Professional Documents
Culture Documents
4d Recon
4d Recon
7, 2021
Abstract—Four-dimensional (4D) ultrasound reconstruction can Traditional 2D ultrasound slice cannot completely track this
greatly extend the spatial and temporal range of two-dimensional time-varying motion of a 3D organ. Hence, time-dependent
(2D) ultrasound in clinical practice. However, uneven breaths may structural reconstruction, also known as 4D reconstruction, is
yield a considerable motion artifact in the reconstructed time
sequences of volume ultrasound. In this paper, a system with sparse significant for capturing respiration-induced liver motion.
respiratory signal matching is proposed to realize accurate 4D ul- The 4D reconstruction of ultrasound image slices can be
trasound reconstruction by effectively estimating uneven breaths. regarded as 3D reconstruction at different respiratory phases. A
A slippery platform is built to automatically control the ultrasonic large number of ultrasound slices acquired in different positions
probe for acquiring 2D ultrasound slice sequences in different
and at different phases are assembled as multiple 3D volumes
abdominal areas. Based on these acquired sequences, the signals
of the respiratory phases are first extracted. Second, the sparse according to their spatial and temporal correlations. For sim-
matched signal pairs are determined through the signal distribu- plification, test subjects can be asked to hold the breath during
tion and the corresponding image connectivity. After matching is ultrasound acquisition. In the method [4], the 2D ultrasound
conducted, signals in different positions are aligned in a common images are obtained under breath holding at multiple phases in
space wherein the outliers of the signals caused by the uneven
a cycle. By holding the breath, the temporal correlation between
breaths are excluded. Third, images with the same signal value
in different positions are collected as the 3D image at the phase the slices acquired in different positions is specific, and the
corresponding to the signal, and the 3D images at different phases influence of the motion artifact is reduced in the reconstructed
are employed to reconstruct the final 4D image. The accuracies images. Nevertheless, images of a large abdominal area cannot
of the reconstructed 4D image are evaluated and compared with be captured in a short time during breath holding. Moreover,
those of five existing methods. Experimental results demonstrate
previous experiments [4] indicated that the number of phases is
that the spatial and temporal continuities of the 4D ultrasound
image reconstructed by our method outperform those obtained by limited by breath holding. These problems may result in a few
other methods. phases and a small field covered by the reconstructed 4D image.
Different from the breath holding-based reconstruction, tem-
Index Terms—4D reconstruction system, ultrasound image,
respiratory signal, sparse matching. poral correlation is computed in reconstruction with slices ac-
quired under a free breath. The temporal correlation is indicated
by the respiratory signals of slices. Respiratory signal can be
I. INTRODUCTION viewed as a one-dimensional feature to indicate the state of the
LTRASOUND imaging is widely used in the diagnosis respiratory motion and phase in a cycle. The actual respiratory
U and treatment of liver diseases because of its high real-time
quality, accessibility and safety [1], [2]. Affected by respiration,
phase of a slice is selected on the basis of signal, and the slices at
the same phase constitute a structural 3D image. The 3D images
the liver periodically moves along the head-foot direction [3]. at the multiple phases constitute the final 4D image. Therefore,
respiratory signal extraction and respiratory phase selection of
Manuscript received November 3, 2020; accepted January 19, 2021. Date the ultrasound slices are the main steps in 4D reconstruction.
of publication January 26, 2021; date of current version February 25, 2021. A common method of extracting respiratory signals involves
This work was supported in part by the National Key R&D Program of China the use of external devices, such as spirometers, optical and
under Grant 2019YFC0119300, in part by the National Science and Technology
Major Project of China under Grant 2018ZX10723-204-008, and in part by electromagnetic tracking devices [5]–[11]. The outputs of these
the National Science Foundation Program of China under Grants 61971040, external devices indicate different physical meanings. If the
81627803, 81871374, and 61771056. (Corresponding author: Danni Ai; Jian output is high-dimensional data for each slice, then principal
Yang.)
Tianyu Fu, Jingshu Li, Jiaju Zhang, Danni Ai, Jingfan Fan, and Jian Yang component analysis (PCA) is applied to obtain a value as the
are with the Beijing Engineering Research Center of Mixed Reality and Ad- respiratory signal. Given that additional equipment is needed,
vanced Display, School of Optics and Electronics, Beijing Institute of Technol- these extraction methods are not widely applicable. To avoid this
ogy, Beijing 100081, China (e-mail: fty0718@163.com; jshul2020@163.com;
1205766491@qq.com; danni@bit.edu.cn; fjf@bit.edu.cn; jyang@bit.edu.cn). problem, the feature in slice is extracted as a respiratory signal
Hong Song is with the School of Computer Science, Beijing Institute of [12]–[17]. The position of the diaphragm is a common feature.
Technology, Beijing 100081, China (e-mail: anniesun@bit.edu.cn). Hwang et al. [15] used an adaptive threshold and polynomial
Ping Liang is with the Department of Interventional Ultrasound, Chinese PLA
General Hospital, Beijing 100039, China (e-mail: liangping301@126.com). fitting to segment the diaphragm in a slice sequence. In a fixed
Digital Object Identifier 10.1109/TCI.2021.3054527 area, the position of the diaphragm is detected and used as
2333-9403 © 2021 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 235
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
236 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
TABLE I
PLATFORM PARAMETERS
TABLE II
IMAGING PARAMETERS
Fig. 3 shows the flowchart of the proposed reconstruction where Ii,m and Ii,n are the ultrasound slices acquired in the
method, which contains three parts: signal extraction and prepro- position i; si,m and si,n are the points in the low dimensional
cessing, sparse matching and phase-independent reconstruction. space; · indicates the geodesic distance. In the low dimen-
The inputs are the 2D ultrasound slice sequences in the N sional space, each ultrasound slice could be expressed as a
abdominal positions acquired by the constructed system. In low dimensional feature. If the dimension equals to one, si,m
the first part, the dimensional reduction is used to extract the can be regarded as the respiratory signal of the corresponding
respiratory signals of the 2D ultrasound slices in each posi- ultrasound slice Ii,m .
tion. The directions of the extracted signals are revised among The original signals obtained by Isomap will be directionally
the sequences for consistency. And the signals are smoothed revised, smoothed and segmented for further sparse matching.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 237
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
238 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
images at the same phase acquired in the different positions ⎨Λ1 = I1,c |c ∈ [1, C1 ]
will be matched for signal sequence alignment. The scales and Λ50 = I 50 |c ∈ [1, C1 ]
, (9)
⎩ 190 1,c 90
translations of the signal sequences in the adjacent positions Λ1 = I1,c |c ∈ [1, C1 ]
will be normalized and compensated through the matching.
where C1 is the number of completely half cycle in S1 . In these
Moreover, only a few signals in a sequence are selected as the
sets, the corresponding signals of images are grouped as three
matching points to improve the efficiency and accuracy of the
signal sets as follows:
alignment. ⎧ 10 10
In the first position, the images with the signal value closest to ⎨Ψ1 = s1,c |c ∈ [1, C1 ]
10%, 50% and 90% of the range are selected as I110 , I150 and I190 , Ψ50 = s50 |c ∈ [1, C1 ]
. (10)
⎩ 190 1,c
respectively. These images are used as the initial templates for Ψ1 = s901,c |c ∈ [1, C 1 ]
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 239
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
240 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
In the first position, the image with the closest signal value to TABLE III
MODULES USED IN THE IMPLEMENTATION OF COMPARATIVE METHODS
the average of the chosen segment is selected as the first slice
in the reconstructed image. In the other positions, the image
connectivity and signal similarity are considered in the selection
to improve the accuracy of reconstruction. In the proposed
method, L2 distance is used to evaluate the connectivity between
two ultrasound image slices. If L2 is small, the connectivity
is high. From the chosen segment, the image with the highest
connectivity to that in the previous position is selected in the
position i as the ith slice Iip in the reconstructed image. The
process of an inspiratory phase can be formulated as follows:
p
Iip = arg min Ii−1 − Ii,m 2 subject to si,m ∈ Dp+ , (13)
where Ii,m is an image with the signal value si,m belonging respectively. If the L2 between the adjacent slices is large, the
the phase Dp+ and m ∈ [1, M ]. After the reconstruction at each selected slices to be reconstructed may be not come from the
respiratory phase, the 4D ultrasound image is obtained. same respiratory phase, and then the boundary of the tissue in
the reconstructed image is not smooth enough. The respiratory
III. EXPERIMENT line is the signal sequence of all 3D images in the in the
reconstruction result obtained by Isomap. If the respiratory line
In the experiment, the proposed method is evaluated on the is similar to a Gaussian curve, the temporal continuity of the 4D
2D ultrasound images which acquired from six test subjects reconstruction result is high.
using the built system. The approval No. of ethics committee
is S2020-464-01 provided by Ethics Committee of Chinese
B. Implementations of Comparative Methods
PLA General Hospital. The assessment criteria is shown in the
section A. Based on the same ultrasound sequences acquired by There are mainly four modules involved in the implementa-
the slippery platform, the reconstruction results obtained by the tions of the comparative methods including respiratory signal
proposed method are compared with those of the methods in extraction, dense matching of the signal, similarity constraint
Georg et al. [18], Li et al. [23], Dikaios et al. [22], Low et al. and separate reconstruction. In the module named image sim-
[5] and Wachinger et al. [19]. The implementations of these ilarity constraint, image similarity is used as a constraint for
methods are detailed in the section B. The evaluation contains reconstruction. Meanwhile, the respiratory signal extraction can
three parts corresponding to the sections C, D and E. In the be divided into external device utilization and manifold learning.
section C, the signals obtained by the external devices is used The different modules are used to implement the comparative
as the ground truth, and compared with those obtained by the methods as shown in Table III. In Table III, “” and “×” mean
proposed method. Meanwhile, the sparse matching results of that the module is used and not used in the method, respectively.
the respiratory signal sequences from the adjacent positions are And “–” means that the module is not involved in the method.
assessed. The spatial and temporal continuities of the reconstruc- Therefore, the comparisons of Dikaios et al. [22] and Low
tion result are evaluated in the sections D and E, respectively. et al. [5] are used to test the advantages of the proposed signal
In the experiment, the used CPU and RAM are i7-9700K and extraction and processing. The comparisons of Li et al. [23],
16 G, respectively. The proposed method was implemented in Georg et al. [18] and Wachinger et al. [19] are used to examine
Python and C ++ programming languages. the advantage of the proposed sparse matching.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 241
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
242 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
Fig. 11. Selection ratio at each phase. (a)-(e) are the ratios of six test subjects. The first and second rows are the ratios in inhalation and exhalation processes,
respectively.
subjects. The left and right columns show the sparsely and fully compared in Fig. 11. This figure shows the selection ratio at each
matched sequences. In each sub-figure of Fig. 10, the vertical and phase obtained by different methods. Figs. 11 (a)–(e) present
horizontal axes represent the position and the respiratory signal the ratios of six test subjects. The first and second rows are the
value, respectively. The points in each position of the vertical axe ratios in the inhalation and exhalation processes, respectively.
are the horizontally projected points of the matched sequence in Inhalation and exhalation processes are not separated in the
the corresponding position. Meanwhile, the projected points of method of Georg et al. [18], so its selection ratios are the same
the matched sequence in a position are colored accordingly to in the sub-figures of the first and second rows. The selection
the ratio between the signal range of the sequence and the mean ratio obtained by the proposed paper is generally high at each
range. The closer to 1 the ratio is, the closer to green the color of phase. Due to the significant shift identified between the matched
the points will be. Otherwise, the color is closer to red or blue. sequences, the selection rate obtained by Georg et al. [18] is
Therefore, if matching is accurate, then the mapped points in very low. A higher selection ratio means that more slices in the
the different positions are distributed nearly in the range of the reconstructed image are actually acquired from the ultrasonic
template sequence. The ranges of all densely matched sequences apparatus, and the tissue structures shown in the reconstructed
are quite different, which is indicated by numerous red or blue image is closer to the truth.
points as shown in the first column of Fig. 10. Meanwhile, Figs. 12 and 13 show the sagittal planes of the 4D ultrasound
accumulative errors in the matching between the sequences in images of two test subjects reconstructed using different meth-
the adjacent positions result in a significant shift between the ods. In these figures, the left, middle and right parts show the
matched sequences. By contrast, the colors of the sequences reconstructed 3D ultrasound images at the start, middle and
matched by the proposed sparse matching is closer to green, and end phases of the processes. In each part, the left and right
the mapped points cover the same signal value range. However, columns are the images in inhalation and exhalation processes,
the ranges of the matched sequences in the few positions are respectively. The first to sixth lines show the results obtained by
away from the mean range because of the uneven breath. Georg et al. [18], Li et al. [23], Dikaios et al. [22], Low et al.
[5], Wachinger et al. [19] and the proposed method. The yellow
arrows point the liver vessels in the reconstructed images. As
D. Evaluation of the Spatial Continuity shown in these figures, the boundaries of the liver vessels in
Different from Li et al. [23], the ultrasound slices are se- the reconstructed image obtained by the proposed method are
lected in the positions according to their signal values at each smoother and more continuous than those obtained by the other
respiratory phase in the reconstruction of the other compara- methods. It is worth noting that the reconstructed images in
tive methods and the proposed method. However, the different Figs. 12 and 13 are the results of interpolation, which is used to
selections in various methods may result in the lack of slices generate a slice for the position without a selected slice. Because
in some positions. For the position without the selected slice, of the too low selection ratio from Georg et al. [23], the 4D
interpolation is used to generate a slice and obtain a complete 3D ultrasound image can not be completely reconstructed even after
ultrasound image at the phase. The selection ratio, which is the the interpolation. Therefore, the black segment is identified in
quotient result of the numbers of the positions with the selected the image reconstructed by Georg et al. [23].
slices and all positions, is utilized to evaluate the methods. A To quantitatively evaluate the spatial continuity of the recon-
higher selection ratio indicates that more slices are selected and struction results obtained by the different methods, the values
few slices are interpolated to reconstruct. Meanwhile, because of the L2 distances between the selected adjacent slices in the
the spatial continuity constraint changes the selected slices rather reconstructed 3D image at different phases are computed and
than the number of them, only the selection ratios obtained by shown in Fig. 14. The slices generated via interpolation are
Georg et al. [18], Dikaios et al. [22] and the proposed method are excluded in the computation of the spatial continuity. Fig. 14
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 243
Fig. 12. Reconstruction result on sagittal plane of subject 1. The left, middle and right parts show the reconstructed 3D ultrasound images at the start, middle
and end phases of the processes. In each part, the left and right are the images in the inhalation and exhalation processes. The yellow arrows point the liver vessels
in the enlarged view of the reconstructed images.
TABLE IV
L2 OF THE RECONSTRUCTED 4D ULTRASOUND IMAGE IN INHALATION
TABLE V
L2 OF THE RECONSTRUCTED 4D ULTRASOUND IMAGE IN EXHALATION
shows the L2 in the inhalation and exhalation of the first test means and standard deviations of the L2 distances of all phases
subject, respectively. In Fig. 14, the boxplots colored in red, in inhalation and exhalation processes for each test subject,
green, blue, purple, black and yellow are the L2 obtained by respectively. All the L2 for each test subject are normalized to
Li et al. [23], Georg et al. [18], Dikaios et al. [22], Low et al. make the results intuitively comparable. Because inhalation and
[5], Wachinger et al. [19] and the proposed method. A small L2 exhalation processes are not separated in the method of Georg
means a high continuity of the reconstructed image. With the et al. [18] and Li et al. [23], their means and standard deviations
continuity constraint, the proposed method achieves a smaller of L2 are the same as those in the two tables. The results
L2 than that of the other methods. Tables IV and V show the show that the means and standard deviations of L2 obtained
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
244 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
Fig. 13. Reconstruction result on sagittal plane of subject 4. The left, middle and right parts show the reconstructed 3D ultrasound images at the start, middle
and end phases of the processes. In each part, the left and right are the images in inhalation and exhalation processes. The yellow arrows point the liver vessels in
the enlarged view of the reconstructed images.
Fig. 14. Resultant boxplot of L2 between the selected adjacent slices in the reconstructed ultrasound image of test subject 1. The left and right are the results of
inhalation and exhalation, respectively.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 245
TABLE VI
NORMALIZED SIGNAL VALUES OF THE 3D IMAGES AT THE PHASES OF THE
START-IN, END-IN, START-EX AND END-EX IN THE RECONSTRUCTION RESULTS
E. Evaluation of the Temporal Continuity Table VI shows the normalized signal values of the start-
The 4D reconstruction result covers the entire respiratory inspiratory (start-IN), end-inspiratory (end-IN), start-expiratory
cycle. If the temporal continuity is high, the respiratory line (start-EX), end-expiratory (end-Ex) images in the reconstruction
joining the signals of all 3D images in the in the reconstruction results. For the proposed method, the obtained signal values
result should be similar to the Gaussian curve. Meanwhile, the of start-IN and end-EX are approximately equal and close to
inspiratory and expiratory lines should be similar to the each −1. The signal values of start-EX and end-IN obtained by the
other, which indicates by the symmetry of the respiratory line. proposed method are approximately equal and close to 1. The
We use the Isomap to extract the respiratory signal of each 3D results of Georg et al. [18] and Li et al. [23] are not collected
image in the reconstruction result. The respiratory lines of the in Table VI because of their undistinguished reconstruction for
different methods are shown in Fig. 16. In this figure, (a-f) show inhalation and exhalation. In Fig. 17, the ratio between the
the respiratory lines using the data from the six test subjects. area under the inspiratory curve (AUIC) and the area under
The signal value range of each line is normalized to [1[,-1]. the expiratory curve (AUEC) to evaluate the symmetry of the
And the range is divided into 15 segments. Compared with the obtained respiratory line. Inhalation and exhalation processes
other methods, the respiratory lines obtained by the proposed are not separated in the method of Georg et al. [18] and Li et al.
method are more similar to a Gaussian curve without the zigzag. [23], so their ratios of them are equal to 1. The ratio achieved
Meanwhile, the values of 1 and −1 are the end-inspiratory (or by the proposed method is closer to 1 than that obtained by the
start-expiratory) and end-expiratory (or start-inspiratory) signal methods of Dikaios et al. [22], Low et al. [5] and Wachinger
values in the ideal respiratory line, respectively. et al. [19].
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
246 IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, VOL. 7, 2021
TABLE VIII
RECONSTRUCTION TIME (MINUTE)
V. CONCLUSION
4D ultrasound image reconstruction is significant for the esti-
mation of respiration-induced liver motion. A sparse matching-
based method is proposed in this paper to reconstruct a 4D
wide-field ultrasound image. The original ultrasound slice se-
quences from the different abdominal positions of a subject are
automatically acquired through a system including a slippery
platform. The manifold learning is used to directly extract
the respiratory signal from each ultrasound slice without any
IV. DISCUSSION additional markers. The directions of the extracted signals are
firstly revised and then divided into inhalation and exhalation.
A. Acquisition Time The signal sequence in different positions are aligned by sparse
The acquisition time of the six test subjects is shown in matching with the three selected pairs. After the matching, the
Table VII. The gender of each test subject is presented in the signals with the same values from different positions belong
first column of Table VII. The acquisition time is only related to to the same respiratory phase, and the corresponding ultra-
the number of acquisition positions. Therefore, the acquisition sound slices could be assembled for the phase-independent
time can be adjusted for the different subjects. For the fat and reconstruction. The assembled images at different phases are
males, the number of the acquisition position should be large grouped as the 4D ultrasound image. In the experiment, the
to cover the large liver and larger respiratory amplitude, and the ultrasound slice sequences are acquired from six subjects to test
corresponding acquisition time is long. For the thin and females, the proposed method, and the results are compared with those
the number of the acquisition position is small according to the of five existing methods. The high correlation value between the
small liver and respiratory amplitude, and the corresponding extracted signals and the signals provided by markers indicate
acquisition time is short. that the proposed method can accurately extract the respiratory
signals without the additional markers. In the 4D ultrasound
image reconstructed by the proposed method, the small L2 value
B. Reconstruction Time between the adjacent slices indicates that a higher connectivity
The time consumption of the three steps in the proposed correlation than that obtained by other comparative methods.
method is computed in Table VIII. For the test subject with And the respiratory line extracted from the reconstruction result
a large position number, the time consumption of the three steps obtained by the proposed is more similar to a Gaussian curve
is long. Meanwhile, the time consumption of sparse matching than that obtain by other comparative methods. Therefore, the
in the second step is compared with that of dense matching. In proposed method can reconstruct a 4D ultrasound image with
contrast to dense matching, a few signals are used to align the high spatial and temporal continuities.
adjacent sequences in sparse matching, and the efficiency of the
proposed method greatly improves.
REFERENCES
[1] W. Cong et al., “Global patch matching (GPM) for freehand 3D ultrasound
C. Entire Time Optimization reconstruction,” Biomed. Eng. Online, vol. 16, no. 1, pp. 124–149, 2017.
[2] J. Dong et al., “Multiresolution cube propagation for 3-D ultrasound image
The signal sequence is extracted and preprocessed indepen- reconstruction,” IEEE Trans. Comput. Imag., vol. 5, no. 2, pp. 251–261,
dently in each position, and only the adjacent sequences are Jun. 2019.
sparsely matched in the proposed method. Therefore, acquisition [3] D. Ai et al., “Nonrigid registration for tracking incompressible soft tissues
with sliding motion,” Med. Phys., vol. 46, no. 11, pp. 4923–4939, 2019.
and reconstruction can be performed in parallel after the images [4] J. M. Blackall, G. P. Penney, A. P. King, and D. J. Hawkes “Alignment
are acquired in the first position. Table IX shows the comparison of sparse freehand 3-D ultrasound with preoperative images of the liver
between the parallel and serial processing time. The mean reduc- using models of respiratory motion and deformation,” IEEE Trans. Med.
Imag., vol. 24, no. 11, pp. 1405–1416, Nov. 2005.
tion rate reaches 27.965% through the parallel processing, and [5] D. A. Low et al., “A method for the reconstruction of four-dimensional
the entire time of obtaining the 4D ultrasound image is shorter synchronized CT scans acquired during free breathing,” Med. Phys.,
than 17 minutes. vol. 30, no. 6, pp. 1254–1263, 2003.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.
FU et al.: FOUR-DIMENSIONAL WIDE-FIELD ULTRASOUND RECONSTRUCTION SYSTEM WITH SPARSE RESPIRATORY SIGNAL MATCHING 247
[6] J. Ehrhardt et al., “An optical flow based method for improved recon- Jingshu Li is currently working toward the Ph.D.
struction of 4D CT data sets acquired during free breathing,” Med. Phys., degree with the School of Optics and Photonics,
vol. 34, no. 2, pp. 711–721, 2007. Beijing Institute of Technology, Beijing, China. Her
[7] C. Gianoli et al., “A multiple points method for 4D CT image sorting,” current research interest includes ultrasound image
Med. Phys., vol. 38, no. 2, pp. 656–667, 2011. reconstruction.
[8] F. Ernst et al., “Correlation between external and internal respiratory
motion: A validation study,” Int. J. Comput. Assist. Radiol. Surg., vol. 7,
no. 3, pp. 483–492, 2012.
[9] T. H. Wagner et al., “Optical tracking technology in stereotactic radiation
therapy,” Med. Dosimetry, vol. 32, no. 2, pp. 111–120, 2007.
[10] M. Yang et al., “Ultrasound fusion image error correction using subject-
specific liver motion model and automatic image registration,” Comput. Jiaju Zhang is currently working toward the Ph.D.
Biol. Med., vol. 79, pp. 99–109, 2016. degree with the School of Optics and Photonics,
[11] J. R. Mcclelland et al., “4D motion models over the respiratory cycle for use Beijing Institute of Technology, Beijing, China. His
in lung cancer radiotherapy planning,” Proc. SPIE, vol. 5744, pp. 173–184, current research interests include reconstruction and
2005. registration for ultrasound images.
[12] J. Wölfelschneider et al., “Examination of a deformable motion model for
respiratory movements and 4D dose calculations using different driving
surrogates,” Med. Phys., vol. 44, no. 6, pp. 2066–2076, 2017.
[13] D. Zhou et al., “A feasibility study of intrafractional tumor motion esti-
mation based on 4D-CBCT using diaphragm as surrogate,” J. Appl. Clin.
Med. Phys., vol. 19, no. 5, pp. 525–531, 2018.
[14] A. Giger et al., “Ultrasound-driven 4D MRI,” Phys. Med. Biol., vol. 63, Danni Ai received the B.E. and M.E. degrees from
no. 14, pp. 145–155, 2018. Xi’an Jiaotong University, China, in 2005 and 2008,
[15] Y. Hwang et al., “Ultrasound image-based respiratory motion tracking,” respectively, and the D.E. degree from Ritsumeikan
Med. Imag. 2012: Ultrason. Imag. Tomogr. Ther., 2012, pp. 83200. University, Japan, in 2011. She joined the Beijing
[16] M. Von Siebenthal et al., “4D MR imaging of respiratory organ motion Institute of Technology as a Postdoctoral, in 2013.
and its variability,” Phys. Med. Biol., vol. 52, no. 6, pp. 1547–1564, 2007. She is currently an Associate Professor with the
[17] J. Wu et al., “Fast and robust extraction of surrogate respiratory signal from School of Optics and Photonics, Beijing Institute of
intra-operative liver ultrasound images,” Int. J. Comput. Assist. Radiol. Technology.
Surg., vol. 8, no. 6, pp. 1027–1035, 2013. Her research interests focus on medical image pro-
[18] M. Georg et al., “Manifold learning for 4D CT reconstruction of the cessing and augmented reality.
lung,” in Proc. IEEE Comput. Soc. Conf. Comput. Vis. Pattern Recognit.
Workshops, 2008, pp. 1–8.
[19] C. Wachinger et al., “Manifold learning for image-based breathing gating
in ultrasound and MRI,” Med. Image Anal., vol. 16, no. 4, pp. 806–818,
Jingfan Fan received the Ph.D. degree in optical
2012.
engineering from the Beijing Institute of Technology,
[20] C. Wachinger, M. Yigitsoy, and N. Navab “Manifold learning for image-
China, in 2016, and worked as a Postdoctoral Re-
based breathing gating with application to 4D ultrasound,” in Proc. Int.
searcher with University of North Carolina at Chapel
Conf. Med. Image Comput. Comput.-Assist. Intervention, 2010, pp. 26–33.
Hill from 2017 to 2019. He is currently an Associate
[21] J. Wu et al., “A manifold learning method to detect respiratory signal from Professor with the School of Optics and Photonics,
liver ultrasound images,” Computerized Med. Imag. Graph., vol. 40, 2015,
Beijing Institute of Technology. His research interests
pp. 194–204.
focus on medical image processing, computer vision,
[22] N. Dikaios et al., “MRI-based motion correction of thoracic PET: Initial
and augmented reality.
comparison of acquisition protocols and correction strategies suitable for Hong Song received the Ph.D. degree in computer
simultaneous PET/MRI systems,” Eur. Radiol., vol. 22, no. 2, pp. 439–446,
science from the Beijing Institute of Technology,
2012.
China, in 2004. She is currently a Professor with the School of Computer Science,
[23] C. Li et al., “Subject-specific and respiration-corrected 4D liver model
Beijing Institute of Technology. Her research interests focus on medical image
from real-time ultrasound image sequences,” Comput. Methods Biomech. processing, augmented reality, and computer vision.
Biomed. Eng.: Imag. Visual., vol. 6, no. 1, pp. 7–16, 2016.
[24] J. B. Tenenbaum, V. De Silva, and J C. Langford “A global geometric
framework for nonlinear dimensionality reduction,” Science, vol. 290,
no. 5500, pp. 2319–2323, 2000.
[25] P. Fischer, T. Pohl, and J. Hornegger “Real-time respiratory signal extrac- Ping Liang is currently the Chairman of Depart-
tion from X-ray sequences using incremental manifold learning,” in Proc. ment of Interventional Ultrasound, Chinese PLA
IEEE 11th Int. Symp. Biomed. Imag., vol. 29, pp. 915–918, 2014. General Hospital. Her research interests focus on
[26] S. Carroll M, C. Viemari J, and M. Ramirez J. “Patterns of inspiratory microwave ablation and radiofrequency ablation
phase-dependent activity in the in vitro respiratory network,” J. Neuro- treatment of solid tumors in multiple organs
physiol., vol. 109, no. 2, pp. 285–295, 2013. (liver, kidney, adrenal gland, spleen, thyroid, breast,
[27] A. Sarwal, F. O. Walker, and M S. Cartwright “Neuromuscular ultra- metastatic lymph nodes, etc.).
sound for evaluation of the diaphragm,” Muscle Nerve, vol. 47, no. 3,
pp. 319–329, 2013.
Authorized licensed use limited to: San Francisco State Univ. Downloaded on June 22,2021 at 18:43:59 UTC from IEEE Xplore. Restrictions apply.