Professional Documents
Culture Documents
Koenig 1980
Koenig 1980
Koenig 1980
Koenig
Professor of Mechanical Engineering
and Orthodontics.
Introduction
A primary goal of Orthodontic research is to design ap- analysis, any appliance may rapidly be investigated to
pliances which will deliver predetermined force systems. In calculate the force system which is developed upon activation
order to achieve this goal, a system must be developed of the loop. The maximum stresses in the wire may also be
whereby a complex loop design may be analyzed to determine determined to insure that the activation is within the elastic
the forces and moments which will be delivered upon ac- range. Deactivation forces and moments may also be
tivation. Moreover, the capability of redesigning any ap- calculated upon insertion of the resulting tooth movement.
pliance which does not deliver the correct forces, must be Redesign of the appliance takes place employing an in-
available. Two distinct approaches are possible. teractive graphics technique which is coupled to the basic
One may rely on a purely experimental system to achieve analysis. A perturbation to the basic shape may be made and
the goal of force system orchestration. In this method, a the new appliance is reanalyzed. This process continues until
highly sophisticated set of experimental equipment must be the final loop design is achieved. Basic geometric parameters
constructed so that each individual appliance can be analyzed such as interbracket distance, loop lengths, loop heights may
to determine the activated force system. be easily varied so that the effect of their changes may be
Experimental studies for a particular configuration are studied. Future appliance designs can use the knowledge
performed varying design parameters such as loop height, obtained from these studies.
width, placement of helices and interbracket distances. Data In this paper, the experimental and analytical techniques
can be plotted for interpolation and a predictive mathematical are compared with one another for two separate loops un-
model can be constructed which is used for design im- dergoing a total of four separate activations. A T-loop and a
provement. New designs are then tested experimentally. rectangular loop are activated both experimentally and
Experimental approaches have the advantage that actual analytically. Results show close agreement between the two
force systems are measured which may include variables that techniques over a wide range of activations.
are not included in an analytical model. Unfortunately,
current experimental devices are not completely isometric Experimental Method
and, hence, are inaccurate for high load-deflection ap-
pliances. A device for measuring uniplanar forces and moments (Fig.
A second distinct direction towards loop design may be 1) delivered by orthodontic appliances was developed in the
pursued. This, the analytical approach, is based upon the Bioengineering Laboratory of the Department of Or-
development of an accurate computer program which is based thodontics, School of Dental Medicine, University of Con-
upon fundamental principles of engineering. Using this necticut Health Center (Solonche, et al. [8, 9] and Vanderby,
et al., [10]). Forces and moments are converted to linear and
angular displacements, respectively, and then transduced to
Contributed by the Bioengineering Division for publication in the JOURNAL
OF BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering
electrical signals. These signals are Analog to Digital con-
Division, October 16,1979; revised manuscript received July 24, 1980. verted and fed to a mini-computer for analysis. The appliance
- - - - Nomenclature
.19614 n
Analytical-Experimental Comparison
If the ultimate goal of the researcher is to have a predictive
tool with which to analyze, design and redesign appliances for
clinical application, confidence must be gained in the
analytical approach described in the foregoing. The best
means of achieving this goal is to simultaneously activate
certain appliances using the experimental apparatus described 0.25 0.50 0.75 1.0 1.25 15 1.75 2.0
in the foregoing while simulating these activations using the DEFLECTION-W -mm 2
computer code, which is derived from the theoretical ap Fig. 6 Theoretical versus experimental comparison of mesial force (T-
proach. loop)
Figure 3 shows a T-loop which is activated by applying a
0.196-N vertical load on one end while attempting to com end rotations, activation force, and end moments, respec
pletely restrain motion at the other end. The cross section of tively, for both the experimental and analytical methods. The
the wire is nominally 0.01 in. x 0.022 in. (0.25 mm x 0.56 small deviations which result encourage the future use of the
mm), but has been measured for its exact properties as shown. computer code as a predictive tool. The numerical values for
The rotation of the transducer chuck at the fixed end was the T-loop are shown in Table 3.
measured as 0.11 deg. This end rotation was also inserted in Figure 8 shows a rectangular loop which was chosen for
the computer code. Table 2 shows that a maximum deviation further comparisons. Two sets of experiments were per
of 4.06 percent results between the analytical and ex formed. A negative displacement of the wire in the intrusive
perimental output. direction yields the comparisons shown in Figures 9-11 and
Figure 4 shows the geometric parameters of a T-loop which demonstrate the accuracy of the theoretical development for
is given an experimental activation of 2 mm. This activation predictive purposes. Figure 12 shows a distal force which is
for a 7-mm interbracket appliance closely simulates the predicted by the computer program. Since this force has a
clinical practice for cuspid retraction. Figures 5-7 show the value which is approximately 36 percent of the intrusive force,
I
2 -
Table 3 T-Ioop
• R (= I mm TypJ
F i g . 11 A n a l y t i c a l v e r s u s e x p e r i m e n t a l c o m p a r i s o n of a n -
terior/posterior m o m e n t (rectangular loop, negative displacement)
- EXPERWENmL^,,
• EXPERIMEN1AL&
THEORETICAL
F i g . 9 Analytical versus experimental comparison of e n d rotations Fig. 12 Analytical description of mesial force (rectangular loop,
(rectangular loop, negative displacement) negative displacement)
-
< /&'
cfr%
/ ?/ /
1
I
a A>'
/ #//
2
5 /?" __-^^=e=-—
jSs* — — " ^ ^
SJ3 -~ ~"1!H—^— ~~
jfesfi^ 'i —i 1 1 1 1 1 ZM
I.7S
<+>
Fig. 13 Analytical versus experimental comparison of end rotations
(rectangular loop, positive displacement)
Fig. 15 Analytical versus experimental comparison of an-
terior/posterior moment (rectangular loop, positive displacement)
-^-EXPERIMENTAL(Nornlrial)
THEORETICAL
y 60
K
c
I
U.* 40
I
ds E3A
+
\34"/L\a1G13/Sln \<fc/
dWb
<t>n=- + TW„ (11) 1 tf(a,G13)„:
ds 2
sin20
• ( • • • • )
(a,G 1 3 ) 2 ds
dW„
+ «uV.-TV, (12) 2
ds (\ a-,^G-,,W
/
^ H
\ ds /
- L V
Va, G-n /
^l^>cos
ds
0l(27)
J
3,=*, (13)
<n-(^k,>o r_(s)=x(s)i+y(s)j+Z(s)k
dr
,. ,- r,,,
e , = -f =x'i+y'j + Z'k
(32)
(33)
i d
( a l G 1 3 ) . _ o 22 ds ~
cos 0
(«,G 1 3 ) 2 ds de ,
(34)
2 ds
+ (\a G' / W^W^V^^sin
\ds/ \a G / ds
*
2 2i 2 2i de_,
(35)
/sin20\/_J 1_W / & \ ~ds~
V 2 /\a,G„ axGn) ds \A ) de
•• — ne t + re (36)
1 1 ds
+ COS26
(fs)( of,G 13 a2G23- /) V
( !.4) deb
(37)
/sin20\/g
Q„6\\ rr --7
7 d(a!G13) ds
2
\ 2 / \ ^ 4 /L(a,GI3) ds e
.,xe„=eb (38)