Professional Documents
Culture Documents
Assesment Tool Equipment and Instruments
Assesment Tool Equipment and Instruments
2. Inventory of equipment shall be done every first month of the calendar year.
PROCEDURE:
1. The supplier should inform the laboratory of the expected date and time of the
delivery. The delivery should preferably be done during regular office hours.
3. Inspect the packing for any damages in the presence of the service representative
and the supplier.
4. Check whether the delivered instrument meets the specifications of the quotation.
Any discrepancies should be notified to the supplier.
5. If the packing is intact and no damage to the equipment is anticipated, unpack the
equipment carefully. Retain all the packaging, labels, supports and booklets.
7. The service representative shall assemble and perform a ‘’test run’’ using the
installation and operation manuals.
8. The training of the technical staff of the laboratory should be carried out under the
supervision of the service representative of the company.
12. Only the authorized trained personnel should use the instrument.
Republic of the Philippines
Province of Misamis Oriental
Local Government of Claveria
PROCEDURE:
1. CALIBRATION
a. Equipment calibration should be done according to its specification.
2. PREVENTIVE MAINTENANCE
a. Ensure availability of Med.Tech. or trained personnel to check the functionality of the
equipment.
Microscope
PROCEDURE: DAILY
1.Check the physical condition of the power cord, plug & grounding
system (including other parts)
2. Clean the unit
3. Lubricate rollers and other movable parts (monthly)
4. Check the adjustment of the iris diaphragm & condenser assembly
NOTE :
1. Avoid exposing the unit to high temperature, dust and vibration
2. Cover the unit when not in use
3. Report abnormalities or malfunctions to the maintenance
Department.
NOTE:
1. Maintain proper balancing of load & check speed indicator, if
applicable
2. Unplug the unit when not in use
3. Cover the unit when not in use
4. If the unit is refrigerated, check temperature & leaks
5. Report abnormalities or malfunctions to the Maintenance
Department
Laboratory incubators/oven
PROCEDURE: WEEKLY
NOTE:
1. If the inside chamber is stained, wipe off with a moist
cloth. Never wash it with water
2. Install or store the unit in a location away from direct
sunlight and splashes of reagents and water
3. Avoid exposing the unit to shocks and vibrations
4. Report abnormalities or malfunctions to the
Maintenance Department
Refrigerator
PROCEDURE: DAILY
NOTE:
1. Check if the door closes tightly after every opening
2. Do not splash water on the surface to avoid short circuit
3. Monitor the temperature reading daily
4. Unplug the unit before cleaning
5. In case condensation forms on the front glass due to a very
humid environment, wipe it off with a soft dry cloth
6. Report abnormalities or malfunctions to the Maintenance
Department
Spectrophotometer
PROCEDURE: DAILY
1. Check the physical condition of the power cord, plug, meter grounding
system, knobs and switches
2. Clean the unit
3. Check calibration
NOTE:
1. Check the automatic voltage regulator (AVR) meter
2. Make sure that the unit is installed away from high magnetic
field strength or high frequency radiation and vibration
3. Replace blackened bulb
4. When solvent or sample solution is spilled in the sample
chamber, wipe it immediately with cloth which has been
soaked in soap and water
Republic of the Philippines
Province of Misamis Oriental
Local Government of Claveria
5. When not in use, the instrument should be turned off and
covered
6. Report abnormalities or malfunctions to the Maintenance
Department
NOTE:
1.Report abnormalities or malfunctions to the Maintenance Department
3.REPAIR
PROCEDURE:
a. Breakdown of equipment should be reported immediately to the MHO.
b. Clients in waiting should also be informed. Inform concerned clients of the
unavailability of examinations due to equipment breakdown.
c. Refer patient to other laboratory under MOA or to any laboratory of patient’s
choice.
d. Letter of request for the repair or replacement for a new one must be submitted to
the Lab manager.
e. If equipment is still under the company warranty, inform the supplier.
f. It is essential that all devices are cleaned and decontaminated before being sent for
repair or calibration.
g. In case of equipment loss due to robbery or theft.
A. Identify equipment lost
B. Make a report and document in the incident/accident logbook.
C. Submit report to the MHO.
D. File blotter report to the municipal police force.
The laboratory ensures quality CBC result using automated CBC analyzer. In case of a machine
breakdown or unavailability of reagent/s, manual CBC procedure is not an option.
PROCEDURE:
1. In case of CBC machine breakdown and/ or unavailability of reagent/s:
- The laboratory will not accept CBC request.
- The Laboratory personnel will inform the patient of the unavailability of the test.
- Refer the patient to any licensed laboratory of their choice.
2. When machine breaks down, the Medtech will contact the service engineer/ technician for
Republic of the Philippines
Province of Misamis Oriental
Local Government of Claveria
technical assistance.
3. Inform the Municipal Health Officer or the Physician in charge that no CBC examination will be
done for the day.
PROCEDURE:
1. Do an inventory of equipment.
2. Identify equipments that are obsolete, dangerous or no longer functioning to the standard
required.
3. All instruments and general items, which are unserviceable, should be removed from the
inventory and document on the specified Equipment Preventive Maintenance Form. This has
to be done in accordance with the government regulations.
4. Make a list of the equipment that is to be removed from the inventory and submit to the
Owner.
Republic of the Philippines
Province of Misamis Oriental
Local Government of Claveria
It is a policy of the laboratory to comply with the standards of Hematology procedure to give accurate
and reliable result. The laboratory implements the non-performance of manual CBC procedure. All
Complete Blood Count Examination (WBC Count, RBC Count, Hemoglobin/ Hematocrit
determination, RBC Indices and Platelet count) are performed in the CBC analyzer except for WBC
differential count where manual accounting may be performed.
PROCEDURE:
1. Receive patient’s laboratory request.
2. Perform the implemented CBC laboratory procedure according to the Manual Operating
Procedure. Refer to technical procedures.
3. No manual CBC procedure will be performed except for WBC differential count.