FK SERIES FK-202F TRANSDUCER
TRANSDUCER

SPECIFICATIONS
Page 2 of 2

CONFIGURATION

Sensor Connector Extension Cable


Driver

System Cable Length :


FK-202F1 : 5m
FK-202F2 : 9m

• Dimension of target [recommended for rotational speed measurement]


D BA D A D B
C C
Dimension of A≥6
C
Target B≥7
[recommended] C ≥ 2.5
(mm) D ≥16
Set gap
[recommended] 1.0 to 1.5
(mm)

6H14-042 Rev.9
Issued : Feb. 2015
Revised : Apr. 2019
CV SERIES CV-86 PIEZOELECTRIC VELOCITY TRANSDUCER
TRANSDUCER

SPECIFICATIONS

Model Code

CV-861 /EX CV-862 /EX


Cable Type Intrinsically Safety Cable Type Intrinsically Safety
1 Connector Type 1 TIIS (Ex ia A T4 X) With integral cable type *1 1 TIIS (Ex ia A T4 X)
2
5 ATEX (Ex ia A T4 Ga) (Coax. cable 3m standard) 7 NEPSI (Ex ia A T4 Ga)
7 NEPSI (Ex ia A T4 Ga)
8 KTL (Ex ia A T4)

Standard type CW- 0 5 0 E-EF- 0 0- 0


Note) *1 Cable length 3m (end cut off),
Drip-proof type CW- F-FF- 0- CW cable not required.

*2 M3.5 crimping terminals at the


Cable length Lug type*2 Armor
cable ends are standard.
050 5m 0 Cut off 0 Without
100 10m R Ring lug 1 With*3
*3 With armor : Cable length 50m
150 15m Y Spade lug

SPECIFICATIONS
Sensitivity 3.94mV/mm/s (100mV/in/s REF.)pk ± 5% at 100Hz, 25°C (77°F REF.)
Max. Velocity 1,270mm/s (50in/s REF.)pk
2
Vibration Limit 2,450m/s (250g REF.)pk
2
24,500m/s (2,500g REF.)pk (non-Intrinsically Safe)
Shock Limit 2
23,520m/s (2,400g REF.)pk (Intrinsically Safe)
Max. Shock Energy 4J (Intrinsically safe)
Linearity ± 1% of F.S.
Natural Frequency 15kHz
Frequency Response 2.5 to 3,500Hz ± 10%, 2 to 7,000Hz ± 3dB
Transverse Sensitivity Max.5%
Output Impedance 200(typical)
Grounding Case isolated, internally shielded
Temperature Response ± 10% (around the operating temperature range)
Power Supply 18 to 30VDC, 2 to 10mADC (constant current)
Operating Temperature Range -50 to +120°C (-58 to +248°F REF.)
Range of Temperature at Explosion proof EX1,7 : -20 to +60C (Transducer, Cable)
construction EX5,8 : -50 to +120C (Transducer, Cable)
Relative Humidity 100%RH
Case Material Stainless Steel
Sealing Hermetic
Protection Rating IP67(CV-861 & CW-□□□F-FF, CV-862)
Weight Approx.145g(CV-861) Approx.250g(CV-862(including cable))
Accessories Supplied M6 mounting stud
Output Connector*4 MIL-C-5015 2-pin
Matching Connector*4 D/MS3106A10SL-4S
CV-861 : Twisted pair shielded cable
Cabling
CV-862 : Coax. cable (intergral cabling type)
2 2
Recommended Cable Shielded Specifications AWG No.20 to No.16 gage (0.5mm to 1.25mm ) Twisted pair shielded cable
*4 CV-861 TYPE SPECIFICATIONS
【NOTICE】
The instructions manual contains important information such as conditions necessary for safe handling of the system.
Such information and conditions are important and indispensable for ensuring safety.
Therefore, be sure to read the instructions manual thoroughly before handling the system.
6H14-062 Rev.5
Issued : Feb. 2015
Revised : Oct. 2019
0 5 0 0
VM-25 VM-25 CONDITION MONITORING SYSTEM
MONITORING SYSTEM

SPECIFICATIONS Page 1 of 5

Model Code

Vibration measurement function


Recorder Output
[VM-25F01] Relay Output
(Isolated) Model Code
Recorder Output [VM-25F21]
Input Channel Monitor Output [VM-25F26]
(Non-isolated)
2ch 2ch 2ch 0ch 0ch VM-25M00-151-01-00-00-00-00
2ch 2ch 2ch 0ch 4ch VM-25M00-151-01-00-21-00-00
2ch 2ch 2ch 4ch 0ch VM-25M00-151-01-00-26-00-00
2ch 2ch 2ch 4ch 4ch VM-25M00-151-01-00-26-21-00
4ch 4ch 4ch 0ch 0ch VM-25M00-151-01-01-00-00-00
4ch 4ch 4ch 0ch 4ch VM-25M00-151-01-01-21-00-00
4ch 4ch 4ch 0ch 8ch VM-25M00-151-01-01-21-21-00
4ch 4ch 4ch 4ch 0ch VM-25M00-151-01-01-26-00-00
4ch 4ch 4ch 4ch 4ch VM-25M00-151-01-01-26-21-00
4ch 4ch 4ch 4ch 8ch VM-25M00-151-01-01-26-21-21
6ch 6ch 6ch 0ch 0ch VM-25M00-151-01-01-01-00-00
6ch 6ch 6ch 0ch 4ch VM-25M00-151-01-01-01-21-00
6ch 6ch 6ch 8ch 0ch VM-25M00-151-01-01-01-26-26
8ch 8ch 8ch 0ch 0ch VM-25M00-151-01-01-01-01-00
※ Each combination includes standard housing[VM-25F61], DC 24V power supply function[VM-25F55], Modbus/TCP communication function[VM-25F41] and
display function[VM-25F71] as standard.

Specification

GENERAL MONITOR SPECFICATION Operation Contact Input


Display : Display measurement value and alarm set value on Alarm reset(RES.) : Reset the SELF-HOLD Alarm.
7-segment LED. Sequence (SEQ.) : Prevent the alarm output during machine startup.
Vibration measurement function When sequence circuit is in progress, SEQ. lamp on
: Measure vibration of displacement or velocity, front panel is lit.
acceleration. (The number of input channels is Contact type : Dry contact
selected from Model Code.)
Isolated recorder output
Temperature Range
: Output a signal (isolated) proportional to
measurement value (The number of output channels Operating temperature : -20 to +65°C
is selected from Model Code.) Storage temperature : -30 to +85°C
Relay output : Output contact signal from relay. Relative humidity : 20 to 90%RH(non-condensing)
(The number of output channels is selected from
Model Code.)
Power Supply Voltage
Modbus/TCP communication
: Output of data by Modbus protocol Rated Power Supply Voltage : 24VDC
Power Supply Voltage Range : 22 to 26VDC
Power consumption : 24W(Max.)
Display
Power supply : Green LED
Dimensions
Normal : On
Stopped : Off Approx. 158.2(W) x 99(H) x 112.85(D) mm (Excluding the projection parts)
Alarm : Red LED
When DANGER alarm occurs : On
Mounting Method
When ALERT alarm occurs : Flashing
Normal : Off 35mm DIN rail
Sequence : Yellow LED
Operating : On
Mass
Stopped : Off
DANGER Bypass : Green LED All full load : Max. 0.9kg(2.0lb)
Operating : On
Stopped : Off
Communication : Green LED
Connected : On
Communicating : Flashing (100msec interval)
Disconnected : Off

6H16-032 Rev.9 [C]


Issued Jun. 2017
Revised Dec 2018
VM-25 VM-25 CONDITION MONITORING SYSTEM
MONITORING SYSTEM

SPECIFICATIONS Page 2 of 5

Specification

DISPLAY FUNCTION Sequence Function*1


Used to prevent alarm output that is caused by excessive vibration during
Display
machine startup. Block off the DANGER/ALERT alarm, or switch the alarm
Display : 4 digits 7-segment Red LED display. setup value to another number magnified by setup number.
Character height : 8mm Sequence Setup Value : 0(Block off), 1 to 10 (0.5step)
Display Accuracy
Display Contents
:
:
±(I/O conversion accuracy + 1 digit) at 25°C
Measurement value △ ! WARNING
GAP/Bias voltage value The sequence function may leave the machine unprotected.
DANGER Alarm set value In case the SEQ. magnification number is set up “0“, machine is
ALERT Alarm set value unprotected as alarms are disabled.
OK Alarm set value In case the SEQ. magnification number is set up from 1.5 to 10.0, and
Display Mode SEQ. function is in effect, alarm set points are multiplied by the set value.
Display Mode Description If the number is more than 150% of the monitor range the alarm may not
All-channel cycle display Measurement values for all input channels are output.
displayed in turns at intervals of five seconds.
Specific channel cycle Measurement values for specific input channel Suppression Function*1
display or alarm setting values are displayed in turns at If the vibration value is less than the setup value, this function is forced to
intervals of five seconds. suppress the measured vibration value and recorder output*2.
Specific channel fixed Measurement values for the specific input Suppression Setup Value*3: 0 to 5% ( 0.1% step)
display channel are displayed continuously.
100(%)
All-channel maximum The maximum measurement value for all the
value display input channels is displayed.

DISPLAY
OUTPUT
※ Selectable by front switch.

2CHANNEL VIBRATION MEASUREMENT FUNCTION 0


0 * 10 Input 100(%)
3

Input/output characteristic
Input Points
Input points : 2 channel / slot
Burnout Function*1
*1
Input Transducer When an OK alarm occurs (Output defeat function: ON) or at the monitor
startup, the recorder output and the measurement value are controlled as
Eddy current transducer : FK-202F
follows.
Displacement vibration mesurement
Burnout function setting
Velocity transducer : CV-86, CV-87
Velocity vibration measurement or Burnout
Burnout setting value
displacement vibration measurement Recorder output value Mesurement value
Acceleration transducer : CA-302 Down Scale 0% Equivalent to “0” of the 0
Acceleration vibration measurement or measurement value
velocity vibration measurement Down Scale 0mA / 0V 0V or 0mA 0
Other transducer : Voltage signal
Conversion Accuracy
Vibration Measurement
I/O Conversion Accuracy :±1.5% of F.S. at 100Hz at 25°C
Monitor type*1 : Displacement vibration, Velocity vibration, ±3.0% of F.S. at 100Hz at -20 to +65°C
Acceleration vibration
Monitor range*1 : 1 to 1000 Output
Unit : μm, mils, mm/s, in/s, m/s2, g
Measurement detection : pk-pk、pk、rms Recorder(Non-Isolated) Output*1*4 :
Rectification : Avarage value Current / Voltage output proportional to
Measurement range(AC) : 0 to 9V pk measurement value.
The input voltage of 100% of F.S. at 100Hz should be Measurement value of each channel can be
within the range of 39.4mV pk to 9V pk. assigned to any output channel of its own slot.
Measurement range (DC) : 0 to 22VDC(CV-86, CV-87, CA-302 input) The recorder output gain (1 time or 2 times) can be
: -22 to 0VDC(FK-202F input) changed.
Input voltage range : -24 to 24VDC Number of output points : 2 channel
Input impedance : Approx. 50kΩ Output range : 4 to 20mADC, 1 to 5V
Max. load resistance : 600Ω (current mood)
Output impedance: Approx. 250Ω (voltage mood)
Frequency Response*1 Monitor output : Input signal is output via buffer amplifier.
4 pole high-pass filter : 2Hz, 5Hz, 10Hz, 20Hz (-3db) Location : BNC(Front) and Terminal block(Bottom)
2 pole low-pass filter : 500Hz, 1kHz, 4kHz, 10kHz (-3db) Output impedance : Approx.100Ω
Output voltage : Max.5mA
Transducer Power Supply :
Proximity transducer (FK-202F) : -24VDC±10% / 25mA Max.
Piezoelectric transducer (CA-86, CA-302)
: +24VDC±10% / 4mA (constant current)

6H16-032 Rev.9 [C]


Issued Jun. 2017
Revised Dec 2018
VM-25 VM-25 CONDITION MONITORING SYSTEM
MONITORING SYSTEM

SPECIFICATIONS Page 3 of 5

Specification

Alarm*1 4 CHANNEL RELAY OUTPUT FUNCTION


Alarm Set Point : DANGER Alarm
Output Points
1 point, from 0 to 100% of monitor range
ALERT Alarm Output points : 4 channel / slot
1 point, from 0 to 100% of monitor range
Alarm Delay Time : 0 to 5sec.(0.5sec step)
Alarm*7
Alarm Reset : AUTO-RESET / SELF-HOLD
Logic : Changeable
Display Energization method : Normally de-energized / Normally energized
Contact type : Dry contact (SPDT)
OK Alarm : Green LED / Red LED Contact capacity : 250VAC/2A, 30VDC/2A
Normal : ON(Green)
When OK alarm occurs : Flashing(Green)
When DANGER alarm occurs : ON(Red) Insulation Resistance
When ALERT alarm occurs : Flashing(Red) Input, Power, GND*8 – Recorder output (isolated) – Contact
When Channnel Bypass occurs, : 100MΩ at 500VDC
or the channel disables : Off
Notices Withstand Voltage
All of the terminal on this function should not be connected to earth. Input, Power, GND*8 – Recorder output (isolated) –Contact
: 100VAC for 1minute.
*1 User can change the setting by VM-25S01 Device Config. *7 User can change the setting by VM-25S01 Device Config.
*2 Suppression function also applies to the output from VM-25F26 4 channel isolated *8 Input, Power and GND are not isolated from each other.
recorder output function.
*4 When the Recorder (Non-isolated) output terminal is connected to a instrumentation
device, the input of the device should be an isolated type or an isolation amplifier should MODBUS / TCP COMMUNICATION FUNCTION
be provided between the terminal and the device.
Communication Function
4 CHANNEL ISOLATED RECORDER OUTPUT FUNCTION Network : Ethernet 10Base-T / 100Base-TX
Protocol : Modbus® Based on AEG Modicon PI-MBUS-300
Output Points Reference Manual.
Output points : 4 channel / slot Transmission mode : RTU (Remote Terminal Unit) mode
Connector : RJ-45 (Shared to DEVICE CONFIG)
Output
Recorder output (Isolated)*5 Input / Output Data
Current / voltage output proportional to Data transmitted from VM-25 to host network
measurement value. : Measurement value, Peak mesured value,
Measurement value of each channel can be Gap/Bias voltage value,
assigned to any output channel of its own unit. DANGER Alarm status,
The recorder output gain (1 time or 2 times) can be ALERT Alarm status, OK Alarm status,
changed. DANGER Bypass status, Setting of DANGER,
Output range : 4 to 20mA, 1 to 5V Setting of ALERT, and OK Limit
Max. Load Resistance : 600Ω(current mood))
Output impedance : Approx. 250Ω (voltage mood) Data received by VM-25 from host network
: Date and Time Data
Insulation Resistance
Function of VM-25 that can be controlled from host network
Input, Power, GND*6 – Recorder output (isolated) – Contact
: Channel Bypass status ON/OFF
: 100MΩ at 500VDC
DANGER Bypass status ON/OFF
Withstand Voltage Peak Hold reset
Alarm Reset
Input, Power, GND*6 – Recorder output (isolated) –Contact
Sequence Mode ON/OFF
: 100VAC for 1minute.
*5 User can change the setting by VM-25S01 Device Config.
*6 Input, Power and GND are not isolated from each other.

Alarm Contact Operation

Power ON
Contact type Energization method Power OFF
Normal state Alarm state
NORMALLY DE-ENERGIZED OPEN OPEN CLOSE
N.O. contact
NORMALLY ENERGIZED OPEN CLOSE OPEN
NORMALLY DE-ENERGIZED CLOSE CLOSE OPEN
N.C. contact
NORMALLY ENERGIZED CLOSE OPEN CLOSE

6H16-032 Rev.9 [C]


Issued Jun. 2017
Revised Dec 2018
VM-25 VM-25 CONDITION MONITORING SYSTEM
MONITORING SYSTEM

SPECIFICATIONS Page 4 of 5

Default Value

Alarm Isolated Recorder output


DANGER set value : 80% of monitor range Recorder Output gain : 1
ALERT set value : 60% of monitor range Recorder Allocation :
OK set value : [FK-202F] -1.4V(Low), -18.8V(High) Number of input channel : 2ch
[CV-86] 2.1V(Low), 22.0V(High) Slot 3
[CA-302] 2.1V(Low), 22.0V(High) [Recorder1] Measurement channel 1 of slot 1
Alarm delay time : 3.0 sec.(DANGER,ALERT) [Recorder2] Measurement channel 2 of slot 1
Alarm reset : Auto Reset [Recorder3] Measurement channel 1 of slot 1
[Recorder4] Measurement channel 2 of slot 1
Non-Isolated Recorder output
Number of input channel: 4ch
Recorder Output gain : 1 Slot 3
Recorder Allocation : [Recorder1] Output of measurement ch1 of own slot [Recorder1] Measurement channel 1 of slot 1
[Recorder2] Output of measurement ch2 of own slot [Recorder2] Measurement channel 2 of slot 1
[Recorder3] Measurement channel 1 of slot 2
Contact output
[Recorder4] Measurement channel 2 of slot 2
Relay logic : [RELAY1] OR logic (DANGER of all channels) Number of input channel : 6ch
[RELAY2] OR logic (ALERT of all channels) Slot 4
[RELAY3] OR logic (NOT-OK of all channels) [Recorder1] Measurement channel 1 of slot 1
[RELAY4] None [Recorder2] Measurement channel 2 of slot 1
Enagization method : NORMALLY DE-ENERGIZED [Recorder3] Measurement channel 1 of slot 2
[Recorder4] Measurement channel 2 of slot 2
Communication
IP address : 192.168.8.8 Slot 5
Subnet mask : 255.255.255.0 [Recorder1] Measurement channel 1 of slot 3
Port number : 8888 [Recorder2] Measurement channel 2 of slot 3
[Recorder3] Measurement channel 1 of slot 3
Others
[Recorder4] Measurement channel 2 of slot 3
Sequence : 1.0 (DANGER,ALERT)
Suppresstion : 0.0%
Timed OK channel defert : ON
Burnout : Down Scale 0mA

Functional Specification Code (Default)

 2ch VIBRATION MEASUREMENT FUNCTION


VM-25F01 - ∗ ∗∗ ∗ ∗ ∗ ∗0–0

Channel1, 2 Recorder Output


Functional Specification Code
Monitor Range Transducer Type HPF LPF (Non-isolated)
VM-25F01-D03D1230-0 0 to 200 µm pk-pk FK-202F 5Hz 4kHz 4 to 20mA
VM-25F01-D01V1320-0 0 to 100 µm pk-pk CV-86 10Hz 1kHz 4 to 20mA
VM-25F01-V01V1320-0 0 to 25 mm/s pk CV-86 10Hz 1kHz 4 to 20mA
VM-25F01-V01A1320-0 0 to 25 mm/s pk CA-302 10Hz 1kHz 4 to 20mA
VM-25F01-V03V1320-0 0 to 20 mm/s rms CV-86 10Hz 1kHz 4 to 20mA
VM-25F01-V03A1320-0 0 to 20 mm/s rms CA-302 10Hz 1kHz 4 to 20mA
VM-25F01-A05A1340-0 0 to 5 g pk CA-302 10Hz 10kHz 4 to 20mA

 4 CHANNEL ISOLATED RECORDER OUTPUT FUNCTION


VM-25F26-0 0 0 0

Isolated recorder output


Recorder1,2,3,4
0000 4 to 20mA

 4 CHANNEL RELAY OUTPUT FUNCTION


VM-25F21
6H16-032 Rev.9 [C]
Issued Jun. 2017
Revised Dec 2018
VM-25 VM-25 CONDITION MONITORING SYSTEM
MONITORING SYSTEM

SPECIFICATIONS Page 5 of 5

Functional Layout

VM-25M00-151-01-00-00-00-00 VM-25M00-151-01-01-21-00-00 VM-25M00-151-01-01-01-00-00


1. Vib. measurement 1. Vib. measurement 1. Vib. measurement
2. Blank 2. Vib. measurement 2. Vib. measurement
3. Blank 3. Relay Output 3. Vib. measurement
4. Blank 4. Blank 4. Blank
5. Blank 5. Blank 5. Blank
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5

VM-25M00-151-01-00-21-00-00 VM-25M00-151-01-01-21-21-00 VM-25M00-151-01-01-01-21-00


1. Vib. measurement 1. Vib. measurement 1. Vib. measurement
2. Blank 2. Vib. measurement 2. Vib. measurement
3. Relay Output 3. Relay Output 3. Vib. measurement
4. Blank 4. Relay Output 4. Relay Output
5. Blank 5. Blank 5. Blank
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5

VM-25M00-151-01-00-26-00-00 VM-25M00-151-01-01-26-00-00 VM-25M00-151-01-01-01-26-26


1. Vib. measurement 1. Vib. measurement 1. Vib. measurement
2. Blank 2. Vib. measurement 2. Vib. measurement
3. Recorder Output 3. Recorder Output 3. Vib. measurement
4. Blank 4. Blank 4. Recorder Output
5. Blank 5. Blank 5. Recorder Output
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5

VM-25M00-151-01-00-26-21-00 VM-25M00-151-01-01-26-21-00 VM-25M00-151-01-01-01-01-00


1. Vib. measurement 1. Vib. measurement 1. Vib. measurement
2. Blank 2. Vib. measurement 2. Vib. measurement
3. Recorder Output 3. Recorder Output 3. Vib. measurement
4. Relay Output 4. Relay Output 4. Vib. measurement
5. Blank 5. Blank 5. Blank
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5

VM-25M00-151-01-01-00-00-00 VM-25M00-151-01-01-26-21-21
1. Vib. measurement 1. Vib. measurement
2. Vib. measurement 2. Vib. measurement
3. Blank 3. Recorder Output
4. Blank 4. Relay Output
5. Blank 5. Relay Output
1 2 3 4 5 1 2 3 4 5

※ The specifications and other items indicated herein are subject to change without notice.

6H16-032 Rev.9 [C]


Issued Jun. 2017
Revised Dec 2018
VM-25 VM-25 DEVICE CONFIG
MONITORING SYSTEM
(Configuration software for VM-25 condition
SPECIFICATIONS monitoring system) Page 1 of 1

No entry if additional
Model Code / Additional Spec. Code spec, code is not specified.

VM-25S01 /SU0
Setup*1
*1 A PC is required. Setup option includes initial setup of the PC, installation of the software and PC address setting.

Specification

SYSTEM REQUIREMENTS CONTROL FUNCTIONS


ALARM CONTROL : Alarm reset, Peak-hold reset
HARDWARE REQUIREMENTS
BYPASS CONTROL : DANGER bypass, Channel bypass
PC/AT compatible personal computer LED CONTROL : All LEDs on/off
Processor 1GHz or higher recommended
Memory 1GByte or higher recommended OTHER FUNCTIONS
Display 1280 x 768 or higher –resolution is recommended Password protection : Password authentication on start-up, write
Hard disk drive 1GByte of available hard-disk space configuration to the device and adjust the device.
Drive CD-ROM drive Firmware update : Update a firmware of VM-25.
Network Ethernet 100BASE-TX Device setting file output : Output a configuration of VM-25 setting as text file

SOFTWARE REQUIREMENTS SOFTWARE DISTRIBUTION MEDIA


Category Model Edition Version Media : CD-ROM
OS Microsoft® Windows® 7 (32/64bit) Professional Any
or higher SYSTEM CONFIGURATION
Microsoft® Windows® 8.1(32/64bit) Pro or higher Any
Microsoft® Windows® 10 (32/64bit) Pro or higher Any Ethernet
Others Microsoft® .NET Framework 4.0 100BASE-TX

CONNECTION
Supported devices : VM-25 Monitor VM-25 Monitor VM-25S01 Device Config
Number of simultaneous connections : 1device*3
*3 Connected device can be changed by altering the IP address, if the connection is made
via Ethernet. Example of Screen

COMMUNICATION Ethernet
Network : Ethernet 100Base-TX
Protocol : TCP/IP

SCREEN
Screen display language : English

DEVICE CONFIGURATION MANAGEMENT


Configuration change : Change monitor type (displacement vibration,
velocity vibration or acceleration vibraton)
Paramter change : Parameter setting of vibration measurement
(Measurement, Alarm), Recorder output, Relay,
Parameter setting such as communication
Adjustment : Gap/Bias voltage value, measurement value, etc
Others : Channel disablement,
Time setting, Simulated value setting,
Modbus communication setting
(Data copy setting, scaling setting)

DISPLAY FUNCTION
Actual screens may differ from the images shown in this document
MEASUREMENT VALUES
Gap/Bias Voltage Indications, Digital indication of measurement values,
Peak measurement values

ALARM STATUS
DANGER alarm, ALERT alarm, OK alarm

OTHER DISPLAY
The specifications and other items indicated herein are subject to change without notice.
Time etc. All company and product names in this brochure are trademarks or registered trademarks.

6H16-044 Rev.4 [C]


Issued Jun. 2017
Revised Dec. 2018