Technical Specification for ventilator

S.No Description

1 Screen : Should be 12.1 to 12.5” with Color TFT Touch Screen

2 Should display 3 waveforms and 2 loops with vital settings , patient parameters and alarm
settings in single view

3 Should have 2 level user interface.

4 Direct Access knob for vital setting

5 Tidal volume : 2 to 300 ml

Minute Volume : 0.3 – 55 l/min

Inspiratory pressure : 0 – 115 cm H2O

Apnea time: 1-45 Sec

6 Should support both invasive and non-invasive type of ventilation

7 Breath Rate should be 5 to 145 bpm

8 Inspiratory peak flow should be 0 to 195 lpm with 3 flow patterns

9 Following are equivalent modes should be there in the ventilator

Control Mode: Pressure control, volume control, pressure regularized volume control

Combined Mode: SIMV (PC) + PS, SIMV (VC)+PS, SIMV(PRVC)+PS, Bi-Vent

Spontaneous Mode: Volume Support, Pressure Support, CPAP

10 I:E ratio should be 1:10 – 4:1; PEEP should be 0 to 50 cmH2O

11 Should Support special gas delivery - Heliox

12 Should Continue ventilation even if one gas supply source fails

13 Should support special type of ventilation HBO (Hyperbaric Oxygen Therapy)

14 Should have lung protective ventilation packages

15 Gas measuring method for both oxygen and flow sensor should be of ultrasonic based

16 No gas/bleed wastage should be there

17 Bias flow triggering should be in the range of 0 to 4 lpm

18 Air/oxygen blender should be of electronic module

19 Ventilator should support upgradation of modes such as Open Lung Tool (Breath-to-Breath
Method),Stress Index,Etco2 monitoring, NAVA (Neurally Adjusted Ventilator Assist)

20 Should have following lung mechanics

Static & Dynamic compliance, Total PEEP, Plateau pressure (Pplat), Auto PEEP, Inspiratory &
Expiratory Resistance, Elastance, Time Constant, P0.1 measurement, Work of Breathing
patient, Work of Breathing ventilator, Shallow Breathing Index (SBI)

21 Should have special software such as Suction support, Preset program for Higehr O2
concentration for 2 Mts. pre-suction, 1 Mt. window for suction & 1 Mt. post-suction,
Compliance compensation, Tube Resistance Compensation

22 Should have inbuilt nebulizer and provide humidifier and mini compressor along with it.

23 Ventilator should be FDA and CE certified

24 Ventilator should have battery back more than 160 min