Professional Documents
Culture Documents
Nanochemistry Characterization AFM SEM
Nanochemistry Characterization AFM SEM
angstroms to 100 microns. Binnig, Quate and Gerber invented the first AFM in 1986 and
Gerd Binnig with Heinrich Rohrer developed the precursor to the atomic force microscopy
(AFM), called, the scanning tunneling microscope (STM) that earned them the Nobel Prize
AFM is operated by measuring attractive or repulsive forces between a tip and the
sample. Depending on the situation, forces that are measured in AFM include mechanical
contact force, Van der Waals forces, capillary forces, chemical bonding, electrostatic
forces, magnetic forces [in Magnetic force microscope (MFM)], etc. A schematic view of
an AFM system is shown in Fig. 1. The technique involves imaging a sample using a
probe, or tip. The tip is located at the free end of a cantilever that is 100 or 200 μm long.
The tip is typically made of silicon or silicon nitride with a tip radius of curvature on the
order of nanometers (~ 10 – 30 nm). The tip (or the sample) is scanned using piezoelectric
tubes. Piezoelectric scanners are designed to be moved precisely in any of the three
perpendicular axes (x,y,z). When the tip is brought into proximity of a sample surface,
forces between the tip and the sample lead to a deflection of the cantilever according to
Hooke's law. This deflection is measured using the position of a laser light reflected from
the cantilever head as the angular deflection of the cantilever causes a twofold larger
angular deflection of the reflected light. A two or a four-segment position sensitive photo-
detector detects the laser spot. By measuring the difference signal of the photo-detector
segments, changes in the bending of the cantilever can be measured. The measurements of
stage, which controls the movement of the sample under the tip.
Photodetector
B
Mirror D
A
Laser C
Control Unit
Cantilever
with tip
Scanner
tube
The AFM can be operated in two basic modes (a) with feedback control and (b)
without feedback control. If the electronic feedback is switched on, then the positioning
piezo, which is moving the sample (or tip) up and down can respond to any change in
force, which is detected, and alter the tip-sample separation to restore the force to a pre-
determined value. This mode of operation is known as constant force or height mode. If the
feedback electronics are switched off, then the microscope is said to be operating in
constant height or deflection mode. This is particularly useful for imaging very flat samples
at high resolution. The way in which image contrast is obtained can be achieved in many
ways.
The three main classes of interaction of tip to sample surface are contact mode,
tapping mode and non-contact mode. Contact mode is the most common method of
operation of the AFM. As the name suggests, the tip and sample remain in close contact as
the scanning proceeds. By "contact” it means that the tip sample interaction is in the
repulsive regime of the inter-molecular force curve (see Fig. 2). One of the drawbacks of
remaining in contact with the sample is that a large lateral force exists on the sample and
Force Intermittent
Contact
Contact
Non contact
Tapping mode is the next most common mode used in AFM. When operated in
air or other gases, the cantilever is oscillated at its resonant frequency (often hundreds of
kilohertz) and positioned above the surface so that it only taps the surface for a very small
fraction of its oscillation period. This is still in contact with the sample in the sense defined
earlier, but the very short time over which this contact occurs means that lateral forces are
dramatically reduced as the tip scans over the surface. When imaging poorly immobilized
or soft samples, tapping mode may be a far better choice than contact mode for imaging.
by AFM. The cantilever must be oscillated above the surface of the sample at such a
distance that we are no longer in the repulsive regime of the inter-molecular force curve.
This is a very difficult mode to operate in ambient conditions with the AFM. The thin layer
of water contamination that exists on the surface on the sample will invariably form a small
capillary bridge between the tip and the sample and cause the tip to "jump-to-contact".
number of other modes, including magnetic force, electrical force and force-distance mode
etc. Magnetic force mode imaging utilizes a magnetic tip to enable the visualization of
magnetic domains on the sample. In electrical force mode imaging a charged tip is used to
locate and record variations in surface charge. In force-distance mode, the sample is
oscillated beneath the tip, and a series of force-distance curves are generated. This permits
the separation of sample topography, stiffness and adhesion values producing three
The scanning electron microscope (SEM) uses electrons rather than light to form an
image. There are many advantages of using the SEM instead of a light microscope. The
SEM has a large depth of field, which allows a large area of the sample to be in focus at
one time. The working principle of SEM is shown in the Fig. 3. When the
Backscattered 一
Electrons (BSE)
一 SE
一 一 一
一
一 一
Excitation volume
for SE emission
Z
一 一
一
一 一
一 一
SE
Specimen
primary electron beam interacts with the sample, the electrons lose energy by repeated
random scattering and absorption within a teardrop-shaped volume of the specimen known
as the interaction volume, which extends from few nm to around few µm into the surface.
The size of the interaction volume depends on the electron's landing energy, the atomic
number of the specimen and the specimen's density. The energy exchange between the
electron beam and the sample results in the reflection of high-energy electrons by elastic
A beam of electrons is generated in the electron gun, located at the top of the
column. This beam is attracted through the anode, condensed by a condenser lens, and
focused as a very fine point on the sample by the objective lens. The electron beam, which
typically has an energy ranging from a few hundred eV to 40 keV, is focused by one or two
condenser lenses to a spot about 0.4 nm to 5 nm in diameter. The beam passes through
pairs of scanning coils or pairs of deflector plates in the electron column, typically in the
final lens, which deflect the beam in the x and y axes so that it scans in a raster fashion over
Generally, the secondary electrons are collected to form the image in the SEM
mode. The beam current absorbed by the specimen can also be detected and used to create
images of the distribution of specimen current. Electronic amplifiers of various types are
used to amplify the signals which are displayed as variations in brightness on a cathode ray
tube (CRT).