Professional Documents
Culture Documents
Manual Oxymon - Oxysoft v30103 1604
Manual Oxymon - Oxysoft v30103 1604
Einsteinweg 17
6662 PW Elst
The Netherlands
2
Thank you for purchasing the Oxymon Mk III Near Infrared
Spectrophotometer from Artinis Medical Systems B.V. The Oxymon Mk III
Near Infrared Spectrophotometer (Oxymon) is a continuous wave near
infrared spectroscopy (NIRS) system with transmitters with each 2 or more
wavelengths of emitting light. The device is developed to measure
concentrations of substances using near infrared light. More information
about the Oxymon can be found in chapter 10. The software concerning to
this device is Oxysoft, which performs both data acquisition and analysis.
For more information about Oxysoft, see chapter 0.
Please read the entire manual thoroughly, including the safety information
(chapter 3) in order to get acquainted with the principles and usage of the
Oxymon.
For quick help use the index (page 173) and the table of contents (page 4). In
Oxysoft, Menu Help a help file is available. This file is still under
construction. In the Oxysoft directory, folder “Data examples” an example
project (“Measurement examples.oxyproj”) is available. Use this example to
see how Oxysoft works; how to do calculations, how to work with cyclic
operators, how to add events, how to make new graphs, etc.
If installed, Oxysoft help files can be found at the C:\Artinis Oxymon”. Here
are this manual and “teamviewer support. If these files are not available on
the computer, look at the software CD/USB delivered with the system. With
“teamviewer support” we can assist remotely how to use Oxysoft.
If you still have question after reading the manual and watching the
instruction videos, you can always contact Artinis. Use the e-mail address of
your contact person or use the contact information at the front page of the
manual.
3
Manual Oxymon MkIII .................................................................................................1
For Oxysoft 3.0.103 and higher ................................................................................1
1 Introduction .........................................................................................................3
2 Content .................................................................................................................4
3 Safety information ..............................................................................................8
3.1 General Safety ........................................................................................ 8
3.2 Electrical safety ...................................................................................... 9
3.3 Optical safety ....................................................................................... 10
3.4 Servicing ............................................................................................... 11
3.5 Cleaning ............................................................................................... 11
3.6 Symbols ................................................................................................ 11
4 Introduction to Near Infrared Spectroscopy ............................................ 12
4.1 Optical Oximetry................................................................................... 12
4.2 Near Infrared Spectroscopy ................................................................... 13
4.3 Principles of NIRS .................................................................................. 14
4.4 Spectral extinction coefficients of the chromophores ............................. 15
4.5 The pathlength correction factor ........................................................... 17
4.6 Occlusion methods in NIRS .................................................................... 18
4.7 Venous occlusion technique .................................................................. 18
4.8 Arterial occlusion technique .................................................................. 19
4.9 Quantitation of absolute tissue blood volume........................................ 19
4.10 Quantitation of absolute blood flow ...................................................... 20
4.11 References ............................................................................................ 21
5 Hardware and software installation ............................................................ 25
5.1 Hardware installation ........................................................................... 25
5.2 Software installation ............................................................................ 27
5.3 Filesetup installation............................................................................. 30
5.4 Oxymon connection .............................................................................. 33
5.4.1 Installing on Windows 7 or 8 ................................................... 33
5.5 Polhemus Patriot Connection (Optional) ................................................ 37
6 Oxymon ............................................................................................................. 38
6.1 Oxymon ................................................................................................ 38
6.2 Interconnecting multiple Oxymons ........................................................ 41
6.2.1 Cables ...................................................................................... 42
4
6.2.2 Interconnections ...................................................................... 42
6.2.3 Software ................................................................................... 42
6.3 AD input board ..................................................................................... 44
6.4 AD boxes .............................................................................................. 44
6.4.1 AD input box ............................................................................ 44
6.4.2 AD output box .......................................................................... 45
6.5 The PortaSync ....................................................................................... 49
6.6 Polhemus Patriot .................................................................................. 51
7 Oxysoft ............................................................................................................... 53
7.1 The program ......................................................................................... 53
7.2 Start Oxysoft ........................................................................................ 55
7.3 Close Oxysoft ........................................................................................ 55
7.4 About Oxysoft....................................................................................... 55
7.5 License ................................................................................................. 56
7.6 Program options ................................................................................... 56
7.6.1 Data collection ......................................................................... 57
7.6.2 Graph options .......................................................................... 58
7.6.3 Template file ............................................................................ 59
7.6.4 Default coefficient tables ......................................................... 60
8 Measuring ......................................................................................................... 61
8.1 How to do a measurement .................................................................... 61
8.1.1 Preparation .............................................................................. 61
8.1.2 Start experiment ...................................................................... 62
8.2 Connect the optical fibers ...................................................................... 63
8.3 The optode holder................................................................................. 65
8.4 Data Collection ..................................................................................... 67
8.5 DAQ status view ................................................................................... 68
8.6 Template DAQ STATE ............................................................................ 69
8.7 DAQ value view .................................................................................... 69
8.8 DAQ settings......................................................................................... 71
8.8.1 Laser output level .................................................................... 71
8.8.2 Gain settings ............................................................................ 71
8.8.3 Out of range ............................................................................. 73
8.9 Calibration............................................................................................ 74
9 Data management .......................................................................................... 76
9.1 Projects ................................................................................................ 76
9.1.1 A new project ........................................................................... 76
9.1.2 Open a project ......................................................................... 76
9.1.3 Save a project........................................................................... 77
9.1.4 Project properties .................................................................... 77
9.1.5 Close a project ......................................................................... 78
5
9.2 Measurements...................................................................................... 78
9.2.1 A new measurement ............................................................... 78
9.2.2 DAQ measurement wizard ...................................................... 78
9.2.3 Create offline measurement ................................................... 85
9.2.4 Measurement properties ........................................................ 87
9.2.5 Rename a measurement ......................................................... 87
9.2.6 Duplicate a measurement ....................................................... 87
9.2.7 Remove a measurement ......................................................... 87
9.2.8 Digitize optode positions (3D plugin only) .............................. 87
9.3 Graphs ................................................................................................. 91
9.3.1 Create (a regular) Graph .......................................................... 92
9.3.2 Create all graphs ...................................................................... 96
9.3.3 Create graphs from a template ............................................... 96
9.3.4 Topo Graph .............................................................................. 97
9.3.5 Spectrum Graph..................................................................... 102
9.3.6 Indicator view ........................................................................ 103
9.3.7 Graph properties ................................................................... 104
9.3.8 Open a graph ......................................................................... 104
9.3.9 Rename a graph..................................................................... 105
9.3.10 Duplicate a graph................................................................... 105
9.3.11 Clear a graph.......................................................................... 105
9.3.12 Freeze a graph ....................................................................... 106
9.3.13 Arrange graphs ...................................................................... 106
9.3.14 Close a graph ......................................................................... 106
9.3.15 Remove a graph ..................................................................... 106
9.3.16 Printing .................................................................................. 107
9.3.17 Zoom ...................................................................................... 107
9.3.18 Time span .............................................................................. 107
9.4 Traces................................................................................................. 107
9.4.1 Create a trace ........................................................................ 108
9.4.2 Create all traces ..................................................................... 113
9.4.3 Trace properties .................................................................... 114
9.4.4 Duplicate a trace.................................................................... 114
9.4.5 Remove a trace ...................................................................... 114
9.4.6 Bias a trace ............................................................................ 114
9.5 Data import and export ...................................................................... 115
9.5.1 Open *.oxy3 files ................................................................... 115
9.5.2 Import an Oxysoft measurement .......................................... 115
9.5.3 Import other data files (ASCII) ............................................... 115
9.5.4 Export data ............................................................................ 119
9.5.5 Export data Real time (ASCII writer) ...................................... 123
9.5.6 Export data Real time (FieldTrip buffer) ................................ 123
10 Data analysis................................................................................................... 125
10.1 Events ................................................................................................ 125
10.1.1 Define event keys .................................................................. 125
10.1.2 Insert an event during measuring ......................................... 126
10.1.3 Insert an event after measuring ............................................ 126
10.1.4 Generate events automatically ............................................. 127
6
10.1.5 Event properties .................................................................... 129
10.1.6 Event list................................................................................. 129
10.1.7 Change or remove an event................................................... 130
10.2 Periods ............................................................................................... 130
10.3 Cyclic operators .................................................................................. 131
10.4 Calculation ......................................................................................... 135
10.5 Filters ................................................................................................. 139
10.5.1 Selection of an appropriate filter........................................... 139
10.5.2 Add a filter to a trace ............................................................. 139
10.5.3 Remove a filter from a trace .................................................. 139
10.5.4 Add a filter to all traces of a measurement ........................... 139
10.5.5 Remove a filter from all traces of a measurement ................ 140
10.5.6 Moving average filters ........................................................... 140
10.5.7 FIR filters ................................................................................ 142
11 Optode templates ......................................................................................... 148
11.1 Introduction........................................................................................ 148
11.2 Setup examples .................................................................................. 149
11.2.1 1 Channel setup ..................................................................... 149
11.2.2 2 Channel Setup ..................................................................... 150
11.2.3 4 Channel Setup ..................................................................... 150
11.2.4 8 Channel Setup ..................................................................... 151
11.2.5 24 channel Setup ................................................................... 152
12 Tissue saturation index ................................................................................ 155
12.1 Spatial resolved spectroscopy.............................................................. 155
12.2 Theory ................................................................................................ 155
12.2.1 Parameters............................................................................. 156
12.3 TSI Fit Factor ....................................................................................... 157
12.4 TSI Accesories ..................................................................................... 158
12.5 TSI Calibration .................................................................................... 158
12.6 TSI measurement ................................................................................ 160
12.7 Overview of MLB and SRS theories ...................................................... 161
13 Frequently Asked Questions ....................................................................... 162
13.1.1 General................................................................................... 162
13.1.2 Installation and license key .................................................... 163
13.1.3 Using the software ................................................................. 165
14 List of shortcuts ............................................................................................. 169
15 Abbreviations ................................................................................................. 170
16 Warranty policy .............................................................................................. 171
17 Index ................................................................................................................. 173
7
BEFORE OPERATING THE OXYMON:
1.1 THE USER MUST ALWAYS CHECK THE EQUIPMENT PRIOR TO USE
AND ENSURE IT IS SAFE AND PROPER USE.
1.2 THE OXYMON MAY ONLY BE USED BY PEOPLE WHO READ THE
MANUAL THOROUGHLY.
O O
1.9 USE THE OXYMON BETWEEN 10 C AND 40 C. FOR SPECIAL
CONDITIONS (LOW TEMPERATURE ETC.) PLEASE CONTACT
ARTINIS FOR FURTHER INSTRUCTIONS.
8
1.10 BE AWARE WHEN APPLYING THE DEVICE FOR MULTIPLE HOURS.
TOO MUCH SWEAT, STRESS AND/OR WARMTH CAN CAUSE SKIN
IRRITATION. DO NOT APPLY TOO MUCH PRESSURE ON THE SKIN
WITH THE PROBE ATTACHMENT AND STOP AND REMOVE PROBE
IF SUBJECT FEELS IRRITATION OF THE PROBE ATTACHMENT.
2.6 ONLY USE THE SUPPLIED ADAPTER AS POWER SUPPLY FOR THE
ADBOX.
9
2.8 THE OXYMON IS A “CLASS I” LASER PRODUCT. THE LASER
PRODUCTS HAVE BE CLASSIFIED ACCORDING TO “IEC 60825-1
1993 AMENDMENT 2 2001”.
2.12 THE LASERS FIRE AT 1000, 2000 OR 4000 HZ (16 AND MORE
CHANNELS, 8 CHANNELS, 1-4 CHANNELS).
2.13 THE PULSE TIMES ARE ABOUT 70 NS (HALF WIDTH FULL MAX).
2.18 THE LASERS ONLY WORK WHEN A FIBER FROM ARTINIS MEDICAL
SYSTEMS IS CONNECTED TO BOTH LASERS OF ONE
TRANSMITTER BECAUSE OF A SAFETY INTERLOCK MECHANISM.
THE RED “MEASURE” LAMP AT THE FRONT OF THE OXYMON
TURNS ON IF THE LASERS ARE FIRING.
10
In case of problems, unplug the power cords and contact the
manufacturer. No user-serviceable parts are inside the Oxymon. Service
contracts are available from the manufacturer.
Turn off the Oxymon, and disconnect from the power supply. Clean the
Oxymon by using a soft cloth moistened with a non-aggressive general
purpose cleaner. Do not spill any liquid into the instrument.
Potential equalization
Type B equipment
M Manufacturing date
11
This chapter introduces the history of NIRS as well as a basic background on
how to use the technique. For a brief introduction into NIRS principles the first
chapter of the ‘Near Infrared Spectroscopy: Toy or Tool?’ thesis by dr. W.N.J.M.
Colier is included below. This part is not specific for the Oxymon.
During the second half of the 19th century it was discovered that blood
cells contain a coloring substance whose spectrum is influenced by various
blood gases [Hoppe, 1862]. Stokes discovered that this colored substance
was the oxygen carrier of the blood. Felix Hoppe-Seyler, a German chemist,
named the substance hemoglobin. He found that when shaking a solution
of hemoglobin with air this resulted in a spectral change. He called the
component responsible for this change "oxyhemoglobin". In 1876 Karl von
Vierordt published the first study in which spectroscopy was used to study
hemoglobin in tissue. He found a transition from oxyhemoglobin to
deoxyhemoglobin in the tissue of the finger after the arm was fully
occluded, a technique still practiced today to study oxygenation in muscles.
The oximeter of Millikan was improved by Wood and Geraci [1949]. Further
developments on reflection oximetry were done by Brinkman and Zijlstra.
They developed the "Cyclops", a device measuring oxygen saturation on the
forehead of a subject [Brinkman et al. 1950]. In the following years,
oximetry was used more and more in clinical situations. The title of Zijlstra's
thesis [1951]: "Fundamentals and applications of clinical oximetry"
12
indicated that they were far ahead of their time.
In his article in Science [1977] Jöbsis reported that biological tissues have a
relatively good transparency for light in the near infrared region (700-1300
nm). Therefore it is possible to transmit sufficient photons through organs
for in situ monitoring. In this region hemoglobin, which can be divided into
its main components oxyhemoglobin (O2Hb) and deoxyhemoglobin (HHb),
shows oxygen dependent absorption. The main attention by Jöbsis and his
co-workers however was given to the redox state of cytochrome a,a3
(Cyt.Ox.), the terminal enzyme of the mitochondrial respiratory electron
transport chain. Approximately 90% of the intracellular oxygen
consumption is catalyzed by this enzyme. It has been shown that the
absence of oxygen results in a complete reduction of Cyt.Ox. This reduction
can be monitored by measuring the change in absorption band of Cyt.Ox.
in the near infrared region (770-880 nm) [Chance 1966]. Information
gained from the near infrared spectrum can therefore give insight into the
oxygen availability at the intracellular level. Combined with the information
from the hemoglobin signal it is now possible to monitor both circulatory
oxygen supply and intracellular oxygen consumption of the tissue.
In the years following the first publication of Jöbsis many studies were
carried out focusing on the assessment of Cyt.Ox. within the brains of
animals or humans, especially neonates [Jöbsis 1979, Giannini et al. 1982,
Keizer et al. 1985, Proctor et al. 1985, Brazy et al. 1985, Brazy and Lewis
1986, Jöbsis-VanderVliet et al. 1987, Brazy 1988, Jöbsis-VanderVliet et al.
1988]. Due to technical improvements NIRS became a more widespread
method at the end of the eighties and into the nineties. Most clinical
studies still are in the neonatal field [Brazy 1991, Edwards et al. 1991, Livera
et al. 1991, Skov et al. 1991, Wickramasinghe et al. 1992, Skov et al. 1992,
McCormick et al. 1993, Bucher et al. 1993], some of them performed by the
Nijmegen group [Liem et al. 1992, 1994, 1995]. A short history of NIRS in
Milestones is given in Table 4.1.
13
Table 4.1. The history of NIRS in milestones
The technique of NIRS relies on the Lambert-Beer law [Beer 1851], given
by:
𝐼0
𝑂𝐷𝜆 = 𝑙𝑜𝑔 ( ) = 𝜀𝜆 𝑐𝐿
𝐼
∆𝑂𝐷𝜆
∆𝑐 =
𝜀𝜆 𝐿𝐵
This equation is valid for a medium with one chromophore. In the case of
more chromophores we need to measure at least as many wavelengths as
there are chromophores present. This results in a set of linear equations.
The solution of this set leads to the algorithm used in most NIRS systems.
14
Figure 4-1. A scattering medium with an incident and transmitted light ray. The chromophore is
symbolized by black dots. Light ray A is scattered, and therefore travels a distance which equals
the pathlength correction factor times the physical pathlength L. Light ray B is absorbed
completely after being scattered.
15
Figure 4-2. Extinction coefficients of adult oxy- and deoxyhemoglobin (O2Hb and HHb
respectively), determined by Wray [1988] (solid lines) and Zijlstra [1991] (• and ▲ symbols), and
cytochrome oxidase (Cyt.Ox, ● symbol), also from Wray [1988].
Up until now there exists no consensus about which of the data sets for
hemoglobin and cytochrome oxidase come closest to reality. It can
therefore be concluded that the existence of several spectral data sets
results in a non-uniform algorithm for NIRS.
16
Table 4.2. Values of the Differential Pathlength Factor (DPF) for various types of tissue determined
by either "time-of-flight" (TRS) or "frequency-resolved" (FRS) measurements. All DPF values are
For the quantitation of the absorption changes not only the physical
pathlength L, is needed, but also the DPF. The DPF can be accessed via
different ways. The most common way is by "time-of-flight" measurements
[Delpy et al. 1988, Wyatt et al. 1990a, Ferrari et al. 1992, van der Zee et al.
1992]. An ultra-short laser pulse is fired into the tissue. The pulse is
detected by an ultrafast camera. The time of flight t can then be converted
into a travelled distance d using the formula:
𝑐𝑡
𝑑=
𝑛
where c is the velocity of light and n the refractive index of the tissue.
Division of d by L gives the DPF.
17
between the incident and transmitted light. From the phase shift the mean
pathlength can be obtained. A good set of pathlength data obtained with
this technique for brain and muscle tissue has recently been published by
Duncan [1995]. This technique still is cumbersome, relatively expensive and
insensitive compared to continuous wave techniques. An overview of
studies to DPF values is shown in Table 4.2.
Figure 4-3. An example of a NIRS tracing. In this case two consecutive venous occlusions (V.O.)
were applied to the upper arm. The NIRS optodes were attached to the brachio-radialis muscle of
the forearm. During V.O. there is arterial inflow, but no venous outflow. This results in an increase
of both oxy- and deoxyhaemoglobin. From the rise in the tHb signal the blood flow into the limb
can be calculated.
18
The blood flow into a limb can be completely stopped by inflating a blood
pressure cuff to a pressure of more than 250 mmHg. It is then possible to
calculate the local oxygen consumption in rest [Cheatle et al. 1991, de Blasi
et al. 1993] or during (isometric) exercise [Colier et al. 1995] from the
gradient of the subsequent decrease of the O2Hb signal.
After the pressure of the cuff is released, the tissue will show a hyperemic
reaction. The recovery time of the re-saturation observed with NIRS can be
used as measure for, e.g., the vascularisation of the leg in patients with
peripheral vascular disease [McCully et al. 1994, Komiyama et al. 1994].
By combining the NIRS data with arterial saturation (SaO2), measured for
example by pulse oximetry, it is possible to quantitate the absolute value of
blood volume of the examined tissue. This method was first described by
Wyatt et al. [1990b] for the quantitation of cerebral blood volume.
The effect of a small, gradual and transient change in SaO2 on the O2Hb
concentration is monitored. A decrease in SaO2 of approximately 10%,
induced by lowering the inspired oxygen concentration, is sufficient to
calculate the blood volume. Provided blood flow, volume and oxygen
consumption remain constant during the procedure, the tissue blood
volume (TBV) in ml/100 g is given by:
∆(𝑂2 − 𝐻𝐻𝑏)
𝑇𝐵𝑉 = 𝑐𝐻𝑏 𝜌𝜏 𝑘
2𝑅∆𝑆𝑎𝑂2
3
where cHb (mM) is the hemoglobin concentration of whole blood, ρT (g/cm )
the specific density of the tissue, k a constant reflecting metric conversions
and R is, in case of cerebral tissue, the large-to-small vessel hematocrit ratio
with a value of 0.69 [Lammertsma et al. 1984]. The difference between the
O2Hb and HHb concentration is taken to obtain a better signal-to-noise
ratio. This difference is also called the hemoglobin difference signal (HbDiff).
19
The principle of measuring organ blood flow with NIRS is based on the Fick
principle which states that the accumulation of a tracer in an organ equals
the difference between the inflow (arterial concentration x flow) and
outflow (venous concentration x flow). If we measure within the transit time
of the tracer through the organ the venous concentration will be zero. In
NIRS the tracer used is a bolus of O2Hb, which can be induced by suddenly
increasing the inspired oxygen concentration. The concentration of the
bolus can be measured by attaching a pulse oximetry probe onto the
organ. The increase of O2Hb as measured by NIRS represents the
-
accumulation of the bolus into the organ. The blood flow (BF, in ml●100g
1 -1
●min ) through the organ is then given by the change of O2Hb divided by
-1
the product of the arterial hemoglobin concentration (cHb, in g●ml ) times
the integral of change in arterial saturation (SaO2, in %) :
𝐾∆(𝑂2 𝐻𝑏)
𝐵𝐹 = 𝑡
𝑐𝐻𝑏 ∙ 10−2 ∙ ∫0 ∆(𝑆𝑎𝑂2 )𝑑𝑡
20
For more literature about NIRS, see the literature list.
Assendelft van, OW. Spectrophotometry of Haemoglobin Derivatives. Springfield, IL. CC
Thomas 1970; 355–359,
Aoyagi T, Kishi M, Yamaguchi K, Watanabe S. Improvement of the earpiece oximeter. In:
Abstracts of the Japanese Society of Medical Electronics and Biological
Engineering 1974; 90-91.
Beer A. Versuch der Absorptions-Verhältnisse des Cordierites für rothes Licht zu
bestimmen. Ann Physik Chem 1851; 84: 37-52.
Benaron DA, Gwiazdowski, Kurth CD, Steven J, Delvoria-Papadopoulos, Chance B. Optical
pathlength of 754nm and 816nm light emitted into the head of infants. IEEE Eng
Med Biol Soc 1990; 12: 1117-1119.
Brazy JE, Lewis DV, Mitnick MH, Jöbsis-VanderVliet FF. Noninvasive monitoring of cerebral
oxygenation in preterm infants: preliminary observations. Pediatrics 1985; 75:
217-225.
Brazy JE, Lewis DV. Changes in cerebral blood volume and cytochrome aa 3 during
hypertensive peaks in preterm infants. J Pediatr 1986; 108 983-987.
Brazy JE. Cerebral oxygen monitoring with near infrared spectroscopy: clinical application
to neonates. J Clin Monit 1991; 7: 325-334.
Brazy JE. Effect of crying on cerebral blood volume and cytochrome aa3. J Pediatr 1988; 112:
457-461.
Brinkman R, Zijlstra WG, Koopmans RK. A method for continuous observation of
percentage oxygen saturation in patients. Acta Chir Neerl 1950; 1: 333-344.
Bucher HU, Edwards AD, Lipp AE, Duc G. Comparison between near infrared spectroscopy
and 133Xenon clearance for estimation of cerebral blood flow in critically ill
preterm infants. Ped Res 1993; 33: 56-60.
Chance B. In: The Biochemistry of Copper (Peisach G, Aisen P, Blumberg WE, eds.) pp: 293-
301. Academic Press, New York. 1966.
Cheatle TR, Potter LA, Cope M, Delpy D, Coleridge Smith PD, Scurr JH. Near-infrared
spectroscopy in peripheral vascular disease. Br J Surg 1991; 78: 405-408.
Colier WNJM, Liem D, Oeseburg B. The effect of fetal hemoglobin on the algorithm used in
near infrared spectroscopy. In: Proc Int Soc Oxygen Transport to Tissue, P76,
1992.
Cope M. The application of Near Infrared Spectroscopy to non-invasive monitoring of
cerebral oxygenation in the newborn infant. PhD-Thesis of the University of
London, Department of Medical Physics and Bioengineering, UK, 1991.
Crowe J. Absorption spectra of cytochrome oxidase. In: Newsletter Concerted Action NIRS
& I, 1994; 3: 3-4.
De Blasi RA, Cope M, Elwell C, Safoue F, Ferrari M. Noninvasive measurement of human
forearm oxygen consumption by near infrared spectroscopy. Eur J Appl Physiol
1993; 67: 20-25.
De Blasi RA, Ferrari M, Natali A, Conti G, Mega A, Gasparetto A. Noninvasive measurement
of forearm blood flow and oxygen consumption by near infrared spectroscopy. J
Appl Physiol 1994; 76: 1388-1393.
Delpy DT, Cope M, Zee P van der, Arridge S, Wray S, Wyatt J. Estimation of optical
pathlength through tissue from direct time of flight measurements. Phys Med Biol
1988; 33: 1433-1442.
Duncan A, Meek JH, Clemence M, Elwell CE, Tyszczuk L, Cope M, Delpy DT. Optical
pathlength measurements on adult head, calf and forearm and the head of the
newborn infant using phase resolved optical spectroscopy. Phys Med Biol 1995;
40: 295-304.
Edwards AD, Brown C, Cope M, Wyatt JS, McCormick DC, Roth SC, Delpy DT, Reynolds EOR.
Quantification of concentration changes in neonatal human cerebral oxidized
cytochrome oxidase. J Appl Physiol 1991; 71: 1907-1913.
Edwards AD, Wyatt JS, Richardson CE, Delpy DT, Cope M, Reynolds EOR. Cotside
measurement of cerebral blood flow in ill newborn infants by near-infrared
spectroscopy (NIRS). Lancet 1988; 2: 770-771.
21
Elwell CE, Cope M, Edwards AD, Wyatt JS, Reynolds EOR. Measurement of cerebral blood
flow in adult humans using near infrared spectroscopy - methodology and
possible errors. Adv Exp Med Biol 1992; 317: 325-245.
Elwell CE, Owen-Reece H, Cope M, Wyatt JS, Edwards AD, Delpy DT, Reynolds EOR.
Measurement of adult cerebral haemodynamics using near infrared
spectroscopy. Acta Neurochir 1993; 59(Suppl): 74-80.
Essenpreis M, Elwell CE, Cope M, van der Zee P, Arridge SR, Delpy DT. Spectral dependence
of temporal point spread functions in human tissues. Appl Optics 1993; 32: 418-
425.
Ferrari M, Wei Q, Carraresi L, De Blasi RA, Zaccanti G. Time-resolved spectroscopy of
human forearm. J Photochem Photobiol 1992; 16: 141-153.
Giannini I, Ferrari M, Carpi A, Fasella P. Rat brain monitoring by near infrared spectroscopy:
an assessment of possible clinical significance. Physiol Chem Phys 1982; 14: 295-
305.
Hardman KD, Eylar EH, Ray DK, Banaszak LJ, Gurd FRN. Isolation of sperm whale myoglobin
by low temperature fractionation with ethanol and metallic ions. J Biol Chem
1966; 241: 432-442.
Harris AP, Sendak MJ, Donham RT, Thomas M, Duncan D. Absorption characteristics of
human fetal haemoglobin at wavelengths used in pulse oximetry. J Clin Monit
1988; 4: 175-177.
Horecker BL. The absorption spectra of haemoglobin and its derivatives in the visible and
near-infrared regions. J Biol Chem 1943; 148: 173-183.
Hüfner G. Ueber die gleichzeitige quantitative Bestimmung zweier Farbstoffe im Blute mit
Hilfe des Spectrophotometers. Arch Anat Physiol 1900; Physiol Abt 39.
Jöbsis FF. Noninvasive , infrared monitoring of cerebral and myocardial oxygen suffiency
and circulatory parameters. Science 1977; 198: 1264-1267.
Jöbsis FF. Oxidative metabolic effects of cerebral hypoxia. Adv Neurol 1979; 26: 299-318.
Jöbsis-VanderVliet FF, Fox E, Sugioka K. Monitoring of cerebral oxygenation and
cytochrome aa3 redox state. In: Advances in Oxygen Monitoring (Tremper KK,
Barker J, eds.) pp: 209-230. Little, Brown and Company, Boston 1987.
Jöbsis-VanderVliet FF, Piantadosi CA, Sylvia AL, Lucas SK, Keizer HH. Near-infrared
monitoring of cerebral oxygen sufficiency. I: Spectra of cytochrome c oxidase.
Neurol Res 1988; 10: 7-17.
Keizer HH, Jöbsis-VanderVliet FF, Lucas SS, Piantadosi CA, Sylvia AL. The near infrared (NIR)
absorption band of cytochrome aa3 in purified enzyme, isolated mitochondria
and in intact brain in situ. Adv Exp Med Biol 1985; 191: 823-832.
Kelleher JF. Pulse oximetry. J Clin Monit 1989; 5: 37-62.
Komiyama T, Shigematsu H, Yasuhara H, Muto T. An objective assessment of intermittent
cladication by near infrared spectroscopy. Eur J Vasc Surg 1994; 8: 294-296.
Kramer K. Fortlaufende Registrierung der Sauerstoffsättigung im Blute an uneröffneten
Blutgefässen. Klin Wochenschr 1934; 13: 379-380.
Lammertsma AA, Brooks DJ, Beaney RP, Turton DR, Kensett MJ, Heather JD, Marshall J,
Jones T. In vivo measurement of regional cerebral haematocrit using positron
emission tomography. J Cereb Blood Flow Metab 1984; 4: 317-322.
Liem KD, Hopman JCW, Kollée LAA , Oeseburg B. Effects of repeated indomethacin ad-
ministration on cerebral oxygenation and haemodynamics in preterm infants:
combined near infrared spectrophotometry and Doppler ultrasound study. Eur J
Pediatr 1994; 153: 504-509.
Liem KD, Hopman JCW, Oeseburg B, de Haan AFJ, Festen C, Kollee LAA. Cerebral
oxygenation and hemodynamics during induction of extracorpereal membrane
oxygenation as investigated by near infrares spectroscopy. Pediatrics 1995. (in
press).
Liem KD, Oeseburg B and Hopman JCW. Method for the fixation of optrodes in near
infrared spectrophotometry. Med Biol Eng Comput 1992; 30:120-121.
Livera LN, Spencer SA Thorniley MS, Wickramasinghe YABD, Rolfe P. Effect of hypoxaemia
and bradycardia on neonatal cerebral haemodynamics. Arch Dis Child 1991; 66:
376-380.
Matthes K, Gross F. Fortlaufende Registrierung der Lichtabsorption der Farbe des Blutes in
zwei verschiedenen Spektralbezirken. Arch Exp Pathol Pharmacol 1939; 191: 369-
380.
Matthes K. Untersuchungen über den Verlauf der Oxyhämoglobinreduktion in der
menschlichen Haut. Pflügers Arch 1942; 246: 70-91.
22
McCormick DC, Edwards AD, Brown GC, Wyatt JS, Potter A, Cope M, Delpy DT, Reynolds
EOR. Effect of indomethacine on cerebral oxidized cytochrome oxidase in
preterm infants. Ped Res 1993; 33: 603-608.
McCully KK, Halber C, Posner JD. Exercise-induced changes in oxygen saturation in the calf
muscles of elderly subjects with peripheral vascular disease. J Gerontol 1994; 49:
B128-B134.
Millikan GA. The oximeter, an instrument for measuring continuously the oxygen
saturation of arterial blood in man. Rev Sci Instrum 1942; 13: 434-444.
Nakajima S, Hirai Y, Takase H, Kuse, Aoyagi T, Kishi M, Yamaguchi K. Performances of new
pulse wave earpiece oximeter. Resp Circ 1975; 23: 41-45.
Nicolai L. Fortgesetzte Untersuchungen über den Verlauf der Oxyhämoglobinreduktion in
der menslichen Hut. Arch Ges Physiol 1931; 230: 328-245.
Patterson MS, Chance B, Wilson BC. Time resolved reflectance and transmittance for the
non-invasive measurement of tissue optical properties. Applied optic 1989; 28:
2331-2336
Proctor HJ, Cairns C, Fillipo D, Jöbsis-VanderVliet FF. Near infrared spectrophotometry:
potential role during increased intracranial pressure. Adv Exp Med Biol 1985; 191:
863-871.
Rea PA, Crowe J, Wickramasinghe Y, Rolfe P. Non-invasive optical methods for the study of
cerebral metabolism in the human newborn; a technique for the future? J Med
Eng Technol 1985; 9: 160-166.
Samejima T, Yang JT. Optical Rotary dispersion of sperm-whale myoglobin and its
derivatives. J Mol Biol 1964; 8: 863-871.
Sevick EM, Chance B, Leigh J, Nioka S, Maris M. Quantitation of time- and frequency-
resolved optical spectra for the determination of tissue oxygenation. Anal
Biochem 1991; 195: 330-351.
Skov L, Greissen G. Apparent cerebral cytochrome aa3 reduction during cardiopulmonary
bypass in hypoxaemic children with congenital heart disease. A critical analysis of
in vivo near-infrared spectrophotometric data. Physiol Meas 1994. 15: 447-457.
Skov L, Pryds O, Greissen G. Estimating cerebral blood flow in newborn infants:
comparison of near infrared spectroscopy and 133Xe clearance. Ped Res 1991; 30:
570-573.
Skov L, Rydling J, Pryds O, Greissen G. Changes in cerebral oxygenation and cerebral blood
volume during endotracheal suctioning in ventilated neonates. Acta Pediatr 1992;
81: 389-393.
Theorell H, Åkeson Å. Reversible splitting of a homogeneous horse myoglobin. Ann Acad
Scient Fennicæ 1955; 60: 303-312.
Thorniley MS, Houston R, Wickramasinghe YA, Rolfe P. Application of near-infrared
spectroscopy for the assessment of the oxygenation level of myoglobin and
haemoglobin in cardiac muscle in vivo. Biochem Soc Trans. 1990; 18: 1195-1196.
Vierordt K. Die quantitative Spektralanalyse in ihrer anwendung auf Physiologie, Chemie
und Technologie. Tubingen: H. Laup p'sche Buchhandlung. 1876.
Wickramasinghe YABD, Livera LN, Spencer SA, Rolfe P, Thorniley MS. Plethysmographic
validation of near infrared spectroscopic monitoring of cerebral blood volume.
Arch Dis Childhood 1992; 67: 407-411.
Wickramasinghe YABD, Palmer KS, Houston R, Spencer SA, Rolfe P, Thorniley MS, Oeseburg
B, Colier WNJM. Effects of fetal haemoglobin on the determination of neonatal
cerebral oxygenation by near-infrared spectroscopy. Pediatr Res 1993; 34: 15-17.
Wood E, Geraci JE. Photoelectric determination of arterial oxygen saturation in man. J Lab
Clin Med 1949; 34: 387-401.
Wray S, Cope M, Delpy DT, Wyatt JS, Reynolds EOR. Characterisation of the near infrared
absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive
monitoring of cerebral oxygenation. Biochim Biophys Acta 1988; 933: 184-192.
Wyatt JS, Cope M, Delpy DT, Zee van der P, Arridge S, Edwards AD, Reynolds EOR.
Measurement of optical pathlength for cerebral near infrared spectroscopy in
newborn infants. Dev Neurosci 1990a; 12: 140-144.
Wyatt JS, Cope M, Delpy DT, Richardson CE, Edwards AD, Wray S, Reynolds EOR.
Quantitation of cerebral blood volume in newborn human infants by near
infrared spectroscopy. J Appl Physiol 1990b; 68: 1086-1091.
Yamazaki I, Yokota K, Shikama K. Preparation of crystalline oxymyoglobin from horse
heart. J Biol Chem 1964; 239: 4151-4153.
23
Zee van der P, Cope M, Arridge SR, Essenpreis M, Potter LA, Edwards AD, Wyatt JS,
McCormick DC, Roth SC, Reynolds EOR, Delpy DT. Experimentally measured
optical pathlengths for the adult head, calf and forearm and the head of the
newborn infant as a function of inter optode spacing. Adv Exp Med Biol 1992.
316: 143-153.
Zijlstra WG, Buursma A, Meeuwsen-van der Roest WP. Absorption spectra of human fetal
and adult oxyhemoglobin de-oxyhemoglobin, carboxyhemoglobin and
methemoglobin. Clin Chem 1991; 37: 1633-1638.
Zijlstra WG. Fundamentals and applications of clinical oxymetry. PhD-Thesis University of
Groningen, The Netherlands, 1951.
24
In this chapter the installation of both the hardware and the software is
described. It is important to read this chapter carefully. It is also important not
to switch on the Oxymon before the software is installed. Windows 7 and 8 are
supported.
Before using the Oxymon, the system should be properly installed and
connected. Connections should only be changed when the instruments are
off.
Make sure that the mains voltage suits the needed operating voltage! This
value is at the power supply information. The Oxymon is equipped with a
self-switching power supply, operating between 110 and 230 VAC, either 50
or 60 Hz. Also, check the mains voltage supply of the additional instruments
or computers. A safety transformer can be used, but is not necessary
(Figure 5-1). Plug the power supply cable into an electric power supply and
connect it to the power supply inlet of the Oxymon.
25
Figure 5-2. How to change the “Power Save” options of the computer
The Oxymon should be installed such way, that a free airflow is guaranteed.
No obstructions should be placed before an air in- and outlet.
For installation turn on the power of the computer and computer monitor
but DO NOT yet turn on the Oxymon or connect the USB cable. Wait until
all the software has been installed.
Note that for proper functioning, the 'Power Save' mode of the PC should
be de-activated. This can be found by right-clicking the mouse while
situated over the desktop. The “Display Properties” screen will appear. Set
“Select screensaver” to none. Click on the button “Power” and set all
settings for use with the Oxymon to “Never” (Figure 5-2).
26
Insert the CD/USB with the software delivered with the Oxymon system into
the PC. If the computer does not automatically open the CD/USB, go to “My
Computer” and open the CD/USB by double clicking. Open “ReadMe.txt”
and follow the instructions in the file. This paragraph will describe the steps
one by one. Note that the version number of the software can change
according to the version delivered.
Open “Oxysoft_x.msi”, where “x” is the version number of Oxysoft. The
“Setup wizard” will open (Figure 5-3). Click on “Next” to start the “Setup
Wizard”.
27
Figure 5-4. End-User License Agreement
28
Figure 5-6. Start installation
29
To install device specific files and some help files, open “FileSetup.exe” of
the installation CD/USB. When you run this file, it might be that Windows
asks for permission. Please click yes.
Select the language during installation in the next window (Figure 5-8). The
“Setup wizard” will open Figure 5-9. Click on “Next” to start the “Setup
Wizard”. Read the installation instructions on Figure 5-10 and click “Next”.
30
Figure 5-10 Installation instructions
Click on “Next”. In the following window, you can place the shortcuts in the
windows menu (Figure 5-12)
31
Figure 5-12. Select start menu folder
Click on “Next”. In the following window, the installation will start by clicking
on “Install” (Figure 5-13). After the installation, it will finish by clicking on
“Finish” (Figure 5-14).
32
Figure 5-14. Finish installation
At page 38 all elements of the Oxymon are described. Now connect the
Oxymon to the PC or laptop using the USB cable. Put one end into the USB
connection of the Oxymon (Figure 5-15). Connect the other end to a free
USB port on the PC. This can be either a type 1.1 or 2.0 USB connection.
Turn the Oxymon on by the main switch at the back of the Oxymon. The
power indicator on the front of the instrument will turn green. A balloon will
come up in the right corner of the monitor. Do NOT remove the CD/USB.
The PC will detect new hardware.
Once you connect the USB to your computer, Windows 7 and 8 will install
the device automatically. Please notice the information you see in the
bottom right of your screen (Figure 5-16).
The following steps might be necessary for Oxymon devices before 2015
on windows 7. Please contact Artinis if you wish to install such a device on
Windows 8.
33
Figure 5-16. Automatic driver installation notification
If Windows gives an error which says “Device driver software was not
successfully installed” (Figure 5-17), the driver needs to be updated.
34
Figure 5-19. Choose to browse a file on my computer
Figure 5-20. Select the folder to search for the FTDI driver
35
Figure 5-21. Safety warning. Please select to install anyway
36
For installation on Winodws 7 or higher, insert the included Polhemus DVD
and the autostart screen will start. Click on the “Install Host Software”
button and follow the on-screen instructions. This will install both the driver
and some proprietary Polhemus software tools. You can uncheck the
“Polhemus SDK Files” option. Make sure that the Patriot device is not
connected before of while installing the software. After having successfully
installed the software, connect the Patriot device to a USB-port of your
computer. For more information, please see the Patriot manual also
included on the Polhemus DVD
37
This chapter introduces the Oxymon and explains how to connect it to the PC,
the subject and other devices. Please, follow the installation procedure as
described in chapter 5 before connecting the Oxymon(s) for the first time to the
computer.
38
Figure 6-1. Oxymon, front side (above) and back side (below)
The lasers of the Oxymon transmit light in the near infrared spectrum.
Usually, an Oxymon has two wavelengths about 850 nm and 760 nm. The
lasers of 850 nm are the upper lasers; the lasers of 760 nm are the lower
lasers. Two lasers with each a different wavelength forms a transmitter. If
the Oxymon has three or four wavelengths, the lasers are arranged in a
different way. In this case, the three or four lasers with each a different
wavelength forms a transmitter. Two lasers above each other form a
transmitter block, which is equal to a transmitter if just two wavelengths are
available in the Oxymon. Both lasers concerning to one transmitter block
have to be connected to an optical fiber before the lasers can be enabled.
The machine receives the light with an avalanche photodiode (APD), the
receiver. The lasers fire one by one, in that way the system can distinguish
between lasers. This principle is called the time-sequenced principle and also
works when multiple Oxymons are coupled.
The light is transmitted from the laser to the skin and from the skin to the
receiver by optical fibers containing glass fibers. A fiber ending connecting
to the Oxymon is called connector; a fiber ending at the skin is called
optode, transmitting to or receiving light from the skin. The optodes are
fixated in an optode holder (Figure 6-2). The distance between the optodes
is called the source-detector distance or the inter-optode distance. A
transmitter and receiver optode together form an optode combination or
channel.
39
Figure 6-2. Example of an optode holder. Optode holder with optodes to create a single
measurement point. The optodes stick out to get a good optical fiber-skin contact.
Some configurations work with split optical fibers, to reduce the number of
needed transmitters and/or receivers and the complexity of the optode
holder. A split optical fiber has two or more optodes. However, the optodes
of a split transmitter fiber have to be combined with optodes of two
different receivers (or vice versa); otherwise, the system cannot distinguish
the optode combinations by the time sequence principle. A split fiber will
reduce the amount of light transmitted through the fiber with about 50%.
Using a combination of a split receiver fiber and a split transmitter fiber will
reduce the amount thus to 25%. Cover the optodes of the split fibers,
which are not used during a measurement or calibration with the black
shields or cloths.
The power level of the lasers can be set to “standard" or “dimmed" which is
using about half of the power of the laser. Various types of optical fibers
are available with respect to length and width of the fibers. Specially
designed optical fibers that are suitable for simultaneous measurement
with NMR/MRI are also optional.
The system can measure at, 1, 2, 5, 10, 25 and 50, which is called the
sample rate or frequency. If the system is designed for 250 Hz, it is possible
to measure at 125 and 250 Hz as well.
40
The Oxymon can be switched on and off by the main switch. The power
indicator will turn green. The status lamp should blink green and orange
during starting up. After a while, it turns green and the device is ready for
data acquisition. If the status lamp continues blinking green/orange, please
contact Artinis. The measure lamp turns on if the lasers are firing!
The master Oxymon controls the slave Oxymons by the master-slave and
asynchronous cables. Therefore, the time sequence principle still works
using multiple Oxymons.
41
Make sure that the correct cables are available for connecting the cabinets.
The connector cables look like normal Ethernet (network) cables, but are
not the same. An asynchronous connector cable has two identical ends,
marked with an “A”. A master-slave connector cable has different ends and
it is important that the one marked “M” is connected to a master device and
the other end to a slave device. The USB cable is a standard USB cable
(Figure 6-3).
To change the setup by plugging in or out an Oxymon: please switch off the
power supplies, then turn them on again and restart Oxysoft.
42
other systems as the PortaMon and the PortaLite, in one Oxysoft package.
However, the Oxysoft version 3 or higher is limited to a combination of 8
receivers.
3 2 1 0
Oxymon master A
M M M
Oxymon slave 1A 1
*
S
Oxymon slave 2A
S
*
Computer A
Oxymon slave 3A
*
S
Oxysoft 3+
3 2 1 0
Oxymon master B
M M M A
Oxysoft 2.1.6
If full
Oxymon slave 2B
S
*
Computer B
*
S
43
Analog-Digital channels (AD Channels) can be applied to connect an Oxymon
to other devices and measure synchronized with these devices. Using the
input AD Channels give the possibility to analyze all data in Oxysoft. The
input of the AD Channels can be shown in Oxysoft with traces; when an AD
Channel is selected as “type” at the channel A tab. The channels are only
available if ordered!
Analog-Digital (AD) boxes give extra possibilities for the input and output of
data from the Oxymon. AD boxes are configured as input or as output box.
Important: If it is not possible to connect with the device in Oxysoft (or if the
inputs or outputs are not working), check your cables and connections and
see page 72. Make sure that you turn on the power of the Oxymon and the
AD box at the same time.
Using an AD input box instead of the AD channels of the Oxymon has the
following advantages:
- The signal of the AD Box is filtered at half the sample rate set in Oxysoft
for the Oxymon, which reduces the aliasing effects. The AD box has a
cutoff frequency with a maximum of 70 Hz.
- An input AD Box has 16 inputs, while an AD board has only eight
inputs.
44
6.4.1.1 Installation
The AD input box can be installed as a second Oxymon, using only the USB
connection on the AD box to the computer. The AD box needs to be
selected when making a connection to the Oxymon as shown in Figure 6-5.
Figure 6-5. Connect to an AD input box by selecting it before connecting to both systems.
45
6.4.2.1 Installation
The AD box has to be installed in combination with the Oxymon in this way:
- Connect the end marked “M” (master) of the cable to the AD box (in the
third position from the left, Figure 6-7).
- Connect the end marked “S” (slave) of a master slave cable to the
Oxymon (in the third position from the left).
- Connect the USB cable between the AD box and the computer (not to
the Oxymon!).
- Insert the power supply to the AD box.
- Turn on the power to the Oxymon and the AD box at the same time
(the status lamp and power indicator will turn green).
Figure 6-6. AD box – front side. Green lamp is turned on when the AD box is powered on. Blue
lamp is turned on during data acquisition, and blinks when AD output traces are being biased.
Bias button (press 1-2 s) will bias all the output signals
Figure 6-7. AD box – backside. The analog inputs or outputs (1-16) can be found on the backside,
and connectors for USB, Oxymon and power supply.
46
1 3 5 7 9 11 13 15
2 4 6 8 10 12 14 16
With the black button (bias button) at the front side of the AD box is
pressed for a couple of seconds, the output of the AD box is biased to 0
Volt. The Oxymon only measures relative concentrations (see paragraph
9.4.6), therefore the absolute value of the concentrations are not relevant,
only the relative changes.
Remark: A receiver gain or laser output change will result in a jump in the
voltage. The calibration of the Oxymon is not valid for the AD Box. Please,
set the gain and output before the start of the experiment and do not
change it during the experiment.
47
Table 6.1. Example of trace mapping of AD outputs
template 2 x 1 channel template 8 channel (split)
AD
output: Receiver: Transmitter: Transform: AD output: Receiver: Transmitter: Transform:
1 R1a - T1 O2Hb 1 R1 - T1 O2Hb
2 R1a - T1 HHb 2 R1 - T1 HHb
3 R1b - T2 O2Hb 3 R1 - T2 O2Hb
4 R1b - T2 HHb 4 R1 - T2 HHb
5 R1 - T3b O2Hb
template 4 channel square 6 R1 - T3b HHb
AD
output: Receiver: Transmitter: Transform: 7 R1 - T4b O2Hb
1 R1 - T1 O2Hb 8 R1 - T4b HHb
2 R1 - T1 HHb 9 R2 - T2 O2Hb
3 R1 - T2 O2Hb 10 R2 - T2 HHb
4 R1 - T2 HHb 11 R2 - T3a O2Hb
5 R2 - T1 O2Hb 12 R2 - T3a HHb
6 R2 - T1 HHb 13 R2 - T4a O2Hb
7 R2 - T2 O2Hb 14 R2 - T4a HHb
8 R2 - T2 HHb 15 R2 - T1 O2Hb
16 R2 - T1 HHb
template 4 channel cross
AD
output: Receiver: Transmitter: Transform: template 1 channel TSI Fit Factor + 2 x 1 channel (split)
1 R1 - T1 O2Hb AD output: Receiver: Transmitter: Transform:
2 R1 - T1 HHb 1 R1a - T1 O2Hb
3 R1 - T2 O2Hb 2 R1a - T1 HHb
4 R1 - T2 HHb 3 R1a - T2 O2Hb
5 R2 - T3 O2Hb 4 R1a - T2 HHb
6 R2 - T3 HHb 5 R1a - T3 O2Hb
7 R2 - T4 O2Hb 6 R1a - T3 HHb
8 R2 - T4 HHb 7 R2 - T4a O2Hb
8 R2 - T4a HHb
9 R1b - T4b O2Hb
10 R1b - T4b HHb
It is possible to have the TSI as output. For this you need to set the gain and
choose the AD output channel. This can be found under the TSI trace
properties (right mouse click on the trace) under the tab “transformation A”
(Figure 6-9. AD output signal gain settings.). Please be aware that the TSI
only updated 1/s and has 0.5 second delay.
48
Figure 6-9. AD output signal gain settings.
The PortaSync (Table 6.2) is a separate Bluetooth device which can be used
for generating an event signal. The system has two buttons, each creates a
distinct signal (2.5V and 5V).
The PortaSync also has a BNC input connection for analogue event signals
of other measuring modalities and a BNC output, to send the signal from
the buttons to any other measuring modality.
The PortaSync has the same Bluetooth range as the PortaMon and
PortaLite. However, the PortaSync must first be enabled for Bluetooth
before connecting. You can do so by pressing both buttons for 3 seconds.
The Event generator uses 2 AAA batteries, which last about 8 hours with
Bluetooth connection (blue LED). The orange power LED will blink when the
battery is almost empty.
Bluetooth has a delay of 100-200 milliseconds.
49
Table 6.2. Overview of PortaSync function
50
The Patriot device is a 3D digitizer made by Polhemus Inc.
(http://www.polhemus.com) and is supported to be used in conjunction
with the Oxymon through Oxysoft. The Patriot device is used to measure
the shape of the participant’s head, to align (coregister) it with a template
brain in order to determine the locations of the optodes relative to the
template brain. For more information on this coregistration procedure see
paragraph 9.2.8. The Patriot device consists of several parts that need to be
connected to the main station. See Figure 6-10 for an overview.
Figure 6-10 Parts of the Polhemus Patriot device. To the left is the power supply. On the top left is
the main station, to which the power supply, source and the sensors are connected. On the top
right is the source (grey cuboid) and the stylus (pen-shaped tool) constituting sensor 1. To the
lower right is sensor 2. On the lower left is the headband, to which sensor 2 can be mounted and
two adhesive, foam inlays to increase wearing comfort.
51
position, its button has to be pressed. Oxysoft then registers the current
position and automatically continues to the next step (see paragraph 9.2.8).
The headband including the second sensor is intended to be worn by the
participant. Data from the second sensor is then used to account for small
head movements during the digitization process. Consequently, the source
has to be connected to the Source port of the Patriot main device, the
stylus has to be connected to the Sensor 1 port of the Patriot main device
and the headband sensor has to be connected to the Sensor 2 port of the
Patriot main device.
52
Oxysoft is the program which can be used to record, manage and analyze data
measured by the NIRS device(s). This chapter illustrates how to work with
Oxysoft in order to handle data. The data acquisition is described in chapter 8.
The menu contains several submenus, like “File”, “Measurement” and “Data
Collection”. If one of these is clicked, the submenu will open. Each submenu
contains functions to control the program. In the toolbar all frequently
used functions are available as buttons for a quick access. The toolbar can
be customized by right clicking on the bar and selecting “Customize…”.
The DAQ value view and DAQ status view are important during the data
acquisition. In the DAQ status view the status of the NIRS device(s) is
displayed. In the DAQ value view it is displayed how much light is measured
by the NIRS device(s). The status bar shows the current state of the NIRS
53
device(s). The status bar displays the value in the graph at the mouse
pointer and displays if the data is acquired (DAQ) or data is recorded (REC).
The mouse pointer displays information about the trace that is under the
mouse point.
When you drag a window small boxes appear on each side of the screen as
well as a cross in the middle. If you drag a window on one of these figures,
the window will be fit to the chosen side.
54
The acquisition program “Oxysoft” can be started by clicking on the
shortcut on the desktop or by the Windows start button, “All
programs”, “Artinis Oxymon”, “Oxysoft X” or by going to the program
directory and click on “Oxysoft.exe”.
Oxysoft can be closed by Menu File Exit. First, save the project to save
all measurements and graphs if needed.
Information about the used Oxysoft version and options is in Menu Help
About Oxysoft (Figure 7-2): the copyright, license, version and contact
information. The license document “040303 AMS Software License
Agreement.rtf” is available in the Oxysoft installation folder.
55
If the software is not available with the (correct) license key plugged in the
USB of the PC, close Oxysoft and restart it. The license key has to be
plugged in before Oxysoft is started. If the software is still not working,
please try the following:
If Windows 7 or 8 is used, it can be necessary to first run (or use are repair)
a driver installation manually for the dongle. The drivers can be found on
the CD/USB with software. It could be possible the driver does not work
immediately. If not; go to the Windows start menu and go to devices and
printers. Here you will see the ROCKEY. Right click and go to properties. Go
to tab hardware and click properties and update the driver.
The data can look different in newer Oxysoft versions. The data may look
different in different versions of Oxysoft due to different offsets, but most
probably, the settings have changed. Check the settings, which have an
immediate impact on the graphical displaying. For example, “source-
detector distance” and “DPF” in the measurement properties and “bias” in
trace properties.
56
Figure 7-3. Program options – Data collection
Startup options
Start recording manually: Check if the recording has to start manually. If not
checked, the recording will start with the start of data acquisition. Bias
traces at start: Traces are biased at the start of the measurement. For more
information about biasing traces, see paragraph 9.4.6.
57
Calibration
Select all gains high for TSI calibration to set the receiver gains to high
before the TSI calibration starts, also if you have set the receiver gains too
low for the measurements. We strongly recommend to use high gains
during calibration.
Graph font
The font used in the graph for labels, legends, etc. in graphs.
Editor font
The font used for the DAQ Value view.
Background color
Click on this button to change the background color of the graph.
Text color
Click on this button to change the text color of the graph.
58
Show legend
When checked, a legend of the traces is shown in the right of the graph.
This tab (Figure 7-5) defines the Measurement template file path selecting
the file which defines standard templates used to create measurements,
graphs and traces from a template. The template files are called *.oxypt.
59
After creating/opening the project, create a measurement and name it like
the desired template. Make all graphs and traces and arrange the graphs
as desired. Save the project with the name of the desired template file and
close Oxysoft. Browse to the folder were the project is saved, copy and
paste it. Rename the project from “*.oxyproj” to “*.oxypt”. Start Oxysoft and
change the template file in Oxysoft in Menu View at the tab “Template
file” to the new template file.
The DPF (differential pathlength factor) table can be used to correct the
DPF for its wavelength dependency. The table can be selected by clicking
on “Browse”. The DPF tables are saved as “DPF_*.xml” file in the Oxysoft
directory in the folder “Tables”. The default table is “no correction”, which
we recommend for most applications.
60
This chapter describes how to do a measurement with Oxysoft and which
parameters are important during the measurement. Take care of the
measurement quality with the DAQ value. The last paragraph describes how to
calibrate an Oxymon, in the other paragraphs is assumed that the system is
calibrated.
Like in every scientific set-up you should, in advance, have an idea of the
questions you want to solve as it determines the design of the protocol.
Concerning NIRS measurements, it means that you must choose the task
(intervention), the anatomic area of interest, and decide whether you prefer
a qualitative or quantitative approach. Especially the latter has distinct
consequences for the protocol as it can only be done under specific
conditions.
Good preparation before data acquisition will save time and increase the
accuracy of the measurements. We advise to create all desired graphs
(paragraph 9.3) and to add all necessary traces (Paragraph 9.4) to the
graphs before you start the measurement. Graphs and traces can also be
made during and after the measurement. For instance, when you program
your expected and possible unexpected occurrences (e.g. “movement
artifact”) prior to the measurement, you will have the hands and mind free
to react to other things that come up. Accurate and frequent instructions
for the subject are also essential for a good result. Especially instructions
when not to move can be very important for the determination of
quantitative variables afterwards, e.g. at the start or after cessation of an
occlusion. Movement artifacts can easily obscure the data and rule out the
possibility of quantitative calculation, but can mostly be avoided with
adequate instruction.
1. Put the power supply cable into the Oxymon(s) (paragraph 6.1).
2. If necessary: Connect the Oxymon to other Oxymons (paragraph 6.2)
3. Put the USB cable to the USB port of the Oxymon (paragraph 6.1). If an
AD box is used, see paragraph 5.4 how to connect the Oxymon in
combination with an AD box.
4. If applicable: Connect the AD cables to the AD board or box (paragraph
6.3)
5. Connect the optical fibers to the Oxymon(s) (paragraph 8.2)
6. Connect the optical fibers to the optode holder(s) (paragraph 8.2).
Measure the distance between the optodes of each holder to
determine the source-detector distances.
7. Turn the Oxymon(s) on by the main switch (paragraph 6.1). Wait until
the status lamp is green (paragraph 6.1).
8. Start and create a measurement in Oxysoft (paragraph 8.2) and set
DPF, D(mm), (delta d(mm), h and k for TSI templates) and appropriate
sample rate
9. Warm up the machine for about 20 minutes for a TSI calibration (if
applicable) (paragraph 12.5)
10. Make graphs (paragraph 9.3) with traces.
11. Fixate the optode holder to the subject (paragraph 8.3)
12. Fixate the optical fibers to the subject and the environment
13. Start the data acquisition (paragraph 8.4). Check the ‘Measure’ lamp at
the Oxymon (paragraph 6.1).
14. Check the DAQ status view before the start of the experiment
(paragraph 8.5)
15. If necessary, adjust the gain settings in Oxysoft to get a good
measurement quality (paragraph 8.7)
16. If all settings are set right, bias the traces and start recording.
17. Indicate the start of the measurement with an event (paragraph 10.1)
for two reasons. 1: to indicate the start, 2: to be sure the data is being
recorded. If the event cannot be inserted, make sure you have started
the recording (paragraph 10.1.2).
18. Check the DAQ value view or Template DAQ state during the
measurement to keep a good measurement quality (paragraph 8.7)
19. Insert events during the measurement if needed(paragraph 10.1)
20. End the data acquisition (paragraph 8.4)
21. Analyze the data (chapter 10)
8.1.2.1 Problems
If no changes are visible during the DAQ in the traces, the traces are flat,
zero or gone, there can be several reasons behind this observation. If
analyzing data make sure that there is:
- attached a data file with data (check the file information (paragraph
9.2.3))
- the ‘measure’- lamp at the Oxymon is turned on (paragraph 6.1). If not,
restart Oxysoft, connect to the Oxymon and start the measurement
62
NOT by pressing the recording button, but the green start button or via
the data collection measurement wizard.
- filled in the correct inter-optode distance and DPF (paragraph 9.2.4)
- the correct light source-to-optode assignment (paragraph 9.2.4)
- a correct graph scale and zoom or set to auto scale (paragraph 9.4 and
9.3.17)
- done ‘bias all traces in measurement’ to get better offsets (paragraph
9.4.6)
- non-filtered traces (paragraph 10.5)
In Figure 8-2 it is illustrated how the fibers should be connected if only two
wavelengths are available in the device. Connect the first fiber to both
lasers of Tx1; connect the second fiber to both lasers of Tx2, etc. Usually,
an Oxymon has two wavelengths about 850 nm and 760 nm. The lasers of
850 nm are the upper lasers; the lasers of 760 nm are the lower lasers.
Two lasers with each a different wavelength form a transmitter and have to
be combined into one fiber. If the Oxymon has three or four wavelengths,
the lasers will be arranged in a different way. In this case, the three or four
lasers with each a different wavelength forms a transmitter and will be
connected to one fiber. Two lasers above each other form a transmitter
63
block, which is equal to a transmitter if just two wavelengths are available in
the Oxymon. Both lasers concerning to one transmitter block have to be
connected to an optical fiber before the lasers can be enabled.
Figure 8-2. How to connect the fibers to the Oxymon. Connect both connectors of a fiber to one
transmitter block!
It is also possible to connect the fibers in other ways, but only do this if you
have more than two wavelengths (see below) or if you are sure that it is
right. Please contact Artinis before you do this! If you have connected the
fibers in the wrong way during a measurement, the data of the
measurement might be useless. Always check (and change) the lasers to
optode mapping in the measurement properties if you are using another
way of fiber connecting.
If three different wavelengths are used for one transmitter, an optical fiber
with three connectors and an empty connector (a connector without the
fiber) has to be used. This empty connector is necessary because the
transmitter blocks only emit light if both lasers are connected. The first two
wavelengths at the first transmitter block (Tx1) are enabled by making the
connection to the three-ending fiber. The third laser at the second
transmitter block (Tx2) is not enabled by connecting the three-ending fiber,
because the fourth laser is not connected to a fiber. To enable the third
laser, connect the empty connector to the fourth laser and fasten the
screw. If four different wavelengths are used for one transmitter, use an
optical fiber with four connectors and connect those to the lasers.
If three or four wavelengths are available, also only two wavelengths can be
used for a measurement. If both desired wavelengths are at one
64
transmitter block, connect these lasers to an optical fiber with two
connectors. If the desired wavelengths are at two different transmitter
blocks, connect the desired lasers to the optical fiber with two connectors
and connect the other lasers to empty screws.
Screw Optode
(holes) holes
Fixation
frames
Figure 8-3. One Channel optode holder. The optodes stick out a bit at the lower side of the holder
The fixation of the optode holder can be done just before starting the data
acquisition. Prepare the optode holder in the following way:
1. In case of hair on the skin:
2. A little hair will not influence your measurement.
3. Shave the skin if the fixation with tape is a problem or use the straps.
4. Dark and much hair between the skin and the optode can diminish the
measurement quality. Part the hair if necessary and place the optodes
at the parting.
65
5. Determine which optode holder is necessary for the optode
configuration (or optode holders)
6. Clean the optodes, the holder and the skin with alcohol
7. Determine the optode distance between the receiver and transmitter
optodes and choose applicable holes for the optodes
8. Stick the double-sided electrode stickers around the chosen holes at
lower side of the holder and place it at the region of interest (Figure 8-4
left). We advise to use extra (medical) tape for better skin contact
(Figure 8-4 right).
9. Put the optodes in the chosen holes of the holder. Notice the distance
between the optodes. To get a good skin contact when applying the
holder to the skin, it is better that the optodes stick out a little bit from
the holder at the lower side. Light leaking between the optode and the
skin will cause undesired reflections and light lose. If the optodes stick
out too far, it will hurt the subject.
10. Fasten the screws (in the screw holes) in the optode holder with a small
screwdriver until the optodes are fixated.
Figure 8-4. Apply optode holder to the skin using e.g. double sided sticky discs. Apply extra
medical tape for better skin contact
11. The holder (and optodes) should not move during the measurement.
However, a too tight fixation of the holder can hurt the subject or
obstruct the blood flow. Especially during exercise, caution has to be
taken that no friction on optode-skin contact arises because of
movements of the fibers. Avoid movement of the measurement volume
by accurate support of optical fibers. The optical fibers can be fixated
to the environment (like a table, chair or bike) with tape and to the
clothing of the subject with the supplied clip to prevent for movement
artifacts. Weight of the fibers on the holder itself can be reduced by
making a strain release with extra tape (Figure 8-5)
66
Caution should be taken with the placement of optical fibers and body
parts under investigation, in order to obtain sound results. Take care that
circulation is not impaired due to incorrect positioning of the subject. For
instance, in case of lower arm or lower leg, avoid contact of the
measurement site with the ground to prevent circulatory obstructions
other than those which are controlled (e.g. occlusion). Avoid venous pooling
of the blood in the limb, possibly by placing the limb at heart level and in an
o
upward angle of ±30 .
More general: take care that the subject is comfortable in his/her position
and able to relax, as unwanted muscle tonus must be avoided as much as
possible.
Covering of the optode holder from environmental light may improve the
measurement quality (especially when measuring TSI).
Figure 8-5. Apply a strain release (in top) to reduce effect of movement of the fibers on the
holder.
The lasers are firing by clicking on the green start button in the toolbar or
by Menu Data Collection start device. Data is recorded automatically or
by clicking on or by Menu Data Collection start recording. If the
lasers are firing, the measure lamp at the Oxymon will blink red. The data
67
acquisition ends by clicking on or by Menu Data Collection stop
device.
Watch the device status of all devices during the measurement (open by
Menu View DAQ status view, Figure 7.7). If an error or warning occurs,
try to find out what causes this error or warning. If the cause is unclear,
contact Artinis Medical Systems and do not continue the measurements. The
warning “TX 2/3/4 Fiber not connected” is a normal warning and means that
no optical fiber is connected to this transmitter block.
The clock as shown in Figure 8-7 displays the duration of the recording
(00:50, min:sec). The file size is the size of the saved *.oxy3 file. “00:23 since
A” means that 4 seconds are expired since the last event, which was event
“A”.
68
The Template DAQ State shows the used optode template belonging to the
devices and warns with red marks if one of the optode combinations
receives too less (See Rx1 - Tx4 and Rx2 – Tx4 in Figure 8-8).
Figure 8-8. Template DAQ state showing the used optode template with optode combinations
out of range (Rx1 - Tx4 and Rx2 - Tx4)
Monitor the DAQ value view (open by Menu View DAQ value view,
Figure 8-9) during the measurement to control the percentage of received
light. The value view describes:
- the percentage of received light per optode combination
- the wavelength of each laser
- the used laser output level (dimmed or standard)
- the used receiver gains (low or high)
The DAQ value view shows all possible optode combinations, independent
from the chosen optode template and belonging optode combinations.
The first row shows the corresponding device. For one stack of Oxymons
the device numbers are shown in the order of master, slave 1, slave 2 and
slave 3.
The second row shows the wavelengths of the Oxymon per laser. The
wavelengths can be a little bit different from the ordered wavelengths. In
case a stack of Oxymons are used, the lasers of the slaves are numbered
successive to the master lasers.
69
Figure 8-9. Data Collection value view. All laser outputs are set to standard. Receiver 1 is used
with high gain, receiver 2 with low gain. The light of L1 to L4 is not received by R2, the value view
warns with red fields. If light is not received, but is also not included in the measurement
template, the will not turn red (L5 to L8 for R2). Switch between gains (“low” and “high”) by clicking
on the gain field per receiver-laser combination above the DAQ value or for all lasers for one
receiver by clicking on the ‘select’. The laser output power (“off”, “dimmed” and “standard” can be
changed in the same way by clicking on the power field.
The third row shows the power level of the lasers, which are all set to
“standard” in Figure 8-9. Switch the power level of a laser to “off” or
“dimmed” (for less laser power if “standard” saturates the receiver) by
clicking on the belonging power field in the DAQ value view.
The fourth row shows the gain of receiver 1 (R1) for each laser, which are all
“high” in Figure 8-9. Switch the gain of a receiver to “low” by clicking on the
belonging gain field in the DAQ value view. The sixth row shows the gain of
receiver 2 (R2) for each laser, which are “low”. In case a stack of Oxymons is
used, the receivers of the slaves are numbered successive to the master
receivers (R3, R4, etc.).
The fifth row shows the percentage of received light in percentages for the
optode combinations belonging to receiver 1. The percentage of received
light is displayed in percentage (“Perc”) or in AD counts (“Raw”); switching
between those can be done by clicking in the left field of the first line of the
DAQ value view (in Figure 8-9 “Perc”). The percentage is between 0-100%
and the counts between 0 and 65535, where 65535 counts equals to
100%. If a DAQ value field is red-colored, the percentage of light is out of
range, it is too low or too high to get a high quality measurement. Set the
limits for out of range in the program options the limits for out of range
(paragraph 7.6).
If the percentage is higher than 75% the DAQ value field becomes red.
When the percentages get higher than this value the measurement is
unreliable due to non-linearity.
If the percentage is close to 0%, the tissue absorbs (almost) all light. Try to
remove dark hair. Most probable the light of one transmitter is completely
70
absorbed. Check if you can see the heart beat in the signals or something
else you are expecting, see paragraph 8.7 for more information.
The seventh row shows the percentage of received light in percentages for
the optode combinations of receiver 2.
The settings during data acquisition are important for the quality of the
measurement. An important factor for the quality is the percentage of
received light. If a field in the DAQ value view is red, the measured signal is
out of range. Change the settings to improve the signal and the
measurement quality as described in the next paragraphs. The device
settings are the sample rate, the light source power level and the receiver
gain. The sample rate cannot be changed during the measurement. The
right settings are dependent to the individual subject, measurement and
need. Therefore, these have to be set for each measurement. It is handy to
insert an event after adjustment of the settings. However, we advise to
choose the settings before the start of the experiment to have a good
quality during the whole experiment!
The gain defines the degree of amplification of the received signal and can
be “low” and “high”. Older systems have also a “medium” gain setting. The
gain of the system can be changed per optode combination.
71
The effect of using a low or high gain is shown by an example. Using the
low gain, the percentage of received light is 4% (Figure 8-10), which is about
2600 counts. This is a good value for measuring; the values should be
between 150 and 50000 counts. If the counts are too low, the DAQ value
view will warn you by a red field. The counts can be increased by using high
gain. This results in an increase of the percentage (to 34% and 35%), as
illustrated in Figure 8-11. This figure shows the same measurement as
Figure 8-10.
However, the concentrations measured with low and high gain are equal!
The system compensates for the gains by using the calibration information.
The adjustment to the high gain takes some time, which gives the short
spike at 24 sec in Figure 8-11. Because of this spike and the “never-perfect”
calibration, we advise to set the gain before the start of the measurement
and not changing it during the measurement. For the same reason, we
advise to use the auto-gain function only before the measurement to see
what the best gain is for the measurement.
Care should be taken when using the highest gain setting: if environmental
light comes onto the optical probes or detector, this may increase the
detected intensity value. This will only be problem if the signal measured is
very weak, for example when a large source-detector distance is used in
brain measurements. It might help to reduce the ambient light in the
measurement area, and to shield the fiber of the transmitter.
Figure 8-10. Example of using low gain at a 1 channel measurement (Rx1- Tx1). Receiver 1
receives 4% of the transmitted light of laser 1 and laser 2. Laser 3 and 4 are not used.
72
Figure 8-11. High Gain, same measurement as Figure 8-10. The gain is switched from low to high
around 24 seconds, which is visible by a short peak. This short peak is because of the adjustment
to the high gain takes some time. Receiver 1 receives 34% and 35% of the light of respectively
laser 1 and laser 2.
Another phenomenon may occur when there is too much light on the
detector from the lasers at the highest gain. Saturation in this gain setting
may cause compression of the detected signal, resulting in a lower value
than could be expected. If suspicious, lower the gain to see what happens.
If nothing changes in whatever channel it is best to restart the Oxymon and
the Oxysoft.
During the gain switch it is possible to get a spike in the signals, therefore it
is recommend to switch not gains during the (important parts of the)
measurement.
Watch during the whole measurement the DAQ value view (paragraph 8.7)
to see if the percentage of received light is not out of range. If the
percentage of received light is too low or too high, try the following:
- Check the power level of the laser. The power for every laser has to be
set to off (no output), dimmed (low output) or standard (high output).
73
- Check the gain. The gain of the receiver can be set at low gain and high
gain (older systems have also the option medium gain). High gain
amplifies the received light more than low (or medium) gain.
- Check the fixation of the fibers to the skin. Are all fibers connected in
the right way? Is the optode holder still fixated? Are the optodes or is
the skin polluted? Is there hair between the optodes and the skin?
- Are all fibers still in the machine?
- Diminish or increase the source-detector distance in the optode
holder. A smaller distance will increase the percentage of received light.
- Dark skin and hair color can diminish the percentage of light received.
The date of the last calibration is in the calibration file. Go to the Oxysoft
installation directory, Folder “Device” and open the calibration file. Open it
with internet explorer; the calibration date is marked as “timestamp”.
74
- Connect an optical fiber to receiver 1 and connect a fiber to laser 1.
(Laser 1 only transmits light if laser 2 also is connected to an optical
fiber.)
- Take the calibration tool (Figure 8-12). The receiver and a transmitter
optical fiber can be plugged in the two openings at the endings of the
tool. Use the small screws to fasten the optical fibers in the calibration
tool.
- Connect the Oxymon by USB to the PC and start the Oxymon.
- Start Oxysoft. Go to Menu Data Collection Calibration. Follow the
procedure of the calibration as described in the window. The large
screws of the calibration tools can change the percentage of received
light.
- Save the results when the calibration is ready! The calibration file is
saved in the Oxysoft directory in the folder “device” as *.xml file.
Small screw
Large screw
13
75
This chapter describes how to manage your data in Oxysoft: How to make a
new measurement and how to display, save and export the measured data.
There are examples available in the folder “Data Examples” in the installation
folder.
Open a new project by Menu File New project file. The project view will
refresh and show the new project. The name of the project is automatically
set to “New Project”. It can be changed by saving the project under another
name. The first time that Oxysoft is used, a new project without any
measurements is shown in the project view.
Open an existing project by Menu File Open project file. Project files
have the extension *.oxyproj. Measurements can be added and changed.
Use Menu File to quickly select the last used projects.
Figure 9-1. Project properties
Save a project by Menu File Save Project file as or Menu File Save
Project File. The latter will overwrite the last saved version of the current
project. If the project is not saved before closing Oxysoft, the project
settings will be lost. In that case, the measured data is not lost, because it is
always saved in the *.oxy3 files.
Show the properties of the project (Figure 9-1) by Menu File Project
Properties or by right mouse clicking on the project in the project view and
selection of “Project properties” or by double clicking at the project in the
project view, or by selection of the project in the project view and clicking
on the toolbar button . The project properties define the used
coefficients: extinction coefficient table, the DPF table and the project
settings.
9.1.4.1 Coefficients
The extinction coefficient table of this project can be selected by clicking on
“Browse”. The extinction coefficient tables are saved in a *.xml format in the
Oxysoft installation directory in the folder “Tables”. You can also add your
own table. The default table is from Cope [Cope 1991]. The default used
tables can be selected in the program options (paragraph 7.6).
The DPF (differential pathlength factor) table can be used to correct the
DPF for its wavelength dependency. The table can be selected by clicking
on “Browse”. The DPF tables are saved as “DPF_*.xml” file in the Oxysoft
directory in the folder “Tables”. The default table is “no correction”, which
we recommend for most applications.
77
Close a project by Menu File Close Project File
78
Figure 9-2. Create new measurement
Name
This specifies the name of the measurement. This name is for own
reference and will be shown in the project view.
Data file
This specifies the data file and the saving location. The file (*.oxy3 or the
older *.oxy/*.evt combination) is the primary source of data for all graphs
and calculations accomplished in this measurement. A standard name is
given as the copy of the measurement name. The field itself is not editable,
but can be changed by pressing the browse button that opens the file
browse dialog. Click on browse, search a location to save and type a name
for your file. Older files cannot be overwritten.
We strongly advise to save the data to the local computer. Oxysoft does not
warn for lost data if the data is lost during transmission over a network to
save the data at another computer.
Description
The description field is a free text field reserved for your own comments.
Quick start
This will skip the next screens if you are sure that no other settings needs
to be adjusted, compared with the copied settings. After selecting the
device (Figure 9-3) the device will be ready to start the measurement.
79
Next, Oxysoft will try to make a connection to the device(s) (Figure 9-3).
Enable the system you want to use. If only one system is present, it will
automatically connect. It is possible to connect multiple systems (also
Oxymon and portable systems) up to a maximum of 8 receivers or 7
portable systems. Right mouse click will give the question to deselect all.
“Save Selection” will save the selected devices for a following connection.
New devices can be added in the file in C:\Users\Public\Documents\Artinis
Medical Systems BV\common\PortaSoft.ini by copying another (similar)
system and change the number and Bluetooth address of the wireless
devices.
Figure 9-3. Connect to a device screen. Enable the system you want to use. If only one system is
present, it will automatically connect.
Optode template
This specifies the predefined optode template to use for this
measurement. The choices of templates may be simplified by filtering the
list. If the desired template cannot be found, make sure that the category is
not already filtered out. The selected optode template is shown in the
figure on the right.
Filter
When clicked, a dialog pops up that allows filtering the list of optode
templates by category: single channel, multichannel, TSI, PortaMon and
PortaLite.
80
Subtemplate
The subtemplate data grid is designed to define parameters for calculating
concentrations per subtemplate. A template may consist of several
independent subtemplates. Each subtemplate is parameterized
individually. The following parameters can be set:
- DPF: Differential pathlength factor. Choose DPF according to literature.
DPF can be changed before and after the data has been recorded and
it will influence the graphical data representation of the concentration
changes. Some reference values can be found in Table 4.2.
- d (mm): source-detector distance or nominal distance of the template
grid. Change d by clicking on the “35” and choosing another value or
typing in that field. Measure the distance at the optode holder. The
value for source-detector distance can be changed before and after the
measurement.
81
This formula is valid for measurements on BRAIN tissue for 17-50 year old
human subjects, derived from the data of Duncan et al. [1996]. Fill in the
age, click on “Copy to Clipboard” and “Return”. The window will close. Paste
the DPF in the DPF field by right mouse clicking in the field and “Paste”.
Restore
This function will restore the measurement and optode template settings
to the settings used during recording.
82
Figure 9-6. Laser to optode mapping
Laser 1 and 2 are in transmitter block 1 (Tx1 at the Oxymon), laser 3 and 4
are in transmitter block 2 (Tx2 at the Oxymon), laser 5 and 6 in transmitter
block 3, laser 7 and 8 in transmitter block 4, laser 9 and laser 10 in
transmitter block 1 of the second Oxymon (Tx1 at the second Oxymon), etc.
If you have to change the mapping, make sure you have connected the
optical fibers in the same way as you configure this mapping (see
paragraph 8.2)!
After choosing the right optode mapping, click on “Next”.
83
Sample rate
This specifies the available sample rates of the system. The system can
measure at, 1, 2, 5, 10, 25 and 50, which is called the sample rate or
frequency. If the system is designed for 250 Hz, it is possible to measure at
125 and 250 Hz as well.
Power
This specifies all the laser output level of system. A laser can be turned
"off", "standard" or "dimmed” which is using about half of the laser power.
Please be aware that split fibers can only emit one output power, so do not
use different powers one different subtemplates with the same split
transmitters.
Subtemplate
The laser output levels are selected for all lasers arranged according to the
subtemplates.
Optode
The power levels are selected for all lasers arranged according to the
optodes.
Laser
The power levels are selected according to a single laser.
After choosing the device settings, click on “Next” to get further options
(Figure 9-8). If the measurement has to start after finishing the wizard,
select “Start measurement after finishing wizard”. If the measurement has
to start later, select “Do not … for data-acquisition”. Finish the wizard by
clicking on “Finish”.
Next, Figure 9-9 will come up to help you to create some graphs
automatically. It will create graphs for all measurement channels with the
84
selected traces. For more information about graphs and traces, see
paragraph 9.3 and 9.4. After finishing the wizard, the new created
measurement (and graphs) is available in the project view.
85
Figure 9-11. Create a measurement
If you have very old or downloaded (from portable Artinis systems) *.oxy
files Click on drop down menu to switch between *.oxy3 and *.oxy, select
the file and click on “Open”. Press “Apply” to transform the data of an *.oxy
file to *.oxy3 file and to see the data file info as described below. Click on
“OK” to close the window and make graphs to display the data. The *.oxy3
file made by transformation of an *.oxy file will be named like the *.oxy file.
Opening an *.oxy file with the same name as an existing *.oxy3 file (in the
same folder) will open the already existing *.oxy3 file instead of the *.oxy
file. To open the *.oxy file, rename the *.oxy file (and the *.evt file) or the
existing *.oxy3 file or move it to another folder.
86
9.2.3.1 Data file info
This is a read only field with information about the data recording stored in
the data file (*.oxy3). The available information is the used version of
Oxysoft, name of the data file, recording date, system information,
measurement settings and the used wavelengths (Figure 9-12). Clicking on
“Apply” will refresh the information.
87
can either use a Polhemus digitizer, or you can manually determine the
optode positions using a 2D topographic plot.
1
http://www.bic.mni.mcgill.ca/ServicesAtlases/ICBM152NLin2009
88
Figure 9.9-14 The five anatomical landmarks (fiducial points) on the human head used in Oxysoft
for digitization and coregistration. The nasion is defined as the valley on the nasal bone. The
inion can be located by searching for the occipital skull bone at the back of the head and using
the lower bulging of this bone. The left and right preauricular points are located just above the
upper end of the tragus of the ears. The vertex is located right exactly in the center of the
circumferential line between nasion and inion and in the center of the circumferential line of the
two preauricular points.
89
Figure 9.9-15 Digitization and coregistration procedure of the an OctaMon optode template. Rx1
has already been digitized and the view automatically rotated to the digitized position. Next, Rx2
has to be digitized, as it is the highlighted optode in green on the right in the optodetemplate.
After having digitized all optodes of the optodetemplate, all 3D topo views
of that measurement will use these digitized and co-registered optode
positions. If such digitized 3D positions are available, an inverse
transformation and projection will be performed to calculate the 2D
positions in the 2D topo view (for more information on different view
modes, see paragraph 9.3.4). This is done by coregistering the 3D fiducial
points to the fiducial positions of the standard 10-20 system2. Note that
due to the use of an anatomical template model, the shown locations of
activation do not allow for any conclusive statements about the true
location of brain activity during the measurement.
2
Jasper, HH (1958). "Report of the committee on methods of clinical examination in
electroencephalography: 1957”. Electroencephalography and Clinical Neurophysiology 10
(2): 370–375. doi:10.1016/0013-4694(58)90053-1
90
three coordinates that the program shows change significantly, the
current setup is not working well, and the position of the source
has be changed.
Graphs show the measured data by traces. The graph is the window; the
trace is for instance the O2Hb concentration measured by one optode
combination. The available graphs are:
- (regular) Graph: Time view (paragraph 9.3.1)
- Topo Graph: Spatial view (paragraph 9.3.4)
91
- Spectrum Graph: Frequency spectrum (paragraph 9.3.5)
- Indicator view: Numerical view (paragraph 9.3.6)
A regular graph is a time view of the measured data. The time is plotted
along the x-axis in seconds and the measured signal is plotted along the y-
axis. The scroll bar in the graph can be used to show the concentrations
over time. Graphs with the icon are regular graphs. A regular graph can
be created in several ways:
- Menu Graph Create graph or Create All Graphs or Create graphs
from Optode Template.
- Right-mouse clicking the project view on the measurement and select
“Create Graph”, “Create All Graphs” or “Create graphs from
Measurement Template”.
- Via “Create all graphs” at the end of the DAQ measurement wizard.
The function “Create Graph” creates an empty graph without traces. “Create
all graphs” creates graphs for the optode template with traces for the
desired concentrations according to the optode template (chapter 11).
“Create Graph from a template” creates graphs from a measurement
template (paragraph 7.6.3). This measurement template shows the data in
a clarifying way.
This paragraph describes how to create a graph with the function “Create
Graph” (Figure 9-16).
92
9.3.1.1 Graph properties
The properties in this tab affect the name of graph, event display mode,
coloring and comments. More graph properties can be found in Menu
View Options, tab “Graph Options”.
Graph Name
This specifies the name of the graph.
Show events
To show events in the graph, set this to 'show events'. For more
information about events, see paragraph 10.1. In case of cyclic data the list
of options is expanded with traces based on cyclic data (paragraph 10.3).
Selecting one of these, events will be projected on the timeline of that
trace. To show/hide events as default in any graphs, use the program
options use the program options (paragraph 7.6).
Show Legend
When checked, a legend of traces is shown to the right of the graph (Figure
7-1). To show/hide as default the legend in all graphs, use the program
options (paragraph 7.6).
Background color
This button changes the background color of the graph. To change the
default background color, use the program options (paragraph 7.6).
Text color
Click on this button to change the text color of the graph. To change the
default text color, use the program options (paragraph 7.6).
Comments
The comments field is a free text field reserved for extra information.
93
9.3.1.2 X-axis
The properties in this tab affect scaling settings of the time axis (Figure
9-17. Graph properties – the x-axes). During data acquisition the x-axis is
set by the time span settings in the program options.
Auto scale
If checked, the time scale is set to automatically: the axis will start at the
minimum time of traces and end at the maximum time.
Manual scale
When checked the manual scale settings are used instead. The scale is set
by:
- Minimum: The axis will start at the minimum value entered here. Fill in
the minimum.
- Time: “1”, “10”, etc. The axis will start at 1 or 10 seconds.
- A unique event label: “A1”, “B2”, etc. The axis will start at event “A1”
or “B2”.
- An event label: “A”, “B”, etc. The axis will start at the first event “A” or
“B”.
- A combination of time and event labels: “A+10”, “B-10”, etc. The axis
will start 10 seconds after event A1 or 10 seconds before event B1.
(Warning: “10+A” does not work). The code “<” smaller than and “>”
bigger than: “A>100”, “B<500”. The periods will start with the first
event A after 100 seconds or event B before 500 seconds (handy to
define the maximum of the x-axis).
94
- The code “|” OR: “A1|A3”. The axis will start at the first of A1 and A3.
The input “A1|A3>100” will start at the first of A1 and A3 which is larger
than 100.
- The “!” operator subtracts the given set from the set containing all
events. For example: !U2 takes all events except U2 or !U2 & !U4 takes
all events except U2 and U4 Note the equivalent is !(U2|U4).
- Maximum: The axis will end at the maximum value entered in this field.
The maximum should be larger than the minimum. See the minimum
for how to define the maximum.
- Ticks: Enter the axis-tick interval in this field. Leave 0 to let the program
decide for a reasonable value.
9.3.1.3 Y axis
The properties in this tab (Figure 9-18) affect scaling settings of the value
axis
Auto scale
If checked, the value scale is set to automatically: the axis will start at the
minimum value of traces and end at the maximum value and some margin
will be included. During data acquisition, the scaling will adapt the y-axis
dynamically in order to show the current values. The minimum range of the
scale can be set by:
- Minimum range: When in auto scale mode, the range of the scale will
not be smaller than this value.
Manual scale
When checked the manual scale settings are used instead. The scale can
be set by:
- Minimum: The axis will start at the minimum value entered.
- Maximum: The axis will end at the maximum value entered. The
maximum should be larger than the minimum.
- Ticks: Enter the axis-tick interval. Leave “0” to let the program choose
for a reasonable value.
- Unit: Enter the unit of the y-axis. This unit will be shown along the y-axis
-6
and when the y-axis value is reported: Relative concentration (µM, 10
mol/l), AD Input Voltage (Volt), TSI (%)
95
checked, the factorization is suppressed and the values are shown in an
unscaled way, at the expense of a wider space reserved for the y-axis
numbers. To suppress scale factorization in all graphs, change the program
options (paragraph 7.6).
9.3.1.4 Gridlines
This tab defines if any gridlines is to be shown in the graph (Figure 9-19)
Gridlines X-axis/Y-axis
If checked, gridlines on the x-axis/y-axis are shown.
This function will remove all earlier created graphs from the measurement.
Therefore, a warning (Figure 9-20) is given.
96
clicking in the project view on the measurement and selection of “Create
graphs from a template”. This function will remove all earlier created graphs
from the measurement. Therefore, a warning (Figure 9-20) is given.
A nice example is in the folder “Sample” in the Oxysoft directory. Open the
project “Measurement examples.oxyproj” and open the available
measurement.
97
Figure 9-21. Topo graph for a 4 channel square template measurement
98
Figure 9.9-22 Example of a 3D glasshead view of an n-back task using the Octamon.
99
Figure 9-24. Topo graph properties – color scale
Show Labels
When checked, labels are shown in the graph. The o in a topo graph shows
the used optode combination points, which can be labeled by using this
option.
Show sensors
When checked, indicates the optode positions in the topo graph.
9.3.4.2 Transformation A
The properties in this tab are identical with defining a regular trace
(paragraph 9.3.1).
100
will be included. During DAQ, the scaling will adapt the y-axis dynamically in
order to show the current values.
Color mapping
This defines the color mapping of the topo graph. The options are:
- Linear color mapping: defines the color scale of the topo graph by going
linear in the sRGB color space, from the selected minimum color to the
selected maximum (see below).
- Hue: is an attribute, which defines an element in the color wheel (which
starts with red, orange, yellow... and then comes back to red again).
Positive hue means the color mapping will follow the color wheel in
positive direction from the selected minimum color to the selected
maximum (see below), and negative hue in negative direction. RGB and
hue are standard tools to describe perceived color in color spaces.
The scale can be set by:
- Minimum: The color mapping will start at the minimum value entered
here. Select by clicking on the color box.
- Maximum: The color mapping will start at the maximum value entered
here. Select by clicking on the color box.
101
Figure 9-25. Create movie
9.3.4.5 Movie
The topo graphs can be exported to a movie by Menu Playback Movie
capture. A window shown in Figure 9-25 will open. Recording a movie is only
possible in plain 2D mode.
Movie specification
Choose a file name by browse. The window size specifies the size of the
windows of the movie. Small sizes can result in a cut off legend, but require
less space to save. To add two graphs, choose a window size with A and B!
The frame rate specifies number of frames per seconds (fps). A high frame
rate gives more detail, a low frame rate requires less space. The frame rate
is restricted to the sample rate of the measurement; higher frame rates
give a frame rate equal to the sample rate. The movie duration defines the
duration of the movie; a three-hour measurement of can be reduced to a
ten-minute movie.
The graphs have scrollbars if the complete measurement does not fit in the
screen, which are also used to review data. Use the zoom buttons to look
at details.
102
Figure 9-26. Spectrum graph
Fourier transformation. The frequency is plotted along the x-axis and the
amplitude along the y-axis. Graphs with the icon are spectrum graphs. A
spectrum graph can be opened by Menu Measurement Create
Spectrum Graph or right mouse clicking at the measurement in the project
view and selection of “Create Spectrum Graph”.
103
Figure 9-27. Indicator view with two traces and legend, horizontal bars and vertical bars.
104
A closed graph can be opened by selection of the graph in the project view,
Menu Graph Open Graph, by right mouse clicking in the project view at
the graph and selection of “Open Graph” or double clicking in the project
view at the graph. If you cannot find the graph, it might be hidden behind
another graph or minimized.
105
A graph can be frozen/de-frozen during the measurement by clicking on
in the toolbar or by Menu View Freeze Graphs or by the shortcut <F8>.
The measurement continues recording during the freeze period, only the
graph is frozen. This function can be handy to make nice screenshots
during the measurements.
An open graph can be closed by right mouse clicking in the project view at
the graph and selection of “Close Graph” or at the exit cross in the upper
right corner.
106
Figure 9-30. Toolbar – zoom
A graph can be printed by Menu File Print. Only the selected graph will
be printed. Choose Menu File Print preview to preview the printed
graph. Select Menu File Print setup to change the print settings.
Copy a graph to another program by right clicking in the graph and
selection of “Copy to Clipboard” or use Ctrl+C.
The used time span can be switched to another time span (rollover mode
and trend view) by Menu Data Collection Toggle graph(s) to trend mode
or clicking on in the toolbar. The trend mode has a longer time span
then the rollover mode. The duration of spans can be changed by the
program options (paragraph 7.6).
107
automatically by creating graphs from a template or by the function “create
all graphs”. Add traces to a graph by the function “create trace” or “create
all traces”. The first creates a trace without specifications. The second
creates traces belonging to the optode template. Traces can be created in
regular graphs, indicator views and spectrum graphs.
A trace can be created in a graph if the graph is selected in the project view,
Menu Trace Create trace or by right-mouse clicking the project view on
the graph and select “Add trace”. The window shown in Figure 9-31 will
come up. Trace coordinates are shown by moving the mouse over the
trace. A trace can be selected by clicking on it in the graph or in the project
tree.
9.4.1.1 Channel A
The properties in this tab specify the name of trace, and the primary source
of data.
Label
This specifies the name of the trace. This name is automatically generated
from the source and the transformation applied. Overrule this by
deselecting “Auto” checkbox and enter an alternative name. The label is for
own reference and will be shown in the project view.
108
Figure 9-32. TSI optode template
Type source
Choose the type of data source. This option list will show “optode
combination”, the available AD Channels and external data file sources. If
an external data file source is added, select one of the defined data
columns in “column” (see paragraph 9.5.3).
Channel
If the optode combination is selected as type source, select one of the
optode combinations available in the optode template as data source. In
Figure 9-31 the white circles in the optode template are representing valid
optode combinations. Select a channel by clicking on one of these
combinations in the graphical display. To show the relative concentrations,
choose “Rx1-Tx1”, “Rx1-Tx2” or “Rx1-Tx3” as source channel. Each shows
the measured concentrations of that optode combination. The picture of a
TSI optode template has a square between the first and second channel
(two channel measurement) or behind the second channel (three channel
measurement), which is called the TSI combination source (Figure 9-32).
With selection of this square, you can show TSI or TSI Fit Factor and the
absolute concentrations calculated with the SRS method. Switch between
TSI and Tx2 by clicking again at the square.
Transformation
This defines which type of data to display. The available transformations
depend on the NIRS system and the selected data source at the “Channel
A” tab. The following transformations are explained:
- O2Hb, HHb, HbDiff and tHb are standard traces and display
concentration changes of oxygenated, de-oxygenated, difference
(O2Hb-HHb) and total (O2Hb+HHb) amount of hemoglobin and
Myoglobin [Mb]. These can be selected when an optode combination is
selected as source. ). If the TSI combination is selected as data source,
the estimated absolute concentrations are given.
- CtOx (or others) is available for Oxymons with a transmitter of three
wavelengths (or more) and can be selected when an optode
combination is selected as source. This trace displays concentration
changes of cytochrome oxidase.
- TSI% and TSI Fit Factor are absolute traces and display index of tissue
saturation in % and a figure for the Fit Factor. The Oxymon, setup and
channel must qualify for a spatially resolved spectroscopy
109
measurement. Select the TSI combination as source in at the Channel A
tab. For more information about TSI and the Fit Factor, see chapter 12.
- User defined transformation displays optical densities measured by the
Oxymon at a specific wavelength (and independent on DPF and source-
detector distance). Note that multiplication factors must be defined in
the entry box beside the available wavelength of interest, use “1” for
standard OD’s for one wavelength at the time and “0” for the other
wavelength (Figure 9-34). If the TSI combination is selected as data
source, the calculated absorption coefficient µa is displayed.
- User defined * 1/(dpf*dist) is the user defined transformation above, but
with the DPF and source-detector distance settings taken into account
(divided by these factors).
Wavelength entries
If “User defined” transformation is selected, a wavelength entry selects the
wavelength and multiplication factor. "O" means unselected, "1" means
selected and any positive number can be used for multiplication (Figure
9-34).
This defines the multiplication factor of the trace from an optode
combination. It can also be assigned negative values.
110
Figure 9-34. Trace properties - transformation A - User defined
Scale
This defines the multiplication of a channel. This is standard set to 1.
Bias
This defines the offset y-value added to the entire trace from an optode
combination. For more information about the bias, see paragraph 9.4.6.
This parameter is automatically defined by the program when the ‘bias all
traces’ function is used or ‘bias traces at start’ is selected in the program
options (paragraph 7.6 but you can also type any number (positive and
negative)).
Filter
A filter can be applied to the trace. For more information, see paragraph
10.5.
111
Cyclic operator
Cyclic operators can be used to analyze repetitive measurements. For more
information about the cyclic operators, see paragraph 10.3.
Voltage
If an AD Channel is selected as “Channel A” data source, the voltage entries
will be available. Min Scale and Max Scale define the range of the input in
voltages, which should be specified by the input devices supplier or have to
be measured.
Scaled value
If an AD Channel or imported data file is selected as “Channel A” data
source, these scaled value entries will be available (Figure 9-35). Min Scale
and Max Scale define the (defined) scaling values. For example: If 0 Volt
should represent 10 bpm and 3 Volt should represent 80 bpm, then fill in
“0” for Min Scale of the Voltage, “3” for the Max Scale of the Voltage, “10” for
the Min Scale of the Scaled value and “80” for the Max Scale of the Scaled
value.
Delay (sec)
If AD Channel or imported data file is selected as “Channel A” data source,
these scaled value entries will be available. Fill in the delay of the AD
Channel signal in seconds.
9.4.1.3 Operator
The properties in this tab (Figure 9-36) affect the operator between the
primary source of data (Channel A) and secondary source (Channel B).
Operator
This operator enables the combining of Channel A and Channel B source
into the same trace. By selecting no operator "None (Trace A only)" the
program only uses the defined settings of trace A. When an operator is
selected, also the properties of Channel B (next tab) need to be defined.
The available operator options are an addition (+), multiplication (*) and
division (/).
9.4.1.4 Channel B
This only seen if an operator is selected. The properties in this tab affect
the second source of data. These settings are only in use when an operator
is selected for combining Channel A and Channel B source into the same
trace.
112
The properties in the Channel B tab are identical with defining settings of
Channel A.
9.4.1.5 Transformation B
The properties in the transformation B tab are identical with defining
settings of transformation A.
9.4.1.6 Trace
The properties in this tab (Figure 9-37) affect the color and style of the
trace.
Color
This specifies the color of the trace. Select by clicking on the color box.
Style
This specifies the thickness of the trace. Select an available option from the
list.
By Menu Trace Add All Traces or by right mouse clicking in the project
view at the current graph and selection of “Create all traces” the traces to
show the concentrations (O2Hb, HHb, tHb, HbDiff, TSI and TSI Fit Factor)
can be created at once for an optode combination of the used optode
template. The window is shown in Figure 9-38.
113
Show the trace properties by Menu Trace Properties, by right mouse
clicking on the trace in the project view and selection of “Trace properties”,
by double clicking on the trace in the project view or by selecting of the
trace in the project view and clicking on the toolbar button . The
windows belonging to this function are equal to the windows belonging to
trace creation (paragraph 9.4.1).
To remove a trace, select the trace in the project view; go to Menu Trace
Remove Trace or by right-mouse clicking in the project view on the trace
and selection of “Remove trace”.
9.4.6.1 Biasing
Biasing is setting a baseline of the data. Biasing all traces once in the
beginning of data acquisition is usually preferred, or to a point considered
as the baseline or reference of the measurement. Since we are always
looking into relative changes the starting point of a concentration value will
never matter, and the “Bias Trace” or “Bias All Traces …” is only for own
convenience and graphical displaying during the measurement or analysis.
The “biasing” will never affect the real recorded data and when afterwards
(off-line) traces can be biased to any point (for example in the “Trace
properties”). When using cyclic operator, the options with "detrend" will
bias traces to an average over a specified period, such as for example the
first 5 seconds.
114
in the toolbar or <F6> is only available during measuring and will bias all
traces and clear the graphs at once. The trace(s) will be biased to the
selected data point.
Import data files (*.txt) by Menu Measurement Add external file or right
click at the measurement folder in the project view and select “Add data
file”. The imported data file will be added to the current measurement in
the project view as shown in Figure 9-40. The data file import wizard is
shown in Figure 9-42. The result of importing the example (Figure 9-43)
with the settings from Figure 9-42 is shown in the graph of Figure 9-40.
115
Figure 9-39. Import measurement
Name
Name shown in the project view for the added data file. In the example
(Figure 9-41) the name is “import file”.
File path
Select the *.txt file, which should be added, by clicking on “Browse”. The
example import file is called “import.txt” (Figure 9-43).
Type
This specifies the type of column separation:
- Fixed width: fixed width of the column (as in the example)
- Delimited: columns are separated by a delimiter:
- “,”
- “space”
- “tab”
116
Figure 9-41. Import data file
Columns
This specifies the data columns in the data file:
- Name: Choose a name for the data column (“value” and “sample nr” in
the example).
- Start: start position of the data column in the data file (Fixed width).
- Width: Width of the data column (number of characters) in the data file
(Fixed width).
- Column number: number of the data column in the data file (Delimited).
The buttons “Add”, “Delete” and “Preview” also belong to this field
- Add: add an extra column.
- Delete: delete a column.
- Preview: give a preview of the imported data.
Time column
This field specifies the time column. The column names as specified at the
columns will be displayed. Select “row number” to use sample numbers.
Time scale
This field can be used to up or down sample the imported data to the same
frequency as your Oxysoft data. If the imported data is measured with 10
Hz the time scale value has to be set to 0.1.
Further it can be used to convert your time column data to seconds if it has
a different unit than seconds. Use for example a value of 0.001 to go from
ms to s.
117
Figure 9-42. Example file
9.5.3.2 Properties
Change the properties of the imported data file by right mouse clicking in
the project view on the data file and select “Properties”.
118
9.5.3.3 Duplicate imported data file
Duplicate an imported data file by right mouse clicking in the project view
on the data file and select “Duplicate”.
If an error comes up during export, usually the amount of data is too big
for the program used to export to (usually MS Excel), or the file used to
export to is already opened in another program.
It is also possible to read the raw date (*.oxy3 files) with Matlab directly.
Contact Artinis for the m-files.
119
Figure 9-44. Export – export type
120
9.5.4.2 Export file destination
The file name and destination of the exported file can be selected in the
next window, shown in Figure 9-45. Click on “Browse”, a new window will
open (Figure 9-46). Choose the file name and destination and click on
“Save” (Figure 9-46). Click on “Next” in the window of Figure 9-45.
Data transform
- Graph data of open graphs: exports concentrations and TSI shown in
the opened graphs. This can also export OD values and ADC voltages if
these are shown in an opened graph.
- OD values and ADC voltages: exports OD values and ADC voltages. This
can be used for (re)calculation and analysis.
- Raw data: gives an indication of the measured AD counts.
We recommend using only graph data of open graphs for normal
applications. This will export the measured concentrations and TSI. OD
values and ADC Voltages and Raw data is only useful for special applications.
Make sure you have all graphs with data open before you start the export.
Null field
separation of data (delimiter) fields can be specified by “Empty”, “Space”, “0”,
“-“ “*” or “NULL”.
Click on “Next”.
121
Figure 9-48. Export – export time span
Use:
- Time: “1”, “10”, etc. The period will start or end at 1 or 10 seconds.
- An event label: “A1”, “B2”, etc. The period will start or end at event “A1”
or “B2”.
- An event label: “A”, “B”, etc. The period will start or end at events A1 or
B1. You cannot export different periods to one file by using multiple
events (A1, A2, etc.).
- The general code: ‘*’ instead of an event label, which will select any
existing event fulfilling the criteria.
- A combination of time and event labels: “A+10”, “B-10”, etc. The period
will start or end 10 seconds after event A1 or 10 seconds before event
B1. Warning: “10+A” does not work.
By example, a down sample factor “10” and a sample rate of 10 Hz will lead
to a new sample frequency of 1 Hz. A down sample factor “1” will leave the
sample rate to the sample rate during the measurement. The export
sample rate is shown at the right side of the field if filled in correctly.
Start Export
Click to start the export.
122
The Ascii writer is a protocol that allows third party applications to receive
online data from OxySoft. Communication is through sockets.
The ASCII writer can be enabled in Portasoft.ini by adding in top of the file:
[Asciiwriter]
Enable=1
Port=7777
If the ASCII writer is enabled, it writes the sample data to a local TCP socket.
To connect to the ASCII writer a standard TCP socket can be opened to the
host computer on port 7777.
Sample number, (RX1 - TX1), (RX1 - TX2), (RX2 - TX1), (RX2 - TX2) ...
The order of the data sent by the ASCII writer depends on the used
measurement template. For a split template, the data would be for
example:
Sample number, (RX1a - TX1), (RX1b - TX1), (RX1a - TX2), (RX1b - TX2) …
The data is always written as plain text (ASCII), and should be parsed by the
user as only raw values are given (relative changes should be calculated by
using a previous value as a starting point).
The FieldTrip buffer is an open source TCP/IP protocol, which allows clients
to write and read data from (for a more extensive description see
http://www.fieldtriptoolbox.org/development/realtime/buffer_overview).
The open source architecture allows a number of different systems to write
123
data to the shared memory segment, so that not only data from within
Oxysoft but also e.g. EMG or EEG data can be written to the buffer. The
FieldTrip buffer is platform and OS independent, thus all Windows, Mac and
Linux clients can connect to the buffer, and the software reading from or
writing to the buffer can be in any programming language. The original
buffer implementation is natively written in C and C++, but
implementations for reading and writing are freely available for Matlab. Java
and Python.
To activate the FieldTrip buffer, the following lines need to be added to the
portasoft.ini:
[FieldTrip]
Enable = 1
StartServer=1
124
This chapter describes how to work with events, periods, cyclic operators and
how to do a calculation or to apply a filter.
Use events to indicate the events during the measurement like the start
and end of an occlusion. Events will help to analyze measured data; they
are included in export files. Furthermore, they can be used for calculations,
to set the cyclic operator properties, to define the export period, to define
the period used for creation of a movie and adjustment of the x-axis. An
event is shown by a vertical line in the graph (Figure 10-1) and marked with
an event label (“E” in this example); however, it is only visible in trend
display mode and after the measurement is finished. The time since the
last event is marked in the DAQ status view.
The events made during online recording are defined by Menu Analysis
Event keys (Figure 10-2). The events are labeled by the keys, like “A”, “B”,
etc. Give an “event text” to describe the event as shown. It is possible to use
the same event label for multiple events. For instance, “A” indicates the
start of an exercise and “B” indicates the end. If the subject does the same
exercise twice in one measurement, the program will refer to the first start
event “A1” and the second “A2” and call the end events “B1” and “B2”. Using
the events to define a certain period for calculation, export, etc. call them
together as “A” and “B” and separate as “A1” and “A2”.
100
0
-100
-200
-300
39.2 39.4 39.6 39.8 40.0 40.2 40.4 40.6 40.8 41.0 41.2 41.4 41.6
Figure 10-1. Graph with event E at 40.24 seconds
125
Figure 10-2. Define event keys
126
The time coordinate of the event can be adjusted, also the event label and
the event text related to the event. The event is displayed as a blue vertical
line. Events inserted after the measurement are NOT saved in the *.oxy3
file but in the measurement. This can be done by using the export function
(8.5.d) and export to a new *.oxy3 file.
10.1.4.1 Absolute
The absolute mode looks to the data values of one trace. An event will be
defined when the trace fulfills two conditions in a certain time span. The
event will be generated in the middle of the moments when the first
condition is fulfilled and the second condition is fulfilled. An example is
shown in Figure 10-4.
Select trace
Select a trace by clicking on it in the project view or in the graph and drag
this to the button. This trace will be used to generate events.
Event label
A, B, C, etc. labels the event. Previous defined events with the same label in
the measurement will be removed if “remove previous defined events with
this label” is checked.
Pre-event value-window
This defines the first condition. The trace has to be between the minimum
and maximum to fulfill to the first condition.
Post-event value-window
This defines the second condition. The trace has to be between the
minimum and maximum to fulfill to the second condition.
Width of test
This defines the time span wherein both conditions have to be fulfilled. The
width has to be specified in seconds.
127
Figure 10-4. Generate events - absolute, see Figure 10-5 for the results
Figure 10-5. Generated events with the parameters from Figure 10-4 (events C, where the data is
zero after the peak) and Figure 10-6 (events I, used to find only slopes with a steep slope).
Time offset
This adds a time offset to the event, it shifts the events forward in time. The
offset has to be specified in seconds.
Inhibit period
This defines the shortest possible time allowed between two events. The
period has to be defined in seconds.
128
Figure 10-6. Generate events - relative
10.1.4.2 Relative
The relative mode looks to the slopes one of a trace. An event will be added
if the trace has a certain slope defined in the window as shown in Figure
10-6. Most properties are equal to the absolute mode, with exception of
“First derivate value window (unit/s)”.
The event properties can be shown by right clicking on the event in the
graph and selection of “Event properties”. It is only possible to see the
properties of events inserted after the measurement, because those can
be changed. To see properties of all the events, open the event list.
129
possible by typing in this window, but after closing the event list, the
adjusted information is not saved.
A period (or two) can be selected and used for calculation during an online
measurement or after each measurement or to make a spectrum graph
over a specified period. For a calculation, a period A can be used to define
the start and end of the calculation period (see paragraph 10.4). For the
compare function, also select a period B. After the measurement: use the
periods (period A) to make a spectrum graph over that period. Define the
periods by:
130
- start of period A: Press or Menu Analysis Set start of period A.
- end of period A: Press or Menu Analysis Set end of period A.
- start of period B: Press or Menu Analysis Set start of period .B
- end of period B: Press or Menu Analysis Set end of period B.
Show/hide the period by pressing or Menu Analysis Show period.
The periods can be moved in the graph clicking at the middle of the grey
area (Figure 10-9) and drag the area to the desired time coordinate. It can
be enlarged or decreased by clicking at the start or end of the grey area
and drag the start or end of the area to the desired time coordinate. The
periods are not saved or transformed to events.
-4
-8
-12
. 65 67 69 71 73 75 77 79 81 83 85 87 89 91 93
Period
To change the cyclic operator of a trace, select the trace and open the trace
properties (paragraph 9.4.3). Open the tab “Transformation A” and click on
“cyclic operator” (Figure 10-11). Choose an appropriate operator. The result
is equal if the operator is applied before, during or after a measurement.
However, during the measurement the trace is shown without the cyclic
operator. To remove the cyclic operator, select “No operation” as cyclic
operator.
131
- Average: the average of the cyclic traces will be displayed as one trace.
- SD: the standard deviation of the average trace will be displayed, in
another brighter color than the average trace.
And all combinations of these options.
First use only cyclic operators without averaging, look into the graph and
make sure the right cyclic periods and number of traces are present.
200
150
100
50
0
tHb
-50
HHb
-100 O2Hb
a. 0 15 30 45 60 75 90 105 120 135 150 165 180 195 210 225 240
150
100
50
0
tHb
-50
HHb
-100 O2Hb
b. 0 10 20 30 40 50 60
Figure 10-10. Example of cyclic measurements. a. measured concentrations (green tHb, red
O2Hb, blue HHb). b. Cyclic average of the measured concentrations with cyclic period is “60”
seconds, cyclic offset “0” and operator “cyclic average”.
132
Detrend front period (sec)
This defines the window in seconds if the “detrend front cyclic operator” is
selected.
133
Figure 10-11. Trace properties – transformation A
134
Oxysoft can apply several calculations to the data over a certain period or
more periods by the calculation function or by applying a cyclic operator to
the trace (paragraph 10.3). To open the calculation function, go to Menu
Analysis Calculate (Figure 10-13) press in the toolbar or use the
shortcut <F5>. Select a trace, a start and end of the calculation period, a
function and a channel to perform a measurement. After this, press
“Calculate”.
Trace
Add the trace to which the calculations have to be applied. It is not possible
to use traces from indicator views. Press the button and click in the project
view on the trace or drag-and-drop a trace from a graph. Select to apply
the calculation to the trace itself or the graph or traces in all opened graphs
of the measurement belonging to the selected trace in the field under
“Channel”.
135
- An event label: “A”, “B”, etc. including all events, which are present with
that label (“A1”, “A2”, etc.). The periods will start or end at events A or B.
In this way, more periods can be defined at once.
- The general code: ‘*’ instead of an event label, which will select any
existing event fulfilling the criteria.
- A combination of time and event labels: “A+10”, “B-10”, etc. The period
will start or end 10 seconds after event A1 or 10 seconds before event
B1. Warning: “10+A” does not work.
- The code “<” smaller than and “>” bigger than: “A>100”, “B<500”. The
periods will only start or end with events A after 100 seconds or events
B before 500 seconds.
- The code “|” OR: “A1|A3”. The period will start or end at A1 and A3. The
input “A1|A3>100” is actual “(A1|A3)>100”; if A1 or A3 is after 100
seconds since the start of the measurement, it is used as start/end of
the period.
- The “!” operator subtracts the given set from the set containing all
events. For example: !U2 takes all events except U2 or !U2 & !U4 takes
all events except U2 and U4 Note the equivalent is !(U2|U4).
Select trace B
For the function “Compare” two traces are required or two periods of the
same trace. Press the button and click in the project view on the trace or
drag-and-drop a trace to the button. Define the start and end by using
period B or time and/or seconds like described at “Start and end”.
Function
The functions are the calculations, which can be applied to the data. The
functions are:
- Average: calculates average and standard deviation
- MinMax: calculates the minimum, maximum and the range (maximum
minus minimum).
- Average & MinMax: combination of Average and MinMax.
- Compare: compares two traces (or two periods of the same trace)
giving averages, standard deviations and the t-value of a student 2-
tailed paired t-test with the probability (prob) and alpha = 5%.
- Correlation: Select trace 1 and time period and choose trace 2, to value
calculate correlation with on same time period using Pearson
2
regression. Results are goodness of fit value: R , and it linear prediction
values with intercept and slope. The slope is the measure of
correlation.
- Regression: applies a linear regression to the data (y= a+bt). Results are
a, b, and r^2: a is the offset, b the slope and r^2 the fit quality ([0...1], 1
is a perfect linearity).
- Loglinear regression: applies a logarithmic linear regression to the data
(log10(y)=a+bt). Results are a, b, and r^2: a is the offset, b the slope and
r^2 the fit quality ([0...1], 1 is a perfect logarithmic linearity).
- VO2: volume consumed Oxygen, used for occlusion analysis; calculated
as the change of [O2Hb], [HHb] and [tHb] over time; “b” means the
slope from a regression fit (see regression). Both traces of O2Hb and
136
HHb are needed for this (function cannot be used on ‘trace only’) so it
can also calculate a reliability factor according to RF= 100% * (1-abs(
b(tHb)/b(O2Hb) ) ). Note this function calculates the slope needed to
calculate VO2, not VO2 itself. More information about VO2 is found in
the thesis of Dr. M. van Beekvelt.
- Area: area under the trace. The area under the trace of a venous
occlusion is this equal to the blood flow (see chapter 4), arterial should
have no blood flow. Carefully notice is needed with biased traces.
- Graph data: no calculation, this gives the graph data in the specified
period.
- Harmonized graph data: resamples the specified period to your wishes.
If the function “compare” gives "0" as result, please make sure to open the
graphs, select the trace, drag-and-drop it to the correct button under
“Trace”. Both of the two buttons need to be updated with the correct
traces.
Channel
The calculations can be applied to more traces by specifying this field:
- Selected trace only.
- Selected graph only: the graph containing the selected trace.
- All opened graphs, non-cyclic traces.
- All opened graphs, cyclic traces only.
- All opened graphs.
Type
If the option “selected trace only” is not selected, this field is enabled. The
type data, which can be used for calculation, is specified here.
Calculate
Press to calculate. After finishing a calculation, just choose a new function,
trace or period and click on “calculate” to do a new calculation. If you have
selected “output to MS Excel file” it will write the output in the selected
sheet.
Save settings
Press to save the settings. The calculation is saved and can be reopened by
clicking in the project field on the calculation indicated with this icon .
The result of the calculation will be shown if the button “Calculate” is
pressed.
Copy to clipboard
This copies the result to the clipboard, which can be pasted in other
programs.
137
“Calculate”. Never have the Excel file already open. If an error comes up,
usually the amount of data is too big for the MS Excel program, or that the
file used to export to is already opened in MS Excel or by another program.
10.4.1.1 Format
This tab specifies the lay-out of the calculation results (Figure 10-14).
10.4.1.2 MS Excel
This tab specifies the MS Excel file where the calculation results are saved
(Figure 10-15). This is only possible if MS Excel is installed to the computer
and the Excel file is not opened!
File
MS Excel file and location
Sheet
Sheet of the MS Excel file
138
On top of sheet/append to sheet
If the sheet already contains information, this information will be
overwritten by the calculation results if the first option is chosen. The
second option appends the data to the sheet.
To add a filter to one trace, select the trace and open the trace properties
(paragraph 9.4.3). Open the tab “Transformation A” and click on “filter”.
Choose an appropriate filter. Fill in the filter width (seconds) or the cut-off
frequencies (Hertz). Click at “OK”. The program filters the data now. The
result of a filter is equal if the data is filtered before, during or after a
measurement. Filtering after measuring has as advantage that the effect of
the filter to the measured data will be clear. Applying a filter to a trace
before the measurement shows the result immediately.
To remove a filter from a trace, select the trace and open the trace
properties (paragraph 9.4.3). Open the tab “Transformation A” and click on
“filter”. Select “No filter” at “Filter”. Click at “OK” and the trace will be
unfiltered.
139
width or the cut-off frequencies. Check if “Modify” is selected for the filter
type and the filter width/cut-off frequencies. Click at “OK”. The program
filters all data of the measurement.
i t
n
Figure 10-16. Moving average filter with filter width (n) over time (t).
Figure 10-17. Measured O2Hb concentration (red) and filtered O2Hb concentration filtered by a
moving average filter with 0.3 seconds filter width (black).
140
Figure 10-18. Measured O2Hb concentration (red) and O2Hb concentration over time filtered by
a moving average filter with 10 seconds filter width (black).
1 FWHM
n , where fs is the sample frequency.
f s 2 2 ln( 2)
Example: Figure 10-20 is the same as Figure 10-18, but the O2Hb
concentration is filtered with Moving Gaussian with a 0.3 s filter width
(blue). This filter result is much smoother compared to the filter result of
the moving average with the same filter width (black). Increasing the filter
width of the moving average still gives a less smooth result (green, filter
width is 0.8 s) compared to the Gaussian filter result, while the amplitudes
are equal.
141
Gain Amplification H(f)
i t
n
Figure 10-19. Moving Gaussian average filter with a Gaussian window with filter width (n) over
time (t).
Figure 10-20. measured O2Hb concentration (red), O2Hb concentration filtered by a moving
average filter with 0.3 seconds filter width (black), filtered by a moving Gaussian filter with 0.3
seconds filter width (blue) and filtered with a moving average filter width 0.8 seconds filter width
(green).
The low pass, high pass, band pass and band stop filters are filters which
attenuate a certain frequency component of the signal. The filters
parameters are estimated by the Parks-McClellan algorithm for FIR (Finite
Impulse Response) filter design. FIR filter designs need a certain amount of
samples for good characteristics, and this may result in "lack of data points"
in the beginning and end of the trace result. FIR filters are described by the
filter type, the gain, the band pass ripple, the transition band and the cut-
off frequencies. The cut-off frequencies are limited at 1/100 of the used
sample rate. If the cut-off frequency is too low, the error shown in Figure
10-21 will come up. If the cut-off frequency limit is reached, you can try to
downsample the data via the export function and use this downsampled
data for a lower cut-off frequency.
142
Which frequencies are present in the data can be shown by making a
spectrum graph (paragraph 9.3.5).
0
Gain [dB]
-30
fL
fL/4
Figure 10-22. Low pass filter characteristics: attenuation is 30dB, band pass ripple is 1 dB and
transition band is Lo Freq/4 Hz
143
Gain [dB] 0
-30
fH
fH/4
Figure 10-23. High pass filter characteristics: attenuation is 30dB, band pass ripple is 1 dB and
transition band is Hi Freq/4 Hz
Figure 10-24. Measured O2Hb concentration (red) and high frequency O2Hb concentration
variations (black) filtered by a high pass filter, cut-off frequency =2 Hz
144
0
Gain [dB]
-30
fL/4 fL fH fH/4
Figure 10-25. Band pass filter characteristics: attenuation is 30dB, band pass ripple is 1 dB, low
transition band is Lo Freq/4 Hz and high transition band is Hi Freq/4 Hz
Figure 10-26. Measured O2Hb concentration (red), O2Hb concentration without slow variations
(black) filtered by a high pass filter, cut-off frequency = 0.1 Hz, O2Hb concentration without slow
variations and high frequency noise (green) filtered by a band pass filter, low cut-off frequency =
0.1 Hz and high cut-off frequency= 0.9 Hz
Example: To reduce low frequent noise and high frequency noise of data, a
band pass filter is applied (Figure 10-26). The resulting data of the band
pass filter are the O2Hb concentrations fluctuations resulting from the
pulse. The picture is partly enlarged in Figure 10-27.
145
filter designs need a certain amount of samples for good characteristics,
and this may result in "lack of data points" in the beginning and end of the
trace result.
0dB
-30dB
fL/4 fL fH fH/4
Figure 10-28. Band stop filter characteristics: The attenuation is 30dB, the band pass ripple is 1
dB, the low transition band is Lo Freq/4 Hz and the high transition band is Hi Freq/4 Hz
Figure 10-29. Measured O2Hb concentration (red), O2Hb concentration without fluctuations
resulting from the pulse (black) filtered by a band stop filter, low cut-off frequency = 0.1 Hz and
high cut-off frequency = 2 Hz.
146
energy is equal to 1/2√2 * amplitude. The filter characteristics are equal to
a normal band pass filter, but with an extra application.
Figure 10-30. Amplitude modified sinus wave (blue), which is RMS band pass filtered (orange).
147
This chapter gives information about the setup of optodes and optode
templates. The optode template describes the optode configuration.
Figure 11-1. 1 channel optode template and 4 channel square optode template using split
transmitter and receiver fibers. The square template is a used for mapping measurements.
Optode template “2x1 channel” is a two channel optode template with two
subtemplates and with each one channel. A 4 channel measurement can
be applied with four subtemplates (optode templates: 4x1 channel (split) I
and II), but also as a mapping measurement using one subtemplate
(optode templates: 4 channel cross and square). Mapping measurements
can be accomplished with two dimensional optode templates (see
paragraph 11.2.3 to 11.2.5) and can be displayed with a topo graph
(paragraph 9.3.4). Important for mapping measurements is an equal
source-detector distance between all transmitter and receiver optodes
forming an optode combination of the optode template.
How to connect optical fibers to the Oxymon and the optode holder is
described in the paragraph 8.2 and 8.3. Some examples of setups of
optode configurations are given in this chapter.
Figure 11-2. Schematic view of the setup of a 1 channel system with a one channel Oxymon
cabinet and the one channel optode configuration. Rx means receiver, Tx means transmitter. The
number “1” is the channel number.
The 1 channel system consists of one Oxymon cabinet with at least one
receiver and one transmitter as shown in Figure 11-2. The channel is
formed by the combination of the transmitter and the receiver optode. To
connect the optode holder to the Oxymon, two fibers are necessary: a
transmitter and receiver fiber. Connect both fibers to the Oxymon and
fixate the optodes in the optode holders. Measure the distance between
the optodes to determine the source-detector distance.
149
Figure 11-3. Schematic view of the setup of a 2 channel system with a two channel Oxymon
cabinet and the two channel optode configuration. Rx means receiver, Tx means transmitter. The
number “1” and “2” are the channel numbers.
Figure 11-4. Schematic view of the setup of a four channel square system with a four channel
Oxymon cabinet and the 4 channel optode configuration. Rx means receiver, Tx means
transmitter. The number 1-4 are the channel numbers.
150
Table 11.1. The 4 channels will then be situated at the following receiver-transmitter pairs.
Tx1 Tx2
Rx1 1 2
Rx2 3 4
The 8 channel system consists of a full Oxymon cabinet with two receivers
and four transmitters (Figure 11-5). All transmitter optodes are combined
with all receiver optodes, which makes in total 8 channels. In this setup the
receiver fibers are unsplit, the fibers of Tx1 and Tx2 are unsplit as well, and
the fibers of Tx3 and Tx4 are both split. The 8 channels will then be situated
at the receiver-transmitter pairs shown in Table 11.2. In terms of single
diodes and position in exported data files, the pairs are shown in Table
11.3. This optode template is called “8 channel split” in Oxysoft and can be
specified in the measurement properties.
Table 11.2. The 8 channels will then be situated at the following receiver-transmitter pairs.
Table 11.3. The 8 channels will be filed the following way in terms of single diodes and position in
exported data files
151
Figure 11-5. Schematic view of the setup of an 8 channel system with an eight channel Oxymon
cabinet and an eight channel optode configuration. Rx means receiver, Tx means transmitter.
The number 1-8 are the channel numbers.
The 24 channel system consists of a 2 full Oxymon cabinets with each two
receivers and four transmitters (Figure 11-6). These cabinets have to
connect to each other as described in paragraph 6.2.2. The optode
configuration for the 24 channel setup is shown schematically in Figure
11-7. The fibers necessary for this setup are 8 unsplit transmitter fibers and
4 split receiver fibers. Take care that all fibers go to the right position on the
head setup. The 24 channels will then be situated at the receiver-
transmitter pairs shown in Table 11.4. In terms of single diodes and
position in exported data files, the pairs are shown in Table 11.5.
152
Figure 11-6. 24 channel Oxymon cabinets with a master and a slave cabinet
Tx1 Rx2a
Rx1a 1 2 3
Tx2
4 5 6 7
8 9 10
Tx3 Rx4a
Tx4
Rx3a
11 12 13 14
15 16 17
Tx6
Rx4b
Tx5 Rx1b
18 19 20 21
22 23 24
Rx3b
Tx7 Rx2b Tx8
Figure 11-7. Schematic view of a 24 channel optode configuration. Rx means receiver, Tx means
transmitter. The number 1-24 are the channel numbers.
Table 11.4. The 24 channels will be filed with the following optode combinations
153
Table 11.5. The 24 channels will be filed in terms of single diodes and position in data files as
shown in this table
154
This chapter will help to set up a measurement for tissue saturation index (TSI)
in Oxysoft. Note that you need special TSI equipment for the Oxymon.
Figure 12-1. Schematic view of a TSI measurement: light through tissue with three transmitters,
rho, ua, us and absorbers
155
The most significant absorbers in tissue that determine the absorption
coefficient are oxyhemoglobin, deoxyhemoglobin and water. The amount
of water in tissue is assumed to be constant and its value can be set in
Oxysoft. The correction for absorption due to water is as follows:
where 𝑐𝐻2𝑂 , 𝑐𝑂2𝐻𝑏 and 𝑐𝐻𝐻𝑏 are the absolute concentrations of water,
oxyhemoglobin and deoxyhemoglobin respectively. 𝜀𝐻2𝑂 (), 𝜀𝑂2𝐻𝑏 () and
𝜀𝐻𝐻𝑏 () are the extinction coefficients of water and the different types of
hemoglobin at the Oxymon wavelengths. Using two different wavelengths
enables you to distinguish between the oxygenated and deoxygenated
forms of hemoglobin. The tissue saturation index is an estimate of the
oxygen saturation of the tissue (StO2) and is defined as:
𝑐𝑂2𝐻𝑏 𝑐𝑂2𝐻𝑏
TSI [%] = × 100% = × 100%
𝑐𝑡𝐻𝑏 𝑐𝑂2𝐻𝑏 + 𝑐𝐻𝐻𝑏
Table 12.1. Example of scattering coefficient for three tissue types found by Matcher et al. (1997).
-1 -1
Tissue type 𝒌 [mm ] 𝒉 [nm ]
Forearm 1.1 4.6 ∙ 10−4
Head 1.45 4.5 ∙ 10−4
Calf 1.63 5.5 ∙ 10−4
The following parameters are needed in a TSI measurement and can be set
in Oxysoft in the optode template tab in the measurement properties
screen:
156
Note that the DPF is not needed in the calculation of the TSI, but the DPF is
needed in the calculation of the concentration changes, so always set this
value as well.
Figure 12-2. Screenshot of tab Optode Template in measurement properties where all
parameters can be set.
The TSI is typically 60-80% in muscle tissue under normal conditions. Note
that the TSI theory will only be valid under certain conditions. Furthermore,
the Oxymon is assumed to be properly calibrated under the conditions it
will be used during measurement (power level, gain settings and fibers).
When the calculated TSI percentage is negative or larger than 100, it is
truncated to [0,100]. In that case the measurement does not fit into the
mathematical model.
157
The TSI can be measured with an Oxymon containing at least two (we
advise three) transmitters and a special TSI optode holder with small fibers.
Figure 12-3 shows the special TSI optode holder. The average source-
detector distance can be adjusted from 3.5 to 5 cm. The shortest distances
are recommended for a TSI measurement.
The optode holder has a top plate which fixates the transmitting optical
fibers. It can be unscrewed if the fibers need to be replaced or changed to
a deeper/higher position. The position of the optode ends can be changed
to a deeper or higher position by replacing the small plate under or over
the optodes.
The Oxymon has to be calibrated before measuring the TSI. Make sure the
system is calibrated (paragraph 8.9) and do a “System calibration” of the
device if needed. Besides the Oxymon calibration, also a TSI calibration is
necessary. The TSI calibration is preferably performed before every
experiment. When the fibers are re-attached to the machine or any other
condition has changed which will affect the relative light transmission
between Tx1, Tx2 (and Tx3), a new TSI calibration must be done.
158
4. Make sure the transmitter fibers are connected to the TSI holder in the
right order:
a. If you have three transmitter fibers, connect Tx1 to the hole
closest to the hinge of the optode holder, Tx2 to the next and
Tx3 to the last.
b. If you have two transmitter fibers, you can choose where to
connect the fibers, but always connect Tx1 closer to the hinge
of the optode holder than Tx2.
Place the holder at the top of the basket (Figure 12-5)
5. When a TSI measurement with split TSI fibers is performed, cover the
transmitting optodes which are not being calibrated with a black cloth.
6. Start Oxysoft, create a new measurement and set the device open for
data acquisition without turning on the lasers (do not start the
measurement).
7. Choose the optode template which will be used during the
measurement in the “Optode Template” tab of the “Measurement
properties” screen.
8. Special optode templates are designed for TSI measurements. For TSI
measurement with two transmitters in one holder select a template
with only “TSI” in its name and for TSI measurement with three
transmitters in one holder select a template with “TSI Fit Factor” in its
name.
9. It is recommended to pre-set the device settings that are going to be
used during measurement, such as the laser output levels and sample
rate. For a high signal to noise ratio, standard laser power and a low
sample rate (1 Hz) are recommend.
10. Start the TSI calibration via Menu Data Collection TSI Calibration
and follow the steps of the calibration procedure.
159
Figure 12-5. Top view of the calibration basket with TSI optode holder in position
160
Table 12.2. Overview of MLB and SRS theories
Measures: Hemoglobin concentration changes TSI, TSI Fit Factor and absolute hemoglobin
concentrations
Parameter Known: 𝜀𝑂2𝐻𝑏 , 𝜀𝐻𝐻𝑏 and d 𝜀𝑂2𝐻𝑏 , 𝜀𝐻𝐻𝑏 , d and Δd
s: Assumed: DPF 𝜇𝑠 () (ℎ and 𝑘) and 𝑐𝐻2𝑂
Channel selection for Select circle: Select square:
plotting: Example: Rx1 – Tx1 Example: Rx1 – Tx1, Tx2, Tx3
Measurement example Graph 1 (biased at start) Graph 1 – 3 calculated from the MLB law, graph 4
(Figure 12-6) and 5 calculated from the SRS theory
* Note that a TSI measurement will also provide you with the relative
concentrations calculated by the MLB theory as one TSI channel forms 2 or 3
normal channels. Therefore the DPF should also be set for TSI measurements.
** A TSI channel combination of 1 Rx and 2 Tx does not have the Fit Factor option
161
1. What is Near InfraRed Spectroscopy (NIRS) based on?
NIRS is based on absorption of light by certain chromophores. When NIRS
is used on tissue the main absorbing chromophores are hemoglobin and
myoglobin. Both these chromophores have similar absorption spectra and
cannot be distinguished using NIRS. However, most important for NIRS are
the changes in oxygen binding to these chromophores. Very often only
hemoglobin is used to refer to both chromophores, as also by us.
Hemoglobin mainly exists in two forms, oxy- and deoxyhemoglobin. The
NIRS device measures changes in light absorption and uses the modified
Lambert Beer law to calculate changes in hemoglobin concentrations. Our
NIRS devices also have the possibility to calculate absolute concentrations
using spatially resolved spectroscopy. For more information you are
referred to the manual.
162
can do this yourself and are referred to the manual for further instructions.
The PortaMon can be calibrated by us, please contact us at
askforinfo@artinis.com. If you use Tissue Saturation Index (TSI) with the
Oxymon, we advise to do the absolute/TSI calibration each time you use a
TSI.
163
11. My license key does not work and software does not work.
- Try another USB port
- Make sure you are using the right key/USB dongle.
- The key should be plugged in before program start up and during
usage.
- Run the RocKey driver installation scripts on the Oxysoft CD/USB
17. I cannot enable for data acquisition or connect with the machine in
the software?
First try to restart the computer. If this does not work, try the following;
Oxymon: If the ‘create new measurement wizard’ is not available, probably
your FTDI drivers are not working correctly and need to be (re-)installed.
You can find this driver in C:\Program Files\Artinis Medical Systems
BV\Oxysoft. If you have an Oxysoft version 2.1.6 or older, you will need a
new FTDI driver for Windows 7. Contact Artinis at askforinfo@artinis.com
for more information.
PortaMon: Try to connect first via the Bluetooth software. If this does not
work, make sure the Bluetooth dongle is inserted, the PortaMon is on and
try to make a new connection in the Bluetooth software with the PortaMon
and use the new COM port for connection.
164
18. My Bluetooth connection of the PortaMon is very poor and gets
disconnected.
You might be too far from the computer. It depends on the type of antenna
you are using. Standard antennas go up to 100 meter, but we have
antennas which can go up to 1 km in the open field. If you have obstacles
between the PortaMon and laptop the distance will be less. Contact Artinis
at askforinfo@artinis.com for more information. The most likely cause is by
using the wrong (internal) Bluetooth module. This can be checked by
connecting to the PortaMon. While you are doing data acquisition, you
disconnect the USB Bluetooth module. If Oxysoft hangs you are using the
correct module. Please contact Artinis for further assistance. If Oxysoft is
still measuring you are using the internal Bluetooth module. To disable this,
go to ‘my computer’, go to ‘system properties’ and click on ‘device manager’.
Search for the internal Bluetooth module and disable it.
20. Where can I find manuals and information about NIRS, Oxymon or
Oxysoft?
Check the software CD/USB from the delivery of your Oxymon. You will find
various instruction videos, manuals and application notes. If you cannot
find it, please contact us.
22. I do not want to start recording every time I start the data-
acquisition.
You can change this default setting in the “View” menu, select “Options...”. In
the “DAQ”-tab there is a check-box where you can enable/disable the
option to “Start recording manually”.
165
23. I cannot get the calculation function “compare” to work, or I only get
“0”.
Please make sure to open the graphs, select the trace you want to study,
drag-and-drop it to the correct button under “Trace”. Both of the two
buttons need to be updated with the correct traces.
166
of a concentration value will never matter, and the “Bias trace” or “Bias all
traces…” is only for your own convenience and graphical displaying during
the measurement or analysis. The “biasing” will never affect the real
recorded data and when you look at recorded data afterwards (off-line)
traces can be biased to any point (for example in the “Trace properties”).
30. Why are my data traces only ‘flat’ or ‘zero’ or ‘gone’ or ‘wrong’?
There can be several reasons behind this observation. Make sure that you
have 1) attached a data file (.oxy3) with data (check the file information), 2)
filled in the correct optode distance and DPF, 3) the laser-to optode
assignment is correct, 4) that your graph scale and zoom is correctly set, 5)
do ‘bias all traces in measurement’ to get correct offsets. If you see this
while doing a measurement please make sure that 6) your experimental
setup is correct, and all transmitting and receiving fibers and optodes are
well attached (Oxymon), you still have a good Bluetooth connection
(PortaMon), 7) you receive enough light (signal %) by checking “View” menu,
select “DAQ value view” and look at all the signals you are using. If the
received signal is too low, set the power output or the gain settings of the
receiver to high (only possible with Oxymon).
If you see data, but not as expected, take a few steps back and check your
experimental setup. Please never underestimate the simplest 1 channel
measurement for piloting, trial and error!
167
error frequency. If upgrading is not a possibility, reducing your sampling
rate or recording time solves this issue too.
168
In Oxysoft several shortcuts are available for quick handling of Oxysoft.
F1 help
F2 open all graphs
F3 close all graphs
F4 insert event now
F5 calculate
F6 bias all traces in a measurement and clear graphs
F7 clear graphs
F8 freeze graphs
F9 zoom in X
F10 zoom out X
Crtl+F9 zoom in Y
Ctrl+F10 zoom out Y
Crtl+N open a new project
Ctrl+C copy (graph)
Alt+Enter properties
169
These abbreviations are used in the manual and in Oxysoft
Abs absolute
AD analog/digital
Cyt cytochrome
Cyt.Ox cytochrome oxidase
DA digital/analog
DPF Differential pathlength factor
HHb de-oxyhemoglobin
Max maximum
Min minimum
NIRS Near Infrared Spectroscopy
O2Hb oxyhemoglobin
O2Mb oxymyoglobin
OD optical density
StO2 Tissue oxygen saturation
tHb total hemoglobin
TSI Tissue saturation index
170
Product Warranty
The warranty of Artinis Medical Systems products is one year from date of
delivery, except for optical fibers, which are fully excluded from warranty.
These warranties do not cover product abuse, modification, and failure to
adhere to product instructions, improper operations and/or misuse. Artinis
Medical Systems B.V. is not responsible for damage arising from failure to
follow instructions relating to the product’s intended use. Artinis Medical
Systems B.V. is not responsible for injury or loss caused by or associated
with the installation and/or use of equipment in any manner other than in
strict conformance with the instructions set in this manual. Artinis Medical
Systems B.V. does not warrant damages or defects to unauthorized
changes to one of the items or shipping damage (other than original
shipment from Artinis Medical Systems). The warranty is voided if the serial
number of the product is defaced, modified or missing. Software Products
are covered specifically for defective media or manuals only, and are
provided as is. The software license you acquired cannot under any
circumstance by transferred back to Artinis Medical Systems B.V. Artinis
Medical Systems B.V. does not warrant or represent that third-party
software or hardware will function error-free when used in conjunction
with its products.
Artinis Medical Systems devices are not intended to cure, treat, mitigate or
prevent any disease.
Warranty Repair
In the event that any Artinis Medical Systems product becomes defective in
material or workmanship during the warranty period, Artinis Medical
Systems B.V. will determine if the product defect is covered under
warranty. Artinis Medical Systems B.V., at its sole discretion, may replace or
repair the unit determined to be under warranty. The labor and material
costs associated with the repair of the product may be the responsibility of
Artinis Medical Systems if determined to be under warranty. You must
receive pre-approval by Artinis Medical Systems for the labor and material
costs prior to repair or replacement of warranty products. You must
contact Artinis Medical Systems to obtain a Return Material Authorization
(RMA) number. An RMA number may be obtained by contacting Artinis
Medical Systems B.V. online or by telephone. Performance of any repair or
replacement on product under warranty does not renew or extend the
warranty period. For repairs of products obtained via a reseller or
distributor, contact the reseller or distributor for warranty.
Non-Warranty Repair
You can return a product for repair that is not covered by warranty only if
you have received a preapproved RMA number from Artinis Medical
171
Systems. Labor costs and freight charges associated with non-warranty
repair will be the sole responsibility of the customer. For any product that is
repaired outside of the warranty period, extra costs for labor and materials
specific for the needed repair will be offered by Artinis Medical Systems BV
after receiving the product and has to be approved by the customer before
start of the repair. Repairs on products out of warranty carry a half year
warranty, but only for the repaired parts. The warranty becomes effective
the day that you receive the item after repair. For repairs of products
obtained via a reseller or distributor, contact the reseller or distributor for
warranty.
Non-Defective Products
You are notified if, after examining and testing a returned product, Artinis
Medical Systems concludes that the product is not defective. The product is
returned to you and you would be responsible for the freight charges
associated with the return.
172
absorption................................. 12, 15, 17, 98 cut-off frequency ...................... 126, 127, 130
AD box ........................................................ 43 cyclic operator .................... 83, 103, 113, 122
AD Channel ....................... 37, 42, 97, 98, 110 cyclic offset .......................................... 120
board...................................................... 37 cyclic period ......................................... 120
delay .................................................... 101 detrend front period............................. 120
scaled value.......................................... 100 Cyt.Ox. .................................................. 13, 98
voltage ................................................. 100 cytochrome................................... See Cyt.Ox.
air in- and outlet ................................... 24, 37 cytochrome oxidase...................... See Cyt.Ox.
analog interface board.......... See AD Channel DAQ measurement wizard.................. 72, 104
Analog-Digital channels .........See AD Channel DAQ status view.................................... 47, 62
arrange graph DAQ value view ......................... 47, 57, 63, 67
as optode template .......................... 87, 94 data acquisition, open for ........................... 79
cascade .................................................. 94 data file....................................... See oxy3 file
restore graph layout .............................. 94 device settings ............................................ 77
tile .......................................................... 94 Differential Pathlength Factor ........... See DPF
arterial occlusion ........................................ 17 dongle ..................................... See license key
arterial oxygen saturation........................... 13 DPF.......................... 14, 17, 54, 71, 74, 75, 98
asynchronous connector cable ............ 40, 41 table ................................................. 54, 71
average traces........................................... 120 end-user license agreement .........See License
axis environmental light .................................... 64
auto scale ............................. 52, 84, 85, 89 event ........52, 62, 83, 111, 113, 119, 120, 160
common auto scale ................................ 52 change ................................................. 118
manual scale .............................. 52, 84, 89 generate............................................... 115
ticks ........................................................ 85 insert .................................................... 114
unit ......................................................... 89 label ..................................................... 115
x-axis ...................................................... 84 list 117
y-axis ...................................................... 85 properties .................................... 117, 118
background color .................................. 51, 83 remove ................................................. 118
beep ..................................... See out of range evt file ........................................... 70, 73, 108
bias................ 44, 51, 100, 103, 104, 120, 160 export ....................................... 108, 112, 113
blood flow ............................... 17, 18, 19, 124 down sample factor ............................. 111
blood volume ...................... 12, 13, 17, 18, 19 options ................................................. 110
calculation........................... 73, 110, 113, 122 time span ............................................. 111
area ...................................................... 124 extinction coefficient ................ 14, 15, 54, 71
average ................................................ 124 filter ...................................... 91, 92, 100, 126
compare ............................................... 124 add ....................................................... 127
graph data ............................................ 124 band pass ............................................. 129
minmax ................................................ 124 band pass filter .................................... 132
regression ............................................ 124 band stop ............................................. 129
VO2 ...................................................... 124 band stop filter .................................... 133
calibration ............................................. 41, 68 FIR ................................................ 126, 130
basket................................................... 150 high pass .............................................. 129
calibration tool ....................................... 68 high pass filter...................................... 130
TSI150 low pass ............................................... 129
channel ................... See optode combination low pass filter ....................................... 130
chromophore ........................................ 14, 15 moving average ......................... 126, 127
clean ........................................................... 10 moving Gaussian .......................... 126, 128
clock ............................................................ 62 remove ................................................. 127
concentration ....... 14, 19, 37, 81, 86, 98, 103 RMS band pass filter ............................ 133
connect Oxymon ......................................... 73 width ............................................ 127, 128
connect OxyMon............................. 29, 44, 68 FIR filters ................................................... 129
connection .................................................. 40 frequency
connector ................................................... 38 filter ......................................See filter, FIR
contact information ....................................49 frequency domain....................................... 90
copy (to clipboard) ................ 75, 95, 117, 125 frequency spectrum .............................. 90, 92
173
gain ..........................................65, 66, 67, 149 import.................................................. 104
auto gain................................................ 65 new ........................................................ 72
glass fibers .................................................. 38 optode template ................................. 136
graph .............................................. 55, 70, 81 properties .............................................. 81
arrange ................................... See arrange remove .................................................. 81
clear ....................................................... 93 rename .................................................. 81
close ...................................................... 94 template ................................................ 73
close all .................................................. 94 measurement quality ................................. 55
common auto scale ............................... 85 menu .......................................................... 47
common graph properties ............. 92, 103 movie ................................................. 90, 113
create .................................................... 82 myoglobin .................................................. 15
create all .................................... 82, 86, 96 NIRS ...................................................... 12, 98
create from template ............................ 82 null field.................................................... 110
duplicate ................................................ 93 occlusion
font ........................................................ 51 arterial ................................................... 18
freeze..................................................... 94 venous ................................................... 17
gridlines ........................................... 86, 92 volume consumed Oxygen .................. 124
manual scale.......................................... 86 OD ............................. 14, 39, 74, 98, 110, 135
open ...................................................... 93 optical density ..................................... See OD
open all .................................................. 93 optical fiber ..................... 8, 10, 37, 38, 57, 58
options .................................................. 51 split ........................................................ 39
properties ........................................ 83, 92 optical pathlength ...................................... 14
remove .................................................. 95 optode .................................................. 38, 78
rename .................................................. 93 optode combination38, 39, 81, 87, 89, 97, 135
template ................................................ 87 Optode combination
ticks ....................................................... 86 TSI combination............................... 97, 98
unit ........................................................ 86 optode configuration ... See optode template
ground .................................................... 9, 37 optode holder ...................................... 38, 59
hardware installation ................................. 23 optode mapping ......................................... 76
hemoglobin ................... 12, 13, 15, 16, 19, 98 optode template39, 44, 51, 74, 75, 76, 97, 135
import 1 channel ..................................... 135, 136
data files .............................................. 104 12 channel ........................................... 139
data to graph ............................... 101, 107 2 channel ............................................. 137
measurement .. See measurement; import 24 channel ........................................... 140
remove imported data ........................ 108 2x24 channel ....................................... 143
indicator view ................................. 47, 82, 92 4 channel ................................87, 135, 137
interconnection board ............................... 37 8 channel ............................................. 138
inter-optode distanceSee source-detector distance subtemplate .......................................... 74
Lambert-Beer law ................................. 13, 14 out of range......................... 51, 64, 65, 67, 68
laser .....................................17, 37, 38, 57, 61 oximetry .............................. 12, 13, 15, 18, 19
laser module............. See transmitter module oxy file ...........................................70, 73, 108
laser power level .......................64, 65, 67, 77 open ...................................................... 80
legend................................................... 52, 83 oxy3 file ............... 62, 70, 71, 73, 81, 104, 108
license ............................................ 25, 49, 50 open .............................................. 80, 104
license key .................................................. 49 oxygen consumption .................................. 17
light source power level ............................ 65 OxySoft
local oxygen consumption .......................... 18 about ..................................................... 49
magnetic resonance imaging........................ 8 close ...................................................... 49
main switch .................................... 30, 37, 72 directory ............................... 26, 54, 69, 71
mapping ..................................................... 87 start ....................................................... 48
mapping measurement .................... 135, 138 pathlength correction factor ..............See DPF
high density ......................................... 143 period ................................................. 91, 118
master ................................. See master-slave periodic data ............................................ 119
master-slave ........................ 37, 40, 41, 43, 68 power indicator .......................................... 39
cable ................................................ 37, 40 power lamp .....................................30, 37, 43
measure lamp....................................... 37, 61 power supply ......................... 9, 23, 37, 41, 43
measurement ............................................. 70 information...................................... 23, 37
create .................................................... 72 print............................................................ 95
duplicate ................................................ 81 program options.................... 50, 83, 100, 104
174
project......................................................... 70 text color ............................................... 51, 83
close ....................................................... 72 time span ...................... 51, 95, 111, 115, 116
new ........................................................ 70 time-of-flight............................................... 17
open ....................................................... 70 time-sequenced principle ........................... 38
properties .............................................. 71 tissue saturation index ........................See TSI
save ........................................................ 71 toolbar ........................................................ 47
project view .......................................... 47, 70 topo graph ........................ 81, 87, 89, 96, 135
pulse oximetry ...................................... 13, 15 labels ...................................................... 89
QCF ....................................................... 46, 98 properties .............................................. 88
radiation ..................................................... 14 transformation A.................................... 89
incident radiation ...................................14 trace.......................................... 55, 56, 81, 87
transmitted radiation ............................. 14 bias................................................ See bias
raw data .............................................. 70, 110 channel A ......................... 43, 96, 100, 101
receiver ....................................................... 38 channel B .................. See trace, channel A
gain .............................................. See gain color ..................................................... 101
receiver module .......................................... 37 create ..................................................... 96
receiver saturation...................................... 67 create all .............................................. 102
rollover mode ....................................... 51, 95 duplicate .............................................. 102
safety ...................................................... 8, 23 operator ............................................... 101
sample frequency ................. See sample rate properties ........ 45, 92, 102, 103, 120, 127
sample rate ................... 39, 77, 111, 130, 151 remove ................................................. 103
down samplingSee export; down sample factor style ............................................. 101, 102
scale factorization ................................. 52, 86 transformation A45, 92, 98, 103, 120, 121, 127
scattering .............................................. 12, 14 transformation BSee trace,transformation A
service ......................................................... 10 transmitter............................................ 38, 57
slave ............................... 41, See master-slave transmitter block .................................. 38, 57
software installation ...................................24 transmitter module .................................... 37
source-detector distance ........ 38, 68, 75, 136 trend mode ........................................... 51, 95
spectroscopy ..................................... See NIRS TSI46, 60, 77, 87, 98, 110, 147
spectrum graph................. 47, 82, 90, 92, 118 optode holder ...................................... 150
stack ...................................................... 40, 41 slope .................................................... 117
status bar .................................................... 47 TSI QCF ............................................... See QCF
status lamp ..................................... 37, 39, 43 USB ..................................... 29, 37, 41, 43, 68
subtemplate ................................................ 78 USB connection.................................. See USB
Template DAQ State ...................................62 venous occlusion ........................................ 17
template file See measurement: template file wavelength ................... 12, 21, 37, 38, 57, 98
templatefile ................................................ 53 zoom ............................................. 90, 95, 160
175