Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 11

Table of Contents

1. Overview............................................................................................................................................1
2. Common malfunction and remedy....................................................................................................1
2.1 Malfunctions due to improper installation environment............................................................1
2.2 Common malfunction and remedy...........................................................................................1
2.3 Check whether power supply is normal...................................................................................2
3. Malfunction and remedy....................................................................................................................3
3.1 No response when turn on instrument.....................................................................................3
3.2 No display on the screen..........................................................................................................3
3.3 The test results cannot be printed out through the internal printer..........................................5
3.4 Inactive reagent strip sensor....................................................................................................5
3.5 Time on the screen is abnormal...............................................................................................6
3.6 Strip holder moves abnormally.................................................................................................6
3.7 Strip holder doesn’t move........................................................................................................6
3.8 Buttons don’t work....................................................................................................................7
3.9 TROUBLE-01~06.....................................................................................................................7
3.10 TROUBLE-07.........................................................................................................................7
3.11 TROUBLE-10.........................................................................................................................7
3.12 TROUBLE-12.........................................................................................................................8

VERSION: 06/2020
1. Overview

URIT-50 urine analyzer is manufactured under strict supervision and adjusted by high technical
personals. This is a kind of precise analyzer, improper operation or any damaged component may
affect the test results or cause damage to the instrument. The instrument should be repaired only by
professionals with certain knowledge of electronic technique. The method of repair bases on
replacing components. This section of the manual lists various errors and messages, together with
probable causes and remedies recommended to correct the problems efficiently. Consult Table 2 for
common malfunction and remedy and table 3 for malfunction code and remedy.
Caution:
Be sure to turn off the power supply before opening the instrument cover or replacing parts.
Beware of the high voltage inside.

2. Common malfunction and remedy

2.1 Malfunctions due to improper installation environment.

Consult table 1.
Table 1
No Installation Environment Malfunction

1 Strong light Incorrect result

2 Temperature exceeds the recommended range Incorrect result

3 Electromagnetism interference Incorrect result

4 Dusty, moist or caustic gas Cause damage to precise parts

5 Out of horizon Incorrect result


2.2 Common malfunction and remedy

Preparation before troubleshooting: check whether power supply is normal. Consult 2.3
Consult Table 2 for common malfunction and remedy and table 3 for malfunction code and remedy.
Table 2: Common malfunction and remedy
S/N MALFUNCTION CAUSE REMEDY
1. The input voltage falls short of the requirement.
No response when 2. Power connector assembly is loose or defective.
1 turning on the 3. Defective socket. 3.1
instrument 4. Defective power switch.
5. Abnormal adapter output.
2 No display on the 1. Display board 20-core cable is loose or defective. 3.2
screen 2. Improper LCD contrast grade.
3. Damaged LCD module.

1
4. Defective LCD drive circuit.
1. The printer is OFF.
2. Printer interface board 10-core cable is loose or
Test results cannot
defective.
3 be printed out by the 3.3
3. Damaged printer.
internal printer
4. Problem with printer drive circuit.
5. Problem with printer heat circuit.
1. The sensor is OFF.
Inactive reagent strip
4 2. Damaged sensor. 3.4
sensor
3. Problem with sensor circuit.
1. Damaged crystal oscillator.
Time on the screen is
5 2. Damaged timer chip. 3.5
abnormal.
3. Battery loose or out of electricity.
1. Strip holder is broken.
Strip holder moves
6 2. Inner spindle is broken down. 3.6
abnormally.
3. Drive spindle is broken down.
1. Motor wire plug-in not well inserted.
Strip holder doesn’t
7 2. Motor is broken. 3.7
move.
3. Problem with motor control circuit.
1. Connector between panel and main board is not
Buttons don’t work
8 well connected. 3.8
properly
2. The panel is broken.

Table 3: Malfunction code and remedy


CODE CAUSE REMEDY

TROUBLE-01 Reference LED Light is too strong.

TROUBLE-02 Red LED Light is too strong.

TROUBLE-03 Green LED Light is too strong.


3.9
TROUBLE-04 Reference LED Light is too strong.

TROUBLE-05 Red LED Light is too weak.

TROUBLE-06 Green Light is too weak.

TROUBLE-07 Strip placement is incorrect. 3.10


Place correct strips ( Consult
TROUBLE-08 Incorrect strip setting.
operation manual).
TROUBLE-09 No strip is on strip holder. Place strip according to warning.
Instrument is not in good work condition (chip
TROUBLE-10 3.11
on main board is broken).
Strip holder is dirty or something is on the Clean the strip holder or take off
TROUBLE-11
holder. the foreign body.
Driving mechanism, photo-sensor or circuit is
TROUBLE-12 3.12
defective.

2
2.3 Check whether power supply is normal

Consult figure 2.1.


Instrument is in normal operation, the primary condition is power supply is normal and stable, so it is
necessary to check power supply.
1) Solve according to method 3.1, if TP2(+12V) voltage not within range or zero, replace J5.
2) TP3(+8.4V) voltage not within range, replace U4.
3) TP4(+5V) voltage not within range, replace U2.
4) TP5(+3.3V) voltage not within range, replace U7.

U7

J5

U4

TP2/3/4/5

U2

Figure 2.1

3. Malfunction and remedy

3.1 No response when turn on instrument

Measure the voltage of TP2 as figure 2.1 shows, (normal voltage is +12V), check power adapter,
power input socket, power switch, power input connector, power input wire if the voltage not within
the range, and replace the defective parts. If all these components are all right, please check the
voltage as 2.3. If the voltage is OK, please replace the main board.
3.2 No display on the screen

As shown in figure 3.1 and 3.2, Check if the voltage of TP6 is -7±2V when the contrast grade is

3
adjusted to be >>>> on the screen. This voltage can be adjusted through “LCD contrast adjust
potentiometer (RP1)”. Replace the potentiometer or U6 if can not be adjusted. Check whether LCD
20-core cable is well connected if the voltage is normal. Please replace LCD module if problem still
persists by re-plug LCD 20-core cable.

4
U6
J4 J5

TP6

RP1

J16 J3

Figure 3.1
Printer 10-core cable

Printer connection cable

Panel socket

LCD 20-core cable

Figure 3.2
5
3.3 The test results cannot be printed out through the internal printer

Printer board consult figure 3.3.


1) Confirm whether the printer is in good work condition;
2) First, check whether power supply is normal(TP2, TP5); check whether printer 10-core cable(J1)
on printer board and printer connection cable(J4) are well connected, re-plug printer 10-core
cable and printer connection cable;
3) Select print in review if above mention is no problem, printer no response and there is character
on paper, that is U3 insufficient solder or broken;
4) Printer paper not character, first check whether the paper is reversely installed, if paper is
correctly install please test TP1(8V) and Q1(0.6V) voltage during paper skip.
①TP2 and Q1 voltage are normal, that is U3 insufficient solder or broken;
②Q1 voltage is normal, TP1 no voltage, that is U5 insufficient solder or broken;
③Q1 no input voltage, that is U4 insufficient solder or broken.
5) If above mention are normal, that is U2 insufficient solder or broken.

J1 U3 U2 J4 U5 Q1 TP1/2/3/4/5

VP
8V
PS
GND
+5V

U4

Figure 3.3

3.4 Inactive reagent strip sensor

As shown in figure 3.4, check the voltage of TP2 (ensure J3 on main board and JP1 on limit board
are well connected), the voltage of TP2 should not be exceeding 1.5V if there is on strip on the
holder; and should be more than 3.5V if strip is on the holder and under the strip sensor. Please
replace strip sensor if the voltage is not in the range. If problem persists, change U2 (LM393/SO) on
limit board.

6
Figure 3.4
3.5 Time on the screen is abnormal

As shown in figure 3.6, relocate the battery. Check whether the crystal oscillator (Y1 32.768K) and
U10 is normal.
3.6 Strip holder moves abnormally

If strip holder moves abnormally, the most possibly reason are like that: gear of strip holder is
rubdown, inner spindle is rusted and transmission shaft is rubdown. Replace relative part. Strip
holder, inner spindle and transmission shaft are shown in figure 3.5.

Figure 3.5
3.7 Strip holder doesn’t move

Power supply is normal and motor wire socket J1 is well connected, main cause is that motor is
broken or motor control circuit defective. Motor control circuit U1 broken with high probability, consult
figure 3.6 for chip position.

7
3.8 Buttons don’t work

As figure 3.2, re-plug panel socket if button do not work, replace panel if above is invalid..
3.9 TROUBLE-01~06

Select “Menu” in the ready screen then select “System Setting” to get to the system setting screen.
Press “F3” for a few seconds till a “di…”sound is heard, then press “F1-F2-F3-F4” in turn to get to
adjustment mode. Select “Strip”-“LED” to set the values of parameters 720, 620, 550 (usually the
value is set to be 3000), then press “OK”. If the setting is success, there will be a prompt “OK” above
the “Return” key. If the setting is fail, see table-4 to set.
Table-4
CODE CAUSE REMEDY
TROUBLE-01 Reference LED Light is too strong. Reduce the value of parameter 720.
TROUBLE-02 Red LED Light is too strong. Reduce the value of parameter 620.
TROUBLE-03 Green LED Light is too strong. Reduce the value of parameter 550.
TROUBLE-04 Reference LED Light is too strong. Increase the value of parameter 720.
TROUBLE-05 Red LED Light is too weak. Increase the value of parameter 620.
TROUBLE-06 Green Light is too weak. Increase the value of parameter 550.

Instructions:
If “TROUBLE-?” is displayed when turn on the instrument, press any key on the panel till get to the
ready screen when turn on the instrument.
3.10 TROUBLE-07

The strip should be placed in the right position otherwise TROUBLE-07 will be displayed. The strip
should be put in the groove of strip holder, and put it in the end of groove.
CAUTION: When the strip holder moves abnormal, TROUBLE-07 will also be displayed. See 3.6 for
details.
3.11 TROUBLE-10

As picture 3.6,check U10,Y2 on main board, replace relative chip if problem occurs.

8
U10

Y2

U1

J1

J16

Figure 3.6
3.12 TROUBLE-12

Test the voltage of 4 pin in the sensor output connector as shown in figure 3.7. When the white bar
for location is just under the photo-sensor U1, namely the strip holder is back to the initialization
position, the voltage should be more than 1.5V. If the voltage is not in the range, please check if the
white bar for location is broken off or dirty. If the white bar for location is normal, please replace the
photo-sensor (U1).
CAUTION: U1 should be positioned in the left compare with the middle of strip holder. If the position
of U1 is in the middle, the strip sensor will detect the white contrast block and the strip holder will not
get to the initialization position.

9
U1 4 Pin of sensor output connector

Figure 3.7
CAUTION:
The way to get to the adjustment screen: From “Menu” screen get to “About” screen, then
press F3 for 3 seconds, and then press F1, F2 and F3 in turn.

10

You might also like