Professional Documents
Culture Documents
Sensors 2409312 Peer Review r1
Sensors 2409312 Peer Review r1
Fernando Garnacho 1,2, Fernando Álvarez2, Alf Elg3, Christian Mier4, Kari Lahti5, Abderrahim Khamlichi1,2, Álvaro 4
Camuñas2, Joni Klüss3, Armando Rodrigo Mor6, Pertti Pakonen5, JoséRamón Vidal1, Eduardo Arcones2, Jorge 5
Rovira1, Carlos Vera2, Miran Haider3 6
1 FFII-LCOE, Eric Kandel Street 1, Getafe, 28906 Madrid, Spain; fernando.garnacho@ffii.es (F.G.); ak@ffii.es 7
(A.K.); jose.vidal@ffii.es (J.R.V.); jrovira@ffii.es (J.R.) 8
2 Universidad Politécnica de Madrid, Ronda de Valencia 3, Madrid 28012, Spain 9
3 RISE Research Institutes of Sweden, Borås 501 15, Sweden 10
4 Delft University of Technology, Mekelweg 4, 2628 CD Delft, Netherlands; c.mierescurra@tudelft.nl 11
5 Tampere University (TAU), Korkeakoulunkatu 3, 33720 Tampere, Finland; 12
6 Universitat Politècnica de València, Instituto de Tecnología Eléctrica, Camino de Vera s/n, 46022 Valencia, 13
Spain 14
15
Abstract: On-site partial discharge (PD) measurements have turned out to be a very efficient tech- 16
nique to determine the insulation condition in high-voltage electrical grids (AIS, cable systems, GIS, 17
HVDC converters, etc.), however, there is not any standardized procedure to determine the perfor- 18
mances of PD measuring systems. In on-line and on-site PD measurements, high frequency current 19
transformers (HFCTs) are commonly used as sensors, as they allow monitoring over long distances 20
in high-voltage installations. To ensure the required performances, the metrological qualification of 21
the PD analysers is necessary applying an evaluation procedure. A novel evaluation procedure has 22
been established to specify the quantities to be measured (electrical charge and PD repetition rate) 23
and to describe the evaluation tests considering the measuring influence parameters: noise, charge 24
amplitude, pulse width, and time interval between consecutive pulses. This procedure has been 25
applied to different types of PD analysers used for off-line measurements, sporadic on-line meas- 26
urements and continuous PD monitoring. The procedure was validated in a round robin test involv- 27
ing two metrological institutes (RISE from Sweden and FFII from Spain) and three universities 28
(TUDelft from the Netherlands, TAU from Finland and UPM from Spain). With this round robin 29
test the effectiveness of the proposed qualification procedure for discriminating between efficient 30
and inappropriate PD analysers has been demonstrated. Furthermore, it has been shown that the 31
PD charge quantity can be properly determined for on-line measurements and continuous monitor- 32
ing, by integrating the pulse signals acquired with HFCT sensors. In this case, these sensors must 33
have a flat frequency spectrum in the range between several tens of kHz and at least two tens of 34
MHz, where the frequency pulse content is more significant. The proposed qualification procedure 35
can be useful for improving the future versions of the technical specification TS IEC 62478 and the 36
standard IEC 60270. 37
Keywords: partial discharges; cable insulation; GIS; condition monitoring; insulation testing; qual- 38
ification procedure; standardization; HVDC - HVAC transmission and distribution grids; off-line 39
measurement; on-line measurement 40
41
42
43
1. Introduction 44
Although there are still challenges to overcome for the improvement of off-line and on-line PD meas- 45
urements in HVDC grids, PD measurements have proven to be very useful to detect insulation defects in 46
HVAC installations [1-7]. The most critical insulation defects that generate PD activity in electrical grids are 47
internal or void type defects and internal surface defects. Less consideration should be given to external 48
surface defects or to floating potential defects and when the corona effect is detected. Among the inherent 49
drawbacks of PD measurements in HVAC are the lack of applicable standards, absence of reference insula- 50
tion diagnosis criteria, difficulty to detect, separate and locate PD sources under presence of electric noise. 51
These difficulties are also present in HVDC with the added inconveniences of the low generation of PD 52
pulses and the high levels of electronic noise present. In HVDC, except with the corona insulation defect, 53
PDs appear mainly when the voltage value changes, e.g. due to the polarity inversion, voltage fluctuations 54
or when a temporary or transient overvoltage occurs [8-12]. Besides, the existence of pulsating noise gener- 55
ated in the power electronic converters of the substations makes the measurement especially difficult. Mon- 56
itoring applications, particularly those operating in HVDC, require efficient HFCT sensors, digital recorders 57
and PD signal processing tools (hereinafter referred to as PD analysers), for the adequate detection of PDs 58
that may travel distances of several kilometres until they arrive to the measuring point, where the cable 59
sheaths are accessible [13]. In this sense, to assure adequate PD measurements and analysis, the qualification 60
of the PD analysers is necessary. Furthermore, the establishment of a homogeneous criterium for the quali- 61
fication process is also required. In on-site PD tests, special attention must be paid in the influence factors 62
that affect to the measurements. These influence factors can be summarized as follows: the background 63
noise, the PD magnitude variability, the PD pulses width and the minimum time interval between consec- 64
utive pulses. The previous influence factors have been considered in the set of four tests proposed in this 65
research which are described in section 5. For the qualification of PD analysers, robust measurements and 66
appropriate requirements are considered in these tests. The stochastic behaviour associated with PD activity 67
does not allow the correct qualification of PD analysers when real insulation defects generated for example 68
in test cells are used. For the correct performance of these tests, controllable series of PD pulses and noise 69
signals artificially generated are required. In this research an industrially developed portable synthetic PD 70
calibrator was used [14]. Special requirements are established depending on the applicability of the PD an- 71
alyser under qualification: for off-line PD measurements, for on-line sporadic PD measurements or for con- 72
tinuous PD monitoring. To perform the four metrological qualification tests proposed, several PD pulse 73
trains and electric noise signals are generated by means of this synthetic PD calibrator, that uses an arbitrary 74
wave generator with an impedance load of 50 and 400 MHz bandwidth [14]. 75
76
In section 2 the reference PD pulse trains and noise signals used in the metrological tests are described. 77
In section 3 the importance of knowing with accuracy the pulses charge value is explained. Different charge 78
measurement methods are presented in section 4. The metrological tests are described in section 5. The prac- 79
tical cases implemented in the round robin test for the metrological qualification of PD analysers are shown 80
in section 6. Finally, in section 7 the conclusions are presented. 81
82
90
Where ipeak is the peak value of the current pulse i(t) and α and β two time constants. The parameter k 91
is obtained with the following equation. 92
Sensors 2023, 23, x FOR PEER REVIEW 3 of 16
+ +
k= (2) 93
The charge value of this current pulse can be calculated by equation (3): 94
q = i( t ) dt = i peak i p.u.( t ) dt = i peak TPD (3) 95
0 0
107
Figure 1. Reference IDE pulse with a TPD of 75 ns, with 1/=44 ns and 1/=9.9 ns. 108
A summary of the parameters defined for the pulse trains to be generated for the qualification tests is 109
shown in Table 1. Furthermore, for all metrological tests except for the “resolution time test”, the trains 110
were made up of consecutive bursts of four pulses of equal amplitude, generated every 10 ms (400 pulse/s) 111
with 1 ms between them. The “linearity test” is performed generating seven PD trains with the reference PD 112
pulse, but with different charge values from 10 pC to 2,4 nC. For the “resolution time test” the pulses of each 113
train are generated at equal time intervals from 2,500 s to 10 s. Depending on the metrological test to be 114
performed on a specific PD analyser, a different type of noise signal must be superimposed on the PD trains. 115
The characteristics of the noise signals used in the metrological tests are presented in the next subsection. 116
Table 1. Summary of the main parameters of the PD pulse trains and noises used in the metrological tests. 117
(1) The frequency spectrum of noise #3 changes every time interval T during the whole test time. 154
155
The type and amplitude of noise signal to be superimposed to the PD trains for the metrological tests 156
are specified in Tables 1 and 2. The noise amplitudes show in Table 1 are referred to the maximum amplitude 157
of the pulses of the PD trains generated. Noise signals and PD pulse trains used in the metrological tests are 158
generated using the synthetic PD calibrator. This calibrator injects pulsating signals (PD pulses and pulsing 159
noises) and non-pulsing noises into a current loop [14]. The HFCT sensor of the qualifying PD analyser can 160
be installed in the current loop where the pulse trains can be configured with the required charge, PD repe- 161
tition rate and different PD pulse widths. 162
3. Importance of the PD charge value and considerations about the estimation method 163
In PD diagnosis, when a defect is detected its level of criticality is determined analysing among other 164
variables, the pulse repetition rate (PD rate) n and the accumulated charge quantity(qa) over time [31]. Thus, 165
these two variables are very important to assess the insulation condition of HV electrical installations. The 166
charge values of the PD pulses, q, are visualized in PD patterns for HVAC and HVDC measurements and 167
also in graphics and histograms for HVDC measurements. The accumulated charge is determined consid- 168
ering the charge of all pulses exceeding a specified threshold level in a considered time interval. As the 169
charge value of a PD pulse gives to the expert analyst a valuable information to perform accurate diagnosis, 170
the estimation with accuracy of this parameter should be checked in the metrological qualification tests of 171
PD analysers. 172
173
The PD charge measurement along a cable system is a very important parameter because the charge 174
of PD pulses traveling through the cable sheath of a transmission HV cable system remains almost constant. 175
When the PD pulses travel along the cable their amplitude attenuates and their width increases almost in 176
the same proportion, keeping nearly constant its charge. The previous statement can be demonstrated taking 177
into account the current pulse i(t) along the cable in the frequency domain Ix(). 178
()·𝑥
𝐼𝑥 () = 𝐼𝑜 () · 𝑒 − (5)
Where 179
𝑇𝑃𝐷 (𝑛𝑠)
𝑞 (𝑝𝐶) = 𝑚𝑉 · 𝑢𝑝𝑒𝑎𝑘 (𝑚𝑉) (9)
𝑍𝑠 ( )
𝑚𝐴
In this approach the charge value, q, depends on the pulse width (TPD), which is well known in many 209
cases. In addition, it also depends on the frequency spectrum of the sensor transfer impedance Zs, which 210
should remain constant up to the PD pulse cut-off frequency. The TPD of the PD pulse can be estimated using 211
the voltage signal at the output of the HFCT sensor if its transfer function does not provoke significant signal 212
distortion. 213
214
2
10
ZHFCT(s)
1
10
|Zt|(mV/mA)
Zm(s)
0 I(s)
10
TPD=30 ns
TPD=150 ns
-1
10
2 3 4 5 6 7 8
10 10 10 10 10 10 10
f (Hz)
230
Figure 2. Frequency spectrum of two pulses with TPD=150 ns and TPD=30 ns, respectively. Bandpass filter 231
A (frequency range below 1 MHz, f1=610 kHz and f2= 770 kHz). Bandpass filter B (frequency range above 232
1 MHz, f1=1.75 MHz and f2=3.25 MHz). Transfer impedance of a HFCT sensor ZHFCT (s). 233
LCOE
30
f 11.75MHz f 2=3.25MHz
20 f 1=610kHz f 2=770kHz
10
0
(%)
-10
-20
-30
-40
0 50 100 150
TPD(ns)
234
Figure 3. Theoretical charge errors versus pulses with different TPD resulting of applying the bandpass filter A 235
(f1=610 kHz and f2= 770 kHz), red line, and the bandpass filter B (f1=1.75 MHz and f2=3.25 MHz), blue line. 236
𝑞 = 𝐼( = 0) (13)
The integration in the frequency domain is the most accurate approach, but it is more difficult to apply. 242
This is because it requires the characterization of the HFCT sensor to determine the frequency spectrum of 243
its transfer impedance and to perform a complex signal processing using the signal raw data of the PD. If 244
the transfer function of the HFCT sensor, ZHFCT is very flat for the entire PD pulse frequency spectrum, the 245
waveform reconstruction is not needed. This charge estimation method can give very acceptable results as 246
it is shown in section 6. 247
248
5. Metrological tests description 249
For the metrological qualification of PD analysers four tests are proposed. These tests are performed 250
using the synthetic calibrator [14]. Their realization enables the detection of measuring errors caused by PD 251
analysers, due to different influence parameters. 252
Sensors 2023, 23, x FOR PEER REVIEW 8 of 16
253
Before the performance of the metrological tests the PD analyser to be checked must be adjusted and 254
calibrated. In the adjustment process the analyser cut-off frequencies must be selected to perform the meas- 255
urements in the frequency band less affected by the noise signals. This process is carried out generating a 256
train of reference pulses of 500 pC superimposed to the noise signal to be used in the metrological tests 257
(noise type #1, #2 or #3). Then, the analyser must be calibrated with a train of reference PD pulses of 200 pC 258
without the noise signal. 259
260
The first test, called Noise rejection, is carried out to determine the error values in the charge (q) and 261
PD repetition rate (n) values, when various conditions of non-impulsive background noise are superim- 262
posed. The second and third tests, called Linearity and PD Time, respectively, are performed to determine 263
the error values of the charge quantity for different PD pulse amplitudes and widths, respectively. And the 264
fourth test, called Resolution time, is carried out to determine the PD rate error value of a PD analyser under 265
qualification as function of the PD repetition frequency (N). 266
267
the applicability of the PD analyser, the amplitude of the superimposed noises is chosen according to the 304
percentages shown in Table 1. The linearity error influence (εl) is calculated as half the difference between 305
the maximum charge error value and the minimum charge error value. The linearity error is determined 306
for two charge ranges, from 50 pC to 2.4 nC and from 10 pC to 2.4 nC. 307
308
Maximum acceptance errors for the scale factor linearity. In the standard IEC 60270 it is indicated 309
that the maximum deviation of the scale factor k must be lower than ±5 %. This maximum acceptance error, 310
valid for PD analysers used in off-line tests, has been adopted also for PD analysers used in on-line sporadic 311
tests or in continuous monitoring. 312
Three different PD analysers operating for off-line measurements were qualified (A-1, A-2 and A-3). 353
All of them operate according to the wideband method with an upper frequency limit f 2 1 MHz. One of 354
them applies an upper frequency limit f2<500 kHz in accordance with the current IEC 60270. The other two 355
operate with f2 > 500 kHz according to the new revision of IEC 60270. The lower and upper frequency limits 356
of each PD analyser are shown in Table 3. The results of these qualification tests are also shown in Table 3. 357
To carry out metrological tests the non-pulsating noise #1 was superimposed on the generated pulse trains. 358
359
In the noise rejection test, the PD analyser A-3 meets the maximum acceptance errors considered for 360
the charge value (Max ABS q 10%) and for the PD repetition rate (Max ABS n 2%), with 50% of noise #1 361
level. These good results were mainly due to the appropriate selection of the lower and upper frequency 362
limits f1 and f2. The wideband measurement was tuned in a frequency range where the amplitude of the 363
noise frequency spectrum was lower. The results obtained with A-2 and A-3 PD analysers were very affected 364
by the noise signal, due to the measurements are performed in a fixed wideband where the noise spectrum 365
amplitude was significant. As the noise signal #1 is high for frequencies below 500 kHz (see table 2), the 366
errors increase the lower are the values of f1 and f2. Therefore, the results obtained for A-1, with f1=40 kHz 367
and f2=400 kHz, are more affected by the noise signal. 368
369
In the linearity test, the PD analyser A-3 also meets the maximum acceptance error in the two charge 370
ranges considered. The larger errors obtained for A-1 and A-2 are also affected by the noise signal. 371
372
In the PD Time test, all the analysers met the maximum acceptance error in the TPD range considered. 373
The TPD errors (εTPD) for the three PD analysers were 2.5%±0,1%, this figure is consistent with the theoretical 374
error estimated in subsection 4.2 (see Fig.3) for this type of measurements which was ±4%. 375
376
In the Resolution time test, the best result was obtained for the A-3 analyser with a resolution time of 377
10 µ s, for which the PD repetition rate error was 1,3 %, lower than the maximum acceptable error of 2%. 378
379
In conclusion, the PD Analyser A-3 is the only one that meets all the maximum acceptable errors (10% for 380
charge measurements and 2% for PD repetition rate measurements). PD Analysers A-2 and A-3 are very affected 381
by the noise. 382
383
Table 3. Metrological test results of three PD analysers for off-line PD measurements under noise #1. 384
PD Analyser for off-line
Analyser A-1 Analyser A-2 Analyser A-3
measurements
f1-f2 (kHz) 40 – 400 40 – 800 610 – 770
f (kHz) 360 760 160
Charge measurement method Quasi-integration Quasi-integration Quasi-integration
Manual selection
Filtering method None trigger level
of f1, f2 and
Metrological tests
Charge & PD repetition rate errors versus noise level
Noise amplitude
Requirements
(%)
Max q 10%; Max n 2% under 50% noise level
q=248% q=151% q=61.9%
200
n>1000 % n>1000 % n=-0,4%
Noise
rejection q=104.3% q=53,9% q=32.0%
100
q=100 pC n>1000 % n>1000 % n=-0,4%
q=27% q=26% q=2.6%
50
n>1000 % n>1000 % n=-0,4%
q=7,8% q=9,6% q=0,5%
20
n>1000% n=496% n=-0,4%
Sensors 2023, 23, x FOR PEER REVIEW 11 of 16
385
Noise Charge error & PD repetition rate errors versus noise level
amplitude Requirements
(%) Max q 30%, Max n 2% under 200% noise level
q= 8.0 % q=-0.9 % q=-36.5% q=16.4%
200
Noise n=0.0 % n=0.0 % n=0.7 % n=0,0%
rejection q=3.2 % q=-0.4 % q=13.0% q=8.1%
q=100 pC 100
n=0.0 % n=0.0 % n=0.7 % n=-0,0%
q=-0,9% q=-0.2 % q=10.4% q=3.3%
50
n=0.0 % n=0.0 % n=0.7 % n=-0,0%
q=0,2 % q=-0.1 % q=1.7% q=1,0 %
20
n=0.0 % n=0.0% n=0.7% n=-0,0%
Charge Linearity error = l = (Max q value – Min q value) / 2
Linearity range (pC) Requirement l 5% under 35% noise level
under 35% noise 50 to 2,400 ±0.3 % ±0.6% ±13,2 % ±3,6 %
10 to 2,400 ±0.6 % ±12.7% ±14,0 % ±4,1 %
TPD range TPD error = TPD = (Max q value – Min q value) / 2
PD Time (TPD) (ns) Requirement TPD 30% under 35% noise level
under 35% noise 37.5 - 150 ±25.1 % ±2.0 % (*) ±2.8 % ±25.4 %
q=200 Pc Not applicable
8 to 150 ±29.1% ±4.0 % ±29.6%
(**)
Resolution time tres range PD repetition rate error
under 35% noise (s) Requirement Max n 2% under 35% noise level, at least 20s
q=200 pC 2,500 – 10 10 s, n=-0.4 % 20 s, n=0.0 % 2.5 ms, n=0.0 % 10 s, n=0.0 %
(*) In the PD Time test, the direct readings were corrected using the equation (3) considering the nominal TPD values. For this 408
reason, the results obtained with this analyser cannot be considered as independent measurements. 409
(**) It is not applicable because the digital recorder used a sampling rate of 10 ns. 410
411
In the PD Time test, all PD analysers met the maximum acceptance error in the TPD range from 37.5 to 150 ns 412
considered. The maximum TPD error (εTPD) was ±29.6 %, which is slightly lower than the maximum permitted error 413
(30%). The TPD errors (εTPD) obtained with B-1 and B-4 (±29.1 % and 29.6 %, respectively) are consistent with the 414
theoretical one estimated in subsection 4.2 for these measurements, which was ±25.1%. The lowest T PD error was 415
achieved with the B-3 PD analyser, that determines the charge value by integration of the acquired pulses in the 416
time domain. With the B-2 PD analyser reasonable results were achieved, but the readings were corrected with 417
formula (1) using nominal TPD values. For this reason, the results obtained with this analyser cannot be considered 418
as independent measurements. 419
420
In the Resolution time test, the analysers B-1, B-2 and B-4 meet the minimum resolution time required of 421
20 s. The best result of 10 µ s was achieved using the analysers B-1 and B-4. A repetition rate error of 0% was 422
obtained with Analyzer B-4 (see Table 4). 423
424
6.3. Metrological qualification of PD analysers for continuous PD monitoring 425
The same first three PD Analysers used for online sporadic measurements (B-1, B-2, and B-3) along with a 426
new one C were tested as PD analysers used for continuous monitoring. To carry out metrological tests the non- 427
pulsating noise #1 was superimposed on the generated pulse trains. 428
429
The upper and lower frequency limits of B-1 and B-4 were tuned for a better charge measurement, while the 430
filters of PD analysers B-2 and B-3 were the same previously used in sporadic measurements (see table 5). With 431
the new PD analyser “C” an automatic wavelet filter was used for the noise rejection test and the quasi-integration 432
approach was applied for charge measurements. The results of these qualification tests are also shown in Table 5. 433
Sensors 2023, 23, x FOR PEER REVIEW 13 of 16
434
In the noise rejection test, the results obtained for all PD analysers are very similar to those obtained with 435
the on-line sporadic measurements (subsection 6.2). All PD analysers, except B-3 meet the requirements. 436
437
In the linearity test, as occurred with the for on-line sporadic measurements, the most significant errors 438
were obtained for analysers B-2 and B-3. All PD analysers meet the linearity requirements for the charge range 439
from 50 pC to 2,400 pC, but only the PD analyser B-1 and B-4 meet the requirements for the charge range from 10 440
pC to 2,400 pC. 441
442
In the PD Time test, the maximum TPD error (εTPD) was ±29,8 %, which is slightly lower than the maximum 443
permitted error (30%). PD analysers B-2 and B-3 are more affected by the noise signal, especially when low charge 444
values are measured (20 pC). However, as in the case of on-line sporadic PD measurements, the lowest TPD error 445
was achieved with B-3. This analyser determines the charge value integrating the pulses in the time domain. The 446
B-2 PD analyser achieved reasonable results, but the readings were corrected with formula (3) using the nominal 447
TPD values. For this reason, the results obtained with this analyser cannot be considered as independent measure- 448
ments. 449
450
In the Resolution time test, as in the case of on-line sporadic measurements (subsection 6.2), the B-1, B-2 451
and B-4 meet the minimum resolution time required of 20 s. The best result of 10 µ s was achieved using the 452
analysers B-1 and B-4 (see Table 5). 453
454
In conclusion, for continuous PD monitoring, the analyser B-1 and C are the only two ones that meet all the 455
maximum acceptable errors (30% and 2% for charge and PD repetition rate measurements, respectively). How- 456
ever, charge errors can be improved by signal integration similar to the approach applied by Analyser B-3. With 457
the analyser B-2 should calculate by itself the TPD value to meet all the requirements. The PD Analyser B-3 is very 458
affected by the noise. 459
Table 5. Metrological tests results of PD analysers operating for continuous PD monitoring. 460
Metrological tests
Noise Charge error & PD repetition rate errors versus noise level
amplitude Requirements
(%) Max q 30%, Max n 2% under 200% noise level
Noise
rejection q= 4.1 % q=-0.4 % q=-40,0% q=-11,7%
200
q=100 pC n=0.0 % n=0.0 % n=0.7 % n=-0,0%
q=0.8 % q=0.0 % q=16,5% q=-1,6%
100 n=0.0 % n=0.0 % n=0.7 % n=-0,0%
50 q=-0.1% q=0.1 % q=8,7% q=1,7%
Sensors 2023, 23, x FOR PEER REVIEW 14 of 16
7. Conclusions 465
A procedure for the metrological qualification of PD analysers has been proposed and tested, determining 466
measurement errors caused by the influence parameters. A set of metrological tests were defined along with spe- 467
cific acceptance requirements. These tests were defined considering all possible applications of PD analysers, 468
which were classified into three categories: off-line PD measurements, sporadic on-line PD measurements, and 469
continuous PD monitoring. The measurement conditions and acceptance requirements are adapted to the type of 470
application of the analyser. The allowable errors for each test were selected considering the IEC 60270 standard 471
requirements, the Technical Specification IEC 62478 and the inherent noise conditions of off-line, on-line sporadic 472
measurements and continuous monitoring. For the practical validation of the qualification procedure, several PD 473
analysers were characterized in a European round robin test within the framework of a EURAMET research pro- 474
ject. 475
476
For PD analysers operating in off-line measurements, the requirements related to the noise rejection test 477
(Max q 10% and Max n 2% with 50% noise signal) are meet if the cut-off frequency limits are set within a 478
bandwidth according to the future IEC 60270 (1 MHz f2 500 kHz), where the signal-to-noise ratio is larger. 479
480
PD analysers used in on-line sporadic measurements or in continuous monitoring must operate with cut- 481
off frequencies above 1 MHz to avoid as much as possible the influence of noise signals, since these usually have 482
large amplitude levels below 1 MHz. In sporadic on-line measurements the cut-off frequencies adjustment can be 483
manual, as no significant changes are expected in the noise frequency spectrum during a PD measurement. How- 484
ever, in continuous PD monitoring, as the noise behaviour can change over time, an automatic filtering tool should 485
be used, such as an automatic wavelet filter. When on-line PD sporadic measurements are performed, the selection 486
of the cut-off frequencies of the passband filter has a significant influence on the results. 487
488
The realization of the round robin test has been useful for the validation of the proposed qualification pro- 489
cedure. The results of the round-robin test show that the proposed acceptance requirements can be achieved if 490
some technical aspects are considered. For example, the adequate selection of the cut-off frequencies, the use of 491
automatic noise filtering tools or the appropriate current signal integration for the charge value determination. 492
The reference qualification procedure proposed will be useful to improve the features of PD analysers and to 493
support the development of the future versions of PD standards. 494
495
Sensors 2023, 23, x FOR PEER REVIEW 15 of 16
To achieve a complete qualification of PD analysers, the metrological tests presented in this research should 496
be complemented with diagnosis tests for diagnostic tools. In this sense, the authors have developed another 497
procedure, also within the scope of this EURAMET project, for the qualification of PD analysers diagnostic tools 498
(PD recognition, PD clustering and PD location). This procedure will be presented in a future paper. 499
500
Author Contributions: Fernando Garnacho, Abderrahim Khamlichi and Carlos Vera have defined the metrological procedure, 501
the acceptance tests, and the requirements for the qualification of a synthetic generator. Fernando Á lvarez has supported them. 502
Alf Elg, Armando Rodrigo, Kari Lahti, Joni Klüss, Christian Mier, José Ramón Vidal, Á lvaro Camuñas, Pertti Pakonen, Jorge 503
Rovira, Eduardo Arcones and Miran Haider reviewed the proposed qualification procedure. José Ramón Vidal, Christian Mier, 504
Fernando Á lvarez, Joni Klüss and Kari Lahti carried out the round robin metrological tests. Fernando Garnacho carried out 505
the analysis of the round robin results and performed the paper writing. All the authors of the article participated in the 506
discussion of the results and in the revision of the article, a special mention to Fernando Á lvarez. Jorge Rovira has participated 507
in the editing and layout of this paper. 508
Acknowledgments: The research presented in this paper has received funding from the project 19ENG02 FutureEnergy of the 509
EMPIR programme, co-financed by the Participating States and from the European Union’s Horizon 2020 research and inno- 510
vation programme. 511
References 513
514
1. Sikorski, W.; Wielewski, A. Low-Cost Online Partial Discharge Monitoring System for Power Transformers. Sensors 2023, 515
23, 3405. 516
2. Á lvarez, F.; Garnacho, F.; Ortego J.; Sánchez-Urán, M.A. Application of HFCT and UHF sensors in on-line partial discharge 517
measurements for insulation diagnosis of high voltage equipment. Sensors 2015, 15. 518
3. Á lvarez, F.; Albarracín, R.; Garnacho,F.; Granizo. R. Diagnosis of insulation condition of MV switchgears by application 519
of different partial discharge measuring methods and sensors. Sensors 2018, 18. 520
4. Sheng, B.; Zhou, C.; Hepburn, D.M.; Dong, X.; Peers, G.; Zhou, W.; Tang, Z. Partial discharge pulse propagation in power 521
cable and partial discharge monitoring system. IEEE Transactions on Dielectrics and Electrical Insulation 2014, 21. 522
5. Bassan, F.R.; Rosolem, J.B.; Floridia, C.; Penze, R.S.; Aires, B.N.; Roncolatto, R.A.; Peres, R.; Júnior, J.R.N.; Fracarolli, J.P.V.; 523
da Costa, E.F.; Cardoso, F.H.; Pereira, F.R.; Furoni, C.C.; Coimbra, C.M.; Riboldi, V.B.; Omae, C.; de Moraes, M. Multi- 524
Parameter Optical Monitoring Solution Applied to Underground Medium-Voltage Electric Power Distribution Networks. 525
Sensors 2023, 23, 5066. 526
6. Hu, X.; Zhang, G.; Liu, X.; Chen, K.; Zhang, X. Design of High-Sensitivity Flexible Low-Profile Spiral Antenna Sensor for 527
GIS Built-in PD Detection. Sensors 2023, 23, 4722. 528
7. Riba, J.-R. Application of Image Sensors to Detect and Locate Electrical Discharges: A Review. Sensors 2022, 22, 5886. 529
8. P.H.F. Morshuis; Smit, J.J. Partial discharges at DC voltage: their mechanism, detection and analysis. IEEE Transactions on 530
Dielectrics and Electrical Insulation 2005, 12 531
9. Morshuis P.; Jeroense, M.; Beyer, J. Partial discharge. Part XXIV: The analysis of PD in HVDC equipment. IEEE Electrical 532
Insulation Magazine 1997,13. 533
10. Reference: D1-306_2016. SESSION MATERIALS. A development of DC PD pattern recognition method using modified 534
CAPD together with PD founding for gas insulated apparatus under DC voltage. Paris, France, 2016. 535
11. Wang, G.; Jo, H.E.; Kim, S.J.; Kim, S.W.; Kil, G.S. Measurement and analysis of partial discharges in SF6 gas under HVDC, 536
Measurement. 2016, 91, 351-359. 537
12. M. A. Fard, M. E. Farrag, S. G. McMeekin and A. J. Reid, "Partial discharge behavior under operational and anomalous 538
conditions in HVDC systems," in IEEE Transactions on Dielectrics and Electrical Insulation, vol. 24, no. 3, pp. 1494-1502, 539
June 2017. 540
13. Winkelmann, E.; Shevchenko, I.; Steiner, C.; Kleiner,C.; Kaltenborn, U.; Birkholz, P.; Schwarz, H.; Steiner, T. Monitoring 541
of Partial Discharges in HVDC Power Cables. IEEE Electrical Insulation Magazine 2022, 38. 542
14. Garnacho, F.; Khamlichi, A.; Á lvarez, F.; Ramírez, A.; Vera, C.; Rovira, J.; Simón, P.; Camuñas, A.; Arcones, E.; Ortego, J. 543
Best practices for Partial Discharge Monitoring of HVDC Cable Systems and Qualification Tests. CIGRE SCIENCE & EN- 544
GINEERING, a Scopus registered magazine (CSE) 2023, 27, 1-14, 2426-1335. 545
15. Koltunowicz, W.; Plath, R. Synchronous multi-channel PD measurements. IEEE Transactions on Dielectrics and Electrical 546
Insulation 2008, 15, 1715-1723. 547
16. Wang, Y.; Chen, P.; Zhao, Y.; Sun, Y. A Denoising Method for Mining Cable PD Signal Based on Genetic Algorithm Opti- 548
mization of VMD and Wavelet Threshold. Sensors 2022, 22, 9386. 549
17. Ma, X.; Zhou, C.; Kemp, I.J. Automated wavelet selection and thresholding for PD detection. IEEE Electrical Insulation 550
Magazine 2002, 18, 37-45. 551
Sensors 2023, 23, x FOR PEER REVIEW 16 of 16
18. De Oliveira, H.; Chaves, L.; Cunhas, T.; Vasconcelos, F.H. Partial discharge signal denoising with spatially adaptive wavelet 552
thresholding and support vector machines. Electric Power Systems Research 2011, 81, 644-659. 553
19. Kyprianou, A.; Lewin, P.L.; Efthimiou, V.; Stavrou, A.; Georghiou, G.E. Wavelet packet denoising for online partial dis- 554
charge detection in cables and its application to experimental field results. Measurement Science and Technology 2006, 17, 555
2367-2379. 556
20. Wang, Y.; Chen, P.; Zhao, Y.; Sun, Y. A Denoising Method for Mining Cable PD Signal Based on Genetic Algorithm Opti- 557
mization of VMD and Wavelet Threshold. Sensors 2022, 22, 9386. 558
21. Alvarez, F.; Ortego, J.; Garnacho, F.; Sanchez-Uran, M.A. A clustering technique for partial discharge and noise sources 559
identification in power cables by means of waveform parameters. IEEE Transactions on Dielectrics and Electrical Insulation 560
2016, 23. 561
22. Cavallini, A.; Montanari, G.C.; Puletti, F.; Contin, A. A new methodology for the identification of PD in electrical apparatus: 562
properties and applications. IEEE Transactions on Dielectrics and Electrical Insulation 2005, 12, 203-215. 563
23. Ardila-Rey, J.A.; Martínez-Tarifa, J.M.; Robles, G.; Rojas-Moreno. Partial discharge and noise separation by means of spec- 564
tral-power clustering techniques. IEEE Transactions on Dielectrics and Electrical Insulation 2013, 20, 1436–1443. 565
24. International Standard IEC 60270:2000+AMD1:2015 CSV Consolidated version. High-voltage test techniques - Partial dis- 566
charge measurements. TC 42 High-voltage and high-current test techniques. 567
25. Gulski, E. et al., “Guidelines for partial discharge detection using conventional (IEC 60270) and unconventional methods”, 568
CIGRE WG D1.37 technical brochure 662, Electra, 2016, 288. 569
26. Lemke, E. et al., “Guide for partial discharge measurement in compliance to IEC 60270 Std”, CIGRE WG D1.33 technical 570
brochure, 2008, ISBN: 978-2-85873-053-7. 571
27. Lemke, E. A critical review of partial-discharge models, IEEE Elec. Insul. Mag. 2012, 28. 572
28. Cavallini, A.; Montanari, G.C.; Tozzi, M. PD apparent charge estimation and calibration: A critical review. IEEE Trans. on 573
Dielec. and Elec. Insul. 2010, 17. 574
29. International Standard IEC TS 62478:2016. High voltage test techniques - Measurement of partial discharges by electromag- 575
netic and acoustic methods. TC 42 High-voltage and high-current test techniques. 576
30. Gulski, E. et al., “Guidelines for Unconventional Partial Discharge Measurements”, CIGRE WG D1.33 technical brochure 577
2010, ISBN: 978-2-85873-132-9. 578
31. Arcones, E.; Á lvarez, F.; Ortego, J.; Khamlichi, A.; Camuñas, A.; Rosete, A.; Garnacho, F.; Gómez, P. Development, testing 579
and aging of reference insulation defects for the improvement in partial discharges diagnosis. IEEE Electrical Insulation 580
Conference (EIC), Denver, CO, USA, 2021. 581