Professional Documents
Culture Documents
Lec 5-Accuracy (Compatibility Mode)
Lec 5-Accuracy (Compatibility Mode)
ii. Noise X
n
n
X
iii. Calibration N
1
18/10/2016
ii. Range and Uncertainty Errors iii. Hysteresis and Repeatability Errors
-8
1.0x10
-5
5.0x10
-9 3.0x10
I (A)
-5
0.0 2.0x10
G(S)
-9
-5.0x10 -5
1.0x10
-8
-1.0x10
-0.6 -0.4 -0.2 0.0 0.2 0.4 0.6
VDS (mV) 0.0
-30 -15 0 15 30
Vg(V)
Each instrument has finite dynamic
range. Beware of saturation and too Noise limits the accuracy and
small signals! resolution. Beware of too small
Linearity is an idealization. Know the signals! •Will cause error if used as a sensor
range where it works!
2
18/10/2016
Use the right sensor: The sensor should not affect the
• Random errors are perturbations of the measurement either side of
process and the process should not destroy the sensor.
the true value caused by random and unpredictable effects, such
Check the accuracy of each element and determine the that positive errors and negative errors occur in approximately
accumulated accepted error equal numbers for a series of measurements made of the same
Calibrate each instrument quantity.
• Such perturbations are mainly small, but large perturbations occur
Connect system with proper wires
from time to time
Check the system for electrical noise • Cuases of Random errors :
Estimate the total error in the system from all known • when measurements are taken by human observation of an
sources analogue meter, especially where this involves interpolation
between scale points.
Perform a system calibration by measuring the variable • Electrical noise
in a known process. This gives you a single calibration • Can be overcome by taking the same measurement a number of
constant for the entire system. Example: scales times and extracting a value by averaging or other statistical
techniques
3
18/10/2016
From statistical theory about random error we can assume the Examples of random errors
following:
The correlation between the true score and the error term is zero.
There is no serial correlation between the true score and the error term. “Dynamic” error. Measurement "at first glance" for unsteady
The correlation between errors on distinct measurements is zero. state. Often caused by inappropriate time constant.
From these assumptions it follows: Recording and calculation error. Incorrect interpolation between
E(M) = E(T) +E(e) marks on analog meter. Occurs if operator does not know how to write, not paying
= E(T) + 0 attention, not familiar with math, etc
= E(T) Incorrect interpretation error. Trying to measure microvolts on
=T "kiloVolts" scale (or the opposite, which may also result in the damage to the
The smaller the variations in the error term, the more reliable operator/instrument)
the measurements. Misuse of instrument. Measurement of high resistance source using low
M=T+e input resistance meter. Trying to measure Amps on "Hertz" scale. Using meter as a
hammer
var(M) = var(T) + var(e)
where var is variance Misuse of sensor. Using thermometer without appropriate thermal
contact.
Reliability = pm = var(T) / var(M) Malfunction of sensor or instrument. (e.g. loose contact)
= [var(M) –var(e)] / var(M)
= 1 – [var(e) / var (M)]
Statistical analysis of measurements Which of the two measurement sets A and B, and the corresponding
mean and median values, should we have most confidence in?
subject to random errors
In set A, the spread between the smallest (396) and largest (430)
1. Mean and median values value is 34, whilst in set B, the spread is only 6.
398 420 394 416 404 408 400 420 396 413 430 {Measurement set A} Now, mean = 406.5 and median = 406.
Mean = 409.0 and median = 40 Thus the median value tends towards the mean value as the number of
409 406 402 407 405 404 407 404 407 407 408 {Measurement set B} measurements increases.
Mean = 406.0 and median= 407
Example: Calculate δ and V for measurement sets A, B and C above a histogram for set C
4
18/10/2016
5. Estimation of random error in a single measurement Example: Suppose that a standard mass is measured 30 times with
the same instrument to create a reference data set, and the
• If only one measurement can be made, some means of estimating the calculated values of δand α areδ = 0.43 and α = 0.08. If the
likely magnitude of error in it is required. instrument is then used to measure an unknown mass and the
• The normal approach to this is to calculate the error within 95% reading is 105.6 kg, how should the mass value be expressed?
confidence limits, i.e. to calculate the value of the deviation D such that
Using the relation in the previous slide
95% of the area under the probability curve lies within limits of ± D.
• These limits correspond to a deviation of ±1.96δ.
• this only expresses the maximum likely deviation of the measurement
from the calculated mean of the reference measurement set, which is
not the true
• Thus the calculated value for the standard error of the mean has to be The mass value should therefore be expressed as:
added to the likely maximum deviation value
6. Calculations of Error for a Test with Multiple Variables 7. Aggregation of measurement system errors
N
2
max
i 1
i
5
18/10/2016
c. Error in a product
If the power in a circuit is calculated from measurements of voltage and
• If the outputsyandzof two measurement system components are current in which the calculated maximum errors are respectively ±1% and
multiplied together, the product can be written as PDyz. If the ±2%, then the maximum likely error in the calculated power value is:
possible error in y is ±ay and in z is ±bz, then the maximum and
minimum values possible in P can be written as:
d. Error in a quotient
Example:
If the output measurement yof one system component with possible If the density of a substance is calculated from measurements of
error ±ay is divided by the output measurement z of another system its mass and volume where the respective errors are ±2% and
component with possible error ±bz, then the maximum and minimum ±3%, then the maximum likely error in the density value is
possible values for the quotient can be written as:
For a<<1 and b << 1, terms in ab and b2 are negligible compared with
the other terms
Hence
6
18/10/2016
Average deviation
X n X
Signal noise arises both within the measurement circuit itself and DN n
N
also during the transmission of measurement signals
Noise voltages can exist either in serial mode or common mode forms. Standard 1 N
Serial mode noise voltages act in series with the output voltage from a deviation SN (X n X )2 X
N n 1
measurement sensor or transducer, which can cause very significant errors in
the output measurement signal
Signal-to-noise
The extent to which series mode noise corrupts measurement signals is
Ratio
measured by
X X
SNR N
X 1 N
(X n X )2
N n 1
where Vs is the mean voltage level of the signal and Vn is the mean voltage X
level of the noise. SNR N
SNR improves as
In the case of a.c. noise voltages, the root-mean squared value is used as the X
mean.
Inductive coupling
• Noise can be generated from sources both external • This is the primary mechanism by which external devices such as
and internal to the measurement system. mains cables and equipment, fluorescent lighting and circuits
operating at audio or radio frequencies generate noise
• External noise • If signal-carrying cables are close to such external cables or
• Inductive coupling equipment, a significant mutual inductance M can exist between them,
• Capacitive (electrostatic) coupling • This can generate a series mode noise voltage of several millivolts
• Noise due to multiple earths given by: Vn = MI’,
• Noise in the form of voltage transients • where I’ is the rate of change of current in the mains circuit.
• Internal noise includes
• Thermoelectric potentials
• Shot noise
• Electrochemical potentials
• The magnitude of the capacitance between each signal wire and • As far as possible, measurement signal circuits should be isolated
the mains conductor is represented by the quantities C1 and C2 from earth.
• A capacitance can also exist between the signal wires and earth, • However, leakage paths often exist between measurement circuit
represented by C3 and C4 in the figure signal wires and earth at both the source (sensor) end of the circuit
• The series mode noise voltage Vn is zero if the coupling and also the load (measuring instrument) end.
capacitances are perfectly balanced, i.e. if C1 = C2 and C3 = C4. • This does not cause a problem as long as the earth potential at both
• However, exact balance is unlikely in practice, since the signal ends is the same.
wires are not perfectly straight • However, it is common to find that other machinery and equipment
carrying large currents is connected to the same earth plane.
• This can cause the potential to vary between different points on the
earth plane.
• This situation, which is known as multiple earths, can cause a series
mode noise voltage in the measurement circuit.
7
18/10/2016
Earthing Shielding
• Shielding consists of enclosing the signal wires in an earthed, metal
• Noise due to multiple earths can be avoided by good earthing shield that is itself isolated electrically from the signal wires.
practices. • The shield should be earthed at only one point, preferably the signal
• This means keeping earths for signal wires and earths for high- source end.
current equipment entirely separate. • A shield consisting of braided metal eliminates 85% of noise due to
• Recommended practice is to install four completely isolated earth capacitive coupling whilst a lapped metal foil shield eliminates noise
circuits as follows: almost entirely.
• Power earth: provides a path for fault currents due to power • The wires inside such a shield are normally formed as a twisted pair
faults. so that protection is also provided against induced noise due to
• Logic earth: provides a common line for all logic circuit nearby elecromagnetic fields.
potentials.
• Analogue earth (ground): provides a common reference for all
analogue signals.
• Safety earth: connected to all metal parts of equipment to
protect personnel should power lines come into contact with
metal enclosures.
8
18/10/2016
9
18/10/2016
10