Professional Documents
Culture Documents
CC 1.01 Basic Principles of Clinical Chemistry
CC 1.01 Basic Principles of Clinical Chemistry
OUTLINE
1. Units of Measure III. REAGENTS
2. Reagents Most instrument manufacturers make the reagents in a ready-to-use
3. Clinical Laboratory Supplies form or “kit” where all necessary reagents and respective storage
4. Centrifugation containers are prepackaged as a unit requiring only the addition of
5. Laboratory Mathematics and Calculations water or buffer to the prepackaged components for reconstitution.
6. Specimen Considerations
A. Chemicals
LEGEND 1. Analytical Reagents
Remember Lecturer Book Previous Presentation Suitable for use in most analytical laboratory procedures
Trans Impurities must be stated
2. Ultrapure Reagents
Chemicals that have been put through additional purification steps
OBJECTIVES: Used in procedures such as chromatography, atomic absorption,
At the end of the lecture, the student should be able to: immunoassays, molecular diagnostics and other techniques that
Convert results from one unit format to another using the SI and may require extremely pure chemicals
traditional systems. 3. Chemically Pure Reagents
Describe the classifications used for reagent grade water. Indicates that impurity limitations are not stated
Identify the varying chemical grades used in reagent preparation and Preparations of these chemicals is not uniform
indicate their correct use. Not recommended for clinical laboratories unless further
Define primary standard and standard reference materials. purification or reagent blank is included
Describe the following terms that are associated with solutions and, 4. USP and NF grade
when appropriate, provide the respective units: percent, molarity, Used to manufacture drugs
normality, molality, saturation, colligative properties, Redox potential,
Based on the criterion of not being injurious to individuals
and conductivity.
Chemicals are pure enough for use in most chemical procedures
Describe two ways to calibrate a pipetting device.
USP – United States Pharmacopeia, NF – National Formulary
Identify the preanalytic variables that can adversely affect laboratory
results as presented in this chapter.
5. Technical/Commercial Grade
Used for manufacturing
I. CLINICAL CHEMISTRY Should never be used in the clinical laboratory
Clinical – comes from the greek word kline, meaning “bed”
Chemistry – is a basic science that utilizes the specialty of IV. REFERENCE MATERIALS
chemistry to study human beings in various stages of health and STANDARD
disease A substance or solution in which the concentration is determined
It is an applied science when analysis are performed on body It is used in the calibration of an instrument of method
fluids or tissue specimens to provide information for the diagnosis
or treatment of disease.
A. Types of Standard – Atomic Weight Standard
II. UNITS OF MEASUREMENT Pure to the last molecule
Système International d'Unités (SI) Grade A (IUPAC)
It consists of seven independent base units, each unit is represented
by a symbol. B. Ultimate Standard
Adopted internationally in 1960 Grade B (IUPAC)
It is devised to provide the global scientific community a uniform
method of describing physical quantities. C. Primary Standard
Highly purified, concentration is exactly known
A. UNITS At least 99.98% purity
Length meter (m) Used in clinical laboratories
Mass Kilogram (kg) Grade C (IUPAC)
Time Second (s)
Quantity of substance Mole (mol) D. Working Standard
Electric current Ampere (A) <0.05% particulates, 99.95% purity
Temperature Kelvin (K) Grade D (IUPAC)
Luminous Intensity Candela (cd)
Table 1.01 (Units of Measurement) E. Secondary Standard
Substance of lower purity
NOTES TO REMEMBER Grade C (IUPAC)
It has been recommended that analytes be reported using moles of
solute/volume of solution (mmol/L) V. WATER SPECIFICATIONS
Enzyme units are given as the international Unit per liter (U/L), WATER
although katal unit has been adopted previously as SI unit, its use is Considered as the universal solvent
limited It is the most frequently used reagent in the laboratory
pH scale is retained for measurement of hydrogen ion concentrations
Liter is used as the reference volume METHODS OF FILTRATION – A. FILTRATION
SI unit is used because compounds react on a molar basis, and Passage through a filter or other material that prevents passage of
expressions of amounts of substances in such terms allows for a certain molecules, particles or substances
better understanding of the relative proportion of compounds. Can remove particulate matter from municipal water supplies before
any additional treatments
CLINICAL CHEMISTRY 2 of 7
MLS109 Basic Principles of Clinical Chemistry LE 1 TRANS 01
CLINICAL CHEMISTRY 3 of 7
MLS109 Basic Principles of Clinical Chemistry LE 1 TRANS 01
A. Calibration
Done every 6 months
Class A pipettes do not require recalibration
GRAVIMETRIC METHOD
Principle is that weight of water delivered by the pipet is equal to
the volume of liquid delivered
Precision
Determined by pipetting 10 aliquots of water, weighing the
aliquots and calculating the mean and the standard deviation.
Serologic Pipette Accuracy
Determined by correcting the mean for temperature (20 –
Mohr Pipette 26C) and comparing the observed value to the expected value
Calibrated between two marks of the stem It should not deviate more than 5% of the stated value
Graduations do not go all the way to the tapered tip
Accuracy is greatest at full volume SPECTROPHOTOMETRIC METHOD
Requires controlled delivery of the solution between calibration Principle is that a colored compound of known absorptivity is
marks. diluted into 10 test tubes using the pipet being checked
The absorbance of each tube is obtained using measurements
from a spectrophotometer
Precision
Determined by calculating the standard deviation of the aliquotes
Must be less than 5%
Accuracy
Determined by comparison to a reference solution
Measuring/Graduted Pipette
B. Centrifugation
X. MICROPIPETTE The process using centrifugal force to separate the lighter portions
Pipettes used for measurements of microliters of volume of a solution, mixture or suspension from heavier portions of a
Piston operated, positive displacement solution
Uses disposable tips and detipers Used to
Remove cellular elements from biological fluids to provide cell-
Blue Tip – 100 to 1000 uL or 1mL free specimen for analysis
Yellow Tip – 2 to 200 uL To concentrate cellular elements for microscopic examination
White Tip – 0.5 to 10 uL RPM
Revolutions/rotations per minute
Speed of the centrifuge
Checked and calibrated using a tachometer
RCF
Relative centrifugal force
Force that is generated by the centrifuge that is required to
XI. QUALITY CONTROL AND PREVENTIVE MAINTENANCE separate substances in a solution
Routine Formula
RCF = 1.118 x 10-5 x r x (rpm)2
Rinse immediately after using
Wash with low alkali, non-ion detergent Where : r = radius in cm from the center of the head to the
bottom of the tube shield or bucket
Rinse with tap water, then 3-5x with Type I water
Final rinse should have a pH of 5.5 – 5.7 TYPES OF CETRIFUGES
For Blood Clots
1. Horizontal Head Centrifuge
Soak in 1 0% NaOH
For New Pipettes Swinging bucket type
Soak in 5% HCl or 5% nitric acid The centrifuge tubes are held in a vertical position when not
moving but are horizontal when the centrifuge is fully in motion
For Grease
Soak in Generate low speeds only but can produce a tighter pellet of
Any organic solvent precipitate or clotted cells at the bottom of the tube
50% KOH or Contrad 70 Generates a lot of friction
For Permanganate Stain (according to ser wel color violet „to hehe)
Soak in 50% HCl; or
1% FeSO4 in 25% H2SO4
For Bacteriologic Glassware
Soak in 2-4 % cresol solution; followed by
Autoclaving and thorough washing
For Heavy Metal Analysis
Rinse in dilute acid to remove metal ion contaminant
Dichromate - H2SO4 cleaning solution
1M HCl or HNO3
Bromsulhalein Dye
used to check random pieces of residual alkaline detergent
Pink = incomplete removal of detergent
CLINICAL CHEMISTRY 4 of 7
MLS109 Basic Principles of Clinical Chemistry LE 1 TRANS 01
C. Serum
Liquid portion of coagulated blood
It is important that serum samples should be allowed to clot
completely (≈20 minutes) before being centrifuged.
D. Arterial Blood
Properly balance Cetrifuge. Colored Circles represent Used in measuring blood gases (partial pressure of oxygen and
counterbalanced positions for sample tubes. carbon dioxide) and pH
Syringes containing heparin anticoagulant are used instead of
2. Angle Head Centrifuge evacuated tubes because of the pressure in an arterial blood vessel
Has a fixed 25 to 52 degree angle at which tubes are held during
centrifugation E. Other Body Fluids
The sediment packs at an angle bit not as tightly as with a Cerebrospinal fluid
horizontal head centrifuge Paracentesis fluids
Since the sediment is formed at an angle and not tightly packed, Pleural
decantation is not recommended Pericardial
Less prone to heat build up due to air friction than horizontal head Peritoneal
centrifuge heads Amniotic fluid
Urine
F. Hemolysis
Ruptured or destruction of RBC
G. Icterus
Increased bilirubin pigment
H. Lipemia
Increased lipids
Samples should be analyzed within 4 hours to minimize the
effects of evaporation, samples should be properly capped and
kept away from areas of rapid airflow, light, and heat. If testing is XIII. FACTORS AFFECTING COMMON SERUM CONSTITUENTS
to occur after that time, samples should be appropriately stored. A. Prolonged Torniquet Application
For most, this means refrigeration at 4°C for 8 hours. Increased:
3. Ultracentrifuge Cholesterol
Centrifuge head is held at a fixed angle but generates tight Albumin
sediments due to high speeds generated Potassium
Generates the highest speeds Ammonia
Refrigerated to reduce the heat generated by the friction due to Lactic Acid
high centrifugal speeds B. Posture – Supine to Standing
Especially used for lipoprotein separation since refrigeration Increased:
enhances separation of lipoproteins Albumin
Cholesterol
C. CENTRIFUGATION – QC and MAINTENANCE Calcium
Timer and speed to be checked every 3 months
Speed is checked using a tachometer C. Hemolysis
Timer is checked using a stopwatch Increased:
Routinely cleaned everyday Potassium,
Must be cleaned immediately if there are spills Lactate Dehydrogenase (LDH)
Alkaline Phosphatase (ALP)
XII. SPECIMEN CONSIDERATIONS Acid Phosphatase (ACP)
The process of specimen collection, handling, and processing Creatine Kinase (CK)
remains one of the primary areas of pre-analytic error.
Aldolase
Phosphate
A. Phlebotomy or Venipuncture
Magnesium
23- or 21- gauge needle , routinely used needle
Ammonia
IV infusion set or butterfly is used when veins are fragile, small, or
Iron
hard to reach or find
Sites adjacent to IV therapy should be avoided; however if both
D. Alcohol
arms are involved in IV therapy and the IV cannot be discontinued
Increased Decreased
for a short time, a site below the IV site should be sought.
Gamma-Glutamyl Transferase
The initial sample drawn (5mL) should be discarded because it is Glucose
(GGT)
most likely contaminated with IV fluid and only subsequent sample
tubes should be used for analytic purposes Triglycerides
In addition to venipuncture, blood specimens can be collected using
a skin puncture technique ( hell stick, finger stick) E. Diurnal Variation
CLINICAL CHEMISTRY 5 of 7
MLS109 Basic Principles of Clinical Chemistry LE 1 TRANS 01
CLINICAL CHEMISTRY 6 of 7
MLS109 Basic Principles of Clinical Chemistry LE 1 TRANS 01
CLINICAL CHEMISTRY 7 of 7