I-CAL-GUI-017 V 4.1 Calibration Guideline No. 17
I-CAL-GUI-017 V 4.1 Calibration Guideline No. 17
I-CAL-GUI-017 V 4.1 Calibration Guideline No. 17
2 Methods of calibration
If appropriate, the procedure of calibration should allow according to the client's
requirements the evaluation of the hysteresis, the linearity and the repeatability of the
instrument to be calibrated.
The applied procedure depends on the target uncertainty of the instrument according to
the client's requirements and can limit the lowest achievable uncertainty.
Generally, the calibration is to be carried out in pressure points uniformly distributed over
the calibration range. In the Sections 5.2.1, 5.2.2 and 5.2.3, there are exemplary
recommendations for a calibration range from 0 % FS (full scale) to 100 % FS including
the zero point as one point to illustrate the procedures.
The comparison between the measurement values of the instrument to be calibrated and
the reference standard can be performed by two different methods: