicates that the database is created successfully. Now
you can access the ELAN 30s application (Refer section 4.8.7 Accessing
ELAN 30s Software in more details).
4-35
Operator Manual ELAN 30s
4-36
Operator Manual ELAN 30s
4-37
Operator Manual ELAN 30s
Select the required upgrade script file, and then click Open.
The database up-gradation process will be started and the following
message will displayed “All Corresponding Databases Upgraded
Successfully” after completion.
4-38
Operator Manual ELAN 30s
5. Choose the path to store the backup, and then click Save.
On clicking, the following screen will be displayed to confirm the successful
completion of database backup.
4-39
Operator Manual ELAN 30s
4-40
Operator Manual ELAN 30s
4-41
Operator Manual ELAN 30s
This may result into loss of data, in case you do not have
latest backup.
8. Click Save to save the database backup.
On clicking, a copy of database will be saved in the desired location.
IMPORTANT: Note the path at which you have saved the backup. It is
recommended to copy this Database backup file on other media such as
CD or DVD.
4-42
Operator Manual ELAN 30s
9. The following screen will be displayed after successful deletion of the database.
Browse for the file and select the location where the database backup file is
saved.
4-43
Operator Manual ELAN 30s
11. Select the desired database backup file, and then click Open.
The following screen will be displayed.
4-44
Operator Manual ELAN 30s
The following user login screen will be displayed. This screen will be
displayed every time when you start the application software.
Enter the following details, and then click OK. You may use this Login
details for the first time and create the other Login IDs with appropriate
Access Rights.
4-45
Operator Manual ELAN 30s
At every startup of software, after login the software checks the compatibility
between application and embedded software. In case of mismatch between the
two, following message is displayed:
4-46
Operator Manual ELAN 30s
5. Functional Description
5.1. System Description
The images used in the manual are of indicative nature and may
not reflect the final component due to continuous development of
the product.
The SRGT and STRIP units are always covered by the plastic tray.
5-1
Operator Manual ELAN 30s
5-2
Operator Manual ELAN 30s
5-3
Operator Manual ELAN 30s
The strip tray can be removed by the user for strip loading, cleaning, and changing
the photometer lamp.
5-4
Operator Manual ELAN 30s
5-5
Operator Manual ELAN 30s
5-6
Operator Manual ELAN 30s
5.1.5. Photometer
This module is used for reading the optical density of the samples in the micro
wells of the Strips. The analyzer consists of the movable reader.
The photometer consists of seven filters (4 standard + 3 optional). The
measurement is performed with any combinations of 2 wavelengths selected
among the following 7 wavelengths:
405nm, 450nm, 492nm, 550nm, 595nm, 630nm, 690nm
5-7
Operator Manual ELAN 30s
5-8
Operator Manual ELAN 30s
5-9
Operator Manual ELAN 30s
5-10
Operator Manual ELAN 30s
5.2.1. Tests
It consist of three options:
New: It consists of one option that is used for defining new tests
parameters. The options is:
Manual Entry: This option can be used for manually entering the Test
Details from Kit Insert provided with the Kit.
There are three tests available for selection;
Qualitative Test, Quantitative Test and Semi-Quantitative Test
Open: This screen is used for editing Test Parameters, Test Sequence,
Calculation details, External Control Details and Deleting Tests.
Profiles: This screen is used for defining the new test profiles.
5-11
Operator Manual ELAN 30s
5.2.2. Search
This screen is used for searching the Patient Results.
5.2.4. Service
This screen is used for performing the service action by the service engineers.
5.2.5. Patient
This screen is used for entering the patient related entries like Sample Ids,
Selecting Tests etc.
Scan Barcode: Used for entering Sample Ids either by scanning barcode on
sample test tubes or by auto increment.
Test Selection: For selecting Tests corresponding to each sample Id.
Work List: This button is used for viewing the list of Pending Tests.
Patient Entry: For entering Patient related information.
5.2.6. WorkBench
This screen is used for selecting the tests to be performed by analyzer in the
current batch.
5.2.7. Results
This screen is used for displaying the results of the current and previous batches.
5.2.8. Maintenance
This screen is used for performing Maintenance operations at the start and end of
the day.
5.2.9. Shutdown
This screen is used to turn off the application software.
5-12
Operator Manual ELAN 30s
6. Routine Task
This chapter describes the routine tasks that you need to perform to keep your
instrument running.
6-1
Operator Manual ELAN 30s
6-2
Operator Manual ELAN 30s
6-3
Operator Manual ELAN 30s
6-4
Operator Manual ELAN 30s
7. Patient
The Patient screen is used for entering the patient related details into the system.
It can be accessed through Patient screen present on the toolbar. The patient
screen will be displayed as shown in figure below:
The following are some of the important functions performed by this module.
Managing patient specific data.
Managing reference doctor data
Linking patient data with medical requirement such as tests (scheduling)
etc.
Keeping account of the tests to be performed and also the list of pending
tests.
The following sections explain the functioning of the buttons provided in the Patient
screen.
7-1
Operator Manual ELAN 30s
5. Now, place the scanned sample tube on position number 1 on the SRGT tray.
6. Similarly, for placing the remaining barcoded samples follow the same step
from 3 to 4.
7. Click on Save to save the IDs.
7-2
Operator Manual ELAN 30s
7.2. Manual
This option is used to add the multiple samples manually by entering the starting
position number. The sample details are added in the incremental order. To do this
you have to enter the start position, sample ID, and the number of samples to be
added. To add the samples, use the following step:
1. Open Patient screen.
2. Click on Manual button.
On clicking, the following screen will be displayed:
3. Now enter the appropriate position of the sample in the Start Pos. field, sample
ID in ID filed, and the number of sample to be placed in the No. of Samples
field.
4. Click on Ok.
On clicking, the sample ID will be displayed on the screen.
Now, enter the appropriate tube type for each sample positions.
7-3
Operator Manual ELAN 30s
5. Now, click on the appropriate sample position. This will rotate the SRGT tray to
the required sample position.
6. Place a required Sample tube on the sample position.
Before placing the sample, make sure that you are placing
the correct Sample tube having same ID number.
7. Repeat the same step from 5 to 6 for the remaining sample tube.
8. Click on Save to save the IDs.
7-4
Operator Manual ELAN 30s
For example, position defined for reagents R2, R3 & R4 as sample tube as
shown in below screen:
7-5
Operator Manual ELAN 30s
7. User can not define the sample position more than 40, if option "Use sample
positions (41-58) for reagents" is selected. In case, if sample positions are
more than 40, then the following message will be displayed.
7-6
Operator Manual ELAN 30s
7.3. Edit
This option is used to edit the sample ID of a sample by specifying the position
number.
1. To edit, click on this button and enter the sample position number, then click
OK.
7-7
Operator Manual ELAN 30s
Once the sample IDs are defined and the sample tubes are placed on the tray, you
can assign the required number of tests to the samples by clicking on the list of
available tests. Also, you will be able to program the number of replicates for each
sample for a particular test by clicking scroll button of the mouse on the space
provided.
Here you will be able to select the tests very quickly for the scanned samples
which are entered in the Patient screen.
7-8
Operator Manual ELAN 30s
7-9
Operator Manual ELAN 30s
The screen is divided into two windows. On the left side, the available tests
are listed, and on the right side, sample details with their assigned position
are listed.
4. Select the required tests to be performed by clicking checkboxes provided
against each tests as shown below.
7-10
Operator Manual ELAN 30s
6. You can select a test for multiple samples by clicking on the cells for a test
against any sample ID and then right click & drag the mouse down for the
required samples. For example, to perform LH EM test on sample IDs 1235,
1236, 1237 click on cell for LH EM against 1235 and then right click & drag the
mouse down up to sample ID 1237. The test LH EM will now be selected for
these sample IDs.
7. You can also define the required number replicates to each test, if required.
To do this, use the following procedure:
a. Place the cursor over any test (for example sampleID 1237 and test
HIV), and single click Scroll button of the mouse. This will show
Replicates option.
7-11
Operator Manual ELAN 30s
b. Click on the option and enter the number of replicates to be defined, and
click Ok.
8. In case, if you can select the listed Test profiles (this will automatically select all
the tests listed in that particular Profile).
7-12
Operator Manual ELAN 30s
1. To enter the patient details, click on the New button. Enter the required details
about the patient and click on Save to save the details.
Patient ID will be generated automatically.
2. Fields shown in red color are mandatory fields, which should not be left blank.
3. To enter the details for a particular sample, use the following procedure:
a. In the Patient Entry screen, click on New.
b. Click on the Sample ID text box. On clicking, a list of sample ID will be
displayed with dates.
c. Click on the appropriate sample ID. This will add the sample ID
d. After this, enter the appropriate details, and click Save to save the
details.
4. To edit the details for a particular sample, use the following procedure:
a. In the Patient Entry screen, click on the Sample ID text box.
b. Select the sample ID to be edited by clicking over it.
c. Now the previously entered details for that particular sample ID are
displayed. Now you can add the new details and then click Save to save
the settings.
7-13
Operator Manual ELAN 30s
5. You can browse through the records using the arrow keys provided as shown
below.
7.7. Worklist
Click on the Worklist button on the Patient screen. Once the Tests are
programmed in the Test Selection screen, the details of the completed tests and
pending tests will be enlisted.
The screen is divided into two windows.
Patient Worklist
Pending Worklist
In the Patient Worklist, you can view the status of the all the scheduled samples
with their sample ID and the test to be performed on those samples.
In pending Worklist, you can view the status of the samples after at least one
complete Run of a batch. Once a batch of Tests is Run, the samples, programmed
for the remaining tests (i.e. the tests which were not present in the Batch, whose
Run is completed or the pending tests) are listed with their samples I.D’s.
The samples whose tests were kept pending are displayed in Yellow color and the
samples whose tests are complete are displayed in Green color.
7-14
Operator Manual ELAN 30s
8. Test Programming
This screen is used for programming or entering the test parameters for the tests
to be performed by the analyzer. This is performed through the Test menu
available on the main screen.
The test programming menus consist of the following sub menus:
New Test
To create/program a new test into the system.
Open
To view/edit the tests, those are already programmed into the system.
Profile:
To view an already existing profile/create a new profile.
8-1
Operator Manual ELAN 30s
8.2. Open
On clicking this option from the Tests menu, the following screen will be displayed:
8-2
Operator Manual ELAN 30s
3. Click on Modify.
4. Enter the required details and click on Save to save the details.
8-3
Operator Manual ELAN 30s
The Test for which Kit Details are not saved will appear in
red color indicating Incomplete status. After saving Kit
Details, the color of test will change depending on type of
test i.e. Green for Qualitative, Blue for Quantitative and
Pink for Semi-Quantitative.
8-4
Operator Manual ELAN 30s
You can add the new tests in the kit through the Profiles screen
It should be noted that the Mfg. Date should be less than Expiry
Date.
If you select Test Type as Qualitative for Qualitative tests, the
units will be O.D. (by default).
8-5
Operator Manual ELAN 30s
On clicking, the test parameters will appear with all the parameter Blank or
Pre-Set.
2. Click on Modify.
8-6
Operator Manual ELAN 30s
Wavelength
Primary and Reference: It is used to enter the Primary (compulsory) and
Reference wavelength (if required). The available wavelength are 405,
450, 492, 630 (Optional three wavelength are 550nm, 595nm, 690nm )
ORR
Normally, plates are read first with the primary filter, then with the
secondary filter, and afterwards, final readings are calculated subtracting
the reading obtained with the secondary filter from the reading obtained
with the primary filter. For few parameters the third filter is used: that
filter is referred as the ORR filter.
First, the read operation is performed with primary filter. If available, the
dual (differential) wavelength measurement is performed with the
secondary (reference) filter. Now, suppose it is known to the
manufacturer of the kit that the readings of the particular standard is
excessively high, while reading with the primary filter and those
excessively high values must be rectified using ORR filter (If the reader
8-7
Operator Manual ELAN 30s
8-8
Operator Manual ELAN 30s
User then has to enter the volume of the reagent required per test in the
Vol. /Test (µl) field. Select the bottle type as required from the Bottle Type
field. User can also define reagent volume for a blank well by selecting
Disp. in Blank and for a control by selecting Disp.in Control. The reagent
volume to be dispensed has to be entered in Blank Disp. Vol. and must be
within the range 25 and 200µl.
Click ADD button to add a reagent.
Washing: In this section, washing specification will be entered by the
operator.
You are also able to program the No. of Wash cycles (Maximum 8 cycles,
from 3 to 10), Strip Type, Wash Type, and Soak Time (in seconds).
By default, the soak time is set to 30 seconds
8-9
Operator Manual ELAN 30s
2. Click on Modify.
On clicking, the screen will be displayed as below.
8-10
Operator Manual ELAN 30s
3. The screen displays the test parameters details of the selected test with the
defined test sequence. Click Modify to modify the test sequence. To clear
the test sequence, click on Clear All.
8-11
Operator Manual ELAN 30s
2. Now click on the parameter against the selected test sequence. It can be
Sample or Reagent.
3. On clicking the required sequence will be added and displayed in the Test
Sequence section. See figure below.
8-12
Operator Manual ELAN 30s
During the execution of the test, this sequence is performed from top to bottom.
To delete a programmed process, right click on the required sequence, and select
Delete.
In case, if you want to replace or change a specific process, right click on the
required sequence and choose Clear.
Once a test parameter has been used it cannot be reused and is disabled
immediately after selection except Read, Washing, and Incubation.
8.2.4. Calculations
This module is used to interpret the results. To do this, use the following steps:
1. Click on the Calculations.
2. Click on the required test. If we select Quantitative test e.g. TSH EM.
8-13
Operator Manual ELAN 30s
8-14
Operator Manual ELAN 30s
4. Define the Calibration Curve; enter the values and interpretation details as
per the instruction provided in kit literature.
5. Enter your remarks for the test, which will appear in the final report.
6. Click on Save.
7. Else, if you select Qualitative test e.g. HIV EM and click on Modify. The
following screen will be displayed.
8. Enter the Cut off Formula, Gray Zone % and Cut off Sides details.
9. Select required Interpretation and enter Remarks.
10. Click on Save button to save the details.
11. If you select the Semi-Quantitative test in test parameters then following
screen will be displayed;
8-15
Operator Manual ELAN 30s
12. Click Modify, select the Calibration Curve, enter the Index formula and
Values with their corresponding Interpretation and Remarks.
13. Click Save to save details.
14. If CUSTOM UNIT CONVERT option is selected then screen will displayed
as below;
8-16
Operator Manual ELAN 30s
15. Click Modify, select the Calibration Curve, enter the Index Formula,
Result and Values with their corresponding Interpretation and Remarks.
16. Click Save to save details.
17. Alternatively on clicking Modify, if USING STD. CURVE option is selected
then the screen will be displayed as below;
18. Select the Calibration Curve, enter the Index Formula, Values with their
corresponding Interpretation and Remarks.
19. Click Save to save details.
8.2.5. Calibration
This module is used to see the previous calibration data and curve details in offline
mode. To view this, use the following steps:
1. Click on Calibration button.
2. Click on the required test, e.g. TSH EM.
8-17
Operator Manual ELAN 30s
3. Click on Modify.
On clicking, the following screen will be displayed.
8-18
Operator Manual ELAN 30s
4. Select the respective calibration curve and click Save button. Following
message will be displayed. Click OK.
5. The selected calibration curve will be used for result calculation of patient
sample and controls.
8-19
Operator Manual ELAN 30s
8-20
Operator Manual ELAN 30s
8.3. Initialize
This button will save the default parameters. When the test parameters are edited
as per test requirement, you can use this button to save the default parameters.
8.4. Print
This button is used to print the required test parameter. To print, click on the Print
button, and then click on the required test name.
8.5. Delete
This button is used to delete the required test. To delete, click on the Delete
button, and then click on the test name to be deleted.
Now enter the new test name, the test name should be strictly less than 10
characters and click OK. This will create a new test of same type with the copied
parameters and is displayed on the screen.
To create a new test of specific type (Quantitative, Qualitative,
Semi-Quantitative) copy/select a test of same type.
8.7. Download
This button will download and save the test parameters details into the xml file at
location C:\Elan30s\DOWNLOAD PARAMETERS. To do this, click on the
Download button, and then click on the required test.
8.8. Upload
This button will upload test by uploading test parameters details xml file. To do
this, click on the Upload button, and browse for the required xml file. To upload a
test, you need to first download a test by clicking Download button and then select
the required test. This will download test parameters details xml file at location
8-21
Operator Manual ELAN 30s
8.9. Exit
This button is used to exit the Patient screen
8-22
Operator Manual ELAN 30s
9. WorkBench
Before the Start of the RUN user should make sure that the Waste
Can and Tip Bin are emptied. The waste should be disposed as per
the Local Regulatory rules for waste disposal. During RUN, if waste
gets filled up then the RUN will be terminated to prevent
overflowing of waste.
This screen is used to schedule the run. On clicking this screen, it will display the
entire test that is available in the Test > Open. The group of test that operator
plans to perform in the current batch will be selected from this screen.
The WorkBench section is consisting of following section:
Available test
Selected test
Batch ID
See figure below.
9-1
Operator Manual ELAN 30s
In this screen you have option for selecting Standards and External Controls, Tips
for probes, Wash Fluids for the test Run in the current Batch. By default all
standards are selected. You can choose the tip for the Sample probe by selecting
the Sample check box under Tips For Probe label. By default wash bottle B1
filled with Distilled water is selected for Probe wash during the Run. User can
select the Buffer Bottle (From B1 to B3) in which the required Buffer solution is
stored. If the bottle is already selected for another test, a message will be
displayed.
9-2
Operator Manual ELAN 30s
Click Yes if you want to use the same wash solution. Select No for using different
wash solution and place the bottle at remaining position. Then click OK button to
save the settings.
It also displays the number reagents for the selected test. By default, large bottles
are used to place the reagents on tray but if you want to keep particular reagent in
a sample tube, then select Sample Tube option.
If you select the Mix Ctrl in Internal & External control details in test parameter s
following message will be displayed for min control volume of 500ul
9-3
Operator Manual ELAN 30s
If you are using Sample Tube, then before starting the run, you
must calibrate the sample tube for dead volume.
9-4
Operator Manual ELAN 30s
If you want to run any Quantitative test for the first time, then it is
mandatory to run the standards. If this is not done i.e. Standards
check box is not ticked in WorkBench screen, then the message
“This test was not calibrated, Please select All Standards” will be
displayed and the run will not proceed.
9-5
Operator Manual ELAN 30s
You can remove a selected test from the Selected Test section, by right clicking
and select Delete option. After the User selects delete option that test will be
removed.
9-6
Operator Manual ELAN 30s
In case, during the selection of tests in the WorkBench module, the number of
selected tests exceeds six, a message, indicating the same will pop up.
Once the required tests are selected and you click on Save button, a message
box will be displayed. Select Yes if you want to define common reagent for
different tests, else No, as appropriate.
See section 9.6 Common Reagent for more details.
9.3. Batch ID
In this section, the Batch ID for the WorkBench is displayed. The software
automatically generates the Batch ID for the saved Batch. This Batch ID is unique
for a Batch.
You can any time “Exit” the WorkBench module by clicking Exit button. The Exit
option brings back the User to the main menu.
Save option on the WorkBench brings the control on the Reagent layout screen.
9-7
Operator Manual ELAN 30s
Click on Fluidic Check button to start the fluidic check. You will be prompted to
put the SRGT cover in position before placing reagent/sample/control in position.
After placing SRGT cover in position, click OK. Following message will be
displayed.
9-8
Operator Manual ELAN 30s
You need to place the special Fluidic Check test tube at Sample position 58, 1.5 ml
of Probe Clean solution at position 60.
At the beginning of every RUN user needs to dispose of the old solutions at
Position 60 and fill the test tube with fresh solution. In case if test tube is not
placed at Position 58 or it is not emptied, then an error message will be displayed
and will prompt the user to place an empty Special Fluidic Check test tube.
The user will be also prompted to check the aspiration and dispensing of Washer,
user need to check the level of the fluids in all wells. Volume should be same in all
wells at Washer Priming position.
9-9
Operator Manual ELAN 30s
After completion of Fluidic Check the user needs to remove the Special Fluidic
Check test tube and place the respective sample.
9-10
Operator Manual ELAN 30s
Click on Place ST Cover button to move the Robotic Arm towards SRGT tray to
place the Strip Tray cover. Similarly, click on Place SRGT Cover button to move
the Robotic Arm towards Strip Tray to place the SRGT cover.
This screen gives the description about the following:
9-11
Operator Manual ELAN 30s
This sheet gives the details about the Reagents, Standards, and Controls used for
that test.
Reagent details are displayed in the following column:
Position: Displays the reagent position number on tray.
Reagent Name: Displays the selected reagent name for the test.
Bottle Type: Displays the bottle type to be used for placing the
reagents.
Reagent Volume: Displays the required reagent volume in ml including
dead volume.
Use Sample Tube: Displays Yes or No, indicating the bottle type (Small
or Large) to be used for the reagents. It displays “Yes” if the option
Sample Tube option is selected for the particular reagent. See page 9-3
for details.
Reagents are depicted by following colors in Test Requirement Sheet and SRGT
tray layout as shown in above screen:
9-12
Operator Manual ELAN 30s
9-13
Operator Manual ELAN 30s
You need to place the Tips starting from the position close to the sample position
number 1 (this is indicated by a dark line) on SRGT tray near the tip starting
position. Number of Tips required are 10 more than the actual requirement of tips.
This is required to account for the Tips that are disposed off; in case of an error
Strips.
On clicking OK, the run will proceed.
If tip is already present on the tip position 1, it will be asked to
remove the tip which is indicated by a following message.
You can click on the Click button next to the strip which will bring Strip Tray on the
front side that is at its loading position. You must place the strip carefully such that
the strip rests properly on the Tray. In case if all 8 wells are not be used then you
can break the strip (minimum 2 wells) and place a dummy broken strip and should
9-14
Operator Manual ELAN 30s
have the remaining wells (dummy strip should be of same size as of the actual
strip).
For example, if only three wells are required then user needs to break the actual
strip which 3 wells and place a dummy strip of same type that has 5 wells. It
should be noted that whenever a broken strip is used then it should have at least
one side of the strip (i.e. holding position) intact.
If you are using the broken strip, then make sure all the wells (New wells and
dummy wells) should be at a same level. If all wells are not in the same level,
then washer will not wash the wells properly and there are chances that
Washer will leave high residual volume which may affect the result.
9-15
Operator Manual ELAN 30s
One well should not be used on the tray positions. There should be at least 2
wells (along with 6 dummy wells).
Washer is 8 needle manifold and during washing process, it will wash all 8
wells simultaneously. In case, if number of wells are less than 8, then for any
of the strips, ensure that the dummy wells (preserved or used strips, which
can be used as dummy wells when patient samples are not in multiples of 8)
are placed in the empty position. If this is not done, then this may cause
spillage and may damage the instrument.
In case, during the batch run, if Waste can is full to its capacity,
then batch run will be terminated automatically.
On clicking the Run button message will be displayed for asking the user to empty
the waste bottle.
9-16
Operator Manual ELAN 30s
When you click on the Start button, he will be prompted to clean the Tip bin, in
case, if Tips are going to be used for the Test.
You need to remove the Tip bin, clean it, and place it again into its location. A
buzzer will be sounded when Tip Bin is placed in its position. Refer section 9.5.6
Placing Tips.
A level scan will be performed which will check the adequacy of Reagent and
controls/Calibrators / Standards required for the Test, it will also scan for the
number of tips placed on SRGT.
9-17
Operator Manual ELAN 30s
Also, if the numbers of tips are not adequate for the RUN, a message box will pop
up indicating the error.
If room temperature is out of range following error will occur. Click Yes to continue
or No to cancel the run.
On clicking Yes, AGC and Gain during run is checked. After completing the
checking of AGC and Gain during run, user will be prompted to perform strip
checking of the Strip Tray.
9-18
Operator Manual ELAN 30s
On clicking OK, you will be directed back to Run screen. Replace the dirty strips
(wells) with new strips. The analyzer will perform all the above steps again and ask
the user to perform strip checking. Click No to proceed with the Run.
On completion of Run, Batch Completed message is displayed.
9-19
Operator Manual ELAN 30s
9-20
Operator Manual ELAN 30s
2. Click on any particular test to display the results with calibration curve for
that particular test.
9-21
Operator Manual ELAN 30s
3. Click View Print button. The results can be viewed/printed with two options:
a. PrintTablewiseResults
b. PrintStripwiseRawData
You can also view the TableWise and StripWise report using
Search options after completing the batch run. Refer section 11
Search for more details.
4. By default, calibration curve is saved automatically with default test name.
The saved calibration curve can be viewed clicking Curve Calibration
button. Select check box to display the curve.
9-22
Operator Manual ELAN 30s
5. The curve can also be saved manually with another name. Click on
Memorise Curve button. Give the desired name to the curve and click OK.
9-23
Operator Manual ELAN 30s
7. The new curve name will now be displayed in Memorized Curves grid.
9-24
Operator Manual ELAN 30s
9-25
Operator Manual ELAN 30s
9-26
Operator Manual ELAN 30s
1. After the completion of tests, click on View Results button. In this case,
only O.D. is displayed and no data is displayed in Standard Details, Conc,
and Result field in the grid.
9-27
Operator Manual ELAN 30s
3. Select any calibration curve from the saved calibration curves for that
particular test. Click Using Selected Calibration button. Following
message will appear. Click OK.
9-28
Operator Manual ELAN 30s
9-29
Operator Manual ELAN 30s
9-30
Operator Manual ELAN 30s
Click View Print button. In this case (results calculated using earlier
saved calibration), PrintStripwiseRawData option is disabled.
9-31
Operator Manual ELAN 30s
9-32
Operator Manual ELAN 30s
9-33
Operator Manual ELAN 30s
Common
Reagents with
common position
Enter the reagent position manually in the RgtPosition column for those tests
having the common reagents. As shown in above screen, same reagent positions
are defined 1, 2, 3 for the tests T3 EM and TSH EM for common reagents.
After entering the positions click Save. This will displayed the Reagent Layout
screen as follows:
9-34
Operator Manual ELAN 30s
Common Reagents
Common
reagent with
common position
9-35
Operator Manual ELAN 30s
Common Reagents
Make sure that the reagent name should be same for defining the common
reagent. In case, if it not same, a error message will be displayed. See the
following screen.
9-36
Operator Manual ELAN 30s
9-37
Operator Manual ELAN 30s
5. Run:
Here you can select the Batch and perform Run.
All validations are checked before going to Run a Batch.
Validation to be checked:
a. Initialize robotic arm
b. Checking Bottle placement
c. Checking tips quantity
d. Initialize photo meter
e. Checking strip placements
f. Checking wash fluids
g. Checking Reagent volumes
If there are no errors then analyzer will continue with the Run.
9-38
Operator Manual ELAN 30s
10. Results
The Result screen available on the main screen is used for viewing and printing
the patient’s result. On clicking the Report, the following screen will be displayed.
There are four different options available to generate the patient result. They are:
Test wise Results
Batch wise Results
Sample wise Results
Batch wise Results with Date
The options are explained below:
10-1
Operator Manual ELAN 30s
Figure 10-1
The following report will be generated on clicking Print Results.
10-2
Operator Manual ELAN 30s
Figure 10-2
Click the Export Report button to save the report in *.rpt, pdf, xls, doc or rtf format.
10-3
Operator Manual ELAN 30s
Figure 10-3
10-4
Operator Manual ELAN 30s
Figure 10-4
The following report will be generated on clicking Print Results
Figure 10-5
10-5
Operator Manual ELAN 30s
Figure 10-6
The following report will be generated on clicking Print Results.
10-6
Operator Manual ELAN 30s
Figure 10-7
10-7
Operator Manual ELAN 30s
10-8
Operator Manual ELAN 30s
11. Search
This screen is used to search the Patient and Sample details. Three options are
available for selection:
Patient /Sample
Test Details
Doctor
TableWise Report
StripWise Report
To open this screen, click on the Search from the main screen. The following
window will be displayed:
11-1
Operator Manual ELAN 30s
3. Select the check box to select the From Date and To Date, if required.
4. Then click View Result.
On clicking, the details will be displayed in the grid.
11-2
Operator Manual ELAN 30s
11-3
Operator Manual ELAN 30s
If you select the test from the Select Test fields, then only the
records which are related with that Test will be searched and
displayed.
If you select the dates from the From Date and To Date fields,
then the records between those two dates will be searched and
displayed.
If you select the tests along with From Date and To Date, then the
records of that test between those two dates will be displayed.
11.3. Doctor
This option is used to search the record of all patients referred by the doctor.
Use the following step to search the details:
1. Select the option Doctor.
2. Select both checkboxes if required, and enter the name of the doctor in
Enter Doctor Name and select the dates in From Date and To Date text
box.
11-4
Operator Manual ELAN 30s
If you only select Doctor then the records of all Patients referred
by that Doctor will be searched and displayed.
If you select only From Date and To Date then the records of
Patients between those dates will be searched and displayed.
If you select Doctor with From Date and To Date then the records
of Patients referred by that particular Doctor between those dates
will be displayed.
11-5
Operator Manual ELAN 30s
11-6
Operator Manual ELAN 30s
11-7
Operator Manual ELAN 30s
11-8
Operator Manual ELAN 30s
11-9
Operator Manual ELAN 30s
11-10
Operator Manual ELAN 30s
11-11
Operator Manual ELAN 30s
11-12
Operator Manual ELAN 30s
The Quality Control screen is useful for viewing the results in Graphical format.
The user should either rerun the controls again or recalibrate the test and run the
controls.
Following are the steps to view the results and chart:
1. Select required test from the Test dropdown list.
2. Select the option By OD/Conc. or By Index.
12-1
Operator Manual ELAN 30s
12-2
Operator Manual ELAN 30s
13. Maintenance
This section provides the necessary and minimal maintenance procedures in order
to ensure that the analyzer operates correctly and provides the accurate
measurement results.
It is recommended that the analyzer be maintained by the user with proper
instructions from the service personal to get the optimum output and to avoid any
possible damage to the analyzer.
Voltages present in the ELAN 30s can produce severe
electrical shocks. Place the analyzer and the computer in the
OFF (0) position and disconnect all line chords before
attempting any maintenance procedure.
13-1
Operator Manual ELAN 30s
There are various types of Maintenance operations. Some of them are very easy
and may be performed by the OPERATOR. Others, however, should only be
carried out by the authorized QUALIFIED TECHNICIAN only.
Among the operations that are carried out by the OPERATOR are those of
checking, cleaning, and decontamination of the equipment.
13-2
Operator Manual ELAN 30s
13-3
Operator Manual ELAN 30s
Average use life of water filter is 3 months. Replacement of water filter depends on
quality of DI water used.
13-4
Operator Manual ELAN 30s
13-5
Operator Manual ELAN 30s
13.3.1. Routine
Routine tab is used for performing analyzer’s routine maintenance. To open this
screen, Click on the Maintenance menu, and then click on Routine tab.
13-6
Operator Manual ELAN 30s
13-7
Operator Manual ELAN 30s
Auto Shutdown: This command is used to perform the auto shutdown activities
such as probe cleaning, washer cleaning and software shutdown.
On clicking Auto Shutdown option following message is displayed.
13-8
Operator Manual ELAN 30s
Follow the on screen instructions to flush out water from all tubing before
transportation of analyzer.
Washer Decontamination: Following screen is displayed.
13-9
Operator Manual ELAN 30s
13.3.2. Records
This tab is used to maintain the records. You can save the Date of change and
write the comments or reason for Probe, Lamp, Plunger, Syringe, Filter, or any
other Change. User can view the previously saved records by using the Display
option.
The following screen will be displayed on clicking Maintenance > Records.
13-10
Operator Manual ELAN 30s
13.3.3. Backup
The backup tab is used to take the backup of the installed database. The following
screen will be displayed on clicking Maintenance > Backup.
13-11
Operator Manual ELAN 30s
13-12
Operator Manual ELAN 30s
Administrator of the system can create, edit, and delete the user ids as required,
and also provide access rights to new and existing users.
13-13
Operator Manual ELAN 30s
13-14
Operator Manual ELAN 30s
3. Now, you can enter the new details in the required fields.
4. After entering the details, click on SAVE to save the settings.
3. Click OK.
This will delete the user ID from the grid.
13.3.5. Configuration
The Configuration screen allows you to configure the behavior of the analyzer
hardware and software. This screen is grouped into three parameters:
ASTM settings
Report Header Settings
Reader Settings
13-15
Operator Manual ELAN 30s
System Settings
To open this screen, click on the Maintenance > Configuration.
Click on Settings.
A new window named Settings will open which is used to establish
communication between the analyzer and the host.
The host communicates with LIMS (Interfacing Laboratory Information
Management System) only from the PC where the ELAN 30s software is installed.
Separate serial port is required to enable the communication through serial cable.
13-16
Operator Manual ELAN 30s
13-17
Operator Manual ELAN 30s
When Perform Strip Checking option is checked, during batch run, system will
perform strip checking without asking user confirmation. See page number 9-19.
While performing strip check, if the detected OD <=0.01 then it is considered as
no/missed well and if the OD>=3 then the well is considered to be dirty.
NOTE: ‘Strip Checking’ option is enabled by default..
When Auto Startup option is selected, analyzer will perform the auto startup
activities after initializing.
13-18
Operator Manual ELAN 30s
This option is used to set the tip eject mechanism. Two types of eject mechanism
are provided.
At Arm
At Eject Location
By default, option At Arm is selected. This option is factory set and can be
changed depending upon the tip eject mechanism.
13-19
Operator Manual ELAN 30s
13-20
Operator Manual ELAN 30s
3. If selected tests are exceed more than one following message will be
displayed
4. Select the single test & click Save. This will displayed the reagent Layout
screen as shown below:
13-21
Operator Manual ELAN 30s
3. Click OK.
4. On clicking, the captions on the screen will appear in the selected
language.
13.3.6. Span
This screen is used to check the gain of the photometer for 405,450, 492 and 630
wavelengths. The range is 1.0 to 2.0. Click on Start button to start. ON clicking
Show Log button, the log details will be displayed on the screen.
If the gain obtained for any wavelength is not within the range, it is highlighted with
red background else if within range then with green background.
The following screen will be displayed on clicking Maintenance > Span > Show
Log.
13-22
Operator Manual ELAN 30s
13-23
Operator Manual ELAN 30s
The date can be selected by changing the date from the dropdown list by
selecting the checkbox provided.
You can select the operation (Run, Service, Maintenance) during which
the errors occurred.
Select the batch id by selecting the BatchId dropdown list.
Once the above selection is done, click on Show button. The error
messages will be displayed according to the operation selected.
In the grid following are the different column present:
ErrorCode – This column displays the error code number to identify the
type of error occurred.
ErrorMessage – This column displays the error code description.
ErrorDateAndTime – This column displays the date and time of the
occurrence of error.
BatchId – This column displays the batch id in which the error occurred.
Refer section 14 Error Codes list for error description and there corrective actions in
more details.
13-24
Operator Manual ELAN 30s
13-25
Operator Manual ELAN 30s
13-26
Operator Manual ELAN 30s
f. In case if there is no suction after dispensing then increase the time and
incase if there is a high amount of liquid being sucked back then
decrease the time and click on save.
g. Repeat step 4 – 6 till a point such that all the wells are filled properly and
min amount of liquid is sucked back by the aspirating needles.
13-27
Operator Manual ELAN 30s
13-28
Operator Manual ELAN 30s
a. Place an empty clean strip at position 31, this strip should be of the type
for which calibration is to be done.
b. Click on start.
c. The washer will perform a wash cycle and will ask to check the residual
volume.
d. If residual volume is Ok (as per the volume specified in the message
box) then click on Yes.
e. After this user will be asked whether he wants to, save the calibration,
click Yes to save and No to discard.
f. In case if residual volume is not OK then click on No when it asks to
check the residual volume in step 3.
g. Repeat the above steps till the residual volume is obtained.
This calibration is to be performed for all the strips that are going to be used
in the analyzer.
13-29
Operator Manual ELAN 30s
13-30
Operator Manual ELAN 30s
14-1
Operator Manual ELAN 30s
Errors Related to ST
40 Not Initialized
41 Positional error
42 Probe Not at Home (Interlock)
43 Washer Not at Home (Interlock)
Misc Errors
80 Tip Exhausted
81 Unable to Dispose Tip
82 Proper Tip Fixing Failed
SS - Strip Number
Value Possible Cause
Strip Number 1 01
Strip Number 30 30
14-2
Operator Manual ELAN 30s
W - Well Number
Value Possible Cause
Well Number 1 01
Well Number 8 08
14-3
Operator Manual ELAN 30s
14-4
Operator Manual ELAN 30s
Obstruction Detect
error) then please
call service
engineer.
9 322SSW !S Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Sample aspirating 2. If there is no
aspiration Sample obstruction but still
it’s detecting VOD
(Vertical
Obstruction Detect
error) then please
call service
engineer.
10 422SSW R! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Reagent dispensing 2. If there is no
dispense Reagent obstruction but still
it’s detecting VOD
(Vertical
Obstruction Detect
error) then please
call service
engineer.
11 422SSW C! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Standards/Co dispensing 2. If there is no
ntrols Sample obstruction but still
dispensation it’s detecting VOD
(Vertical
Obstruction Detect
error) then please
call service
engineer.
12 422SSW S! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Sample dispensing 2. If there is no
dispense Sample obstruction but still
it’s detecting VOD
(Vertical
Obstruction Detect
error) then please
call service
engineer.
13 1522SSW PW! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Probe Wash priming 2. If there is no
obstruction but still
it’s detecting VOD
14-5
Operator Manual ELAN 30s
(Vertical
Obstruction Detect
error) then please
call service
engineer.
14 1622XXX Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
liquid level liquid level check 2. If there is no
check obstruction but still
it’s detecting VOD
(Vertical
Obstruction Detect
error) then please
call service
engineer.
15 280SSW !T Error during This error may 1. If there are
281SSW Tip Fixation get reported sufficient Tips and
when Tip is used still this error was
282SSW for Sample / reported then
Control / Reagent service engineer
dispensing needs to be
consulted.
16 581SSW T! Error during Error when the 1. Open the Robotic
Tip eject Tip is not arm cover section
properly ejected and check for any
after dispensing obstructions for
of Sample / Probe movement.
Control / Reagent 2. If no obstructions
is observed then
service engineer
needs to be
consulted.
17 422SSW S! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Sample dispensing 2. If there is no
dispense Sample obstruction but still
it is detecting VOD
(Vertical
Obstruction Detect
error) the please
call service
engineer.
18 422SSW R! Vertical Obstruction 1. Check for any
obstruction detected when obstructions to
error during probe goes for Probe movement.
Reagent dispensing 2. If there is no
dispense Reagent obstruction but still
it is detecting VOD
(Vertical
Obstruction Detect
error) then service
engineer needs to
14-6
Operator Manual ELAN 30s
be consulted.
14-7
Operator Manual ELAN 30s
14-8
Operator Manual ELAN 30s
14.2.2. SRGT
Serial Error Error
Flag Possible Failure Corrective Actions
Number Code Message
1 130SSW @SRGT SRGT This error is 1. Check for any
230SSW Initialization generated if obstructions in
Error SRGT is not SRGT movement.
330SSW able to 2. If there is no
430SSW initialize obstruction but
530SSW still its generating
630SSW this error then call
service engineer.
730SSW
830SSW
930SSW
14-9
Operator Manual ELAN 30s
1530SSW
1630SSW
2 831SSW @SRGT SRGT Error 1. Check for any
931SSW Rotational generated if obstructions in
Error during required SRGT movement.
1231SSW aspiration position opto 2. If there is no
slots are not obstruction but
detected still its generating
when SRGT this error then call
moves for service engineer.
aspiration
3 332SSW @PSRGT SRGT Error 1. Check for any
832SSW Rotational generated in obstructions to
Error during case if probe Probe movement
932SSW aspiration has not 2. If there is no
1232SSW when probe initialized and obstruction but
is not at SRGT gets a still its generating
HOME command to this error then
move. This is service engineer
an interlock needs to be
used to consulted
prevent any
damage to
probe
4 1631XXX @SRGT SRGT Error 1. Check for any
Rotational generated if obstructions to
Error during required SRGT movement
LONGREAD position opto 2. If there is no
slots are not obstruction but
detected still its generating
when SRGT this error then
moves for service engineer
Liquid Level needs to be
Check consulted
14-10
Operator Manual ELAN 30s
14.2.4. Heater
Serial Error Error
Flag Possible Failure Corrective Actions
Number Code Message
1 2000001 @TMP Heater Cancel Batch and
Communicat report the Results
ion Error with the Flag.
2 2000002 @TMP Heater Cancel Batch and
Sensor Error report the Results
with the Flag.
14.2.5. Washer
Serial Error Error
Flag Possible Failure Corrective Actions
Number Code Message
1 850SSX @W Washer This error is 1. Check for any
950SSX Initialization generated if Washer obstructions to
Error (can't is not able to Washer
1150SSX get to home initialize movement
1250SSX position) 2. If there is no
obstruction but
still its generating
this error then
14-11
Operator Manual ELAN 30s
service engineer
needs to be
consulted
14.2.6. Photometer
Serial Error Error
Flag Possible Failure Corrective Actions
Number Code Message
1 860SSX @PH Photometer This error is 1. Check for any
960SSX Initialization generated if obstructions to
Error Photometer is Photometer
1260SSX not able to movement
1860SSX initialize 2. If there is no
obstruction but
still its generating
this error then
service engineer
needs to be
consulted
2 861SSX @PH Photometer Error 1. Check for any
961SSX Linear generated if obstructions to
Movement required Photometer
1261SSX Error position opto movement
1861SSX slots are not 2. If there is no
detected obstruction but
when still its generating
Photometer this error then
moves for service engineer
reading needs to be
consulted
3 ?? L-FAIL Photometer --
Lamp OFF
during
Photometer
Reading.
14.2.7. Syringe
Serial Error Error
Flag Possible Failure Corrective Actions
Number Code Message
1 670SSW @SYG Syringe This error is Call service engineer.
770SSW Initialization generated if
Error Syringe is not
870SSW Initialization able to
970SSW Error initialize
1570SSW
1670SSW
14-12
Operator Manual ELAN 30s
14-13
Operator Manual ELAN 30s
14-14
Operator Manual ELAN 30s
10 High-OOR This flag is displayed when the O.D. is out of range i.e. >3
15-1
Operator Manual ELAN 30s
15-2
Operator Manual ELAN 30s
16. Appendix
16.1. TBM Image Installation
This section explains you the procedure of installing the TBM image (Windows
Embedded 7 Operating System) into the analyzer PC.
The analyzer PC comes with preinstalled TBM image.
For any reason, if the TBM image is not installed or you want to upgrade the
existing operating system (OS) into TBM image, please read the following
information carefully for installing the TBM image to analyzer PC.
16.1.1. Scenario
There could be two scenarios under installing the TBM image:
New (Clean Installation)
Upgrade/ Reinstall
16.1.1.2. Upgrade/Re-installation
It is kind of an upgrade installation in which the analyzer PC already has some
operating system (for example: previous TBM image) installed. In this case, the
TBM image setup will only format the C: drive and install the TBM image on the
same. This type of installation must be used if you want to install TBM image
on C: drive but want to keep the data on other drives untouched.
16-1
Operator Manual ELAN 30s
1. Restart the PC and continuously press F12 (or key that is mentioned on the
screen to go to boot menu) to go to boot menu. Select system setup in the
boot menu to find all the options for the system setup.
2. Find the boot sequence option in the system setup and make sure that the
sequence of boot for CD-ROM is higher than hard drive.
3. If any changes are made to the existing sequence, click Apply to save the
changes and restart the PC.
Use the following steps to install the TBM image available on DVD to the analyzer
PC:
1. Insert the DVD containing TBM image setup into the DVD drive and restart
the PC.
2. It will show a message “Press any key to boot from CD or DVD”. As
suggested, press any key so as to boot from the DVD we entered.
3. Now, the PC will boot from the DVD containing TBM image. It will show the
Loading Windows... progress bar message on the screen.
4. After loading windows, wait till it shows command prompt window i.e.
X:\Windows\System32>.
5. Now, you need to get to the DVD and existing OS drive letter so that we can
start the TBM setup.
(Refer section 16.1.3 Identifying Various Drives for details on finding the
DVD and OS drive letter on the analyzer PC. Note down DVD drive letter
and Drive on which you want to install TBM image of OS).
6. Type the appropriate DVD drive letter followed by colon ‘:’ to enter into the
DVD drive.
For example, if E is the DVD drive letter then type the following command
and press Enter key.
X:\Windows\System32> E:
7. Now, type the following command in the command prompt window and press
Enter key.
TBMSetupWithClean
This will delete all the data from the PC.
16-2
Operator Manual ELAN 30s
Type y for yes to continue with the installation or type n to cancel the
installation.
On clicking n, the installation will be canceled and the command prompt will
be displayed.
8. After confirmation, it will clean the entire hard disk and then create two
partitions i.e. C and D drive with C drive size as 35 GB approximately, and
the remaining as D drive. It will then start applying the TBM image on the
PC. This process may take few minutes depending on the configuration of
PC.
9. Once the installation of TBM image is completed, a confirmation message
for the same will be provided on the screen and then the target computer will
restart automatically as shown below.
16-3
Operator Manual ELAN 30s
10. After restart, it might again ask to boot from DVD. Do not press any key and
let the PC continue to load the windows automatically. It is ok to take the
boot DVD out.
11. It will apply the required settings and may boot again.
12. Finally, it will ask to enter the product key. Enter the product key and click
Next button.
13. Follow and accept the remaining instructions to complete the TBM image
setup on the target computer.
TBM setup image contains driver for specific hardware
configuration, especially for Network Interface and Display. If
the analyzer PC has different hardware components, you may
need to install the corresponding drivers from the PC’s
motherboard utility CD.
16-4
Operator Manual ELAN 30s
7. Now, type the following command in the command prompt window and
press Enter key.
TBMSetup
A message will be asked to type the OS drive letter followed by colon.
For example, if C is the OS drive letter then type C: in the command
prompt and press Enter key. If you do not want to continue with the
installation then type exit and press Enter key.
After drive letter is entered in the command prompt, it will ask to confirm
the TBM image installation on the specified drive, type y for yes in the
command prompt to continue with the installation. If you do not want to
continue with the installation type n to exit and return to command prompt.
The same has been shown in the figure.
8. This will format the drive that was provided in step above and then start
installing the TBM image on the PC. This process may take few minutes
depending on the configuration of analyzer PC.
9. Once the installation of TBM image is completed, a confirmation message
for the same will be provided on the screen and then the analyzer PC will
restart automatically as shown below.
16-5
Operator Manual ELAN 30s
10. It might again ask to boot from DVD. Do not press any key and let the PC
continue to load the windows automatically. It is ok to take the boot DVD
out.
11. It will apply the required settings and may boot again.
12. Finally, it will ask to enter the product key. Enter the product key and click
Next button.
13. Follow and accept the remaining instructions to complete the TBM image
setup on the target computer.
In the upgrade installation scenario, if the previous
installation is Windows 7, it is possible that after
installing TBM image, it will show one extra volume
which was the hidden partition in previously installed
Windows 7 OS. The typical size of this partition or
volume is normally 100 to 200 MB.
TBM setup image contains driver for specific hardware
configuration, especially for Network Interface and
Display. If the analyzer PC has different hardware
components, you may need to install the corresponding
drivers from the PC’s motherboard utility CD.
16-6
Operator Manual ELAN 30s
6. Cancel and Close the Notepad window after confirming the DVD
drive letter.
16.1.3.2. Finding the Existing Operating System (OS) Drive Letter
Use the following steps to find the drive letter pertaining to DVD drive. You can
skip the steps (1 to 3) if already done:
16-7
Operator Manual ELAN 30s
16-8
Operator Manual ELAN 30s
Cancel and Close the Notepad window after confirming the DVD drive letter.
16-9
Operator Manual ELAN 30s
16-10
Operator Manual ELAN 30s
Addendum release.
2012.01.01 1-Jan-2013 Accessories checklist updated with new item SD
code and description for Reagent bottle
2013.01 -- Not used --
2013.02 -- Not used --
2013.03 7-June-2013 Updated as per ELAN30s software version SD
2013.03.
Following are the brief description:
Message to view results after batch
completion is added.
Provision to keep reagents on sample tube
using adaptor
Edit button removed from the Workbench
screen.
After saving the test, extra column will be
displayed that is Use Sample Tube. It
indicates small bottle is selected for the
reagent.
DI water requirement added
Decontamination procedure added
Accessories checklist
Strip fitment instructions
2013.04 16-Dec-2013 Updated as per software version 2013.04 PS
Common Reagent
Accessories list updated
2014.01 09-Jan-2014 Updated as per software version 2014.01 PS
Sample positions (41-48) for reagents
2014.01.06 05-May-2014 Following changes are as per software version PS
2014.01.06
Raw data in maintenance.
On line printing & Analyzer as a reader
mode in configuration.
Tablewise & Stripwise report with date.
17-1
Operator Manual ELAN 30s
17-2
Contact Us
We welcome your feedback about the product. You can reach us at the following.
Transasia Bio-medicals Limited, Transasia House, 8 Chandivali Studio Road, Andheri (East),
Mumbai-400072, India