151 - Daae29 - qtl0054 Rev B Breathtracker Model SC Manual
151 - Daae29 - qtl0054 Rev B Breathtracker Model SC Manual
INSTRUMENT MANUAL
1. INTRODUCTION...................................................................................................5
2. THE PRINCIPLE BEHIND THE CO2 CORRECTION FACTOR................................5
3. DESCRIPTION OF CONTROLS...........................................................................6-7
A. FRONT PANEL COMPONENTS
B. REAR PANEL COMPONENTS
4. ACCESSORIES..........................................................................................................7-8
5. ENVIRONMENTAL CONDITIONS.........................................................................8
6. INSTALLATION...........................................................................................................8
7. CALIBRATION PROCEDURE..................................................................................9-14
8. ANALYZING PATIENT SAMPLES PROCEDURE........................................................15
9. SELF-CORRECTION FEATURE..................................................................................16
10. OTHER FEATURES AND DISPLAYS ASSOCIATED WITH NORMAL OPERATION. .17
FLOW MONITORING FEATURE
AUTOMATIC FLOW MONITORING FEATURE
13. TROUBLESHOOTING........................................................................................19
“ERR” DISPLAY
“OVR” DISPLAY
“MAX” DISPLAY
WARRANTY INFORMATION.........................................................................................22
CONTACT INFORMATION......................................................................................24
PRODUCT INFORMATION................................................................................25-32
When contacting the Customer Service Department, please indicate the device type:
BREATHTRACKER SC.
***BreathTracker SC Users***
Sample Drying Tube Conditioning Procedure:
Flush 60 mL of QuinGas-3 (QT07230-G) through each freshly filled sample flush drying tube.
This procedure will condition the Drierite and negate the effects of possible CO2
absorption in the drying tube for subsequent use in drying the patient samples.
Fresh indicating Drierite (QT01156-C) used in QuinTron’s sample flush drying tube
(QT01135-K) can absorb CO2 from a patient sample. Since the BreathTracker SC
utilizes CO2 concentration as a reference for correcting H2 and CH4, absorption of CO2
may lead to inaccurate correction factors unless the Drierite is conditioned.
This procedure must be followed each time a sample drying tube is refilled with Drierite.
Please do not develop or modify any collection techniques or devices without consulting QuinTron’s Customer
Service Department. QuinTron shall not be held responsible for any patient samples that have been jeopardized
or damage to your instrumentation by use of products not supplied by QuinTron.
We know for certain that syringes manufactured by “BD” have been found to be incompatible with our
instrumentation. When using products other than those supplied by QuinTron, it is very difficult for the
end- user to detect if the resulting sample is inadequate. If users are found to be using products not supplied
by QuinTron (i.e. stopcocks or syringes), QuinTron cannot support any interpretation help, servicing issues,
or technical support until all supplies are proven to be from QuinTron.
For an updated list of products that have been discovered to be incompatible with our instrumentation, please
visit our web site: www.quintron-usa.com
Please review all the information provided for collection and analysis of patient samples prior to attempting
collection or analysis of actual patient samples to minimize potential errors.
The BreathTracker SC is a stand-alone analyzer, which measures both hydrogen and methane in a gas
sample at levels of parts per million (ppm). The trace gas concentrations of the sample, and the values
calculated for alveolar concentrations, are presented in ppm on the 2-line instrument display.
CO2 is the physiological regulator of breathing and the whole breathing system is dedicated to keeping
the alveolar CO2-pressure (PACO2) constant at 40 mm Hg (torr). Therefore, CO2 is the most reliable
“normalizing” component in the sample because it ordinarily has the most constant alveolar composition of
any gas in the sample.
Alveolar PCO2 remains constant at 40 torr among normal individuals if ventilation is normal. The percent
of CO2 in an alveolar sample is affected by the barometric pressure (altitude) at which the sample is col-
lected. Alveolar air with a PCO2 of 40 torr in Miami (at sea level) will have a CO2 concentration of about
5.5% in dry air (40/(760-37)), while alveolar air in Denver (where barometric pressure is closer to 625 torr)
will be near 6.8% (40/(625-37)). Significant differences in barometric pressure exist at different altitudes,
as demonstrated by Miami and Denver. However, using a single correction factor, alveolar concentration
will simplify the process without introducing significant error, because all the samples will be normalized to
the
same (constant) CO2 level.
1 5
6
2
7
3
4
8
2. The UP/FLOW CHECK key is a dual function switch. In the Data mode, this switch increases the value of
the displayed cal gas concentration for H2, CH4, or CO2. In the Run or Cal modes, this switch allows the flow
rate to be checked.
3. The FUNCTION key is normally used only in the Run mode to select the Linearizing (½ cal) Function;
which is performed with a 50% dilution of the cal gas. Other functions can be performed using the Function
switch, but are not recommended for access in the normal operation of this instrument.
4. The DOWN/CORRECTED key is a dual function switch. In the Data mode, this switch decreases the value
of the displayed cal gas concentration for H2, CH4, or CO2. In the Run mode, this switch allows the user to
toggle the display back and forth showing the gas concentration values or the “corrected” gas values.
5. The ENTER key selects which mode in the Menu to enter (Run, Cal, or Data). It also steps the system
through the startup sequence, which is used to enter data values, and to place the system to the Run or Cal
mode.
6. The START key initiates a sample analysis when pushed, and returns the system to the “Ready”
mode when an analysis has been completed.
8. The ZERO key operates in the Run and Cal mode and is used to zero the values for H2, CH4, and CO2.
9. The SAMPLE PORT is used to inject the sample into the instrument.
1 2. Air-In Port
(Draws room air in through the SivRite-4 Bottle)
4
2 ON OFF
4. Main Power Switch
(Turns the instruments main power ON/OFF)
6
ON OFF
5 5. Power Connection
Power cord should be used for main disconnection (Insert/Remove the power cord)
6. Symbol Legend
Figure-3 - Rear Panel (Displays the symbols and their meanings)
4.
ACCESSORIES
YOU SHOULD RECEIVE ONE SET OF EACH ITEM BELOW: (NOT ALL ITEMS ARE SHOWN)
1 5 8
6
9
2
7
3 10
5. ENVIRONMENTAL CONDITIONS
This instrument has sensitivity to temperature and humidity changes.
Please keep the instrument away from:
• Direct sunlight
• Drafts from ventilation systems and windows
Careful consideration of the instrument’s environmental c
If you need turn off the instrument temporarily, remember to only turn off the pump switch and leave the
main power switch turned on. You should only turn the MAIN POWER switch off if you do not plan to use
the instrument for an extended period of time or need to relocate the instrument.
If you turn off the pump switch only, you need to allow the instrument to warm up for at least eight hours
before putting it into operation.
Leaving the pump switch turned on indefinitely allows for no warm up period, but may cause wear on the
pump apparatus. QuinTron recommends that if you have a breath-test scheduled, turn the instrument on the
day before the test to allow for the best warm up. If there are no further tests scheduled for the week then you
may turn off the pump switch.
1. Following the sufficient warm-up period, press the ENTER ( ) key. The following menu mode is displayed.
At this point, the instrument is in the menu mode. There are three functions that can be selected in the menu
mode, Run (analyzing samples), Cal (the system will automatically calibrate the instrument to the cal
values entered in the Data mode), and Data (where calibration values are entered).
Pressing the UP ( ) key steps the instrument to the next mode.
Refer to the label on the calibration gas tank for calibration values of H2, CH4, and CO2.
QuinGas 3 [QT07230-G] is recommended for calibrating the instrument.
3. Press the ENTER ( ) key to select the Data mode, then the following will display: [Figure-8]
4. Press the UP( ) key to increase the calibration value for H2 or press the DOWN ( ) key to decrease the
value. When the value on the display matches the calibration gas tank, press the ENTER ( ) key to
advance to the next screen.
5. Press the UP( ) key to increase the calibration value for CH4 or press the DOWN ( ) key to decrease the
value. When the value on the display matches the calibration gas tank, press the ENTER ( ) key to
advance to the next screen. [Figure-9]
6. Press the UP( ) key to increase the calibration value for CO2 or press the DOWN ( ) key to decrease the
value. When the value on the display matches the calibration gas tank, press the ENTER ( ) key to
advance to the next screen. [Figure-10]
7. [Figure-11] Press the UP ( ) key to increase the alveolar value for CO2 or press the DOWN ( ) key to
decrease the value. When the value on the display equals 5.5% press the ENTER ( ) key to exit the
Data mode, the instrument will automatically enter the Cal mode. See the section titled The Principle Behind
the CO2 Correction Factor in the front of this manual to determine if a setting other than 5.5% is appropriate.
8. The instrument can now be calibrated using gas from the calibration cylinder. If the display does not
indicate zero for each gas, press the ZERO ( ) key to zero all gas values.
9a. Insert the stopcock and syringe into the sample port on the lower panel of the BreathTracker SC. [Figure-
13] Open the stopcock and inject at least 20 mL of gas.
10. After injecting the calibration gas, press the START ( ) key on the front panel to start the
analysis. The following “CAL WAIT” display appears. Note the LED color is amber to indicate
processing.
11. As the system processes each gas, an audible train of sounds will indicate sensor tracking. The individual
gas concentrations will appear in their locations on the screen once they have finished analyzing.
When the instrument is finished processing the data, the following display will appear: [Figure-
15]
Note: Your numbers should match the values indicated on your calibration cylinder.
Figure-15 - Full Calibration Done
12. Note the LED is blinking red…this indicates the instrument is in a recovery period. When this recovery
period is over (approximately three minutes), the LED will start blinking green indicating the instrument is
ready for another sample. Press the START ( ) key to advance to the next step in calibration.
If the display does not indicate zero for each gas, press the ZERO ( ) key to zero all gas values. [Figure-17]
15. Inject the full 30 mL of this gas into the Sample Port and press the START ( ) key.
Normally, the BreathTracker instrument will automatically advance the user from the full calibration
analysis into the ½ calibration step. The ½ calibration process can also be accessed by entering the Run
Mode of the instrument and then pressing the Special Function key, then pressing the Enter key.
If during your ½ calibration procedure the “Flow Warning” screen appears, you
must adjust the flow rate on the rear of the instrument back to 120 mL/min (+/- 2
The entire calibration procedure must be performed
mL/min). Press the Enter Key to exit the flow warning screen.
before you run any patient samples for the day.
The instrument will now display the Run Ready screen and you should press the Special
Function Key to re-enter the ½ calibration screen.
You may calibrate as often as you wish during the day. On average, QuinTron recommends at least 2 calibrations per
day if you are analyzing multiple patients over an eight (8) hour period.
If you are experiencing any difficulties or have any questions regarding calibration please contact QuinTron.
Remember to only use QuinTron products to ensure proper collection and accurate analysis results.
If the display does not indicate zero for each gas, press the ZERO ( ) key to zero all gas values.
Inject at least 20 mL of the patient sample through the Sample Drying Tube and into the sample port of the
instrument and press the START ( ) key. [Figure-21]
Water/Dust Trap
Patient Sample Sample Drying Tube
As the system is processing each gas, an audible train of sounds will indicate sensor tracking, and the individual
gas concentrations will appear in their locations as each one has finished analyzing.
When the instrument has completed processing the data, the Run Done display will appear: [Figure-22]
These are your raw values and still need to be checked for sample contamination. (See Section 10)
NOTE: These numbers are for demonstration purposes only; your actual values will vary from sample to sample.
To use the self-correction feature of the BreathTracker SC, press the DOWN ( ) key. The following display
will appear indicating the corrected values for H2 and CH4 and the correction factor that was applied. In this
case, the correction factor is 1.5, H2 is 100 (67 x1.5) and CH4 is 50 (33x1.5). Note: The display for CO2 is
unchanged. (Pressing the DOWN ( ) key again will toggle the display back to the original data above).
[Figure-23]
Record the patient’s sample values, then press the START ( ) key when the status LED begins to blink
green. After pressing the Start Key the instrument automatically steps into the Run Ready mode and is ready for
more patient samples. Follow sections 9-10 until you have finished all your patient samples.
In the case where the correction factor is too high, the following display is shown when the DOWN ( ) key is
pressed. [Figure-24]
The BreathTracker SC has built-in flow monitoring to continually monitor the internal flow rate (which is set
to 120 mL/min +/- 3 mL/min) and displays the following warning when the flow rate falls out of this range.
When the above display type appears, the operator must adjust the flow rate using the Flow Control on the rear
of the panel to 120 mL/min +/- 3mL/min. When the flow rate is set to the correct value as in this display above,
the operator can press the ENTER ( ) key to return to the Run Mode. [Figure-20]
After an analysis session is completed and no further samples will be analyzed for the rest of the day, you may
turn the PUMP SWITCH off, which will place the instrument in standby. [Figure-3]
You should leave the MAIN POWER SWITCH on at all times even when the instrument is in standby, unless
you do not plan on running any gas samples for at least two days.
If regular preventative maintenance is not performed, your instruments warranty will be voided.
Water/Dust Trap:
- Make sure that the water/dust trap is always in place on the Sample Port (located on the front of the
instrument) and on the SivRite-4 Bottle.
Replace both traps every 6 months or more frequently if the inside of the trap becomes discolored or dirty.
SivRite-4 Bottle:
- Replace the SivRite-4 Bottle when the silica beads turn from orange to green to the line indicated on the label,
Do not dispose of your SivRite-4 Tubing assembly.
If your flow rate suddenly drops significantly, first check to make sure that both port plugs on the SivRite-4
bottle have been removed. Then verify that the tube from the SivRite-4 bottle is not obstructed or bent. If the
tubing from the bottle to the machine is deemed adequate and you are still experiencing an extremely low
flow
rate, proceed with the following instructions:
Before making any radical flow rate adjustments with the screwdriver, turn the main power to the instrument off
for ten seconds and then back on, resetting the unit.
If the flow rate warning does not automatically appear, press the UP ( ) key to check the flow rate and verify
that it is in proper operating range.
If the instrument is still not in the proper operating range, you may adjust the flow rate with the QuinTron
screwdriver by inserting it into the flow rate adjust slot on the back of the instrument.
When the flow rate returns to the necessary operating value (120 mL/min +/- 2 mL/min), press the ENTER ( )
key to return to the Run mode.
FLOW WARNING APPEARS DURING 1/2 CALIBRATION:
If the flow rate warning appeared during the 1/2 calibration procedure, you should adjust the flow rate to the proper
setting, then press the Enter key when finished.
Now the instrument display should indicate Run Ready.
From this display, press the Special Function key to show the “Perform 1/2 Cal” screen.
“ERR” DISPLAY:
Should an operator accidently press the START ( ) key without flushing Cal gas into the instrument, the
display will indicate ERR. Wait until the Status light starts blinking green and try again.
If you have experienced this error and are certain it is not user error, please contact QuinTron.
“OVR” DISPLAY:
If a patient sample contains H2 or CH4 above levels that the instrument can normally read, the display will
indicate OVR.
It is recommended that this sample be diluted by 50% with room air and re-analyzed.
Example: If the syringe has 20 mL of a sample, draw in 20 mL of room air so that the total sample in syringe to equal 40 mL.
If you are still experiencing the “OVR” display after attempting the above instructions, please contact QuinTron.
“MAX” DISPLAY:
This code is displayed during the patient sample correction feature activation. If this code appears it means that
the pre-correction (Raw) values multiplied by the CO2 correction factor is above the limit the BreathTracker’s
microcontroller can accurately determine. Users should revert back to and record the raw values for the
physician to use during interpretation. Users should also right MAX on the sample sheet so the physician
understands that the sample analyzed was above the limit the correction feature can accurately calculate.
QTL0054 REV B (BT-SOFT 2 COPYRIGHT
©2009 QUINTRON INSTRUMENT COMPANY ,
HOW TO USE YOUR QUINGAS™ TRANSFER VALVE
The QuinGas transfer valve is used to remove calibration gas from the QuinGas cylinder for analysis in your
BreathTracker breath-testing instrument.
The transfer valve should also be removed from the top of the QuinGas cylinder when not in use.
FAILURE TO REMOVE THE TRANSFER VALVE WHEN NOT IN USE MAY LEAD TO LOSS OF CALIBRATION GAS.
Figure 2a Figure 3a
Varying local regulations exist across the country regarding recycling and what may or may not be
acceptable for land-fill sites. The best thing is to recycle the cylinders so the metals can be reclaimed, but
this may not be practical for your office. You will need to contact your local waste management company for
more specific instructions if you require them. Always disengage the valve for safe disposal.
Protective Earth
Any terminal which is intended for connection
to an external protective conductor for protection Attention, consult accompanying documents
against electric shock in case of a fault.
Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC)
and needs to be installed and put in to service according to the EMC information provided:
a. Portable and mobile radio frequency (RF) communications equipment can affect medical
electrical equipment. This product is intended for use in the electromagnetic environments
specified.
b. The end user of this product should assure it is used in such an environment.
i. Portable and mobile RF Communications equipment (cell phones) should not be used at
close distances.
ii. Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or hospital environment.
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS
ONLY IN ACCORDANCE WITH UL 60601-
1, IEC/EN 60601-1. CAN/CSA C22.2 No.601.1
3MFU
Shipping and Storage conditions:
Temperature range within -40C to +70C;
Relative humidity range within 10% - 100%;
Atmospheric pressure range within 500 to 1060 hPa
Disclaimer
The Products are specifically designed for the purpose as set forth in the Owner’s Manual for the Products and for no other
purpose. The Products should only be used and operated by those skilled individuals that have read and fully understand the
Owner’s Manual and are of legal age and sound mind, and only in a manner consistent with its specifically designed purpose
as set forth in the Owner’s Manual.
THE REMEDY OF REPAIR AND REPLACEMENT PROVIDED FOR HEREIN SHALL BE THE CUSTOMER OR
END-USER’S SOLE AND EXCLUSIVE REMEDY IN THE EVENT OF A BREACH OF THIS LIMITED WARRANTY.
ADDITIONALLY, UNDER NO CIRCUMSTANCES SHALL QUINTRON BE LIABLE FOR ANY INDIRECT,
SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, INCLUDING BUT NOT LIMITED TO, LOST PROFITS AND PUNITIVE
QT07220-G -
QuinGas-2 (~150ppm H2, 75ppm CH4)
X Reorder when psi reaches below 50psi.*
QT07225-G -
QuinGas-2 (~150ppm H2, 6% CO2)
X Reorder when psi reaches below 50psi.*
QT07230-G -
QuinGas-3 (~150ppm H2, 75ppm CH4, 6% CO2)
X Reorder when psi reaches below 50psi.*
All products listed above do not need to be replaced when analyzing multiple patient samples.
Maintenance materials are used to ensure longevity of the BreathTracker system.
Failure to replace components when needed can result in damage to the product and/or adversely affect your patient samples.
If there are any questions regarding these products or their use please consult with QuinTron’s Customer Service
Department.
U.S. Office
QuinTron Instrument
Company 3712 West Pierce
Street Milwaukee, WI 53215
USA
Customer Service:
(800) 542-4448 (Toll-Free US & Canada
Only) (414) 645-4222
Fax: (414) 645-3484
www.QuinTron-USA.com E-mail: Sales@QuinTron-USA.com
Technical Support:
(800) 542-4448 (Toll-Free US & Canada
Only) (414) 645-3778
E-mail: Support@QuinTron-USA.com
QuinTron-EU
via Vico Vigano, 55
Roma, Italia 00133
Phone: +39-06-4067873
Fax: +39-06-4065151
E-mail: Support@QuinTron-EU.com
www.QuinTron-EU.com
Drierite is used to remove water vapor in patient breath samples prior to analysis or storage of samples.
All QuinTron instrumentation utilize Drierite in some fashion. The use of Drierite is critical to keeping
moisture out of your instrument. By using no Drierite (or allowing Drierite to expire), the following can occur:
damage to the internal components, delays/damage in the instruments sensor recovery and adverse impact on
patient sample results. If an instrument is damaged due to neglect in properly using this product, any service
work may not be covered under warranty.
Drierite is typically poured into the Patient Sample Drying Tube (QT01135-K) via the 35 mL plastic funnel
(QT02658) provided in the BreathTracker Accessory Kit or can be purchased separately. When the Drierite
is 3/4 expired (pink) the technician/end-user will empty the drying tube and replace the contents with fresh
Drierite. Drierite absorbs small amounts of carbon dioxide (CO2) unless conditioned properly before use.
Users which have instrumentation that utilize the CO2 sensors are required to condition the drying tube that has
been filled with fresh Drierite with 60mL of calibration gas each time the expired Drierite is replaced to ensure
that the Drierite does not absorb the CO2 in the patient samples.
FRESH EXPIRED
10/20 Mesh Indicating
Drierite Desiccant -
QT01156-C
KIDSAMPLER™ The KidSampler system is similar to the GaSampler, except it can utilize a
mouthpiece or a face mask for pediatric collection. This system is designed
only for in-house testing on any BreathTracker.
This system uses multiple parts that do not come together in a kit; all parts
are sold as disposable except the multi-patient foil laminate collection bags
which can be used multiple times over multiple patient collections.
Other pediatric collection systems are available for babies and neonatal
patients.
ALVEOSAMPLER SYSTEM:
The AlveoSampler System is an economical and disposable device used to collect alveolar air samples for
subsequent analysis. It permits one-patient use of a modified Haldane-Priestley tube.
The alveolar sample is drawn into a syringe from the end-expiratory air blown through the device. During
expiration through the mouthpiece, a vented polyethylene bag with a medium-resistance leak is filled to
indicate that adequate dead space volume has been exhaled. As exhalation continues, air is then steadily drawn
into the syringe by the operator. The bag serves as a check-valve to prevent contamination of the syringe
sample with atmospheric air as long as the patient keeps the AlveoSampler mouthpiece in his/her mouth.
Use of the AlveoSampler removes the danger of inter-patient cross-infection, and will save time and money by
eliminating the costs of cleaning and sterilizing reusable products.
STORAGE LIMITS:
After the alveolar air sample is collected and the syringe has been removed from the collection mouthpiece,
the sample may be immediately analyzed. The syringe will only be able to hold a sample for a maximum of
2 hours. If you cannot analyze the sample within that time, please transfer the samples into a Sample Holding
Bag (QT00842-P).
After the alveolar air sample is collected and the Collection Bag cap is securely in place, the sample can be
immediately analyzed by withdrawing the sample from the stopcock attached to the small port.
There are two types of GaSampler collection bags: Single-Patient and Multi-Patient.
STORAGE LIMITS:
Single-Patient Bags can only hold a sample for a maximum of 3 hours.
Multi-Patient Bags can only hold a sample for a maximum of 6 hours.
If you cannot analyze within this time frame you must transfer the samples into Sample Holding Bags (QT00842-P).
(Your Port/Port Caps may differ in color and/or style.)
• Once finished with the test the Collection Bag should be discarded.
Never reuse any products that has been used on a patient with a communicable or infectious disease!
We expect the multi-patient collection bags to last for a respectable period of time, say for 10-20 patients, but
NOT indefinitely. The number of uses for the bags is dependent on how the bags are handled.
Continuously filling the bags to their maximum volume causes the development of creases and wrinkles and
also stretches the seams along the edges, resulting in both damaging the product and adversely impacting the
integrity of the sample. Careful handling of the bags increases the length of use of the product and ensures the
validity of the breath study.
The recommended usage limits indicated below are based on QuinTron’s in-house testing; usage limits may not
be the same for all facilities depending on how the bags are handled. Inter-patient cleaning is still
recommended for all products that are sold/used as Multi-Patient. Never use single-patient products inter-
patient.
* Multi-Patient Collection Bags require special attention, see “Cleaning/Sterilization Practices” on laminated bags for more information.
4. Insert a stopcock into the small port on the Sample Holding Bag.
5. Insert your Sample Drying Tube into the stopcock on the
Sample Holding Bag.
6. Attach the stopcock on the Collection Bag to the other end of the Sample Holding Bag Collection Bag
Patient Sample Drying Tube.
7. Open both stopcocks and gently apply pressure to the Collection Bag until an adequate sample volume is
transferred into the Sample Holding Bag.
8. Close the stopcock on both the Sample Holding Bag and Collection Bag to ensure no sample expires.
9. Remove the Patient Sample Drying Tube and syringe from the Sample Holding Bag.
10. Repeat the same procedure with a new Sample Holding Bag for each sample you wish to store.
11. You may reuse the Sample Drying Tube inter-patient until it has fully expired.
TEE-MOUTHPIECE ASSEMBLY:
This assembly completes the GaSampler when used with a Collection Bag and
Discard Bag. It includes a plastic Tee, a removable one-way valve and mouthpiece.
These are sold as disposables. Each component of the Tee-Mouthpiece Assembly
can also be ordered individually.
Catalog #: QT00854-P - Tee-Mouthpiece Assembly (sold as each)
DISPOSABLE MOUTHPIECES:
Mouthpieces are inexpensive enough to be disposable, but they can be cleaned,
disinfected and reused if desired.
Catalog #: QT00991-P - Plastic Mouthpiece (sold as each)
BENZ-ALL DISINFECTANT:
Benz-All® is a concentrated disinfectant for the reusable components of the
GaSampler. One 40cc bottle of Benz-All is sufficient to produce one full gallon of
ready to use disinfecting solution. Rubber and metal instruments need only be soaked
for 15 minutes, but may be left in the solution longer if necessary. And while not a cold
sterilizer, Benz-All may be used for the storage of heat sterilized instruments.
Catalog #: QT00993-P - Benz-All disinfectant (1 box = 15 bottles)
Benz-All® is a registered trademark of Xttruim Laboratores, Inc.
*If you plan to reuse the bags for multiple patients, please review the section “General Cleaning/Sterilization Practices”.