1.

PURPOSE

1.1. Establishes working policy and procedure for the control, calibration, and
maintenance of inspection, measuring, and test equipment; and defines control of
the process for calibrating product.

2. SCOPE

2.1. This document applies to all Corporation facilities.

2.2. Facilities other than Corporate Irvine may also have local versions of this
procedure, or supplements to this procedure. Local versions are the primary
document for the respective facility.

2.3. In the event of a conflict between this procedure and a local version, or if a local
procedure does not address a requirement, this procedure has preference for the
applicable paragraph(s).

2.4. Calibration performed under the scope of ISO/IEC Guide 25 or ISO 17025 are
addressed by the Accredited Calibration procedure, NPP-99171.

3. DEFINITIONS
CALIBRATION
The set of operations which establish, under specified conditions, the relationship
between values indicated by a measuring instrument or measuring system, and the
corresponding standard or known values derived from the standard.

MEASURING AND TEST EQUIPMENT (M&TE)


All of the measuring instruments, measurement standards, reference materials, and
auxiliary apparatus that are necessary to perform a measurement. This term includes
measuring equipment used in the course of testing and inspection, as well as that used
in calibration.

MEASUREMENT STANDARD
A measuring instrument, reference material or system intended to define, realize,
conserve or reproduce a unit or one or more known values of a quantity to serve as a
reference.

QUALITY
Correctness of calibration result and meets the customer’s need.

TRACEABILITY
The property of a result of a measurement whereby it can be related to appropriate
standards, generally national or international standards, through an unbroken chain of
comparisons, with uncertainty stated at each link in the chain.

SIGNIFICANT OUT OF TOLERANCE (OOT) CONDITION


Calibration Page 2

At least twice the identified tolerance of any measurement.

4. POLICY

4.1. Industry Standards

4.1.1. Calibration shall comply with requirements of:


 Calibration Laboratories and Measuring and Test Equipment - General
Requirements (ANSI/NCSL Z540-1-1994)
 Quality Systems - Model for Quality Assurance in Production, Installation,
and Servicing (ISO 9002:1994) section 4.11, Control of inspection,
measuring and test equipment.
 Quality assurance requirements for measuring equipment - Part 1:
Metrological confirmation system for measuring equipment (ISO 10012-
1:1992).
 General Requirements for the Accreditation of Calibration and Testing
Laboratories (CAN-P-4C:1991 a.k.a. ISO/IEC Guide 25)

4.2. Conformance to the requirements of the calibration system

4.2.1. Quality Assurance is responsible for establishing and maintaining


documented procedures to evaluate the adequacy of the calibration system
and to ensure compliance with the requirements of this document.

4.3. Commitment and Objectives

4.3.1. The Corporation Quality Policy is the corporate position on quality. The
calibration program at Corporation, and specifically for product
calibration, supports the corporate Quality Policy with specified
commitments and objectives.

4.3.2. The calibration program is committed to providing outstanding service to


our Customers by maintaining customer satisfaction and by providing high
quality in calibration services to our customers, internal or external.

4.3.3. Both parts of the commitment will be measured by tracking a combination


of calibration related customer complaints and by tracking internal errors
in calibration and/or documentation. The organization(s) actually
performing the calibration will track errors. The Corporation Customer
Complaint Review Board will track customer complaints per NPP-99066.

4.3.4. This measure will be presented to the Quality Council a minimum of once
per year per.

4.4. For the purposes of this procedure, Quality Assurance and Corporate Calibration
Manager refer to personnel within the Corporation Corporate Division responsible
for the duties and/or functions described.

5. PROCEDURE

5.1. Identification of inspection equipment and M&TE to be calibrated. This includes


Calibration Page 3

measuring equipment belongings to employees.

5.1.1. The production area manager or designee that possess inspection


equipment and M&TE, with the assistance of Quality Assurance, will be
responsible for determining:
 Which equipment requires calibration
 Identifying the measurements to be made
 Accuracy (trueness and precision) required
 Selection of the appropriate inspection, measuring, and test equipment
 Selection of the appropriate calibration cycle
 Ensure delivery of the equipment to be calibrated on schedule
 Provide space and power for the calibration provider
 Verify function of equipment received from calibration

5.2. The production area manager or an assigned representative will work with Quality
Assurance to determine the calibration process (subcontracted or in-house) for the
M&TE and/or measurement standards.

5.3. Records

5.3.1. Quality Assurance is responsible for the records documenting that


established schedules and procedures are followed to maintain the
accuracy of all M&TE and measurement standards.

5.3.2. The record will reflect each item requiring calibration including
measurement standards and will provide a description or identification of
the item, calibration interval, date calibrated, identification of calibration
source, calibration procedure used, calibration results and calibration
actions taken.

5.3.3. If accuracy is reported via a calibration certificate or report, it shall be


stated on the report.

5.3.4. If needed, Quality Assurance will assign an ID # sufficient to link the


calibration record (see Exhibit A) with the unit calibrated to maintain
traceability.

5.3.5. All documentation (reports, certificates, or data sheets) that is related to


calibration will be appropriately located in Quality Assurance.

5.3.6. Local facilities may also maintain, or have copies of, the calibration
database and calibration management program as appropriate for the
maintenance of the calibration program at that facility.

5.3.7. Customer confidentiality, proprietary rights and trade secrets ( or


customer), and related types of information shall be protected by all
company personnel.

5.4. Calibration sources and vendors


Calibration Page 4

5.4.1. Quality Assurance is ultimately responsible for the calibration sources and
vendors.

5.4.1.1. The Corporate Calibration Manager (by whatever name) is


responsible to evaluate calibration providers and provide
recommendations regarding usage of their services.

5.4.1.2. The Corporate Calibration Manager is responsible to manage


calibration contracts, including billing, scope of service, delivery
of service, and other related functions and issues.

5.4.2. All M&TE and measurement standards shall be calibrated utilizing


measurement standards whose calibration is traceable.

5.4.2.1. For occasions when traceability is not possible, there must be at


least one sufficiently comprehensive test using traceable
calibrated equipment to check the validity of tests performed with
non-traceable equipment.

5.4.2.2. Traceable equipment may also be used to monitor appropriate


parameters tested by non-traceable equipment.

5.4.3. All measurement standards used in the calibration system shall be


supported by certificates, reports, or data sheets attesting to the description
or identification of the item, the calibration source, date of calibration,
calibration assigned value, statement of uncertainty and environmental or
other conditions under which calibration results were obtained.

5.4.4. Purchasing

5.4.4.1. Purchase orders issued for new equipment requiring calibration


and/or items sent for calibration shall request a certification that
the items have been calibrated under a Quality System that
conforms to the requirements of ANSI/NCSL Z540-1-1994 and/or
ISO Guide 25.

5.4.4.2. In the event that conformance to ANSI Z-540 and/or ISO Guide
25 is either cost and/or time prohibitive, or is not available from a
vendor, the requirement may be left off the purchase order,
provided the MAPICS system indicates the item must be
calibrated before being placed into service. Calibration shall be
performed by an approved vendor.

5.4.4.3. Purchase Orders to a National Standards body (e.g. NIST, NRC)


do not require the ANSI Z-540 or ISO Guide 25 statement.

5.4.5. Quality Assurance

5.4.5.1. Outside agencies used for calibration purposes will be approved


by Quality Assurance, and will be surveyed by qualified
personnel to verify the calibration system conforms to the
requirements of ANSI/NCSL Z540-1-1994 and/or ISO Guide 25.

5.5. Calibration status

5.5.1. All M&TE and measurement standards shall be labeled to indicate


calibration status. The label shall include the date calibrated (day, month,
Calibration Page 5

and year), the serial or identification number, the specific calibration due
date, and who performed the calibration.

5.5.2. Items not calibrated to their full capability, or have limitations of use,
shall be labeled or otherwise identified as to the limitation. For impractical
applications of calibration label, the label will be affixed to instrument
container or some other suitable measures will be used to reflect
calibration status.

5.5.3. Tamper-resistant seals shall be affixed to control adjustment or


panels/screws leading to control adjustment on M&TE and measurement
standards.

5.5.4. If the seal is removed, Quality Assurance shall be informed and a Quality
Assurance representative shall determine if removal of the seal affects the
performance of the M&TE.

5.5.5. Quality Assurance and production area managers are responsible for
maintaining all M&TE and measuring standards to a calibration status.
This includes new equipment.

5.5.5.1. Area managers, or their designee, are responsible for delivering


equipment due for calibration to Quality Assurance in the time
and manner prescribed by Quality Assurance.

5.6. Adequacy of measurement standards

5.6.1. All measurement standards used for calibration of M&TE and other
measurement standards shall have the accuracy, stability, range and
resolution required for the intended use, and be traceable to the National
Institute of Standards and Technology or other national standards body.

5.6.2. The collective uncertainty of the measurement standards shall not exceed
25% of the acceptable tolerance for each characteristic being calibrated.

5.7. Acceptable tolerance

5.7.1. Equipment that is in a significant OOT condition will be impounded until


it can be recalibrated, repaired, replaced, destroyed, or identified as "For
Indication Only".

5.7.2. Equipment in an OOT condition that is not significant may receive a


limited calibration, specifying calibration parameters, providing the
equipment is not used for product certification or other critical
measurements.

5.8. Environmental controls

5.8.1. All M&TE and measurement standards shall be calibrated and utilized in
an environment controlled to the extent necessary to assure continued
measurements of required accuracy, giving due consideration to
temperature, humidity, vibration, cleanliness, and other controllable
factors.

5.8.2. M&TE in Receiving Inspection and manufacturing areas shall be kept in a


temperature environment between 15° C and 30° C (60 deg F to 85 deg F)
Calibration Page 6

and relative humidity (RH) of 20% to 70%.

5.8.3. Special working areas where temperature or RH is critical shall be


identified.

5.8.4. Where the environment departs from acceptable conditions, compensating


corrections shall be applied to the calibration results obtained or the
calibration activity shall be stopped until the environmental condition can
be corrected. Compensating conditions may be physical or software
enabled.

5.9. Intervals of calibration

5.9.1. The intervals of calibration shall not exceed 3 years.

5.9.2. Quality Assurance and production area manager will be responsible for
determining the intervals of calibration.

5.9.3. M&TE and measurement standards shall be calibrated at periodic intervals


established and maintained to assure acceptable accuracy and reliability.
The Intervals are based on accuracy, reliability, purpose and degree of
usage.

5.9.4. Intervals shall be shortened or may be lengthened, when the results of


previous calibrations indicate that such action is appropriate to maintain
acceptable reliability. Newly purchased equipment shall be placed on the
next calibration interval unless certification is provided with the
equipment. An established recall system for the mandatory recall of the
M&TE and measurement standards to assure timely recalibrations shall be
incorporated.

5.10. Recall of M&TE

5.10.1. The computer database software GAGEtrak will be used by Quality


Assurance to coordinate the recall of equipment requiring recalibration.

5.10.2. Equipment requiring recalibration shall be calibrated within 4-weeks from


the calibration due date. Equipment that exceeds calibration due date for
more than 4-weeks shall be impounded until it is calibrated.

5.10.3. The manager of the area where M&TE is located, or designee, shall be
responsible for compiling the equipment requiring calibration, with
assistance from QA.

5.10.4. When equipment is turned in for calibration, the calibration manager or


designee shall employ adequate safeguards to ensure equipment is
controlled until the equipment is returned to the owning department, and
to ensure the equipment is correctly returned.

5.11. Calibration procedures

5.11.1. The affected production area manager, with the assistance of Quality
Assurance will be responsible for the generation of calibration procedures
for all M&TE and measurement standards.

5.11.2. Procedures shall be utilized for the calibration of all M&TE, measurement
standards, and product. The calibration procedure shall specify the
Calibration Page 7

measurement standards and details of equipment type, identification


number, frequency of checks, check method, the required parameter,
range, and accuracy of the measurement standards and the acceptable
tolerance of each instrument characteristic being calibrated.

5.11.2.1.Procedures used may be those available from DOD or other U.S.


Government agencies, available in manufacturer's manuals, or
published standard practices, provided they meet the requirements
of this document.

5.11.2.2.Test equipment measured or tested out of tolerance shall be


reviewed by a quality engineer to decide to what extent the
condition will affect the product performance.

5.11.3. The quality engineer with the assistance of others, if necessary, shall
determine if any corrective action is necessary, and the results shall be
recorded on the certificate of calibration form.

5.12. Test fixtures

5.12.1. When using test hardware (e.g., jigs, fixtures, templates, and patterns) or
test software as suitable forms of inspection, they shall be checked to
prove that they are capable of verifying the acceptability of product prior
to use.

5.12.2. Hardware/software shall be rechecked at periodic intervals. The extent and


frequency of such checks shall be established through usage and records
shall be maintained as evidence of control.

5.12.3. Process Assurance Engineers (PAEs) are responsible to identify the


periodic interval (cycle) at which fixtures shall be checked. The cycle
shall not exceed 2 years. PAEs will provide QA with appropriate
information once the scheduled checks are completed. QA will maintain
the records per NPP-99012.

5.12.4. The safeguarding of inspection, measuring, and test facilities for both
hardware and software from adjustments which would invalidate the
calibration setting shall be incorporated.

5.12.5. Measurement design data shall be made available, for verification that it is
functionally adequate.

5.13. Handling and storage

5.13.1.1.All M&TE and measurement standards shall be handled, stored


and transported in a manner which will not adversely affect the
calibration or condition of the equipment.

5.14. Positive recall action

5.14.1. All products inspected, measured, and tested with a measuring and test
equipment that is malfunctioning shall be held until appropriate corrective
action has been taken.

5.14.2. If an item of M&TE is found to be out of calibration by more than 30


days, or during calibration was found to be out of tolerance, then the
Calibration Page 8

department which owns the M&TE shall investigate the validity of


certifications and / or calibrations performed using that item of M&TE.

5.14.3. If the certifications and / or calibrations are found to be invalid, the


department, in cooperation with QA and any other departments that are
directly affected, shall determine whether to recall product, and the extent
of such a recall.

5.14.4. The department shall document the results of the investigation per NPP-
99037. Records shall be maintained per the Quality Manual.

5.15. Thread gauges

5.15.1.1.Thread gauges used to inspect products shall be tracked by usage


and/or time. Records of usage shall be maintained as evidence of
control, where appropriate.

5.15.1.2.Thread gauges with estimated usage over 100/year will be


identified and calibrated every 12 months. Thread gauges with
estimated usage under 100/year will be identified and calibrated
every 36 months.

5.15.1.3.Thread gauges shall be permanently identified with a unique


identification such that gauges can be recorded separately. The
identification is never duplicated, even for identical replacements.

5.15.1.4.The threaded portion of thread plugs may be replaced with new


items without requiring a new identification number. The new
threads must be purchased with a calibration certificate compliant
to Z540-1, and retained on file with other calibration certificates.

5.16. Product calibration

5.16.1.1.Calibration of product for sale to customers shall include a


Certificate of Calibration and a Calibration Sticker. The
Calibration Sticker shall have the S/N, By, and Date areas
completed. The Due area may be left blank. This will be
determined by appropriate management personnel for the product
being produced. A typical example of a calibration sticker is
provided in Exhibit B. Exact format may differ.

5.16.1.2.If the due area of the calibration sticker is left blank, a note shall
accompany the calibrated product stating that at time of
production, the item was calibrated to the applicable standard,
and that our Quality System and storage methodologies safeguard
the integrity of the calibration until shipment to the customer. The
note will also inform the customer that for tracking purposes the
customer should fill in the due date as one year from delivery of
the product.

5.16.1.3.Actual calibration of product is detailed in lower level calibration


procedures per.

5.17. Product recalibration

5.17.1.1.When a customer sends product for recalibration, after receipt the


Calibration Page 9

product will be routed to the appropriate technician who will


determine whether the product can be calibrated in-house, or
must be sent to the manufacturer or calibration vendor. The
product will be routed and calibrated in accordance with the
technician’s determination. An appropriate record must be made
of the determination and routing.

5.17.1.2.Upon completion of the calibration, the technician shall perform


an appropriate inspection of the calibrated product. An
appropriate record must be made of the inspection, including
criteria used to perform the inspection, any deficiencies noted,
and whether the product is accepted or rejected.

5.18. Once the product is inspected and the technician has accepted the product, it shall
be returned to the customer. A Certificate of Calibration and a Calibration Sticker
shall accompany the product. The Calibration Sticker shall have all areas
completed, including Due.

5.19. Calibration certificates

5.19.1. Calibration Certificates provided to the customer shall meet the


requirements of ISO Guide 25.

5.19.2. Calibration Due area of the calibration certificate may be left blank, at
management discretion, to compensate for differences in order and
delivery methodologies between different products.

5.19.3. Recalibrations require Condition on Arrival and Condition on Departure


be documented, regardless of product.

5.19.4. Any Out Of Tolerance condition noted during a product recalibration must
be documented, including test type, nominal (desired) reading, actual
reading, and tolerances or uncertainty. The information may be provided
as an attachment to the calibration certificate.

5.20. Customer Complaints

5.20.1. Customer complaints shall be addressed via the Corporation Customer


Complaint Review Board per.

5.20.2. If, in the Board’s opinion, there is legitimate doubt about the integrity of
the calibrations, or adherence to procedures, or other reasons as defined by
the Board, an audit shall be performed by personnel as assigned by the
Board. The audit shall be documented per.

5.21. Quality Assurance and the production area manager or appropriate representative,
will ensure that the measurement equipment and M&TE are calibrated within the
interval of calibration.

5.22. Quality Assurance will control the out-of-tolerance conditions of equipment. If


needed, Quality Assurance will be responsible for taking the appropriate action for
notification of the out-of-tolerance condition to the respective user.

5.23. Quality Assurance will maintain calibration records (reports, certificates, or data
sheets) of M&TE and/or measurement standards in an appropriate location.
Calibration Page 10

5.24. Quality Assurance will generate a list of equipment due for calibration.

5.25. Quality Assurance will file the calibration record as a hardcopy file or in a data
base system. A typical example is shown in Exhibit A. Exact format may differ.

5.26. Subcontracted calibration

5.26.1. Sub-contracted calibration refers to calibration provided by an


organization external to Corporation, and providing calibration
certificates representing Corporation.

5.26.2. Sub-contracted calibration providers must be evaluated by Quality


Assurance, approved, and provide calibration compliant with ANSI/NCSL
Z540-1.

5.26.3. The organization responsible for the product receiving sub-contracted


calibration is also responsible to provide a cover calibration certificate
informing the customer that the calibration was provided by an approved
calibration provider.

5.27. In-house calibration

5.27.1. Quality Assurance, the production area manager, and/or assigned qualified
personnel:
 Ensure that the measurement equipment and M&TE are calibrated within
the interval of calibration.
 Perform the calibration procedure.
 Label the M&TE or measurement standard.
 Document the calibration actions of appropriate calibration records.
 Formally notify any departments affected of the completion of calibration
action(s). This notification may be by e-mail or other formal method.

5.28. Impound

5.28.1. Quality Assurance and the production area manager


 Recall and transport M&TE or measurement standards that are out of
calibration status to the calibration impound.
 Tag with a sign indicating no usage is allowed until corrective actions
have been taken the M&TE that are out of calibration status and difficult
to remove from the working area.

5.29. Asset Control

5.29.1. Departments turning in equipment for calibration may have QA or the


calibration company sign for the equipment. A sample format for a
signature form is provided in Exhibit C, exact format may differ.

5.29.2. Equipment turned in for calibration, and signed for, must have positive
control. The methods may vary, but should be appropriate to the type of
equipment turned in.
Calibration Page 11

5.29.3. Equipment maintained in the Impound Area will be reviewed at least


annually. This review will attempt to determine if items impounded may
be used within the company, or must be written off per NPP-99049. Items
that are written off may be donated to local non-competitive organizations
such as High Schools or Colleges, or disposed of.

5.29.4. Equipment that appears to be useable, but is not identified as being owned
by any department, will be maintained in the impound area.

6. EXHIBITS
Calibration Page 12

7.1. Exhibit A - Calibration Record


DESCRIPTION_______________________________________ I.D.__________
CALIB. INTERVAL_____________ ACTIVE__________ INACTIVE___________
CALIB. SOURCE____________________________________________________
CALIB. PROC._____________________________________________________

CAL. DATE BY DUE RESULTS ACTION REMARK

7.2. Exhibit B - Calibration Sticker


Note: Phone number is not required on Calibration Sticker.

CALIBRATION
SN___________
By_____Day___
DUE__________
Calibration Page 13

6.3 Exhibit C – Calibration signature log sample

Date:

Item No. Part Number S/N or Asset No. Print/Sign Organization