10.total Cholesterol
10.total Cholesterol
Note:
COMPOSITION
Reaction volumes can be proportionally changed.
REAGENT A:
For concentration > of 700 mg/dl (18.1 mmol/l) dilute
Good buffer, pH 6.7 50 mmol/l sample 1:10 with NaCl (9 g/l) solution and multiply the result
Cholesterol oxidase (CHOD) ≥ 100 U/l by 10.
Cholesterol esterase (CHE) ≥ 300 U/l
Hydroxybenzoic acid 12 mmol/l
CALCULATION OF RESULTS
4-Amminoantipirine 0.3 mmol/l
Peroxidase (POD) ≥ 500 U/l
A sample
Sodium Azide ≤ 0.095 g/l Cholesterol, mg/dl = x 200
A standard
Sodium Hydroxide ≥ 0.1 %
Conversion factor
STANDARD: 1x5 ml
Cholesterol [mg/ml] x 0.02586 = Cholesterol [mmol/l]
Cholesterol 200 mg/dl
Superfactant ≥ 5%
Verified against NIST reference material. REFERENCE VALUES
Cholesterol values according to a study on a population of
Preparation adults in absence of coronary disease(3) are the following:
Reagents are liquids ready to use.
Recommended values: < 200 mg/dl (5.17 mmol/l)
Storage and stability Upper limit: 200÷239 mg/dl (5.2-6.2 mmol/l)
Store at 2-8 °C. Do not freeze the reagents! The reagents High values: ≥ 240 mg/dl (6.21 mmol/l)
are stable up to the expiry date stated on the label if
contamination and evaporation are avoided, protected from Each laboratory should establish reference ranges for its
light. own patients population.
The above conditions are valid if the vials are opened just
only for the time to take the reagent, closed immediately ANALYTICAL PERFORMANCES
with their cap and stored at the indicated conservation Precision
temperature. Within-run and between-run coefficients of variation have
been calculated on replicates of three controls at different
A slight pink colouring of Reagent A does not interfere with total cholesterol concentration. The obtained results are
the results. Reagent A has to be limpid, eliminate turbidity, reported in the following table:
if present.
Within-run Between-run
ANCILLARY EQUIPMENT Mean
Sample SD %CV SD %CV
Automatic pipettes (mg/dl)
Photometer Serum 1 32.5 0.57 1.8 1.02 3.1
Analysis cuvettes (optical path = 1 cm)
Serum 2 93.7 1.35 1.4 3.08 3.3
Temperature controlled water bath
NaCl solution 9 g/l Serum 3 205.9 4.41 2.1 5.45 2.6
SAMPLES
Serum, heparin or EDTA plasma. Do not use fluoride, citrate
and oxalate as anticoagulant.
Separate serum from cells as soon as possible.
Stable 6 days at 2-8 °C
Pag. 1/2
GA4340 00 – KL4340 00 – BK4340 00 TOTAL CHOLESTEROL
Interferences
Bilirubin > 15 mg/dl
Hemoglobin > 500 mg/dl
Triglycerides > 1000 mg/dl
PRECAUTIONS IN USE
The Reagent A (REF GB034000) is harmful (Warning -
GHS07: H315, H319).
Skin Irrit. 2, Eye Irrit. 2
Hazard statement Codes: H315 - Causes skin irritation;
H319 - Causes serious eye irritation.
Supplemental Hazard statement Code:
EUH032 - Contact with acids liberates very toxic gas.
Precautionary statements: Prevention P280 - Wear
protective gloves/protective clothing/eye protection/face
protection. Response P305+P351+P338 - IF IN EYES: Rinse
cautiously with water for several minutes. Remove contact
lenses, if present and easy to do. Continue rinsing.
P332+P313 - If skin irritation occurs: Get medical
advice/attention. P337+P313 - If eye irritation persists: Get
medical advice/attention. P362 - Take off contaminated
clothing.
The Standard (REF GB034100) is harmful (Danger -
GHS05: H318).
Eye Dam. 1
Hazard statement Code: H318 - Causes serious eye
damage.
Precautionary statements: Prevention P280 - Wear
protective gloves/protective clothing/eye protection/face
protection. Response P305+P351+P338 - IF IN EYES: Rinse
cautiously with water for several minutes. Remove contact
lenses, if present and easy to do. Continue rinsing. P310 -
Immediately call a POISON CENTER/doctor.
Reagent A and Standard are considered harmful according
to (CE) n. 1272/2008 directive about classification,
packaging and labelling of dangerous substances. The NOTE: Changes Highlighted
reagents should be handled with caution.
The use of laboratory reagents according to good laboratory
practice is recommended.
Waste Management
Please refer to local legal requirements.
BIBLIOGRAPHY
1. Jakobs DS, Kasten Jr. BL, Demmott WR, Wolfson WL:
“Laboratory Test Handbook”, Lexi-Comp and Williams &
Wilkins Ed. 2nd Edition (1990).
2. Allain CC, Poon LS, Chan CSG, Richmond W, Fu PC:
Enzymatic determination of total serum cholesterol.
Clin Chem, 20:470 (1974).
3. Bachorik PS, Ross JW: National Cholesterol Education
Program Recommendations for measurement of low-
density lipoprotein cholesterol: executive summary.
Clin Chem 41:1414-1420 (1995).
4. Chitto G, Fabi A, Franzini C, Galletta G, Leonardi A,
Marelli M, Morelli AM: Variabilità biologica intra-
Pag. 2/2