Professional Documents
Culture Documents
Open Canal Fittings Ten Take Home Tips.6
Open Canal Fittings Ten Take Home Tips.6
Open Canal Fittings Ten Take Home Tips.6
Fitting hearing aids involves innovation and compromise. ple areas where misconceptions exist. In this paper, we’ll pro-
This is especially true in deciding the tightness of the fit, vide some tips on OC fittings, which we hope will clarify
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
Figure 1. A sample patient’s REUR, and the resulting REORs Figure 2. Mean REUG and REOG findings for 14 individuals
for two different earmold fittings. fitted with an open-canal fitting tip for the REOR measurements.
OC fittings? This has prompted some to big it was. At this point, it doesn’t matter systematically influence your REAR mea-
question if it is still okay to subtract the if the REOG, before the hearing aid was sures or REIG calculations.
REUG and use REIG calculations. The turned on, was -30 dB or 0 dB. It is unnec-
quick answer is yes, but this too will be essary to make up for this negative REOG ISSUE #2: EARTIP OCCLUSION
explained more in a later section. with amplifier gain. As mentioned, one potential benefit of
The third use of the term “insertion An example of this is shown in Figure OC fittings is the use of an eartip in the
loss”--the misleading use we’re concerned 1. The top curve shows this patient’s ear canal that reportedly is non-occlud-
with here--relates to the change in ear canal REUR, and the two lower curves repre- ing. Manufacturers have designed a vari-
SPL for a given input when a hearing aid sent the REOR for two different earmolds. ety of eartips for this purpose. But it’s
(or earmold) is placed in the ear, with the Earmold #2 was deeper and tighter fitting, important to ask, how non-occluding are
hearing aid turned off. This is assessed by which is why this curve is lower in ear canal these new eartips, and what is the result-
conducting the real-ear occluded response SPL. If we fall into the trap of subtracting ing ear canal resonance when they are in
(REOR) and calculating real-ear occluded these REORs from the REUR, we would place?
gain (REOG), which often is negative.11,12 find that in the 2000-Hz range, the UR- To answer these questions, we con-
It is tempting then to subtract the REOG OR difference is about 23 dB for earmold ducted probe-mic testing on 14 adults,
from the REUG and think of this as inser- #1 and 30 dB for earmold #2. But this seven males and seven females, ranging in
tion loss. Because the REUG in the high doesn’t mean this patient needs 7 dB more age from 25 to 58. We measured the
frequencies is usually around 15-17 dB, amplifier gain if he is fitted with earmold REUR for the right ear of each partici-
and the earplug effect often results in an #2. In fact, because earmold #2 is a deeper pant. Each person was then fitted with an
REOG of -10 dB to -20 dB in the highs, fit, and there is a smaller residual ear canal OC product (turned off ) using an open
the UG-OG difference can be quite large. volume, he actually might need less ampli- eartip that was sized to give each person a
This calculation has led some to conclude fier gain with this fitting. “tight” fit. An REOR was then conducted.
that in a closed fitting, the amplifier must So, as you can see, the term “insertion The resulting REUG and REOG mean
add as much as 40 dB of gain to com- loss” has been used in different ways and findings are shown in Figure 2. Observe
pensate.13 Why is this logic wrong? This misuse can lead to confusion. While sub- that the average REUG is fairly typical,
amount of insertion loss would only apply tracting the REOR from the REUR does and not too different from the average val-
if the patient never turned on the hearing provide guidance concerning the “open- ues used in common prescriptive meth-
aid. We’ll explain. ness” of the OC fitting, which probably ods or as an average-ear reference in
Consider that when a relatively closed will help predict the degree of occlusion probe-microphone equipment. The mean
fitting is used, the hearing aid is an earplug. effect, it is not a very useful measure for REOG values indicate that, indeed, these
In the turned-off measurement, sound fitting the hearing aid. The term “inser- eartips were not very occluding, as the
must pass through the hearing aid (or ear- tion loss” was perhaps handled best in the resulting gain for the “occluded” ear was
mold) and around the hearing aid, or ANSI Real-Ear Standard S3.46-1997. Of not too much different from the open ear.
through the vent (if there is one), to be the 50 or so terms defined in this stan- Note that the typical REUG peak at 3000
picked up by the probe mic in the ear dard, insertion loss was not one of them. Hz was reduced by 3 dB, but there now
canal. But, once the hearing aid is turned is an REOG peak at 2000 Hz. Although
on, a new pathway is established through Take-Home Tip #1: REOGs are not a we haven’t investigated this systematically,
the microphone inlet port and the hear- true measure of insertion loss. They sim- we have observed in OC fittings, as might
ing aid. Hence, the total “earplug effect” ply indicate how tight the fit is (or how be expected, that the REAR peak is aligned
is immediately eliminated, no matter how vented). The size of the REOG does not with the patients REOR peak. In case
26 The Hearing Journal Open-Canal Fittings: A Special Issue November 2006 • Vol. 59 • No. 11
you’re wondering, we have no explanation little or no occlusion with another eartip, hearing aid occlusion effect. Recent
for why the placement of the eartip in the and MacKenzie16 found no difference research has shown this to be true. Most
ear canal moved the resonant peak to a among three different eartips when he used recently, MacKenzie compared the occlu-
lower frequency. a modified REOR measure (with the sion effect for the OC product of three
We obtained these results only using patient’s own voice as the input). To date, manufacturers.16 In all cases, the occlu-
eartips made by a single manufacturer, but our UG-OG comparisons have not sion effect was absent or so small that it
we would predict similar results with other included OC instruments that place the was not considered annoying.
eartips. For example, Yanz and Olson, receiver in the canal, but indirect data sug- Eliminating the occlusion effect can
using a different eartip, show essentially gest that the eartip for at least one of these make a dramatic difference for some
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
the same UG-OG difference (also with a models also is not very occluding.17 patients, and can spell the difference
trend toward a lower average peak for the If the UG-OG difference is small, it between hearing aid acceptance and rejec-
REOG).14 Meanwhile, Kuk et al. showed would seem there would be little or no tion of amplification. An example of this
is shown in Figure 3.
This represents the probe-mic mea-
surement of the occlusion effect for a
patient experiencing occlusion problems.
The patient’s own voice, the “eee” vocal-
ization, was used as the input signal, and
the hearing aid was turned off during test-
ing. Observe that with his own standard
earmold, the ear canal SPL was 25-30 dB
higher in the low frequencies compared
with the open-ear condition. With the
OC eartip, the occlusion effect was nearly
eliminated and only a few dB remained
around 200 Hz. This improvement is
noticeable and significant and nearly
always leads to improved patient satisfac-
tion (assuming that the fitting can deliver
appropriate gain).
Figure 3. The measurement of the hearing aid occlusion effect Figure 4. The output of an ER-2 earphone measured in a
for the “ee” vocalization by the patient. Shown is the open-ear Zwislocki coupler for three different depths (relative to the
SPL, with the resulting occlusion effect for an open-canal fitting opening of the coupler). The earphone tube was placed in the
tip and for a closed earmold. coupler without the usual occluding foam tip.
this sound delivery is not too far from the KEMAR. Specifically, a steady-state noise tion depths relative to the opening of the
TM. We wondered if the same residual was delivered into the coupler using an Zwislocki coupler (2 mm, 12 mm, 25
resonant properties observed in the mean Etymotic ER-2 earphone. This earphone mm). The output was then calculated
REOG data in Figure 2 would still be pre- terminates in a tube in a similar fashion based on digital recordings made through
sent when the signal is delivered in the ear to common insert earphones. However, an Etymotic ER-11 microphone coupled
canal rather from the sound field. in this case we did not place a foam tip to the Zwislocki coupler.
To examine this, we conducted mea- on the end of the tube as is usual. The tip As Figure 4 shows, there was little or
sures using a Zwislocki coupler in the of the tube was then placed at three inser- no change in the frequency location of
November 2006 • Vol. 59 • No. 11 Open-Canal Fittings: A Special Issue The Hearing Journal 29
the peak or the magnitude of the output higher frequencies, however, the two sig- needs 15-20 dB or more of gain, the sum-
when the signal was delivered at the micro- nals interact, which results in summing ming and cancellation effects are mini-
phone (25 mm; eardrum location) or at and cancellation effects observed as a rip- mized, as the hearing aid path signal will
the entrance of the ear canal (2 mm). This ple in the combined signal. This ripple is have significantly higher amplitude than
suggests that REOR data are a good pre- most evident from about 500 to 1500 Hz, the direct path.
dictor of ear canal characteristics for a sig- the region where the signals from the two
nal presented from a hearing aid and pathways are similar in amplitude, result- Take-Home Tip #4: Summing effects do
delivered in the ear canal itself. ing in notches greater than the 6-dB not occur in the absence of cancellation
“bumps” observed at other frequencies. It effects and are largest when the hearing
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
Take-Home Tip #3: The residual ear canal should be noted however, that the aid is providing little or no gain. As a result,
resonance observed in the REOG of an “notches” are generally narrower than the summing effects are not expected to play
OC fitting is also present when the signal bumps, and they may therefore have less a significant role in achieving desired real-
is delivered in the ear canal. perceptual impact on signal level. ear gain and output.
Those of you who conduct probe-mic
ISSUE #4: SUMMING AND measures may or may not have observed ISSUE #5: “FREE”
CANCELLATION EFFECTS these notches, depending on the sampling HIGH-FREQUENCY GAIN
Consider that with an OC fitting, con- rate and smoothing of the system you use. We often hear clinicians say that people
siderable sound enters the ear canal In fact, on some systems, the notches fitted with OC products “don’t need much
through the direct pathway (i.e., between the bumps are smoothed out, high-frequency gain” or read reports that
unprocessed by the hearing aid), which is making it look as if there is simply a little state that with OC products manufactur-
then mixed with the amplified sound. more gain. ers can take advantage of the 15-20 dB of
When these two inputs (from the same From a clinical standpoint, it is impor- natural amplification of the unaided ear.19
source) are fairly equal in amplitude, sum- tant to consider the significance of the Is there really “free gain” in the high fre-
mation and cancellation can occur across effects shown in Figure 5 for routine OC quencies with these products? And if so,
frequencies, depending on the phase rela- fittings. First, it seems unlikely that the how much?
tionship at the tympanic membrane. That summing effects will provide much added The issue of free gain is related to the
is, for frequencies that are in phase, there benefit. They are largest when the two sig- insertion loss issue that we discussed ear-
will be summing effects (in theory, as much nals are similar in level, which is when the lier. On one hand, with an OC fitting we
as 6 dB). hearing aid is providing minimal gain, and have the residual ear canal resonance in
This sometimes is reported as a plus they never exceed 6 dB. If the hearing aid the 2000- to 4000-Hz range, which will
for OC fittings. We’ve heard the comment: was programmed to provide minimal gain, enhance amplifier gain, thus giving an
“You don’t need as much amplifier gain then we’d presume the patient needs only advantage to the OC. On the other hand,
because of summing effects.” However, minimal gain, and doesn’t need the added with a closed fitting, especially using a
for frequencies that are out of phase, there summation effects. deep earmold or deep custom-fitted instru-
can be cancellation effects. It has been sug- In addition, perception of the sum- ment, we have a reduced residual volume
gested that this could be a negative con- ming effects under these conditions is and impedance changes which also increase
sequence of OC fittings, because of expected to be at least partially offset by high-frequency gain, which gives an advan-
distortions.18 the cancellation effects. When the patient tage to the closed condition. Who wins?
To help explain this, we will We examined this in two
use data collected at the Uni- different ways. The first
versity of Iowa Auditory method was to examine out-
Research Laboratory by Ruth put in a Zwislocki coupler for
Bentler and Yu-Hsiang Wu “closed” versus “open” condi-
(see Figure 5). This figure tions. A steady-state noise was
shows the ear canal SPL for output into the coupler using
three different conditions: the an Etymotic ER-2 earphone.
direct path, the hearing aid The tip of the tube was then
output path, and the com- placed at an insertion depth of
bined. The input signal was approximately 25 mm so we
the speech-shaped babble of could examine the output for
the CST. an “open” condition. Without
Observe that for the lower moving the tube, we then
frequencies, the only path is measured output for the
the direct path. That is, “closed” condition by sealing
because of the open fit, there the entrance of the Zwislocki
is no hearing aid gain for this Figure 5. Data showing SPL for the direct path, the hearing aid path, coupler with Fun-Tac. We
region and no summing or and the combined results for on OC fitting. Input signal was the sealed only the entrance of the
speech-shaped babble of the CST.
cancellation is present. For the coupler because we were inter-
30 The Hearing Journal Open-Canal Fittings: A Special Issue November 2006 • Vol. 59 • No. 11
ested in avoiding big changes in the cou- is a good reminder of the amount of
pler volume; this is different from an amplifier gain that is needed to obtain
actual hearing aid fitting where the ear just a few dB of real-ear gain in the low
canal volume becomes smaller when a frequencies for an OC fitting. This was
closed fitting is used. The output was especially noticeable for Hearing Aid A.
again calculated based on digital record- The minimal low-frequency gain for
ings made through an Etymotic ER-11 Hearing Aid B apparently is due to man-
microphone coupled to the Zwislocki ufacturer design, as gain for this region
coupler. was programmed at “max” (important
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
The difference in output between to know if one were ever to use this prod-
these open and closed conditions was uct with a closed earmold). Also, due to
calculated by subtracting the closed some feedback problems, Hearing Aid
response from the open response, and is B had slightly less average REIG—not
shown in Figure 6. Not surprisingly, a concern, however, as we were only
given the resonance frequency of the Figure 6. The difference in output for an ER-2 interested in relative differences.
coupler (which, of course, is intended earphone in a Zwislocki coupler calculated by Our question related to free gain,
to mimic the average ear), up to 10 dB subtracting the level measured with the tube sealed however, is best answered, or at least
more output was achieved for the open in the coupler from the level measured with the visualized, by calculating a difference
condition in the high frequencies, with tube simply placed in the coupler at the same depth. curve (open REIG versus closed REIG)
a peak in the 3000-Hz range. This for both products, and this is shown in
advantage is consistent with the well-estab- probe-mic system, we fitted both prod- Figure 8. Note that for the frequencies of
lished ear canal resonance pattern (minus ucts to the NAL-NL1 targets for a 65- 1000 Hz and above, the results for the
the pinna and concha resonances) dB input for the open condition, or as two different products were essentially
described in detail by Shaw.20 Equally close to target as possible without feed- identical—an expected finding, but one
unsurprising, up to 30 dB less output was back. The reference microphone was dis- that adds to the reliability of the measures.
achieved in the open condition for fre- abled for all testing (more on this when Note that in the 1500- to 4000-Hz
quencies under approximately 2000 Hz, we get to Issue #8). REAR measures were region, there is increased gain for the OC
which is consistent with well-established conducted and REIGs were calculated fit, with a mean peak difference of about
venting effects. using average REUGs. The closed ear- 5 dB. Figure 8 is somewhat misleading,
We also believed it was important to molds were then placed on the instru- however, as the OC advantage occurred
examine open versus closed fitting for real ments, and the testing was repeated. The for different frequencies for the different
ears. For this, we used two popular mini- hearing aids were not fitted to target for subjects, which is obscured in the mean
BTE OC products, and had half-shell the closed condition, but simply tested data—four of the five subjects had an OC
closed earmolds (medium canal depth) at the same settings as were used for the advantage of 8 to 10 dB at some frequency
custom-made for five subjects (three males OC fit. in the range of 2000 to 4000 Hz.
and two females) using the same tubing Figure 7 (panels a and b) shows the While we are conducting more of
and connector as used for the OC fit. An average results for the two different prod- these measures on a larger and more
audiogram ranging from 20 dB at 250 ucts for the two different conditions. As diverse sample, these finding are similar
and 500 Hz down to 70 dB at 4000 Hz expected, considerably more gain in the to what was observed in the coupler (see
was used for all participants. low frequencies was obtained for the Figure 6). Although we only have lim-
Using the modulated speech-shaped closed condition. The unusual pattern of ited data to support this next statement,
composite noise signal of the Fonix 7000 the REIG values for the closed condition we suspect that the OC advantage (“free
gain”) will be the greatest for
individuals with large ear
canals and large REOGs, as
they would lose the most and
gain the least when their ear
canal is closed.
ments? As reported by Mueller, industry Note the REAR that we obtained for essarily a closed earmold. The point is,
leaders tend to rate the OC product the OC fitting (Figure 9). The unusual however, that we didn’t select the OC
slightly higher than other available prod- peak in output in the 4000-Hz range was because it provided the best high-fre-
ucts for providing greater high-frequency the hearing aid going into feedback at the quency gain. Rather, it provided the best
gain, and also improved speech intelligi- time of our measurement, showing that, compromise between desired high-fre-
bility (presumably because of the high- indeed, we were at maximum gain for the quency gain, cosmetics, and the reduc-
frequency gain).7 The same impression open condition. Notice that it was cap- tion of the occlusion effect.
appears to be shared by a sizable group of tured in the REAR, but was not on long
dispensers.21 But is it true? enough to be reflected in the OC REIG Take-Home Tip #6: While it is possible
We haven’t conducted probe-mic test- calculations below. For this patient, as to achieve more high-frequency gain with
ing with this topic in mind, but OC products on individual
there are a few obvious variables patients, we believe that in many
that we can discuss. First, the cases, because of feedback issues,
answer probably will depend some- maximum high-frequency gain will
what on the maximum gain of the be obtained with a closed fitting.
instrument. As shown in Figure 8,
there is an OC advantage in the ISSUE #7: PRESCRIPTIVE
high frequencies for lower gain set- FITTINGS
tings. If the maximum gain of the Evidence-based practice reviews
instrument is also quite low and and published fitting guidelines
max gain can be reached in the consistently recommend use-vali-
open fitting, then, as shown in Fig- dated prescriptive fitting methods
ure 8, it’s probable that the open when hearing aids are selected and
fitting wins. If the instrument’s verified.22,23 Following a validated
maximum gain is high, then the prescriptive method may be even
closed condition would be more more important for OC fittings,
apt to win. Figure 8. The difference in REIG for two commercial OC because as illustrated by Bentler et
That takes us to the second fac- instruments between standard (open) and occluding (closed) al., the manufacturer’s “first fit”
earmold tips using the same tubing.
tor, the effectiveness of the prod- may be sadly inadequate.24
uct’s adaptive feedback-cancellation Some of you may be familiar
system. The more effective the with this statement from the gain
feedback-cancellation algorithm, verification section of the recent
the greater the probability that American Academy of Audiology
maximum gain in the highs will be Hearing Aid Fitting Guidelines:
obtained for the OC fitting. “…some prescriptive formulas for
Finally, the open-closed difference open fittings may be inappropri-
will depend somewhat on the ate as there is no need to correct
patient’s residual ear canal reso- for the insertion loss created by
nance (with eartip in place). including an earmold or hearing
While it’s certainly possible to aid shell in the fitting process.”23
achieve more high-frequency gain We are puzzled by this state-
with an open than a closed fit, we ment. It could be that the authors
would consider it risky to general- were referring to the proprietary
ize this to all fittings. A good exam- fitting methods of some manufac-
ple is seen in Figure 9, which shows turers. Some, however, have inter-
one of the five participants from preted the statement to mean it’s
the study we reviewed earlier, who Figure 9. REAR and REIG findings for an open versus closed inappropriate to use validated fit-
for some reason was particularly eartip for an OC product. Peak in 4000-Hz region for open ting methods, such as the NAL-
prone to having feedback prob- REAR is hearing aid going into feedback at time of testing. NL1 or the DSL 5.0 with OC
32 The Hearing Journal Open-Canal Fittings: A Special Issue November 2006 • Vol. 59 • No. 11
fittings. Indirectly, this notion is supported
by many manufacturers who have devel-
oped new gain and output fitting targets
for open instruments. What’s wrong with
the fitting methods we use for all other
styles of hearing aids?
If one considers that the verification of
the prescriptive fitting targets is based on
the amplified (or unamplified) signal deliv-
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
ear to exceed the signal delivered from the residual resonance). The bottom line is two ways to work around the problem
loudspeaker. This will cause the control that you may conclude that a hearing aid without the new software.
loop of the system to reduce the input, has less gain or output than it really has. One method, available in the speech
mapping mode, is to mute the hearing
aid after the calibration chirp occurs, but
just before the recorded speech signal
begins. The system will then use the stored
equalization. Another method is to use
the CROS test mode, and then use the
reference microphone on the opposite ear.
In using stored equalization, it’s impor-
tant that the patient not move his head,
and when the CROS method is employed,
the listener must face directly toward the
loudspeaker (to prevent any head dif-
fraction or shadow effects).
We tested using both methods and
compared our results with the standard
concurrent equalization method (for this
system the reference microphone is
located about an inch below the opening
of the canal). We obtained essentially the
same difference values for both methods,
and what is shown in Figure 10 (a and
b) is for the CROS mode method. As
expected, the difference increased as hear-
ing aid gain increased. For this reason,
we continued to increase hearing aid gain
in 2-dB steps until we found the region
where differences occurred, and then con-
tinued to increase gain until feedback was
reached.
As seen in Figure 10, when measured
REIG was in the range of 20-25 dB, there
was no difference between the standard
concurrent and contralateral CROS equal-
ization methods (panel a). Above that
level, we started to observe a difference,
and an approximate difference of 5 dB
(panel b) was noted when the REIG was
in the range of 25-30 dB (max gain before
feedback for this fitting).
Using the same hearing instrument
and patient, we then conducted similar
testing with the Frye Electronics Fonix
7000. In this case, we simply turned the
reference mic “on” and “off ” for the two
conditions. The reference microphone
was located about 1.5 inches above the
36 The Hearing Journal Open-Canal Fittings: A Special Issue November 2006 • Vol. 59 • No. 11
the ICRA input signals, and with the
MedRx Avant REM Speech system using
ICRA signals or live speech mapping. We
suggest checking with the manufacturer
for details.
If you have an older system that has
only concurrent equalization, there really
isn’t a work-around. But it shouldn’t be a
major problem. For starters, if the refer-
Downloaded from http://journals.lww.com/thehearingjournal by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4
XMi0hCywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8KKGKV0Ymy+78= on 05/26/2023
November 2006 • Vol. 59 • No. 11 Open-Canal Fittings: A Special Issue The Hearing Journal 37
measure is used. Assuming that your input
signals are chosen carefully, a good REIG
fit should also be a good REAR fit, and
vice versa.
Figure 14. Illustration of the front-and-back REAR measures for Figure 15. Illustration of the REAR for DNR-on versus DNR-off
a directional OC fitting. Front-to-back difference is displayed for an OC product with noise-reduction technology. Difference
on lower portion of chart. calculation displayed on lower portion of chart.
instrument being fitted, and the need for different hearing aid styles are carefully 13. Martin RL: Rules for successful open fittings. Hear J
2006;59(9):54-58.
improved performance in noise must all weighed for each patient, the OC prod- 14. Yanz J, Olson L: Open-ear fittings: An entry into hear-
be weighed when considering the poten- uct will be the winner in many cases. ing care for mild losses. Hear Rev 2006;13(2): 48-52.
tial for directional benefit in OC fittings. 15. Kuk F, Keenan D, Ludvigsen C: Efficacy of an open-
fitting hearing aid. Hear Rev 2005;12(2):26-32.
Another special feature typically H. Gustav Mueller, PhD, is Professor of Audiology at Vanderbilt
16. MacKenzie DL: Open-canal fittings and the hearing
employed in OC fittings is digital noise University, Senior Audiology Consultant to Siemens Hearing Instru- aid occlusion effect. Hear J 2006;59(11):50-56.
ments, and Contributing Editor to The Hearing Journal. Readers may
reduction (DNR). As with directional tech- 17. Otto W: Evaluation of an open canal hearing aid by
contact him at gus.mueller@mindspring.com. Todd A. Ricketts, experienced users Hear J 2005;58(8):26-32.
nology, the impact of DNR will be reduced PhD, is Associate Professor at the Vanderbilt University Bill Wilkerson 18. Kuk F, Keenan D: Fitting tips: How do vents affect
in OC fittings for many of the same rea- Center for Otolaryngology and Communication Sciences. hearing aid performance? Hear Rev 2006;13(2):34-
sons. If it was the low-frequency noise 42.
19. Martin RL: Why open fittings are the next big thing
reduction that provided the “wow effect” REFERENCES
Hear J 2005;58(5):50-52.
for a closed fitting, a given patient might 1. Harford E, Barry J: A rehabilitative approach to the prob-
20. Shaw E: Transformation of sound pressure from the
lem of unilateral hearing impairment: Contralateral
not notice this effect much with an OC routing of signals (CROS). J Sp Hear Dis 1965;30:121-
free field to the eardrum in the horizontal plane. J
Acoust Soc Am 1974;56:1848-1861.
product. Figure 15 shows the effects of 138.
21. Johnson EE: Segmenting dispensers: Factors in select-
DNR for “Max” versus “Off” for a typi- 2. Dodds E, Harford E: Modified earpieces and CROS for
ing open-canal fittings. Hear J 2006;59(11):58-64.
high-frequency hearing losses. J Sp Hear Res 1968;
cal OC instrument. Observe that, as with 1:204-218.
22. Mueller HG: Fitting hearing aids to adults using pre-
scriptive methods: An evidence-based review of effec-
the directivity measures, there is little or 3. Green D: Non-occluding ear molds with CROS and
tiveness. JAAA 2005;16(7):448-460.
no DNR effect in the lower frequencies. IROS hearing aids. Arch Otolaryngol 1969;89:96-
23. American Academy of Audiology: Guidelines for the
102.
Audiologic Management of Adult Hearing Impairment,
4. Dodds E, Harford E: Follow-up report on modified ear-
2006. Online at www.audiology.org/publications.
Take-Home Tip #10: Directional and pieces and CROS for high-frequency hearing losses.
24. Bentler R, Wu Y-H, Jeon J: Effectiveness of directional
DNR advantages are present in OC fit- J Sp Hear Res 1970;13:41-43.
technology in open-canal hearing instruments. Hear
5. Mueller HG, Schwartz D, Surr R: The use of the expo-
tings, but not to the extent available in a nential acoustic horn in an open mold configuration.
J 2006;59(11):40-47.
25. Dillon H: What’s new from NAL in hearing aid pre-
closed fitting. Hear Instr 1981;32(10):22-24.
scriptions? Hear J 2006;59(10):10-16.
6. Lee L, Humes L, Wilde G: Evaluating performance with
26. Verifit User Guide Version 2.8. Dorchester, Ontario:
high-frequency emphasis amplification. JAAA 1993;
CLOSING COMMENTS 4(2):91-97.
Audioscan, 2006.
27. Madsen P: Insertion gain optimization. Hear Instr
This is an exciting time in the hearing aid 7. Mueller HG: Open is in. Hear J 2006;59(11):11-14
1986;37(1):28-32.
dispensing world, and OC fittings seem 8. Erickson F, Van Tasell D: Maximum real-ear gain of in-
28. Jensen O, Olsen S, Lantz J: Real-ear measurements and
the-ear hearing aids. J Sp Hear Res 1991;34:351-359.
to be leading the charge. As with any new 9. Killion M, Monser E: CORFIG: Coupler response for
feedback cancellation—the road to openREM. GN
Otometrics Application Note 2005 (March).
product, we learn as we go, so we thought flat insertion gain. In Studebaker G, Hochberg I, eds.,
29. How to perform real-ear testing on open-fit hearing
it was time to toss out a few things that Acoustical Factors Affecting Hearing Aid Performance.
instruments in two steps. Hear Rev 2006; 13(7):51.
Baltimore: University Park Press, 1980: 147-168.
we’ve been thinking about. As we con- 10. Killion M, Revit L: CORFIG and GIFROC: Real ear
30. Olsen S, Hernvig L: Objective evaluation of maximum
stable gain level and the performance of digital feed-
tinue researching this product, we’ll no to coupler and back. In Studebaker G, Hochberg I,
back suppression in hearing aids. Presentation at the
doubt have more tips to add, and hope- eds., Acoustical Factors Affecting Hearing Aid Performance,
21st Danavox Symposium, Kolding, Denmark, 2005.
2nd ed. Boston: Allyn and Bacon, 1993: 65-85.
fully we won’t have to retract any of the 11. Mueller HG: Terminology and procedures. In Mueller
31. Lantz J, Jensen O, Olsen S, Haastrup A: Special con-
siderations regarding real-ear measurement verifica-
ten we just discussed. H, Hawkins D, Northern J, Probe Microphone Mea-
tion in open, non-occluding hearing instruments.
As we said in the beginning, fitting surements. San Diego: Singular Publishing Group: 41-
Proceedings, International Congress of Hearing Aid
66, 1992.
hearing aids is all about compromise. But 12. Sullivan R: An acoustic coupling-based classification
Acousticians, Nuremburg, Germany, 2005.
32. Ricketts TA: Directivity quantification in hearing aids:
from our test results and experience, we system for hearing aid fittings. Hear Instr 1985;
Fitting and measurement effects. Ear Hear 2000;21:45-
believe that when the pros and cons of 36(9):25-28.
58.
November 2006 • Vol. 59 • No. 11 Open-Canal Fittings: A Special Issue The Hearing Journal 39