Professional Documents
Culture Documents
Advmet CM 028 IR Thermometer
Advmet CM 028 IR Thermometer
Advmet CM 028 IR Thermometer
Calibration Laboratory
مؤسسة القياس المتقدم
مختبر المعايرة
Advmet-CM-028 Page 1 of 7
1.0 Responsibility
1.1 Calibration Engineer (Mechanical) or as delegated to other staff as appropriate.
2.0 Procedure
2.1 Receive IR Thermometer to be calibrated.
2.2 Take Thermal radiation source, ambient temperature thermometer and distance
measuring device as appropriate.
2.3 Check validity of the Thermal radiation source, ambient temperature thermometer and
distance measuring device.
2.4 Check the range and resolution of instrument under calibration and record the details in
the observation sheet. Also check identification sticker on the instrument / box, as
appropriate having reference of calibration service request no. / sr. no.
2.5 A device should be used to determine measuring distance. Measuring distance is the
distance from the radiation source to the infrared thermometer. The distance measuring
device will typically be a tape measure or a measuring rod.
3.2 Before the calibration, the infrared thermometer should be allowed enough time to
reach the temperature of the laboratory, typically 15 minutes. This is an especially
important consideration when bringing a thermometer in from the outside.
3.3 The temperature in the laboratory should be maintained within reasonable limits. The
ambient temperature during the calibration or the laboratory’s temperature limits should
be stated on the report of calibration.
4.0 Calibration
4.1 Before start calibration the following points should be noticed:
First, never set up a laboratory so that a heat source is facing the radiation source.
Second, make sure that the temperature of the walls facing the radiation source is that
of the laboratory.
Third, the position of the technician performing the calibration should be considered,
since he/she influences the reflected temperature.
Another concern in setup of a laboratory is ambient air flow. In no case should any
forced air be near the surface of the radiation source. This means care should be taken
not to set up the radiation source near (or beneath) any HVAC vents and doorways.
4.2 The radiation source should be set to the desired calibration temperature and should
be allowed to stabilize.
4.3 The customer should determine the calibration points used. They should be based on
the customer’s wants and needs, If the customer does not know what calibration points
he/she wants, the calibration laboratory may offer advice. If the infrared thermometer is
used over a narrow range of temperature, one calibration point may be enough. For an
infrared thermometer used over a wide range of temperatures, at least three points
should be used. These points should represent at least the minimum, maximum and
mid-range of the infrared thermometer’s usage range.
4.4 The emissivity setting of the infrared thermometer should be the same as the source’s
calibrated emissivity.
4.5 Align the infrared thermometer, first set the measuring distance; the measuring
distance is set from the flat plate surface to the front housing of the infrared
thermometer, second the infrared thermometer should be centered on the calibrator
surface. This can be done by using the laser provided with the infrared thermometer, or
by maximizing the signal by moving the infrared thermometer up and down and side to
side. When alignment is completed, the line from the infrared thermometer to the
calibrator surface should be no more than five degrees from normal (perpendicular).
4.6 The measurement should be initiated. The measurement time should be ten times
longer that the infrared thermometer’s response time, typically five seconds, the
measurement is made by holding the trigger for five seconds. The resulting final
Prepared By Approved By Signature
readout temperature should be recorded as the readout temperature for the calibration.
In spite of the complex sounding nature of this method, the procedure is actually quite
simple. For one measurement, it should take the calibration technician no longer than
15 seconds. The UUT temperature indicator must agree within the accuracy listed in
the manufacturer manual.
4.7 Repeat steps 4.5 and 4.6 for each of the remaining temperature values selected in
step 4.3.
4.9 Calibration complete. Set all POWER switches to OFF or minimum, disconnect and
secure all equipment.
4.10 Report average observation in the calibration certificate issued to the customer.
4.11 Calculate the error for each calibration point.
4.12 Record the details in raw data sheet.
4.13 Complete the calibration activities and do status stickering on instrument.
4.14 After that send instrument at outgoing area.
4.15 Put appropriate identification mark on the instruments as given by the customer / as per
Adv-met own system, if not provided by the customer.
4.16 Ensure that handling, storing, transportation and preparation of items are done as per
procedure no. QP/17.
5.0 Status
1. If agreed with the customer during review of calibration request regarding providing calibration
validity and status of instrument after calibration, follow the below mentioned steps.
2. If error found within the acceptance criteria specified by the customer, affix ‘Pass’ status
sticker after calibration.
3. If any of the readings are found outside the specified acceptance criteria by the customer,
then identify it as ‘rejected’ and send back to customer with the ‘Fail’ status sticker on
it.
6.0 Affix appropriate status sticker with following details
No. E/QMS/01-28
As found AVR
Nominal Value Error Result
As left
(Pass/Fail)
Measured Value
Temperature: Humidity:
Calibrated By:
Prepared By Approved By Signature