U50 - User Manual 10 11 14G V1.05

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Table of Contents

Copyright and Declaration....................................................................................................................................1

Part 1: Operation Introduction............................................................................................................................. 3

1. Intended use...................................................................................................................................................... 3

2. Specifications.....................................................................................................................................................3

3. Unpacking.......................................................................................................................................................... 4

4. Structure and Principle..................................................................................................................................... 4

4.1 Structure.......................................................................................................................................................4
4.2 Principle........................................................................................................................................................4
4.3 Contrast of semi-quantitative symbol and concentration..................................................................... 5

5. Outer Appearance and Components............................................................................................................. 7

5.1 Appearance..................................................................................................................................................7
5.2 Components................................................................................................................................................ 7

6. Function.............................................................................................................................................................. 8

6.1 Self-checking and detecting......................................................................................................................8


6.2 Print...............................................................................................................................................................8
6.3 Display..........................................................................................................................................................9
6.4 Data storage and review............................................................................................................................9
6.5 External printer............................................................................................................................................9
6.6 Communication with computer................................................................................................................. 9
6.7 Connection with a bar code scanner....................................................................................................... 9

7. Installation........................................................................................................................................................ 10

7.1 Environment requirement........................................................................................................................10


7.2 Installation of printer paper..................................................................................................................... 10
7.3 Connection with Power Cord.................................................................................................................. 10
7.4 Connection with External Printer............................................................................................................10
7.5 Connection with computer.......................................................................................................................11

8. Operation..........................................................................................................................................................12

8.1 General information..................................................................................................................................12


8.2 Operation................................................................................................................................................... 12
8.3 Calibration and Quality Control.............................................................................................................. 15
8.4 Preparation before Testing......................................................................................................................16
8.5 Routine Test...............................................................................................................................................16
8.6 Serial Interface..........................................................................................................................................18
8.7 PS/2 port.................................................................................................................................................... 19

9. Cleaning and maintenance............................................................................................................................20


9.1 Precautions for Maintenance..................................................................................................................20
9.2 General cleaning...................................................................................................................................... 20
9.3 Daily cleaning............................................................................................................................................20
9.4 The dismounting of strip holder..............................................................................................................21

10. Storage and Transportation........................................................................................................................ 22

10.1 Storage Condition.................................................................................................................................. 22


10.2 Transportation Condition.......................................................................................................................22

Part 2: Service Manual....................................................................................................................................... 23

1. Introduction...................................................................................................................................................... 23

2. Trouble-shooting............................................................................................................................................. 23

Appendix 1: The meanings of marks that appear on the instrument or label are as followings............. 24

Appendix 2: Exchangeable Parts......................................................................................................................25

Appendix 3: Format of test results....................................................................................................................26


Copyright and Declaration
Copyright: C URIT MEDICAL ELECTRONIC CO., LTD. (URIT)

Declaration:

Thank you very much for your purchase of the model URIT-50 urine analyzer.

All contents in this manual were strictly compiled according to the related laws and regulations in

china, as well as the specific condition of URIT-50 urine analyzer, covering all the updated

information before printing. URIT is fully responsible for the revision and explanation of the

manual, and reserves the right to renovate the relevant contents without separate notice. Some of

the demonstration pictures in this manual are for reference and subject to real object if any

differences.

All the information included is protected by copyright. No any part of this document may be

reproduced, stored or transmitted in any form or by any means unless written authorization by

URIT.

All instructions must be followed strictly in operation. In no event should URIT be responsible for

failures, errors and other liabilities resulting from user's noncompliance with the procedures and

precautions outlined herein.

Limitation of Liability:

URIT warrants to the original purchaser that this instrument will be free from defects in materials

and workmanship for a period of one year from the later of the date of original purchase or

installation.

URIT assumes no liability in the following situations even during the period of warranty.

1. Failure due to abuse the instrument or neglect the maintenance.

2. Use reagents and accessories other than manufactured or recommended by URIT.

3. Failure due to operation not under the instructions described in the manual.

4. Replace accessories not specified by URIT, or after maintenance or repair by a service agent

not approved or authorized by URIT.

NOTE:

URIT makes no warranties, either expresses or implied, as to product quality, performance,

and value as a commodity or applicability for any particular purpose.

1
Technical service and troubleshooting are provided by URIT. If the instrument has

malfunction, please contact the agency authorized by URIT.

URIT Medical Electronic Co., Ltd.


Address: No.D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004,

P.R.China

Tel: +86(773)2288586

Fax: +86(773)2288560

Web: www.urit.com

Email: service@uritest.com

VERSION: 02/2022

2
Part 1: Operation Introduction

1. Intended use
The instrument can be used in combination with urine reagent strips for semi-quantitative or
qualitative detection of biochemical components in urine samples, which can provide reference for
clinical examination and diagnosis. The items that can be detected are Urobilinogen (URO),
Bilirubin (BIL), Ketone (KET), Blood (BLD), Protein (PRO), Nitrite (NIT), Leukocyte (LEU),
Glucose (GLU), Specific Gravity (SG), pH, Microalbumin (MA), Creatinine (CR), Vitamin C (VC)
and Calcium (Ca), etc.

NOTE:
Only URIT Urine Reagent Strips guarantee the precision.
NOTE:

The instrument is for professional, in vitro diagnostic use ( ).

2. Specifications
1) Reagents: URIT 10G, URIT 11G, URIT 14G Urine Reagent Strips produced by URIT Medical
Electronic Co., Ltd. (The reagent strips type which can be used have been set to the instrument
by manufacturer before delivery).
2) Measuring Principle: Reflectance Photometry.
3) Throughput: 60 tests / hour or 125 tests / hour (max).
4) Measuring Mode: Continuous test and single test modes. It can sense the strip automatically.
5) Urinary parameters analyzed: Leukocyte, Ketone, Nitrite, Urobilinogen, Bilirubin, Protein,
Glucose, Specific Gravity, Blood, pH, Ascorbic Acid, Creatinine, Calcium and Microalbumin.
6) Display: LCD display with English prompt menu.
7) Print: Thermal printer built-in to print the results. It can also work with external printer (serial or
parallel port).
8) Operating Environment: 15℃~30℃, RH≤80% (Recommended);
9) Control Function: Self-checking, test and trouble-shooting judge are controlled by the
microcomputer.
10) Lumination: ≤2000lx.
11) Communication port: communication with the computer via the RS232 or parallel port, and
PS/2 port to bar code.
12) Memory: 5000 samples could be reserved.
13) Weight: About 2 kg.
14) Outer Dimension: 290mm×200mm×100mm.
15) Power Requirement: Instrument: 12V , The Adapter: 100V-240V~ 50/60Hz.
16) Power Consumption: 30VA.
17) Environment-Friendly Use Period: 10 years.

3
3. Unpacking
1) Carefully unpack the shipping carton and take out URIT-50 instrument and accessories.
2) Check the contents with following Packing List for quantity and any breakage.
3) Please notify the carrier or distributor at once if any damage or loss exists.

4. Structure and Principle


4.1 Structure

The structure is as outlined in figure 1. This instrument is composed of optical-electronic detector


system, transfer-part, I/V converter, CPU, urine draining system, printer, LCD display, keyboard
and power supply.

Figure 1 Structure of the instrument


4.2 Principle

The analyzer measures change in reflectance of the reagent strip pads. A detector integrated in
the system is composed of a light source and a light receiver, the light from which goes through
the spherical integrator and reflects at the reagent pads on the strip. The absorbance (and
reflectance) varies according to the color of the reagent pads. The darker the reagent pad is, the
higher the absorbance is, and less light is reflected. Conversely, the lighter the reagent pad is, the
lower the absorbance is, and more light is reflected; i.e. the degree of color development is
proportional to the concentration of an analyte in urine.
The reflected light goes into the optical-electronic detector system, which transforms the optical
signal into electrical signal. The strength of the electricity correlates with reflectance. Then the
electric signals will be processed by CPU after being transformed by I/V converter, and the test
results can be printed out by printer.

4
The reflectance is calculated using the following formula:

Tm · Cs
R%= ×100%
Ts · Cm

R: Reflectance
Tm: Reflective light amount at the reacted pad with the measurement wavelength
Ts: Reflective light amount at the reactive pad with the reference wavelength
Cm: Reflective light amount at the calibration pad with the measurement wavelength
Cs: Reflective light amount at the calibration pad with the reference wavelength

However, ACR is measured and calculated with the below formula.

"R"ACR = (100 – "R"MA) / "R"CR

These reflectance “R” are compared against calibration curves stored in the instrument and

output as measurement results. Also, temperature is corrected to eliminate changes caused by

environmental temperature.

4.3 Contrast of semi-quantitative symbol and concentration

The factors that may affect the test results are freshness of urine sample, drinking & eating habits
of the patient, as well as pre-sports & post-sports activities. Therefore, the semi-quantitative value
is good enough for clinical diagnosis. The test results are rounded off. It is permissible that two
relevant values may vary slightly in repeated tests.
Table-1 shows the test method and data.
TABLE-1
Analyte Semi-Quantitative Symbol and Concentration
Semi-Quantitative - +- +1 +2 +3
LEU leu/μL(Conv.) 0 15 70 125 500
CELL/μL(SI) 0 15 70 125 500
Semi-Quantitative - +- +1 +2 +3
KET mg/dL(Conv.) 0 5 15 40 80
mmol/L(SI) 0 0.5 1.5 4 8.0
NIT Semi-Quantitative - +
Semi-Quantitative +1 +2 +3
URO mg/dL(Conv.) Normal 2.0 4.0 8.0
μmol/L(SI) 33 66 131
Semi-Quantitative - +1 +2 +3
BIL
mg/dL(Conv.) 0 0.5 2.0 6.0

5
μmol/L(SI) 0 8.6 33 100
Semi-Quantitative - +- +1 +2 +3
PRO mg/dL(Conv.) 0 15 30 100 300
g/L(SI) 0 0.15 0.3 1.0 3.0
Semi-Quantitative - +- +1 +2 +3 +4
GLU mg/dL(Conv.) 0 5 100 250 500 1000
mmol/L(SI) 0 2.8 5.5 14 28 55
SG 1.005 1.010 1.015 1.020 1.025 1.030
Semi-Quantitative - +- +1 +2 +3
BLD mg/dL(Conv.) 0 0.03 0.075 0.24 0.6
CELL/μL(SI) 0 10 25 80 200
pH 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0
Semi-Quantitative - +- +1 +2 +3
VC mg/dL(Conv.) 0 10 25 50 100
mmol/L(SI) 0 0.6 1.4 2.8 5.6
mg/dL(Conv.) 10 50 100 200 300
CR
mmol/L(SI) 0.9 4.4 8.8 17.6 26.4
mg/dL(Conv.) 4 10 20 30 40
Ca
mmol/L(SI) 1.0 2.5 5.0 7.5 10
mg/dL(Conv.) 1 3 8 15
MA
mg/L(SI) 10 30 80 150
Semi-Quantitative Normal +1 +2
ACR mg/mmol <3.4 3.4~33.9 >33.9
mg/g <30 30-300 >300

6
5. Outer Appearance and Components
5.1 Appearance

The appearance of URIT-50 is shown in figure 2.

Figure 2
5.2 Components

1) Display screen: To display the operating prompts and test results. Press the corresponding key
to finish testing.
2) Function Key: There are 4 keys under the screen. Press the keys to perform routine testing
according to the operating prompts displayed on the screen.
3) Printer and printer cover: Thermal printer for printing test result; lift the printer cover to load
thermal printer paper.
4) Strip holder: to hold strips to test.
5) Power switch: To turn the power supply on or off.
6) Power Socket: Do use the adapter packing with the instrument to ensure the instrument runs at
correct voltage.
7) Serial interface: Standard 9 pin Serial interface, RS232 port.
8) PS/2 interface: Standard 6 pin mini DIN socket for bar code scanner.
9) Parallel Interface: Standard 25 pin socket for external printer.

Caution
Be sure to install the external interface of the instrument when the instrument is turned off,
so as not to cause the instrument damage or malfunction.

7
6. Function
6.1 Self-checking and detecting

When the instrument is ready for testing after self-checking, it can detect a strip and test
automatically.
6.2 Print

The display screen will show the results after test, and the printer will print the results
automatically.
An example of printed result is as following (figure 3):

Figure 3
Note: The printing order is different as different strip type. The detail of the order is shown in the
label on urine reagent strip packing.
Line 1: ID:XXXXXXXXXXXXX is the serial number of bar code input.
Line 2: NO.000001 is the serial number of sample. 2005-01-10 is the date (Year-Month- Day).
Line 3: 10:15:56 is the time (Hour: Minute: Second).
Line4 to Line 14: Analyte names, positive/negative notations, test results and units.
The first column: Abbreviation of analytes. Refer to Table-2 for the full names.
The second column: Semi-quantitative symbols. "-" stands for negative and"+"stands for positive.
Degrees of positive are represented by+2, +3....
The third column: Objective values of the analytes, which are displayed by semi-quantitative
method. The figures of this column correspond to the plus system notation in the second column.
e.g., a “+3” means a high concentration.
The fourth column: SI international unit.

Caution
If the test results are positive, there will be an asterisk “*” before the analyte name as a
prompt.

8
6.3 Display

All test results will be shown on the LCD display after each testing. The first display shows the

first six items, and the next display shows the rest. You can review them by pressing the “ ”
button.
The display format varies with the measuring unit that you can select from the System menu.
The full names of test items are as shown in table-2.
Table-2 Abbreviation of Test Items
No. Full name Abbreviation
1 LEUKOCYTE LEU
2 KETONE BODY KET
3 NITRITE NIT
4 UROBILINOGEN URO
5 BILIRUBIN BIL
6 PROTEIN PRO
7 GLUCOSE GLU
8 SPECIFIC GRAVITY SG
9 BLOOD BLD
10 pH VALUE pH
11 ASCORBIC ACID (VITAMIN C) Vc
12 CREATININE CR
13 CALCIUM Ca
14 MICROALBUMIN MA
15 MICROALBUMIN-TO-CREATININE RATIO ACR (MA/CR)

6.4 Data storage and review

This instrument can store 5000 records for reviewing and printing. The stored data won't be lost
when the power supply is cut off. When the data exceed 5000, the first one will be replaced.
6.5 External printer

The instrument can work with an external printer to print test results.
6.6 Communication with computer

Connect the RS232 interface to the COM1 or COM2 of a computer so that the instrument can
communicate with the computer. See 8.6 for details.
6.7 Connection with a bar code scanner

URIT-50 urine analyzer can connect a bar code scanner with the PS/2 interface. See 8.7 for
details.

9
7. Installation
7.1 Environment requirement

1) Install the instrument on a clean, flat and firm surface, which is free from direct sunlight, strong
magnet and water.
2) To avoid the instrument exposing to moisture to excess or high temperature to excess. Install
the instrument in a room with air-conditioner so that there will be a stable temperature and
humidity for the instrument. To make sure the precise of test results, please keep the environment
condition as the strips request.
3) Keep away from sunlight, oven, radiation, heat source or explosive air. Avoid excessive dust
and vibrate platform like refrigerator.
4) To make sure the ground terminal of power socket is well grounded when power supply is on.
7.2 Installation of printer paper

1) Press the locker of the printer cover and lift the cover.
2) With your index fingers, press the rubber roller toward yourself and take away the roller
upward.
3) Lay down the printer paper. Then, put the rubber roller on the notch and press the roller
downward until it clips to the notch.
4) Pull the paper through the printer cover and replace the cover.

Figure 4

Caution
The thermal printer can just print on the obverse surface. If the paper is installed in
reverse, the printer will print nothing.
Ensure that the paper is dry before installation. If the paper is wet, exchange with a dry roll
to avoid blocking.

7.3 Connection with Power Cord

One end of the power cord should be connected with the DC 12V power adapter and the other
end with a well grounded AC electrical outlet.

Caution
Only use the adapter which is packed with the instrument.

7.4 Connection with External Printer

Turn off the power switch and connect the instrument with external printer via the 25-pin parallel

10
interface or the 9-pin serial interface. These printers may have suitable cable. If there is no cable,
please buy in addition.
7.5 Connection with computer

Turn off the power switch and connect the RS232 interface to the COM1 or COM2 on a computer
with a 9-core cable. See 8.6 for details.

Caution
Equipments and peripherals connected with the instrument should be consistent with
safety criterion.

11
8. Operation
8.1 General information

Operator should operate the instrument by pressing the keys on the instrument. The last line
information shown on the display screen is function prompts for the keys. For convenience, every
function prompt on the screen corresponds with the key below.
If a key is active when pressing it, a short tone will be heard and you will see a change either in
the display or in the instrument operation.
Besides the prompts on the bottom of screen, the high light character in the screen means the
effective choice.

Caution
Untrained operation may cause erroneous test result, and result in wrong diagnosis.
Operator should be trained before using the instrument.

Biological risks
DO NOT touch the urine sample or used strip directly as they may carry potential
infections. Put on protective gloves during operation. If the sample spatters on body, wash
it off with water immediately.

8.2 Operation

8.2.1 Self-checking
Turn the power switch on. The instrument starts with a self-checking. After that, the display
changes to the Ready screen (Figure 5) and is ready for routine test. If the strip holder is not clean
enough, a message of “Dirty, please clean it.” will be shown on the display. In this case, turn the
power switch off. Take the strip holder out and clean it. After installation, turn the power on again.

Caution
There should be no urine strip or any foreign objects on the strip holder, when turning on
the power switch. Otherwise the system will warn that the strip holder is not clean.

URIT-50 2010-08-03

Strip: 11G 04:51:35


Seq Number: 000001 Mode: Once

Figure 5

On the Ready screen, you may see the icon and/or displayed on the upper of the screen:

: Strip detector is closed.

12
: Sensitivity has been adjusted.
8.2.2 Settings
Press the MENU key from the Ready screen to change to the MAIN MENU screen (Figure6):

System Contrast Initialize


DataBank Rs232 Info
Printer Date&Time GreyMeas

Figure 6
8.2.2.1 System
Press the System key from the MAIN MENU to display the System menu (Figure 7):

Mode: Once Speed: Fast


Sensit: Def Units: SI
Language: ENG Detecter: ON

Figure 7
A) Mode:
Continuous and single test modes are available.
Continuous Mode: The instrument performs testing continuously with the prompts for user to
prepare the next sample 30 seconds or 60 seconds in advance.
Single Test Mode: The instrument will stop the movement when a sample testing is completed.
B) Speed:
There are Normal (60 seconds) and Fast (30 seconds) for selection. When the test mode is
continuous, normal means a sample test is 60s; when the mode is once, the speed selection is
unavailable.
C) Sensitivity:
There are two options: Default setting, Manual for selection. The operator is allowed to adjust
every single test item in Adjust option. For positive ratio, "+" denotes raising reflectance ratio and
decreasing positive ratio, "-" denotes decreasing reflectance radio and raising positive ratio. For
linearity, "+" denotes raising reflectance radio of high concentration,"-" denotes decreasing
reflectance radio of high concentration. (The adjustable range in Adjust option is from +10 to -10,
this figure is just a relative value of adjustable range, not a percentage).
Press the following keys to adjust the sensitivity:
[+]: To increase the reflectance rate.
[-]: To reduce the reflectance rate.
[ ]: Switches between Positive and Linear.

[ ]: Displays the next item to be adjusted. Keeping pressing it for 3 seconds will save the
adjustment and exit.

13
Caution

There is only one kind of compensate for every item. 0 means no adjustment. Press [ ] for
3 seconds to save the setting.

Caution
It's recommended that operator should collect enough QC test data before using the
function of sensitivity adjustment, and check the conformance with clinical test after
adjustment.

When there is an icon “ ” displayed on the upper of screen, it means the sensitivity setting is
not default. Then the instrument will prompt whether to resume after self-checking. [Yes] means to
resume, and [No] means to keep the setting.
D) Units
Operator could select SI unit ,CONV unit or English unit.
E) Language:
English is default.
F) Detector:
URIT-50 urine analyzer can automatically detect the presence of strip placed on the strip holder. If
the detector is ON, the instrument will automatically start the testing procedure when a strip is
placed on the strip holder, not necessarily pressing the START key.

Caution

If the detector is OFF, the icon will display on the Ready screen, and you need to
press the START key to begin the testing.

8.2.2.2 Contrast (LCM):


Press “[+]” and “[-]” to regulate the contrast of the Liquid Crystal Display.
8.2.2.3 Initialize:
Press [OK] all test results stored in memory will be cleaned and reset all features to default; press
[DATA], only test results stored in memory will be cleaned.

Caution
All records will be cleared up when initializing the instrument. Please select it carefully.

8.2.2.4 Databank:
All records stored in memory can be recalled from the Databank. Pressing [↓], the screen will
show the previous stored record, pressing [↑] will show the next stored record. For quick scrolling
by an interval of 10 records, keep [↑] or [↓] pressed the 10 previous record or 10 later record is
recalled.
Press [OUTPUT] to display four options for data output. These are:

14
1) [RS232]: Test result will be exported through the serial interface.
2) [PRINT]: Test result will be printed out through the internal thermal printer.
3) [LPT]: Test result will be exported through the parallel interface.
4) [ALL]: Test result will be exported through the serial interface, printer and parallel interface.
When selecting any one of the four options above, the display changes to the corresponding data
screen. Pressing the prompt key (RS232, PRINT, LPT or ALL), the current data will be exported. If
you want to export all data (from current one to the newest one) at one time, keep pressing the
prompt key for 2 seconds until a “Di..” sound is heard. To stop the export, press it again during the
continuous exporting process.
8.2.2.5 RS232 (Serial interface)
1) RS232 OUT: Turns on or off the RS232 interface. OFF means the instrument won't export data
when testing is completed.
2) Format: Data export format, which is suitable to different computer software. At present, only
U50 form is effective.
3) BAR CODE: The bar code scanner option is only available for the single test mode. For this,
the bar code scanner option has to be enabled.
The bar code ID will also be shown on the report of internal printer, RS232 exporting and LPT
exporting, bur not be stored in the instrument memory, i.e., there won't be bar code ID in the
Databank screen.
8.2.2.6 Info (System information)
Manufacturer and instrument information can be found here.
8.2.2.7 Printer
1) Printer: Turn on or off the internal printer.
2) Item Name: English is default for the item name.
3) LPT: Turns on or off the parallel interface.
8.2.2.8 Date&Time: Adjust the date and time.
8.2.2.9 GreyMeas: Only for factory.
8.2.3 SN (Serial Number)
Press the SN key form the Ready screen to display the SN screen. Move the cursor left or right to
the digit that needs to be changed by pressing the [ ] key. Then, press [+] to increase the digit,
and [-] to decrease the digit. After modification, press [OK] to save. The serial number will always
be reset to“000001”each time turning on the instrument.
8.2.4 Strip
Press the STRIP key from the Ready screen to display the Strip Setting screen. There are 10G,
11G and 14G urine strips for selection.
8.2.5 Start
Press the START key from the Ready screen to begin the testing procedure manually.
8.3 Calibration and Quality Control

8.3.1 Calibration
Operator should use URIT check strip to perform calibration for the analyzer frequently. There are
two check strips that packing with the analyzer. One is for daily use, and the other for standby use.
Compare the check strip results with the reference value. If the results are consistent, it means
the analyzer is in order and can be used for routine testing. If not, use the other check strip for
calibration again. If the results are not consistent yet, check whether malfunction occurs and solve

15
it. It is recommended that QC solution testing be run under the following situations (Please
contact the manufacturer to get the QC solution):
 When to use a new bottle of reagent strips
 When to replace an operator
 When to question about the test results

Caution
1. The check strips are used for daily calibration.
2. One is for daily use, the other is backup.
3. Don't dip it in any liquid.
4. Keep it clean.
5. Please compare the inspected results of the check strip with the data labeled on the
bottom. Consistent data indicate normal instrument and inconsistent data indicate some
errors.
6. Please note the data labeled on the bottom is to check the instrument only but not for
clinic reference.

8.3.2 Operation for calibration


Make sure the strip type is 11G and the instrument is ready to test. If the detector is ON, put the
check strip into the strip holder, push the strip inside until a “D” sound is heard, then the
instrument is going to test. After 60s, the test result will be displayed and printed. If the detector is
OFF, press the START key on the ready screen, after the sound of prompt, put the check strip into
the strip holder as the request. After 60s, the test results will be displayed and printed.

Caution
The check strip should not be soaked into urine or water. If the strip holder is dirty, clean it
first to avoid contaminating the check strip.

8.4 Preparation before Testing

1. Make sure that your strips being used are the URIT 10G, 11G or 14G strips. Use of any reagent
strips other than the specified URIT 10G, 11G or 14G will cause erroneous results.
2. Make sure the strip being used matches with the strip type displayed on the Ready screen.
Otherwise, “TROUBLE-8” will be displayed if a non-corresponding reagent strip is used.
3. Make sure the operator knows the position of strip should be put; Make sure the operator
knows the request of testing time.
4. Check the date, time and printer. Change them when needed.
8.5 Routine Test

Biological risks
Make sure to put on protective gloves to prevent exposure to infectious microbes.

1) Connect the adapter to the instrument and the electrical outlet. Push the power switch at “1”

16
position. The instrument is ready for test after self-checking.
2) Prepare the filter paper (for removing excess urine), urine sample and urine reagent strip.
3) Completely immerse the reagent strip into fresh, well-mixed, and uncentrifuged urine sample,
then slowly draw it out along the container wall, using filter paper for excess urine, but not too dry.
Place the top end of reagent strip in the groove of the strip holder, with pads upward. Then, push it
along the groove until the buzzer goes off.
A. Single Testing Mode
If the Detector is ON, the instrument will automatically begin the testing procedure. A message of
“Testing is going on, please wait...”is displayed and the 60s countdown timer starts to work,
decreasing 1 every second. You can input bar code right now if the Bar Code is ON. When the
timer shows“0”, test results is displayed and printed out.
When the strip holder completely moves out, remove the used strip and place a new one, the
instrument will start the next testing automatically.
If the Detector is OFF, press the START key to begin the testing procedure manually. The display
screen shows 65s countdown. When counting down to 60 seconds, the buzzer goes off and the
instrument begins testing.
When the strip holder completely moves out, remove the used strip and place a new one. Press
the START key to perform the next testing.
B. Continuous Testing Mode (Normal Speed)
If the Detector is ON, the instrument will automatically begin the testing procedure. A message of
“Testing is going on, please wait...”is displayed and the 60s countdown timer starts to work,
decreasing 1 every second. When the timer shows“0”, test results is displayed and printed out. At
the same time, the buzzer goes off, prompting to prepare the next strip. When the strip holder
completely moves out, remove the used strip and place a new one, the instrument will start the
next testing automatically. Afterward, the buzzer will go off every 60s to prompt operator to
prepare the next strip.
If the Detector is OFF, press the START key to begin the first testing manually. The rest of
operation is the same as above.
C. Continuous Testing Mode (Fast speed) ---Not Recommended
If the Detector is ON, the instrument will automatically begin the testing procedure. A message of
“Testing is going on please wait...”is displayed and the 60s countdown timer starts to work,
decreasing 1 every second. When the timer shows“30”, the buzzer goes off, prompting to prepare
the second strip. When the timer shows “0”, test result of first strip is displayed and printed out.
At the same time, the buzzer goes off, prompting to prepare the third strip. When the strip holder
completely moves out, remove the first strip and place the second one, the instrument starts next
testing automatically. Afterward, the buzzer will go off every 30s to prompt operator to prepare the
next strip.
If the Detector is OFF, press the START key to begin the first testing manually. The rest of
operation is the same as above.

Caution
1) In the continuous testing mode, the strip should not be replaced until the strip holder
has moved out completely.
2) When the strip holder moves out completely, replace the strip as soon as possible.

17
3) When the Detector is on, if you don't replace the strip in time, the instrument will show
“The strip has not been changed” and stop testing; When the Detector is off, if you don't
replace the strip, the instrument will continue testing the same strip. It will not stop until
there is no strip on the holder.

Caution
Dispose of waste solution, used consumables, used gauze, used cotton swabs and used
gloves according to local regulations.

The position of strip should be put correctly. See figure 8 for details. Practice several times to
handle the operation.

Figure 8
4) Stop a test
During the testing process, pressing the STOP key, the instrument will stop testing and the strip
holder will remove out. If the instrument detects no reagent strip on the holder when testing, it will
also stop testing automatically.
5) Turn off power
Push the power switch at “0” position when all the testing is complete. Clean the strip holder as
directed in Daily Cleaning of Part 2 Maintenance.

Caution
If the instrument is disused for a long time, take the power cord out from the AC electrical
outlet.

8.6 Serial Interface

All the test results stored in memory can be exported to computer or external printer via the serial
interface (RS232). You can turn on or off the interface (see 8.2.2.5 RS232).
A RS232 connector pin assignments
 The second pin RXD: Receive Data
 The third pin TXD: Transmit Data
 The fifth pin GND: Ground
 Other pins are blank.
B Parameter setting

18
 Baud rate: 9600BPS
 Data bit: 8
 Parity bit: None
 Stop bit: 1 bit
C Exporting data form (Appendix)
 Regular result exporting form
 Trouble (error) exporting form
8.7 PS/2 port

The PS/2 port is mainly used for connecting with the bar code scanner to enter identification
number from bar coded label. You can turn on or off the port (see 8.2.2.5 RS232).
A PS/2 port connector pin assignments
 The first pin: DATA
 The third pin: GND
 The fourth pin: Power supply +5
 The fifth pin: CLK
 The second and the sixth pins are blank

B Parameter setting
 Exporting character: LC
 Exporting Finis character: Enter (CR)

Caution
Only when the test mode is single, the bar code can be input.

19
9. Cleaning and maintenance
URIT-50 is a precise instrument; it’s very import to keep maintenance periodically to keep the
instrument work properly. If there’s something wrong with the instrument, please contact URIT.

Biological risks
Make sure to put on protective gloves and clothes to prevent exposure to infectious
microbes.

9.1 Precautions for Maintenance

1) Carefully read through the Operating Manual, as well as the reagent Instruction before
maintenance.
2) The instrument should be installed in the place where no direct sunlight, no swash, no strong
magnet but has power input, otherwise will lead inconvenience.
3) Only URIT urine strip grantee the precise.
4) The power request: Instrument: 12V , The Adapter: 100V-240V~ 50/60Hz.
5) The instrument should be well ground.
6) Keep the instrument clean to guarantee the good performance.
7) Cover the instrument with dust cloth if it is disused for a long time and perform Quality Control
analysis when resuming the instrument.
8) DO NOT dismantle the instrument at will.
9) If there is any trouble when operating, see table 3 for details. If problem persists, please contact
URIT.
9.2 General cleaning

Keep the instrument clean and dust-free. If cleaning is needed, use wet, lint-free fabric. If the
stains are stubborn, use a mild detergent. Any oil, ester, silica gel and lubricant are not advisable
for use on the instrument.
Use only soft fabric to clean the screen. Water, ketene and aromatic solvent are not advised for
use.
9.3 Daily cleaning

Clean the instrument at least once every day. If uncleanness affects test results, clean the
instrument at any time.
If there is too much urine dirt on the strip holder’s groove or flank, the test result might not be
exact. In this case, remove out the holder as the steps shown in 9.4. Clean the strip groove with
water and dry it with soft, lint-free cloth.

Caution
DO NOT wipe across the white block and stick on the top end of the holder. If dust exists,
use a clean fabric to clean the block and the stick.

20
9.4 The dismounting of strip holder

1. Switch off the power supply. Gently pull out the holder toward operator.
2. After cleaning, gently push the holder in the entrance until the end of the holder parallels to the
instrument edge.
3. Switch on the power supply. The instrument will automatically adjust the holder to its proper
position.

Caution
The power should be turned off when dismounting the strip holder. If the operator feels
hard to push, please pull the strip holder out and then readjust the position to push again.
Don’t push by force.

21
10. Storage and Transportation
10.1 Storage Condition

Storage environment: -20℃~55℃;RH≤ 95%.


Atmospheric pressure: 75kPa~106kPa.
10.2 Transportation Condition

The URIT-50 urine analyzer is a precise instrument that must be transported with care, avoiding
moisture, direct sunlight and bump.
Transportation environment: -20℃~55℃;RH≤95%.
Atmospheric pressure: 75kPa~106kPa.

22
Part 2: Service Manual

1. Introduction
URIT-50 urine analyzer is manufactured under strict supervision. This is a kind of precise analyzer,
improper operation or any damaged component may affect the test results or cause damage to
the instrument. The instrument should be repaired only by professionals with certain knowledge of
electronic technique. This section of the manual lists various errors and messages, together with
probable causes and remedies recommended to correct the problems efficiently. The
maintenance instructions in this manual are based on the replacement of parts. Exchangeable
parts see appendix 2.

Caution
There are some high temperature components in the instrument. Beware of scald!

2. Trouble-shooting
No. Symptom Possible Cause
TROUBLE-01 The reference light of LED is too strong.
TROUBLE-02 The red light of LED is too strong.
TROUBLE-03 The green light of LED is too strong. Please contact your local distributor
TROUBLE-04 The reference light of LED is too weak. or URIT.
TROUBLE-05 The red light of LED is too weak.
TROUBLE-06 The green light of LED is too weak.
TROUBLE-07 The position of strip is wrong. See 8.5 for details.
TROUBLE-08 The setting of strip type is not correct. Set the strip type correctly.
TROUBLE-09 There is no strip on the holder. See 8.5 for details.
The instrument is not working well. (The
TROUBLE-10 Check the voltage or repair.
chip of main board is defective.)
The strip holder is dirty or there is some Clean the strip holder or take away
TROUBLE-11
foreign body on it. the foreign body.
Check if the white stick on the strip
TROUBLE-12 Defective mechanism holder is off or the strip holder is
installed.
When to call for assistance:
 If the problem continues after performing the steps described in troubleshooting chart;
 If the accessories are damaged when using;
 If additional assistance is required concerning an instrument problem;
 Main Parts of the Instrument are broken apparently.

Caution
After the instrument scrapped, please according to the regulation of waste electrical and
electronic equipment recycling to deal with.

23
Appendix 1: The meanings of marks that appear on the

instrument or label are as followings

Danger! High voltage Power off

Caution Power on

Biological risks Grounded

In Vitro Diagnostic medical device Serial number

Electrical and electronic products


Manufacturer
recycling symbol
Protect from heat and radioactive
No rolling
sources

Up Fragile, handle with care

Keep dry Stacking limit

Consult instructions for use Temperature limitation

Serial Port Parallel Port

Authorized representative in
Direct current
European Community

Polarity of d.c. power connector Alternating current

Environment-Friendly Use Period Date of manufacture

Use by date PS/2 To connect a bar code scanner

24
Appendix 2: Exchangeable Parts
SN ITEM REMARK
1 LCD With Socket
2 Thermal Printer
3 Power switch
4 Adapter
5 Step Motor With Plug
6 Power Cord
7 CR2032 Battery
8 Key Pad
9 Strip Holder
10 Optical-electric sensor system
11 Connection Line With Plug
12 Main Board
13 Print Drive Board

25
Appendix 3: Format of test results
14G

Remark:
1) Digits denote serial numbers of ASCII code which are exported via the serial port.
2) The second code is internal communication code, which can be taken as a space code.
3) The black squares “■”denote test results. The blank spaces denote space codes.
4) Other urine reagent strips types of data output format consistent with the 14G output format,
but the project sequence is different, the order of the project on the order of the test paper tube.

26

You might also like