Professional Documents
Culture Documents
D Dimer Exclusion I I 30455
D Dimer Exclusion I I 30455
bioMérieux SA English - 1
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
bioMérieux SA English - 2
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
bioMérieux SA English - 3
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
The influence of 47 analytes was also detected in vitro: no interference was observed.
Amikacin sulfate 10.4 mg/dL Diazepam 0.5 mg/dL Nifedipine 0.04 mg/dL
Ampicillin 5 mg/dL Digoxin 0.0006 mg/dL Penicillin G 2500 U/dL
sodium salt
Ascorbic Acid 6 mg/dL D-L methyl dopa 1.8 mg/dL Pentobarbital 9 mg/dL
(vitamin C) hydrochloride
Atenolol 1 mg/dL Dopamine 0.1 mg/dL Phenobarbital 10 mg/dL
hydrochloride
Caffeine 6 mg/dL Erythromycin 6 mg/dL Phenytoine 5 mg/dL
Captopril 0.5 mg/dL Ethanol 400 mg/dL Primidone 4 mg/dL
Carbamazepine 3 mg/dL Ethosuximide 25 mg/dL Propranolol 0.2 mg/dL
hydrochloride
Chloramphenicol 5 mg/dL Furosemide 6 mg/dL Theophylline 4 mg/dL
Chlordiazepoxide 1.1 mg/dL Gentamicin sulfate 1 mg/dL Urea 500 mg/dL
hydrochloride
Chlorpromazine 0.2 mg/dL Heparin lithium salt 300 U/dL Uric acid 24 mg/dL
hydrochloride
Cimetidine 2 mg/dL Heparin sodium salt 300 U/dL Valproic acid 60 mg/dL
sodium salt
Cinnarizine 3 mg/dL Ibuprofen 50 mg/dL Verapamil 0.2 mg/dL
hydrochloride
Lidocaine 1.2 mg/dL Warfarin 1 mg/dL
Specimen stability Before each new lot of reagents is used, specifications (or
Plasma samples separated from the pellet can be stored factory master calibration data) must be entered into the
at 2-8 °C in stoppered tubes for 3 days. instrument using the MLE data. If this operation is not
The assay can be also performed on plasma frozen for performed before initiating the tests, the instrument will
6 months: plasma must be frozen at -25 ± 6°C not be able to print results. The master lot data need only
immediately after decantation. It must be thawed rapidly at be entered once for each lot.
37°C when assaying and the assay should be performed It is possible to enter MLE data manually or automatically
rapidly after thawing. Two freeze/thaw cycles are depending on the instrument (refer to the Operator’s
acceptable. Manual).
INSTRUCTIONS FOR USE Calibration
For complete instructions, see the Operator’s Manual. Calibration, using the calibrator provided in the kit, must
VIDAS PTC protocol data entry be performed each time a new lot of reagents is opened,
When using the assay for the first time, and before after the master lot data have been entered. Calibration
reading the MLE data, scan the bar code(s) (at the end should then be performed every 28 days. This operation
of the package insert) using the instrument’s external bar provides instrument-specific calibration curve and
code reader. This reading will allow VIDAS PTC protocol compensates for possible minor variations in assay signal
data to be transferred to the instrument software for its throughout the shelf-life of the kit. The calibrator, identified
update. These data should only be read the first time the by S1, must be tested in duplicate (see Operator’s
assay is used. Manual).
The calibrator value must be within the set RFV "Relative
Master lot data entry Fluorescence Value" range. If this is not the case,
Note: When using the assay for the first time, enter recalibrate.
the VIDAS® PTC protocol (bar codes at the end of the
package insert) before reading the MLE data. If the
®
MLE data have been read before the VIDAS PTC
protocol, read the MLE data again.
bioMérieux SA English - 4
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
6. For this test, the calibrator, control and sample QUALITY CONTROL
test portion is 200 µL. Two controls are included in each VIDAS D-Dimer
7. Insert the “DEX2” SPRs and “DEX2” strips into their Exclusion II kit. These controls must be performed
appropriate position on the instrument. Check to make immediately after opening a new kit to ensure that reagent
sure the color labels with the assay code on the SPRs performance has not been altered. Each calibration must
and the STRs match. also be checked using these controls. The instrument will
only be able to check the control values if they are
8. Initiate the assay as directed in the Operator’s Manual.
identified by C1 and C2.
All the assay steps are performed automatically by the
Results cannot be validated if the control values deviate
instrument.
from the expected values.
9. Reclose the vials and return them to the required
temperature after pipetting. Note
It is the responsibility of the user to perform Quality
10. The assay results are available within
Control in accordance with any applicable local
20 minutes. After the assay is completed, remove the
regulations.
SPRs and STRs from the instrument.
11. Dispose of the used SPRs and STRs into an
appropriate receptacle.
PERFORMANCE
Studies performed using VIDAS® D-Dimer Exclusion II gave the following results:
Measurement range
The measurement range of the VIDAS D-Dimer Exclusion II starts at 45 ng/mL (FEU) with an upper limit of quantification
of 10 000 ng/mL (FEU).
Linearity
The linearity has been determined on plasmas according to the recommendations of CLSI® document EP6-A volume 23,
number 16. The results obtained have shown that VIDAS D-Dimer Exclusion II is linear from 45 to 10,000 ng/mL (FEU).
However, for a few isolated samples, VIDAS D-Dimer Exclusion II may not be linear due to the matrix.
Detection limit
Defined as the smallest concentration of D-dimer significantly different from the zero concentration with a risk α of 5%:
< 45 ng/mL (FEU). The results of the detection limit were determined according to the standard CLSI® EP17-A.
bioMérieux SA English - 5
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
Hook effect
No hook effect was found up to D-dimer concentrations of 400,000 ng/mL (FEU).
Specificity
Interference was studied according to the recommendations of the CLSI document EP7-A2.
Cross reactant Concentration (spiked samples) Cross reactivity
Fibrinogen <10 g/L No
Fibrinogen degradation products X < 10 µg/mL No
Fibrinogen degradation products Y <10 µg/mL No
Fibrinogen degradation products D 10 - 100 µg/mL Yes*
* : such high levels of fibrinogen degradation products D do not occur in the target population of suspected VTE patients.
They may be observed when patients are subjected to therapeutic lysis with thrombolytic agents (13).
NB : the specificity of the two antibodies used in this assay has not been tested against fibrinogen degradation
products E, therefore cross-reactivity cannot be ruled out.
Precision
Three samples were tested in duplicate in 40 different runs (2 runs per day) with 2 reagent lots at 3 sites (n=240).
The repeatability (intra-run precision), and between lot reproducibility were calculated using this protocol, based on the
®
recommendations of CLSI document EP5-A2 :
Study 2:
378 citrated plasmas that came from laboratory routine activity of 1 European site were tested using two lots of
VIDAS® D-Dimer Exclusion II (ref. 30 455) and two lots of VIDAS D-Dimer Exclusion (ref. 30 442) according to the
recommendations of CLSI document EP9-A2. The concentration range of plasma samples investigated is from 46.70 to
9,913.53 ng/mL (FEU). The following results were obtained :
• Passing & Bablok regression and correlation coefficient (r):
Site 4 n Slope Intercept r
Europe 374* 1.09 -29.35 0.991
* 4 statistical outliers were removed from the analysis
In conclusion, these two studies show a mean slope of 1.14 between VIDAS D-Dimer Exclusion II (ref. 30 455) and
VIDAS D-Dimer Exclusion (ref. 30 442), with slopes between 1.09 and 1.19 depending on the lots.
bioMérieux SA English - 6
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
Clinical performance
Clinical performance using the VIDAS D-Dimer Exclusion II assay (ref. 30 455) in conjunction with a patient’s
pretest probability to exclude VTE (Venous ThromboEmbolism)
A study was performed using 315 frozen characterized samples from patients from previous prospective clinical
studies (14, 15).
The overall prevalence of VTE in the total population studied was 23.5% (74/315). The sensitivity, specificity, negative
predictive value (NPV) and positive predictive value (PPV) using a clinical cut off of 500 ng/mL (FEU) are summarized
below with the corresponding 95% confidence intervals (CI).
Agreement study between the two methods (VIDAS D-Dimer Exclusion II and VIDAS D-Dimer Exclusion) and the
clinical status are presented below (n = 315 frozen samples):
Clinical Status
VIDAS D-Dimer Exclusion II VIDAS D-Dimer Exclusion Total NEGATIVE POSITIVE
ref. 30 455 ref. 30 442 samples (NO VTE) (VTE)
< 500 ng/mL 85 85 0
< 500 ng/mL
≥ 500 ng/mL 1 1 0
< 500 ng/mL 6 6 0
≥ 500 ng/mL
≥ 500 ng/mL 223 149 74
Total samples 315 241 74
There is no significant difference at a risk level of 5% between the sensitivity and specificity of the two assays.
Clinical performance using the VIDAS D-Dimer Exclusion (ref. 30 442) assay in conjunction with a patient's
pretest probability
To exclude DVT
Frozen samples collected from patients enrolled in a multi-center, prospective cohort study (16) were used to validate
the diagnostic utility of VIDAS D-Dimer Exclusion to exclude a diagnosis of deep vein thrombosis (DVT).
Consecutive eligible outpatients (n = 556) with a first suspected DVT episode were evaluated at three hospitals during
the course of the study. Using the Wells model to estimate the probability of DVT, patients were classified as having a
high, intermediate or low pretest probability (PTP) of DVT (17).
The VIDAS D-Dimer Exclusion assay was performed without knowledge of the PTP assessment results and the clinical
outcome of the patients from which the samples were derived. A clinical cut off value of 500 ng/mL (FEU) was used.
A D-dimer result of ≥ 500 ng/mL (FEU) was considered positive and a result of < 500 ng/mL (FEU) was considered
negative.
bioMérieux SA English - 7
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
This study was designed as a management clinical trial and patients were grouped according to PTP. Those patients
having a negative D-dimer test result and a low or intermediate PTP of DVT underwent no further diagnostic testing and
were followed up for 3 months for development of DVT. Patients with a positive D-dimer test result and/or high PTP
underwent serial compression ultrasound (CUS).
The overall prevalence of DVT in the total population studied was 10.1% (56/556). One sample from the original study
was not tested due to volume limitations. The sensitivity, specificity and negative predictive value (NPV) of the VIDAS D-
Dimer Exclusion assay using a clinical cut off of 500 ng/mL (FEU) are summarized below with the corresponding 95%
confidence intervals (CI).
To exclude PE
The diagnostic utility of VIDAS D-Dimer Exclusion to exclude a diagnosis of PE was validated in a multi-site
prospective patient management study using fresh specimens from 1290 consecutive patients presenting to the
emergency department with suspected PE (18). Of these 1,290 patients enrolled, 325 were eventually excluded for a
total of 965 patients included in the final analysis.
All patients enrolled were classified using the WICKIE model (18) as having a high, intermediate or low pretest probability
(PTP) of PE. These patient specimens were tested using the VIDAS D-Dimer Exclusion test. A cutoff value of 500 ng/mL
(FEU) was utilized with D-dimer results of ≥ 500 ng/mL (FEU) considered positive and results of < 500 ng/mL (FEU)
considered negative.
All patients with negative D-dimer test results received no treatment and underwent no further diagnostic testing.
Patients with positive D-dimer test results were further evaluated using ultrasound, Helical CT scans, and /or
angiography. These patients were then treated based upon the results of the additional diagnostic testing. All patients
were followed up for a period of three months to evaluate any possible Venous thromboembolic events (DVT or PE) and
episodes of bleeding.
The overall prevalence of PE in the total population studied was 23% (222/965). The sensitivity, specificity, and negative
®
predictive value of the VIDAS D-Dimer Exclusion™ assay using a clinical cut off of 500 ng/mL (FEU) are summarized
below with the corresponding 95% confidence intervals (CI).
bioMérieux sa English - 8
VIDAS® D-Dimer Exclusion II™ 14219 B - en - 2011/06
bioMérieux, the blue log, VIDAS and SPR are used, pending, and/or registered trademarks belonging to bioMérieux SA or one of its
subsidiaries.
CLSI is a trademark belonging to Clinical Laboratory and Standards Institute, Inc.
Any other name or trademark is the property of its respective owner.
bioMérieux, Inc
bioMérieux SA Box 15969,
RCS LYON 673 620 399 Durham, NC 27704-0969 /
69280 Marcy-l'Etoile / France USA
Tel. 33 (0)4 78 87 20 00 Tel. (1) 919 620 20 00
Fax 33 (0)4 78 87 20 90 Fax (1) 919 620 22 11
http://www.biomerieux.com