Professional Documents
Culture Documents
Grafidy3 Procedure
Grafidy3 Procedure
Grafidy3 Procedure
Grafidy3 Procedure
(Click Here for Summary Version)
1 Personnel Requirements
2 Preliminary Requirements
2.1 Tools and Test Equipment
2.3 Safety
WARNING
POISON HAZARD!
THE PHANTOM CONTAINS CUSO4. DO NOT INGEST.
DISPOSE OF AS A HAZARDOUS WASTE ACCORDING TO STATE AND FEDERAL REGULATIONS.
3 Procedure
3.1 Hardware Set-Up
1. Unfold the 6-channel grafidy fixture and lock the arms into place (90o from center plate).
2. Connect Hypertronics cable to T/R board and use screws to secure. Place the fixture at the front (Head coil end) of
the MR table with the connection box toward the bore.
3. Plug the Hypertronics connector into the Port A. SeeIllustration 1 . (The Green Coil ID light will come on.)
Starting non-proprietary service tools: From the Common Service Desktop, select [Calibration], [Grafidy3]
from the calibration menu, and select [Click here to start this tool] to start the procedure.
2. If you receive the message shown inIllustration 3 about “Short TC Cal Values”, click [Yes].
3. Click on the [Auto Mode] tab. See Illustration 4. All three-coil axes will be selected. Click the [Scan] button to
start the calibration. (When the scan is active, you can stop it by pressing the [Abort] button and the “STOP”
button on keyboard.)
Position.
SNR.
Signal Saturation.
4. If you receive the message shown in Illustration 7 about “Short TC Cal Values”, click [Yes]. See Illustration 5.
5. The Grafidy 3 software runs multiple scans to bring each term into specification. It iterates up to the maximum
specified (default is 5 iterations). The WHOLE or ZOOM mode usually takes about 2 hours.
As each component finishes a scan/analysis iteration, the residual of the eddy current of the measurement
displays in the results table. The results are color-coded, see Table 1.
6. Review the numeric output values from the auto mode scan. Verify that the results of final iteration for each
component are below specs (green). SeeIllustration 6 for help in reading output values.
Illustration 6: Reading The Output : The specs and the Actual values (Smoothed) are shown circled.
7. If a component is not in spec after the maximum number of iterations you may need to use manual mode to
calibrate that component. See Step .
Use Manual Mode when Auto Mode fails to bring one or more of the terms into specification. Manual Mode lets you
iterate through the process yourself, and decide when to stop.
4. If a “good fit” is shown, click the [Accept] button. The calibration values will be saved to disk, and these values will
be used in the next scan. Repeat this process by going back to Step .
5. If a “bad fit” is shown, click [Cancel]. This will leave the last accepted values in the calibration files. Do not expect
any improvement on future scans when the fit doesn’t match the data even if you accept the cal parameters.
6. The last iteration for any component must be a check only, and you should not accept the fit parameters. You will
not be allowed to exit immediately after updating the cal parameters.
7. In order to know more about Grafidy, refer toGrafidy 3 Tools. or if problems are encountered, see Grafidy 3
Troubleshooting.
4 Finalization
1. Remove Fixture from the system.